D-1
Small Tool Instruments Calipers Height Gages Depth Gages
About the charge function (Super Caliper)Th60 lux. -
Caliper can be used
The charge function allows the operator to use this Super
illumination (less than the minimum necessary illumination intensity).
the charging conditions. If this Super Caliper is left unused in an illumination of 500 lux (usual for manufacturing
reach full charge.
IP67 protection level
No ingress of dust.
immersion in water.
effects shall not be possible when the enclosure is temporarily immersed 1 meter in water under standardized conditions of pressure and time (30 min.).
Improvement in surface finish of beam
Direction of the length Direction of the length
Super Caliper
Ra: 0.126µm Ra: 0.07µm
D-3
Technical Data
Resolution: .0005”/0.01mm or 0.01mm Repeatability: .0005” / 0.01mmDisplay: LCDLength standard:ABSOLUTE electrostatic capacitance type linear encoderMax. response speed: Unlimited
FunctionOrigin-set, Data hold, Data output,
Optional Accessories959143: Data hold unit (SPC output model only)959149: SPC cable with data switch (40” / 1m) 959150: SPC cable with data switch (80” / 2m)
off for the entire life of the caliper. At 60 Lux and up the ABSOLUTE solar caliper is turned on ready to start measurement.
Range L a b c Mass (g)0 - 4" / 0 - 100mm 182 40 21 16.5 1250 - 6" / 0 - 150mm 233 40 21 16.5 1500 - 8" / 0 - 200mm 290 50 24.5 20 180
FEATURES
low light intensity.
durability.
DIMENSIONS AND MASSUnit: mm
500-474
D-4
ABSOLUTE Coolant Proof CaliperSERIES 500 — with Dust/Water Protection Conforming to IP67 Level
Technical Data
Resolution: .0005"/0.01mm or 0.01mmRepeatability: .0005” / 0.01mmDisplay: LCDLength standard: ABSOLUTE electromagnetic induction linear encoderMax. response speed: Unlimited
938882Battery life: Approx. 3 years under normal use
FunctionOrigin-set, Zero-setting, Automatic power on/off, Data output,
Optional Accessories05CZA624: SPC cable with data switch (40” / 1m)05CZA625: SPC cable with data switch (80” / 2m)Connecting cable for U-Wave-T02AZD790A: SPC cable for U-Wave (160mm)
FEATURES
exposed to coolant, water, dust or oil.
scale.
system.
500-763-10
500-752-10
with a water-resistant SPC cable.
05CZA624
Built-in ABS (absolute) scale means that these calipers are ready to use immediately after power-on without origin resetting. It’s as easy as vernier caliper measurements.
D-5
Air
Valve Valve
Master Capsule
Differentialpressure gage
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Remarks0-150mm 500-702-10* +/-0.02mm 0.01mm0-150mm 500-712-10 +/-0.02mm 0.01mm0-150mm 500-719-10 +/-0.02mm 0.01mm dia. 1.9mm rod depth bar0-150mm 500-721-10 +/-0.02mm 0.01mm carbide-tipped jaws for ID measurement0-150mm 500-723-10 +/-0.02mm 0.01mm carbide-tipped jaws for OD & ID measurement0-200mm 500-703-10* +/-0.02mm 0.01mm0-200mm 500-713-10 +/-0.02mm 0.01mm0-200mm 500-722-10 +/-0.02mm 0.01mm carbide-tipped jaws for ID measurement0-200mm 500-724-10 +/-0.02mm 0.01mm carbide-tipped jaws for OD & ID measurement0-300mm 500-704-10* +/-0.03mm 0.01mm0-300mm 500-714-10 +/-0.03mm 0.01mm
Air leakage detection system used for water-proof testing
resistance. Testing begins by placing a measuring tool into
none of the air in the capsule seeps into the measuring tool,
master, and the differential pressure gage will continue to
measuring tool, it will create an air pressure difference in the amount indicated by the differential pressure gage. Thus, detection of air pressure differences is used as a criterion for
calipers and Coolant Proof micrometer is tested this way for
IP67 protection level
No ingress of dust.
immersion in water.
effects shall not be possible when the enclosure is temporarily immersed 1 meter in water under standardized conditions of pressure and time (30 min.).
