Top Banner
Features • Simplified user interface • Paramagnetic O 2 sensing • Non-Invasive ventilation (Optional) • Secure access to central stations • Sophisticated power management control with battery backup • Auxiliary pressure sensor • Airway Resistance Compensation Integrated Ventilation and Monitoring • Advanced ventilation • INview Suite: SpiroDynamics and FRC INview • Plug and play modules • Patient Spirometry • Gas monitoring with metabolics and energy expenditure • Optional use of proximal Neo Flow Sensor with Neonatal ventilation Exceptional Design • Adaptable to your environment • Flexible and moveable display • Transferable module bay • Quick-release expiratory valve • Multiple trolley configurations Aerogen Aeroneb ® Pro • Built-in advanced nebulization system • Operated in-line or independently for infants through adults Engström Carestation Breathing life into critical care GE Healthcare
8

Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Apr 26, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Features• Simplifieduserinterface

• ParamagneticO2sensing

• Non-Invasiveventilation(Optional)

• Secureaccesstocentralstations

• Sophisticatedpowermanagementcontrol withbatterybackup

• Auxiliarypressuresensor

• AirwayResistanceCompensation

Integrated Ventilation and Monitoring• Advancedventilation

• INview™Suite:SpiroDynamics™andFRCINview

• Plugandplaymodules

• PatientSpirometry

• Gasmonitoringwithmetabolicsandenergyexpenditure

• OptionaluseofproximalNeoFlowSensor withNeonatalventilation

Exceptional Design• Adaptabletoyourenvironment

• Flexibleandmoveabledisplay

• Transferablemodulebay

• Quick-releaseexpiratoryvalve

• Multipletrolleyconfigurations

Aerogen Aeroneb® Pro• Built-inadvancednebulizationsystem

• Operatedin-lineorindependentlyforinfants throughadults

EngströmCarestation™

Breathinglifeintocriticalcare

GEHealthcare

Page 2: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Physical Specifications

Dimensions

Height: 44.5cm/17.5in(Displaydown) 67.5cm/26.6in(Displayup)

Heightincludingcart: 122cm/48in(Displaydown) 145cm/57.1in(Displayup)

Width: 38cm/15in

Depth: 36cm/14in

Weight: 31kg/68.3lb(notincludingcart); 76kg/167.6lb(includingcart)

Display motion

Verticaltilt: 160°inraisedposition 60°inloweredposition

Heightadjustment: 23cm/9.1in

Modes of Ventilation

VolumeControlled(VCV)

PressureControlled(PCV)

PressureControlled,VolumeGuaranteed(PCV-VG)

SynchronizedIntermittentMandatoryVentilation,

VolumeControlled(SIMV-VC)

SynchronizedIntermittentMandatoryVentilation,

PressureControlled(SIMV-PC)

SynchronizedIntermittentMandatoryVentilation,Pressure

Controlled,VolumeGuaranteed(SIMV-PCVG)(optional)

BiLevelAirwayPressureVentilation(APRVcapable)

BiLevelwithVolumeGuaranteed(BiLevel-VG)(optional)

Non-InvasiveVentilation(NIV)(optional);nCPAPavailablewithNeonataloption

ConstantPositiveAirwayPressure/PressureSupportVentilation(CPAP/PSV)

ApneabackupavailableinSIMV-VC,SIMV-PC,BiLevel,SIMV-PCVG,BiLevel-VG,CPAP/PSVandVG-PS(institutionallyselectabledefaults)

VolumeGuaranteePressureSupport(VG-PS)availablewithNeonataloption

Control and Ranges

Maximumpeakflow: 200L/min

Flow: 0.2to30L/min(0.004to0.5L/sec) 2to90L/min(0.04to1.5L/sec) 2to160L/min(0.04to2.6L/sec)

Incrementalsettings: 0.2to5L/min(incrementsof0.1L/min) 5to30L/min(incrementsof0.5L/min) 2to40L/min(incrementsof1L/min) 40to90L/min(incrementsof5L/min) 40to160L/min(incrementsof5L/min)

FiO2: 21to100%O2

Rate: 3to150breathsperminuteforVCV,PCV, PCV-VGandBiLevel(incrementsof1breath perminute)

3to120breathsperminuteforVCV,PCV, PCV-VGandBiLevel(incrementsof1breath perminute)

2to60breathsperminuteforSIMV-VC, SIMV-PC,SIMV-PCVG,BiLevel-VG (incrementsof1breathperminute)

1to60breathsperminuteforSIMV-VC,SIMV-PC,SIMV-PCVGandBiLevel-VG(incrementsof1breathperminute)

Key: Available only when Adult patient type is selected

Available only when Pediatric patient type is selected

Available only when Neonatal patient type is selected

Note: Neonatal software is an optional feature. If not specified with the icon, features listed in this specifications sheet apply to Adult/Pediatric units and patient population selections.

