Density, Refractive Index, pH and Color Multiparameter Analysis Flavors & Fragrances
Density, Refractive Index, pH and ColorMultiparameter Analysis
Flav
ors
& F
ragr
ance
s
2
Mul
tipar
amet
er A
naly
sis
Flexible product approach
A wide range of samples can be measured, from low to high viscosity products. Sam-pling and cleaning cycles and result limits can be programmed individually for each type of sample. Once analysis is complete, sample recovery allows the sample to be re-tained in the vial as a reference sample.
Various samples – multiple parameters
All analysis results are shown on the in-strument touchscreen. A computer is only required when handling a large product database or when data needs to be ex-ported to LIMS or ERP. With the LabX client/server software, the computer can be set up remotely, for example in the office of the lab manager.
The fully automatic analysis of samples saves time, increases throughput, and prevents data transcription errors. The modularity of the systems allows existing instruments to be upgraded with additional modules at any time to add further measurement parameters to the analysis.The standard solution provides reliable and comparable results, wherever the system is used.
Our standard system does not require a computer
3
Our standard system does not require a computer
www.mt.com/liquiphysics-flavors-and-fragrances www.konicaminolta.eu
Find more Information:
3
DM45 Density Meter + RX40 Refractive Index Cell + CM-5 Spectrophotometer + SC30 Sample Changer
Partnership Programbetween:
• Density / specific gravity• Refractive index• pH• Automation• Software
• Color
4
DensityThe LiquiPhysics DM density meter typically acts as the master and controls all other instruments, including the automation unit. One click on the selected product or measure-ment method starts the measurement on all connected instruments. The measurement is performed automatically and all results are displayed on the touchscreen. All measure-ment data can be analyzed and transferred to an LIMS or ERP system.
Automation The SC1 and SC30 sample delivery and cleaning units completely automate the mea-surement of all parameters, including sampling, rinsing and drying. Using compressed air allows high viscous samples to be measured. There is no evaporation or contamina-tion of the samples. SC1 is for single samples, SC30 for 30 samples.
Refractive IndexLiquiPhysics RX refractive index measuring cells can be connected to the DM density meter to measure refractive index or related values in a multi-parameter system. Benchtop LiquiPhysics RM refractometers are also available for standalone measure-ments.
A standalone instrument can be upgraded with different modules at any time to increase throughput and add further measuring parameters to the analysis. The SC1 and SC30 sample delivery and cleaning units allow the processes of sampling, draining, rinsing and drying to be automated.
Simultaneous measurement of density, refractive index, pH/conductivity, and color. The open architecture allows further instruments, such as polarimeters, to be added. The LabX software makes it easy to manage all instruments, users with different profiles, product database, methods, and measurement data, with full integration in your IT environment.
From a single analysis to a fully automated multi-parameter system
Mod
ular
ity
Density Density + single-sample automation Density + refractive index + multi-sample automation
5
pHThe SevenCompact S220 pH/ion meter can be connected to the LiquiPhysics DM density meters or RM refractometers, using a flow cell to measure pH.
ColorThe compact Konica Minolta CM-5 desktop spectrophotometer can be used either within a multi-parameter system or as standalone instrument to measure and analyze the color of liquid samples in transmission and solid samples in reflectance. When equipped with a special flow cell, the CM-5 becomes a fully integrated and controlled instrument for color measurement of liquids within a multi-parameter analysis system.
SoftwareOur LabX software conveniently manages instruments, users, methods, measurement results, system suitability tests, and adjustments. Reports can be adapted to specific corporate layout formats. The results of multiple instruments are stored in the same data-base and can be exported to LIMS or ERP systems. LabX fully supports software valida-tion and compliance with relevant regulations, including 21CFR Part 11.
The perfect modular system to meet your needs
Density + refractive index + color + pH + multi-sample automation + software
6
Save up to 70% labor timeAll four measurements are taken fully automati-cally by pressing just one button or scanning the barcode. Rather than spending a lot of time per-forming measurements and then cleaning each individual instrument, operators can schedule their valuable time for more demanding tasks.
Increase throughput by 50% All parameters are determined simultaneously in all instruments, rather than each individual analysis being performed in sequence one after the other. The SC30 sample changer can be loaded with up to 30 samples. A built-in barcode reader identifies each sample and starts the corresponding analysis automatically.
The dedicated solution for fully automatic analysis of all parameters helps increase efficiency and ensures high data quality. Operator-independent measurements ensure maximum reproducibility and save labor costs.
A single system to cover all bases
Adva
ntag
es
7
Manage raw materials and final productsA product database with all raw materials and finished products can be imported from an existing database into the LabX software. The measure-ment results are automatically compared against the target values and limits stored in the product database. Fully customized reports or quality cer-tificates can be created with LabX.
No transcription errorsThe whole workflow takes place without any hu-man interaction, thereby eliminating the risk of transcription errors. The sample is identified by barcode, the corresponding measurement method is selected accordingly, and a comparison of the results with the target values is then performed. The "pass/fail" result is stored in the database with all measurement data and can be transferred to an LIMS or ERP system (e.g. SAP).
For more informationwww.mt.com/liquiphysics-flavors-and-fragrances
Mettler-Toledo AG, AnalyticalCH-8603 Schwerzenbach, SwitzerlandTel: +41-44-806 77 11Fax: +41-44-806 73 50
Subject to technical changes© 06/2012 Mettler-Toledo AGPrinted in Switzerland Global MarCom Switzerland
General
Min. sample volume per vial 10 ml for measurement of all parameters
Max. sample viscosity 30,000 mPas corresponds to the viscosity of liquid honey
Typical analysis time per sample 5 min for analysis of four parameters (density, RI, pH, color), including complete cleaning and drying of the whole system, for samples at around 2000 mPas (e.g. edible oil).
Density DM40 DM45 DM50
Measuring range [g/cm3] 0.0000 … 3.00000 0.00000 … 3.00000 0.00000 … 3.00000
Repeatability [g/cm3] 0.00005 0.000005 0.000005
Error limit [g/cm3] 0.0001 0.00002 1) 0.00002 3)
0.00005 2) 0.00004 4)
Refractometry RX40 RX50
Measuring range [nD] 1.3200 ... 1.7000 1.32000 ... 1.58000
Repeatability [nD] 0.00005 0.00001
Error limit [nD] 0.0001 0.00002
pH S220
Measuring range [pH] –2.000 … 20.000
Resolution 0.001 / 0.01 / 0.1
Accuracy ± 0.02
Color 5) CM-5 CR-5
Wavelength range [nm] 360–740 Tri-Stimulus-Meter
Wavelength pitch [nm] 10 -
Repeatability, spectral [%] 0.1 [400–740 nm] -
Repeatability, colorimetric [dE*] 0.04 0.04
1) 0.7 … 1 g/cm3, 15 … 20°C 2) 0 … 3 g/cm3, 0 … 91°C 3) 0 … 1 g/cm3 4) 0 … 3 g/cm3 5) Provided by Konica Minolta
Technical specifications for multiparameter systems