Page 1
SiemensWalz
SOMATOM Definition
Hardcopy Cameras
Installation and Startup
© Siemens, 20All documents may only bservices on Siemens Healdocument in electronic foonce. Copy and distributioments and hardcopies is will be liable for damages. served.
CT
08e used for renderingthcare Products. Anyrm may be printedn of electronic docu-prohibited. OffendersAll other rights are re-
2008
Print No.: CT00-000.814.30.02.02Replaces: CT00-000.814.30.01.02
ED
nglishoc. Gen. Date: 05.09
n.a.H WS CS SD CO
Camera information with SOMARIS/7 VA20;VA26
© Siemens, 2008This document is valid for the following CT systems:SOMATOM DefinitionSOMATOM Definition ASSOMATOM Definition AS+SOMATOM Definition AS20
0774076910430603
Page 2
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2008“ refers to the copyright of a Siemens entity such as Siemens Aktienge-sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-spond to the version at the time of system delivery and/or printout. Versions to hardcopy documentation are not automatically distributed. Please contact your local Siemens office to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under any theory of law.
The installation and service of equipment described herein requires superior understand-ing of our equipment and may only be performed by qualified personnel who are specially trained for such installation and/or service.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 2 of 260
Page 3
Table of Contents 3
0Table of Contents
1 _______ General Hardcopy Camera Information ______________________________ 7
HCC Information for SOMATOM Definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Scope of applicability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Image quality and optical density check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18General image quality information (LUT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Density verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Reference films . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Description of Correction LUT Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Additional and unnecessary LUT files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Postscript Hardcopy Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Restrictions and Known Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Postscript Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Look-Up Tables (LUT). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2 _______ Specific Camera Information - Agfa________________________________ 30
Agfa Drystar AXYS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . . 33Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Agfa Drystar 5503. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 41Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Agfa Drystar 5500. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 49Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Agfa Drystar 5302. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 57Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Agfa Drystar 5300. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 63Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Agfa Drystar 3000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 70Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 3 of 260
Page 4
4 Table of Contents
Agfa LR5200 / MG3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 76Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Agfa LR5200 / LR DICOM Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 82Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
3 _______ Specific Camera Information - Codonics____________________________ 87
Codonics Horizon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 90Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
4 _______ Specific Camera Information - Fuji_________________________________ 94
Fuji Drypix 7000 / 5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 97Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Fuji Drypix 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 105Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Fuji Drypix 3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 112Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Fuji Drypix 2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 119Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Fuji FM DPL/FN-PS551 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 128Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
5 _______ Specific Camera Information - Kodak _____________________________ 134
Kodak DryView 6800. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 4 of 260
Page 5
Table of Contents 5
SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 137Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Kodak DryView 5800 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 145Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Kodak DryView 8900 (small border) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 154Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Kodak DryView 8700-Plus / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 159Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Kodak DryView 8700-Dual / PACS Link MIM 9410. . . . . . . . . . . . . . . . . . . . . . . . . . . . 164General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 166Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Kodak DryView 8200 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 172Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
Kodak DryView 8100 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 179Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Kodak DryView 8150 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 187Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Kodak KELP 2180/PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 194Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
6 _______ Specific Camera Information - Konica Minolta ______________________ 199
Konica Minolta DryPro 832. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 202
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 5 of 260
Page 6
6 Table of Contents
Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Konica Minolta DryPro 793 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 212Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
Konica Minolta DryPro 771 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 219Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Konica DryPro 751/752 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 227Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
7 _______ Specific Camera Information Sony________________________________ 232
Sony UP-DF 550 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 235Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
Sony UP-DF 500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 242Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244
8 _______ Specific Printer Information Epson _______________________________ 245
Epson AcuLaser C 4200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245Printer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246Modality Configuration (Postscript Driver) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246Hardcopy Printer Identification (provided by OEM) . . . . . . . . . . . . . . . . . . . . . . . . . 249Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250
9 _______ Specific Printer Information Hewlett Packard _______________________ 252
Hewlett Packard Color Laser Jet 4700N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252Printer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Modality Configuration (Postscript Driver) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Hardcopy Printer Identification (provided by OEM) . . . . . . . . . . . . . . . . . . . . . . . . . 257Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258
10 ______ Changes to Previous Version ____________________________________ 260
Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 6 of 260
Page 7
General Hardcopy Camera Information 7
1- 1General Hardcopy Camera InformationHCC Information for SOMATOM Definition 0
General information 0
Acronyms and abbreviations
Block Diagram SOMATOM Definition
Fig. 1:
System references
• Refer to Siemens Medical Solutions Intranet, UPTIME Services
AE Title Application entity title
BU Business Unit (Modality)
DCS DICOM Conformance Statement
DICOM Digital Imaging and Communication in Medicine
HW/SW Hardware/Software
HCC Hardcopy Camera
LAN Local Area Network
LUT Lookup Table
OEM Original Equipment Manufacturer
SCP Service Class Provider
SCU Service Class User
TCP/IP Transfer Control Protocol/Internet Protocol
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 7 of 260
Page 8
8 General Hardcopy Camera Information
• Product-specific technical documentation
Hardware components
• Ethernet switch ATI-FS708 (part no.: 30 79 332)
• Cable, straight-through, S/UTP RJ45 (5m.; part no.: 30 74 515)
• Cable, straight-through, S/UTP RJ45 (30 m.; part no.: 30 74 523)
or
• Refer to Siemens Medical Solutions Intranet, UPTIME Services
• Product-specific technical documentation
Restrictions and known effects
NOTE Sending film sizes that are not available at the customer sitemay cause problems at the camera site. For this reason, it isrecommended that you delete these film sizes for portraitand landscape formats.
NOTE The Grayscale Standard Display Function (GSDF) is imple-mented for printing and imaging. No Presentation LUTs areavailable. For this reason, the DICOM Presentation LUT is notreleased.
NOTE The use of the presentation LUT has to be disabled by localservice for configuring LUT files.
NOTE The density values for the image quality were measured witha calibrated densitometer X-Rite 301. The result of the mea-surement from other densitometers used may differ from theabsolute density values as described in the tables.
NOTE Print the 17 graystep test image to check the modality andcamera adjustment according to the description. Verify theresults using the reference values for max. density 2.80 listedin the table. This is also part of the acceptance test.