Note: Ranges and Settings without an icon pertain to both Adult and Pediatric patient types.

Page 3: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Inspiratorytime:0.1to10sec 0.1to1sec(incrementsof0.01) 1to4sec(incrementsof0.1) 4to10sec(incrementsof0.25) 0.25to15sec 0.25to1sec(incrementsof0.05) 1to4sec(incrementsof0.10) 4to15sec(incrementsof0.25)

Thigh: 0.1to10sec 0.1to1sec(incrementsof0.01) 1to4sec(incrementsof0.1) 4to10sec(incrementsof0.25) 0.25to15sec 0.25to1sec(incrementsof0.05) 1to4sec(incrementsof0.1) 4to15sec(incrementsof0.25)

Tlow: 0.25to18sec 0.25to1sec(incrementsof0.01) 1to4sec(incrementsof0.1) 4to18sec(incrementsof0.25) 0.25to18sec 0.25to1sec(incrementsof0.05) 1to4sec(incrementsof0.1) 4to18sec(incrementsof0.25)

Tsupp: 0.1to0.8sec(incrementsof0.01) 0.25to4secforNIV 0.25to1sec(incrementsof0.05) 1to4sec(incrementsof0.1)

Expiratorytime: 0.25to59.75sec 0.25to29.9secInvasiveventmodes 0.5to59.75secforNIV

Risetime: 0to500msofinspiratoryperiodforeitherfloworpressuredependingonthemodeselected.ActiveinVCV,PCV,PCV-VG,SIMV-VC,SIMV-PC,SIMV-PCVG,BiLevel-VG,NIVandBiLevel(incrementsof50ms)

PSVrisetime: 0to500msofinspiratoryperiodforpressuresupportedbreathsonly.ActiveinSIMV-VC,SIMV-PC,SIMV-PCVG,BiLevel,BiLevel-VG,CPAP/PSVandVG-PS(incrementsof50ms)

Triggerwindow:0to80%ofexpirationtime(incrementsof5%)

Flowtrigger: 0.2to1L/min(incrementsof0.05L/min) 1to3L/min(incrementsof0.1L/min) 3to9L/min(incrementsof0.5L/min)

Pressuretrigger:–10to–3cmH2O(incrementsof0.5cmH2O) –3to–0.25cmH2O(incrementsof0.25cmH2O)

Biasflowrate: 2to15L/min(incrementsof0.5L/min) fornCPAP 2to10L/min(incrementsof0.5L/min) 8to20L/minforNIV(incrementsof0.5L/min)

Control and Ranges (continued)

Minimumrate: 2to60breathsperminuteforVG-PS(incrementsof1breathperminute) 0to60breathsperminuteforCPAP/PSVand0to40breathsperminuteforNIV(incrementsof1breathperminute)

Inspiratory/expiratoryratio: 1:199to40:1inBiLevel 1:9to4:1(ventilatorsetting) 1:79to60:1inBiLevel

Tidalvolumerange: 2to350mL 20to2000mL

Incrementalsettings: 2to50mL(incrementsof0.5mL) 50to100mL(incrementsof1mL) 100to350mL(incrementsof5mL) ForVCV,PCV-VG,SIMV-VC,SIMV-PCVG, VG-PSandBiLevel-VG 20to50ml(incrementsof0.5ml) 50to100ml(incrementsof1ml) 100to300ml(incrementsof5ml) 300to1000mL(incrementsof25mL) 1000to2000mL(incrementsof50mL) ForVCV,PCV-VG,SIMV-VC,SIMV-PCVG andBiLevel-VG

Patientweight: 0.25to1kg(incrementsof0.01kg) 1to7kg(incrementsof0.1kg) 0.5to2lb(incrementsof0.02lb) 2to15lb(incrementsof0.2lb) 5to15kg(incrementsof0.5kg) 15to100kg(incrementsof1kg) 100to200kg(incrementsof2kg) 10to34lb(incrementsof1lb) 34to220lb(incrementsof2lb) 220to440lb(incrementsof5lb)

