Top Banner
Digital Slit Lamp eVO 500/500D
22

eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

May 16, 2018

Download

Documents

dangdat
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

Digital Slit Lamp

eVO 500/500D

Page 2: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  1 

 

Contents

1. Introduction 6-7

2. Safety Instruction 8

3. Installation 9-11

4. Operating Procedure 12-13

5. Lamp and Fuse Replacement 14-15

6. Care & maintenance 16-18

7. Technical Specification 19

Page 3: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  2 

 

eVO500D

 

Page 4: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  3 

 

Parts Description

1. On/Off Button.

2. Slide Plate

3. Joystick

4. Light Intensity Regulator

5. Guide Rail Cover

6. Instrument Base Locking Knob

7. Microscope arm Locking Knob

8. Illumination arm locking arm

9. Lamp access door

10. Filter Dial

11. Slit Length Dial

12. Slit width rotation knob.

13. Yellow Filter

14. Magnification knob

15. Focusing rod

16. Back Light

17. Target Light

18. Binocular Head Locking Knob

19. Eyepiece

20. SD card slot

21. Monitor

22. Inlet for Lamp Supply

23. AC inlet

24. Inlet for Back light

25. Fuse Holder

Page 5: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  4 

 

eVO500

 

Page 6: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  5 

 

Parts Description

1. On/Off Button.

2. Slide Plate

3. Joystick

4. Light Intensity Regulator

5. Guide Rail Cover

6. Instrument Base Locking Knob

7. Microscope arm Locking Knob

8. Illumination arm locking arm

9. Lamp access door

10. Filter Dial

11. Slit Length Dial

12. Slit width rotation knob.

13. Yellow Filter

14. Magnification knob

15. Focusing rod

16. Back Light

17. Target Light

18. Binocular Head Locking Knob

19. Eyepiece

Page 7: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  6 

 

1 INTRODUCTION

Congratulations on the purchase of your new eVO500/ eVO500D!

This instruction manual is designed as a training and reference manual for the operation and

maintenance of your instrument. We recommend that you read it carefully prior to use and

follow the instructions to ensure optimum performance of your new instrument.

Please retain this manual for future reference and to share with other users. Additional copies

can be obtained from your authorized LABOMED dealer or from the LABOMED service

department. Contact information is provided at the end of this manual.

LABOMED eVO500/ eVO500D Slit Lamp is an all-in-one digital ophthalmic instrument for

observation, measurement and diagnosis of Contact lens fitting, Anterior eye segment,

Cataract, Glaucoma, Retina etc.

Its optimally structured design also allows the eVO500/ eVO500D to be used in laser therapy;

this has been achieved by the possibility of operating the Slit projector in the central position

either with the left hand or right hand. The excellent transmission of the optical system

minimizes light loss during therapy. Together with the excellent image quality, this provides

ideal conditions for crisp, crystal clear images of even the finest structure.

Considering standard work procedures, the design of this system have been modified to carry

an in-builtcamera. The integrated module is optically corrected unit, to deliver the best results

in its class. An eVO500D also has a micro SD slot and an HD touch screen panel, which provides

you with an optimum, ergonomically corrected configurations and is low on fatigue while in use.

The perfection of the observation system along with an outstanding operating convenience

ideally complements each other. The outstanding benefits of this design include an effortless

operation, short working distance, comfortable arm and body posture, easy grip controls for the

sensitive setting of the Slit length, width and rotation. Using the joystick, you can control the

eVO500/ eVO500D quickly and precisely in all three coordinates and set it exactly in any

position with a quick action locks. The joystick also has a push button at the top which is very

easy to use to click photos and videos in eVO500D.

Page 8: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  7 

 

Its slit angle of incidence makes it easier to view the inner side by rotating the illumination arm

(0° - 90°), giving it a 3 dimensional rigid view of the eye interrelated with slit housing. It makes

and rotate the slit at its 0° - 180° position with fine click stop smooth movement option for the

corneal view, vertically/horizontally reducing or increasing the size of slit from 0-14mm.With the

continuous slit measurement of 1.5 to 11mm and variable fixed measurement at 0.5, 5, 8,

12mm it has enhanced red free, cobalt blue and heat absorbing filters.