Inch/Metric
Range Order No. Accuracy Resolution Remarks0-6”/0-150mm 500-752-10* +/-.001” .0005”/0.01mm0-6”/0-150mm 500-762-10 +/-.001” .0005”/0.01mm0-6”/0-150mm 500-768-10* +/-.001” .0005”/0.01mm .075” rod depth bar0-6”/0-150mm 500-769-10 +/-.001” .0005”/0.01mm .075” rod depth bar0-6”/0-150mm 500-731-10* +/-.001” .0005”/0.01mm carbide-tipped jaws for OD measurement0-6”/0-150mm 500-735-10 +/-.001” .0005”/0.01mm carbide-tipped jaws for OD measurement0-6”/0-150mm 500-733-10* +/-.001” .0005”/0.01mm carbide-tipped jaws for OD & ID measurement0-6”/0-150mm 500-737-10 +/-.001” .0005”/0.01mm carbide-tipped jaws for OD & ID measurement0-8”/0-200mm 500-753-10* +/-.001” .0005”/0.01mm0-8”/0-200mm 500-763-10 +/-.001” .0005”/0.01mm0-8”/0-200mm 500-732-10* +/-.001” .0005”/0.01mm carbide-tipped jaws for OD measurement0-8”/0-200mm 500-736-10 +/-.001” .0005”/0.01mm carbide-tipped jaws for OD measurement0-8”/0-200mm 500-734-10* +/-.001” .0005”/0.01mm carbide-tipped jaws for OD & ID measurement0-8”/0-200mm 500-738-10 +/-.001” .0005”/0.01mm carbide-tipped jaws for OD & ID measurement0-12”/0-300mm 500-754-10* +/-.0015” .0005”/0.01mm0-12”/0-300mm 500-764-10 +/-.0015” .0005”/0.01mm
Range L a b c e Mass (g)0 - 6” / 0 - 150mm 233 40 21 16.5 16 1680 - 8” / 0 - 200mm 290 50 24.5 20 16 1980 - 12” / 0 - 300mm 404 64 27.5 22 20 350
DIMENSIONS AND MASS Unit: mm
IP67 model
IP67 model
D-6
ABSOLUTE Digimatic CaliperSERIES 500 — with Exclusive ABSOLUTE Encoder Technology
Technical Data
Resolution: .0005”/0.01mm or 0.01mmRepeatability: .0005”/ 0.01mm Display: LCDLength standard: ABSOLUTE electromagnetic induction type linear encoder (200mm and smaller models) ABSOLUTE electrostatic capacitance type linear encoder (300mm and larger models)Max. response speed: UnlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3.5 years under normal use
FunctionOrigin-set, Zero-setting, Data output,
Optional Accessories959143: Data hold unit959149: SPC cable with data switch (40” / 1m)959150: SPC cable with data switch (80” / 2m)Connecting cable for U-Wave-T02AZD790C: SPC cable for U-Wave (160mm)
Direction of the length Direction of the length
ABS Digimatic Caliper
Ra: 0.126µm Ra: 0.07µm
Mitutoyo's absolute Digimatic Caliper is the next generation of electronic calipers. It keeps track of its origin point once set.
displays the actual slider position ready to start measurement. No more repeated zero setting is necessary with the absolute encoder technology as well as no more care
FEATURES
be Zero-Set at any slider position along the scale for incremental comparison measurements. This switch will also allow return to the absolute (ABS) coordinate and display of the true position from the origin point (usually jaws-closed point).
to and including 12” or 300mm.
Except 40” / 1000mm supplied in wooden case.
High quality guide surface finish for smooth slider movement
Range L a b c d Mass (g)0 - 4” / 0 - 100mm 182 40 21 16.5 16 1430 - 6” / 0 - 150mm 233 40 21 16.5 16 1680 - 8” / 0 - 200mm 290 50 24.5 20 16 1980 - 12” / 0 - 300mm 403 64 27.5 22 20 350
DIMENSIONS AND MASSUnit: mm
959143
959149
Range L a b c d Mass (g)0 - 18” / 0 - 450mm 630 90 47 38 25 11700 - 24” / 0 - 600mm 780 90 47 38 25 13500 - 40” / 0 - 1000mm 1240 130 61 50 32 3300
500-197-30
500-151-30
The new Mitutoyo ABS Digimatic Caliper line with
measurement dependability by increasing resistance to harsh workshop conditions.
