Differential Scanning Calorimeter (DSC)
The technique was developed by E.S. Watson and
M.J. O'Neill in 1962, and introduced commercially
at the 1963 Pittsburgh Conference on Analytical
Chemistry and Applied Spectroscopy.
The first adiabatic differential scanning calorimeter
that could be used in biochemistry was developed
by P.L. Privalov and D.R. Monaselidze in 1964.
The term DSC was coined to describe this
instrument which measures energy directly and
allows precise measurements of heat capacity.
Introduction
Differential scanning calorimetry or DSC is a
thermoanalytical technique in which the difference in
the amount of heat required to increase the
temperature of a sample and reference is measured
as a function of temperature.
Both the sample and reference are maintained at nearly the same temperature throughout the experiment.
What is Differential Scanning Calorimetry ?
“Measures temperatures and heat flows associated with thermal transitions in a material”
“These are techniques by which the difference in heat flow to or from a sample and to or from a reference is monitored as a function of temperature or time, while the sample is subjected to a controlled temperature program”
DSC
Generally, the temperature program for a DSC analysis is designed such that the sample holder temperature increases linearly as a function of time.
The reference sample should have a well-defined heat capacity over the range of temperatures to be scanned.
A DSC is a twin instrument, comprising of individual sample and reference calorimeters within a common thermal enclosure, the cell.
Instrumentation
Each calorimeter will measure heat flow as a function
of temperature or time. The two calorimeters are
assumed to be identical. The term “calorimeters”
refers to each raised platform and its respective
sensor.
The output of the DSC is the difference between the
heat flows measured by each of the calorimeters.
The cell is heated at a linear heating rate. The sensor
body is made out of constantan (an alloy usually of
55% Cu and 45% Ni), consisting of a thick flat base
and a pair of raised platforms where the sample pan
and reference pan are positioned for the analysis.
The thin side (wall) of the sensor creates the
thermal resistance. This constantan material
provides good thermal conduction to the sample
and the reference.
As the temperature of the furnace is changed, heat
is transferred from the silver base of the enclosure
to the sensor body and thus to the pans. The
temperature of the furnace is controlled by the
refrigerated cooling system (RCS).
Thermocouples on the underside of each platform
are used to measure the temperature of the
sample and the reference. A third detector is used
to measure the temperature of the sensor base.
Pans are made from gold, copper, aluminium,
graphite and platinum or alodine-aluminium.
The lids can either provide a complete enclosure
or can contain a pin hole to release the pressure
that builds up inside, due to evaporation of water as
the temperature is increased.
Most samples can be run in non-hermetically sealed
pans either uncovered, or crimped with an aligned
or inverted cover. Atmospheric interaction is
optimised by using an open (uncovered) pan.
11
SELECTING CRUCIBLES FOR DSC
DSC pans & lids are available in aluminum, gold,
platinum, graphite, and stainless steel versions.
They can be used under a variety of temperature
and pressure conditions.
Samples can be run in the standard DSC mode in open pans, pressed or hermetically sealed pans/lids .
Sample Pans
When a sample undergoes a physical transformation such as a phase transition, more or less heat will need to flow to it than to the reference (typically an empty sample pan) to maintain both at the same temperature. Whether more or less heat must flow to the sample depends on whether the process is exothermic or endothermic.
For example, as a solid sample melts to a liquid it will require more heat flowing to the sample to increase its temperature at the same rate as the reference.
This is due to the absorption of heat by the sample as it undergoes the endothermic phase transition from solid to liquid.
Principle of Operation
Likewise, as the sample undergoes exothermic processes (such as crystallization) less heat is required to raise the sample temperature.
By observing the difference in heat flow between the sample and reference, differential scanning calorimeters are able to measure the amount of heat absorbed or released during such transitions.
DSC may also be used to observe more subtle phase changes, such as glass transitions.
Differential Scanning Calorimetry
• Differences in heat flow are measured as a function of temperature
Sample and reference
Energy difference measured
• Power Compensated DSC
Two furnaces
Heat flow evaluated
Difference in power input monitored
In watts
• Useful for glass transition temperatures
• Purity of drug samples
DSC • Constant Heating Rate
o Initial Temp
o Final Temp
o Heating Rate (°C/min)
• Data
o Heat flow to sample minus
Heat flow to reference Vs
Time (Temp.)
• Measures heat of
crystallization
Polymer without weight change in this temperature range
Detection of Phase Transitions
The basic principle underlying this technique is that when the sample undergoes a physical transformation such as phase transitions, more or less heat will need to flow to it than the reference to maintain both at the same temperature.
