ESTABLISHMENT OF ESTABLISHMENT OF PRESSURE STANDARD USING PRESSURE STANDARD USING PRESSURE BALANCEPRESSURE BALANCE
Samodro, R.A. and Ginanjar, G. Research Centre for Calibration, Instrumentation and Metrology
Maintenance and dissemination of national pressure measurement standards to establish pressure measurement traceability in Indonesia
KIM-LIPIKIM-LIPI
Brief of HistoryBrief of History
In order to increase the competitive excellence of national products in global market and to guarantee the validity of output in measurement research in research centres, the development of infrastructure’s technical standardization is needed; which covers metrology, standardization, testing, and quality (MSTQ).
The infrastructure mentioned before is National Metrology Institute which duty is to assure the traceability of measurement in a country. In Indonesia, the responsibility is obliged to Puslit KIM-LIPI.
The Research Center for Calibration, Instrumentation and Metrology (Puslit KIM-LIPI) is a part of the Indonesian Institute of Science (LIPI). All the facilities belong to Puslit KIM-LIPI in Serpong are housed in an area of 22.000 m2 wide. Puslit KIM-LIPI has a core competence in metrology and instrumentation, which is advanced through research and becomes the basis of the services offered to the public.
Structure Organization of KIM LIPIStructure Organization of KIM LIPI
Scientific Services Division
Scientific Services Division
Co-operation Sub div
Co-operation Sub div
Information and Documentation
Sub div
Information and Documentation
Sub div
KIM LIPIDirector
KIM LIPIDirector
Administration Division
Administration Division
Thermoelectric Calibration
Thermoelectric Calibration
Mechanical Calibration
Mechanical Calibration
CalibrationDivision
CalibrationDivision Instrumentation
Division
Instrumentation Division
Standard and Environment
Instrumentation
Standard and Environment
Instrumentation
Industrial Instrumentation
Industrial Instrumentation
ElectricalLaboratory
ElectricalLaboratory
Radio-PhotometryLaboratory
Radio-PhotometryLaboratory
TemperatureLaboratory
TemperatureLaboratory
AcousticLaboratory
AcousticLaboratory
LengthLaboratory
LengthLaboratory
Mass Sub Div
Mass Sub Div
MetrologyDivision
MetrologyDivision
Mass And Related QuantiesMass And Related Quanties
FlowLaboratory
FlowLaboratory
Viscosity//VolumeLaboratory
Viscosity//VolumeLaboratory
PressureLaboratory
PressureLaboratory
ForceLaboratory
ForceLaboratory
DensityLaboratory
DensityLaboratory
Mass Sub Div
Mass Sub Div
Mass Laboratory
Mass Laboratory
Pressure LaboratoryPressure Laboratory
Development Pressure standard1985~1999
The following pressure standard were purchase
– Pneumatic pressure balance : Bell & Howwel
• Low range(up to 0.035MPa)• Mid range(up to 0.35MPa)• High Range(up to 3.5MPa
Development Pressure standardDevelopment Pressure standard
– Hydraulic Pressure balance: Budenberg
• 4 to 12MPa• (up to 400MPa)• Burdon Pressure
gauge
Development Pressure standardDevelopment Pressure standard
1999~2001 Industrial Pressure
balance : Budenberg– 30MPa– 70MPa– 120MPa
Development Pressure standardDevelopment Pressure standard
Hydraulic Pressure Balance : Ruska 2485– 0.5 to 5MPa– 5 to 50MPa– 10 to 100MPa
Development Pressure standardDevelopment Pressure standard
2008(Present) Pneumatic
Pressure Balance DHI 7302-AMH– Low Range (Up to
0.35MPa)– High Range (Up to
7MPa)
Development Pressure standardDevelopment Pressure standard
Hydraulic Pressure Balance DHI 7601-AMH– Low range (2MPa)– Mid range (20MPa)– High
Range(500MPa
Pressure dissemination scheme using pressure Pressure dissemination scheme using pressure balancebalance
Variable volume (vv)
Loading masses
Added masses
Variable volume (vv)
valve valve
Pressure Balance StandardPressure Balance U T Pressure Measuring Device U T
Pressure balance calibration system PMD calibration system
Comparison performed between two metrology institutes as KIM-LIPI with NML-CSIRO (now NMIA) in 2000 to 2001 on Hydraulic pressure field up to 100 MPa. This comparison using Digiquartz Paroscientific pressure transducer model 760 as the transfer standard.
Bilaeral comparison result
between NML (AUSTRALIA) and KIM-LIPI (INDONESIA),
artefact: Diqiquart Paroscientific
-5.0
0.0
5.0
10.0
15.0
20.0
0 20 40 60 80 100 120
Pressure MPa
Erro
r kPa
KIM-REP pcu 513G 17 May 2001 NML pcu 7763 30 May 2001
increasing pressure
decreasing pressure
Regional ComparisonRegional Comparison
KIM-LIPI Key Comparison APMP.M.P-K7 in range 10 MPa to 100 MPa of Hydraulic Gauge pressure. This comparison was carried out in the period of November 2002 to June 2004 and piloted by NMIJ-Japan. Two pressure monitor which had been use in this comparison were Digiquartz Paroscientific model 785 and RPM3 of DHI, Inc. as the transfer standard. :
Recently KIM-LIPI taking part on APMP DEC Interlaboratory Comparison exercise up to 100 MPa Hydraulic pressure. Followed by 12 NMI and piloted by SIRIM-Malaysia, the comparison scheduled carried out from 2008 to 2009, by using transfer standard HBM model P3M with range 100 MPa its still distributed to each participating NMI.
Future PlanFuture Plan
Had been prepare for expanded scope in pneumatic area and being propose to have assessment in the end of 2008
ResearchResearch
Research being prepared is determination of local gravitation, which is one of many parameter influence the measurement of mass and related quantity including pressure. This research base on the measurement of free falling body acceleration, using laser interferometer in the vacuum chamber. Obtaining accurate value and uncertainty of local gravitation highly expected in this research, it will increase the reliability of laboratory.