Page 1: ELECTROMYOGRAPHY

1

ELECTROMYOGRAPHY

EMG- BIOFEEDBACK

Page 2: ELECTROMYOGRAPHY

2

ELECTROMYOGHAPHY ELECTRO + MYO + GRAPHY

ELECTRICITY MUSCLE WRITE

Page 3: ELECTROMYOGRAPHY

3

DEFINITION Study of motor unit activity Recording of AP of muscle fibres

firing near the needle electrode in a muscle

Page 4: ELECTROMYOGRAPHY

4

INTRODUCTION

Luigi Galvani – 1791EMG

Clinical EMG Kinesiology EMG

Page 5: ELECTROMYOGRAPHY

5

MOTOR UNIT

Page 6: ELECTROMYOGRAPHY

6

SIZE PRINCIPLE

The muscle contraction depends on the “size principle”It states that “the motor neurons are recruited in order of size from small to large”

Page 7: ELECTROMYOGRAPHY

7

Page 8: ELECTROMYOGRAPHY

8

MOTOR UNIT ACTION POTENTIAL

Page 9: ELECTROMYOGRAPHY

9

Page 10: ELECTROMYOGRAPHY

10

Recording an EMG includes 3 phases systemInput phaseProcessor phaseOutput phase

Page 11: ELECTROMYOGRAPHY

11

INSTRUMENTATIONELECTRODESSurface electrodesNeedle electrodesFine wire indwelling electrodes

Page 12: ELECTROMYOGRAPHY

12

FINE WIRE INDWELLING ELECTRODE

Page 13: ELECTROMYOGRAPHY

13

SURFACE ELECTRODES

Page 14: ELECTROMYOGRAPHY

14

NEEDLE ELECTRODESCONCENTRICMONOPOLARSINGLE FIBREMACRO

Page 15: ELECTROMYOGRAPHY

15

3 ELECTRODES ARE USED

ACTIVE ELECTRODE

REFERENCE ELECTRODE

GROUND ELECTRODE

Page 16: ELECTROMYOGRAPHY

17

Page 17: ELECTROMYOGRAPHY

18

Page 18: ELECTROMYOGRAPHY

21

Page 19: ELECTROMYOGRAPHY

24

Page 20: ELECTROMYOGRAPHY

26

Page 21: ELECTROMYOGRAPHY

27

AMPLIFYING THE EMG SIGNAL Amplifier- converts the electrical potential seen by electrodes to a voltage signal large enough to be displayed

Differential amplifier-Rejects common mode voltages which appear between both input terminals and common grounds

Page 22: ELECTROMYOGRAPHY

28

Common mode rejection ratio- It is a measure of how much the desired signal voltage is amplified relative to the unwanted signal

Page 23: ELECTROMYOGRAPHY

29

Signal-to-noise ratio- It is the ratio of the wanted signal to unwanted signalGain- Ratio of output signal level to input level. A higher gain will make a smaller signal appear larger on the display

Page 24: ELECTROMYOGRAPHY

30

Input impedence- It is a resistive property, opposing current

flow, which occurs in alternating current circuits.

Affected by- Electrode material Electrode size Length of the leads Electrolyte

Page 25: ELECTROMYOGRAPHY

31

Frequency bandwidth- Bandwidth is the difference between highest and lowest frequency that will be processed.

Amplifier must be able to respond to signals between 10 and 10000 Hz

Page 26: ELECTROMYOGRAPHY

32

Displaying the EMG signalThe form of output used is dependent on the type of information desired and instrument available

Page 27: ELECTROMYOGRAPHY

33

WAVEFORM DISPLAY2 forms are used

1. Analog oscilloscope display

2. Computer based digital video display

Page 28: ELECTROMYOGRAPHY

34

EMG- RECORDING & DISPLAY

Page 29: ELECTROMYOGRAPHY

35

FACTORS AFFECTING OUTCOME OF RECORDING

Age of the patientsProperties of the muscle under study Limb temperatureElectrical specifications of the needle electrodes and the recording apparatus. e.g. filter settings

