Issued: 2-Sep-2019 Revision: 3.4 Copyright: 2019 Agilent Technologies Select pages for required products Agilent InfinityLab LC Series Installation Checklist Agilent InfinityLab LC Series Installation Checklist Thank you for purchasing an Agilent instrument solution. This checklist is used by the installing engineer to ensure that the instrument and associated systems are correctly installed, upgraded and functioning as designed in your facility. This checklist will be completed at the end of the service and provided to you as a record of the installation. Customer Information Customers should provide all necessary operating supplies upon request of the engineer. Service Engineer’s Responsibilities Only complete/printout sections or pages that relate to the system that has been installed. Complete empty fields with the relevant information. Complete the relevant checkboxes in the checklist using a “X” or tick mark “ ” in the checkbox. Complete Not Applicable check boxes to indicate services not delivered, as needed. Complete the Service Review section together with the customer. In case of diagnostic tests, record test results in section service review (test results). System Information Guidance: Check box if instrument configuration report is attached instead of completing the table below. Instrument System Name/I.D: Instrument Location: Record the list of system component product numbers below. List the serial numbers of the components present in the system below. 1. 1. 2. 2. 3. 3. 4. 4. 5. 5. 6. 6. 7. 7. 8. 8. 9. 9. 10. 10.
24
Embed
Customer Information Service Engineer’s …...The installation of ELSD drivers depends upon the customer's choice of software platform. If OpenLAB version A.02.01 is installed then
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Thank you for purchasing an Agilent instrument solution. This checklist is used by the installing engineer to ensure that the instrument and associated systems are correctly installed, upgraded and functioning as designed in your facility. This checklist will be completed at the end of the service and provided to you as a record of the installation.
Customer Information
Customers should provide all necessary operating supplies upon request of the engineer.
Service Engineer’s Responsibilities
Only complete/printout sections or pages that relate to the system that has been installed.
Complete empty fields with the relevant information.
Complete the relevant checkboxes in the checklist using a “X” or tick mark “ ” in the checkbox.
Complete Not Applicable check boxes to indicate services not delivered, as needed.
Complete the Service Review section together with the customer. In case of diagnostic tests, record test results in section service review (test results).
System Information
Guidance:
Check box if instrument configuration report is attached instead of completing the table below.
Instrument System Name/I.D: Instrument Location:
Record the list of system component product numbers below.
List the serial numbers of the components present in the system below.
Unpack/verify the condition and completeness of shipment. For discrepancies, use the table below.
Product/Part Description
Observation Action
Discuss any specific questions or issues with the customer prior to starting.
Discuss any configuration options with the customer prior to starting.
Check for required service note applicability and firmware updates if connecting to instruments.
When plumbing the customers system, make sure that the relevant capillary kit and instructions are used.
If you use the PC or the Instant Pilot for firmware updates, use the correct date/time.
Upgrades only - Ensure with customer that instrument control settings, data, methods etc. have been properly saved or archived prior to starting any installation procedures.
The installation of ELSD drivers depends upon the customer's choice of software platform. If OpenLAB version A.02.01 is installed then ELSD drivers' version 1.2 is installed as part of the installation. An UPGRADE to 1.3 will be seen during the new install. If a THIRD PARTY software package is used a minimum of ICF A.02.02 is required.
Agilent Information Center Install the Agilent Information Center to the PC controlling the system.
OR
If this is not permitted, install the Agilent Information Center to a different PC.
OR
If this is also not permitted, use the installation medium ( Agilent InfinityLab LC Series User Documentation (G4800-64600)) to view the required user documentation.
Record the location to which the Agilent Information Center has been installed.
OR
If the Agilent Information Center has not been installed, record that the installation medium has been delivered, and keep it close to the instrumentation for future reference.
Section NOT Applicable Connect solvent inlet tubings.
Connect waste and seal wash sensor outlet tubings.
Connect remote control cable (if applicable).
Connect CAN cable.
Connect LAN cable (if applicable).
