1 Flow Injection Analysis (FIA) Flow Injection Analysis Flow Injection Analysis Chemiluminescence – Trace Metals ( e.g. dFe) Spectrophotometric – Nutrients S T H A W Time [S] Flow Injection Analysis Flow Injection Analysis Typical FI detection response output: S is sample injection; W is peak width; A is peak areas; H is peak height and T is residence time.
9
Embed
Flow Injection Analysis - California State University, Los … · · 2013-09-27Flow Injection-Capillary Electrophoresis (FI-CE)-CE: ... Flow Injection Analysis – Capillary Electrophoresis
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
1
Flow Injection Analysis (FIA)
Flow Injection AnalysisFlow Injection Analysis
Chemiluminescence – Trace Metals ( e.g. dFe)
Spectrophotometric – Nutrients
S
T
HA
W
Time [S]
Flow Injection AnalysisFlow Injection Analysis
Typical FI detection response output: S is sampleinjection; W is peak width; A is peak areas; H ispeak height and T is residence time.
2
Co
Sample injection
Flow Injection AnalysisFlow Injection Analysis
Co
Sample injection
C max
Flow Injection AnalysisFlow Injection Analysis
3
Flow Injection AnalysisFlow Injection Analysis
Flow Injection AnalysisFlow Injection Analysis
4
Flow Injection AnalysisFlow Injection Analysis
Sample
Waste
Waste
Acid
Buffer
ICP-MS
column
Flow Injection AnalysisFlow Injection Analysis
5
Flow Injection AnalysisFlow Injection Analysis
Flow Injection AnalysisFlow Injection Analysis
0 50 100 150 200 250 300
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
Absorb
ance (
A.U
.)
Time (sec)
Reproducibility
6
Samplepump
NO
Sample loop
COMM COMM
Waste
Carrierpump
Ultra-purewater
NC NC NO
NC
COMM
NC
NO NO
Reagentpump
Ammoniummolybdate
Tin(II)chloride
Reagentpump
Coil 1 (30 cm)
Sample
T-piece
T-piece
Coil 2 (30 cm)
Coil 3 (90 cm)
Detector
COMM
Flow Injection AnalysisFlow Injection Analysis
- r2 = 0.998
- LOD = 0.67 µM PO4-P
- Linear Range = 0.8 - 50 µM - Sample throughput = 40 samples h-1