1 Energy Device Options – Strategies to Energy Device Options Strategies to Prevent Injury Dean J Mikami, MD. FACS Associate Professor of Surgery Center for Minimally Invasive Surgery The Ohio State University Program Content • Basics of Electricity • Principles of Electrosurgery • Clinical Applications • Electrosurgical Technologies Electrosurgical Technologies • Associated Risks • Recommendations for Safe Practices
29
Embed
Energy Device Options – Strategies to Prevent InjuryEnergy Device Options – Strategies to Prevent Injury Dean J Mikami, MD. FACS ... resistance Properties of Electricity Current
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
11
Energy Device Options – Strategies to Energy Device Options Strategies to Prevent Injury
Dean J Mikami, MD. FACSAssociate Professor of Surgery
Center for Minimally Invasive SurgeryThe Ohio State University
• AORN estimates c. 40,000 patients burned by faulty ES devices per yeary p y
• Up to 70% of ES burns in laparoscopic surgery may be undetected at the time of injury
• ACS survey: 18% of surgeons had experienced insulation failure or capacitive coupling injury, 54% knew a colleague who had a stray electrical burnknew a colleague who had a stray electrical burn
• In 1999, nearly $600 million paid in claims re ES injuries
Source: Outpatient Surgery Feb 2002
Two Types of Energy Used in Surgery
Electromagnetic Energy Mechanical Energy
– Electrosurgery
• Monopolar Electrosurgery
• Bipolar Electrosurgery
– Laser
–Suturing
–Stapling
–Ultrasonic (Ultrasound)
33
Basic Principles of Electricity
Electricity always . . .
– Seeks ground (its source)
– Seeks the path of least resistanceresistance
Properties of Electricity
Current Flow of electrons during a period of time, measured in amperes
Circuit Pathway for the uninterrupted flow of electrons (must be complete/ closed to flow)
Impedance Obstacle to the flow of current measured in ohms
Voltage Force pushing current through the resistance, measured in volts
• Do not use in presence of flammable material (e.g. alcohol, nitrous oxide)alcohol, nitrous oxide)
• Patient not in contact with any metal objects
Generator (ESU) Power Settings
• Use the lowest possible setting to achieve the desired surgical effect
• The setting will depend upon conditions such as:
– Patient size
– Generator power capabilities
– Target tissue type
– Electrode configuration
– Surgeon Technique
– Location of patient return electrode
77
Dispersive Electrode Guidelines
• Contact with patient must be uniform over large surface area
• Avoid the following:– Bony prominences
– Metal implants or prosthesis
– Scar tissue
– Hairy areasHairy areas
– Adjacent to leads/electrodes
– Pressure areas/points
• Never cut to size
Electrode Size and Effect on Power Settings
88
Current Concentration/Dispersal
The more concentrated the current, the greater the potential for a burn.
Patient Burns
Pad site burn
99
MonoPolar Monopolar InstrumentsActive Electrode
Patient
Low Voltage (“Cut”) “Pure”
30
El t d 1
Dispersive Electrode
High Voltage (“Coag”)0
Electrode 1
Electrode 2
Monopolar
• Active electrode at surgical site
• Return electrode at another site
• Current flows through the body between the electrodes
• High voltageg g
– Coag – 3000 – 9000 V
– Cut – 1350 – 4000 V
1010
Mushroom EffectAs the tissue desiccates Impedance Increases
Waveforms(Blend is a CUT mode)
1111
Direct Coupling
• Occurs when one conductive source touches or arcs to another
1. Current may be directed toward non-target tissue
2. Instrument in contact with active electrode may not be completely in view (laparoscopic case) and/or contacting other tissue (bowel or abdominal wall)
Capacitive Coupling
• Capacitance: defined as stored electrical charge when two conductors separated by an insulator
• Capacitive coupling current occurs when the circuit is completed through the dielectric (e.g. insulator)
• Charge stored in capacitor until either generator is deactivated or pathway to complete circuit is achieved
1212
Alternate Site Injuries
• Current delivered must return to generator
• Much less common today with isolated generators (will not deliver more current to electrode if not enough current returns to generator, i.e. leaves via an alternate site)
• Precautions: patient should not be in contact with any objects with high conductivity
Inadvertent Activation
• Beware of creating an injury out of field of view (blind instrument insertion or inadvertent electrode application)
1313
Direct Thermal Extension
• Duodenum often adherent to or in close proximity toto or in close proximity to gallbladder
• Use short activations of electrode (2-3 sec)
Beware the adhesion with• Beware the adhesion with narrow attachment to duodenum
Adhesions between GB and duodenum
Surgeon Burns/Surgical Glove Injuries
• Holes are present in 15% f i l15% of new surgical gloves and in ½ after use in surgery
Surgeon Burns/Surgical Glove Injuries:Mechanisms of Injury
1. High voltages across glove (dielectric) break insulating capacity of glove
2. Decreased glove resistance (with time and exposure to saline (sweating)
3. Capacitive coupling – risk inversely proportional to glove thickness and increases with higher voltage and g glonger contact time (active electrode and touching hemostat)
Helpful Hints to Avoid Hemostat Burns
• Use lowest power setting possible
• Activate low voltage (cut) waveform
• Avoid touching the patient
Hold hemostat with full grip• Hold hemostat with full grip
• Do not activate in open circuit (avoid metal to metal arcing)
Note: Surgical gloves do not insulate against RF current
1515
Bipolar
• Active and return electrodes in theelectrodes in the instrument
• Current flow confined to tissue between electrodes
• Low Voltage
– 320 - 1200 volts
BipolarLow Voltage (“Cut”) “Pure”
30
Patient
High Voltage (“Coag”) 0
30
Electrode 1
Electrode 2
Two Active Electrodes
1616
Bipolar
• Benefits:
• Doesn’t Require Dispersive Electrode
• Energy Primarily stays between the Jaws
• Requires less Voltage and Current (“cut” waveform)Requires less Voltage and Current ( cut waveform)
Limitations of Traditional (simple) Bipolar
• Continuous, uninterrupted delivery of energy
• Surgeon controls the delivery of energy visually
• No feedback mechanism to determine impedance ( thermal Spread, carbonization)
Promotes e cessi e thermal damage• Promotes excessive thermal damage
• Devices coapt tissue poorly
1717
Modern Advance Bipolar Technology
– Smart Generator Technology
• Near real-time Impedance feedback from delivery device
• Pulsing (rapid on/off) energy delivery– Allows for interval tissue cooling
• Audible signal indicates adequate coagulation
• Vessel up to 7 mm vessels
Best Practices
• No tension– Let the instrument do its job– Let the instrument do its job
• Keep Jaws Clean: – Never activate with coagulum build-up on the
inside of the jaws.
