Top Banner
DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation
44

DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Dec 19, 2015

Download

Documents

Jessie Cox
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

DICOM Second Generation Radiotherapy Objects

Part 1

Supplement 147: Physicians Intent and Segment Annotation

Page 2: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Ulrich Busch• Systems Analyst with Varian Medical

Systems• Member of DICOM WG-07 since 2000• Chairman of DICOM WG-07 since 2011• Member of the IHE-RO Technical

Committee since 2005

2DICOM Second Generation Radiotherapy

Page 3: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Christof Schadt• Software Principal with Brainlab AG• Member of DICOM WG-07 since 2005• Vice-Chair DICOM WG-07 since 2011• Member of the IHE-RO Technical

Committee since 2005

3DICOM Second Generation Radiotherapy

Page 4: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Outline

4

Outline• Overview : All of Second Generation• Discuss the particulars of Supplement 147

•Physician Intent •Segment Annotations

• Touch on the Other Supplements that make up Second Generation

• Next steps

DICOM Second Generation Radiotherapy

Page 5: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Introduction

• New Supplements to DICOM Standard

• Defined by manufacturers and clinicians

• Supports and/or influences new technologies and thus new products

5DICOM Second Generation Radiotherapy

Page 6: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

DICOM WG-07

• One of the DICOM Standards Committee’s (DSC) Working Groups

• Focus on maintenance and further development of Radiotherapy extensions of the DICOM Standard (“DICOM RT “)

• Current (regular) industry members: Accuray, Brainlab, Elekta, IBA, Mevion, Philips, RaySearch, Varian (and others)

6DICOM Second Generation Radiotherapy

Page 7: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

What is DICOM RT?

• DICOM RT is just DICOM

• ‘RT’ indicates the DICOM objects dealing with RT.

• Current release of RT Objects is now known as ”1st Generation RT“ (Gen1)

7DICOM Second Generation Radiotherapy

Page 8: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

2nd Generation

• “Second Generation Radiotherapy“• Supplements are about 270 pages• created by DICOM WG-07

• Worked on for over 7 years• Major re-structuring of Gen1 RT objects

8DICOM Second Generation Radiotherapy

Page 9: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Why now 2nd Generation?

9

Introduction – Full RT Second Generation Object Suite• Initial definitions are not

sufficient to describe new technologies

• Use of workflow management will be a focus for the future

• Some definitions needed to be strengthened or restructured for the best utility

DICOM Second Generation Radiotherapy

Page 10: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Why now 2nd Generation?

Advances in Radiotherapy• Modern Clinical Flow• Continuous Adaptation to accommodate Disease Response• Dynamic Processes need Flexibility in Interoperability

• New Treatment Modalities

• Stronger Capability to represent the Clinical Workflow• Systematic approach for Non-Volumetric and Volumetric Information• Design of Modality-Independent versus Modality-Specific Information

• Electronic Workflow Management

Modernization, Efficiency, Systematics(not handled in this Presentation, e.g. Macros as Building Blocks, Optionality, Use of Latest

Standard Capabilities, Integration of non-RT Objects)

DICOM Second Generation Radiotherapy 10

Page 11: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Process Perspective was Too Narrow• 1st Generation:

Result of Treatment Planning System

• 2nd Generation:

Incorporating complete Treatment Workflow

11

Focus

DICOM Second Generation Radiotherapy

Page 12: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

2nd Generation Object Overview

12DICOM Second Generation Radiotherapy

Page 13: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Course

13

• Represents an RT Treatment Course

• May contain several treatment phases

• Is a container collecting all major objects which are relevant to this course

DICOM Second Generation Radiotherapy

Page 14: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Physicians Intent

14

• Describes how to achieve the curative or palliative therapy

• Total and fractional doses, fractionation schemes, treatment sites, dosimetric objectives and other information is included

DICOM Second Generation Radiotherapy

Page 15: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Conceptual Volume

15

• Description of an anatomical region or point, which may or may not be associated to one or more volumetric representations

DICOM Second Generation Radiotherapy

Page 16: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Segment Annotation

16

• Represents RT specific annotations on a Segmentation, Surface Segmentation or RT Structure Set

DICOM Second Generation Radiotherapy

Page 17: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Radiation Set

17

• Defines a radiation therapy treatment fraction, through a set of RT Radiations, which may be applied one or more times.

• Now handles newer technologies, and is extensible.

DICOM Second Generation Radiotherapy

Page 18: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Radiation Record

18

• Records actual treatment parameters and attributes

• Now handles newer technologies and is extensible.

DICOM Second Generation Radiotherapy

Page 19: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Dose Image

19

• Represents a 3-dimensional dose distribution using the multi-frame and functional group paradigms.

DICOM Second Generation Radiotherapy

Page 20: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Dose Histogram

20

• Describes a dose-volume histogram

• Based on a volumetric dose calculation

• References a Conceptual Volume and an RT Segment Annotation

DICOM Second Generation Radiotherapy

Page 21: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Dose Samples

21

• Represents dose point data, which are calculated or measured.

