DR-7000D Semi-automatic Chemistry Analyzer DR-7000D
Semi-automatic Chemistry Analyzer
Instruction:Thanks for purchase DR-7000D Semi-automatic
Chemistry Analyzer.
In order to use the product correctly, please read the user
manual carefully before operation.
Please keep the user manual safely for your anytime
reference.
Caution:Colorimetric cuvette should be marinated by the alcohol
at least 60 minutes after self-checking for the first time working,
and then use distilled water to rinse it time by time for 10
minutes. If not, the test result might be inaccurate.
Content1Chapter 1 Introduction
11.1 About DR-7000D
11.2 Technical specification
21.3 Warning Mark
3Chapter 2 The Basic Principle and Functional Feature
32.1 Basic principle
42.2 Functions
5Chapter 3 Installation and Application
53.1 Instrument Installation
63.2 Main parts and Application
7Chapter 4 Basic Operation
74.1 Electrification
84.2 Sample testing
124.3 Program measurement
204.4 Quality control
224.5 Setup
244.6 Record search
254.7 Rinse
26Chapter 5 Instrument Maintenance
265.1 Instrument surface cleaning
265.2 Colorimetric cuvette cleaning
265.3 Lamp-house replacement
275.4 Fuse replacement
275.5 Pump pipette replacement
285.6 Print paper replacement
285.7 Printer crankshaft maintenance
29Chapter 6 Transportation and Storage
296.1 Transportation requirements
296.2 Storage requirements
30Appendix A
31Appendix B
33Appendix C
37Appendix D
39Appendix E
Chapter 1 Introduction
1.1 About DR-7000DDR-7000D was designed in response to the needs
of clinical tests. It is a newly intelligentized filter chemistry
analyzer.The built-in dot matrix LCD display provided with
apheliotropic board ensures the clear show of all test results,
functional status and operational instruction. The English
indication system guides users operation, and improves man-machine
communication. 1.2 Technical specification
a) Lamp-house: 20W /12v Halogen tungsten lampb) Wavelength:
340405450492505546578630 nm optionalc) ABS: -0.3~3.0 Absd)
Wavelength accuracy: 1nm
e) Spectrum bandwidth6~12nmf ) Resolution: 0.1m Absg)
Linearity2Abs2%1Abs0.5%
h) Stability 20 minutes for warm-up and then 2mAbs/h
i ) Blank: Auto zero upon the colorimetric cuvette with full of
water
j ) Interface : RS-232 serial portparallel printer port k)
Memory
Quality control data (valid period: 1 month) 200 testing
programs, 3000 testing results
l) Test methods
End-point Kinetic
Two-point kinetic Absorbance Multipoint
m) Measurement time: 0~999 seconds programmable
n) Delay time: 0~999 seconds programmable
o) FlowcellQuartz flowcell, measuring volume is 32lp)
Temperature control
Semiconductor device control , 253037 and room temperature
optionalAccuracy 0.1
q) Sampling system
Peristaltic pump, controlled by step motorSampling volume:
100l~9999l programmable1l/step sampling volume programmable
r) Power supply: 100~240VAC 50/60Hzs) Environment condition:
Ambient Temperature15~30, optimal temperature: 15~25Relative
Humidity85%RHt) Dimension462mm418mm232mm
u) Weight: 6.5kg1.3 Warning Mark
There are two biohazard marks on the equipment, one is Biohazard
mark, another is Force electricity mark, they are all on the
backboard of the equipment. As shown in Figure 1-1.
Warning mark
Figure 1-1 Warning Mark There might be a potential infectivity
for the sample & waste reagent, please make sure to take
protection action during the testing, cleaning process and
instrument maintenance. Samples and waste reagent should be
disposed in a suitable package conforming to the local regulation
for laboratory.Chapter 2 The Basic Principle and Functional
Feature
2.1 Basic principle
The principle of the instrument is based on the phenomenon of
different wave band absorbance from substance, which is in line
with Lambert-Bill Law.
A=log(I/I)=log(1/T)=bc T= I /IT Transparency c Solution layer
thickness(cm)I Transmitted light intensity b Solution Moore
concentration(mmol/ml)I Incident light intensity Solution Moore
absorbance coefficient(mlmmol-1cm-1)A Absorbance2.1.1 Optical
system
The prismatic part of DR-7000D Semi-automatic Chemistry analyzer
adopts the filter in unique research and manufacture techniques. It
is featured by moist-resistant, heat-resistant, longevity of
service, little central wavelength excursion, high optical
signal-noise ratio etc.. The principle diagram is shown as Figure
2-1.
Amplifier
Lamp-house Prismatic Flowcell Detector
System Figure 2-1 Principle diagram of optical system 2.1.2
Electronic system Flowcell Detector Amplifier Monochromator
Driver control Computer system Temperature control
VG
Lamp-house VI Display/Keyboard Printer
Vn
Figure 2-2 Electronic principle diagramThe principle diagram of
the electronic system is shown above in Figure 2-2. When passing
through the monochromator the ray of the lamp-house will be changed
into monochromic light. The result will be then received by the
photoelectric detector when it passes through the flow colorimetric
cuvette, and further transformed into electric signal. The signal
will be amplified and filtered, and then transmitted into A/D
transformer. When the information is transformed into numbers, it
will be sent into data collection system of the computer for kinds
of calculation and transaction. Users can either read the final
result on the display or alternatively get it printed out.2.2
Functions
1.Measurement methodsa) End-point b) Two-point kineticc) Kinetic
d) Absorbancee) Multipoint2. Internal and external printer is
available for uses options.