NOTE For additional Postscript Hardcopy Restrictions and KnownEffects, refer to the Chapter on Postscript Hardcopy Informa-tion as well. (Restrictions and known effects / p. 8)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 8 of 260
Page 9
General Hardcopy Camera Information 9
Errors and warning messages
Only certain errors and warnings have been tested for each hardcopy camera.
These errors and warnings are described in the corresponding camera information.
.
Scope of applicability 0
Modality/system
Hardcopy Cameras
This table includes all hardcopy cameras using DICOM basic print.
Device type Modality (SCU)
Product name SOMATOM Definition
Modality host name/console CT host name
ICS (syngo Acquisition Workplace)
IES (syngo CT Workplace; optional)
IRS (Image Reconstruction System)
Manufacturer Siemens Med CT
SW revision Somaris 7 VA20A (syngo-based VE32A)
Somaris 7 VA26 (syngo-based VE32B)
Conformance statement EDM 10094810-EKL-YK0-01
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 9 of 260
Page 10
10 General Hardcopy Camera Information
Manufac-turer
HCC nameSoftware validity
VA20 VA26 V.... V....
AGFA Drystar AXYS X X
AGFA Drystar 5503 X X
AGFA Drystar 5500 X X
AGFA Drystar 5302 X X
AGFA Drystar 5300 X X
AGFA Drystar 3000 X X
AGFA LR5200 / LR DICOM Controller X X
AGFA LR5200 / MG 3000 X X
CODONICS Horizon X X
FUJI Drypix 7000/5000 X X
FUJI Drypix 4000 X X
FUJI Drypix 3000 X X
FUJI Drypix 2000 X X
FUJI FM DPL / FN-PS 551 X X
KODAK DryView 6800 X X
KODAK DryView 5800 X X
KODAK DryView 8900 (small border) X X
KODAK DryView 8700-Plus /
PACS Link MIM 200X X
KODAK DryView 8700-Dual /
PACS Link 9410X X
KODAK DryView 8200 /
PACS Link MIM 200X
X
KODAK DryView 8150 X X
KODAK DryView 8100 /
PACS Link MIM 200X X
KODAK KELP 2180 /
PACS Link MIM 200X X
KONICA MINOLTA
DryPro 832X X
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 10 of 260
Page 11
General Hardcopy Camera Information 11
X = hardcopy camera is released
N/A = hardcopy camera is not released
NOTE The use of non-released hardcopy cameras in conjunctionwith BU products is prohibited and does not meet therequirements specified.
Hardcopy Printer
This table includes all hardcopy cameras using Postscript.
For configuration of one of these printers - refer to the specific documents for these printers
X = hardcopy printer is released
N/A = hardcopy printer is not released
NOTE The use of a non-released hardcopy printer in conjunctionwith CT products is prohibited and does not meet the require-ments specified.
KONICA MINOLTA
DryPro 793X X
KONICA MINOLTA
DryPro 771X X
KONICA DryPro 751/752 X X
SONY UP-DF550 X X
SONY UP-DF500 X X
Manufac-turer HCC name
Software validity
VA20 V... V.... V....
EPSON AcuLaser C4200 X X
Hewlett Packard
Color Laserjet 4700NX X
Xerox Phaser 6360 DP n.a. n.a.
Manufac-turer HCC name
Software validity
VA20 VA26 V.... V....
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 11 of 260
Page 12
12 General Hardcopy Camera Information
DICOM associations
A DICOM association is a network connection between 2 DICOM applications. The asso-ciation is established and released by the SCU and may be aborted by the SCU or SCP. The maximum number of associations for each HCC device is defined in the DCS. The maximum number of associations depends on the configuration and may be extended by increasing the number of gateways connected to the printer device.
In a DICOM network environment, there are two ways to connect a modality (SCU) to an HCC device (SCP).
Case 1: The association will always be opened. Assuming that this behavior applies to all modalities (connected to the same HCC device), the number of modalities has to be equal to or less than the maximum number of supported associations.
Case 2: The association will be opened (and closed) on demand. Assuming that this behavior applies to all modalities (connected to the same HCC device), the number of modalities may be greater than the maximum number of supported associations. This num-ber will be shared by the modalities.
The association open-close behavior is unknown to other non-SIEMENS modalities. Under optimum circumstances, the number of modalities is equal to or less than the maximum number of supported associations.
The association is always opened by the SCU (case 1)
If the associations are always opened by the SCU for each configured HCC device, the association is reserved for this SCU only.
If the maximum number of associations = 1, and another modality never releases the asso-ciation, this HCC device is always reserved for this modality.
The association is opened and closed on demand (case 2)
The associations will be opened (and closed) on demand by the SCU for each configured HCC device via the product name.
Special cases (compare also to case 1):
If the number of modalities configured in the HCC device is greater than the maximum num-ber of supported associations, some modalities will not be able to print. The printing func-tion depends on the access time and release mechanism of each connected modality (e.g., the maximum number of supported associations = 1, the number of modalities = 2. During access by the first modality, access by the second modality is denied and connectivity fails, and vice versa).
If other modalities do not release the associations, the shareable number of associations decreases.
Modality 0
Hardware and software prerequisites
All hardware and software prerequisites are included in the <modality> product delivery volume.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 12 of 260
Page 13
General Hardcopy Camera Information 13
Parameters
For software configuration, HCC data/parameters and possibly additional film matrix prop-erties (syngo: filming properties) are required for software configuration.