The user expects the slit lamp microscope to provide optimum stereoscopic observation with

selectable magnification. The size of the field of view and the depth of field are expected to be

as large as possible, and there should be enough space in front of the microscope for

manipulation on the eye. eVO500/ eVO500D is a powerful instrument with 5 step magnification

(6.5x, 10x, 16x, 25x& 40x) in combination with a binocular straight head, 12.5x/16mm wide

field eyepieces (Diopter adjustment +/-6 ) and a smooth adjustable IPD 48.5-80mm.

An eVO500D slit lamp is designed for all digital imaging requirements. Its fatigueless

examination with integrated 5.0 mega pixels (1080p) digital module supporting, image

capturing and option of recording live on the screen panel in HD quality makes the job even

easier while performing examination on the spot. Coupled with an integrated capture button to

capture images and make videos which can often be useful to explain to the patient, findings or

the condition and fit of a contact lens during an examination. This saves time-consuming

theoretical explanations later. This modern technique is well suited for documentation,

information and educational purposes and due to its low weight. It does not impair the mobility

and ease of handling of the slit lamp.

NTRODUCTION

Page 9: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  8 

 

2 Safety Instructions

DUCTION

• This instrument is manufactured according to the safety norms as per CE regulations.

• This instrument is intended for use only as prescribed in this manual.

• Servicing and repair are allowed through authorized persons only.

• Replace burnt out fuses by fuses of the same type only.

• Do not keep SD cards in the slots for longer periods.

• Do not touch or clean the panel with wet or rough cloth.

• Use main plug and main socket with protective earth conductors only.

• Do not use force while fixing connectors. If the male and female plugs do not readily

connect, make sure that they are appropriate for one another. If any of the connectors

are damaged, please contact LABOMED representative.

• This instrument is meant to be used in dry rooms only. Care must be taken to ensure

that no product containing liquid be kept on the instrument at any time to avoid spillage

and seepage of liquids inside the instrument.

• The manufacturer will not accept any liability for damage caused by unauthorized

persons tampering with the instrument; this will also forfeit any rights to claim warranty.

• It is recommended to use the instrument only with the accessories supplied. In case you

wish to use other accessory make sure that LABOMED has certified that its use will not

harm the safety of the instrument.

NOTE : Its requested to retain in the Serial Number of the instrument for identification

by the service personnel.

Page 10: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  9 

 

3 Installation

eVO500/eVO500D Slit Lamp is shipped in a multi-layer foam container and it is divided

into major components: Microscope body, Binocular straight head, Base Platform, HD

Panel and Chin Rest Assembly, Illumination Column.

1. Take out the various parts after removing

the wrapping and save the packing.

Fig. 1

2. Mount the chin rest at the base platform

with the socket-head screws, hexagonal

wrench (provided) as shown in the Fig 2.

3. Plug the wire from the chin rest assembly

into the 5 pin connector on the back of the

power supply unit. Fig. 2

4. Mount the base platform on appropriate instrument stand arm or table base.

5. Position the Slit Lamp body gently on the base platform so that the geared rollers match with the teeth on the guide rails. When the instrument is pushed forward or pulled backwards, the wheels should roll freely and evenly over the guide-rails (Refer Figure 3). Fig. 3

Page 11: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  10 

 

6. Tighten the microscope base locking

screw. Place the guide-rail covers over the

guide-rails (Refer Fig. 4).

Base Locking Knob

Rail Cover

Fig. 4

7. Mount the microscope head on the

microscope carrying arm by sliding it into

position, making sure it is up against the

stop. Then, tighten lock-knob located on

the top of the microscope (Refer Fig.5).

Fig. 5

8. Plug the microscope illumination cable with

4 pin adaptor into the connector on the

back of the power supply as well as on the

back side of the microscope base (Refer

Fig. 6).

9. Connect the 3 pin inlet power cord to

power supply and plug into a wall outlet or

the motorized table stand arm.

Fig.6

Page 12: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  11 

 

10. Mount the touch screen panel on the

microscope arm along with the knob in the

mount. Screw the mount to the knob to

hold it tight incase of eVO500D(See Fig 7).

11. Connect the USB output at the back of the

panel and the power cord incase of

eVO500D Fig. 7

Page 13: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  12 

 

4 Operating Procedure

1. Turn the power using the ON/OFF switch located on the front of the power supply.

Brightness can be adjusted by rotating the illumination level knob situated on the

microscope base.

NOTE: The maximum position is for intermittent use only. Continuous use will

shorten lamp life.