Inspiratorypressure(Pinsp)range: 1to98cmH2O(incrementsof1cmH2O)

Phigh: 1to98cmH2O(incrementsof1cmH2O)

Plow: Off,1to50cmH2O(incrementsof1cmH2O)

Pressurelimit(Plimit)range: 7to100cmH2OforVCVandSIMV-VC

(incrementsof1cmH2O)

Max.inspiratorypressure(Pmax)limit: 7to100cmH2O(incrementsof1cmH2O) 9-100cmH2O(incrementsof1cmH2O) inNIVandnCPAP

PEEP: Off,1to50cmH2O(incrementsof1cmH2O) 2-15cmH2O(incrementsof1cmH2O) innCPAP 2-20cmH2O(incrementsof1cmH2O)inNIV

Page 4: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

EtO2: Low: Off,10to99% High: 11to100%,Off

EtCO2: Low: Off,0.1to14.9%orOff, 0to114.5mmHg

High: 0.2to15%,off or0.5to115mmHg,Off

Ventilationsoftlimitindicators: Whenadjustingselectedventilator

parameters,colorindicatorsshowwhenparametersareapproachingtheirsettinglimits.

Parameterswithsoftlimits: Pmax,PEEP,Pinsp,Psupp,Tinsp,RR,I:E,Phigh,Plow,

ThighandTlow

Alarm System

Escalatingalarms: Highpriorityalarmsescalatetoahigherpitch

ifunattendedforspecifiedtime

Adjustableto: 0,10,20and30sec,Off

Autolimits: Alarmlimitscalculatedonthecurrentmeasuredvaluesforselectedparameters

Procedures

Suction

Programroutine:Automatic

Pre-oxygenation: ≤2minuteswith100%O2with

automaticdisconnectiondetection*

Standbypause: ≤2minuteswithautomaticpatient(re-connection)detection

Post-oxygenation: ≤2minuteswith100%O2*

Note:FIO2canbesettolevelotherthan100%

*Note:5to75%abovecurrentFiO2setting

Manual breath

IntrinsicPEEP(includesPEEPiVolume)

LungMechanics: PØ.1 NIF VitalCapacity

Inspiratoryhold:2to15sec(incrementsof1sec)

Expiratoryhold: 2to20sec(incrementsof1sec)

SpontaneousBreathingTrial(SBT)(Adjustablerange:2to120minutes)

Control and Ranges (continued)

Insp.pause: 0to75%ofinspirationtime(incrementsof5%)

Tpause: 0to7.5sec 0to1sec(incrementsof0.05) 1to4sec(incrementsof0.1) 4to7.5(incrementsof0.25) 0to11sec 0to1sec(incrementsof0.05) 1to4sec(incrementsof0.1) 4to11(incrementsof0.25)

PressuresupportfromPEEPlevel: 0to60cmH2OforSIMV-VC,SIMV-PC, SIMV-PCVG,BiLevel,BiLevel-VGand CPAP/PSV(incrementsof1cmH2O) 0to30cmH2OforNIV(increments of1cmH2O)

Endflowlevel: 5to80%ofpeakflowforNIV,SIMV-VC,SIMV-PC,SIMV-PCVG,BiLevel,BiLevel-VG,VG-PSandCPAP/PSV(incrementsof5%)

Alarm Settings

Tidalvolume: Low: Off,1to345mL Off,5to1950mL High: 3to350mL,Off 10to2000mL,Off

Minutevolume: Low: 0.01to10L/min 0.01to40L/min High: 0.02to40L/min 0.4to99L/min

Respiratoryrate: Low: Off,1to99/min High: 2to150/min,Off 2to120/min,Off

Inspiredoxygen(FiO2): Low: 18to99% High: 24to100%,Off

Pmax: High: 7to100cmH2O 9-100cmH2O(incrementsof 1cmH2O)inNIVandnCPAP

Ppeak: Low: 1to97cmH2O

PEEPe: Low: Off,1to20cmH2O High: 5to50cmH2O,Off

PEEPi: High: 1to20cmH2O,Off

Plimit: 7to100cmH2O

Apneaalarm: Useradjustable: 5to20sec 10to60sec

Circuitleak: 10to90%,Off

Page 5: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

FRC INview (optional)Note:NotavailablewhenNeonatalpatienttypeisselected

• FunctionalResidualCapacitymeasurement

– Wash-inandWash-outmethodprovides2separate FRCmeasurements

• FRCdisplayedbothnumericallyandgraphically

• Themostrecent5FRCproceduresdisplayed

• PEEPeandPEEPidisplayedwitheachFRC

• FRCEventLogrecords:

– FRCmeasurements

– Ventilatorsettingsandproceduresthatmayaffectthe FRCprocedure

• Programmabletimeintervalsforautomatic FRCmeasurements

PEEP INview (available with FRC INview)Note:NotavailablewhenNeonatalpatienttypeisselected

• MeasuresFRCatupto5differentPEEPlevels

• GraphicandnumericdisplayofFRCvalues

• UserselectablebeginningandendingPEEPlevels

– VentilatorevenlyspacesadditionalPEEPlevels

– PEEPlevelscaneitherincreaseordecrease

• PEEPeandPEEPidisplayedduringeachFRCmeasurement

Lung INview (available with SpiroDynamics and FRC INview)Note:NotavailablewhenNeonatalpatienttypeisselected

• IntegratesSpiroDynamicsandFRCINviewwithinthe PEEPINviewprocedure

• MeasurestheamountofvolumebetweentheDynostatic curvesateachFRCmeasurement

• Estimateofrecruitmentvolume

INview Vent Calculations (optional) Note:NotavailablewhenNeonatalpatienttypeisselected

DatafromEngströmCarestationandexternallabresultsareusedtoprovidethefollowingvalues:

• PAO2–Alveolarpartialpressureofoxygen

• AaDO2–Alveolararterialoxygendifference

• Pa/FiO2–Oxygenationindex

• PaO2/PAO2–Alveolararterialoxygenpressuredifference

• CO–Cardiacoutput

• Vd/Vt–Deadspaceventilation

• Vd–Deadspacevolume

• VA–Alveolarventilation

Spirometry

Datasource: VentilatororCompactAirwayModule* (M-COV,M-COVX,M-CAiOV,M-CAiOVX;E-COV, E-COVX,E-CAiOV,E-CAiOVX)

Looptypes: Pressure-Volume,Pressure-FlowandFlow-Volume

Savedloop: Uptosixloopscanbesaved

Referenceloop: Asavedloopcanbeselectedasthereferencelooptocomparewiththecurrentloopbeingdisplayed

Cursor: FreezescurrentloopsandprovidesnumericdisplayofXandYaxisascursormovesacrossloops

Pulmonarymechanics: Ppeak,Pplat,Pmean,PEEPe,PEEPi,TVinsp,TVexp,MVinsp,

MVexp,ComplianceandResistance

Auxiliary Pressure

Auxiliarypressure(Paux): Measuredrange:-20to+120cmH2O Alarmrange:12to100cmH2O

Purgeflow: LowflowtoclearthePauxline,canbe turnedOff

SpiroDynamics (optional)Note:NotavailablewhenNeonatalpatienttypeisselected

• TrachealPressure–Volumeloopdisplayed

• DynostaticCurvedisplayscalculatedalveolarpressure

• TrachealpressuremeasuredviaGE’sintratracheal pressuresensor*

–ConnectsdirectlytoEngström’sauxiliarypressureport

• 3pointcompliancemeasurement

– at5-15%ofthebreath

– at45-55%ofthebreath

– at85-95%ofthebreath

• Storeupto6SpiroDynamicloops

• Storeupto6Dynostaticcurves

• Overlayupto2separateloopsand/orcurvesover currentloop

• Cursoravailableacrossalldisplayedloopsandcurves

– Pressureandvolumevaluesdisplayedatcursorposition

• Ppeak,PEEPeandRawdisplayed

*Forcompletespecifications,seeproductspecificationsheets.

Page 6: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Airway Resistance Compensation (ARC)

Note:NotavailableinNeonataloption

Typeofcompensation: Electronictubecompensation

Compensationfor: Endotrachealandtracheostomytubes

Tubediameter: 5to10mm

Levelofcompensation: 25to100%

Mode Families

Allowsuseradjustmenttospecifycertainparametersthatalignwiththehospital’scurrentventilatorusage.