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Fill each channel with isopropanol from the solvent bottle to the internal degassing unit inlet.
Purge each channel for 3 minutes at 3 mL/min.
Condition the pump.
Run the pump in normal operation and ensure the tuning is between ±0.2 .
Connect remote control (ERI) cable (if applicable).
Connect seal wash tubing (if applicable).
Connect CAN cable.
Connect LAN cable (if applicable).
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Install the solvent cabinet (if applicable).
Prepare HPLC grade water for Channel A (A1).
Prepare HPLC grade acetonitrile for Channel B (B1).
Prepare HPLC grade isopropanol for Channels C&D (A2&B2).
Turn on the pump.
Purge each channel for 3 min at 3 mL/min.
After completing the stack installation, purge the complete system.
G7129A/B/C Vialsampler Section NOT Applicable Remove transport foam.
Set up the Vialsampler in the required stack configuration.
Connect capillaries.
Connect tubings for needle wash.
Connect tubings for waste/leak handling.
Install the Multidraw Kit (if applicable).
Mount the Column Identifier Kit (if applicable).
Mount the External Tray (if applicable).
Connect remote control (ERI) cable (if applicable).
Connect CAN cable.
Connect LAN cable (if applicable).
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Prime the needle wash tubings with an appropriate wash solvent (for example, 10 % isopropanol in water).
G7130A Integrated Column Compartment (ICC) Section NOT Applicable
Connect the Integrated Column Compartment to the Vialsampler.
Note down the part number and serial number of the ICC:
Install wash solvent inlet (peristaltic pump in case of Standard Hydraulic Box, or SSV in case of Multiwash Hydraulic Box).
Install Drawer Kit(s) (if applicable).
All positions must be filled either with dummies or drawers. Drawers must be installed from the bottom to the top.
Install Multidraw Kit (if applicable).
Connect waste tubing (it must be in a straight line down, no bends or kinks).
Connect remote control (ERI) cable (if applicable).
Connect CAN cable.
Connect LAN cable (if applicable).
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Turn on Multisampler and perform Auto Referencing before stacking other modules on top.
Connect capillaries.
Fill the available channels with the appropriate wash solvent from the solvent bottles.
Configure "Prime" for the available channels.
Activate "Prime" for the available channels.
Perform wellplate assignment (Assign Wellplate in Chromatography Data System or Set Plate Type in Agilent Lab Advisor Software) (if applicable).
Perform Auto Referencing under operating temperature, after stacking other modules on top.
Turn on the detector and lamp as soon as possible (one hour minimum warm-up time required).
Install solvent inlet/outlet tubings.
Install waste tubing.
Install remote control cable (if applicable).
Connect analog output cable (if applicable).
Connect CAN cable.
Connect LAN cable.
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Turn on the detector and lamp as soon as possible (one hour minimum warm-up time required).
Install solvent inlet/outlet tubings.
Install waste tubing.
Install remote control cable (if applicable).
Connect analog output cable (if applicable).
Connect CAN cable.
Connect LAN cable.
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Turn on the detector and lamp as soon as possible (one hour minimum warm-up time required).
Install solvent inlet/outlet tubings.
Install waste tubing.
Install remote control cable (if applicable).
Connect analog output cable (if applicable).
Connect CAN cable.
Connect LAN cable.
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Section NOT Applicable Install solvent inlet/outlet tubings.
Install waste tubing.
Install remote control cable (if applicable).
Connect analog output cable (if applicable).
Connect CAN cable.
Connect LAN cable.
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Flow cell:
New instruments are shipped with a flow cell installed and have a factory wavelength calibration.
If a flow cell other than the installed/shipped flow cell is used, then:• Verify that the installed flow cell is fixed tightly.
• Perform a Wavelength Calibration using Glycogen (part of FLD Calibration Kit (G7121-68001)).
Perform Lamp Intensity Test and compare profile to a reference profile listed in the Service Manual.
Perform Wavelength Accuracy Test.