3. Sampling volume, test method is programmable.
4. Users can determine whether to execute real time print out of
the test results.
5. If any operational mistakes or other false results are found
during the process of measurement, users can choose the repeat key
() to retest, the original serial number of the test remains.6.
Rinsing program is available between two sample tests to avoid
cross contamination.
7. Up to 1 month quality control tracking, and the printing
results provided.
8. Integrated test report printing is available.9. Reaction
curve display:Reaction curve for sample and standard liquid will be
displayed on the screen during the testing, as shown in Figure 2-3.
ABS LIMIT
Absorbance Absorbance ABS LIMIT 0 0
t1 time t1 time
(a)Positive reaction curve bNegative reaction curve
Figure 2-3 Activity reaction curveNote: 1. Press +/- to check
the reaction curve after testing the standard liquid or sample.
2. Press Exit to test next sample.
Chapter 3 Installation and Application
3.1 Instrument Installation
3.1.1 Installation A. Move all the spare parts out after opening
the package and then check the spare parts quantity according to
the packing list, if there are some spare parts broken please
contact the delivery forwarder or manufacturer.
B. Place the equipment on the selected work bench where is no
vibration.
C. The equipment should not be placed close to the chemicals,
caustic gas and electromagnet.
D. Prevent direct sunshine, humidity and high temperature.E.
Operating temperature range is 1530 (optimal range: 1525),
relativity humidity is less than 85%RH.F. The power supply of the
instrument is 220V/50Hz and the plug is on the back-board, see
Figure 3-1.
The instrument is provided with special two cores electric wire,
which should be connected with well grounded receptacle after
plugging in.
1. Power switch 2. Power socket 3. Fan 4. Fuse socket
5. Warning mark 6. Serial port 7. Parallel Port 8. Biohazard
mark 9. Waste pipette Figure 3-1 Back-board outline3.1.2 Cautions
for use Read the user manual carefully before working , especially
for the caution & hazard info. Pre-heat for 20 minutes before
working and use distilled water to wash the equipment before any
test. Wash over 3~5 times immediately after each test. After finish
working, wash time after time before turning off the
instrument.
Make sure to maintain some cleaning agant in the colorimetric
cuvette when it stops working to avoid the sample remain in the
colorimetric cuvette.
The colorimetric cuvette should be washed and dipped everyday,
firstly use cleaning agent or distilled water to wash it for
several times, and then remain some cleaning agant in it.
When doing sample feeding operation, please pay attention to the
sucking pipette and waste pipette, to avoid damaging and
blocking.
Prevent vibration and direct sunshine and keep way from the wall
at least 20cm.
3.2 Main parts and Application A. Display The display of
DR-7000D Chemistry analyzer is dot-matrix large angle of view LCD,
which adopts the latest STN technology. Its dimension is 108x58mm
with dot-matrix 240x128. In addition, the apheliotropic board
ensures clear reading in a poor light available.
BKeyboardThe keyboard comprises 19 key-presses, i.e. 12 number
keys (including and radix point), 5 functional keys and EXIT and
ENTER keys.
Number key Entering into submenu or numeric parameter input.
ENTER Press to confirm the executive command.
EXIT Press to return to the former menu.
F1 F5 (Functional key) Press to execute the relevant functions
indicated in the five small squares at the bottom of the
display.
C. Buzzer1. As the key-press on the control board is made of
pellicle without clear tactility, buzzer reproducer is used to help
inform operator as soon as the key-press is touched.
2. The buzzer will remind the user to move the sample away once
the equipment finishes sampling.
D. Sample feeding switch and operationThe white sucking pipette
is on the left of the instrument, and the switch of sample feeding
is under it. Sample feeding operation: During the measurement, user
can follow the instructions on the screen to do this operation,
first put the sucking pipette into the bottom of prepared samples
(including distilled water, reagent blank, standard liquid, sample
blank, sample, quality-control sample), press the switch button and
then the samples will be sucked automatically. Dont move the sample
until 2-3 seconds later after the buzzer shows sample feeding is
finished.
E. PrinterBoth of internal printer and external printer are
available for the instrument. The user can select one of them for
operation. There are 3 status optional: ON, OFF and AUTO. Real-time
printing is available once select AUTO status.
Refer to 5.6 for the replacement of thermal sensitive print
paper. Chapter 4 Basic Operation4.1 Electrification The power
switch of the instrument is on the back-board (see Figure 3-1).
Turn on the power switch to get the instrument in the state of
electrification, and then the info in Figure 4-1-1 and 4-1-2 will
be displayed on the screen orderly.
Figure 4-1-1
Figure 4-1-2Note: Under the Self checking interface(as shown in
Figure 4-1-2), press the sucking button to enter into the submenu
as shown in Figure 4-1-3.
Figure 4-1-3
Figure 4-1-4Reminder info (see Figure 4-1-5) will be shown on
the screen after self checking. The user also could press Exit to
omit it to enter into the test interface directly (To avoid
inaccurate test result, please dont omit this process).
4-1-3Figure 4-1-5Key in the final selection and then enter into
the program main menu, as shown in Figure 4-1-6.Note: If select
Yes, all the test results will be deleted except for QC results and
the sample serial number will be reordered from 1 automatically.If
select No, the system will back to the main menu, as shown in
Figure 4-1-61. Sample testing 2. Parameter setup3. Quality
control4. Setup
5 .Record search
6. RinseFigure 4-1-6There are 6 submenus under the main
menuPress 1 to enter Sample testing mode to do normal testPress 2
to enter Parameter setup mode to edit and modify relevant test
parameters.
