EZ-Blocker ® Jan. 2010. EZ-Blocker ® A bronchoscope is mandatory Bronchoscopic control for all in-, and deflations.

Dec 11, 2015

Download

Documents

Liana Higson
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZ-Blocker®EZ-Blocker®

Jan. 2010

Page 2: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZ-Blocker®EZ-Blocker®

A bronchoscope is mandatory

Bronchoscopic control for all in-, and deflations

Page 3: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZ-Blocker®EZ-Blocker®

ETT Endotracheal tube

EZB EZ-Blocker®

DLT Double Lumen Tube

MPA Multiport Adaptor

Page 4: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

PreparationsPreparations

Page 5: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

PreparationsPreparations

Sterile cloth

ETT Lubricant Syringe Marker EZ-Blocker® set

EZB Multiport Adaptor Closing caps CPAP connection piece

Page 6: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

PreparationsPreparations

Inspect for damage Remove protection shaft of the

EZB carefully by pulling the labelat the top of the shaft

Fix 2 blue closing caps on CPAP ports

Inflate cuffs to check for leakage Deflate completely vacuum Lubricate

Page 7: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

PreparationsPreparations

(*)

Page 8: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

PreparationsPreparations

Compare the lengths of EZB withETT plus MPA,place mark on the proximal shaftor remember distance in cm’s

The EZB should wedge on the carina within 8 cm from this mark (*). 4 cm towards the carina and 4 cm to wedge upon the carina

Page 9: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

IntubationIntubation

Page 10: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

IntubationIntubation ETT cuff directly behind the vocal cords

ETT tip to carina at least 4 cm

Connect MPA to ETT and start ventilating through MPA Approx. 100% O2

Page 11: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

IntubationIntubation

Distal end ETT

4 cm

Carina

EZB extensions need 4 cm to spread before wedging the carina

Page 12: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB placementEZB placement

(*)

Page 13: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB placementEZB placement First action after intubation Check by

bronchoscope for abnormalities in anatomy Confirm: depth of ETT (4 cm to carina)

Insert EZB through MPA Check under vision closing main stem bronchus

and/or RUL -> if necessary manipulate cuff in right position

Page 14: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB placementEZB placement

Fiberscopic inspection of cuffs

Inflate the cuff in thetarget bronchus until Minimum Occlusive Volume (MOV) has been reached

Deflate the cuff vacuum

Page 15: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.
Page 16: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.
Page 17: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.
Page 18: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB Procedure lung collapseEZB Procedure lung collapse

Ventilate with deflated cuffs. Approx. 100% O2

Position the patient Disconnect ventilation from MPA as soon as the

surgeon enters the thoracic cavity Lung will collapse

Page 19: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB Procedure lung collapseEZB Procedure lung collapse If needed, the surgeon manipulates the lung to

the size of his desire After successful collapse, the cuff is inflated

under vision, start One Lung Ventilation Through resorption lung collapse will improve If collapsed lung starts ventilating again, deflate

cuff and disconnect ventilation -> re-collapse! When lung has the right size -> re-inflate cuff and

restart ventilation -> Always check with scope!

Page 20: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EZB removalEZB removal

End of operation deflate cuff. Vacuum!

Ventilate carefully to remove all atelectases

Block the other lung for bi-lateral procedure, or remove the EZB

The same ETT is used postoperatively

Page 21: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EssentialsEssentials

Page 22: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

Essentials Essentials

ETT cuff must be introduced directly behind the vocal cords

Page 23: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EssentialsEssentials

First action after intubation Bronchoscopy

Confirm: Depth of ETT (4 cm from carina) and location of right upper lobe

If a cuff is not inflated, it should be deflated completely (vacuum)

Both cuffs should never be inflated at the same time †

Page 24: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EssentialsEssentials

After the EZB extends from the ETT, the EZB will wedge on the carina within ± 8 cm

After wedging the EZB on the carina, the ETT may be advanced a few cm for even more stability

No wedge? Both extensions situated in the same bronchus Check with scope, retry and/or withdraw ETT for

a few cm to provide space for spreading

Page 25: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EssentialsEssentials

ALWAYS withdraw your scope first

After that removethe EZ-Blocker

NEVER withdrawthe EZ-Blocker first

Page 26: EZ-Blocker ® Jan. 2010. EZ-Blocker ®  A bronchoscope is mandatory  Bronchoscopic control for all in-, and deflations.

EssentialsEssentials

Deflate cuffs completely (vacuum) Carefully ventilate until all atelectases are gone PEEP is needed before wound closure Check lung expansion Remove Bronchoscope first, then the EZB Never jet-ventilate a patient with a EZB in place

-> The lung could be damaged