Modality data and configuration parameters
HCC Data and Configuration Parameters1
1 The values for these parameters are part of the specific HCC information
Example: Kodak DryView 8300
Properties Example Supplier
Port number 104 Default entry
AE title PR_”Hostname” Default entry
TCP/IP address xxx.x.x.xxx Network administrator
Application Entity
PropertiesDisplay Supplier
AE TitleK8300
Network administrator or HCC service
Port number 5040 HCC Service
HC Device /
General settings Display Supplier
Type DICOM printer Default entry
Class (printer spec file name)
kodak8300 Default entry
DICOM Node dv8300 Default entry
Filming properties Display Supplier
Hold printed Film Jobs 10 Default entry
Min. density 20 Default entry
Pixel Size [1/1000 mm] 85.0 x 85.0 Default entry
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
8" x 10"
2256 x 2676Default entry
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
8" x 10"
2736 x 2136Default entry
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 13 of 260
Page 14
14 General Hardcopy Camera Information
Film matrix properties
The settings for filming properties depend on the class selected (DICOM camera). A pull-down menu is provided for DICOM-class cameras. Select your DICOM camera type. A specific file for this type is loaded and the "filming properties" are shown. Changing the "filming properties" will affect the camera site and may not meet the requirements. Do not make any changes, unless otherwise instructed.
NOTE Filming properties are "pixel size" and "media sheet size";this information is required for image scaling and cannot bechanged.
Software configuration
• The following information is required:
- Host Name
- TCP/IP address
- Application entity title
- Port number of the DICOM remote host (DICOM laser camera)
The AE title, port number and camera type have to be provided by the local camera service.
syngo software configuration and laser camera
1. Select options in the menu line
2. Navigate to local service
The Siemens Service Software Authentication platform is displayed
3. Enter the password and click ”OK”
The Home Menu Platform of the syngo Service Software is displayed
4. Click ”Configuration”
The Local Service Configuration List of system options is displayed
5. Click DICOM ”General”
6. Click ”>”
The local Application Entity Title platform is displayed
Medium type Blue film Default entry
Film destination Processor Default entry
Color appearance Grayscale 8 bit Default entry
Background Black Default entry
Transformation No magnification Default entry
Filming properties Display Supplier
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 14 of 260
Page 15
General Hardcopy Camera Information 15
7. Provide the Local Camera Service with the information in the field ”Print (SCU) AE Title PR_”Hostname”
8. Click ”Save”; ”>”; ”Finish”
The Configuration Setup platform is displayed
9. Click Local Host ”TCP/IP LAN”
The local Service Configuration TCP/IP LAN platform is displayed
10. Provide the Local Camera Service with the information in the ”IP address” field (scroll down menu for IP address but do not select it)
11. Click ”Finish”
The Configuration Setup platform is displayed
12. Select DICOM Print Devices
The Local Service Hardcopy Device platform is displayed
13. Enter the host name and TCP/IP address of the laser camera (provided by the camera manufacturer)
14. Click ”Test” to test IP communication (ping)
15. Click ”Save”
Host properties successfully saved
16. Click ”OK”
17. Click ”>”
The General Note Properties Dialog platform is displayed
18. Under logical name: Enter the logical name of the laser camera
19. Select the host and choose the previously defined host name in the hardcopy device platform
20. Under application entity properties: Enter the AE title in the edit AE title field (provided by the camera manufacturer)
21. Enter the Port Number (provided by the camera manufacturer)
22. Click ”Add”
23. Click ”Save”
DICOM node properties successfully saved
24. Click ”OK”
25. Select the logical name again and click ”Verification” to test the DICOM communication (mc3echo)
26. Click ">"
The camera-specific parameter platform is displayed.
27. HC device: Enter the camera name
28. General settings: /type: DICOM printer
29. /class: e.g., agfadrystar3000, or refer to specific HCC information
30. Filming properties:
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 15 of 260
Page 16
16 General Hardcopy Camera Information
- Hold printed film job, 10 = default; (max. = 10) (films can be printed again)
- Min. density 20 = default; (can be changed in increments of 1)
- Pixel size (changes not allowed)
- Film Sheet Formats Portrait (do not modify and/or add film sizes, select and delete film sizes that are not available at the customer site)
- Medium type, blue film = default; clear film; paper
- Film destination, processor = default; magazine
- Color appearance, monochrome = default; (select color for printers that support and use color media only, e.g., Drystar2000; Codonics)
- Background, black =default; white
- Transformation no magnification = default;
do not change this value; except for Kodak DV8900; replicate;
(do not change any other values under filming properties)
31. Click "Save"
Caution: Please wait for the message "device properties successfully saved"
32. Click ”OK”
33. Click ”Finish”
The Configuration Setup platform is displayed
34. Click ”>” to proceed and select the next marked DICOM ”HC Overview”
35. Click ”Finish”
The Configuration Setup platform is displayed
36. Click LUT files
The LUT platform is displayed.
NOTE If new LUT files for one or more modalities are available on aremovable medium, e.g., floppy or CD, these files can beselected on the source drive and copied to the dependentLUT directories on the local host. Unnecessary LUT files canbe removed from the LUT directories on the local host.
37. Click ">"
Usage of Presentation LUT platform is displayed
38. Select disable
39. Modifications are necessary for additional image impressions only.
NOTE Only the topic CT image and SC image are relevant. Depend-ing on the camera, specific entries have to be made. If noth-ing else is mentioned, set the CT image and SC image to"CTLUT" and "2.80" for maximum density. The entries for BUimage and SC image have to be identical.
40. Overall maximum density:
- 3.00 = default (depends on the max. possible value of the max. density of all image types)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 16 of 260
Page 17
General Hardcopy Camera Information 17
41. Interpolation for printing:
- Bilinear mode = default; #Smoothing Type
- Replicate #Smoothing Type
- Cubic 0, 2, 3, 4 = concerns cubic mode only
42. When changes are made, click "Save"
43. Click "Finish"
44. Click "Home"; the system automatically restarts the application
NOTE This may take some time. Confirm warnings with "OK."
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 17 of 260
Page 18
18 General Hardcopy Camera Information
Image quality and optical density check 1.1
General image quality information (LUT) 0
Only standard LUT files are loaded. For importing/removing additional/unnecessary LUT files (see section ”Additional/unnecessary LUT files”) (LINK).
The image data from the modality to the HCC is transformed/corrected by the LUTs, once in the modality and once in the HCC. As as result, the LUT has to be set on both sides:
• Refer to the Siemens Medical Solutions Intranet, UPTIME Services
• Product-Specific Technical Documentation
Correction LUT is selected
Density verification 0
There are two ways to check the LUT via the film density:
The modality is configured to a specific correction LUT and the HCC is configured to linear curves. As an alternative, the modality is configured to a linear LUT and the HCC is config-ured to linear curves.
