Top Banner

of 16

Experiment 4 LIPID

Jun 02, 2018

Download

Documents

hasiqmal
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/10/2019 Experiment 4 LIPID

    1/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    1.0Abstract

    Lipids are usually defined as those components that are soluble in organic solvents but

    are insoluble in water. The lipid fraction of a fatty food therefore contains a complex mixture of

    different types of molecule. Even so, Triacylglycerolsare the major component of most foods,

    typically making up more than 95 to 99% of the total lipids present. In this experiment, it

    demonstrates how to separate and identify lipids from sample and the amount of lipid from palm

    oil and vegetable oil was determined. 1L of the sample on the TLC plate was spotted on to the

    TLC plate was placed in the container containing the solvent system in a leaning position to the

    wall of the container. The solvent was allowed to rise until it reaches the solvent front. Differentcompounds in the sample mixture travel at different rates due to the differences in their attraction

    to the stationary phase. By changing the solvent, the separation of components (measured by the

    Rf value) can be adjusted. The Rfof olive oil which is 0.69 was higher than the Rfof palm oil;

    0.31. Compound with the larger Rfvalue is less polar because it does not stick to the stationary

    phase as long as the polar compound, which would have a lower Rfvalue. Lastly, The Rf value

    can be improved by reducing the eluent polarity so that the spot will be lower down on the plate

    and therefore the value will become smaller. Factor that affects the Rf value of TLC is the

    polarity of the solvent. If the solvent is too polar all of the compounds can run right up the plate

    resulting in poor separation. Other important factors that can affect Rf values is the absorbent

    uniformity on the thin layer plate, same concentration, room temperature during the mobile

    phase and development distance of the solvent during the mobile phase.

  • 8/10/2019 Experiment 4 LIPID

    2/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    2.0Introduction

    Lipids are organic compounds that contain the same elements as carbohydrates: carbon,

    hydrogen, and oxygen. However, the hydrogen-to-oxygen ratio is always greater than 2:1

    (Moulton, 2004). All lipids are hydrophobic, which means they are not soluble in water since

    lipids are nonpolar molecules and water molecules are polar but soluble in nonpolar organic

    solvent like chloroforms. This group of molecules includes fats and oils, waxes, phospholipids,

    steroids (like cholesterol), and some other related compounds. Nearly all of the energy needed by

    the human body is provided by the oxidation of carbohydrates and lipids. Whereas carbohydrates

    provide a readily available source of energy, lipids function primarily as an energy reserve. Theamount of lipids stored as an energy reserve far exceeds the energy stored as glycogen since the

    human body is simply not capable of storing as much glycogen compared to lipids. Lipids yield

    9 kcal of energy per gram while carbohydrates and proteins yield only 4 kcal of energy per gram

    (Anonymous, n.d).

    Chromatographic separations take advantage of the fact that different substances are

    partitioned differently between two phases, a mobile phase and a stationary phase. A thin layer

    chromatography (TLC) plate is a sheet of glass, metal, or plastic which is coated with a thin layerof a solid adsorbent usually silica or alumina (Carey, 2013). A small amount of the mixture to be

    analyzed is spotted near the bottom of this plate. The TLC plate is then placed in a shallow pool

    of a solvent in a developing chamber so that only the very bottom of the plate is in the liquid.

    This liquid, or the eluent, is the mobile phase, and it slowly rises up the TLC plate by capillary

    action. As the solvent moves past the spot that was applied, equilibrium is established for each

    component of the mixture between the molecules of that component which are adsorbed on the

    solid and the molecules which are in solution. In principle, the components will differ in

    solubility and in the strength of their adsorption to the adsorbent and some components will be

    carried farther up the plate than others.

  • 8/10/2019 Experiment 4 LIPID

    3/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    When the solvent has reached the top of the plate, the plate is removed from the

    developing chamber, dried, and the separated components of the mixture are visualized. If the

    compounds are colored, visualization is straightforward. Usually the compounds are not colored,

    so a UV lamp is used to visualize the plates.

    Figure 1: The Thin Layer Chromatography

    TLC System consists of:

    a) TLC plates preferably ready made with stationary phase: These are stable and chemically inert

    plates on to whose surface a thin layer of stationary phase is applied. The stationary phase on the

    plates is of uniform thickness and consists of fine particle size.

    b) TLC chamber: This is used for the development of TLC plate. The chamber maintains

    uniform environment inside for proper development of spots. It also prevents the evaporation of

    solvents and keeps the process dust free.

  • 8/10/2019 Experiment 4 LIPID

    4/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    c) Mobile phase: This comprises of a solvent or solvent mixture recommended for the purpose.

    The mobile phase used should be particulate free and of highest purity for proper development of

    TLC spots. The solvents recommended are chemically inert with the sample, stationary phase.

    d) A filter paper moistened in the mobile phase, to be placed inside the chamber. This helps

    uniform rise in mobile phase over the length stationary phase.