D-7
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Remarks0 - 100mm 500-150-30 ±0.02mm 0.01mm ø1.9mm rod depth bar0 - 150mm 500-151-30 ±0.02mm 0.01mm —0 - 150mm 500-154-30 ±0.02mm 0.01mm Carbide-tipped jaws for OD measurement0 - 150mm 500-155-30 ±0.02mm 0.01mm Carbide-tipped jaws for OD & ID measurement0 - 150mm 500-158-30 ±0.02mm 0.01mm ø1.9mm rod depth bar0 - 200mm 500-152-30 ±0.02mm 0.01mm —0 - 200mm 500-156-30 ±0.02mm 0.01mm Carbide-tipped jaws for OD measurement0 - 200mm 500-157-30 ±0.02mm 0.01mm Carbide-tipped jaws for OD & ID measurement0 - 300mm 500-153 ±0.03mm 0.01mm —0 - 450mm 500-500-10 ±0.05mm 0.01mm0 - 600mm 500-501-10 ±0.05mm 0.01mm0 - 1000mm 500-502-10 ±0.07mm 0.01mm
Function of Digital ModelOrigin-set, Zero-setting, Data output,
Optional Accessories for Digital Model959143: Data hold unit959149: SPC cable with data switch (40” / 1m)959150: SPC cable with data switch (80” / 2m)05CZA624: SPC cable with data switch (40” / 1m) for IP67 model05CZA625: SPC cable with data switch (80” / 2m) for IP67 model
Round surface of jaws for accurate ID measurement
550-311-10
D-16
L
d
c
e
b
a
T
Technical Data
FEATURES
features.
Long Jaw Vernier CaliperSERIES 534
Range L a b c d e T0 - 12” / 0 - 300mm 445 110 90 12 20 7 3.80 - 20” / 0 - 500mm 682 225 200 18.5 25 12 60 - 30” / 0 - 750mm 995 232 200 18.5 32 12 80 - 40” / 0 - 1000mm 1230 232 200 18.5 32 12 8
DIMENSIONS Unit: mm
534-110
534-114
Round surface of jaws for accurate CD measurement
SPECIFICATIONSMetric
Order No. Accuracy Graduation Mass (g)Lower Scale Upper Scale
Resolution: 0.01mm or .0005"/0.01mmDisplay: LCDLength standard: ABSOLUTE electrostatic capacitance type linear encoderMax. response speed: UnlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3.5 years under normal use
Function of Digital ModelOrigin-set, Zero-setting, Data output,
Optional Accessories for Digital Model05CZA624: SPC cable with data switch (1m / 40”) IP6705CZA625: SPC cable with data switch (2m / 80”) IP67
Digimatic Carbon Fiber CaliperSERIES 552 — with optional jaw attachments
Technical Data
Resolution: .0005”/0.01mm or 0.01mmDisplay: LCDLength standard: Electrostatic capacitance type linear encoderMax. response speed: unlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3,000 hours in continuous use
SPECIFICATIONSMetric
Order No. Accuracy Resolution Remarks0(20)-450mm 552-302-10 +/-0.04mm 0.01mm0(20)-450mm 552-150-10 +/-0.06mm 0.01mm long jaws 200mm0(20)-450mm 552-155-10 +/-0.04mm 0.01mm ceramic jaws0(20)-600mm 552-303-10 +/-0.04mm 0.01mm0(20)-600mm 552-151-10 +/-0.06mm 0.01mm long jaws 200mm0(20)-600mm 552-156-10 +/-0.04mm 0.01mm ceramic jaws0(20)-1000mm 552-304-10 +/-0.05mm 0.01mm0(20)-1000mm 552-152-10 +/-0.07mm 0.01mm long jaws 200mm0(20)-1500mm 552-305-10 +/-0.09mm 0.01mm0(20)-1500mm 552-153-10 +/-0.11mm 0.01mm long jaws 200mm0(20)-2000mm 552-306-10 +/-0.12mm 0.01mm0(20)-2000mm 552-154-10 +/-0.14mm 0.01mm long jaws 200mm
Inch/Metric
Order No. Accuracy Resolution Remarks0(.5”)-18” 552-312-10 +/-.002” .0005”/0.01mm0(.5”)-18” 552-160-10 +/-.0025” .0005”/0.01mm long jaws 7.9”0(.5”)-18” 552-165-10 +/-.002” .0005”/0.01mm ceramic jaws0(.5”)-24” 552-313-10 +/-.002” .0005”/0.01mm0(.5”)-24” 552-161-10 +/-.0025” .0005”/0.01mm long jaws 7.9”0(.5”)-24” 552-166-10 +/-.002” .0005”/0.01mm ceramic jaws0(1”)-40” 552-314-10 +/-.002” .0005”/0.01mm0(1”)-40” 552-162-10 +/-.003” .0005”/0.01mm long jaws 7.9”0(1”)-60” 552-315-10 +/-.004” .0005”/0.01mm0(1”)-60” 552-163-10 +/-.0045” .0005”/0.01mm long jaws 7.9”0(1”)-80” 552-316-10 +/-.005” .0005”/0.01mm0(1”)-80” 552-164-10 +/-.0055” .0005”/0.01mm long jaws 7.9”
FEATURES
CFRP (Carbon Fiber Reinforced Plastics) in the beam and jaws.
