Fan Yang, M.D., Ph.D. Fan Yang, M.D., Ph.D. Ichiro Ikuta, M.D., Ichiro Ikuta, M.D., M.M.Sc. M.M.Sc. Head Holder Image Head Holder Image Artifact: Artifact: Shadows in the midst of Shadows in the midst of Light, Light, and How We Fixed It. and How We Fixed It. 34 Maple Street 34 Maple Street Norwalk, CT 06856 U.S.A. Norwalk, CT 06856 U.S.A. [email protected][email protected]Twitter: @radiology_ninja Twitter: @radiology_ninja
13
Embed
Fan Yang, M.D., Ph.D. Ichiro Ikuta, M.D., M.M.Sc. Head Holder Image Artifact: Shadows in the midst of Light, and How We Fixed It. 34 Maple Street Norwalk,
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
DisclosuresDisclosuresFan Yang, M.D, Ph.D.: No disclosures.
Ichiro Ikuta, M.D., M.M.Sc.: No disclosures.
PurposePurposeWe will elucidate the etiology of a head CT image artifact which may hinder image
interpretation, and how to avoid creation of this image artifact.
A noncontrast head CT was ordered in the setting of head trauma to assess for acute
traumatic injury such as intracranial hemorrhage. While this clinical scenario is
commonplace and the imaging indicated, we began to notice a recurrent image
artifact in the midst of the axial images that was occurring for multiple head CT
examinations.
A review of the physics of x-ray beam attenuation assists in the understanding of
this phenomenon.
Case ReportCase Report• A single slice in the axial stack of images had uniform low density values
compared to immediately adjacent slices. While sometimes this darkened image
occurred near the floor of the anterior fossa or middle fossa, sometimes it would
occur in the middle of the fronto-parietal parenchyma without a significantly
attenuating calvarial landmark within the image or any adjacent image.
• Multiple maintenance efforts were attempted, but the image artifact persisted.
Finally, it was determined that the artifact was most prominent at certain gantry
angles, and when the head was located near the base of the CT head holder. It
was determined that the x-ray beams were traversing through a thicker part of
the CT head holder holder material (carbon fiber) on affected axial slices,
resulting in an attenuated image.
• To avoid this image artifact, our protocols now call for the patient to be
positioned with the head as far into the head holder as possible, avoiding the
head holder all together on head CT examinations.
The middle slice demonstrates diffuse attenuation, while the left image (cephalad) and right image (caudal) lack this attenuation. The image artifact is most prominent within the midline as the CT head holder attenuates both the projected x-ray beam and x-rays intended to be detected.
Cephalad slice Attenuated slice Caudal slice
Image post-processing with reconstructed multiplanar images demonstrate the x-ray beam attenuation. A surface shaded display also demonstrates focal abrupt increased attenuation.
At this one slice, the X-ray beam traverses tangentially through a thick portion of the base of the head holder, thereby resulting in attenuation of 81%.Attenuation by
carbon fiber head holder of approximately 15%
Attenuation by carbon fiber head holder of approximately 15%
CT head holder artifact at the slice shown
CT head holder
CT head holder base
Physics DiscussionPhysics Discussion• CT head holders are designed to be radiolucent.
• X-ray attenuation proportional to the atomic number cubed (Z3). • Carbon in Carbon fiber has a Z of 6, which is much lower than the Z of
commonly used metals.
• Zcarbon = 6, Zaluminum = 13, Ziron = 26
• Given the same thickness in design, an aluminum head holder would
have 10x the attenuation compared to carbon fiber, and an iron/steel head
holder would have 81x the attenuation of carbon fiber.
• Zcarbon3 = 216, Zaluminum
3 = 2197, Ziron3 = 17576,
• X-ray attenuation is also proportional to the thickness and density of the
material used.• The relation of attenuation to thickness has an exponential relationship
that depends on the linear attenuation coefficient (μ) and thickness of the
material used.• Attenuated intensity /original beam intensity = e (μ * thickness)
• Although the carbon fiber head holder has minimal x-ray attenuation
relative to metals, the shape/design of the CT head holder results in
differing attenuation values based on the thickness of the carbon fiber that
the x-ray beam must traverse. • CT head holder thickness is approximately 5mm, which attenuates the beam
by approximately 15%. An oblique x-ray beam through estimated 5cm of
carbon fiber (at certain gantry angles) at the base of the CT head holder
attenuates the beam by up to 81%.• Linear attenuation coefficient for carbon at 100KeV, u = 0.335 cm-1.
• Attenuated intensity at 0.5cm thickness = e (-.335 * 0.5) = 0.856 (15%)• Attenuated intensity at 5.0cm thickness = e (-.335 * 0.5) = 0.187 (81%)
• This artifact therefore results in increased x-ray attenuation on a single
slice or a few consecutive slices.• Studies in radiation therapy using high energy values have found a similar
phenomenon in cases where high gantry angles are associated with x-ray
attenuation values from 2% normally to 9% when at an oblique angle1. 1. McCormack S1, Diffey J, Morgan A. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. Med Phys. 2005 1. McCormack S1, Diffey J, Morgan A. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. Med Phys. 2005 Feb;32(2):483-7.Feb;32(2):483-7.
ConclusionConclusionPatient positioning is critical to avoid image artifacts from the head holder.
Educating the CT technologists to position the patient with the head at the maximal
cephalad position within the head holder will help avoid this image artifact.
Properly positioned patient, without gantry tilt.
CT head holder artifact at the slice shown in red, due to patient positioning and gantry tilt
SuggestionsSuggestionsThe following are suggestions to minimize this artifact:
•Position the patient's head cephalad relative to the angled base of the head
holder to move the artifact outside the field of view, however this would not be
ideal if patient were to also obtain a CT of the cervical spine.
•Scan axial slices of CT brain without tilting of the gantry to minimize the x-ray
beam traversing through the base of the CT head holder.
•Reduce hyper-extension of the head if possible.
•Educate the CT technologists of these protocol procedures.
ReferencesReferences
JT Bushberg, JA Seibert, EM Leidholdt and JM Boone, The Essential JT Bushberg, JA Seibert, EM Leidholdt and JM Boone, The Essential Physics of Medical Imaging. Williams and Wilkins, Baltimore, MD 1994Physics of Medical Imaging. Williams and Wilkins, Baltimore, MD 1994
Barrett JF1, Keat N. Artifacts in CT: Recognition and Avoidance. Barrett JF1, Keat N. Artifacts in CT: Recognition and Avoidance. Radiographics. 2004 Nov-Dec;24(6):1679-91.Radiographics. 2004 Nov-Dec;24(6):1679-91.
McCormack S1, Diffey J, Morgan A. The effect of gantry angle on McCormack S1, Diffey J, Morgan A. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. Med megavoltage photon beam attenuation by a carbon fiber couch insert. Med Phys. 2005 Feb;32(2):483-7.Phys. 2005 Feb;32(2):483-7.