Adjustableparameters: FlowandInspiratorytiming

Family1: FlowcontrolisOn/Insp.TimingisI:EFamily2: FlowcontrolisOff/Insp.TimingisI:EFamily3: FlowcontrolisOn/Insp.TimingisTinspFamily4: FlowcontrolisOff/Insp.TimingisTinspFamily5: FlowcontrolisOn/Insp.TimingisTpause

Ventilator Monitoring

Airwaypressure –20to+120cmH2O

Patientflow 0.1to32L/min 1to200L/min

Tidalvolume 0.5to1,000mLwithNeonatal FlowSensor 1to1,000mLwithouttheNeonatal FlowSensor 5to2,500mL

Minutevolume 0to99.9L/min

CO2 0to30%/0to225mmHg

Compliance 0.1to150mL/cmH2O

Resistance 1to500cmH2O/L/s

RQ 0.6to1.3

VO2 50to1000mL/min

VCO2 50to1000mL/min

Rate 0to150breathsperminute (incrementsof1breathperminute) 0to120breathsperminute (incrementsof1breathperminute)

FiO2 10to100%

RapidShallowBreathingIndex(RSBI) 1to999bpm/L

Note:NotavailableinNeonataloption

Non-Invasive Ventilation (NIV) (optional)

Maskventilation:Yes

Integrateduniqueleakrecognitionalgorithm

Automatic Patient Detection (APD)

Patientre-connection: Automaticdetectioninstandby

Detectionby: BackpressuretoBias-flow

100% O2 ( O2)

Delivers5to75%abovecurrentFiO2settingfor≤2minutes

Delivers100%O2for≤2minutes

CanbeadjustedtootherO2%

Take Snapshot

ImmediatecaptureandstorageofcriticaldatacurrentlyontheEngström’sdisplay

Storeddata: 3waveformsegments Alarmmessages(upto5,currentlyactive) Allmeasuredparameters Allsetventilatorparameters

MaximumstoredSnapshots: 10mostrecent

Cursor: Abilitytocursoracrosswaveforms forspecificmeasuredvalues

Ventilator Preferences

Back-upMode: EstablishesthespecificventilatormodeandparametersusedintheeventthattheventilatorswitchestoBack-upventilation

ARC: Allowscontrolandsettingoftheairwayresistancecompensation

AssistControl: AllowstheusertoturntheAssistControlcapabilityOnorOff

LeakCompensation: AllowstheusertoturntheLeak

CompensationcapabilityOnorOffTriggerCompensation: AllowstheusertoturnOnorOff

compensationforflowtriggeringTVBasedConditions: AllowssettingbetweenATPD(Ambient

TemperaturePressureDry)orBTPS(BodyTemperaturePressureSaturated)

Page 7: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Nebulization

Nebulizer: AeronebNebulizerSystembuilt-in

Nebulizertechnology: Electronicmicropump

Nebulizerruntime:10,15,20or30minutes

Auto-repeatcapability: Cycles:1to10 PauseTime:30secto8hr 1to5minutes(incrementsof1minute) 5to55minutes(incrementsof5minutes) 1to8hours(incrementsof0.5hour)Nebulizervolumesetting: 2.5,3,5or6mL

Particlesize: AeronebPro:average2.1micronsMMAD (MeanMassAerodynamicDiameter) AeronebSolo:3.4micronsMMADResidualvolume: AeronebPro:average0.3mL AeronebSolo:average<0.1mL

Performancemayvarydependinguponthetypeofdrugused.ForadditionalinformationcontactAerogenordrugsupplier.

Pnuematic nebulizer

Flowcompensation: 1to4L/min(incrementsof0.5L/min) 1to12L/min(incrementsof0.5L/min)

Monitor Module

Modulecapacity 4singleslotor2doubleslotmodules

Compactairwaymodulecompatibility M-C,M-CO,M-COV,M-COVX,M-CAiO, M-CAiOV,M-CAiOVX,E-CO,E-COV, E-COVX,E-CAiO,E-CAiOV,E-CAiOVX, M-miniC,E-miniC

Note:TheEngströmCarestationdoesnotutilizetheAi(inhaledanesthesia)featureofthecompactairwaymodulesatthistime.

Note:TheEngströmCarestationdoesnotutilizeanyofthecompactairwaymoduleswhentheNeonatalOptionisinuse.