by using the optional attachments.
pressing the Offset key.)
starting point.
and magnetic products.
552-314-10
552-155-10
FunctionOrigin-set, Zero-setting, Presetting, Offsetting, Data hold, Data output,
)
Optional Accessories05CZA624: SPC cable with data switch (40”/ 1m)05CZA625: SPC cable with data switch (80”/ 2m)
914055914056914057914058914053: Attachment clamps (for models up to 24” / 600mm range)914054 24” / 600mm range)
type.
Centerline Attachments Pointed ID Measuring Attachments
Attachment Clamps
D-19
Ceramic jaws ID point attachmentsCenterline attachments
552-151-10
DIMENSIONS AND MASSUnit: mm
Standard models, Ceramic jaw modelsRange L a b c d e f Mass (g)0 - 18" / 0 - 450mm 640 100 18 41.2 91.8 25 10 (.25") 7150 - 24" /0 - 600mm 790 100 18 41.2 91.8 25 10 (.25") 7900 - 40" / 0 - 1000mm 1230 150 24 62.8 113.8 32 10 (.5") 1,7600 - 60" / 0 - 1500mm 1740 150 24 62.8 113.8 32 10 (.5") 2,1600 - 80" / 0 - 2000mm 2250 150 24 62.8 113.8 32 10 (.5") 2,560
( ): inch/mm models
Long jaw modelsRange L a b c d e f Mass (g)0 - 18" / 0 - 450mm 680 200 24 89 91.8 25 10 (.25") 1,2150 - 24" / 0 - 600mm 830 200 24 89 91.8 25 10 (.25") 1,2900 - 40" / 0 - 1000mm 1280 200 24 110 113.8 32 10 (.5") 2,0900 - 60" / 0 - 1500mm 1790 200 24 110 113.8 32 10 (.5") 2,4900 - 80" / 0 - 2000mm 2300 200 24 110 113.8 32 10 (.5") 2,890
( ): inch/mm models
10
50 φ1
9510
0
10
φ110
40°
PointedID MeasuringAttachments
CenterlineAttachments
Unit: mm
D-20
Centerline Type
Order No. Components a b07CZA057 07CZA039 x 2pcs. 30mm 30mm07CZA060 07CZA047 x 2pcs. 1.2" 1.2"
Range Order No. Accuracy Graduation Mass (g)0 - 150mm 536-101 ±0.05mm 0.05mm 1500 - 200mm 536-102 ±0.05mm 0.05mm 2000 - 300mm 536-103 ±0.08mm 0.05mm 400
Technical Data
linear encoder
938882
Function of Digital ModelOrigin-set, Zero-setting, Power On/Off, Data output,
)
Optional Accessories for Digital Model05CZA624: SPC cable with data switch (40” / 1m )05CZA625: SPC cable with data switch (80” / 2m)Connecting cable for U-Wave-T02AZD790A: SPC cable for U-Wave (160mm)
DIMENSIONS Unit: mm
573-701
Range a b c d0 - 6” / 0 - 150mm 95 10 40 30.4 (30)0 - 8” / 0 - 200mm 95 10 50 40.4 (38.5)0 - 12” / 0 - 300mm 135 15 64 51
( ) Digital Model
536-101
Digital model
Digital model
D-23
FEATURES
distance measurements on the same and offset planes.
Offset Centerline CaliperSERIES 573, 536 — ABSOLUTE Digimatic and Vernier Type
1.5mm - 10mm (.06" - .4").
