European Standards Laboratory CALIBRATION-CERTIFICATE Yokogawa Europe Amersfoort, The Netherlands Certificate number Page 1 Applicant Applicant Instrument Instrument Manufacturer Yokogawa Type 760304-40-MV-Q/G6/EC/C7/C5/FL Serial number Serial number Inventory number Inventory number ID numbers used standards CSE820 CSE976 Procedure CALWT3000_2A_2A_2A_2A Version: 1 of 09-05-2012 Calibration method The Powermeter was compared to the Yokogawa Power Calibration System via a phantom technique. During the calibration the distortion of the applied voltage and current were below 0.1 %. Before calibration the device was powered on for at least 24 hours. Environmental conditions Temperature 22.4ºC ±1ºC Relative Humidity 52% ±4% Date of Calibration 10-May-12 Result The results of the calibration are shown on the next pages Traceability The results of the calibration services of Yokogawa European Standards Laboratory are traceable to primary and/or (inter)nationally accepted measurement standards. Date 10-May-12 Name E.J. Kroon Function Supervisor Standards Laboratory Yokogawa Europe BV This certificate is issued provided that Yokogawa European Standards Laboratory does not assume any liability. P.O. Box 163, 3800 AD Amersfoort (NL) Phone +31 (0)88 46410000 Reproduction of the complete certificate is permitted. website http/tmi.yokogawa.com Part of the certificate may only reproduced after written permission of Yokogawa.
33
Embed
European Standards Laboratory CALIBRATION-CERTIFICATE ...cdn6.us.yokogawa.com/content/Example_WT3000_2A_1.01.pdfEuropean Standards Laboratory CALIBRATION-CERTIFICATE Yokogawa Europe
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
European Standards Laboratory CALIBRATION-CERTIFICATEYokogawa EuropeAmersfoort, The Netherlands
Certificate numberPage 1
Applicant Applicant
Instrument InstrumentManufacturer YokogawaType 760304-40-MV-Q/G6/EC/C7/C5/FLSerial number Serial numberInventory number Inventory numberID numbers used standards CSE820 CSE976Procedure CALWT3000_2A_2A_2A_2A Version: 1 of 09-05-2012
Calibration method The Powermeter was compared to the Yokogawa Power Calibration Systemvia a phantom technique. During the calibration the distortion of the applied voltage and current were below 0.1 %. Before calibration the device was powered on for at least 24 hours.
Environmental conditions Temperature 22.4ºC ±1ºCRelative Humidity 52% ±4%
Date of Calibration 10-May-12
Result The results of the calibration are shown on the next pages
Traceability The results of the calibration services of Yokogawa European Standards Laboratoryare traceable to primary and/or (inter)nationally accepted measurement standards.
Date 10-May-12
Name E.J. KroonFunction Supervisor Standards Laboratory
Yokogawa Europe BV This certificate is issued provided that YokogawaEuropean Standards Laboratory does not assume any liability.P.O. Box 163, 3800 AD Amersfoort (NL)Phone +31 (0)88 46410000 Reproduction of the complete certificate is permitted.website http/tmi.yokogawa.com Part of the certificate may only reproduced after written
permission of Yokogawa.