Trends

Trenddata: Setparametersandmeasureddata

Trendstyles: Measuredandgraphic

Maximumtrending:14days(336hours)

Trendscaling: 12min,1h,2h,4h,6h,8h,10h,12h,24h, 36h,48hand72h

Resolution: 1minuteintervalsformostrecent12hours, 5minuteintervalsfor12to48hours, 30minuteintervalsafter48hours

Mini-Trends: Waveformvaluescanbedisplayed asatrendinasplitscreenview

Oxygen Monitoring

Technology: DynamicParamagneticOxygen monitoringsystem

Lifespan: Unlimitedoperatinglifeduetotheuseof non-depletingtechnology

Screen

Displaytype: 30.5cm/12inchtouchscreenfullcolorLCDadjustableviewingangle

Waveformsinscreen: Threeatatime

Waveformparameters: Pressure,flow,volume,CO2,O2

andauxiliarypressure

Graphicscaling: Automaticscalingforoptimalsizeor independentscaling

Data: Controlparameters,patientdata,alarmsettingsandmessages

Statusindicator: Ventilationmode,batterylevel,clock

Favorites: 23Hyperlinkshortcutstochoosefrom 7selectableatonetime

Monitoring Accuracy**

Pressurereadings:±2cmH20

Volumereadings: ±10%or±1mL,whicheverisgreater (withproximalneonatalflowsensor) ±10%or±5mL,whicheverisgreater (nCPAP) ±10%or±15mL,whicheverisgreaterO2concentrationmonitor: ±3%

Delivery Accuracy**

Inspiredpressurecontrol: ±2cmH2O

Oxygen–Airmixing: ±3%V/Vofsetting

Tidalvolumedelivery: ±10%ofsettingor±1mL, whicheverisgreater (withproximalneonatalflowsensor) ±10%ofsettingor±5mL, whicheverisgreater

**Ventilationdeliveryspecificationsrequirements: •OperatingatEN794andASTMF1100patientconditions •Operatingat21°Candat1000mbarambientpressure •Allvolumesareatambienttemperatureandpressure,dry(ATPD)

Page 8: Engström Carestation - GE Healthcaregerespiratorycarecentral.com/pdf/EngstromCarestationSpec...Engström Carestation Breathing life into critical care GE Healthcare Physical Specifications

Gas supply

Singlegasoperation: Yes

Emergencyairvalve: Built-in

Oxygen supply

Pressurerange: 240to641kPa/35to94psi

Flow: 160L/min

Air supply

Pressurerange: 240to641kPa/35to94psi

Flow: 160L/min

Environmental Specifications

Thermal

Operatingrange: 10°to40°C

Storagerange: –20°to65°C

Humidity

Operatingrange: 15to95%RHNon-condensing

Storagerange: 15to95%RHNon-condensing InaccordancewithIEC68-2-3

Vibration and shock

Randomvibration: 9.5grms@30minunpacked 2to5000Hz

Altitude

Operatingrange: –440to3565m/500to800mmHg

Storagerange: –440to5860m/375to800mmHg

Trends (continued)

Mini-Trendsparametersarebasedonthewaveformdisplayed: Paw(Ppeak,PplatorLeak) Flow(MVexp,RR) Volume(SpontMVorMechMV, SpontRRorMechRR) Paux(Ppeak) CO2(EtCO2) O2(EtO2,FiO2)

External Communications

Communicationports: Serialport(RS-232),RS-485port,RS-422port, 1USBport,Ethernetport,Compactflash cardsocket,nursecall

EView (optional)

DataAvailable: 10snapshots 7daysofventdata Optionalbreathtobreathwaveformdata

Electrical Specifications

Line supply

Linevoltage: 85to132Vac,47/63Hz 190to264Vac,47/63Hz

Powerconsumption: <200W

Battery supply

Back-upbattery:Built-in

Type: Leadacidgel

Batteryback-uptime: 120minutestypical,30minuteminimum, batteryfullycharged

GEHealthcareP.O.Box7550Madison,WI53707-7550U.S.A.

www.gehealthcare.com/respiratorycare

©2011GeneralElectricCompany–Allrightsreserved.

GeneralElectricCompanyreservestherighttomakechangesinspecificationandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.ContactyourGErepresentativeforthemostcurrentinformation.

GEandGEMonogramaretrademarksofGeneralElectricCompany.

AeronebisaregisteredtrademarkofAerogen,Inc.

Engström,Carestation,INviewandSpiroDynamicsaretrademarksofDatex-Ohmeda,Inc.

GEHealthcare,adivisionofGeneralElectricCompany.

NOTCOMMERCIALLYAVAILABLEINALLMARKETS.NOTCLEAREDORAPPROVEDBYTHEU.S.FDA.

RSP-0149-03.11-EN-US