Top Banner
CARDIOLINE ® Monitor Defibrillator for advanced life support Safety Classification IEC 60601, CF type Dimensions 195 mm (h) x 249 mm (l) x 310 mm (w) Weight with disposable electrodes: 6.3 Kg; with external paddles: 7.2 Kg Display LCD 120 x 89 mm; 320 x 240 pixels (1/4 VGA) Battery NiMH (rechargeable) Capacity More than 120 shocks at 200 J; more than 150 minutes of monitoring Battery charging time Approximately 3 hours Power supply AC: 90-264 V, 47-63Hz; DC: 10-16 V Printer High resolution thermal printer 50 mm Paper Speed 10, 25 and 50 mm/sec ± 5 % Manual printing 1 ECG channel; with SpO 2 option: 1 ECG + 1 SpO2 Automatic printing 8 seconds before and after events and defibrillation discharge Reports HR trends and % SpO 2 (optional); manual and automatic tests results; configuration parameters; stored events and relevant ECGs. Monitoring ECG channels Paddles and disposable electrodes, 5 lead cable; 3 or 10 lead cables (optional) Lead Indication on-screen icon ECG amplitude 5, 10, 20 and 40 mm/mV ECG speed 25 mm/sec Wave visualization time 4.5 seconds Alarms Max/Min HR; Max/Min SpO 2 (optional); VT/VF alarm (AED Version) Heart Rate 30-300 ppm ± 10% Frequency response 0.5-150 Hz CMRR 90 dB. Defibrillation Waveform Biphasic truncated exponential, compensated by patient’s impedance Charging time 6 seconds with new battery Energy external paddles 1-2-3-5-7-9-10-15-20-30-50-70-100-125-150-200 J Paddles Options Reusable paddles (only manual mode) disposable electrodes with reusable or disposable cable Energy selection, charge/discharge Front panel and external paddles button Charge Indication Charging and End-of-charge tone, blinking shock button Synchronization F ront panel button and on-screen indication Help On-screen and sound messages Detection (only AED) VF, VT, NSR sensitivity complies with AHA Tmax Start Analysis-Ready to shock (AED) 21 seconds with new battery Memory (AED) Extractable Compact Flash Data (AED) Continuous ECG, clinical parameters and alarms; last 100 events associated to shocks, pacing or recordings. Pacemaker option Operating modes Fixed; on-demand Waveform Constant current rectilinear with 40 msec pulse Parameters Amplitude: 0-150 mA ; HR: 30-160 bpm. Pulse Oximetry option SpO 2 Parameters 0-99%; ± 2% of 70-100% Pulse Parameters 30-254 ppm; ± 2ppm or ± 2%. Versions Manual standard and synchronized biphasic defibrillation shocks mode Manual & AED standard mode, synchronized biphasic defibrillation shocks mode, semi-automatic mode, storing in a Compact Flash. Technical specifications elife700 66529245 0705 ITA2 01 3.0M DM NV et medical devices SpA Marketing & Sales Head Office Strada Rivoltana Nuova, 53 - 20060 Vignate (MI), Italy Tel. +39 02 95 05 18.1 Fax +39 02 95 66 013 e-mail: [email protected] CARDIOLINE ® is an et medical devices SpA brand et medical devices SpA si riserva il diritto di apportare modifiche tecniche ed estetiche al prodotto descritto nel presente dépliant senza obbligo di preavviso. Rivolgersi al proprio rivenditore di fiducia per informazioni più dettagliate ed aggiornamenti. Altri nomi di prodotto o ditta sono marchi registrati dei rispettivi proprietari. Quality management system certified according to EN ISO 9001: 2000 – EN ISO 13485: 2003 - Certificate N. 800278 N e m k o C e r t i f i c a t i o n NORWEGIAN ACCREDITATION QUAL001 www.cardioline.biz elife700 0318
2

dp elife700 cardioline 01 eng1 - Cardiac Direct · Alarms Max/Min HR; Max/Min SpO2 (optional); VT/VF alarm (AED Version) Heart Rate 30-300 ... VF, VT, NSR sensitivity complies with

Jul 20, 2018

Download

Documents

vothu
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: dp elife700 cardioline 01 eng1 - Cardiac Direct · Alarms Max/Min HR; Max/Min SpO2 (optional); VT/VF alarm (AED Version) Heart Rate 30-300 ... VF, VT, NSR sensitivity complies with

CARDIOLINE®

Monitor Defibrillatorfor advanced life support

Safety Classification IEC 60601, CF typeDimensions 195 mm (h) x 249 mm (l) x 310 mm (w)Weight with disposable electrodes: 6.3 Kg; with external paddles: 7.2 Kg Display LCD 120 x 89 mm; 320 x 240 pixels (1/4 VGA)Battery NiMH (rechargeable) Capacity More than 120 shocks at 200 J; more than 150 minutes of monitoringBattery charging time Approximately 3 hoursPower supply AC: 90-264 V, 47-63Hz; DC: 10-16 VPrinter High resolution thermal printer 50 mmPaper Speed 10, 25 and 50 mm/sec ± 5 %Manual printing 1 ECG channel; with SpO2 option: 1 ECG + 1 SpO2Automatic printing 8 seconds before and after events and defibrillation dischargeReports HR trends and % SpO2 (optional); manual and automatic tests results; configuration parameters; stored events and relevant ECGs.