Modality Image Type Correction LUT Max. density
CT Image CT LUT 2.80
SC Image CT LUT 2.80
HCCRefer to the specific Camera Information "DICOM printer settings".
The HCC Service performs the setting "LINEAR LUT"
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 18 of 260
Page 19
General Hardcopy Camera Information 19
Density verification for CT
Since these two checks are equivalent; it is sufficient to check the “specified density”.
Checking the “HCC linearity” may be necessary if the desired density verification (1) shows deviations and the HCC setting needs to be changed.
NOTE Even if the HCC service does not require a test image of themodality for the HCC setting, it is easier to check this setting- and possibly further HCC corrections - with the HCC linear-ity verification (2) than with the desired density verification(1). Only the HCC linearity LUT is effective. A graphical dis-play would now show the current HCC behavior only. Thiscould be helpful for HCC Service.
Density values for CT
• Measure the density values in 17 gray levels in the IEC test image (Fig. 2 / p. 19) (for each film size).
W:256; C:0;
Fig. 2: Test image
Check no.
Verification for Modality adjustmentReference
valueCT
ApplicationImage Type Correction
LUTMax.
Density
1Desired density CT Image CT LUT 2.80 Refer to the
table for CTLUT
2HCC linearity CT Image Linear 2.80 Refer to the
table for lin-ear LUT
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 19 of 260
Page 20
20 General Hardcopy Camera Information
The density difference of the BU LUT (Ref. LUT) between steps 1, 2, and 3 has to be a minimum density value of 0.02. Test image 3 (SMPTE image) can be used for visual veri-fication (the 95% gray square has to be visible).
Reference films 0
Reference film
To create a reference film with the desired LUT adjustment, proceed as follows:
1. The layout is configurable under Option/Configuration/Filming layout
2. Use format 3.2; default center/width, for the film sheet
Step
CT LUT Linear LUT
Ref. LUTMaximum Density
Minimum Density Ref. LUT
Maximum density
Minimum density
1 0.20 0.23 0.17 0.20 0.23 0.17
2 0.23 0.26 0.20 0.36 0.40 0.33
3 0.28 0.32 0.24 0.53 0.56 0.49
4 0.35 0.39 0.31 0.69 0.73 0.64
5 0.43 0.48 0.38 0.85 0.90 0.80
6 0.52 0.57 0.47 1.01 1.06 0.96
7 0.63 0.69 0.57 1.18 1.23 1.12
8 0.75 0.81 0.69 1.34 1.40 1.28
9 0.88 0.95 0.82 1.50 1.57 1.44
10 1.04 1.11 0.97 1.66 1.73 1.59
11 1.21 1.28 1.14 1.83 1.90 1.75
12 1.41 1.49 1.33 1.90 2.07 1.91
13 1.64 1.72 1.56 2.15 2.23 2.07
14 1.88 1.97 1.79 2.31 2.40 2.23
15 2.13 2.22 2.04 2.48 2.57 2.38
16 2.44 2.54 2.34 2.64 2.73 2.54
17 2.80 2.90 2.70 2.80 2.90 2.70
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 20 of 260
Page 21
General Hardcopy Camera Information 21
3. Select the "Viewing" tab, select Patient Browser, navigate to the folder "CT Reference Images" and view the test images:
Use test images (technical images) without text and the three CT Medical Ref images (cerebrum, pancreas, (LWS) lumbar spine) with text and print the film sheet. (the DICOM command to the printer is always "magnification none")
4. Measure IEC test image 1 (Fig. 2 / p. 19) and verify the CTLUT using the table "Density values for modality CT”
Segment layout:
5. Repeat this procedure with the linear LUT adjustment
6. If the reference images are not available, please load the reference images from
H:\SITE_DATA\Reference-images\* in the storage device (import from offline)
Description of Correction LUT Files 0
Description of Correction LUT Files for CT
1 2
3 4
5 6
Segment Image Window Center
1 IEC test image 1 (17 gray levels) 256 0
2 IEC test image 2 (50% gray field) 256 0
3 Test image 3 (SMPTE test pattern) 100 0
4 Pancreas (Abdomen) 520 104
5 Cerebrum (Head) 80 53
6 LWS (Spine) 440 56
Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80
Linear Linear LUT
CT LUT Standard LUT; default for all cameras; adapts Dmin. of 0.2
CT02 vs. CT LUTDark area = less contrast; darker; softer; white area = more con-trast
CT03 vs. CT02Dark area = less contrast; darker; softer; white area = more con-trast
CT04 vs. CT LUTDark area = more contrast; darker; sharper; white area: no changes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 21 of 260
Page 22
22 General Hardcopy Camera Information
Additional and unnecessary LUT files 0
If new, released LUT files are available on a removable medium (e.g., floppy or CD). These files may be selected on the source drive and copied to the dependent LUT directories on the local system. Unnecessary LUT files may be removed from LUT directories on the local system.
Importing LUT files
1. In Local Service Configuration platform, select LUT files
2. Select the (source) drive and the path where the appropriate LUT files are located
3. Select the destination LUT directory (correction LUT directory) which depends on the image type (CTImage).
4. By pushing the copy button, the selected LUT file will be transferred to CTImage type.
5. Repeat this procedure using the appropriate LUT files for SC Images.
6. Select the destination LUT directory (correction LUT directory) which depends on the image type (SCImage).
7. Pushing the copy button transfers the selected LUT file to SCImage type.
8. Confirm the message "Files successfully copied" by clicking "OK"
Removing LUT files
Unnecessary LUT files can be removed from LUT directories on the local host by selecting the appropriate LUT file and clicking the "Remove" button.