Page 30: ELECTROMYOGRAPHY

36

INDICATIONSNeurogenic disordersNeuromuscular junction disordersMyogenic disorders Metabolic disorders

Page 31: ELECTROMYOGRAPHY

37

CONTRAINDICATIONSRecurrent systemic infectionsBleeding disorders – hemophilia, thrombocytopenia, patient on anticoagulant therapyLocalized inflammationSkin lesions

Page 32: ELECTROMYOGRAPHY

38

serum creatinine kinase level- repeated examination causing inflammation and focal myopathic changes.Transient bacterimia following needle examination may lead to endocarditic in patients with valvular disease or prosthetic valves.

Page 33: ELECTROMYOGRAPHY

39

ELECTROMYOGRAPHY EXAMINATION

Select the muscle

Locate the needle insertion point

Insert the needle quickly in a relaxed state

Patient briefly activates the muscle to confirm the placement

Page 34: ELECTROMYOGRAPHY

40

ACTIVITIESInsertional ActivitySpontaneous ActivityMinimal Volitional ActivityMaximal Volitional Activity

Page 35: ELECTROMYOGRAPHY

41

INSERTIONAL ACTIVITYIt is the measure of muscle excitability.Spontaneous burst of potentials (muscle fibre depolarization) usually lasting less than 300ms after needle movement ceases.Positive and negative high frequency spikes in a cluster are seen.

Page 36: ELECTROMYOGRAPHY

42

Page 37: ELECTROMYOGRAPHY

47

SPONTANEOUS ACTIVITYMINIATURE END PLATE POTENTIALS(MEPP): End plate noiseEND PLATE SPIKES- Nerve potentials

Page 38: ELECTROMYOGRAPHY

53

ABNORMAL SPONTANEOUS ACTIVITYFIBRILATION POTENTIALPOSITIVE WAVESCOMPLEX REPETITIVE DISCHARGESMYOTONIC DISCHARGESFASCICULATION

Page 39: ELECTROMYOGRAPHY

54

MYOKYMIC DISCHARGESCRAMPSNEUROMYOTONIA

Page 40: ELECTROMYOGRAPHY

74

MINIMAL VOLITIONAL ACTIVITY

Motor unit action potential- voluntary contractionRecruitment

Page 41: ELECTROMYOGRAPHY

75

MOTOR UNIT ACTION POTENTIAL

Page 42: ELECTROMYOGRAPHY

76

PHYSIOLOGIC FACTORSAge Inherent properties of motor unit itself Spatial relationship between the needle and the individual muscle fibresResistance and capacitance of intervening tissuesIntramuscular temperature

Page 43: ELECTROMYOGRAPHY

77

NON PHYSIOLOGIC FACTORS

Type of needle electrodeSize of the recording surfaceElectrical properties of the amplifierChoice of oscilloscope sensitivitySweep or filtersMethods of storage and display

Page 44: ELECTROMYOGRAPHY

78

MUAPs

Morphology -Amplitude, duration, phase, initial deflection StabilityFiring characteristic

Page 45: ELECTROMYOGRAPHY

79

AMPLITUDEIt is measured from peak to peak .100V-2mVIt is determined by

Primarily by limited number of fibres located close to the electrode tip.

Size and density of the muscle fibre Synchrony of firing .

Page 46: ELECTROMYOGRAPHY

80

Age of the subject

Muscle examined

Muscle temperature-Decreasing muscle temperature results in higher amplitude and longer duration of MUPs

Page 47: ELECTROMYOGRAPHY

81

DURATIONMeasured from initial take off to the point of return to the baseline.Usually 5-15 msIt reflects the number of muscle fibres within a motor unit and is the most reliable measure to use when judging MUAP morphology.

