Top Banner
Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale EPIC Calibration Meeting, K. Dennerl Spatial variation of the energy scale Energy correction for pixels with an incorrect value in the offset map Investigation of additional spatial offsets
24

EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Dec 21, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Spatial variation of the energy scale

Energy correction for pixels with an incorrect value in the offset map

Investigation of additional spatial offsets

Page 2: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF O-K rev 448

Page 3: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF O-K rev 533

Page 4: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF detector coordinates

150 ks

Page 5: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF detector coordinates

150 ks

Page 6: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Spectral model for the Vela SNR

Page 7: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF O-K

Page 8: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Residual charge map for Ge-L

1.2 keV

Page 9: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Deviation from exp. charge loss, Ge-L

1.2 keV

Page 10: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Residual charge map for Ge-K

9.9 keV

Page 11: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Deviation from exp. charge loss, Ge-K

9.9 keV

Page 12: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Deviation from exp. charge loss, Ge-L

1.2 keV

Page 13: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Printed circuit board

Page 14: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Vela SNR FF O-K

Page 15: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Standard areas for checks

Page 16: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Page 17: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Page 18: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Page 19: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Page 20: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Line energy [keV]

Page 21: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Error in line energy [keV]

Page 22: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Line width [sigma; keV]

Page 23: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 332, Mn-K

Line energy [keV]

Page 24: EPIC Calibration Meeting, K. Dennerl Saclay/France, 2003 Sept 24-25 Spatial variation of the energy scale Energy correction for pixels with an incorrect.

Saclay/France, 2003 Sept 24-25Spatial variation of the energy scaleEPIC Calibration Meeting, K. Dennerl

Calclosed, rev 080, Mn-K

Line energy [keV]