2. Insert the projection rod in the pivot post of the instrument body to make rough PD

and focus adjustments. Position the light onto the flat surface of the projection rod

and adjust the pupillary distance and focus of the eyepieces till a comfortable

viewing is obtained. Remove the projection rod.

3. To position a patient, adjust the chinrest height by turning the control knob on the

post of the chin rest assembly until the patient’s canthus is in line with the canthus

mark on the chin rest post.

4. Microscope elevation is adjusted by rotating the joystick and observing the Slit

image through the microscope until the slit is centered on the patient’s cornea.

5. Move the Slit lamp with the joystick held firmly and slightly angled towards the

patient, until the slit appears sharply on the cornea. The accuracy of this rough

adjustment should be checked by the naked eye. The fine adjustment is performed

while observing the slit through the microscope.

6. Tilt the joystick, which is now held lightly at its upper end, until the slit appears

sharply at the depth of the eye which is to be observed. The horizontal motion of

the base can be locked by tightening the base locking screw. Lock the base

whenever the slit lamp is not in use.

7. The slit width can be adjusted by rotating the slit width control knob on either sides

of the instrument.

8. The angle between the illumination system and the microscope can be varied

between 0° to 90° to either the left or to the right. The angle is indicated on the

scale of the slit lamp arm.

Page 14: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  13 

 

9. Magnification is altered by rotating the dial on the microscope head. The

magnification of each click stop position is engraved on the dial.

10. Three filters are provided for cool and comfortable view. This can be achieved by

rotating the filter dial.

An eVO500 is a basic and eVO500D is the high end model. So, technically what

separates evo500D from evo500 is the broad spectrum of optional accessories that

offer solutions best suited for the comfort level of both-the patient and the user of

the slit lamp. Following are the key features worth mentioning which makes an

eVO500D superior over its basic version eVO500.

1. Mount the touch screen panel that allows the user to view live images. Use click

button provided on the joystick to capture images and record videos.

2. Select the options on the touch screen panel as per your requirements. These

options are “Capture”, “Zoom In/Out” & “Exit”. A Capture option once selected

will show “Capture image” and “Capture Video”, this works only if the SD card is

inserted in the memory card slot. Exit selection will make all the options disappear

on the panel.

3. Insert an SD card to capture live images and recording videos which can be used

later on to retrieve images and videos obtained for the patient with the slit lamp.

4. A PixelPro* control software supports all the above features meant for eVO500D slit

lamp when attached with USB port.

Page 15: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  14 

 

5 Lamp & Fuse Replacement

1. Open the lamp house by loosen the screw (1). See figure 8.

1

Fig. 8

2. Take out the bulb holder (1) by pulling the metal pins on the holder(see fig 9)

1

Fig. 9

3. Loosen the screw and pull the bulb out

of the house assembly.(Refer Fig. 10)

1

Fig. 10

Page 16: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  15 

 

4. Remove the bulb from the housing.

Replace the blown lamp with new one.

Insert the bulb in the housing and

secure the same in the housing.

5. Insert the holder in the bulb slowly and

tighten the lamp cover after putting it

in its place.

Replacement of Fuse

6. Open the fuse holder provided on the

back side of the power supply. See fig.

12

7. One spare fuse is being provided in

the box. Replace the dead fuse with

the live one. Insert the fuse holder

back to it’s position.

Fig. 11

Fig. 12

Fig. 13

Page 17: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  16 

 

6 Care & Maintenance

• Maintenance

1. Headrest/Chinrest Cleaning

For hygienic reasons, wipe the headrest and chin rest with an alcohol, wipe after every patient.

2. Changing Illumination Lamp

Ensure the Slit Lamp is switched off and remember that some parts may be hot. Do not

touch the glass of the new bulb.

Open the bulb access door. Swing the retaining spring towards the arm pivot and pull

the socket and bulb from the housing. Note the position of the notch in the metal collar

of the bulb. Replace the bulb with a new one and make sure the notch is in the same

position. Insert the bulb and the socket into the lamp housing and move the retaining

spring back into position. Close the bulb access door.

3. Projection Pros, Cleaning

Grasp the narrow shank of the projection prism and pull upwards. When cleaning the

prism, blow with clean dry air, then, gently wipe with a soft linen cloth.

4. Cleaning the Slide Plate

If the plate is dirty, it may cause a rough feeling when maneuvering the base of the

Slit Lamp. Clean the slide plate with soft cloth lightly dampened with a mild soap and

water solution.