• Vessels up to 7 mm– Be careful of calcified vessels.
1818
Best Practices
OVERLAPPING SEALS:– Consider overlapping seals in areas of anatomical– Consider overlapping seals in areas of anatomical
tension.– When overlapping seals, do so by 30-50%.
• Knife blade damage– Do not attempt to fuse or cut over clips, staples or
sutures
– Do not use the knife for cutting suture
Thermal Spread with Bipolar
• With Modern Bipolar Instrument– Seen with all electrosurgical instruments
– Tissue damage occurs at 60 degrees Celsius
– Range 1-4 mm
– Be aware of residual heat.
1919
Ultrasonic Technology: The Hand Piece
Electrical energy from the generator
MECHANICAL MOTION VIDEO
is converted to mechanical motion in the hand piece
Piezoelectric ceramics Mount Mechanical Wave
Actions of Ultrasound on Tissue
CoagulationCutting Cavitation
All surfaces of the ultrasonic blade are active.
BACK SCORING/DRILLING VIDEO CAVITATION VIDEO
2020
Factors affecting cutting and coagulation
1. Blade pressure
2. Tissue tension
3. Power level
4. Blade sharpness
Blade activation temperatures
ULTRASONIC
BIPOLAR
0 50 100 150 200
°c
2121
Thermal Spread
Blade activation temperatures
2222
Benefits and Risks of Ultrasonic Coagulation
V l l d ld d t th• Vessels are sealed or welded together
• Minimal spread of energy, but the blade is HOT
• Coagulum does not stick to blade
• Minimal smoke generation more water vapor
• No neuromuscular stimulationNo neuromuscular stimulation
• Uniformly coagulates 5mm vessels
Current Leakage
• Active electrode cords sho ld not be rappedshould not be wrapped around metal instruments
• Active electrode and other electrical device
d h ld bcords should not be bundled together
2323
Pacemakers
• Consult pacemaker manufacturep
• Use electrosurgery with care
• Use ultrasonic technology
• Use low power setting (bipolar)
• Avoid current flow through heart and pacemaker
• Keep cords away from pacemaker and leads
AORN Recommended Practices 2009
Automatic Implantable Cardioverter Defibrillators
• Use of Electrosurgery on a patient ith anon a patient with an activated AICD may trigger an electrical shock to the patient
• The AICD device should b d i d b fbe deactivated before the ESU is activated
AORN Recommended Practices 2009
2424
Active Electrodes
• When not in use, always place active electrodes in a non-conductive holster
• Active electrode tips should be securely seated into the hand piece (increase risk of sparking or burn to targeted tissue)
AORN Recommended Practices, 2009
Smoke/By-products
• Types of products from electrosurgery
– Toxic vapors and gases (e.g. benzene, hydrogen cyanide, formaldehyde)
– Chemicals and irritants (potentially mutagenic or carcinogenic)
– Bioaerosols including blood fragments
– Viruses
– Methemoglobin and carboxyhemoglobin in laparoscopic surgerysurgery
• Surgical masks filter size = 5 microns, 77% of surgical smoke contents are 1.1 microns or smaller
2525
Effects of Smoke/By-products
• High concentrations of smoke cause ocular and upper g pprespiratory irritation in health care personnel. – Note: No documentation of cancer cases from OR
smoke exposure.
• OSHA recommendation: smoke evacuation systems should be used to reduce acute and chronic risks to patients and health care personnelp p
• Do not activate ignition so rces in the presence ofsources in the presence of flammable agents
• Avoid pooling of prep
• Drape patient after vapors ape pa e a e apo sfrom flammable agents have dissipated
2626
OR Fires
• Rare but potentially devastating
• ECRI estimate: 550-650 cases/yr (similar to # wrong site surgery cases)
• 95% are minor and result in no injury
• 20-30 serious with disfiguring or disabling injuries
ECRI Institute is an independent, nonprofit organization that researches the best approaches to improving the safety, quality, and cost-effectiveness of patient care.