Page 48: ELECTROMYOGRAPHY

82

The duration of MUP is a measure of Conduction Length of muscle fiber Membrane excitability Synchrony of different muscle fibres

of a motor unit

Page 49: ELECTROMYOGRAPHY

83

Duration increases with age and cold temperature.

Distal muscles have longer duration MUAPS than proximal ones .

Page 50: ELECTROMYOGRAPHY

84

RISE TIME OF MUPIt is the duration from initial positive to subsequent negative peakIndicator of the distance of needle electrode from the muscle fibre<500µs is acceptable

Page 51: ELECTROMYOGRAPHY

85

A greater rise time is due to resistance and capacitance of the intervening tissue which acts as high frequency filter and results in dull sound on the loudspeaker of EMG equipment. This indicates the need to reposition needle closer to the muscle fibres

Page 52: ELECTROMYOGRAPHY

86

PHASE OF MUPIt is defined as the portion of MUP between departure and return to the baseline.i.e no. of baseline crossings +1.Typical shape of an MUAP is diphasic or triphasic.

Page 53: ELECTROMYOGRAPHY

87

A MUP more than 4 phases is called as polyphasic.The normal MU will fire up to 15/sec with strong contraction.

Page 54: ELECTROMYOGRAPHY

88

SATTELITE POTENTIALThey are the late potentials which are time locked to the main motor unit potential.The satellite potential is generated by a muscle fibre in a motor unit with a long nerve terminal, narrow diameter or distant end plate region

Page 55: ELECTROMYOGRAPHY

89

StableSemi-rhythmic firing pattern

Page 56: ELECTROMYOGRAPHY

90

MAXIMAL VOLITIONAL ACTIVITY Interferance Pattern

Page 57: ELECTROMYOGRAPHY

91

The spike density and average amplitude of the summated response are determined by a number of factors:

Descending input from the cortex Number of motor units capable of

discharging Firing frequency of each motor unit Waveform of individual potentials Probability of phase cancellation

Page 58: ELECTROMYOGRAPHY

92

VARIABILITY IN MUPShort durationLong durationPolyphasicMixed patternDoublets and multiplets

Page 59: ELECTROMYOGRAPHY

93

SHORT DURATIONThe short duration MUPs are those with a duration shorter than that for the muscle of corresponding age

Usually have low amplitude and have rapid recruitment at minimal effort

Page 60: ELECTROMYOGRAPHY

94

They are found in the disorders associated with loss of muscle fibres2 chief pathologies are- Myopathies and neuromuscular

junction disorders Early stage of reinnervation after

nerve damage

Page 61: ELECTROMYOGRAPHY

95

LONG DURATION MUPSThe duration of MUPs exceeds the normal values for the corresponding muscle & ageHave high amplitude & poor recruitmentAssociated with increase in fibre density, loss of synchrony of firing of muscle fibre & increase in the number of muscle fibre

Page 62: ELECTROMYOGRAPHY

96

Seen in Motor neuron disease Axonal neuropathies with collateral

sprouting Chronic radiculopathies Neuropathies Chronic myositis –polymyositis

Page 63: ELECTROMYOGRAPHY

97

POLYPHASIC MUPThere is 4 or more phases in a MUPSeen in myopathies where there is regeneration of fibres and increased fibre density.

Page 64: ELECTROMYOGRAPHY

98

MIXED PATTERNComprises of short ,long, polyphasic MUPsFound in both Myogenic and Neurogenic abnormalities

Page 65: ELECTROMYOGRAPHY

99

DOUBLETS OR TRIPLETSMUPs are fired 2 or more times at an interval of 10-30 msSeen when there is hyperventilation, tetany, motor neuron disease, other metabolic diseases ischemia

Page 66: ELECTROMYOGRAPHY

100

CLINICAL IMPLICATIONS OF EMG

Page 67: ELECTROMYOGRAPHY

101

REFERANCESPHYSICAL REHABBILITATION: ASSESSMENT AND TREATMENT

Susan B.O’sullivanCLINICAL NEUROPHYSIOLOGY UK MishraDIAGNOSTIC TESTING IN NEUROLOGY Randolph W. EvansELECTRODIAGNOSTIC TESTING Kimura

Page 68: ELECTROMYOGRAPHY

102

EMG-BIOFEEDBACK

Page 69: ELECTROMYOGRAPHY

103

DEFINITION A technique of using equipment (usually

electronic) to reveal human beings some of their internal psychological events, normal and abnormal, in the form of visual and auditory signals in order to teach them to manipulate these otherwise involuntary or unfelt events by manipulating the displayed signals.

