CARDIOLINE ® Monitor Defibrillator for advanced life support Safety Classification IEC 60601, CF type Dimensions 195 mm (h) x 249 mm (l) x 310 mm (w) Weight with disposable electrodes: 6.3 Kg; with external paddles: 7.2 Kg Display LCD 120 x 89 mm; 320 x 240 pixels (1/4 VGA) Battery NiMH (rechargeable) Capacity More than 120 shocks at 200 J; more than 150 minutes of monitoring Battery charging time Approximately 3 hours Power supply AC: 90-264 V, 47-63Hz; DC: 10-16 V Printer High resolution thermal printer 50 mm Paper Speed 10, 25 and 50 mm/sec ± 5 % Manual printing 1 ECG channel; with SpO 2 option: 1 ECG + 1 SpO2 Automatic printing 8 seconds before and after events and defibrillation discharge Reports HR trends and % SpO 2 (optional); manual and automatic tests results; configuration parameters; stored events and relevant ECGs. Monitoring ECG channels Paddles and disposable electrodes, 5 lead cable; 3 or 10 lead cables (optional) Lead Indication on-screen icon ECG amplitude 5, 10, 20 and 40 mm/mV ECG speed 25 mm/sec Wave visualization time 4.5 seconds Alarms Max/Min HR; Max/Min SpO 2 (optional); VT/VF alarm (AED Version) Heart Rate 30-300 ppm ± 10% Frequency response 0.5-150 Hz CMRR 90 dB. Defibrillation Waveform Biphasic truncated exponential, compensated by patient’s impedance Charging time 6 seconds with new battery Energy external paddles 1-2-3-5-7-9-10-15-20-30-50-70-100-125-150-200 J Paddles Options Reusable paddles (only manual mode) disposable electrodes with reusable or disposable cable Energy selection, charge/discharge Front panel and external paddles button Charge Indication Charging and End-of-charge tone, blinking shock button Synchronization F ront panel button and on-screen indication Help On-screen and sound messages Detection (only AED) VF, VT, NSR sensitivity complies with AHA Tmax Start Analysis-Ready to shock (AED) 21 seconds with new battery Memory (AED) Extractable Compact Flash Data (AED) Continuous ECG, clinical parameters and alarms; last 100 events associated to shocks, pacing or recordings. Pacemaker option Operating modes Fixed; on-demand Waveform Constant current rectilinear with 40 msec pulse Parameters Amplitude: 0-150 mA ; HR: 30-160 bpm. Pulse Oximetry option SpO 2 Parameters 0-99%; ± 2% of 70-100% Pulse Parameters 30-254 ppm; ± 2ppm or ± 2%. Versions Manual standard and synchronized biphasic defibrillation shocks mode Manual & AED standard mode, synchronized biphasic defibrillation shocks mode, semi-automatic mode, storing in a Compact Flash. Technical specifications elife700 66529245 0705 ITA2 01 3.0M DM NV et medical devices SpA Marketing & Sales Head Office Strada Rivoltana Nuova, 53 - 20060 Vignate (MI), Italy Tel. +39 02 95 05 18.1 Fax +39 02 95 66 013 e-mail: [email protected] CARDIOLINE ® is an et medical devices SpA brand et medical devices SpA si riserva il diritto di apportare modifiche tecniche ed estetiche al prodotto descritto nel presente dépliant senza obbligo di preavviso. Rivolgersi al proprio rivenditore di fiducia per informazioni più dettagliate ed aggiornamenti. Altri nomi di prodotto o ditta sono marchi registrati dei rispettivi proprietari. Quality management system certified according to EN ISO 9001: 2000 – EN ISO 13485: 2003 - Certificate N. 800278 N e m k o C e r t i f i c a t i o n NORWEGIAN ACCREDITATION QUAL001 www.cardioline.biz elife700 0318