InfoCard #: ABMT-GEN-025 FRM2 Rev. 10 Effective Date: 09 Aug 2019 DukeMedicine Division of Cellular Therapy DOCUMENT NUMBER: ABMT-GEN-025 FRM2 DOCUMENT TITLE: Photopheresis Run Sheet FRM2 DOCUMENT NOTES: Document Information Revision: 10 Vault: ABMT-General-rel Status: Release Document Type: General Date Information Creation Date: 02 Aug 2019 Release Date: 09 Aug 2019 Effective Date: 09 Aug 2019 Expiration Date: Control Information Author: MC363 Owner: MC363 Previous Number: ABMT-GEN-025 FRM2 Rev C Change Number: ABMT-CCR-270 CONFIDENTIAL - Printed by: ACM93 on 09 Aug 2019 08:47:54 am
6
Embed
DukeMedicine Division of Cellular Therapyspitfire.emmes.com/study/duke/SOP/Section C Collection/ABMT-GEN-025 FRM2...See Quick Reference section, Table 2: Body Build Factors and Safe
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
ABMT-GEN-025 FRM2 Photopheresis Run SheetInstructions
Name/History #/Date
Therakos Serial Number (SN)
Supplies Pass Visual InspectionKit Lot #, Exp Date, and Card #Saline 500 mL Lot #
ACD-A 500 mL Lot #Treatment #
AgeAccess
Lamp HoursRequired Settings for Single and Double Mode
Single Needle Mode-Double Needle ModeSingle Needle Mode: During prime change collect rate to25 mL/min
Double Needle Mode: During prime change collect rateto 25 mL/min
Double Needle Mode: During prime change return rateto 10 mL/min
Double Needle Mode: Return rate needs to be at least
5 mL/min higher than collect rate
Current HCT%
WBC countEstimated Total Blood Volume
Maximum Safe Extracorporeal Volume RangeCELLEX Predicted ECV
Reserve Volume
Pre photopheresis vital signsEMR (Electronic medical record)
Notify MD for abnonnal values
Important Infomiation when using IV Calcium
Record patient name, history number, and date. (Youmay use patient supplied label).Check the serial number of the machine being operated.Visually inspect all supplies and check Yes or No.Record kit log#, the exp. date, NS, and write in card #.Record saline lot # and exp. Date.Record ACD-A lot # and exp. Date.Record treatment number.
Record patient's current age.Record type of access being used.Record lamp hours at beginning of procedure.Record adjusting the return minimum pressure to-100 (default is-50)
Record adjusting the return bag threshold to 100 (defaultis 200)Select single needle mode or double needle modeHelp reduce system pressure alarms at beginning oftreatment per manufactures recommendation.Help reduce system pressure alarms at beginning oftreatment per manufactures recommendation.Help reduce system pressure alarms at beginning oftreatment per manufactures recommendation.If using double needle mode, return rate at least 5mL/min to reduce patient's fluid balance duringtreatment.
Record current HCT. Must be cun'ent within 48 hours ofprocedure.WBC count on day 1 of procedureRecord weight in kg and Body build factor . Multiplyresults=TBV. See Quick Reference section, Table 2:Body Build Factors and Safe Extracorporeal BloodVolumes (ECV) Limits.Multiply TBV by 10% and 15% and record.Record the CELLEX ECV. See Quick Referencesection, Table 1: Estimated Extracorporeal Volume(ECV) Relative to %Hematocrit, Access Mode andReturn Bag Threshold Value.Subtract the CELLEX estimated ECV from the 15% ofTBV.Record time and vital signs prior to startingphotopheresis to monitor fluid status.Checking In EMR, patient's information will bedocumented in the electronic medical record.
Contact the physician if any of the values are not withinlimits.
When using IV Calcium, the infusion must bediscontinued prior to photoactivation return phase ofprocedure. (Calcium and Methoxsalen are notcompatible)
ABMT-GEN-025 FRM2 Photopheresis Run SheetABMT, DUMCDurham, NC Page 1 of 2
CONFIDENTIAL - Printed by: ACM93 on 09 Aug 2019 08:47:54 am
ABMT-GEN-025 FRM2 Photopheresis Run SheetInstructions (continued)
Photopheresis monitoring
Room Temp/HumidityRoom Temp/Humidity Acceptable?
Uvadex dose
Lot# and exp. dates
Photoactivation time
Fluid balance
Treatment time
Machine cleaned by
Post photopheresis vital signsEMR (Electronic medical record)
Signature and date
Eveiy 15 minutes, record the time, collect pressures,return pressures, and interface # during collect phase ofprocedure. Note any comments and/or alarms thatoccurred and what relevant actions were taken toeliminate the alarms.
Record the current temp/humidity during the procedure.Check whether or not the room temp/humidity areacceptable. If not notify Apheresis Coordinator/designeeand record on the ABMT-GEN-021 FRM3
Photopheresis Temperature and Humidity LogTemperature/RH Troubleshooting Log.Calculate Uvadex dose by using:Treatment volume (mL) multiplied by 0.017= dose ofUvadex (mL).Multiple dose ofUvadex per MD by 20 mcg to find themicrogram dose.
Record lot # and exp. dates for syringes and needle usedduring procedure.Record photoactivation time when photoactivation startsin minutes.
Record fluid balance at end of procedure in mLs.Record treatment time at end of procedure in hours andminutes.
Record initials when machine cleaned at end of
procedure.Record time and vital signs post photopheresisChecking EMR, if patient's information will bedocumented in the electronic medical record.Record signature and date
ABMT-GEN-025 FRM2 Photopheresis Run SheetABMT, DUMCDurham, NC Page 2 of 2