Experiment 10 Spectrum analysis of Co-60 and Na-22 Introduction In gamma ray spectroscopy the intensity of gamma rays are measured as a function of the gamma energy. The detector used must be able to measure the energy of each gamma ray and it must have good detection efficiency. A NaI scintillation detector detects gamma photons with good efficiency up to the MeV energy range and it also has adequate energy resolution to distinguish gamma rays of different energy from each others. A gamma ray spectrometer includes a detector with amplifers and a multi channel analyzer (MCA). The detector produces analog output pulses with heights proportional to the energy of the absorbed gamma rays. The MCA digitizes the pulses and classifies them according to their heights and creates a histogram with pulse height (i.e., gamma energy) as the x-axis and intensity (i.e., number of counts) as the y-axis. The histogram is called an energy spectrum. Gamma ray spectroscopy is applied for example in nuclear medicine, environmental monitoring of radioactivity and in security screening to identify radioactive isotopes. Theory The unstable isotope Co-60 decays through β - emission. Figure 10.1 shows the decay scheme of Co-60. Figure 10.1. Decay scheme of Co-60
12
Embed
Experiment 10 Spectrum analysis of Co-60 and Na-22files.uod-physics-edu.webnode.com › 200000108... · Experiment 10 Spectrum analysis of Co-60 and Na-22 Introduction In gamma ray
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Experiment 10
Spectrum analysis of Co-60 and Na-22
Introduction
In gamma ray spectroscopy the intensity of gamma rays are measured as a function of the
gamma energy. The detector used must be able to measure the energy of each gamma
ray and it must have good detection efficiency. A NaI scintillation detector detects gamma
photons with good efficiency up to the MeV energy range and it also has adequate energy
resolution to distinguish gamma rays of different energy from each others. A gamma ray
spectrometer includes a detector with amplifers and a multi channel analyzer (MCA). The
detector produces analog output pulses with heights proportional to the energy of the
absorbed gamma rays. The MCA digitizes the pulses and classifies them according to
their heights and creates a histogram with pulse height (i.e., gamma energy) as the x-axis
and intensity (i.e., number of counts) as the y-axis. The histogram is called an energy
spectrum.
Gamma ray spectroscopy is applied for example in nuclear medicine, environmental
monitoring of radioactivity and in security screening to identify radioactive isotopes.
Theory
The unstable isotope Co-60 decays through β- emission. Figure 10.1 shows the decay
scheme of Co-60.
Figure 10.1. Decay scheme of Co-60
From the decay scheme we see that in 99.88 % of the decays an electron (β- particle) of E
= 0.31 MeV is emitted followed by two gamma rays of energies Eγ1 = 1.17 MeV and Eγ2 =
1.33 MeV.
The radioactive isotope Na-22 decays through β+ emission. Figure 10.2 shows the decay
scheme of Na-22.
Figure 10.2. Decay scheme of Na-22.
The decay scheme shows that the β+ emission is in 99.94 % of the decays followed by a
gamma ray with an energy Eγ1 = 1.27 MeV. The emitted positron will immediately
annihilate with an electron producing two additional gamma rays of an energy equal to the
electron rest mass energy Eγ2 = 0.511 MeV.
When the radiation emitted by Co-60 or Na-22 is measured and acquired by a NaI
scintillation detector the energy spectrum exhibits the theoretical decay energies shown in
figures 10.1 and 10.2. However, in addition to the decay energies the spectrum will contain
several other features that are related to the detection process. These features are mainly
due to
1. Background radiation
2. Characteristic X-rays
3. Compton scattering from the scintillator material
4. Compton back scattering from materials surrounding the detector.
Figure 10.3 shows a NaI scintillation detector receiving gamma radiation. The produced
energy spectrum reflects all the different detection processes and sources of radiation.
Figure 10.3. Detection of gamma radiation with a NaI scintillation detector.
In figure 10.3 point 1 shows a gamma photon that undergoes photo electric absorption in
the scintillator. This gamma photon releases its total energy in the detector and
disappears. The energy of the gamma rays absorbed through the photo electric process
appears in the energy spectrum as peaks at the specific decay energies. These peaks are
called photo peaks.
Point 2 shows a gamma photon that is Compton scattered from the scintillator atoms.
These gamma rays leaves only part of their energy in the detector and continues their path
in another direction. The Compton scattering process is a random process which means
that the energy absorbed in the detector can take any value between 0 and Emax. The
Compton scattered gamma rays produce a continues background in the energy spectrum
called the Compton continuum. If Eγ is the energy of the gamma ray entering the detector
and interacting with the scintillator then the energy of the scattered gamma ray is given by
1
2
3
4
5
Compton back scattered photon
Background radiation
)cos1(12
0
'
cm
E
EE (10.1)
where m0c2 is the electron rest mass and θ is the scattering angle. The smallest value of
Eγ’ corresponds to θ = 180°. Thus
2
0
'
min, 21
cm
E
EE
(10.2)
Since m0c2 ≈ 0.5 MeV we may rewrite equation 10.2 as
E
EE
41
'
min,
(10.3)
where the energies are given in units of MeV. The maximum energy absorbed in the
detector corresponding to θ = 180° is
E
E
E
EEEEEC
41
4
41
2
'
min,max, (10.4)
This energy is called the Compton edge and it is always less than the photo peak energy
(gamma decay energy) Eγ. Equation 10.4 gives Emax in MeV when Eγ is given in MeV.
Point 3 shows a gamma ray that scatters from materials surrounding the detector. The
energy of these gamma rays is given by equation 10.2. They may hit the detector and
contribute to the Compton continuum.
Point 4 shows a gamma ray that scatters from materials behind the detector with θ = 180°.
