Top Banner
30

Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Dec 18, 2015

Download

Documents

Annis Elliott
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.
Page 2: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile.

Sample Injection

(Mobile Phase 92% buffer pH 3.5, 8% acetonitrile)

Page 3: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Column Testing

The best way to evaluate a column’s performance is to use the same test that was used by the manufacturer before it was shipped. By comparing the efficiency, retention and peak shape of the peaks in the sample and very importantly the pressure under these experimental conditions, you will be able to tell if your column has changed over time.

Page 4: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Column Regeneration

Cleaning Reversed-Phase Columns

Use 10 column volumes of each:100% methanol100% acetonitrile75% acetonitrile/25% isopropanol100% methylene chloride100% hexane100% isopropanol

Aqueous based solvents such as acetic acid, trifluoroaceticacid, urea, guanidine, sodium phosphate, ordimethylformamide can aid in the removal of proteinaceousmaterial from reversed phase columns.

Page 5: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Gradient Testing

Mobile Phase A = MethanolMobile Phase B = 1000 mL methanol with 2 mL acetoneFlow = 2 mL/minDetector UV @ 280nm

The acetone in Mobile phase B is detectable by UV at 280 nm. Therefore, as the amount of Mobile Phase B is increased, the detector response increases.

GRADIENT 0 min 97.5% A and 2.5 % B10 min 2.5% A and 97.5% B12 min 2.5% A and 97.5% B12.1 min 97.5% A and 2.5 % B20 min 97.5% A and 2.5 % B

Page 6: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Evaluation of the Gradient Test.

Gradient TestingRepeat 4-5 Times and note standard deviations oftimes and detector response.

Page 7: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Gradient Testing

Page 8: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Gradient Testing

Should be linear

Not too much curvature

Not too much dwell time

Page 9: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Dwell Time

Dwell Time is the time it takes the mobile phase to movethrough the dwell volume. The dwell volume is the volume of the mobile phase between where the solvents aremixed and the column. This includes the mixing chamber,the injector, and tubing. Depending on the instrument,dwell volumes can be as low as 0.5 mL or as high as 8.0 mL.Differences in dwell volume can cause considerabledifferences in chromatography and may be one reason thatit is difficult to reproduce a separation reported in theliterature. In documenting a method, the dwell volume ofthe instrument should be reported.

Page 10: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Extra-Column VolumeExtra-column volume is volume of the system other than the column that contributes to band broadening outside the column, (in addition to the dwell volume, the detector flow cell, post column reactors, and tubing are the main sources).

The relative amount of allowable extra column broadening is proportionalto the column volume. The relative volumes of 2 columns can be calculated based on the formula for the volume of a cylinder. Length x R2.

Example: (Column 1, diameter = 4.6 mm, extra column volume = 100 L, Column 2, diameter = 2 mm, What is the equivalent extra-column volume?)

Since the radius(R) is the only thing changing in this example, relative volume can be calculated as the ratio of the two radii ( 2.32/12= 5.3) which means that the 2 mm column has a volume 5.3 times less than a 4.6 mm column therefore, the extra-column volume should be 5.3 times less (100 L/5.3 = 18 L).

Page 11: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 1. No Peaks/Very Small Peaks

Page 12: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 2. No Flow

Page 13: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 3. No Pressure or Pressure Lower than Usual

Page 14: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 4. Pressure Higher than Usual

Page 15: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 5. Variable Retention Times

Page 16: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 6. Loss of Resolution

Page 17: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 7. Split Peaks

Page 18: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 8. Peaks Tail on Initial and Later Injections

Page 19: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 9. Tailing Peaks

Page 20: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 10. Fronting Peaks

Page 21: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 11. Rounded Peaks

Page 22: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 12. Baseline Drift

Page 23: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 13. Baseline Noise (Regular)

Page 24: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 14. Baseline Noise (Irregular)

Page 25: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 15. Broad Peaks

Page 26: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 16. Change in Peak Height (one or more peaks)

Page 27: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 17. Change in Selectivity

Page 28: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 18. Negative Peaks

Page 29: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.

Problem 19. Ghost Peaks

Page 30: Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.