Top Banner
PANIC11, 7/24/2011 LUX Experiment A. Lyashenko Current status of LUX Dark Matter Experiment by A. Lyashenko Yale University On behalf of LUX collaboration
25

Current status of LUX Dark Matter Experiment

Dec 21, 2016

Download

Documents

ngokhue
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Current status of LUX Dark Matter Experiment

by A. Lyashenko

Yale UniversityOn behalf of LUX collaboration

Page 2: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

LUX – Large Underground Xenon experiment

LUX Collaboration:Yale, CWRU, UC Santa Barbara,Brown, TAMU, UC Davis, Harvard,LLNL, LIPP Coimbra, Rochester, LBNL, Maryland, South Dakota SMT

Page 3: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

WIMP Dark matter

ρχ=0.3±0.1GeV∙cm-3

~3 WIMPs in 1L

Local Milky Way DM density:

Xe

Low Energy Threshold is vital

0-80keVR Recoil Energy

Assuming ΩDM = Ωχ WIMP DM

WIMP: stable, non-barionic, non-relativistic, el. neutral

mχ=10 GeV-few TeV scatter preferably from nuclei

Page 4: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

WIMP direct detection: two phase Xe

WIMP

S2/S1Gamma>>S2/S1WIMP

Gamma

Drift time

Top PMT Array

Bottom PMT Array

Page 5: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

LUX Detector

Top PMT Array

Bottom PMT Array

• 350 kg LXe (100 kg fiducial)• Ed=2kV/cm, Eextr=5kV/cm, Esc=10kV/cm• 122 PMTs (QE~30% @ 175nm)• Low Radioactivity Materials• >1kW Cooling Power (Thermosyphons)

Titanium Cans

Field Cage andTeflon ReflectorPanels

Thermosyphon

2” Hamamatsu R8778 PMTs

Water Tank

Detector Stand

59cm

49cm

Page 6: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Challenge: Backgrounds

Goal: few events/100kg/year !!!

• Sea level total muon flux: 55.2 m-2∙sec-1 (threshold 300 MeV)

• Ambient radioactivity: ~100 evts/kg/sec

• Human gamma activity 40K: ~4000 gammas/sec

• Walls (U/Th/K): - concrete 25/5.5/640 Bq/kg- rhyolite rock 100/45/900 Bq/kg

• Internal radioactivity

WIMP

Where isWIMP?

Page 7: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Trickery: Going Deep

LUX

4850 feet (1478 m)underground

HomestakeMine

55.2 m-2∙sec-110-5 m-2∙sec-1

~107 fold muon flux suppression

Sanford Underground Research Facility (SURF)

Page 8: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Trickery: Water Shield

Gamma suppression: ~10-7

Neutron suppression (>10 MeV): ~10-3

Neutron suppression (<10 MeV): <10-9

d = 8 m

6 m

20 Tonnesinverted steel pyramid

• 300 Tonnes of Water• Cherenkov ‘Muon Veto’ 20 Ø10” PMTs

Page 9: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Trickery: Understanding internal backgrounds

PMTs: 10/2/65 mBq/PMT (U/Th/K) and 2 n/year/PMTDominant source of background!!!

Titanium cryostat: <0.4 mBq/kg U+Th (very low!)

Cu: 10 ppt U / 12 ppt Th / 18 ppb K [~1% of ER BG budget]

Radon: Cleanroom reduces levels to <40 Bq/m3, Radon Cowl

Kr: present in Xe at ppm level <2 ppt level with charcoal column

PTFE: U/Th/K < 0.004/<0.001/0.053 mBq/kg 6∙10-8

events/keVee/kg/day [~1% of ER BG budget]

X 122

Page 10: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Trickery: Xe self shielding

Expect <0.5 nuclear/electron-recoils in 100 days

PMT gammas

PMT neutrons

Fiducial volume

PMTs are dominant background source

We benefit a lot from scaling up

10 kg

300 kg

Page 11: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Trickery: ER/NR Discrimination

WIMPregion

XENON10 data

Electron Recoils

Nuclear Recoils

Cut 2-3 keVee

High energies 99.5%

Low energies 99.9%

Page 12: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

83mKr (half-life 1.86 hours)

• alpha source• S1 LY and keVee, e- lifetime• Fills entire fiducial volume

LUX calibrationExternal Sources Internal Sources

133Ba, 137Cs, 22Na, 208Tl (collimated):S1 LY, ER energy bands, e- lifetime241Am/Be: NR band

S1

9.4 keV 32.1 keV

Tritium (half-life 12.3 yrs)• Injected as CH3T• Beta source, up to 18.6 keV• Removed by purification system• S2/S1 ER band• Fills entire fiducial volume

L. Kastens et al., Phys. Rev. C80: 045809,2009,A. Manalaysay et al., Rev. Sci. Instr. 81, 073303 (2010)

Page 13: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

LUX Program Timeline

LUX 0.1

LUX Surface Run LUX DM Search Run

2007 - 2009

2010 - 2011

NOW!!!

2012+

Page 14: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Sanford Lab Surface Facility

Page 15: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

May 2011 first cooldown

Held Target Temperature 180K

Flexible Temperature Control

Page 16: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Expectations

WIMP events

LUX 4000 kg-days (MC simulation)

LUX – strong emphasis on WIMP discovery

LUX ~2days

LUX ~60days

LUX ~300days

Targeting at 7∙10-46 cm2 (@ 100GeV)

Page 17: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Future Plans: LZS/LZD ProjectsLZ = joint collaboration of LUX and ZEPLIN

LZS (Sanford) =1500 kg (1200 kg fiducial)

• LUX infrastructure designed to accommodate LZ

• Liquid scintillator veto

• Construction 2012–2013, operation 2014-2015

• σχ = 2x10-47 cm2 in 2 years

LZD = 20 ton concept

• 2 m diameter x 2 m height

• Will be in a deeper site, possibly 8000 feet

• Construction 2014–2016, operation 2017–2020

• σχ = 10-48 cm2 in 3 years

LUX

LZD

Page 18: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

• Dark Matter puzzle convergence of astronomy and particle physics– Evidences point to WIMPs waiting to be discovered

• LUX experiment uses proven liquid Xe technology to reach σχ = 7x10-46 cm2

– Surface operation is in progress

– Dark Matter search run will start in 2012

– Just when the LHC will be searching for New Physics

• LZS/LZD will push the sensitivity to 10-48 cm2 by 2020

• We may know what Dark Matter is within this decade!!!

Summary

Page 19: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Extra Slides

Page 20: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Xe signal generation

Page 21: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Why WIMPs?

Page 22: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

NR scintillation efficiency

Eee

A. Manzur et al., arXiv:0909.1063v4 and references therein

Page 23: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

Tritiated methane (CH3T) removal experiment –Gaseous proportional tube counter

Xe proportional tubefor CH3T counting Xe purifier

• > 99.9% of CH3T removed from proportional tube in one pass through the purifier.• Liquid xenon removal test @ UMD this spring.

CH3T storage bottle

FE-55 x-raycalibration

Univ. of Maryland

Page 24: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko15

Case Western Reserve University

Page 25: Current status of LUX Dark Matter Experiment

PANIC11, 7/24/2011 LUX Experiment A. Lyashenko

1.5 ppt Kr

open leak valve

New analytic technique to detect krypton at the part-per-trillion level

arXiv:1103.2714v1

GOAL: O(10) ppt 85Kr- Leads to an event rate of O(0.1) events in the 2-10 keVee regime after 104 kg-days and 99.5% rejection of e-

recoils

16

Residual Gas Analyzermass spectrometer

LN2