DICOM Second Generation Radiotherapy

Page 22: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Treatment Phase

22

• A Course can be divided into multiple RT Treatment Phases

• Each Phase represents an attempt to reach a treatment goal via RT Radiation Sets

DICOM Second Generation Radiotherapy

Page 23: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

• Workflow Enabled• Because we have dedicated objects• Still possible to work with or without DICOM

Worklist

• Volatile Information outside objects• Status information• Dynamic Relations

23

Solution: Dynamics

DICOM Second Generation Radiotherapy

Page 24: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

• Use new approaches already defined in DICOM

• Volume and Surface Segmentation• Rigid and Deformable Registrations• Some smaller Items (esp. Macros)

• Object-oriented approach• Cohesive objects with different, specific intents• Avoid “Monster” objects, or inappropriate

encapsulation (Plan/RX)

Architectural Principles I

24DICOM Second Generation Radiotherapy

Page 25: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Architectural Principles II

• Flexibility for future expansione.g. open for

• New treatment machines• New segmentation methods• New treatment techniques• New positioning and tracking approaches

25DICOM Second Generation Radiotherapy

Page 26: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Supplement 147 – Prescription and Segment Annotation

26DICOM Second Generation Radiotherapy

Page 27: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

DICOM RT Plan Issues

27

DICOM RT Plan

Prescription

Fractionation

Beams

Patient Setup

• Prescription Module is not precisely documented

• Cannot exist outside an RT Plan object

• Does not represent current technologies well

• Links to volatile information (Imaging)

DICOM Second Generation Radiotherapy

Page 28: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Physician Intent

Contains the following parts:• Physician Intent

• Mostly free-text definition of the physician‘s intent based on a diagnosis

• Prescription• Actual prescription parameters

• Treatment Phase Intent• How the time structure of the delivery is intended

• Exists outside of the RT Plan• Supports real-world timing of Prescription information

28DICOM Second Generation Radiotherapy

Page 29: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Physician Intent Modules

29

IE Module Reference Usage

Include 'RT Second Generation IOD Modules Macro' Table A.VV.1.1.1-1

RT Physician Intent RT Physician Intent C.AA.B1 M

RT Prescription C.AA.B2 U

RT Treatment Phase Intent

C.AA.B3 C

Required if RT Treatment Phase Intent Presence Flag

(30xx,0808) equals YES.

DICOM Second Generation Radiotherapy

Page 30: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Physician Intent Modules

30DICOM Second Generation Radiotherapy

Page 31: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Conceptual Volume (CV)

• Abstract object

• Identifies anatomic region

• May be a combination of other CVs

• Can be utilized without any Segmentation

• Definition can be deferred to later in workflow

31DICOM Second Generation Radiotherapy

Page 32: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Conceptual Volume Modules

32

Page 33: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

CV Example

33

CVol I”Tumor“

RT Structure Set- Region of Interest 1,

“Lesion“

Diagnostic CT

MR

Physician Intent:Prescribe 20Gy to the

tumor.

DICOM Second Generation Radiotherapy

Page 34: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

CV Consistency vs. Change

CV I“Tumor“

RTStructSet 1- ROI 1, “Lesion“

MR CBCT 1

RTStructSet 2- ROI 4, “Target“

CBCT 2

RTStructSet 3-ROI 2, “PTV“

First Day Day 10 Day 2034

Page 35: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

CV Combination

35

CV I“Lung, left“

CV II“Lung, right“

RT Structure Set- Region of Interest 5,

“Left Lung“

Segmentation Storage-Segment 9,

“Lung, r“

CV III“Lungs“

DICOM Second Generation Radiotherapy

Page 36: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Segment Annotation Modules

36DICOM Second Generation Radiotherapy

Page 37: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

RT Segment Annotation

• Separates the RT-specific information from the actual data

• CTV, PTV, OAR, …• Bolus• Density Overrides

• Structure is represented by existing DICOM objects

• Segmentation Storage (3D volume)• Surface Segmentation Storage (3D mesh surface)• RT Structure Set (Contour lines – Gen1)• Others

37DICOM Second Generation Radiotherapy

Page 38: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Other Supplements

• 175 – C-Arm Radiations• 176 – New RT Radiations• 177 – Dose Objects• 178 – RT Course• 179 – RT Explanatory (Part 17)

38DICOM Second Generation Radiotherapy

Page 39: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

How does it affect you?

• Once these objects appear, manufacturers will need to support both Gen1 and Gen2 during the transition

• 8 old objects deprecated • 24 new objects added

• The sooner the Gen2 objects are in place, the earlier support for Gen1 can be reduced (or even stopped)

39DICOM Second Generation Radiotherapy

Page 40: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

How does it affect you?

• Enhanced Clinical Capability will need Gen2

• Detailed prescription sharing • Prescribed Volumes cross-referenced throughout the clinical

process for tracking and adaptation• Complex or high resolution structure sharing• Highly adaptive positioning or treatment

40DICOM Second Generation Radiotherapy

Page 41: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

What will NEMA do to help?

• Provide introduction presentations

• Run hands-on engineering classes in Europe and North America

• Work with IHE-RO to develop the order of adoption for most benefit

• “The Implementer‘s Guide to 2ndGeneration RT Objects“ by WG-07 members

41DICOM Second Generation Radiotherapy

Page 42: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Contact

Stephen VastaghDICOM Secretary NEMA/MITA

[email protected]

Ulrich BuschChairman DICOM WG-07

[email protected]

Christof SchadtVice-Chair DICOM WG-07

[email protected]

42DICOM Second Generation Radiotherapy

Page 43: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Links

• NEMA DICOM Pagehttp://dicom.nema.org

• Participate in DICOMhttp://dicomconference.org/

• DICOM Standard Statushttp://www.dclunie.com/dicom-status/status.html

43DICOM Second Generation Radiotherapy

Page 44: DICOM Second Generation Radiotherapy Objects Part 1 Supplement 147: Physicians Intent and Segment Annotation.

Thanks to

• Christof, Jim Percy (Elekta), Mike Courtney (Philips), Mark Pepelea (Philips), Bruce Rakes (Mevion) for preparing and reviewing the initial content for an AAPM presentation, on which this is based

44DICOM Second Generation Radiotherapy