Technical Data
linear encoder
938882
Function of Digital ModelOrigin-set, Zero-setting, Power On/Off, Data output,
Optional Accessories for Digital Model05CZA624: SPC cable with data switch (40” / 1m)05CZA625: SPC cable with data switch (80” / 2m)Connecting cable for U-Wave-T02AZD790A: SPC cable for U-Wave (160mm)
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Mass (g)10 - 150mm 573-605 ±0.03mm 0.01mm 15710 - 200mm 573-606 ±0.03mm 0.01mm 17710 - 300mm 573-608 ±0.04mm 0.01mm 320
Range Order No. Accuracy Graduation Mass (g)0 - 150mm 536-151 ±0.05mm 0.05mm 1400 - 150mm 536-152* ±0.05mm 0.05mm 140
DIMENSIONSUnit: mm
573-752
536-151 536-152
Point jaw type
Digital model
Digital model
(60°)
φ0.
5φ
2.4
D-27
229
(32)
40
ø3
Vernier model
Digital model
16
231
Mass: 167g
Mass: 150g
(32)
40
ø3
16
FEATURES
facilitates measurements of tube wall thickness.
Tube Thickness CaliperSERIES 573, 536 — ABSOLUTE Digimatic and Vernier Type
Technical Data
Length standard: ABSOLUTE electrostatic capacitance type linear encoder
938882
Function of Digital ModelOrigin-set, Zero-setting, Power On/Off, Data output,
)
Optional Accessories for Digital Model05CZA624: SPC cable with data switch (40” / 1m)05CZA625: SPC cable with data switch (80” / 2m)
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Mass(g)0 - 150mm 573-661 ±0.05mm 0.01mm 167
Inch/Metric
Range Order No. Accuracy Resolution Mass(g)0 - 6" / 0 - 150mm 573-761 ±.002" .0005" / 0.01mm 167
Metric
Range Order No. Accuracy Graduation Mass(g)0 - 150mm 536-161 ±0.05mm 0.05mm 150
DIMENSIONS AND MASSUnit: mm
573-761
536-161
Digital model
Digital model
D-28
Mass: 253g
FEATURES
calipers are ideal for elastic or resilient workpieces such as plastic parts and
ABSOLUTE Low Force CaliperSERIES 573
rubber parts that standard calipers cannot measure.
FEATURES
features a spring-loaded mechanism to
ABSOLUTE Snap CaliperSERIES 573
Technical Data
Resolution: .0005”/0.01mm or 0.01mm Repeatability: .0005” / 0.01mmDisplay: LCDLength standard: ABSOLUTE electrostatic capacitance type linear encoderMeasuring force: 7N to 14N (700gf to 1400gf)Jaw retraction: 2mmMax. response speed: UnlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3.5 years under normal use
FunctionOrigin-set, Zero-setting, Power On/Off, Data output,
)
Optional Accessories959143: Data hold unit959149: SPC cable with data switch (40” / 1m)959150: SPC cable with data switch (80” / 2m)
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Mass (g)0 - 100mm 573-181-20 ±0.02mm 0.01mm 2130 - 150mm 573-182-20 ±0.02mm 0.01mm 233
Resolution: .0005”/0.01mm or 0.01mmDisplay: LCDLength standard: ABSOLUTE electrostatic capacitance type linear encoderMeasuring force: 0.49N to 0.98N (50gf to 100gf)Jaw retraction: 0.3mmMax. response speed: UnlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3.5 years under normal use
FunctionOrigin-set, Zero-setting, Power On/Off, Data output,
Optional Accessories959143: Data hold unit959149: SPC cable with data switch (40” / 1m)959150: SPC cable with data switch (80” / 2m)
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Mass (g)0 - 180mm 573-191-20 ±0.05mm 0.01mm 253
Inch/Metric
Range Order No. Accuracy Resolution Mass (g)0 - 7" / 0 - 180mm 573-291-20 ±.002" .0005” / 0.01mm 253
To measure resilient workpieces take the measurement when the pointer is between the two index lines.
Range L a b c d Mass (g)0 - 100mm 233 40 21 16.5 16 2130 - 150mm 290 50 24.5 20 16 233
Measurement procedures
DIMENSIONS AND MASSUnit: mm
inspection for mass production parts.
DIMENSIONS AND MASS Unit: mm
573-191-20
573-281-20
D-29
DIMENSIONS Unit: mm
FEATURES
scribing on workpiece.