• Click ">"
CT05 vs. CT LUTBrighter; no changes in contrast for large window width; more contrast for head images (small window width)
CT06 vs. CT LUT Brighter than CT5; white area = less contrast
cthead1 LUT for head images, images smoother
cthead2 vs. cthead1 Images smoother
CT015Adapts Dmin of 0.15 (for cameras which are not supported Dmin. command)
CT017Adapts Dmin. of 0.17 (for cameras which are not supported Dmin. command)
CT LUTdryLUT for dry cameras; adapts Dmin. of 0.22 (5% visible in SMPTE test pattern)
CT Lung LUT for lung images
Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 22 of 260
Page 23
General Hardcopy Camera Information 23
Real Size Scaling
This procedure should be performed after system installation and when the Hardcopy Camera is ready to print. The real size behavior has to be performed with each connected Hardcopy Camera.
Real size scaling verification
1. Load the test image (water image phantom) into the viewer (e.g. 20 cm water phantom, generated with the constancy procedure by CT).
2. Insert Crosshair (Viewing Card: Tools) and copy the image to the film sheet (“copy image to film sheet” button).
3. Select film sheet format 1 X 1.
4. Select "Original Image" in the filming card and print the image.
5. Use a calibrated ruler to measure the crosshairs of the printed image.
The results should correspond to the printed value underneath the scale with a tolerance of ±1% in both directions.
Refer to the specific camera information.
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 23 of 260
Page 24
24 General Hardcopy Camera Information
Postscript Hardcopy Information 1.2
Restrictions and Known Effects 0
Restrictions
NOTE Paper printouts are not allowed for medical diagnosis.
NOTE Only one Adobe Postscript Printer can be installed on a64-bit computer.
More than one Postscript Printer Installation will not work.
NOTE The adobe printer driver with the printer-specific descriptionfile has to be used. The paper printer is not linear in greyscaleimage print-out.
Therefore the respective PSLUT Files has to be used to opti-mize the image quality for the print-out.
The PSLUT files are:
• EPSON_C4200_correction.lut
• HP_LaserJet4700_linear.lut
• Kyocera_FS_C5020N_Cor.lut
• inv_CTLUT.lut
NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g calcium scoring) usethe report of the applications to print out the images of theresults.
Known Effects
n.a.
Postscript Driver 0
General information
The Printer Driver Installation of Somaris software includes the particular printer descrip-tion file. Both have to be performed directly in WINDOWS XP. The generic Adobe Post-Script Printer Driver is already preconfigured and also visible in Somaris as “SIEMENS.PD” printer. Only the postscript description file (ppd - file) for the specific printer has to be installed additionally.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 24 of 260
Page 25
General Hardcopy Camera Information 25
Missing Installation of Adobe Postscript Driver V5.2 in Windows XP
Installation of Adobe Postscript
Alternatives for special printer setting
Missing Installation of Adobe Postscript Driver V.5.2 in Windows XP
If the “SIEMENS.PD” is not available (e.g. deleted), a new installation of this generic Adobe driver is required
Log into your system with your administrator account.
Navigate to folder: C:\Somaris\Tools\Postscript\XP64.
Double-click the patch procedure:
Install_Printer_Driver.bat
Installation of Adobe Postscript Printer in Windows XP
Log into your system with your administrator account.
Navigate to folder: C:\Somaris\Tools\Postscript.
Double-click the file PostScriptInst.exe
click <browse>
----> "browse for folder " window appears.
Select the appropriate post script printer name
1. HP Color Laser Jet 4700N
select folder PPD\hp and click OK:
A list of available HP - printers appears
Select "hpmcpd25.ppd" and press Go
The window is closed
Restart system
Login as administrator
2. Epson Aculaser C4200
select folder PPD\epson and click OK:
A list of available EPSON printers appears
Select "epalc420.ppd” and press Go
The window is closed
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 25 of 260
Page 26
26 General Hardcopy Camera Information
“Configure the postscript printer driver as described in the part of printer informa-tion specific for the respective PS printer”.
Restart system
Login as administrator
Configuration of Adobe Postscript Printer in Windows XP
Start --> Settings --> Printers and Faxes
click “SIEMENS.PD” with the right mouse button
select properties
change in “General” the name “SIEMENS.PD” to the appropriate printer name (e.g. HP4700N)
click <OK>
select e.g. "HP4700N"
and with right mouse key
select "Properties"
The "HP4700N Properties"- window appears.
Select: "Ports" and click <Add Port>
"Printer Ports" window
Select: "StandardTCP/IP Port" and "New Port"
"Add Standard TCP/IP Printer Port Wizard" window appears,
click <Next>
Enter the IP address of the printer in the field:
Printer Name or IP Address: xxx.xxx.xxx.xxx
Port Name:
Let Windows choose the port name.
<Next>
"Add Standard TCP/IP printer Port Wizard" window
"Standard" and "Generic Network Card" is selected
Click : <Next>
Click : <Finish>
<Close> "Printer port" window
<Close> "HP4700N Properties" window
2. Epson Aculaser C4200
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 26 of 260
Page 27
General Hardcopy Camera Information 27
Configure in Local Service Configuration
Select "Paper Printer" in Configuration Menu.
Select “Local service” -> configuration -> Paper Printer
Deselect all paper printer media that the printer does not support.
<save>,
click “>”
The “Printer Paper size Mapping” window appears
select the appropriate printer in Configured Printers
a table for mapping of paper sizes appears
Assign the real names of the printer paper sizes to the fixed syngo paper size names.
e.g. Printer Paper Size Mapping
Configured Printer e.g. HP 4700N
Syngo Paper Size Printer Paper Size
Letter Small --> Letter
A4 --> A4
Configure in Local Service Configuration
<save>,
<finish>,
Select "Local service" -> configuration -> PS LUT Files
Go to the PSLUT directory and select the specific Postscript LUT file for this printer (names of PS-LUTs, refer to the description specific to the printer):
Select: "Use LUT for Postscript Printers"
Select: “Use DICOM LUT for Postscript Printers”
Select: "Use floating Zoom factors"
Click : <Save>
Click : <Finish>
Click : <Home>
Restart Somaris
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 27 of 260
Page 28
28 General Hardcopy Camera Information
Look-Up Tables (LUT) 0
CT Reference Film and LUT
To create a reference, film the images of the patient: “Reference Images” have to be used. It contains 3 test images and 3 medical images. Concerning the positions of the images on the paper, the order is from top left to right bottom. Test images cannot contain image text, medical images need the full text.
The measured values are described in the specific Printer Information included in this doc-ument.
