A review of the Stockholm consensus EQALM symposium Berlin 1 July 2009 Quality requirements and quality goals Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009 A review of the Stockholm consensus on analytical quality specifications Per Hyltoft Petersen [email protected] NOKLUS, Bergen, Norway
41

EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Oct 01, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

A review of the Stockholm consensus

EQALM symposium

Berlin 1 July 2009

Quality requirements and quality goals

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

A review of the Stockholm consensus on analytical quality specifications

Per Hyltoft Petersen [email protected]

NOKLUS, Bergen, Norway

Page 2: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Background

Analytical quality specifications based on

Competing approaches to goal-setting in clinical biochemistry

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Analytical quality specifications based on

• clinical and biological use of measurements

• expert groups

• EQAS and PT

• state of the art

Page 3: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Quality Specifications ? History ? Imprecision

1963 David Tonks

ALE = 2CV = [1/4 reference range/mean] x 100% {biological}

1968 Roy Barnett

“Medically significant CV” - opinions of clinicians and laboratory specialists {clinical}specialists {clinical}

1970 Cotlove, Harris and Williams

Biological variation - tolerable analytic variabilityCV < 1/2 CVwithin-subject {biological}

1976 CAP Aspen Conference (1977) {biological}

1978 Wiveka Elion-Gerritzen

“Medically significant CV” - opinions of clinicians {clinical}

Page 4: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

1988 Elizabeth Gowans

Specifications for acceptable bias {biological}

1980s Analysis of clinical situations [Nordic countries] {clinical}

Quality Specifications ? History ? Bias

1991 Sverre Sandberg“Medically significant CV” - opinions of patients {clinical}“Medically significant CV” - opinions of patients {clinical}

1997 Callum G. FraserLevels of quality {biological}

1990s EGE-Lab Working GroupBiological variation and state of the art {biological}

European EQA Organisers Working Group {biological}

ISO TC 212/WG3 ISO 15196“Analytical Performance Goals Based on Medical Needs“ {clinical}

Page 5: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Analytical quality specifications for

coefficient of variation:

The Cotlove goal for imprecision

CVAnalytical < 0.5*CVWithin-Subject

Cotlove et al. Clin Chem 1970;16:1028-32

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Page 6: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Influence of analytical variation, CVAnalytical, on the

total variation, CVTotal-Monitoring, during monitoring .

CVTotal-Monitoring2 = CVWithin-Subject

2 + CVAnalytical2 =

Influence of imprecision on monitoring CV

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

CVWithin-Subject2 + (0.5* CVWithin-Subject)

2 =

1.25* CVWithin-Subject2 1.12* CVWithin-Subject

Increase in CVTotal Monitoringt < 12 %

Page 7: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Specifications for Bias:

Bias < 0.25*CV

The Gowans goal for analytical bias

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Bias < 0.25*CVPopulation

Gowans et al. Scand J Clin Lab Invest 1988;48:757-64

Page 8: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Reference interval with limits

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Page 9: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Confidence intervals for reference limits

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Decreasing CI for increasing N

90 % CI 90 % CI

Page 10: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

IFCC recommendations

Sample size, N,

IFCC recommendations

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Sample size, N, should be > 120

N = 120

90 % CI 90 % CI

Page 11: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Sample size N > 3000

The Gowans goal for analytical bias

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

N = 120

90 % CI 90 % CI

Acceptable bias

Gowans et al. Scand J Clin Lab Invest1988;48:757-64

Page 12: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

With the common reference interval produced

without errors and based on more than 3000

we can allow for bias and imprecision instead of sample size

The Gowans goal for analytical bias

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Thus the common reference interval is as good for all, as if each lab had produced it according to IFCC

and with this analytical quality we can have same reference intervals for homogeneous groups in same regions

Page 13: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

S-PotassiumAction limits for S-Potassium among clinicians in 12 hospitals

“Medically significant CV”

Each clinician gets a questionnaire with the same information about a

Elion-Gerritzen W, Thesis, 1978, Drukkerij J.H. Pasmans, S-Gravenhage

Referenceintervals

Median of low action limits and high action limits

same information about a patient and indicates for which concentration he/she will react