Scribing CaliperSERIES 573, 536 — ABSOLUTE Digimatic and Vernier Type
Technical Data
Display: LCDLength standard: ABSOLUTE electrostatic capacitance type linear encoderMax. response speed: UnlimitedBattery: SR44 (1 pc.), 938882Battery life: Approx. 3.5 years under normal use
FunctionOrigin-set, Zero-setting, Power On/Off, Data output,
Optional Accessories05CZA624: SPC cable with data switch (40” / 1m)05CZA625: SPC cable with data switch (80” / 2m)
SPECIFICATIONSMetric
Range Order No. Accuracy Resolution Mass (g)0 - 150mm 573-676 ±0.02mm 0.01mm 1660 - 200mm 573-677 ±0.02mm 0.01mm 1960 - 300mm 573-679 ±0.03mm 0.01mm 345
Metric
Range Order No. Accuracy Graduation Mass (g)0 - 150mm 536-221 ±0.05mm 0.05mm 1500 - 200mm 536-222 ±0.05mm 0.05mm 1800 - 300mm 536-223 ±0.08mm 0.05mm 355
573-676
Range L a b c d e0 - 150mm 229 46 21.5 17 16 330 - 200mm 288 50 25 20.5 16 430 - 300mm 403 64 27.5 22 20 54
Digital model
Vernier type
D-30
FEATURES
in hard-to-reach places.
ABSOLUTE Inside CaliperSERIES 573, 536 — Knife-edge/Inside Groove/Point Jaw Type
SPECIFICATIONSMetric
Range Order No. Accuracy Remarks Mass (g)10 - 200mm 573-642 ±0.05mm Knife-edge type, Measurable min. hole diameter: ø10mm 22710 - 160mm 573-645 ±0.05mm 14720 - 170mm 573-646 ±0.03mm Point jaw type, Measurable min. hole diameter: ø20mm 157
Inch/Metric
Range Order No. Accuracy Remarks Mass (g).4" - 8" / 10-200mm 573-742 ±.002" Knife-edge type, Measurable min. hole diameter: ø.4" 227.4" - 6" / 10-150mm 573-745 ±.002" 147.8" - 6" / 20-150mm 573-746 ±.0015" Point jaw type, Measurable min. hole diameter: ø.8" 157
Metric
Range Order No. Accuracy Remarks Mass (g)10 - 200mm 536-142 ±0.12mm Knife-edge type, Measurable min. hole diameter: ø10mm 21010 - 150mm 536-145 ±0.05mm 13020 - 150mm 536-146 ±0.05mm Point jaw type, Measurable min. hole diameter: ø20mm 14030 - 300mm 536-147 ±0.08mm Point jaw type, Measurable min. hole diameter: ø30mm 37070 - 450mm 536-148 ±0.10mm Point jaw type, Measurable min. hole diameter: ø70mm 1,25070 - 600mm 536-149 ±0.12mm Point jaw type, Measurable min. hole diameter: ø70mm 1,430
Technical Data
linear encoder
938882
Function of Digital ModelOrigin-set, Zero-setting, Power On/Off, Data output,
Optional Accessories for Digital Model05CZA624: SPC cable with data switch (40” / 1m)05CZA625: SPC cable with data switch (80” / 2m)
573-742
536-142
536-145
536-146
573-746
573-745
Knife-edge type
18” / 450mm and 24”/ 600mm supplied in wooden case.
Inside groove type
Point jaw type
Digital model
Digital model
D-31
288
Mass: 210g536-142
Mass: 130g536-145 Mass: 143g
573-645 / 573-745
Mass: 153g573-646 / 573-746
Mass: 223g573-642 / 573-742
8 (H)
70
(D)
0.3 3
229
35
3 20
5 (H)
3 (I)
3
a
(d)
60°
ø0.5
ø2.4
(ø5)
(ø1)
c g
16
16
b
X‘ X
f e
Section X−X’ Section X−X’
290
8
65
0.3 3
233
35
3 20
5 3
3.5
233
33.5
25
90.5
74.5
58 70
38
40
60°
ø0.5
ø2.4
28.5
16
16
16
X‘
X
R5 R5 R5 R5
5 5
Knife-edge type
Inside groove type
Point jaw type
DIMENSIONS AND MASSUnit: mm
Order No. Range a b c d e f g Mass (g)536-146 150mm 229 16 38 31 5 5 2 140536-147 300mm 403 20 98 89 5 10 2 370536-148 450mm 610 25 145 136 10 25 5 1,250536-149 600mm 750 25 145 136 10 25 5 1,430
D-32
FEATURES
tool for DIY.
readout of measurements.