Certificate number Page: 2
E L E M E N T 1
Voltage Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 15.0000 14.9850 15.0003 15.0150 V DC 0.002 Pass
15V -15.0000 -15.0150 -14.9990 -14.9850 V DC -0.006 Pass
100V 100.000 99.900 100.007 100.100 V DC 0.007 Pass
100V -100.000 -100.100 -99.996 -99.900 V DC -0.004 Pass
150V 150.000 149.850 149.995 150.150 V DC -0.003 Pass
150V -150.000 -150.150 -150.007 -149.850 V DC 0.005 Pass
1000V 1000.000 999.000 1000.119 1001.000 V DC 0.012 Pass
1000V -1000.000 -1001.000 -1000.002 -999.000 V DC 0.000 Pass
Voltage Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 100.0000 99.7000 100.0020 100.3000 V AC 0.002 Pass
150V 150.000 149.550 150.005 150.450 V AC 0.003 Pass
Voltage Calibration RMS at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 15.0000 14.9940 15.0000 15.0060 V AC 0.000 Pass
30V 30.0000 29.9880 30.0025 30.0120 V AC 0.008 Pass
60V 60.0000 59.9760 60.0009 60.0240 V AC 0.001 Pass
100V 100.000 99.960 100.003 100.040 V AC 0.003 Pass
150V 150.000 149.940 150.002 150.060 V AC 0.001 Pass
300V 300.000 299.880 300.037 300.120 V AC 0.012 Pass
600V 600.000 599.760 600.021 600.240 V AC 0.004 Pass
1000V 1000.00 999.60 1000.09 1000.40 V AC 0.009 Pass
See Next Page: 3
Certificate number Page: 3
Voltage Calibration RMS at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 15.0000 14.9880 15.0006 15.0120 V AC 0.004 Pass
30V 30.0000 29.9760 30.0046 30.0240 V AC 0.015 Pass
60V 60.0000 59.9520 60.0039 60.0480 V AC 0.007 Pass
100V 100.000 99.920 100.009 100.080 V AC 0.009 Pass
150V 150.000 149.880 150.012 150.120 V AC 0.008 Pass
300V 300.000 299.760 300.056 300.240 V AC 0.019 Pass
600V 600.000 599.520 600.045 600.480 V AC 0.007 Pass
1000V 1000.00 999.20 1000.14 1000.80 V AC 0.014 Pass
Voltage Calibration RMS at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 100.000 99.850 100.053 100.150 V AC 0.053 Pass
150V 150.000 149.775 150.093 150.225 V AC 0.062 Pass
Voltage Calibration RMS at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 100.000 99.600 100.139 100.400 V AC 0.139 Pass
150V 100.000 99.600 100.223 100.400 V AC 0.223 Pass
Voltage Calibration RMS at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 15.000 14.790 15.027 15.210 V AC 0.179 Pass
100V 100.000 98.600 99.523 101.400 V AC -0.477 Pass
150V 100.000 98.600 99.485 101.400 V AC -0.515 Pass
Voltage Calibration Linearity at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 1.000 0.969 0.998 1.031 V AC -0.200 Pass
100V 10.000 9.960 9.999 10.040 V AC -0.011 Pass
100V 50.000 49.920 50.001 50.080 V AC 0.002 Pass
100V 110.000 109.860 110.003 110.140 V AC 0.003 Pass
Current Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100mA 0.10000 0.09990 0.10000 0.10010 A DC -0.003 Pass
100mA -0.10000 -0.10010 -0.10001 -0.09990 A DC 0.005 Pass
1A 1.0000 0.9990 1.0004 1.0010 A DC 0.037 Pass
1A -1.0000 -1.0010 -0.9998 -0.9990 A DC -0.018 Pass
See Next Page: 4
Certificate number Page: 4
Current Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100mA 100.00 99.70 100.00 100.30 mA AC 0.004 Pass
1A 1.0000 0.9970 1.0001 1.0030 A AC 0.008 Pass
Current Calibration AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