    The retention factor, or Rf, is defined as the distance traveled by the compound divided by the

    distance traveled by the solvent.

    For example, if a compound travels 2.1 cm and the solvent front travels 2.8 cm, the Rfis 0.75:

  • 8/10/2019 Experiment 4 LIPID

    5/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    The Rffor a compound is a constant from one experiment to the next only if the chromatography

    conditions below are also constant:

    solvent system

    adsorbent

    thickness of the adsorbent

    amount of material spotted

    temperature

    Since these factors are difficult to keep constant from experiment to experiment, relative

    Rfvalues are generally considered. "Relative Rf" means that the values are reported relative to a

    standard, or it means that you compare the Rfvalues of compounds run on the same plate at the

    same time.

    The larger an Rfof a compound, the larger the distance it travels on the TLC plate. When

    comparing two different compounds run under identical chromatography conditions, the

    compound with the larger Rfis less polar because it interacts less strongly with the polar

    adsorbent on the TLC plate. Conversely, if you know the structures of the compounds in a

    mixture, you can predict that a compound of low polarity will have a larger Rfvalue than a polar

    compound run on the same plate.

    The Rfcan provide corroborative evidence as to the identity of a compound. If the identity of

    a compound is suspected but not yet proven, an authentic sample of the compound, or standard,

    is spotted and run on a TLC plate side by side (or on top of each other) with the compound in

    question. If two substances have the same Rfvalue, they are likely (but not necessarily) the same

    compound. If they have different Rfvalues, they are definitely different compounds.

  • 8/10/2019 Experiment 4 LIPID

    6/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    2.1Objective

    1) To demonstrate how to separate and identify lipids from sample

    2) To compare the amount of lipids (triglycerides and fatty acid) from palm oil and

    vegetable oil

    3.0Material

    1) Petroleum ether

    2) Chloroform

    3) Diethyl ether

    4)

    TLC plate

    5) Acetic acid

    6)

    Eppendorf tube7) Lipid standards (triglycerides and fatty acid)

    8) Capillary tubes

  • 8/10/2019 Experiment 4 LIPID

    7/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    4.0Procedure

    Cooking oil sample (palm oil or vegetable oil) was mixed in 1ml of petroleum ether-

    chloroform (1:1; v/v)

    The petroleum-chloroform lipid was transferred into an Eppendorf Tube

    The thin layer chromatography plate was prepared using silica gel by making lines to

    indicates the spots of samples to made and solvent

    The solvent system of petroleum ether was prepared: acetic acid (80:30:100, v/v) and

    poured the solvent system in the TLC container with a heavy lid. The depth of the solvent

    should not exceed the line where the sample were spotted

    The TLC plate was placed flat on the table and spot about 10L of the sample using

    capillary tubes or pipette tips. Dried the spot immediately with a hair dryer to avoid

    spreading and forming large diluted spots.

    The TLC place was placed in the container containing the solvent system in a leaning

    position to the wall of the container

    The solvent was allowed to rise until it reached the solvent front

    The plate was removed and allowed the solvent to evaporated at room temperature

    The plate was placed at 110oC for 25 minutes and spots was detected by spraying with

    50% iodine. The lipid the sample was identified

  • 8/10/2019 Experiment 4 LIPID

    8/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    5.0Result

    a) Palm Oil

    Table 1: The retention value of palm oil

    Group 1 2

    Distance Travelled

    for Sample

    2.5 5.5

    Distance Travelled

    for Standard

    5.7 6.0

    Distance of Solvent 18.0 17.2

    Retention value of

    sample

    0.14 0.31

    Retention Value of

    standard

    0.32 0.35

    b) Olive Oil

    Table 2: The retention value of corn oil

    Group 1 2

    Distance Travelled

    for Sample

    3.8 12

    Distance Travelled

    for Standard

    10.1 13.5

    Distance of Solvent 18.9 17.5

    Retention value of

    sample

    0.20 0.69

    Retention Value of

    standard

    0.53 0.77

  • 8/10/2019 Experiment 4 LIPID

    9/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    6.0Discussion

    Thin layer chromatography (TLC) is a method for identifying substances and testing the

    purity of compounds. TLC is a useful technique because it is relatively quick and requires small

    quantities of material. Separations in TLC involve distributing a mixture of two or more

    substances between a stationary phase and a mobile phase. The stationary phase is a thin layer of

    adsorbent (usually silica gel or alumina) coated on a plate. The mobile phase is a developing

    liquid which travels up the stationary phase, carrying the samples with it. Components of the

    samples will separate according to how strongly they adsorb on the stationary phase versus how

    readily they dissolve in the mobile phase.