Lamp Intensity Done
Wavelength Accuracy Excitation Deviation
Emission Deviation
If the test fails, perform Wavelength Calibration (if applicable).
Rerun Wavelength Accuracy Test if Wavelength Calibration has been done (if applicable).
Section NOT Applicable Connect an inlet capillary.
Depending on the detector, different inlet tubings are used (from the Accessory Kit):• G7162A:
• Interfacing capillary (G1362-87300)
• G7162B:
• Capillary ST 0.075 mm x 220 mm (5067-4784) (Sampler to MCT/TCC)
• Capillary ST 0.075 mm x 340 mm (5067-4783) (Column to RID)
Turn on the detector and the heater of the optical unit as soon as possible (one hour minimum warm-up time required).
Install solvent inlet/outlet tubings.
Install solvent recycle.
Install waste tubing.
Waste container needs to be placed above the detector.
Connect analog output cable (if applicable).
Connect CAN cable.
Connect LAN cable.
LAN connection is made between at least one of the Agilent Infinity II modules and the Control PC. If a detector is installed, connect the LAN to this module. If there are multiple detectors with spectral capabilities, consider using additional LAN connections for each detector.
Check Optical Balance (from the RID Tool Screen in Lab Advisor).
Section NOT Applicable Verify that the instrument is the correct module for the regional voltage (120 V/240 V). Do
NOT use instruments which have an incompatible voltage.
Connect the exhaust line to a fume hood or lab extraction line.
Set the correct pressure at the regulator for the pressurized nitrogen line and connect it to the detector.
Install the capillary connection from detector to HPLC.
Install the RS-232 or LAN connection from detector to ChemStation control PC (if applicable).
Install the Aux I/O cable connection from detector to the 1200 stack
Connect the power lead.
Connect the front drain to the waste container/bottle.
Configure the detector in OpenLAB CDS ChemStation Edition or EZChrom Edition. "OpenLAB CDS ChemStation and EZChrom Edition, ELSD Driver version A.01.05 Minimum".
G7102A, G4260B Checkout Switch on the instrument.
Prepare the instrument:
Set the detector nebulizer and evaporator temperature to 70 °C , and gas flow to 1.60 SLM.
Wait until instrument ready status is displayed on front panel or PC.
Record 10 min of baseline noise. This is gas only noise and should be typically ≤0.5 LSU or ≤0.5 mV for G4260B (ASTM baseline noise).
Set up a method for the checkout sample as below:
Flush the system with eluent for 15 min AND
Verify instrument ready status is still displayed on front panel or PC.
In the instrument or system, set up a method using the following parameters:
• Pump: 1.00 mL/min of 100 % water.
Depending on the flow rate setting, the system may overpressure and/or the PEEK fittings may not hold if pressures exceed 350 bar. If this happens, bypass the restriction capillary and guard column (usually for flow rates beyond 2.0 mL/min for UHPLC or 3.0 mL/min for HPLC).
• Injection Volume: 20.0 µL.
The following parameters work well for the recommended default injection volume of 20.0 µL. However, a scouting run is always recommended because run time varies across systems.
• Evaluation Std.: 200 µg/ml caffeine standard (part# 5190-0488).
Attach available reports or printouts to this documentation.
Record the time and date of installation or upgrade completion in the customer’s records or logbook.
Complete the Service Engineer Comments section below if there are additional comments.
Review the installation or upgrade with the customer.
Explain Agilent warranty for instruments.
Explain how to use manuals, guides, online help.
Explain how to get self-help, FAQs from the web.
Explain how to log an instrument service call, support services available.
Advise customer of additional instrument training options.
If the instrument firmware was updated, record the details of the change in the Service Engineer Comments box below or if necessary, in the customer’s IQ records.
Supply the customer with a copy of the Smart Alerts flyer.
If there are any specific points you wish to note as part of performing the installation or other items of interest for the customer, please write in this box.
Firmware was changed (Y/N):
Client was consulted and agreed prior to firmware change (Y/N):
Service Completion
Service Request number…………………. Date service completed………………………