5mA 5.000 4.996 4.999 5.004 mA AC -0.029 Pass
10mA 10.00 9.99 10.00 10.01 mA AC 0.015 Pass
20mA 20.00 19.98 20.00 20.02 mA AC 0.010 Pass
50mA 50.00 49.96 50.00 50.04 mA AC -0.005 Pass
100mA 100.0 99.9 100.0 100.1 mA AC -0.010 Pass
200mA 200.0 199.8 200.0 200.2 mA AC -0.001 Pass
500mA 500.0 499.8 500.0 500.2 mA AC 0.004 Pass
1A 1.0000 0.9960 1.0000 1.0040 A AC 0.004 Pass
2A 2.0000 1.9992 2.0001 2.0008 A AC 0.005 Pass
Current Calibration AC at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
10mA 10.000 9.992 10.004 10.008 mA AC 0.044 Pass
100mA 100.00 99.92 100.00 100.08 mA AC 0.004 Pass
200mA 200.00 199.84 200.02 200.16 mA AC 0.012 Pass
500mA 500.0 499.6 500.1 500.4 mA AC 0.018 Pass
1A 1.0000 0.9992 1.0001 1.0008 A AC 0.015 Pass
2A 2.0000 1.9984 2.0004 2.0016 A AC 0.018 Pass
Current Calibration AC at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
200mA 200.00 199.70 199.99 200.30 mA AC -0.003 Pass
1A 1.00000 0.99850 1.00029 1.00150 A AC 0.029 Pass
Current Calibration AC at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
200mA 200.00 199.20 200.05 200.80 mA AC 0.025 Pass
1A 1.00000 0.99600 1.00023 1.00400 A AC 0.023 Pass
Current Calibration AC at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
50mA 50.000 49.300 49.981 50.700 mA AC -0.038 Pass
100mA 100.00 98.60 99.96 101.40 mA AC -0.035 Pass
200mA 200.00 197.20 199.95 202.80 mA AC -0.026 Pass
1A 1.00000 0.98600 0.99979 1.01400 A AC -0.021 Pass
See Next Page: 5
Certificate number Page: 5
Current Calibration Linearity AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result
1A 0.01000 0.00940 0.00999 0.01060 A AC -0.090 Pass
1A 0.10000 0.09940 0.09998 0.10060 A AC -0.015 Pass
1A 0.50000 0.49940 0.50001 0.50060 A AC 0.003 Pass
1A 1.10000 1.09940 1.09996 1.10060 A AC -0.004 Pass
External Current Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.9500 49.9591 50.0500 mA DC -0.082 Pass
200mV 200.000 199.800 199.989 200.200 mA DC -0.006 Pass
1V 1.00000 0.99900 1.00003 1.00100 A DC 0.003 Pass
2V 2.00000 1.99800 2.00011 2.00200 A DC 0.006 Pass
5V 5.00000 4.99500 5.00030 5.00500 A DC 0.006 Pass
External Current Calibration 10 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.8500 49.9976 50.1500 mA AC -0.005 Pass
200mV 200.000 199.400 200.002 200.600 mA AC 0.001 Pass
1V 1.00000 0.99700 0.99999 1.00300 A AC -0.001 Pass
2V 2.00000 1.99400 1.99998 2.00600 A AC -0.001 Pass
5V 5.00000 4.98500 4.99998 5.01500 A AC 0.000 Pass
External Current Calibration 60 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.8500 49.9945 50.1500 mA AC -0.011 Pass
100mV 100.000 99.960 99.999 100.040 mA AC -0.001 Pass
200mV 200.000 199.920 200.003 200.080 mA AC 0.002 Pass
1V 1.00000 0.99960 1.00000 1.00040 A AC 0.000 Pass
10V 10.00000 9.99600 10.00005 10.00400 A AC 0.001 Pass
External Current Calibration 1 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.9700 50.0110 50.0300 mA AC 0.022 Pass
100mV 100.000 99.940 100.024 100.060 mA AC 0.024 Pass
200mV 200.000 19.988 200.048 200.120 mA AC 0.024 Pass
1V 1.00000 0.99940 1.00020 1.00060 A AC 0.020 Pass
10V 10.00000 9.99400 10.00214 10.00600 A AC 0.021 Pass
