New England Medical Center New England Medical Center Respiratory Care Programs Respiratory Care Programs Changing Compressed Gas Cylinders Changing Compressed Gas Cylinders Training Module Training Module Prepared by William R. Howard, Prepared by William R. Howard, MBA, RRT MBA, RRT Director, Respiratory Care Director, Respiratory Care
34
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Cylinder Chang 97

New England Medical CenterNew England Medical Center Respiratory Care ProgramsRespiratory Care Programs

Changing Compressed Gas CylindersChanging Compressed Gas Cylinders

Training ModuleTraining Module

Prepared by William R. Howard, MBA, RRTPrepared by William R. Howard, MBA, RRT

Director, Respiratory CareDirector, Respiratory Care

Page 2: Cylinder Chang 97

            

This module is intended to:

Provide background information in the changing of compressed gas cylinders

Identify the steps to be demonstrated in the changing of compressed gas cylinders

Cylinder Changing:Training Module

Page 3: Cylinder Chang 97

            

Cylinder Changing:

A gas cylinder should be changed when the contents, as identified by it’s pressure, decreases to about 500 psig.

Page 4: Cylinder Chang 97

            

Cylinder Changing:

Empty Cylinders

Cylinders are never truly "empty."

Empty cylinders shall be handled in the same manner as full and partially full cylinders. 

Page 5: Cylinder Chang 97

            

Cylinder Changing:

Empty Cylinders

Full and empty cylinders are not to be connected or in the same manifold together, (or stored together).

Page 6: Cylinder Chang 97

            

Cylinder Changing:

Empty Cylinders

Empty cylinders are to be labeled:

"Empty"

Page 7: Cylinder Chang 97

            

Cylinder Changing:

Empty Cylinders

Valves are closed on empty cylinders, leaving a positive pressure inside of them.

Valve protective caps, (received with full cylinders), are replaced on empty cylinders.

Page 8: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Identify from your supply room, a full cylinder of gas for your specific needs.

Page 9: Cylinder Chang 97

YOKE STEM

            

Cylinder Changing:

Specific Procedures:

Remove protective cap and/or valve covering material from yoke stem.

Page 10: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

With E-Cylinders and smaller, be sure that a NEW sealing grommet is installed between the yoke stem and the receiving end of the regulator. The OLD grommet must be removed and discarded.

Page 11: Cylinder Chang 97

Grommet installed on regulator:

Cylinder Changing:

Specific Procedures:

Page 12: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

There is no sealing grommet with larger cylinder or threaded ‘nut’-style regulators.

Page 13: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

The cylinder valve is opened momentarily and then closed to blow dirt from the outlet.

                              

Page 14: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

If the gas you are selecting is toxic/poisonous, crack in the safety of an exhaust hood or in a place of adequate ventilation.

                              

Page 15: Cylinder Chang 97

            

Cylinder Changing -

Specific Procedures:

The outlet of an oxygen cylinder valve is never wiped or touched.

This avoids leaving organic residues that might be ignited by exposure to high oxygen pressure, for example.

Page 16: Cylinder Chang 97

            

Cylinder Changing -

Specific Procedures:

Oil or grease must be avoided on the high-pressure side of oxygen cylinders or other cylinders containing oxidizing material. (Otherwise a fire or explosion could

result.)

Page 17: Cylinder Chang 97

            

Cylinder Changing -

Specific Procedures:

Cylinders are only to be used with a regulator.

A regulator lowers the cylinder pressure to safe level.

Page 18: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

A regulator and gauge shall be installed at the point of use to show:

The outlet pressure when the source cylinder is used.

Page 19: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

A regulator and gauge shall be installed at the point of use to show:

The remaining contents of the cylinder

Page 20: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures :

Cylinder valves are closed when not in use.

They are never tampered with, forced, lubricated, or modified.

Page 21: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Cylinders are used only with fittings, valves, regulators, and tubing designated by the manufacturer for the gas being used

Page 22: Cylinder Chang 97

            

Cylinder Changing -

Specific Procedures:

Connections are not forced or made from homemade adapters.

Page 23: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Wear appropriate eye protection

“Crack” cylinder to remove dust.

Page 24: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Attach the closed regulator to the cylinder.

Never open the cylinder valve unless the regulator is completely closed and tight.

Page 25: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

T-or Yoke-style regulators are to be hand tightened only

Large-tank or nut-threaded regulators, (those without sealing grommets), are to be wrench tightened.

Page 26: Cylinder Chang 97

             

Cylinder Changing:

Specific Procedures:

If using a ‘yoke’ style regulator, common to E-size cylinders, do not use the wrench to tighten the regulator onto the cylinder.

Page 27: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Face the pressure gauge away from you when opening cylinder valve.

Page 28: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Identify that the pressure gauge reads full :

Full for most non-liquid cylinders is 2,000-2,200 psig.

Turn the cylinder valve off.

Page 29: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

If the contents gauge does not decline, then the connection is ‘tight’.

Page 30: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Open the administration flowmeter to release pressure.

Close the flowmeter.

            

Page 31: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

The full cylinder is now ready for use.

Follow your Department procedure for appropriate leak-free connection to your equipment.

Page 32: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

The full cylinder is now ready for use.

To deliver therapeutic gas to a patient, the clinician will need to connect to a flow-meter that is specific to the cylinder gas.

Page 33: Cylinder Chang 97

            

Cylinder Changing:

Specific Procedures:

Only licensed personnel under physician authority may dispense or administer therapeutic gases to a patient.