Page 14: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Frequency Distributions for two Distributions

35

0.0

5.0

10.0

15.0

20.0

25.0

0 10 20 30 40 50 60 70 80

Fre

quency d

istrib

utio

n

Prevalence 0.167

Gauss 1: n = 300

mean-1 = 25

s-1 = 5

Gauss 2: n = 60

mean-2 = 45

s-2 = 7

Cut-off = 35

Clinical approach - Classification

Two groups:Healthy and diseased

Prevalence 16.7 %

Cut-off: 35 U/L

0 10 20 30 40 50 60 70 80

Concentration

Prevalence, FP, FN and sum (fraction of all)

Cut-off

0.00

0.05

0.10

0.15

0.20

20 25 30 35 40 45 50

Concentration

FP

, FN

or su

m (fractio

n o

f all)

Prevalence 0.167

Gauss 1: n = 300

mean-1 = 25

s-1 = 5

Gauss 2: n = 60

mean-2 = 45

s-2 = 7

Sum FP+FN

Min sum (%) = 0.028

Min at 36.58

Cut-off = 35

Prevalence and

Optimum of sum

Prevalence 16.7 %

Red curve: FN as function of cut-off

Green curve: FP as function of cut-off

Blue curve: sum of FP and FN

Page 15: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Frequency Distributions for two Distributions

35

0.0

5.0

10.0

15.0

20.0

25.0

0 10 20 30 40 50 60 70 80

Fre

quency d

istrib

utio

n

Prevalence 0.167

Gauss 1: n = 300

mean-1 = 25

s-1 = 5

Gauss 2: n = 60

mean-2 = 45

s-2 = 7

Cut-off = 35

Clinical approach - Classification

Red curve: FN as function of bias

Green curve: FP as function of bias

Blue curve: sum of FP and FN

Effect of bias when Cut-off: 35 U/L

0 10 20 30 40 50 60 70 80

Concentration

Prevalence, FP, FN and sum (fraction of all)

Cut-off

0.00

0.05

0.10

0.15

0.20

20 25 30 35 40 45 50

Concentration

FP

, FN

or su

m (fractio

n o

f all)

Prevalence 0.167

Gauss 1: n = 300

mean-1 = 25

s-1 = 5

Gauss 2: n = 60

mean-2 = 45

s-2 = 7

Sum FP+FN

Min sum (%) = 0.028

Min at 36.58

Cut-off = 35

Prevalence and

Optimum of sum

Blue curve: sum of FP and FN

Effect of Bias, Prevalence and Sum (fraction of all)

Bias = 0

0.00

0.05

0.10

0.15

0.20

-15 -10 -5 0 5 10 15

Bias (Cut-off - Concn)

FP

, F

N o

r su

m (

fra

ctio

n o

f a

ll) Cut-off = 35

Prevalence 0.167

Gauss 1: n = 300

Gauss 2: n = 60

Sum FP+FN

Min at Bias = -1.58

Min sum (%) = 0.028

Bias = 0

Effect of Bias, Prevalence

and Optimum of sumConcentration Bias

Page 16: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Clinical approach - Classification

Effect of Bias, Imprecision and Prevalence

Bias = 0

0.05

0.10

0.15

0.20

FP

, F

N o

r su

m (

fra

ctio

n o

f a

ll) Cut-off = 35

Prevalence 0.167

Gauss 1: n = 300

Gauss 2: n = 60

Sum FP+FN

Min at Bias = -1.58

Min sum (%) = 0.028

Bias = 0

Effect of bias and imprecision when Cut-off: 35 U/L

Can be advancedby use of weight

Imprecision

Prevalence, FP, FN and sum (fraction of all)

Cut-off

0.00

0.05

0.10

0.15

0.20

20 25 30 35 40 45 50

Concentration

FP

, FN

or su

m (fractio

n o

f all)