See Next Page: 6
Certificate number Page: 6
External Current Calibration 10 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.9250 49.9937 50.0750 mA AC -0.013 Pass
200mV 200.000 199.700 200.100 200.300 mA AC 0.050 Pass
1V 1.00000 0.99850 1.00057 1.00150 A AC 0.057 Pass
2V 2.00000 1.99700 2.00119 2.00300 A AC 0.059 Pass
5V 5.00000 4.99250 5.00310 5.00750 A AC 0.062 Pass
External Current Calibration 50 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.8000 50.0697 50.2000 mA AC 0.139 Pass
200mV 200.000 199.200 200.294 200.800 mA AC 0.147 Pass
1V 1.00000 0.99600 1.00136 1.00400 A AC 0.136 Pass
2V 2.00000 1.99200 2.00270 2.00800 A AC 0.135 Pass
5V 5.00000 4.98000 5.00748 5.02000 A AC 0.150 Pass
External Current Calibration 100 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
50mV 50.0000 49.3000 50.0981 50.7000 mA AC 0.196 Pass
200mV 200.000 197.200 200.354 202.800 mA AC 0.177 Pass
1V 1.00000 0.98600 1.00155 1.01400 A AC 0.155 Pass
2V 2.00000 1.97200 2.00302 2.02800 A AC 0.151 Pass
5V 5.00000 4.93000 5.00995 5.07000 A AC 0.199 Pass
Power Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 1A 15.0000 14.9770 15.0049 15.0230 W 0.033 Pass
15V 1A -15.0000 -15.0230 -15.0033 -14.9770 W 0.022 Pass
100V 1A 100.0000 99.8500 100.0499 100.1500 W 0.050 Pass
100V 1A -100.0000 -100.1500 -100.0375 -99.8500 W 0.038 Pass
150V 1A 150.000 149.770 150.043 150.230 W 0.029 Pass
150V 1A -150.000 -150.230 -150.075 -149.770 W 0.050 Pass
1000V 1A 1000.000 998.500 1000.557 1001.500 W 0.056 Pass
1000V 1A -1000.000 -1001.500 -1000.373 -998.500 W 0.037 Pass
See Next Page: 7
Certificate number Page: 7
Power Calibration 60Hz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 10mA 1.00000 0.99940 1.00013 1.00060 W 0.013 Pass
100V 20mA 2.00000 1.99880 2.00000 2.00120 W 0.000 Pass
100V 50mA 5.00000 4.99700 4.99958 5.00300 W -0.008 Pass
100V 100mA 10.0000 9.9940 9.9996 10.0060 W -0.004 Pass
100V 200mA 20.0000 19.9880 20.0007 20.0120 W 0.004 Pass
100V 500mA 50.0000 49.9700 50.0027 50.0300 W 0.005 Pass
100V 2A 200.000 199.880 200.014 200.120 W 0.007 Pass
15V 1A 15.0000 14.9910 15.0001 15.0090 W 0.001 Pass
30V 1A 30.0000 29.9820 30.0037 30.0180 W 0.012 Pass
60V 1A 60.0000 59.9640 60.0029 60.0360 W 0.005 Pass
100V 1A 100.000 99.940 100.005 100.060 W 0.005 Pass
150V 1A 150.000 149.910 150.009 150.090 W 0.006 Pass
300V 1A 300.000 299.820 300.047 300.180 W 0.016 Pass
600V 1A 600.000 599.640 599.999 600.360 W 0.000 Pass
1000V 1A 1000.00 999.40 1000.12 1000.60 W 0.012 Pass
Power Calibration 1 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 10mA 1.00000 0.99900 1.00052 1.00100 W 0.052 Pass
100V 100mA 10.0000 9.9900 10.0023 10.0100 W 0.023 Pass
100V 200mA 20.0000 19.9800 20.0048 20.0200 W 0.024 Pass
100V 500mA 50.0000 49.9500 50.0112 50.0500 W 0.022 Pass
100V 2A 200.000 199.800 200.053 200.200 W 0.026 Pass
15V 1A 15.0000 14.9850 15.0029 15.0150 W 0.019 Pass
30V 1A 30.0000 29.9700 30.0078 30.0300 W 0.026 Pass
60V 1A 60.0000 59.9400 60.0115 60.0600 W 0.019 Pass
100V 1A 100.000 99.900 100.020 100.100 W 0.020 Pass
150V 1A 150.000 149.850 150.033 150.150 W 0.022 Pass
300V 1A 300.000 299.700 300.090 300.300 W 0.030 Pass