MyCAL-LiteSERIES 700 — Digital Caliper for DIY
SPECIFICATIONSInch/Metric
Range Order No. Accuracy Mass (g)0 - 6” / 0 - 150mm 700-113-10 ±.005” / ±0.2mm 1500 - 8” / 0 - 200mm 700-123-10 ±.005” / ±0.2mm 170
Technical Data
Resolution: .001” / 0.1mmDisplay: LCDLength standard: Electrostatic capacitance type linear encoderMax. response speed: 1800mm/sBattery: SR44 (1 pc.), 938882
FunctionZero-setting, Power on/off
700-113-10
4021
16
16.5
231 (287)
DIMENSIONS Unit: mm
( ): 8" Model
700-123-10
D-33
Depth Base AttachmentOptional Accessories for Caliper
Note) Above left, 0.15 mm (2) is read at the position where a main scale graduation line corresponds with a vernier graduation line.
CalipersQuick Guide to Precision Measuring Instruments
Step measuring faces Gib, sliderLocking screw
Screw, gib pressingBeam Stopper, slider
Depth measuring facesMain scale
Reference surfaceThumbwheel
Vernier scaleSlider
Inside measuring faces
Inside jaws
Outside jaws
Outside measuring faces
Screw, gib setting
Depth bar
D-35
■ About Long CalipersSteel rules are commonly used to roughly measure large workpieces but if a little more accuracy is needed then a long caliper is suitable for the job. A long caliper is very convenient for its user friendliness but does require some care in use. In the first place it is important to realize there is no relationship between resolution and accuracy. For details, refer to the values in our catalog. Resolution is constant whereas the accuracy obtainable varies dramatically according to how the caliper is used.The measuring method with this instrument is a concern since distortion of the main beam causes a large amount of the measurement error, so accuracy will vary greatly depending on the method used for supporting the caliper at the time. Also, be careful not to use too much measuring force when using the outside measuring faces as they are furthest away from the main beam so errors will be at a maximum here. This precaution is also necessary when using the tips of the outside measuring faces of a long-jaw caliper.
■ Vernier scaleThis is a short auxiliary scale that enables accurate interpolation between the divisions of a longer scale without using mechanical magnification. The principle of operation is that each vernier scale division is slightly smaller than a main scale division, so that successive vernier graduations successively coincide with main scale graduations as one is moved relative to the other. Specifically, n divisions on a vernier scale are the same length as n-1 divisions on the main scale it works with, and n defines the division (or interpolation) ratio. Although n may be any number, in practice it is typically 10, 20, 25, etc., so that the division is a useful decimal fraction. The example below is for n = 10. The main scale is graduated in mm, and so the vernier scale is 9mm (10 divisions) long, the same as 9mm (9 divisions) on the main scale. This produces a difference in length of 0.1mm (1) as shown in figure A (the 1st vernier graduation is aligned with the first main scale graduation). If the vernier scale is slid 0.1mm to the right as shown in figure B, the 2nd graduation line on the vernier scale moves into alignment with the 2nd line on the main scale (2), and so enables easy reading of the 0.1mm displacement.
Some early calipers divided 19 divisions on the main scale by 20 vernier divisions to provide 0.05mm resolution. However, the closely spaced lines proved difficult to read and so, since the 1970s, a long vernier scale that uses 39 main scale divisions to spread the lines is generally used instead, as shown below.
0 1 2 3 4 5 6 7 8 9 100 2 4 6 8 10
0 10 20 30 40 30 40 50 60 70
39mm19mm
0.05mm
Scale reading 1.45mm Scale reading 30.35mm
Calipers were made that gave an even finer resolution of 0.02mm. Theserequired a 49-division vernier scale dividing 50 main scale divisions. However,they were difficult to read and are now hard to find since Digital calipers withan easily read display and resolution of 0.01mm appeared.