Prevalence 0.167

Gauss 1: n = 300

mean-1 = 25

s-1 = 5

Gauss 2: n = 60

mean-2 = 45

s-2 = 7

Sum FP+FN

Min sum (%) = 0.028

Min at 36.58

Cut-off = 35

Prevalence and

Optimum of sum

Effect of Bias, Prevalence and Sum (fraction of all)

Bias = 0

0.00

0.05

0.10

0.15

0.20

-15 -10 -5 0 5 10 15

Bias (Cut-off - Concn)

FP

, F

N o

r su

m (

fra

ctio

n o

f a

ll) Cut-off = 35

Prevalence 0.167

Gauss 1: n = 300

Gauss 2: n = 60

Sum FP+FN

Min at Bias = -1.58

Min sum (%) = 0.028

Bias = 0

Effect of Bias, Prevalence

and Optimum of sum

0.00

-15 -10 -5 0 5 10 15

Bias (Cut-off - Concn)

FP

, F

N o

r su

m (

fra

ctio

n o

f a

ll)

CV% at cut-off = 14

Effect of Bias, Imprecicion

and Optimum of sum

by use of weight factors for FP and FN

BiasConcentration

Bias

Page 17: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Economical approach - Costs

Cost of follow-up ($) as function of measured Ca-concentration

Bias: + 5 %

Gallaher MP, Mobley LR, Klee GG, Schryver P. The impact of calibration error in medical decision error. NIST 2004

Bias: + 5 %

Page 18: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Editorial:

Fraser CG, Hyltoft Petersen P.

Analytical performance

characteristics should be judged

against objective quality

specifications.

Clin Chem 1999;45:321-3

Page 19: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Stockholm - Consensus Conference

Organiser: Anders Kallner

Page 20: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Kenny et al. SJCLI 1999; 59:585

Page 21: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was achieved in Stockholm?

The consensus from the conference describes a hierarchical structure of approaches to estimation of analytical approaches to estimation of analytical

quality specifications

Page 22: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

The agreement from the Stockholm conference was not

followed directly by ISO/TC 212 - ISO 15 196, because the

chairman, Larry Kapland, completely changed his mind after

the conference

So we gave up and didn’t support the ISO-group furtherSo we gave up and didn’t support the ISO-group further

However, there is now a publication, ISO 15 196, with

recommendations very close to the consensus from the

Stockholm conference, but we do not know who are the

authors – and there is no reference to the Stockholm

consensus or to the editorial which was also very close to the

consensus

Page 23: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

Page 24: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

CoaguChek S, lot 568

30

%

Control of INR-kit

Bias = – 15 %

CVA = 3.1 %

Difference plotDifference (%)30

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

-30

0

0,5 1,5 2,5 3,5 4,5 5,5

Roskilde, INR

CVMatrix = 7.9 %

CVTotal2 =

CVMatrix2 + CVA

2

CVTotal = 8.5 %

-30

0

Reference method (INR)Borrowed from Esther A. Jensen, Denmark

Page 25: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Page 26: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Ordinal scale – dichotomous test

U-hCG

0.99Medical limit = 5

Company limit = 25

2

3

-1 0 1 2 3 4 5 6ln of concentration

Fra

cti

on

of

1

Rankit-plot of Ordinal data (0 and 1), Fraction of 1

Urine-hCG

Fraction of

positive

0.90

0.50

0.10

0.01

31 10 30 100 300

Medical limit = 40

-3

-2

-1

0

1

Ran

kit

(z-v

alu

e)

Fra

cti

on

of

1

Concentration (IU/L)

Lower limit: 5 IU/L Upper limit: 40 IU/L

Hyltoft Petersen et al.

Scand J Clin Lab Invest

2008;68:298-311

Page 27: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Percentage of reported values versus concentrations

28.05.5 14.0 55.0

80.0

100.0

Kit-reference conc

Probability of measuring

Kit-value

Ordinal scale – semi-quantitative test

Urine-glucose

Kit-values

Percentage of measurements with

each Kit-value

0.0

20.0

40.0

60.0

80.0

0.1 1.0 10.0 100.0

True concentration

Pe

rce

nta

ge 0

Kit-value 5.5

Kit-value 14.0

Kit-value 28.0

Kit-value 55.0

Kit-value

5.5

14.0

28.0

55.0Bayer Visual

Urine-

Glucose

Kit-values

0 mmol/L

5.5 mmol/L

14 mmol/L

28 mmol/L

55 mmol/L

Concentration (mmol/L)Hyltoft Petersen et al.