MonitoringECG channels Paddles and disposable electrodes, 5 lead cable; 3 or 10 lead cables (optional)Lead Indication on-screen iconECG amplitude 5, 10, 20 and 40 mm/mVECG speed 25 mm/secWave visualization time 4.5 secondsAlarms Max/Min HR; Max/Min SpO2 (optional); VT/VF alarm (AED Version)Heart Rate 30-300 ppm ± 10%Frequency response 0.5-150 HzCMRR 90 dB.

DefibrillationWaveform Biphasic truncated exponential, compensated by patient’s impedance Charging time 6 seconds with new batteryEnergy external paddles 1-2-3-5-7-9-10-15-20-30-50-70-100-125-150-200 JPaddles Options Reusable paddles (only manual mode) disposable electrodes with reusable or disposable cableEnergy selection, charge/discharge Front panel and external paddles buttonCharge Indication Charging and End-of-charge tone, blinking shock buttonSynchronization Front panel button and on-screen indicationHelp On-screen and sound messages Detection (only AED) VF, VT, NSR sensitivity complies with AHATmax Start Analysis-Ready to shock (AED) 21 seconds with new batteryMemory (AED) Extractable Compact FlashData (AED) Continuous ECG, clinical parameters and alarms; last 100 events associated to shocks, pacing or recordings.

Pacemaker optionOperating modes Fixed; on-demand Waveform Constant current rectilinear with 40 msec pulseParameters Amplitude: 0-150 mA ; HR: 30-160 bpm.

Pulse Oximetry optionSpO2 Parameters 0-99%; ± 2% of 70-100%Pulse Parameters 30-254 ppm; ± 2ppm or ± 2%.

VersionsManual standard and synchronized biphasic defibrillation shocks modeManual & AED standard mode, synchronized biphasic defibrillation shocks mode, semi-automatic mode, storing in a Compact Flash.

Technical specifications

elife700

6652

9245

070

5 IT

A2 0

1 3

.0M

DM

NV

et medical devices SpAMarketing & Sales Head OfficeStrada Rivoltana Nuova, 53 - 20060 Vignate (MI), Italy Tel. +39 02 95 05 18.1 Fax +39 02 95 66 013 e-mail: [email protected]

CARDIOLINE®

is an et medical devices SpA brandet medical devices SpA si riserva il diritto di apportare modifiche tecniche ed estetiche al prodotto descritto nel presente dépliant senza obbligo di preavviso. Rivolgersi al proprio rivenditore di fiducia per informazioni più dettagliate ed aggiornamenti. Altri nomi di prodotto o ditta sono marchi registrati dei rispettivi proprietari.

Quality management system certified according to EN ISO 9001: 2000 – EN ISO 13485: 2003 - Certificate N. 800278

Nemko

Certificati

on

NORWEGIANACCREDITATION

QUAL001

w w w . c a r d i o l i n e . b i z elife700

0318

Page 2: dp elife700 cardioline 01 eng1 - Cardiac Direct · Alarms Max/Min HR; Max/Min SpO2 (optional); VT/VF alarm (AED Version) Heart Rate 30-300 ... VF, VT, NSR sensitivity complies with

CARDIOLINE® elife700

elife700 is available in Manual and Manual/AED version.

Manual version allows standard or synchronized biphasic defibrillation shocks.

Manual & AED version adds the semi-automatic mode, where the unit analyzes ECG of the patient, and determines if the rhythm analyzed can be defibrillated, providing to the user on-screen text messages and audible messages during all the process.

In AED mode, all information are automatically stored in a Compact Flash Card, including ECG signal, clinical parameters and alarms; the last 100 events of defibrillation, pacing or printing are also memorised.

All information can be easily downloaded and analysed on PC.

Main features

Biphasic technology with truncated exponential waveform adapted to patient’s impedance.

LCD back-lit monitor for displaying clinical waveforms, parameters and operation messages.

ECG monitoring with paddles or patient cable.

Integrated thermal printer.

Control of defibrillation and printer directly from paddles.

Paediatric paddles integrated.

Operated from the mains, from a vehicle battery or with its internal battery.

elife700 is the complete solution for hospital ward. Display and printer, rechargeable batteries, standard paddles or disposable electrodes, manual and semi-automatic defibrillation, cardioversion, pacing and pulse oximetry.

Monitor Defibrillatorfor advanced life support.

The

pict

ures

mer

ely

indi

cate

pos

sibl

e w

yas

to u

se th

e pr

oduc

t. Th

e ob

ject

s illu

stra

ted

may

not

be

part

of th

e pr

oduc

t.

Pacemaker option

Integrated pacemaker for external stimulation both fixed or on demand.

This option adds to the unit cardiac pacing features, with programmable stimulation rate and current.

SpO2 option

Pulse oximetry with SIMS BCI® transducer.

This option offers oxygen saturation monitoring by displaying on the screen the percentage value of SpO2 and the plethysmographic waveform, in combination with HR value and electrocardiographic waveforms.