CONTINUOUS POWER SUPPLY

The LABOMED eVO500/eVO500D Slit Lamp can be operated from 100 to 240V AC. This

is achieved by universal power supply. So, there is no need of adjusting the settings

according to the supply line voltage of the destination country.

Page 18: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  17 

 

NOTE: Two fuses are used in this container, one is used in the unit and the other one

is supplied as a spare part. It’s recommended to replace the blown out fuse with the

fuse of same rating and type i.e. T2 5L 250

More information on replacing the fuses can be found in the maintenance section of this

guide.

• Care instructions o Keep accessories away from dust when not in use, e.g. protect them from dust

cover.

o Remove dust with a pneumatic rubber bulb and soft brush.

o Use special optics cleaning cloths and pure alcohol for cleaning lenses and

eyepieces.

o Protect your slit lamp from moist, fumes, acids and cosmetic materials. Do not

store chemicals close to the instrument.

o Protect it from improper handling. Never install other devices sockets or unscrew

optical system and mechanical parts unless explicitly instructed to do so in this

manual.

o Protect the microscope from oil and grease. Never oil or grease the guide

surfaces or mechanical parts.

o Remove coarse contamination using a damp disposable cloth.

o Use disinfectants based on the following ingredients: aldehydes, alcohols,

quaternary ammonium compounds.

o Camera: Keep optical components clean using a lint-free cloth. Soak the cloth

using a little methanol or glass cleaner. Do not use ethanol and spirit.

o Do not clean products with optical components in a cleaning/disinfecting device

or ultrasonic bath.

o LABOMED MaxLite coatings are very resistant. If you clean as described

above, the coatings will not be damaged.

Page 19: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  18 

 

• Tropical environment/fungus:

LABOMED employs certain safety precaution in its manufacturing techniques and

materials. Other preventive measures include:

o Keep optical part clean.

o Use and store them in a clean environment only.

o Store under UV light when not in use.

o Use in continuously climate-controlled rooms only.

o Keep moisture away and cover using a plastic cover filled with silica gel.

• Occupational safety and health protection:

Observe work safety and health protection of persons responsible for processing

contaminated products.

Current regulations of hospital hygiene and prevention of infection must be

observed in the preparation, cleaning and disinfection of the products.

Instructions

• Workplace:

Remove surface contamination with a paper towel.

• Reprocessing:

Recommended: reprocess a product immediately after use.

• Cleaning:

Needed: water, detergent, spirits, microfiber cloth

o Flush the surface with running water (<40˚ C), using a little detergent if

necessary.

o Also use spirits to clean optical components.

o Dry optical components using a microfiber cloth, dry the rest of the product

using a paper towel.

• Service:

Service as and when required should be informed to LABOMED after – sales

service department.

Page 20: eVO 500 User Manual - Labo America · Digital Slit Lamp eVO 500/500D ... Parts Description 1. On/Off Button. ... An eVO500 is a basic and eVO500D is the high end model.

  19 

 

7 Technical Specification

Item Specification

eVO 500 eVO 500D

• Microscope Galilean Galilean

• Magnification 5 step rotation: 6.5x, 10x, 16x, 25x, 40x

5 step rotation: 6.5x, 10x, 16x, 25x, 40x

• Eyepieces Wide field 12.5x, focusable, anti fungus

Wide field 12.5x, focusable, anti fungus

• Interpupillary Distance

48.5 – 80mm, Diopter adjustment ± 6

48.5 – 80mm, Diopter adjustment ± 6

• Slit Length 0 – 14mm 0 – 14mm • Slit Width 0 – 14mm 0 – 14mm • Slit Apertures 0.5, 5, 8, 12mm, 1.5-11mm

continuous, slit measurement

0.5, 5, 8, 12mm, 1.5-11mm continuous, slit measurement

• Slit Rotation 0 – 180° 0 – 180° • Filters Red-Free, Cobalt Blue, Heat

absorbing Red-Free, Cobalt Blue, Heat absorbing

• Longitudinal Movement

90mm 90mm

• Lateral Movement

100mm 100mm

• Vertical Movement

30mm 30mm

• Chin Rest Range 80mm 80mm • Illumination 6V 20W Halogen 6V 20W Halogen • Digital Imaging Integrated 5 MP Imaging