Scand J Clin Lab Invest

2009; in press

Page 28: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no conclusion about absolute and relative quality:

Deviation from a ‘true’ value

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Deviation from a ‘true’ value

Deviation from the method mean

Page 29: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

5

10

15

De

via

tio

n fro

m M

eth

od

Me

an

Validation of Methods and

Validation of Participants

GOOD

METHOD

GOOD

RANGE

WITHIN

METHOD

-10

-5

0

De

via

tio

n fro

m M

eth

od

Me

an

-10 -5 0 5 10 15 Method Bias

GOOD

PER-

FOR-

MER

YOUR

RESULT

Page 30: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no conclusion about absolute and relative quality:

Deviation from a ‘true’ value

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Deviation from a ‘true’ value

Deviation from the method mean

There was no agreement on which level of quality should be achieved

Page 31: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

The effect of imprecision on test result variability

Desirable quality

Optimum

Minimum quality

Optimum

Fraser et al. Ann Clin Biochem 1997;34:8-12.

Page 32: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Effect of bias on reference values

Fraser et al. Ann Clin Biochem 1997;34:8-12.

Page 33: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no conclusion about absolute and relative quality:

Deviation from a ‘true’ value

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Deviation from a ‘true’ value

Deviation from the method mean

There was no agreement on which level of quality should be achieved

There was no agreement on consequences of poor quality

Page 34: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

NORIP project on common reference intervals

Creatinine

Hyltoft Petersen: Stockholm consensus EQALM Berlin 2009

Abbot Bayer Roche Cobas

Roche Hitachi-Modular Vitros

www.furst.no/norip

Page 35: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

1

1,2

1,4

Rela

tive m

ed

ian

Elecsys AutoDELFIA Arckitect Byk Sangtek Access

External control of TSH

Deutsche Gesellschaft für Klinische Chemie

Youden plot

Co

ntr

ol 2

(co

nc)

30 %

30 %

Five dominating methods

0

0,2

0,4

0,6

0,8

2/00 4/00 1/01 1/02 1/03 2/03 3/03 1/04 2/04 1/05 2/05

Rela

tive m

ed

ian

Control 1 (conc)

Co

ntr

ol 2

(co

nc)

± 30 %

±30 %

30 %

Relative median

2000 Year 2005

Petersen et al. Symposium abstracts –IFCC – WorldLab Fortaleza, Clin Chem Lab Med 2008;46 special suppl:S148

Page 36: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no conclusion about absolute and relative quality:

Deviation from a ‘true’ value

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Deviation from a ‘true’ value

Deviation from the method mean

There was no agreement on which level of quality should be achieved

There was no agreement on consequences of poor quality

There was no agreement on the relation between clinical/biological

specifications and specifications for EQAS and PT

Page 37: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

Transferrin: External Quality Assessment

Deutche Gesellschaft

für Klinische Chemie

Youden plot

Transferrin results

Target ± 17 %

Transferrin results

for two controls

Target ± 7 %

Analytical quality specifications according to the EGE-Lab

criteria: ± 7 %

Page 38: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm
Page 39: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm
Page 40: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm
Page 41: EQALM symposium · CV < 1/2 CV within-subject {biological} 1976 CAP Aspen Conference (1977) {biological} ... Bias < 0.25*CV The Gowans goal for analytical bias Hyltoft Petersen: Stockholm

What was not achieved in Stockholm?

There was no conclusion about absolute and relative quality:

Deviation from a ‘true’ value

There was no discussion about matrix-effects and

consequently no specifications for allowable matrix

There was no discussion about measurements on ordinal scale

Deviation from a ‘true’ value

Deviation from the method mean

There was no agreement on which level of quality should be achieved

There was no agreement on consequences of poor quality

There was no agreement on the relation between clinical/biological

specifications and specifications for EQAS and PT