Whether less or more heat must flow to the sample depends on whether the process is exothermic or endothermic.
For example, as a solid sample melts to a liquid it will
require more heat flowing to the sample to increase its
temperature at the same rate as the reference.
This is due to the absorption of heat by the sample as it
undergoes the endothermic phase transition from
solid to liquid.
Likewise, as the sample undergoes exothermic
processes (such as crystallization) less heat is
required to raise the sample temperature.
By observing the difference in heat flow between the
sample and reference, differential scanning
calorimeters are able to measure the amount of heat
absorbed or released during such transitions.
DSC may also be used to observe more subtle physical changes, such as glass transitions. It is widely used in industrial settings as a quality control instrument due to its applicability in evaluating sample purity and for studying polymer curing.
The method of differential scanning calorimetry (DSC) allows the measurement of the energy flow to and from a sample during a temperature controlled program.
The sample is prepared in DSC capsules, which can be either sealed or open, and then submitted to a temperature program.
The actual measurement is performed against a control, which normally is an empty DSC capsule, and the energy required to keep both sample and control at the same temperature is plotted vs. the temperature. The resulting curve is called a thermogram.
DSC thermogram with definitions of onset temperature (Tonset),
peak temperature (Tpeak) and offset temperature (Toffset).
DSC provides rapid and precise determinations of transition
temperatures using minimum amounts of a sample.
Common temperature measurements include the following:
Melting
Glass Transition
Thermal Stability
Oxidation Onset
Cure Onset
Crystallization
Polymorphic Transition
Liquid Crystal
Protein Denaturation
TYPICAL DSC TRACES
Differential scanning calorimetry can be used to
measure a number of characteristic properties of
a sample.
Using this technique it is possible to observe fusion
and crystallization events as well as glass transition temperatures Tg. DSC can also be used
to study oxidation, as well as other chemical
reactions.
Glass transitions may occur as the temperature of
an amorphous solid is increased. These transitions
appear as a step in the baseline of the recorded
DSC signal. This is due to the sample undergoing a
change in heat capacity; no formal phase change
occurs.
As the temperature increases, an amorphous solid will
become less viscous. At some point the molecules may
obtain enough freedom of motion to spontaneously
arrange themselves into a crystalline form. This is known
as the crystallization temperature (Tc). This transition
from amorphous solid to crystalline solid is an exothermic
process, and results in a peak in the DSC signal.
As the temperature increases the sample eventually
reaches its melting temperature (Tm). The melting
process results in an endothermic peak in the DSC curve.
The ability to determine transition temperatures and
enthalpies makes DSC a valuable tool in producing
phase diagrams for various chemical systems.
This composite shows typical shapes for
the main transitions observed in DSC.
An alternative technique, which shares much
in common with DSC, is differential thermal
analysis (DTA).
In this technique it is the heat flow to the
sample and reference that remains the same
rather than the temperature.
Differential Thermal Analysis (DTA)
When the sample and reference are heated
identically, phase changes and other thermal
processes cause a difference in temperature
between the sample and reference.
Both DSC and DTA provide similar information.
Difference of DSC and DTA:
DSC measures the energy required to keep both the
reference and the sample at the same temperature,
whereas
DTA measures the difference in temperature
between the sample and the reference when they
are both put under the same heat.
Differential Thermal Analysis (DTA):
Measures temperature difference (dT)
between the sample and the reference
Differential Scanning Calorimetry (DSC):
Measures heat flow difference (dH/dT)
between the sample and the reference
Difference of DTA and DSC
Both techniques are concerned with the measurement of energy changes in materials.
They are the most generally applicable of all TA methods since every physical or chemical change involves a change in energy.
COMPARISON BETWEEN THERMAL ANALYSIS AND DIFFERENTIAL THERMAL ANALYSIS
Applications of DSC
• Heat capacity and heat of fusion
• Phase changes
• Glass transitions temperature
• Melting and crystallization behavior
• Heat of melting and crystallization
• Purity
• Compatibility
• Oxidative stability
• Polymorphism
• Thermal stability
• Cure / or Cure Kinetics
• With the new DSC offers the users of thermal
analysis a solution with a secure future that exceeds
today's demands
• The development of DTA and DSC instruments
made it possible to determine values faster, more
conveniently and at a lower cost
• Today's instrument is operated by a computer data
station which determines both the transition
temperature and the enthalpy of the transition.
Determining dH values involves drawing the best
base line
• The instrument must be calibrated. This is usually
done with indium since metals can often be
obtained with a higher purity than organic
materials
• However, highly purified liquid crystals have also
been used for calibration. Then an accurately
weighed sample is placed in an aluminum pan,
crimped tightly closed with a sealing press and
placed in the sample chamber of the DSC
instrument.
