Top Banner
Development of Instruments for Real-time Water Monitoring S. Mitra New Jersey Institute of Technology Newark, NJ 07102 [email protected] Acknowledgements: US EPA, NSF, US Army
24

Development of Instruments for Real-time Water Monitoring

Dec 18, 2021

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Development of Instruments for Real-time Water Monitoring

Development of Instruments for

Real-time Water Monitoring

S. Mitra

New Jersey Institute of Technology

Newark, NJ 07102

[email protected]

Acknowledgements: US EPA, NSF, US Army

Page 2: Development of Instruments for Real-time Water Monitoring

On-line Real-time Monitoring

Process

Or

Process Stream

Instrument

or sensor

Page 3: Development of Instruments for Real-time Water Monitoring

Real-time Monitoring of VOCs Ground Water Using Gas Injection Membrane Extraction (GIME)

Gas

Pulse

Heating

Nitrogen

Injector

Membrane Module

Portable GC

Detector

Microtrap

Sample Out

Sample In

Data System

Sample out

1. K. Hylton and S. Mitra. J. Chromatogr. A, 2007, 1152, 199–214.

2. D. Kou and S. Mitra, Anal. Chem. 2001

Page 4: Development of Instruments for Real-time Water Monitoring

Continuous Monitoring by GIME

Compounds MDLs (ug/l) RSD (%)

Benzene 0.1 1.7

Toluene 0.1 2.3

Ethylbenzene 0.9 2.8

Page 5: Development of Instruments for Real-time Water Monitoring

Field Testing of Gime-GC System at a Superfund Site

The on-line real-time monitoring system A groundwater treatment facility at the Naval

and Engineering Station in Lakehurst, NJ,

where the field study was conducted.

Page 6: Development of Instruments for Real-time Water Monitoring

I1 I4I2 I5I3 I6 I7 I9I8

0 3618 9054 10872 126 144

Time (min)

1

2

3

1. 1,1 Dichloroethylene

2. cis 1,2 Dichloroethylene

3. Trichloroethylene

Typical chromatograms from on-line analysis of water entering the treatment facility. Injections were made at I1, I2 etc.

0.00

0.50

1.00

1.50

2.00

2.50

3.00

0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00

Time (min)

Co

ncen

trati

on

(p

pb

)

cis 1,2 Dichloroethylene

0.00

0.20

0.40

0.60

0.80

1.00

1.20

1.40

0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00

Time (min)

Co nc en tra tio n (p pb )

Trichloroethylene

Concentration Profiles over a six hour monitoring period

Continuous Monitoring by GIME-GC

Page 7: Development of Instruments for Real-time Water Monitoring

Steps in Trace Semi Volatile Measurements

Extraction -LLE, SPE, SPME

Concentration -Evaporative, Membrane

Detection -GC, HPLC, MS

Page 8: Development of Instruments for Real-time Water Monitoring

Membrane Extraction

Membrane Feed side Permeate side

Page 9: Development of Instruments for Real-time Water Monitoring

Schematic Diagram of ME-LC

Organic

Extractant

Water

Sample

Detector

HPLC mobile phase

HPLC Column

Solvent Exit

Analyte

Laden

Solvent

Sample Exit

Hollow Fiber

Membrane Module

Injection Valve

X. Guo and S. Mitra, Journal of Chromatography A,904 (2000) 189-196.

Page 10: Development of Instruments for Real-time Water Monitoring

Continuous monitoring by ME-LC

X. Guo and S. Mitra, Journal of Chromatography A, 904 (2000) 189-196

Page 11: Development of Instruments for Real-time Water Monitoring

Simultaneous Extraction and Concentration

Analyte

Acceptor (solvent)

Donor

(water)

Membrane

Extract

Solvent

Analyte

Solvent

Donor

D. Kou and S. Mitra, Analytical Chemistry (2003).

Page 12: Development of Instruments for Real-time Water Monitoring

Solubility and Solvent Loss

Kow

(Log P)

Hexane BA MIBK

IPA

EE EF EE EF EE EF EE EF

PCP 5.12 68.1 36.3 69.5 101.6 43.7 153.3 26.1 203.4

Atrazine 2.61 6.5 3.5 56.2 82.2 N/A N/A 31.6 246.6

Naphthalene 3.30 46.3 24.6 81.3 119 33.9 119 37.4 292

Water Solubility

9.5 mg/l 8.4 g/l 19 g/l 30.9 g/l

Solvent

Loss (%)

26.7 73.3 88.9 95

Ref: D. Kou and S. Mitra, Analytical Chemistry (Nov, 2003)

Page 13: Development of Instruments for Real-time Water Monitoring

Analytical Performance

Compounds MDL (μg/L ) RSD (%)

Atrazine 0.5 4.6

PCP 1.0 7.8

Naphthalene 0.9 6.3

Page 14: Development of Instruments for Real-time Water Monitoring

Total Analyitical System

Integrating Extraction and Concentration

Page 15: Development of Instruments for Real-time Water Monitoring

Extraction with On-line Membrane

Concentration

membrane concentration

Organic solvent

Water sample

Water sample exit

Inert gas

Gas out

Injection valve HPLC

membrane extraction Extract

Concentrated extract

Wang and Mitra, Journal of Chromatography A, 1068 (2005), 237-242

Page 16: Development of Instruments for Real-time Water Monitoring

Wang and Mitra, Journal of Chromatography A, 1068 (2005), 237-242

Page 17: Development of Instruments for Real-time Water Monitoring

Total Analytical System

Organic

Extractant Sample Exit

Extraction Module

Detector

HPLC Mobile Phase

HPLC Column

Injection Valve

Water

Sample

Injec. Loop

Overflow

Concentration

Module

N2

N2 Stripping

Flow Analyte

Laden

Solvent

Extraction Concentration Detection

Wang and Mitra, Journal of Chromatography A, 1068 (2005), 237-242

Page 18: Development of Instruments for Real-time Water Monitoring

Mass transfer in the TAS

Wang and Mitra, Journal of Chromatography A, 1068 (2005), 237-242

Page 19: Development of Instruments for Real-time Water Monitoring

Comparison of EF

in three experimental modes

Mode 1:membrane extraction only;

Mode 2: membrane extraction and pervaporation without N2 stripping;

Mode 3: and membrane extraction and pervaporation with a N2 flow rate of 45 mL/min. Naphthalene

Biphenyl

Mode 1

Mode 2

Mode 3

0

20

40

60

80

100

120

140

160

180

200

EF

Page 20: Development of Instruments for Real-time Water Monitoring

Real Time Monitoring of Haloacetic Acids

Page 21: Development of Instruments for Real-time Water Monitoring

Continuous SLME-HPLC

Wang, Kou and Mitra.Journal of Chromatography A (2005).

Page 22: Development of Instruments for Real-time Water Monitoring

Series of Chromatograms from Continuous

SLME of the Nine HAAs

Page 23: Development of Instruments for Real-time Water Monitoring

Lab-on-a-chip, Total Analytical System

Sample

Extraction

Sample

Concentration

Separation/

Processing

Detection

Page 24: Development of Instruments for Real-time Water Monitoring

On-Chip SLMME