Press 3 to enter Quality control mode to edit the control
parameters or look over control statistics.Press 4 to enter Setup
mode to setup the language, printer status, time, date, ID and
lamp-house luminosity.Press 5 to enter Record search mode for the
search of relevant record.Press 6 to do the rinsing operation.4.2
Sample testing Under the main menu, press 1 to enter item display
submenu. As shown in Figure 4-2-1. There are 10 sorts for the test
items: LIVER, KIDNEY, CARDIAC ENZYME, BLOOD, BLOOD LIPID,
HYDRONIUMGROUP 1, GROUP 2, GROUP 3, GROUP 4; There are 20 items
under each sort and 200 items totally. Funtion key F4 and F5 are
used for page up and page down.
The skip point at the first item after enter into the test item
interface, there are 2 ways to select the item.Way 1: Input the
item number (such as 001).Way 2: Press the funtion key F1 and
F2.
Figure 4-2-1NoteIf inaccurate item number is input, it will be
corrected by key the accurate one under the same interface directly
and the original one will be covered automatically, the item no.
should be 3 digits.
Input the selected Item Number. (such as: 003 or press F2 to
select 003) and then press Enter to enter into the test status. As
shown in Figure 4-2-2.
Rinse Print Calibrate Zero QC Figure 4-2-2Press the sucking
button to suck the distilled water and then the following info (as
shown in Figure 4-2-3) will be shown on the screen. Please follow
the screen reminder info to do the test operation one by one.
Rinse Print Calibrate Zero QC
Figure 4-2-3Note:
a) Press sucking button to make the distilled water, reagent,
sample, standard liquid and quality control liquid are all sucked
into the colorimetric pool by sucking pipette according to the
reminder info one by one, and then test results and ABS value will
be shown on the screen after finishing measurement. If printer
status is ON, press F2 to print the test results out and press 8 to
run the print paper; If printer status is AUTO, the printer will
execute real time test result print-out. Test results will be saved
automatically, if the test results exceed 3000 pieces, the reminder
info THERE IS NO SPACE FOR DATA BASE, CLEAR ALL SAMPLE DATA?, press
EXIT to out and press ENTER to clear it. The equipment will not be
workable unless do the clear operation.b) One sample serial number
is only correspond with one patient, it will be added 1 after each
testing automatically. Further more the operator also could press
ENTER button to enter into sample edit submenu, as shown in Figure
4-2-4, and then input a new sample serial number. (within
001-999).
Figure 4-2-4Press ENTER after inputing the sample number and it
will be the serial number for the next testing sample.
c) If ID STATUS is ON, there are 2 ways to input the ID number,
one is by barcode reader, another is by press number 9 button to
enter into the ID edit interface (as shown in figure 4-2-5), and
then input the ID number (0~9) by hand (the ID number should not
more than 10 digits), otherwise, there will be no ID number.
Note:(1) There is only one ID correspond with one daily sample
number. One time ID number input is avaiable for the samples which
use the same sample serial number.(2) There is no ID for new
sample, repeat ID input is avaiable and the system will keep the
final one as the ID number.d) Repeat test could be done after
pressing the repeat key and the original sample no. will not be
changed.
e) If quality control or re-calibration is required, press F5 or
F3 and then Pump standard liquid again? or Pump control liquid
again? info will be shown on the screen fFor the end point
measurement, reminder info Test the reagent blank again ? will be
shown on the screen after setting parameters again, user could
select the proper info, if select 1 the reagent blank value will be
tested again or it will be calculated according to the original
value by the equipment. As shown in Figure 4-2-6.
Enter into standard liquid test interface after finishing
reagent blank test and the reminder info Test standard liquid
again? will be displayed on the screen, as shown in Figure
4-2-7.
g) For the END point, two poin and Kinetic measurement, if the
test time is not 0, the reaction curve will be displayed on the
screen in real time. After displaying the test results, press
button the enlarge reaction curve will be displayed on the
screen.4.3 Program measurement Under the main menu (as shown in
Figure 4-1-6), press 2 to enter test program submenu, as shown in
Figure 4-3-1.
Figure 4-3-14.3.1 Parameter setupFor example, input 003 or press
F2 () 2 times to select item TP and then press Enter to enter into
its submenu. As shown in Figure 4-3-2.
Figure 4-3-2The content in the bracket is the item instruction
of program (Reminder info.) P1, P2, P3 P4 on the top of the screen
means the page number, Press F4, F5 to do page up and page down
operation, press 1-6 to edit and modify the content of relevant
items.There are totally 5 pages of program submenus, P1 in Figure
4-3-2 means Page 1, press F5 to view Page 2 (see Figure 4-3-3),
Page 3 (See Figure 4-3-4) and Page 4 (See Figure 4-3-5). Figure
4-3-3
Figure 4-3-4If select the Multipoint method, concentration
parameter under P4 should be input, however, it is no need for
other measurement methods.
Figure 4-3-54.3.2 Parameters modificationA) Modifying test
nameUnder the program submenu ( as shown in Figure 4-3-2), press 1
and then press ENTER to enter into Program submenu, as shown in
Figure 4-3-6:
Figure 4-3-6As known from above, corresponds to F1, corresponds
to F3, Delete corresponds to F4, PA( Page down) corresponds to
F5.