600V 1A 600.000 599.400 600.132 600.600 W 0.022 Pass
1000V 1A 1000.00 999.00 1000.25 1001.00 W 0.025 Pass
Power Calibration 10 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 200mA 20.0000 17.9700 20.0136 22.0300 W 0.068 Pass
100V 1A 100.000 99.750 100.078 100.250 W 0.078 Pass
150V 1A 100.000 99.625 100.086 100.375 W 0.086 Pass
See Next Page: 8
Certificate number Page: 8
Power Calibration 50 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 200mA 20.0000 15.9400 19.9706 24.0600 W -0.147 Pass
100V 1A 100.000 99.500 100.167 100.500 W 0.167 Pass
150V 1A 100.000 99.625 100.181 100.375 W 0.181 Pass
Power Calibration 100 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 50mA 5.00000 3.43000 4.95791 6.57000 W -0.842 Pass
100V 100mA 10.0000 6.8600 9.9443 13.1400 W -0.557 Pass
100V 200mA 20.0000 13.7200 19.9123 26.2800 W -0.439 Pass
Power Calibration Linearity 60 Hz PF=1 at constant 1 ARange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 1A 1.0000 0.9600 0.9994 1.0400 W -0.059 Pass
100V 1A 10.0000 9.9580 9.9997 10.0420 W -0.003 Pass
100V 1A 50.0000 49.9500 50.0031 50.0500 W 0.006 Pass
100V 1A 110.0000 109.9380 110.0068 110.0620 W 0.006 Pass
Power Calibration Linearity 60 Hz PF=1 at constant 100 VRange Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 1A 1.0000 0.9600 0.9992 1.0400 W -0.082 Pass
100V 1A 10.0000 9.9580 9.9994 10.0420 W -0.006 Pass
100V 1A 50.0000 49.9500 50.0021 50.0500 W 0.004 Pass
100V 1A 110.0000 109.9380 109.9973 110.0620 W -0.002 Pass
External Current Power Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 50mV 5.00000 4.99250 4.99824 5.00750 W -0.035 Pass
100V 200mV 20.0000 19.9700 20.0028 20.0300 W 0.014 Pass
100V 1V 100.000 99.850 100.020 100.150 W 0.020 Pass
100V 2V 200.000 199.700 200.029 200.300 W 0.014 Pass
100V 5V 500.000 499.250 500.088 500.750 W 0.018 Pass
External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 50mV 5.00000 4.99700 5.00024 5.00300 W 0.005 Pass
100V 200mV 20.0000 19.9880 20.0014 20.0120 W 0.007 Pass
100V 1V 100.000 99.940 100.002 100.060 W 0.002 Pass
100V 2V 200.000 199.880 200.007 200.120 W 0.004 Pass
100V 5V 500.000 499.700 500.018 500.300 W 0.004 Pass
External Current Power Calibration 1 kHz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result
100V 50mV 5.00000 4.99500 5.00047 5.00500 W 0.009 Pass
100V 200mV 20.0000 19.9800 20.0052 20.0200 W 0.026 Pass
100V 1V 100.000 99.900 100.018 100.100 W 0.018 Pass
100V 2V 200.000 199.800 200.040 200.200 W 0.020 Pass
100V 5V 500.000 499.500 500.102 500.500 W 0.020 Pass
See Next Page: 9
Certificate number Page: 9
Power Calibration 60Hz PF=0 laggingRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 1A 0.0000 -0.0230 0.0001 0.0230 W - Pass
100V 1A 0.000 -0.015 0.002 0.150 W - Pass
150V 1A 0.000 -0.225 0.001 0.225 W - Pass
Power Calibration 60Hz PF=0 LeadingRange Target Lowlimit Measured Highlimit Unit Deviation % Result
15V 1A 0.0000 -0.0230 -0.0003 0.0230 W - Pass
100V 1A 0.000 -0.015 -0.005 0.150 W - Pass
150V 1A 0.000 -0.225 -0.002 0.225 W - Pass
Frequency CalibrationRange Target Lowlimit Measured Highlimit Unit Deviation % Result