Image Attributes for syngo-based modalities
Precondition
Segment Image Window Center
1 IEC test image 1 (17 gray levels) 256 0
2 IEC test image 2 (50% gray field) 256 0
3 Test image 3 (SMPTE test pattern) 100 0
4 Pancreas (Abdomen) 520 104
5 Cerebrum (Head) 80 53
6 LWS (Spine) 440 56
Max. Density 2.8
Max. Density overall 3.0
Interpolation for printing Bilinear
Transformation No magnification
Segment lines no
Page number on print-out Not selected
Layout divisions 2x3
Orientation portrait
Image order Horizontal
Paper size A4
Reference image Not selected
Modality
The modality has to be configured for image types “CT Image” and for “SC Image” to CTLUT.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 28 of 260
Page 29
General Hardcopy Camera Information 29
Postscript Printer LUT
Real size scaling
Load the test image (water image phantom) into the viewer (e.g. 20 cm water phantom, generated with the CT constancy procedure ). Insert crosshair with tool in viewer. Send this image from the viewer to the filming tab card and expose it on paper.
Evaluate the printed paper. Search 2 scale lines of the crosshair with the largest possible distance to each other, separate in the x and y direction and noted as the desired value.
Measure the distance with the ruler and compare the measured value with the described values written in the specific Printer Information included in this document.
Image attributes for syngo-based modalities
Printer
The printer must be switched to black & white.
In "Local service" -> configuration -> PS LUT Files
The respective PS LUT file has to be selected
Postscript Printer PS LUT file name
Epson AcuLaser C4200 EPSON_C4200_correction.lut
Hewlett Packard Color Laserjet 4700 HP_LaserJet4700_linear.lut
Kyocera Ecosys FS-C5020N Kyocera_FS_C5020N_Cor.lut
Xerox Phaser 6350 DP inv_CTLUT.lut
Transformation No magnification
Interpolation for printing Bilinear
Segment lines no
Page number on print-out Not selected
Layout divisions 1x1
Orientation Portrait/landscape
Paper size A4/Letter small
Aspect ratio Original image
Scale factor 100%
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 29 of 260
Page 30
30 Specific Camera Information - Agfa
2- 2Specific Camera Information - AgfaAgfa Drystar AXYS 0
General Information 0
Block Diagram
Fig. 3:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10” x 12”Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 30 of 260
Page 31
Specific Camera Information - Agfa 31
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE European EMC Classification is Class A.
NOTE Slight inhomogeneity visible in gray field
NOTE Typical slight line structure of the thermo head is visible intechnical image.
NOTE Two edges of the diagnostic area are cut (approx 1 mm) for8x10, 10x12, 11x14,14x17.
NOTE The background command does not work properly. The back-ground is always white.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / modality software configuration 0
Time 74 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title AXYS Yes
Port number 104 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 31 of 260
Page 32
32 Specific Camera Information - Agfa
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfadrystar5503_AX
YSNo
DICOM Node DSAXYS Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880
In “define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 32 of 260
Page 33
Specific Camera Information - Agfa 33
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Printer Settings
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM printer
Device name: DRYSTAR AXYS
Manufacturer: AGFA
SW Revision: R 1.60C1
Physical interface: Ethernet 10/100 Base-Tx
Network protocol: TCP/IP DICOM
DICOM conformance statement: Document No.: 001150 Revision 1.0
Image memory: 65 GByte
Hard disk: 70 GByte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCP server
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 33 of 260
Page 34
34 Specific Camera Information - Agfa
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / Medium Type is not supported
No
n.a. Print even if Application Type is not supported
No
No Initial memory size allocated per print job (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;
Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Decimate/Crop behavior: Disabled
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value empty
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density empty
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
n.a. Level (%) Empty
Selected Variable name Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 34 of 260
Page 35
Specific Camera Information - Agfa 35
Error / warning messages (examples)
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality site No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported SOP classes Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOP INSTANCE
Basic Grey Image Box SOP Class
Basic color image box SOP class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. LR like off
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open
Tray open or front door open
BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 35 of 260
Page 36
36 Specific Camera Information - Agfa
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Send print job to camera if the camera is in service mode
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Sent print job to cam-era with empty input tray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 391.0 0.26
14" x 17"
Landscape
X 410.0 411.0 0.24
Y 320.0 320.0 0.00
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 320.0 320.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 331.0 0.30
11" x 14"
Portrait
X 260.0 261.0 0.38
Y 330.0 330.5 0.15
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 251.0 0.40
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 271.0 0.37
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 36 of 260
Page 37
Specific Camera Information - Agfa 37
10” x 12”
Landscape
X 280.0 280.5 0.18
Y 220.0 220.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 220.0 220.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 37 of 260
Page 38
38 Specific Camera Information - Agfa
Agfa Drystar 5503 2.1
General information 0
Block diagram
Fig. 4:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film Type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10” x 12”Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 38 of 260
Page 39
Specific Camera Information - Agfa 39
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE European EMC Classification is Class A.
NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.
NOTE Typical line structure of the thermo head is visible.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 70 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5503 Yes
Port number 104 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 39 of 260
Page 40
40 Specific Camera Information - Agfa
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfadrystar5503_AX
YSNo
DICOM Node DRY5503 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 40 of 260
Page 41
Specific Camera Information - Agfa 41
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM printer
Device name: DRYSTAR 5503
Manufacturer: AGFA
SW Revision: R 5.0.0 B2
Physical Interface: Ethernet 10/100 Base-Tx
Network protocol: TCP/IP DICOM
DICOM Conformance Statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5503, SW ver-sion 5.0
Image memory: 2 * 256 Mbyte (RAM memory)
Hard disk: min. 18 Gbyte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCP server
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 41 of 260
Page 42
42 Specific Camera Information - Agfa
No Polarity Normal
No Film destination output tray n.a.