For example, to change the name into GPT: Press F3 (or F1) to
move the cursor to the line where there is G, then press 6 (G is
correspond to 6), G will be displayed in the Name Column. Press F3
(or F1) to move the cursor to the line where there is P, then press
5 (P is correspond to 5) GP will be displayed in the Name Column,
and so forth. When all finished, press ENTER to return to the last
submenu (if the input is more than 8 letters, it will return to
last submenu automatically). If any letter is entered incorrectly,
press F4 to delete the letter. For lowercase entering, press F5 to
look up lowercase on the next page.
BModifying Temperature Press 2 under the submenu on the first
page (as shown in Figure 4-3-2), then press ENTER, the screen is
shown as Figure 4-3-7
Key in the number in front of the selected temperature, then
press ENTER to return to the submenu.
CModifying measure methodPress 3 under the submenu on the first
page, then press ENTER, the screen is shown as Figure 4-3-8.
Figure 4-3-8Select the proper number and then press ENTER to
back the upper menu.
DModifying unitsPress 4 under the submenu on the first page,
then press ENTER. As shown in Figure 4-3-9
F5 is for page down ,as shown in Figure 4-3-10
Figure 4-3-10Key in the number in front of the selected unit,
then press ENTER to return to the submenu (as shown in Figure
4-3-2).
E) Modifying linearity range: Press 5 under the submenu on page
1, then press Enter As shown in Figure 4-3-114-3-12
Figure 4-3-11
Figure 4-3-12Input the proper parameter and then press Enter to
back to the submenu (as shown in Figure 4-3-2).
Note:Input the linearity range according to the specification on
the reagent kits, if the parameters exceed the range during test,
reminder mark ! will be printed on the test report.! means Test the
sample again after dilution and the result should be multiplied the
dilution multiple.FModifying Reagent BlankPress 1 under the program
menu on the second page as shown in Figure 4-3-3 , then press
ENTER, as shown in 4-3-13
Figure 4-3-13Press 0 or 1 for R-blank Yes or No, then press
ENTER to return to the submenu ( as shown in Figure
4-3-3).GModifying sample blank Press 2 under the program menu on
the second page (as shown in Figure 4-3-3), then press ENTER, as
shown in Figure 4-3-14:
Figure 4-3-14HModifying standard liquid concentration: Press 3
under the submenu on the second page (as shown in Figure 4-3-3),
then press ENTER, as shown in Figure 4-3-15
Figure 4-3-15Input the proper standards liquid concentration,
then press Enter to back to the submenu.
Note: The factor will be turned into ---- automatically after
entering the standard liquid concentration.The factor will be saved
automatically and reminder info Confirm the standard liquid again?
will be shown on the screen once the user opens the submenu again,
if select yes, the old coefficient will be replaced by the new
one.I) Modifying FactorPress 4 under the program menu on page 2 (as
shown in Figure 4-3-3), then press Enter. The factor is 00000.
input proper value, then press Enter to return to the submenu. As
shown in Figure 4-3-16.
Figure 4-3-16NoteThe standard concentration will turn into -----
after entering the factor value automatically.JModifying the
sampling volumePress 5 under the program menu on Page2 (as shown in
Figure 4-3-3), then press ENTER. Input the proper number, then
press ENTER to return to the submenu. As shown in Figure
4-3-17.
Figure 4-3-17Note: the min. sample volume is 100l, however, in
order to reduce the cross-contamination of samples with high
concentration, it is better to adjust it to 500l at
least.KModifying the decimal digits of test results: Press 6 under
the program menu on page 2 (as shown in Figure 4-3-3), then press
Enter to input the required decimal digits, then press Enter to
turn to the submenu. As shown in figure 4-3-18. (Decimal digits is
the numbers after the radix point of the test result, the max. one
is 4 digits such as 1.2334).
Figure 4-3-18L) Modifying wavelengthPress 1 under the program
menu on Page 3 (as shown in Figure 4-3-4), then press ENTER. Input
the selected wavelength number, if correct, press ENTER to back to
the submenu (sub. wavelength press 2), as shown in Figure
4-3-19.
Figure 4-3-19Note: There will be a reminder info Input
wavelength again on the screen, once main wavelength value input is
unavailable and the operator should input the wavelength again. M)
Modifying delay timePress 3 under the main menu on page 3 (as shown
in Figure 4-3-4), then press ENTER. Input the proper time, if
correct, press ENTER to back to submenu. (Delay time setup should
be according to the specification on the reagent kits), as shown in
Figure 4-3-20.
N) Modifying test time Press 4 under the main program menu on
page 3 (as shown in figure 4-3-4), then press ENTER, input proper
number, if it is correct, press ENTER to back to the submenu. As
shown in Figure 4-3-21.
Note: Measure time is only applied when using two-point kinetic
and kinetic mode.
O) Modifying reference highUnder the main program press 6 on
page 3 (as shown in figure 4-3-4), then press ENTER. Input the
proper number , if it is correct, press ENTER to back to the
submenuas shown in Figure 4-3-22.
Figure 4-3-22P) Modifying reference lowUnder the main program
press 5 on page 3 (as shown in figure 4-3-4, then press ENTER.