    Different compounds in the sample mixture travel at different rates due to the differences

    in their attraction to the stationary phase, and because of differences in solubility in the solvent.

    By changing the solvent, or perhaps using a mixture, the separation of components (measured by

    the Rf value) can be adjusted.

    Separation of compounds is based on the competition of the solute and the mobile phase

    for binding places on the stationary phase. For instance, if normal phase silica gel is used as the

    stationary phase it can be considered polar. Given two compounds which differ in polarity, the

    more polar compound has a stronger interaction with the silica and is therefore more capable to

    dispel the mobile phase from the binding places. Consequently, the less polar compound moves

    higher up the plate (resulting in a higher Rf value). If the mobile phase is changed to a more

    polar solvent or mixture of solvents, it is more capable of dispelling solutes from the silica

    binding places and all compounds on the TLC plate will move higher up the plate. It is

    commonly said that "strong" solvents (elutants) push the analyzed compounds up the plate, while

    "weak" elutants barely move them. The order of strength/weakness depends on the coating

    (stationary phase) of the TLC plate.

  • 8/10/2019 Experiment 4 LIPID

    10/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    The chromatographic tanks are developed in a 150-mL beaker or jar containing the eluent.

    The tank need to be closed to keep the atmosphere in the beaker saturated with solvent vapor.

    When the TLC plates are removed from the eluent, the solvent is allowed to evaporate. Oilseed is

    colorless. The oilseed separating spots can be visualized by spraying the plate with a reagent that

    will react with one or more of the components of the sample which is iodine solution.

    There are three type of solvent mixture (eluent) with their respective ratio are used in this

    experiment. These three solvents are not always necessary. Sometimes one solvent will be

    appropriate. However, more than one solvent allow us to adjust the eluent polarity easily when a

    developed plate does not give an appropriate result. A suitable eluent is one that gives a

    retardation factor (Rf, also called retention factor) close to 0.5 upon developing the plate. Thus,

    the most suitable solvent mixture (eluent) with respective ratio to extract the oilseed compound is

    petroleum ether: diethyl ether: acetic acid (80:30:100, v/v).

    During the chromatography, the eluent distributes the compounds present in the sample

    over the silica. The compound sample which spotted onto the stationary phase on the plate has

    two choices either adsorbs to the solid phase or being eluted by the mobile phase. Overall, the

    eluent is competed with the sample for a space on the stationary phase) coated on the plate. The

    more polar compound will win this competition and adhere to the stationary phase, while the rest

    of the mixture will move with the mobile phase (upwards). The Rf value is the ratio of the

    distance travelled by the center of the spot to the distance simultaneously travelled by the mobile

    phase.

    In this experiment, two samples were used; Palm oil and Olive oil. Each sample was

    prepared by two groups. Firstly, measured 1 cm from the bottom of the thin layer

    chromatography plate by using a pencil, draw a line across the plate at the 1 cm mark. This is the

    origin where the sample will be spotted. Take care not to press so hard with the pencil that you

    disturb the adsorbent. Then, spotted about 1L of the sample on the TLC plate

  • 8/10/2019 Experiment 4 LIPID

    11/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    . The TLC plate was placed in the container containing the solvent system in a leaning

    position to the wall of the container. The solvent was allowed to rise until it reaches the solvent

    front. As the solvent slowly travels up the plate, the different components of the dye mixture

    travel at different rates and the mixture is separated into different coloured spots. After a

    separation is complete, individual compounds appear as spots separated vertically. Each spot has

    a retention factor (Rf) which is equal to the distance migrated over the total distance covered by

    the solvent. The Rfformula is,

    Rf=

    The Rf value can be used to identify compounds due to their uniqueness to each

    compound. By definition the Rf values are always less than unity. They are usually given to two

    decimal places. From the comparison of the standard that was used, the value of the standard

    should be the same as the standard used by other group but there are certain group having

    different standard value. This is because the plate might be overloaded, in other words the

    solution used to spot the plate is too concentrated or there are simply too many components in

    the mixture to be separated by thin layer chromatography. Moreover from the comparison of

    palm oil from group 1 with the standard, the value of retention factor is too far to the retention

    factor of standard. This is because either the adsorbent has flaked off the sides of the plate or the

    sides of the plate are touching the sides of the container (or the paper used to saturate the

    container) as the plate develops. Crooked plates make it harder to measure Rfvalues accurately.

    The value of reference standard for this experiment was 0.32 from group 1 and 0.35 from group

    2 which as to the sample is palm oil. This showed that the palm oil does fall under required

    standard which the value of retention factor is 0.14 from group 1 and 0.31 from group 2. Same

    goes to the value of olive oil which is the value of the retention factor sample does fall under

    required standard which the value from group 1 is 0.20 and from group 2 is 0.69 with the

    standard value is from group 1 is 0.53 and group 2 is 0.77.

