ECG in Pacemaker Malfunction

Sriram Rajagopal,

Southern Railway Hospital,

Perambur,Chennai

Page 2: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG in “Pacemaker Malfunction”(ECGs in Pacemaker Function

Assessment)

Sriram Rajagopal,

Southern Railway Hospital,

Perambur,Chennai

Page 3: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Implantable pulse generator (IPG)

Lead wire(s)

Implantable Pacemaker Systems Contain the Following Components:

Page 4: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Pseudomalfunctions

Pseudomalfunctions are defined as: Unusual Unexpected Eccentric

ECG findings that appear to result from pacemaker malfunction but that represent normal pacemaker function

Page 5: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 6: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 7: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Pace

• Sense – ( what ? )

• Respond – Inhibit , Trigger or Dual

• Respond to increased metabolic demand

by providing rate responsive pacing

• Provide diagnostic information stored by

the pacemaker

What do pacemakers do ?

Page 8: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Timing Intervals Are Expressed in Milliseconds

• One millisecond = 1 / 1,000 of a second

Page 9: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

60,000

ms BPM

Rate Conversion

• Conversion– Pacing rate in PPM divided by 60,000 = ms

• 60,000 / 60 PPM = 1000 ms

– Interval in ms divided by 60,000 = PPM• 60,000 / 1000 ms = 60 PPM

Page 10: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

VVI / 60

Page 11: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

VVI / 60

Page 12: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

AAI / 60

Page 13: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Noncapture is Exhibited By:• No evidence of depolarization after pacing artifact

Loss of capture

Page 14: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

No Output

• Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate

Pacing output delivered; no evidence of pacing spike is seen

Page 15: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Fusion Beat

• Definition: The combination of an intrinsic beat and a paced beat.

• The morphology varies; in other words, a fusion beat doesn’t really look like a paced beat or an intrinsic beat.

• The pacemaker and the patient contribute to depolarization in Fusion beats.

Page 16: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Fusion

• Ventricular Fusion

Page 17: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Pseudofusion Beat

• Definition: A pacing pulse falls on an intrinsic beat. The pacing pulse is ineffective and the intrinsic complex is not altered.

• The pacemaker does NOT contribute to depolarization in Pseudofusion beats.

Page 18: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Pseudofusion

• Ventricular Pseudofusion

Page 19: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

• One lead implanted in the atrium

• One lead implanted in the ventricle

Dual-Chamber Systems Have Two Leads:

Page 20: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

DDD / 60 / 120

Page 21: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

DDD / 60 / 120

Page 22: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

DDD / 60 / 120

Page 23: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Paced Rhythm Recognition

DDD / 60 / 120

Page 24: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Sensing

• Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring– Pacemakers sense cardiac depolarization by

measuring changes in electrical potential of myocardial cells between the anode and cathode

Page 25: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Sensitivity – The Greater the Number, the Less Sensitive the Device to

Intracardiac Events

Page 26: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

SensitivityA

mp

litu

de

(mV

)

Time

5.0

2.5

1.25

Page 27: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

SensitivityA

mp

litu

de

(mV

)

Time

5.0

2.5

1.25

Page 28: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

SensitivityA

mp

litu

de

(mV

)

Time

5.0

2.5

1.25

Page 29: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Undersensing . . .

• Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately

Intrinsic beat not sensed

Scheduled pace delivered

VVI / 60

Page 30: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Oversensing

• An electrical signal other than the intended P or R wave is detected

Marker channel shows intrinsic

activity...

...though no activity is present

VVI / 60

Page 31: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 32: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Clinical details – age , indication for

pacing, time since implant etc

• Type of pacemaker

• Programmed parameters

• Magnet Behaviour

• Special features

Basic Data :

Page 33: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Magnet Operation

• Magnet application causes asynchronous pacing at a designated “magnet” rate

Page 34: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Rate Responsive Pacing

• An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior

VVIR / 60 / 120VVIR / 60 / 120

Page 35: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Hysteresis

• Allows a lower rate between sensed events to occur; paced rate is higher

Lower Rate 70 ppm Hysteresis Rate 50 ppm

Page 36: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 37: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Identify underlying intrinsic rhythm if any ( in each chamber)

• Verify appropriate sensing ( if possible )

• Verify capture

• Measure base rate and check if appropriate

• Identify any variations in intervals and interpret

• Identify causes of inappropriate sensing or pacing

Single chamber pacing :

Page 38: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 39: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

1

Single Chamber ECG AnalysisProgrammed Parameters

Mode………………………………………….. VVIBase Rate……………………………………….. 70 ppmMagnet Response…………………….. Battery TestHysteresis Rate………………………………… Off ppm

