Top Banner
CTG & Sea Sentry Stephanie Lavelle Maritime Manager Chelsea Technologies Group
15

CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

Dec 31, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

CTG & Sea Sentry

Stephanie Lavelle

Maritime Manager

Chelsea Technologies Group

Page 2: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Chelsea Technologies Group

Application of

sensor

technology

Process control

& monitoring

Medical

diagnostics

Agribusiness

Fresh water &

coastal

monitoring

Offshore pipeline leak detection

Water quality monitoring

NAD

Tactical

oceanography

Acoustic target

simulation

Oceanographic

research

Sensors &

systems

Acoustic

transducers

Fisheries

Exhaust gas

monitoring

Ballast water

monitoring

FerryBox

Water

quality/Treatment

Bio/chem Security

Environmental

Page 3: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

CTG fluorometer products

• Visible• Algae

• Dye tracing

• UV• BTEX

• Polycyclic Aromatic Hydrocarbons

• Tryptophan

• CDOM

• Optical Brighteners

• Active• Photosynthesis

• Algal health

• Ballast water monitoring

Page 4: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Sea Sentry - integrated cabinet solution

Page 5: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Sea Sentry features

• Sensors

• Digital pH probes (pre-calibrated)

• Turbidity sensor, ISO 7027:1999(E) compliant

• Extended range PAHphe eq sensor

• Turbidity, ‘colour’ and temperature correction

• Integrated sensor validation/calibration protocols

• Solid standards for optical sensor validation

• System

• Integrated pump & de-bubbler

• Pneumatic anti-fouling protection

• Single board interface electronics

• Touchscreen Graphical User Interface

Page 6: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Fluidics

pH sensor

controller

pneumatic

solenoid

valves

pneumatic

regulator

de-aerator

pressure

gauge

pump

outlet

isolation valve

pressure relief

valve

flow meter

turbidity sensor

inlet isolation valve

PAHphe eq sensor

pH sensor mount

strainer

drain valve

Page 7: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

De-bubbler design

Aerated

Water

Outlet

Low Pressure

Bleed

Sensors

De-a

era

tor

Page 8: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

• Single interface PCB

• All wiring on removable connectors

• Easy field replacement

• Digital interfaces

• MODBUS via Ethernet

• Dual redundancy

• USB 2.0 for data download

• Analogue interfaces

• Optional 5 off 4-20 mA

Electronics - enclosure

Page 9: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Integrated fluorometer & transmissometer

Fluorometer

• Range: 0.04 to 4500 µg/l, PAHphe eq (corrected)

• Sensitivity: <0.07 µg/l @ 1 Hz

• Excitation wavelength: 255 nm

• Emission wavelength: 365 nm

• Turbidity rejection: ≤5% deviation for 300 µg/l Phenanthrene

over turbidity range 0-1000 FNU (corrected)

Transmissometer

• Measurement range: 0 – 3.5 cm-1

• Resolution: 0.002 cm-1

• Pathlength: 1.25 cm

• Wavelength: 255 nm

PAH+ sensor

Page 10: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Turbidity sensor

• Measurement range: 0 to 1000 FNU

• Sensitivity: 0.15 FNU @ 1Hz

• Measurement principle: ISO 7027:1999(E) compliant

• Wavelength: 860 nm

• Bandwidth: 30 nm

Turbidity sensor

Page 11: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

PAH+ & Turbidity sensors

• Calibration valid for 2 years (unpowered)

• Calibration interval 12 months in use (return to CTG)

• On-site validation using solid sensor calibration

check

pH Sensor

• Calibration interval 1 month in use (on-site)

• Software based calibration protocol using pH4 &

pH7 buffer tablets

Sensor calibration & validation

Optical sensor

calibration check

Page 12: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Touchscreen GUI

Page 13: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

GUI – home screen

Page 14: CTG & Sea Sentry - EGCSA · •Turbidity sensor, ISO 7027:1999(E) compliant •Extended range PAH phe eq sensor •Turbidity, ‘colour’ and temperature correction •Integrated

COMMERCIAL IN CONFIDENCE

Sea Sentry

• Sensors

• Digital pH probes (pre-calibrated)

• Turbidity sensor, ISO 7027:1999(E) compliant

• Extended range PAHphe eq sensor

• Turbidity, ‘colour’ and temperature correction

• Integrated sensor validation/calibration

protocols

• Solid standards for optical sensor validation

• System

• Integrated pump & de-bubbler

• Pneumatic anti-fouling protection

• Single board interface electronics

• Touchscreen Graphical User Interface