  • 8/10/2019 Experiment 4 LIPID

    12/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    When comparing between Palm oil and Olive oil under the same conditions, the Rf of

    olive oil which is 0.69 was higher than the Rfof palm oil; 0.31. The compound with the larger Rf

    value is less polar because it does not stick to the stationary phase as long as the polar

    compound, which would have a lower Rfvalue. The compound with the larger Rf is less polar

    because it interacts less strongly with the polar adsorbent on the TLC plate. This shows that the

    polarity of olive oil is higher than palm oil. Rfvalues and reproducibility can be affected by a

    number of different factors such as layer thickness, moisture on the TLC plate, vessel saturation,

    temperature, depth of mobile phase, nature of the TLC plate, sample size, and solvent

    parameters. These effects normally cause an increase in Rfvalues. However, in the case of layer

    thickness, the Rfvalue would decrease because the mobile phase moves slower up the plate.

  • 8/10/2019 Experiment 4 LIPID

    13/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    7.0 Conclusion

    Chromatography works on the principle that different compounds will have different

    solubility and adsorption to the two phases between which they are to be partitioned. As the

    conclusion, the objectives of this experiment; to demonstrate how to separate and identify lipids

    from sample and to compare the amount of lipids (triglycerides and fatty acid) from palm oil and

    vegetable oil were achieved because thin layer chromatography is a useful technique when trying

    to identify compounds and see how they separate. It is also a useful tool to see how polar or

    nonpolar a compound is. Thin Layer Chromatography is an analytical tool widely used because

    of its simplicity, relative low cost, high sensitivity, and speed of separation.

    How to improve the Rf value? We can reduce the eluent polarity so that the spot will be

    lower down on the plate and the Rf value therefore will become smaller as well. We can either

    choose a different eluent or adjust the solvent ratio by increasing the percentage of the nonpolar

    solvent relative to the polar solvent in the eluent. There are many advantages of calculating the

    Rf value for each spot on the plate. Rf value can be used to identify a particular chemical

    substance. It is like a 'fingerprint' to identify an unknown substance based on known Rf value of

    a list of substance.

    Besides, the factor that affects the Rf value of TLC is the polarity of the solvent. If the

    solvent is too polar all of the compounds can run right up the plate resulting in poor separation.

    The Rf value is also used to established the identity of the spots of TLC plate. In fact the Rf

    value is ratio of the distance the solvent travels to the distance compound travels. Here are some

    other important factors that can affect Rf values; the absorbent uniformity on the thin layer plate,

    same concentration (spotting is too weak or strong), room temperature during the mobile phase

    and development distance of the solvent during the mobile phase

  • 8/10/2019 Experiment 4 LIPID

    14/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    8.0References

    1) Anonymous. (n.d). Thin Layer Chromatography. [Online]. [Accessed 31th October

    2014]. Available from :http://orgchem.colorado.edu/

    2) Carey, J.L. (2013). Retention Factor. [Online]. [Accessed 31th October 2014]. Available

    from :http://chemwiki.ucdavis.edu/

    3) Moulton, R. (2004). Thin Layer Chromatography. [Online]. [Accessed 31th October

    2014]. Available from :http://www.chemguide.co.uk/

    4) RSC. 2010. Palm oil. [Online]. [Accessed 31th October 2014] Available from :

    http://www.rsc.org/

    5) Theoretical Basis of Thin Layer Chromatography (TLC).(2009). [pdf]. Available from :

    file:///C:/Users/Dyan/Downloads/9783642107276-c1.pdf [Accessed 31th October 2014]

    http://orgchem.colorado.edu/http://chemwiki.ucdavis.edu/http://www.chemguide.co.uk/http://www.rsc.org/http://www.rsc.org/http://www.chemguide.co.uk/http://chemwiki.ucdavis.edu/http://orgchem.colorado.edu/
  • 8/10/2019 Experiment 4 LIPID

    15/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    9.0Appendices

    9.1

    Calculation to find Retention Factor

    Rf=

    Distance of solvent =18cm

    1. Sample Palm Oil =2.5cm

    Rf=

    = 0.14

    2. Standard = 5.7cm

    Rf=

    = 0.32

    3. Sample Olive Oil =3.8cm

    Rf=

    = 0.21

    4. Standard = 10.1cm

    Rf=

    = 0.56

  • 8/10/2019 Experiment 4 LIPID

    16/16

    EXPERIMENT 5 : ANALYSIS OF LI PID USING THIN LAYER

    CHROMATOGRAPHY FOOD

    BIOCHEMSITRY

    CFB20203

    9.2 Picture of Result Thin Layer Chromatography for each sample

    Figure1: Result Thin Layer Chromatography of Palm oil

    Figure1: Result Thin Layer Chromatography of Olive Oil