VILSPA/Spain, 2003 Mar 23- 24 Temporal variation of the energy response EPIC Calibration Meeting, K. Dennerl Temporal variation of the energy response determined with the internal calibration source @ Al-K and Mn-K for the FF mode Charge transfer inefficiency (CTI) Energy resolution (FWHM) Peak positions
21

EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

Dec 14, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Temporal variation of the energy response

determined with the internal calibration source @ Al-K and Mn-K for the FF mode

Charge transfer inefficiency (CTI)

Energy resolution (FWHM)

Peak positions

Page 2: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Mn-K: Q0 & Q1Full Frame Mode

Page 3: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Mn-K: Q0 & Q1Full Frame Mode

Page 4: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Mn-K: Q2 & Q3Full Frame Mode

Page 5: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Mn-K: Q2 & Q3Full Frame Mode

Page 6: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @Mn-K: Quadrant 0Full Frame Mode

Page 7: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @Mn-K: Quadrant 0Full Frame Mode

Page 8: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Al-K: Q0 & Q1Full Frame Mode

Page 9: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @ Al-K: Q2 & Q3Full Frame Mode

Page 10: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @Al-K: Quadrant 0Full Frame Mode

Page 11: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

CTI @Al-K: Quadrant 0Full Frame Mode

Page 12: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Peak position @ Al-K: Q0 & Q1Full Frame Mode

Page 13: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Peak position @ Al-K: Q2 & Q3Full Frame Mode

Page 14: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Peak position @ Mn-K: Q0 & Q1Full Frame Mode

Page 15: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Peak position @ Mn-K: Q2 & Q3Full Frame Mode

Page 16: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @ Al-K: Q0 & Q1Full Frame Mode

Page 17: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @ Al-K: Q2 & Q3Full Frame Mode

Page 18: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @Al-K: Quadrant 0

Full Frame Mode

Page 19: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @ Mn-K: Q0 & Q1Full Frame Mode

Page 20: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @ Mn-K: Q2 & Q3Full Frame Mode

Page 21: EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy response determined with the internal calibration source.

VILSPA/Spain, 2003 Mar 23-24 Temporal variation of the energy responseEPIC Calibration Meeting, K. Dennerl

Energy resolution @Mn-K: Quadrant 0

Full Frame Mode