T Temporary programmed value7 Mar 2000 23:201.0 Second

Page 40: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #1

• VVI

• Normal Capture and Sensing

1

Page 41: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

2

Single Chamber ECG AnalysisProgrammed Parameters

Mode………………………………………….. VVIBase Rate……………………………………….. 70 ppmMagnet Response…………………….. Battery TestHysteresis Rate………………………………… Off ppm

T Temporary programmed valueJun 14 1999 2:57 pm1.0 Second

Page 42: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #2

• VVI

• Normal Sensing

• Capture unknown

2

Page 43: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

3

Single Chamber ECG Analysis

Page 44: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #3

• VVI

• Normal Capture and Sensing with initiation of Hysteresis

3

Page 45: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

4

Single Chamber ECG Analysis

Page 46: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #4

• VVI

• Loss of Ventricular Sensing

• Ventricular Undersensing with Functional loss of capture on the second to last beat

4

Page 47: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

5

Single Chamber ECG Analysis

Page 48: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #5

• VVI

• Loss of Ventricular capture with functional loss of sensing

5

Page 49: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

6

Single Chamber ECG Analysis

Page 50: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #6

• VVI

• Normal Capture

• Ventricular Oversensing

6

Page 51: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

10

Single Chamber ECG Analysis

Page 52: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

10

Single Chamber ECG Analysis

Page 53: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #10

• VVI

• Capture Unknown

• Normal Sensing

• Normal initiation of Hysteresis

10

Page 54: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

12

Single Chamber ECG Analysis

Page 55: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

12

Single Chamber ECG Analysis

Page 56: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #12

• VVI

• Normal Capture

• Ventricular Undersensing

12

Page 57: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

13

Single Chamber ECG Analysis

Page 58: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

13

Single Chamber ECG Analysis

Page 59: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #13

• VVI

• Loss of Ventricular Capture

• Normal Sensing

13

Page 60: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

15

Single Chamber ECG Analysis

Page 61: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG #15

• VVIR

• Normal Capture and Sensing under Sensor drive

15

Page 62: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Crosstalk

Page 63: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Basic principles of pacing

• Background information required

• Systematic approach

• Examples with single chamber pacing

• Examples with dual chamber pacing

Overview :

Page 64: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

A Systematic Approach

• Measure Base Rate• Measure AV/PV (PAV/ SAV) Interval• Verify Atrial capture• Verify Atrial sensing• Verify Ventricular capture• Verify Ventricular sensing• Verify Underlying rhythm• Document

Page 65: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 3

Dual Chamber ECG Analysis

Base Rate 60 ppm

MTR 120 ppm

AVD 200 ms

PVARP 250 ms

Page 66: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #3

• Loss of Atrial Capture

• Normal Atrial Sensing

• Normal Ventricular Capture

• Ventricular Sensing Unknown

Base Rate 60 ppm

MTR 120 ppm

AVD 200 ms

PVARP 250 ms

Page 67: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 5

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 150 msPVARP 250 ms

Dual Chamber ECG Analysis

Page 68: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #5• Normal Atrial Capture

– Possible Psuedofusion on 4th atrial output

• Atrial Sensing Unknown• Loss of Ventricular Capture• Normal Ventricular Sensing

– Functional Loss of Ventricular Sensing

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 150 msPVARP 250 ms

Page 69: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 10

Dual Chamber ECG Analysis

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 70: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #10

• Normal Atrial Capture with one beat showing functional loss of atrial capture

• Atrial Undersensing• Normal Ventricular Capture• Ventricular Sensing Unknown

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 71: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 12

Dual Chamber ECG Analysis

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 72: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #12• Normal Atrial Capture

• Normal Atrial Sensing

• Normal Ventricular Capture with two beats of functional loss of capture

• Ventricular Undersensing

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 73: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 20

Dual Chamber ECG Analysis

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 74: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #20

• Initiation of 2:1 block upper rate response

• Normal Upper Rate Behavior Operation

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 200 msPVARP 250 ms

Page 75: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECG # 24

Dual Chamber ECG Analysis

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 150 msMin. PV 88 msPVARP 250 ms

Page 76: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Answer ECG #24• Normal Atrial Capture• Normal Atrial Sensing • Normal Ventricular Capture• Ventricular Sensing Unknown• Initiation of a Pacemaker Mediated

Tachycardia (PMT) with following a PVC

Base Rate 60 ppmMTR 120 ppmAV 200 msPV 150 msMin. PV 88 msPVARP 250 ms

Page 77: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

• Familiarity with basics of pacemaker function

• Knowledge of clinical background and details of

pacing system

• Programmed parameters – essential

• Familiarity with special features

• Systematic approach

Conclusions :

Page 78: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

Thank you for your kind attention !

Page 79: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 80: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 81: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 82: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 83: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 84: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 85: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 86: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers

Page 87: ECG in Pacemaker Malfunction Sriram Rajagopal, Southern Railway Hospital, Perambur,Chennai.

ECGs in Patients with Pacemakers