Input the proper number , if it is correct, press ENTER to back to
the submenuas shown in Figure 4-3-23
Figure 4-3-23Note: All parameters above should be selected
according to the specification on the reagent kit. (If test result
is not within the ref value range, & reminder info will be
marked behind the result.) 4.4 Quality control Under the program
main menu (as shown in figure 4-1-6), press 3 to enter into the
Quality Control submenu , as shown in Figure 4-4-1:
Figure 4-4-14.4.1 Program quality control
Press 1 under the quality control submenu, then press ENTER to
enter into the submenu, as shown in Figure 4-4-2:
1-1
Figure 4-4-2Input the selected Item number then press ENTER to
enter into the submenu, as shown in Figure 4-4-3, then input the
proper number, if it is correct, press ENTER, as shown in Figure
4-4-3.
Input the proper number of control lot, target value and
standard deviation one by one. If they are correct, press ENTER to
back to main menu. Two types of control liquid are available for
the program. 4.4.2 Modifying control statistics Under the program
main menu, press 2 to enter into the Control Statistic submenu, as
shown in Figure 4-4-4.
Press 1 ----- Present control data will be shown on the
screen.
Press 2 -----Individual control data will be shown on the
screen.Press F2 or to select the edited QC item and then press
ENTER to enter into the submenu, as shown in Figure 4-4-5.
Press or to view the control data within 1 month. 4.4.3
Statistic data clearance Press 3 under the program main menu (as
shown in Figure 4-4-1), then press ENTER to enter into the submenu,
as shown in Figure 4-4-6:
Figure 4-4-64.4.3.1 All control data clearance
Press 1 under the main menue (as shown in Figure 4-4-6) to enter
into the submenu, as shown in Figure 4-4-7.
Figure 4-4-7Press ENTER to clear all the statistic data and then
back to upper menu, as shown in Figure 4-4-6.4.4.3.2 Individual
control data clearance
Press 2 under the main menue (as shown in Figure 4-4-6) to enter
into the individual control data submenu as shown in Figure
4-4-8.
Figure 4-4-8Press Enter to enter into the submenu, key in the
selected number and then press ENTER to clear it.4.5 Setup
Under the program main menu, press 4 to enter into the submenu,
as shown in Figure 4-5-1.
4.5.1 PrinterUnder the setup menu (as shown in figure 4-5-1),
press 2 to enter into the printer submenu, as shown in Figure
4-5-2:
Press 1 to select the status of internal printer, then press
ENTER to back to the submenu.
Press 2 to select the status of external printer, then press
ENTER to back to the submenu.Press 3 to select the run paper
distance (per time) and the printed test result line-distance,
there are 5 values optional: 2, 4, 6, 8, 10 line.4.5.2 Date &
TimePress 3 under the setup menu (as shown in Figure 4-5-1) and
then press ENTER to enter into the submenu as shown in Figure
4-5-3:
Input the proper number, then press ENTER to back to the
submenu.4.5.3 Lamp- house replacement Press 4 under the setup menu
(as shown in Figure 4-5-1) to enter into the lamp-house submenu, as
shown in Figure 4-5-4:
Program indicates pump, and set wavelength 340nm automatically,
then read the number of the luminosity continually. Press EXIT to
back to last menu.
4.5.4 ID setup Press 5 under the Setup menu (As shown in Figure
4-5-1) to set up the ID and then press ENTER TO back the upper
menu.
4.6 Record search
Under the program main menu (as shown in Figure 4-1-6), press 5
to enter into the record search submenu, as shown in Figure
4-6-1:
Figure 4-6-14.6.1 Record search
Under the RECORD SEARCH submenu( as shown in Figure 4-6-1),
search function is avaiable once press any key within 1 to 4, such
as press 1 ,the following interface will be shown on the
screen.
Single Print Del Prior Next
Figure 4-6-2Press F1 to select the printing report format
(single or integrate).Press F2 to print the test report.Press F3 to
delete the current note.Press F4 to display the previous
record.Press F5 to display the next record.Press EXIT to back to
the upper menu (as shown in Figure 4-6-1).4.6.2 Record
clearanceUnder the record clearance menu (see Figure 4-6-1), press
5 to enter into the record clearance submenu, as shown in Figure
4-6-3:
Figure 4-5-6Figure 4-6-3Press 1 to clear all the records and
press EXIT to back to upper menu as shown in figure 4-6-1.4.7
Rinse
Under the program main menu (as shown in Figure4-1-6), press 6
to enter into RINSE submenu, as shown in Figure 4-7-1.
Figure 4-7-1Input the rinse time by hand and then press ENTER.
The equiment will do the rinse operation according to the selected
time automatically. (Rinse time scope 1~99)Chapter 5 Instrument
Maintenance5.1 Instrument surface cleaningUse the wet cloth and
cleaning agent to clean the instrument. To avoid damage the DOS
shell and the liquid flow into the equipment, prohibit using any
corrosive impregnants.5.2 Colorimetric cuvette cleaning(1) The
colorimetric cuvette should be washed by the cleaning agent or
distilled water between each test.(2) After finish working
everyday, wash several times by the cleaning agent to rinse the
remains in the colorimetric cuvette.
(3) To avoid serious contamination, adjust the temperature to 37
to accelerate the cleaning speed, first in cleaning agent, then in
distilled water to wash it several times.
(4) Make sure to maintain some distilled water in the
colorimetric cuvette when the equipment stops working everyday. 5.3
Lamp-house replacement (1) Turn off power firstly when replacing
Lamp-house.
(2) Push the sucking pipette out of the instrument smoothly, and
then take down the screw from the bottom of the instrument. Lift
the DOS shell smoothly from the back shell (do not take it off
rapidly), leaving the front on the table. Pull out the sucking
pipette from inside completely before DOS shell is lifted up.