■ Inside Measurement with a CM-type CaliperBecause the inside measuring faces of a CM-type caliper are at the tips of the jaws the measuring face parallelism is heavily affected by measuring force, and this becomes a large factor in the measurement accuracy attainable. In contrast to an M-type caliper, a CM-type caliper cannot measure a very small hole diameter because it is limited to the size of the stepped jaws, although normally this is no inconvenience as it would be unusual to have to measure a very small hole with this type of caliper. Of course, the radius of curvature on the inside measuring faces is always small enough to allow correct hole diameter measurements right down to the lowest limit (jaw closure).Mitutoyo CM-type calipers are provided with an extra scale on the slider for inside measurements so they can be read directly without the need for calculation, just as for an outside measurement. This useful feature eliminates the possibility of error that occurs when having to add the inside-jaw-thickness correction on a single-scale caliper.
■ Small hole measurement with an M-type caliperA structural error d occurs when you measure the internal diameter of a small hole.
øD : True internal diameterød : Measured diametert1, t2: Thickness of the inside jawC: Distance between the inside jaws
Special Purpose Caliper ApplicationsPoint jaw type Offset jaw type Depth type Blade jaw type
For uneven surface measurement For stepped feature measurement For depth measurement For diameter of narrow groove measurement
C-type
Standard outside measurementInside measurement of a stepped holeMeasurement of a stepped part
CN-type
Standard outside measurementMeasurement of a stepped holeMeasurement of a stepped part
9mm (9 graduation lines)
(1) 0.1mm
0.1mm
Divided into10 equal divisions
A
B
(2) Aligned
D-36
■ General notes on use of caliper1. Potential causes of error
a caliper does not conform to Abbe's Principle, differential thermal expansion due to a temperature difference between the caliper and workpiece, and the effect of the thickness of the knife-edge jaws and the clearance between these jaws during measurement of the diameter of a small hole. Although there are also other error factors such as graduation accuracy, reference
the jaws, these factors are included within the instrumental error tolerances.
so that consumers can appreciate the error factors caused by the structure of the caliper before use. These notes relate to the measuring force and stipulate
you measure it with the root or tip of the jaw because a large error could occur in such cases.”
L
L < øD
øD
2. Inside measurementInsert the inside jaw as deeply as possible before measurement.
3. Depth measurement
4. Parallax error when reading the scales
apparent alignment position is distorted by due to a parallax effect caused by the step height (H) between the planes of
height should be no more than 0.3 mm.
A
ΔX
Δf
Δf
AA
BH H
5. Moving Jaw Tilt Error
If the moving jaw becomes tilted out of parallel with the fixed jaw, either through excessive force being used on the slider or lack of straightness in the reference edge of the beam, a measurement error will occur as shown in the figure. This error may be substantial due to the fact that a caliper does not conform to Abbe’s Principle.
f=h =h·a/ℓ
h
h
a
f
Example: Assume that the error slope of the jaws due to tilt of the slider is 0.01mm in 50mm and the outside measuring jaws are 40mm deep, then the error (at the jaw tip) is calculated as (40/50)x0.01mm = 0.008mm.
If the guide face is worn then an error may be present even using the correct measuring force.
6. Relationship between measurement and temperature
The main scale of a caliper is engraved (or mounted on) stainless steel, and although the linear thermal expansion coefficient is equal to that of the most common workpiece material, steel, i.e. (10.2 ±1) ×10-6 / K, note that other workpiece materials, the room temperature and the workpiece temperature may affect measurement accuracy.
7. HandlingCaliper jaws are sharp, and therefore the instrument must be handled with care to avoid personal injury. Avoid damaging the scale of a digital caliper and do not engrave an identification number or other information on it with an electric marker pen.Avoid damaging a caliper by subjecting it to impact with hard objects or by dropping it on a bench or the floor.
8. Maintenance of beam sliding surfaces and measuring facesWipe away dust and dirt from the sliding surfaces and measuring faces with a dry soft cloth before using the caliper.
9. Checking and setting the origin before useClean the measuring surfaces by gripping a sheet of clean paper between the outside jaws and then slowly pulling it out. Close the jaws and ensure that the vernier scale (or display) reads zero before using the caliper. When using a Digimatic caliper, reset the origin (ORIGIN button) after replacing the battery.
ORIGIN
BatteryClose jaws completely
ORIGIN button sets display to zero
10. Handling after useAfter using the caliper, completely wipe off any water and oil. Then, lightly apply anti-corrosion oil and let it dry before storage.Wipe off water from a waterproof caliper as well because it may also rust.
11. Notes on storageAvoid direct sunlight, high temperatures, low temperatures, and high humidity during storage.If a digital caliper will not be used for more than three months, remove the battery before storage.Do not leave the jaws of a caliper completely closed during storage.