John V.Basmajian

Page 70: ELECTROMYOGRAPHY

104

OBJECTIVES AND GOAL

TO IMPROVE MOTOR PERFORMANCE BY FACILITATING MOTOR LEARNING

Page 71: ELECTROMYOGRAPHY

105

PRINCIPLES OF MOTOR LEARNING

Page 72: ELECTROMYOGRAPHY

106

MOTOR LEARNING SCHIMDT defined it as “a set of processes

associated with practice or experiences leading to relatively permanent changes in the capacity for responding”

Page 73: ELECTROMYOGRAPHY

107

Four primary factors that influence motor learning are

1. Stage of learner 2. Type of the task3. Feedback 4. Practice

Page 74: ELECTROMYOGRAPHY

108

TYPES OF FEEDBACK INTRINSIC FEEDBACK EXTRINSIC FEEDBACK-1. KNOWLEDGE OF RESULTS2. KNOWLEDGE OF PERFORMANCE

Page 75: ELECTROMYOGRAPHY

109

EQUIPMENTS USED The basic EMG biofeedback device

includes one ground and two surface electrodes, an amplifier, an audio speaker and a video display.

Page 76: ELECTROMYOGRAPHY

110

The quality of the machine and its output are chiefly governed by Electrodes used Input impedance Common mode rejection ratio Bandwidth Gain Noise level Ability to cope with non EMG artifacts

Page 77: ELECTROMYOGRAPHY

111

TECHNICAL LIMITATIONS RELEVANCY ACCURACY RAPID TO ENHANCE MOTOR

LEARNING

Page 78: ELECTROMYOGRAPHY

112

BIOFEEDBACK IN REHABILITATION

Page 79: ELECTROMYOGRAPHY

113

When using the biofeedback the patient should

Understand Practice Perform

Page 80: ELECTROMYOGRAPHY

114

CLOSE LOOP OPEN LOOP SCHEDULED LOOP

Page 81: ELECTROMYOGRAPHY

115

Biofeedback can assist rehabilitation process by:1. Providing a clear treatment outcome2. Permitting the therapist and patient to

experience with various strategies that generate motor patterns

3. Reinforcing appropriate motor behavior.4. Providing a process oreinted, timely and

accurate KP or KR of the patients’ efforts.

Page 82: ELECTROMYOGRAPHY

116

CONDITIONS STROKE SPINAL CORD INJURIES CP AND TBI MULTIPLE SCLEROSIS DYSTONIAS AND DYSKINESIS PERIPHERAL NERVE DENERVATION PAIN MANAGEMENT

Page 83: ELECTROMYOGRAPHY

117

THERAPEUTIC INTERVENTION

Page 84: ELECTROMYOGRAPHY

118

TREATMENT SESSION PATIENTS FUNCTIONAL

ASSESSMENT PROBLEM IDENTIFICATION THERAPEUTIC INTERVENTION

Page 85: ELECTROMYOGRAPHY

119

REFERANCESPHYSICAL REHABBILITATION:

ASSESSMENT AND TREATMENT Susan B.O’sullivanBIOFEEDBACK PRINCIPLES AND

PRACTICE FOR CLINICIANS John V. BasmajianPHYSICAL MEDICINE AND

REHABILITATION-PRINCIPLES & PRACTICE

Joel A. Delisa

Page 86: ELECTROMYOGRAPHY

120

THANK YOU