(3) Take off the lampshade and pressure pad firstand then take
down the illuminant. (Warning: light bulb and ambient temperature
may be very high, to avoid dangers, please take gloves to do this
operation).
(4) Insert the new bulb (12V, 20W) in the lamp stand without
touching the top of the light bulb, and then screw the pressure
pad. To avoid effecting the test results, please do not touch the
bulb by hand directly. Use alcohol cotton to clean it. (5) Under
such circumstance, switch on the power to enter into setup menu
,then press 4 to enter into the lamp house replacement submenu,
according to the reminder info to pump the distilled water and then
the Luminosity value will be shown on the screen.
(6) Adjust the position of lamp-house until the figure on the
bottom of the display turned into maximum (Reference range
18000~50000).
(7) Ensure the figure on the bottom of the display unchanged,
and tighten all screws.
(8) Turn off the power, and put on the lampshade. Thus the
illuminant replacement is finished.
(9) Insert the pipette properly. Put DOS shell on and tighten
all screws.
Figure 5-3-1
5.4 Fuse replacement The fuse socket is on the back board of the
instrument, below the power switch and on the top of the entrance
of electric wire. As shown in Figure 5-4-1. Pull out the electric
wire, pick off the fuse socket to replace the fuse offered by
manufacturer.
Fuse socket Fuse Figure 5-4-15.5 Pump pipette replacement Open
the DOS shell, the method is same to the lamp-house replacement
(see step1&2). Press the stay bar to open the block board under
the pump, move the pump pipette out and then replace a new one.
To avoid leaking, tighten the screw to tense the pipette
properly.
Pump pipette:
Diameter 2.06mminner
Length90mm
Stay bar Block board pump bearingFigure 5-5-1
5.6 Print paper replacement SPRM printer is provided with paper
scroll before leaving factory, but it is not been inserted. This is
for the safety of the printer during the extended transportation
and storage. Insert the paper scroll before use, refer to the
following steps:
(1) Press the cover below the paper exit gently, the cover will
flick automatically and be open.(2) Cut the paper as Figure
5-6-1.
(3) Switch on the printer power, insert the paper into the paper
entrance, it will enter into the crankshaft automatically.
Paper Entry
Stay bar
5.7 Printer crankshaft maintenanceIn order to ensure the normal
work of the printer, please do not take the crankshaft down
discretionarily.
(1) When replacing the paper, check whether any wastepaper
maintains on the crankshaft, blow off it if there is any. (2)
Ensure the clean of external board.(3) When thermal sensitive
printer is dirty (print results is not clear), clean it with
alcohol cotton.(4) Do not pull the paper emphatically before the
crankshaft is put up.Chapter 6 Transportation and Storage
6.1 Transportation requirements 1. Anti-moisture and anti-water
works should be done well before transportation.
2. Avoid strong shake and extrusion.3. Take care to move, pack
or unpack the instrument.
6.2 Storage requirements1. Indoor
2. Temperature: -40~ +50 3. In a well-ventilated, clean and dry
area, away from chemicals and caustic gas.Cautions:The instrument
should be rinsed several times and the sucking pipette should be
kept empty & dry before transportation or storage. If the
transportation ambient temperature is lower than 0, it is better to
use 95% alcohol to rinse the pipette in order to avoid frost crack.
Appendix A Manufacturers warrantyDear CustomerThanks for purchasing
DR-7000D semi-automatic chemistry analyzer, our company will
provide the following services for you: 1. Technical consultations
are provided at any time. 2. Maintenance free of charge within one
year from the day you purchase the instrument. 3. Maintenance will
be charged in the following conditions.(1) Product which has
expired the warrantee date.
(2) Damage caused by accidental factors or improper use.
(3) Damage caused by the operation that is not according to the
user manual.
(4) Damage caused by your own repair without our companys
permission. 4. We will provide updated service for free according
to the technology improvement.Any technological service requests,
please contact us according to the following address:
DIRUI INDUSTRIAL CO., LTD.95 Yunhe street, New & High Tech.
Development Zone
Changchun, JiLin 130012 P.R. China
Tel: 86-431-85100409
Fax: 86-431-85172581E-mail:[email protected]
http://www.dirui.com.cn
Appendix B Trouble shootingFaultReasonTrouble shooting
1. The fan and W lamp do not work after switching on the
power.The fuse is broken.Replace the fuse (2A).
2. After switching on the power, the fan works but W lamp doesnt
work.W lamp is broken or without voltage.
First check whether the voltage for lamp house is 11.0V. If the
voltage is as normal, then W lamp is broken, replace a new one.
3. Tested result is not good.Sample feeding generates air
bubble, pump delivery and speed is instable.After several rinsing
or sample feeding, if it still doesnt work properly, wash it with
specific cleaning agent for 2-3 times, then with distilled water
for 5-10 times. Re-input the pump parameters.
4. The printer doesnt work and display E1, E2, E3, E4 etc
info.Printer plug falls out or is loose connection.Open the DOS
shell, reinsert plug and check if the light for printer board
flashes.
5. The pump does not run after switching on the sample feeding
power.Switch damaged or loose connection.Open the DOS shell,
reinsert plug or replace the switch.
6. Sampling abnormity after system self-checking.Interface from
the outside electric net.Turn off the power and restart the
equipment after several seconds.
7. Temperature abnormity
after system self-checking.Temperature sensor socket is not
connected well or the temperature exceed the scope of 5~40.Turn off
the power in normal room temperature, restart it after 5 minutes.If
it is not workable, open the equipments shell and refix the sensor
socket.