• Temperature calibration is carried out by running standard materials, usually very pure metals with accurately known melting points
• Energy calibration – using either known heat of fusion for metals (indium) or known heat capacities (e.g. synthetic sapphire – corrundum or aluminium oxide)
• Gas flow inside the furnace:
o To sweep away volatiles
oTo provide required atmosphere
o To assist in heat transfer
ADVANTAGES OF DTA/DSC
• Rapidity of determination – a wide
range of T (minute or hours)
• Small sample size – several mg to
several hundred mg
• Versatility – solids or liquids
• Simplicity and ease of procedure and data analysis
• Applicable to cooling and measurements under high pressure
• Ability to study many different types of chemical reactions
…..ADVANTAGES OF DTA/DSC
DISADVANTAGES OF DTA/DSC
• Relatively low accuracy and precision, 5 – 10% in most cases
• Inability to be used to determine the enthalpy of overlapping reactions
• The need for calibration over the entire temperature range
• Inaccuracies in determining peak areas due to baseline change
Tutorial 4.1
1. What is differential scanning calorimetry (DSC) ?
2. Explain the difference between DSC and DTA.
3. Propose FIVE (5) applications of DSC.
DSC curves
The result of a DSC experiment is a curve of heat flux
Vs temperature or Vs time.
This curve can be used to calculate enthalpies of
transitions. This is done by integrating the peak
corresponding to a given transition.
Figure 2: The area under the curve of a thermal event gives its enthalpy.
The area under the curve of any event in the thermogram is proportional
to the enthalpy of this event. After suitable calibration with standard
materials with known thermal events and their related enthalpy, it is
possible to utilise the integral of a thermal event and obtain very accurate
information about the energetics of thermal events.
Direct integration will give this enthalpy as energy per sample mass
(Joule per gram, J g-1). For better comparison, this value is transformed
to give energy per moles by multiplication with the molar mass. This
value is then given as kJ mol-1 and allows the comparison of enthalpies
for corresponding events detached from the actual measured sample.
The slope of the thermogram baseline gives the heat capacity of
the sample. As this characteristic will change with temperature,
the baseline normally increases with increasing temperature.
where ∆H is the enthalpy of transition, K is the
calorimetric constant, and A is the area under the curve.
The calorimetric constant will vary from instrument to
instrument, and can be determined by analyzing a well-
characterized sample with known enthalpies of transition.
∆H = KA
It can be shown that the enthalpy of transition can be
expressed using the following equation:
How to interpret DSC curve ?
Examples of DSC curve
The solid-state reaction of mullite formation is highly endothermic reaction:
DSC curves of prepared powder of undoped and doped mullite.
The DSC curves reveal that the presence of Fe in the reaction mixture affects
the drop in temperature of mullitization reaction from 1250 to 1150ºC.
It could be ascribed to the existence of Fe which can stabilize the mullite lattice.
The phase transformations are evidenced above 900ºC and between 1100 and
1300ºC.
The 1st endothermic process probably corresponds to the
formation of transitional γ-alumina phase, denoted as a spinel
phase, while the 2nd endothermic peak represents the process of
mullite formation.
Journal of Alloys and Compounds Volume 527, 25 June 2012, Pages 5–9
DSC, ESR and optical absorption studies of Cu2+ ion
doped in boro cadmium tellurite glasses
P. Gayathri Pavani , M. Prasad, V. Chandra Mouli
Differential scanning calorimetry (DSC) is used to
characterize the glasses and to determine glass transition
temperature (Tg), which is useful in suggesting structural
changes that takes place by the compositional changes. This is
because Tg is very sensitive to any change of the coordination
number of the network-forming atoms and also to the
formation of non-bridging oxygen.
Fig. 4: DSC thermogram and inset representing glass transition temperature with CdO mol.%.
Typical DSC plot of boro cadmium tellurite glass system (C3 sample) is shown in Fig. 4.
The inset of Fig. 4 shows variation of glass transition temperature with CdO mol.%.
The transformation to a glass does not take place at on, strictly defined temperature, but
within a temperature range, representing the transformation region. exo
The single endothermic glass transition peak indicates the
homogeneity of the glass. Tg represents the ‘strength’ or
‘rigidity’ of the glass structure.
The glass transition temperature increases with CdO content
which indicates that network connectivity increase. This
result is consistent with molar volume data, which shows that
the network becomes more compact on increasing CdO
concentration.