This gamma ray is called a back scattered photon and it may interact with the detector
through the photo electric process releasing all its energy which equals
E
EEE scatterbackC
41
'
min,, (10.5)
EC,back scatter appears in the energy spectrum as a peak which is called the Compton back
scatter peak. Equation 10.5 gives EC in MeV when Eγ is given in MeV.
Point 5 shows a background gamma ray absorbed by the detector. Background radiation
often appears in the energy spectrum resembling Thorium or Uranium decay energies.
The energy spectrum may also contain characteristic X-ray peaks of the decay products.
Figure 10.4 shows a typical energy spectrum acquired by a NaI scintillation detector. The
different contributions 1 – 5 are indicated in the figure.
Figure 10.4. Energy spectrum of Cs-137.
Experiment
In this experiment a background energy spectrum and the energy spectra of Co-60 and
Na-22 are acquired. The spectra are energy calibrated and the different features of the
spectra are identified. The observed Compton edge and back scatter peak energies are
compared to theoretical values.
ATTENTION! The scintillation detector is a very sensitive and expensive instrument.
Handle it with extra care!
PREPARATION
1. Place the scintillation detector on the soft plastic cover and
connect the detector to the computer with the USB cable.
2. Start the MAESTRO program .
0
5000
10000
15000
20000
0 100 200 300 400 500 600 700 800
Co
un
ts
Energy/keV
1
Background radiation (thorium)
Compton continuum
Gamma photo peak
Compton edge
Compton back scatter peak
X-ray peak
EC,back scatter EC,max
3. Connect to the detector from the Detector/Buffer list.
4. Set the measurement time to 300 s: click Acquire and
choose MCB Properties. Click the Presets tab and enter the
time.
5. Check that the external amplifier gain is 1.0: click
Acquire and choose MCB Properties. Click the Amplifier
tab. Enter 1.0 and click Close.
6. Set the PMT HV to 700 V: click again Acquire and choose MCB Properties. Click
the High Voltage tab and enter the voltage. Click On. Click Close.
7. Make sure that the calibration is off: click Calculate and choose Calibration
. Click Destroy Calibration (if the button is
gray the calibration is already off).
DATA ACQUISITION
8. First a background spectrum is measured. Make sure no source is in front of the
detector. Click Go to start the data acquisition .
9. Adjust the horizontal and vertical scale to see the spectrum. Use the + and –
buttons and the arrow keys.
10. Wait until the measurement is completed (5 minutes).
11. Save the spectrum as a text file: click File and choose Save As .
Select the File type as ASCII and save the file to your flash memory or on the PC.
Name it “Background”.
12. Reset the data acquisition: click the Clear button .
13. Now place the Na-22 source in front of the
scintillator. Acquire a spectrum following steps 8 –
11. Name the spectrum “Na-22”.
14. Replace the Na-22 with the Co-60 source and
repeat step 13. Save the spectrum and name it
“Co-60”.
13
14
ENERGY CALIBRATION
15. From the Co-60 spectrum choose a region of
interest (ROI) around the peak furthest to the
right: click at the left side of the peak and
sweep over it. Then right click and choose
Mark ROI.
16. Double click the red ROI to find the channel
number of the center of the peak. This
channel number corresponds to 1.33 MeV.
17. Click Calculate and choose Calibration
.
18. In the window that appears enter the energy
value in keV: 1333 (this is the exact energy
of the Co-60 peak in units of keV).Click Ok.
19. Clear the ROI: Click ROI and then Clear All.
The Co-60 spectrum is now energy calibrated. You can now check
the energy in keV at each point of the spectrum by moving the
cursor to the desired point and reading the energy at the bottom left
of the program display.
20. Save the Co-60 and close it.
15
16
21. Open the Na-22 spectrum:
22. Destroy the calibration (see step 7).
23. Define ROIs around both photo peaks (see below).
24. Click Calculate and choose Calibration. Enter the first peak channel number and the
corresponding energy: 511 keV. Click Ok. Repeat for the second peak with the
energy 1275 keV. Click Ok. Clear the ROIs (see step 19).
25. Save the Na-22 spectrum and close it.
BACKGROUND SUBTRACTION
Before analysing a spectrum the background has to be subtracted.
26. Open the Co-60 spectrum
27. Click Calculate and choose Strip. Select ASCII, choose the Background file and
click Open.
511 keV
1275 keV
The background is now subtracted from the Co-60 spectrum
28. Save the subtracted Co-60 spectrum. Name it Co-60-b.
SPECTRUM ANALYSIS
29. Study the Co-60 spectrum. Identify the spectrum features as below, check their
energies with the cursor and write your findings in the table of results (see figure
10.4 for help).
3
5
1
2
4
6
E1 E2 E6 E4 E3
30. Close the Co-60 spectrum: .
31. Open the Na-22 spectrum: .
32. Subtract the background from the Na-22 spectrum following step 20.
33. Identify the spectrum features as below, check their energies with the cursor and
write your findings in the table of results (see figure 10.4 for help).
34. Save the subtracted Na-22 spectrum. Name it Na-22-b
35. Close the Na-22-b spectrum.
1
3
2
4
6
E1 E3
E6 E2 E4
Zoomed view
5
36. Switch off the high voltage: click Acquire and choose MCB Properties. Click the
High Voltage tab and click Off. Click Close.
37. Close the Maestro program.
PRINTING THE SPECTRA
38. Open the Co-60-b spectrum: double click on its icon . Click Options and
choose Plot . Change the vertical range to linear, deselect Auto
scale and click Range. . Enter 4000 and
click Ok. Click Ok again.
39. Print the spectrum: . Close the Co-60-b spectrum.
40. Open the Na-22-b spectrum, change the vertical scale and print it (repeat steps 38