8. The buzzer doesnt work.Computer shuts down, which is usually
caused by external interference.Turn off the power, and restart a
few seconds later.
9. some a key does not work.Key cover is used too much.Replace
the key cover.
10. Display with variegation after switching on.It is usually
caused by external interference.Turn off the power, and restart a
few seconds later. Or open case, reinsert all cables
connection.
11. The pump works without sample feeding.
Flow system is blocked or pipette aging or Blocked.Check that if
the input and output pipette are twisted or extruded, and check if
pump pipette is tightened or aging. To replace with a new one or
cut it shorter. And clean and impregnate flowcell.
12. Instrumet alarm.Re-operate based on manual. If it happens
often, please contact the supplier.
Appendix C Communication Interface with ComputerBaud
rate:9600
Data bit: 8 bitsStop bit: 1 bit
Verification: none
Hardware hold hands: noneStart character: 02H
2 bytes space: 20H
Line feed character:0DH0AH
Ending character: 03H
Connection between chemistry analyzer and computer
Test report data output format
1 Data output formatindividual data
123456789101112131415161718192021222324252627282930
STXCRLF
SPSPSPTestingSPReportCRLF
CRLF
SPDate:--SP::CRLF
SPSampleSPNO.:CRLF
SPID:CRLF
SPTestSPName:CRLF
SPREF:SPCRLF
SPSeq.SPSPResultSPSPUnitSPSPHintCRLF
SPSPSPSPCRLF
CRLF
ETX
Note
SP------- space character (20h hex)
X--------- virtual value bitSTX----- start character (02h
hex)
CR------- return character (0dh hex)
LF-------- line feed character (0ah hex)
ETX------ ending character (03h hex)2 Data input
formatIntegrated
data123456789101112131415161718192021222324252627282930
STXCRLF
SPSPQualitySPControlSPReportCRLF
CRLF
SPDate:--SP:CRLF
SPTestSPName:CRLF
SPID:CRLF
SPStandardSPX:SPCRLF
SPStandardSPSD:SPCRLF
SPResult:CRLF
ETX
3Data output formatQC data
123456789101112131415161718192021222324252627282930
STXCRLF
SPSPQualitySPControlSPReportCRLF
CRLF
SPDate:--SP:CRLF
SPTestSPName:CRLF
SPID:CRLF
SPStandardSPX:SPCRLF
SPStandardSPSD:SPCRLF
SPResult:CRLF
ETX
Appendix D Regular Chemistry Reagent ListParameterCode
No.ProductMethodologyWavelengthTemperatureStandard
concentrationFactorUnitMeasurement time(S)Lag time
(S)Reagent
Liver
Function01ALTKinetic340nm37-3376U/L6060double reagent
02GGTKinetic405nm371158U/L6060double reagent
03ALPKinetic405 nm373378U/L6060double reagent
04CHEKinetic405 nm378369U/L6060double reagent
05T-BILEnd point546 nm3789.6umol/Ldouble reagent
06D-BILEnd point546 nm3726umol/Ldouble reagent
07ALBEnd pint630 nm3740g/Lsimple reagent
08TPEnd pint546 nm3780g/Lsimple reagent
09NH32-point kinetic340 nm37200umol/Lsimple reagent
10ASTKinetic340 nm37-3376U/L6060double reagent
11LAPKinetic405 nm372121U/L6060double reagent
12ICDHKinetic340 nm373376U/L60120double reagent
13ADAKinetic340 nm374921U/L60180double reagent
Kidney
Function14BUN2-point kinetic340 nm377.14mmol/L6030double
reagent
15Cr2-point kinetic505 nm37400umol/L6030double reagent
16UAEnd point505 nmroom temperature7.2mmol/Ldouble reagent
Hydronium17CaEnd pint578 nmroom temperature2.5mmol/Ldouble
reagent
18CaEnd pint630 nmroom temperature2.5mmol/Lsimple reagent
19CIEnd point505 nmroom temperature100mmol/Lsimple reagent
20MgEnd point546 nmroom temperature1mmol/Lsimple reagent
21FeEnd point546 nmroom temperature90umol/Ldouble reagent
22PEnd point340 nmroom temperature1.29mmol/Lsimple reagent
Glucose
metabolism23GLUEnd point505 nm375.55mmol/Ldouble reagent
24GLUEnd point340 nm375.55mmol/Ldouble reagent
Code No.ProductMethodologyWavelengthTemperatureStandard
concentrationFactorUnitMeasurement time(S)Lag time
(S)Reagent
25FTSEnd point546 nm371.95mmol/Lsimple reagent
Blood
lipoid26CHOEnd point505 nm377.0mmol/Ldouble reagent
27HDL-CEnd point546 nm3753mmol/Ldouble reagent
28LDL-CEnd point546 nm3777mmol/Ldouble reagent
29TGEnd point546 nm372.2mmol/Ldouble reagent
30LDH-LKinetic340 nm379683U/L6060double reagent
31CKKinetic340 nm378095U/L60120double reagent
32CK-MBKinetic340 nm378254U/L60120double reagent
33HBDHKinetic340 nm374920U/L6060double reagent
Acute
abdomen34AMYKinetic405 nm375138U/L6060double reagent
35AMYKinetic405 nm375138U/L6060double reagent
Note:
1. The above info just for regular items, for the details,
please check the specification on the reagent kits.
2. User also could program the new parameters according to the
specification on the reagent kits.