n.a. Tray 1 Selected
n.a. Tray 2 Deselected
n.a. Tray 3 Deselected
n.a. Tray 4 Deselected
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type “empty”
n.a. Print even if Film Size ID / Medium Type is not supported
No
n.a. Print even if Application Type is not supported
No
No Initially allocated memory size per print job (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;
Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value “empty”
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density “empty”
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
Selected Variable name Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 42 of 260
Page 43
Specific Camera Information - Agfa 43
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) “empty”
n.a. Level (%) “empty”
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality site, No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOP INSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Early processing allowed Yes
n.a. LR like off
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 43 of 260
Page 44
44 Specific Camera Information - Agfa
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Open front door COVER OPEN COVER OPEN Camera cover, drawer or door open
Tray open or front door open
BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine
Sent print job to cam-era with empty input tray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Camera is warming up
PRINTER INIT PRINTER INIT Camera initializing
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 389.5 0.13
14" x 17"
Landscape
X 410.0 409.5 0.12
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 330.0 0.00
Y 320.0 319.5 0.16
14" x 14"
Landscape
X 330.0 329.5 0.15
Y 330.0 330.0 0.00
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 320.0 320.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 250.0 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 44 of 260
Page 45
Specific Camera Information - Agfa 45
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 270.0 0.00
10” x 12”
Landscape
X 280.0 280.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 45 of 260
Page 46
46 Specific Camera Information - Agfa
Agfa Drystar 5500 2.2
General Information 0
Block Diagram
Fig. 5:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10” x 12”Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 46 of 260
Page 47
Specific Camera Information - Agfa 47
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.
NOTE Typical line structure of the thermo head is visible.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 69 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5500 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 47 of 260
Page 48
48 Specific Camera Information - Agfa
Class
(Printer spec file name)agfadrystar5500 No
DICOM Node DRY5500 Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880
In “define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880s
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
HC Device /
General settingsDisplay Changes allowed
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 48 of 260
Page 49
Specific Camera Information - Agfa 49
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer Settings
Device type: DICOM printer
Device name: DRYSTAR 5500
Manufacturer: AGFA
SW Revision: V3.0.0 C1
Physical interface: Ethernet (RJ45)
Network protocol: TCP/IP
DICOM conformance statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5500, Document No.: 737
Image memory: 2 * 256 Mbyte
Hard disk: 18 GByte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCP server
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film destination output tray n.a.
n.a. Tray 1 Selected
n.a. Tray 2 Deselected
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 49 of 260
Page 50
50 Specific Camera Information - Agfa
n.a. Tray 3 Deselected
n.a. Tray 4 Deselected
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / Medium Type is not supported
No
n.a. Print even if Application Type is not supported
No
No Initial memory size allocated per print job (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;
Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value “empty”
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density “empty”
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
Selected Variable name Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 50 of 260
Page 51
Specific Camera Information - Agfa 51
n.a. Level (%) Empty
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality site, No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOP INSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Early processing allowed Yes
n.a. LR like off
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 51 of 260
Page 52
52 Specific Camera Information - Agfa
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Sent print job to cam-era with front door open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Tray open or front door open
BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine
Sent print job to cam-era with empty input tray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Printer warming up PRINTER INIT PRINTER INIT Camera initializing
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 399.0 0.25
14" x 17"
Landscape
X 410.0 409.0 0.24
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 329.5 0.15
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 329.5 0.15
11" x 14"
Portrait
X 250.0 249.5 0.20
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 249.5 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 52 of 260
Page 53
Specific Camera Information - Agfa 53
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 269.5 0.19
10” x 12”
Landscape
X 280.0 279.5 0.18
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 180.0 179.5 0.28
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 169.5 0.29
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 53 of 260
Page 54
54 Specific Camera Information - Agfa
Agfa Drystar 5302 2.3
General information 0
Block diagram
Fig. 6:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10” x 12”Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 54 of 260
Page 55
Specific Camera Information - Agfa 55
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE If the film size was not included in the configuration of thecamera, the connection to the camera is not available.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 67 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5302 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec. file name)agfadrystar5302 No
DICOM Node Agfa5302 Yes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 55 of 260
Page 56
56 Specific Camera Information - Agfa
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.375 x 79.375 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4358 x 5232
14" x 14"
4358 x 4303
11" x 14"
3348 x 4358
10” x 12”
3070 x 3653
8" x 10"
2376 x 3070
In “Define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5232 x 4358
14" x 14"
4303 x 4358
11" x 14"
4358 x 3348
10” x 12”
3653 x 3070
8" x 10"
3070 x 2376
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 56 of 260
Page 57
Specific Camera Information - Agfa 57
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM printer
Device name: DRYSTAR 5302
Manufacturer: AGFA
SW Revision: V1.80 c1
Physical Interface: Ethernet (RJ45)
Network protocol: TCP/IP
DICOM Conformance Statement: Revision 1.1, June 2005
Image memory: 512 Mbyte (internal memory)
Hard disk: 35 GB
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming "Calling AE Title":
PR_hostname
n.a. Use this profile only for local SCP server DEFAULT_SCP
n.a. Host Profile Type GENERIC-GENERIC (custom-ized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID Deselected
No Medium type Deselected
n.a. Print even if Film Size ID / Medium Type is not supported
No
n.a. Print even if Application Type is not sup-ported:
No
No Initial allocated memory size per print job (Mbytes):
100
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 57 of 260
Page 58
58 Specific Camera Information - Agfa
No Kernel SITE (*) Edit Custom:
Interpolation type: Cubic high res. -smoothing factor -0.7
No Magnification Magn: Max. - custom scale fac-tor 1.0
No Pixel Size Ratio: 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value Deselected
No Illumination (cd/m2) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density Empty
No Maximum density Site 300
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
n.a. Level (%) Empty
n.a. Image orientation R0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality Site, No.
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
Selected Variable name Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 58 of 260
Page 59
Specific Camera Information - Agfa 59
Error / Warning Messages (examples)
n.a. Resolution 320
n.a. Supported SOP classes: Edit
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOP INSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. LR-like Off
Action that results in camera error mes-sages
Printer messagesMessage displayed on
SCUDefined DCS (*1)
Sent by SCP (*2)
Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open
Open film supply tray BAD SUPPLY MGZ
BAD SUPPLY MGZ
Problem with supply magazine
Empty input tray SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Sent wrong film size n.a. n.a. Status -> aborted
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 59 of 260
Page 60
60 Specific Camera Information - Agfa
NOTE If film size is not inserted in configuration of camera, connec-tion to camera is not available.