Appendix E Bar code reader installation and operation
1. Installation
The bar code reader is the optional spare parts for DR-7000
series semi-automatic chemistry analyzer, there is a serial port on
analyzer back board for it installation. The operation is very
easy, operator only need to insert the bar code reader connector
plug into the proper serial port.
Notice: The equipment should be powered off before bar code
reader installation.
2. Operation
(1) The bar code reader could read Code-39, Code-128, EAN-8,
UPC-E and Inter leaved 2 of 5 bar code automatically and the total
digit for the bar code number should not more than 10 letters,
otherwise, it will not workable.
(2) The operator could do the scan operation before each test
(ID should be setup ON) and then the bar code number will be saved
into ID number. There is only one bar code number for each sample,
therefore, it is only need to scan one time for the same sample or
the equipment will save the last scan one as the final one.
(3)The bar code reader belongs to far-distance Reader, the
operator only need to aim the scan point at the barcode number and
then press the trigger, so that, the real-time bar code info will
be shown on the screen.
(4)The bar code reader reads the barcode by the light line
center on the top of it, to avoid mistakes, the light line should
cover the whole barcode and keep proper bevel with the trigger
before reading.
Connection of stylus printer
Instrument could connect EPSON LQ300K+ stylus printer. (Other
type stylus printer can be used as well, but it may cause some
change in print speed and character position.)Rev: 06/2009Warning
mark
Biohazard mark
Biohazard mark
A/D
Semi-Automatic
Chemistry Analyzer
Version x.x x
Feb 6 2007
050715
191311
Self checking
Please pump distilled water
050715
191320
Self checking
Please waiting
050715
191351
warm-up for 20 minutes should be done before testing
if not, test results might be inaccurate !!
Clear all sample data?
Yes- press Enter
No- press Exit
001)ALT
002) ALP
003) T-BIL
004)ALB
005)BUN
006) Cr
007) CO2
008) CL
009) Fe
010) GLU
LIVER :
Pre. Next 002 Up Down
011)r-GT
012)CHE
013)D-BIL
014)TP
015)BUN
016)AST
017)Ca
018)P
019)Mg
020)CHO
Test nameALT 050716
ModeKINETIC
Sample No.001
Medium: Sample
ABS
Factor-3376
Result
Please Pump distilled water
Test nameTP 050716
Mode ENDPOINT
Sample No.001 ID:
Medium: sample
ABS
Factor
Result
Please pump sample
__ __ __ __
Input sample No.
__ __ __ __ __ __ __
Input ID No.
Figure 4-2-5
Test nameTP 050716
Mode ENDPOINT
Sample No. ID:
Medium:
ABS0.02
Factor 1-YES
Result 2-NO
Test reagent blank again?
Rinse Print Calibrate Zero QC
Figure 4-2-6
Test nameTP 050716
ModeENDPOINT
Sample No.
Medium
ABS0.002
Factor 1-YES
Result 2-NO
Test standard liquid again?
Figure 4-2-7
Rinse Print Calibrate Zero QC
001)ALT
002) AST
003) TP
004)ALB
005)T-BIL
006) D-BIL
007) TBA
008) ALP
009) GGT
010) LAP
LIVER
Pre Next 001 Up Down
011)AMM
012)CHE
013)GLDH
014)ICDH
015)
016)
017)
018)
019)
020)
1TEST NAME TP
2TEMP -- --
3MODE END POINT
4UNITS g/L
5LLOW:0.0000
6LHIGH:150
UP DOWN
Program P1
1R-BLANK NO
2S-BLANK NO
3STANDARD0.0
4FACTOR 0.1
5VOLUME 500
6. FLOAT: 002
UP DOWN
Program P2
1WL1 000
2WL2 --------
3DelayS: 000
4MEASS: 000
5REF-LOW: 0.1
6REF-HIGH: 0.1
UP DOWN
Program P3
STANDARD1: 0.000
STANDARD2: 0.000
STANDARD3: 0.000
STANDARD4: 0.000
STANDARD5: 0.000
Program P4
P3
UP DOWN
1Test name
0 1 2 3 4 5 6 7 8 9
0 1 2 3 4 5 6 7 8 9
A B C D E F G H I J
K L M N O P Q R S T
U V W X Y Z
Del Page
Program Test name
2TEMP
0
125
230
337
ProgramTemperature
Figure 4-3-7
3MODE
0 ENDPOINT
1 TWOPOINT
2 KINETIC
3 MULTI-POINT
4 ABS
ProgramTest mode
4UNITS
0ABS
1U/L
2kat/L
3mmol/L
ProgramUnits
Page
4.mol/L
5.g/L
6.g/dL
7. mg/dL
Figure 4-3-9
4UNITS
0g/dL
1U/mL
2nmol/L
3mg/L
ProgramUnits
Page
4.g/L
5.ng/L
6.
Input L-LOW value:
Input L-HIGH value:
1.R-BLANK
0 NO
1 YES
ProgramReagent Blank
2.S-BLANK
0 NO
1 YES
ProgramSample blank
Input standard liquid
Input factor
uL
Input sample volume
Decimal digi
Input float 0-4
__
WAVE
1340 5505
2405 6546
3450 7578
4492 8630
n m
Input wavelength 1:
___ ___ ___ S
Input delay time
Figure 4-3-20
___ ___ ___ S
Input measuring time
Figure 4-3-21
________________________
Input REF-High
__________________________
Input REF-Low
Input No.
1Program