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 391.0 0.26
14" x 17"
Landscape
X 410.0 411.0 0.24
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 331.0 0.30
11" x 14"
Portrait
X 250.0 250.5 0.20
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 251.0 0.40
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 271.0 0.37
10” x 12”
Landscape
X 280.0 281.0 0.36
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 170.0 170.5 0.29
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 60 of 260
Page 61
Specific Camera Information - Agfa 61
Agfa Drystar 5300 2.4
General information 0
Block diagram
Fig. 7:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film Type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10” x 12”Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
8" x 10"Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
Other n.a. N/A n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 61 of 260
Page 62
62 Specific Camera Information - Agfa
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 67 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5300 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfadrystar5300 No
DICOM Node DS5300 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 24 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 62 of 260
Page 63
Specific Camera Information - Agfa 63
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Pixel Size [1/1000 mm] 79.375 x 79.375 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4256 x 5174
11" x 14"
3300 x 4256
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5174 x 4256
11" x 14"
4256 x 3300
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM print server/printer
Device name: DRYSTAR 5300
Manufacturer: AGFA
SW Revision: 1.82 C1
Physical Interface: 10/100 Base - T
Network protocol: TCP/IP
DICOM Conformance Statement: Rev.1.2, October 24, 2005
Image memory: 512 Mbyte
Hard disk: None
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 63 of 260
Page 64
64 Specific Camera Information - Agfa
Printer settings
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming “Calling AE title”
PR_hostname
n.a. Use this profile only for local SCP server
All
n.a. Host Profile Type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID Deselected
No Medium type Deselected
n.a. Print even if Film Size ID / Medium Type is not supported
No
No Kernel SITE
Interpolation type: Cubic High Res - Smooth Factor -2.5
No Magnification Magn: Max. - custom scale factor 1.0
No Pixel Size Ratio: 1.0 / 1.0
No Perception LUT Linear
No Kanamori-like value 0
No Custom Value Deselected
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density Empty
No Maximum density Site 300
No. Annotation 1 None
No Annotation 2 None
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 64 of 260
Page 65
Specific Camera Information - Agfa 65
n.a. Window (%) Empty
n.a. Level (%) Empty
n.a. Image orientation R0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo Modality site: No.
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 320
n.a. Supported SOP Classes Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic Color PM Meta SOP Class
Print Queue Management SOP INSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Initial allocated memory size per print job (MBytes)
23
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 65 of 260
Page 66
66 Specific Camera Information - Agfa
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Action that results in camera error messages
Printer messages Message dis-played on SCU
(*2)Defined DCS (*1)
Sent by SCP
Open input tray BAD SUPPLY MGZ
WARNING\BAD SUPPLY MGZ
Camera info.: WARNING
Top cover open COVER OPEN WARNING\COVER OPEN Camera info.: WARNING
Empty input tray SUPPLY EMPTY
WARNING\SUPPLY EMPTY
Camera info.: WARNING
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 401.0 0.25
14" x 17"
Landscape
X 400.0 401.5 0.37
Y 320.0 320.0 0.00
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X 250.0 250.5 0.20
Y 320.0 320.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 250.5 0.20
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 66 of 260
Page 67
Specific Camera Information - Agfa 67
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 67 of 260
Page 68
68 Specific Camera Information - Agfa
Agfa Drystar 3000 2.5
General information 0
Block diagram
Fig. 8:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 4
Format Release status Film Type Medium type
14" x 17"Portrait Yes DT1B Blue-based
Landscape Yes DT1B Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait No No No
Landscape No No No
10” x 12”Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 68 of 260
Page 69
Specific Camera Information - Agfa 69
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
Known effects
NOTE Changing the film size requires manual modifications at thecamera site.
SCU / Modality Software Configuration 0
Time 160 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY3000 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfadrystar3000 No
DICOM Node DRY3000 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 69 of 260
Page 70
70 Specific Camera Information - Agfa
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Pixel Size [1/1000 mm] 79.4 x 79.4 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4224 x 4839
11" x 14"
3268 x 3924
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5142 x 3924
11" x 14"
4224 x 2962
Medium type Blue film Yes
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM printer
Device name: DRYSTAR 3000
Manufacturer: AGFA
SW Revision: MeGa 3.32, cad 32948, ecu 32950, DI301p95, R1.95, aos 32947, dcm 32957, mps 32941, OEM 49535
Physical Interface: 10 base - TP Ethernet
Network protocol: TCP/IP
DICOM Conformance Statement: Rev.2.3, July 31, 2003
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 70 of 260
Page 71
Specific Camera Information - Agfa 71
Printer settings
Error / Warning Messages (examples)
Image memory: 32 Mbyte
Hard disk: 1.2 Gbyte
File Name Content Remarks
pms.ini PR_hostname
IDF=
CFG=D:\MEDCOM2.PMS
No .idf file defined
PMS.IDF is the default
PMS.IDF default STANDARD\1,1 Agfa layout ID:
14x17P ID1411
14x17L ID1511
14x14P ID1611
11x14P ID1211
11x14L ID1311
8x10P ID1011
8x10L ID1111
Medcom2.pms 00
pms_virgin_cfg 0
n-event_enabled 0
trim_flag 0
utl_id 1
is used (DCS), no *.idf file defined
AGFA confirm Level is 0 (warnings)
Disable N-events messages
Disable trim around images
LINEAR LUT
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Switch camera from online to offline
PRINTER OFFLINE Association rejected
Connection to cam-era is not available
Open cover (lower) SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Top cover open n.a. CHECK PRINTER Check camera
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 71 of 260
Page 72
72 Specific Camera Information - Agfa
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 380.0 377.5 0.66
14" x 17"
Landscape
X 400.0 397.0 0.75
Y 300.0 300.0 0.00
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X No No No
Y No No No
11" x 14"
Landscape
X No No No
Y No No No
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 72 of 260
Page 73
Specific Camera Information - Agfa 73
Agfa LR5200 / MG3000 2.6
General information 0
Block diagram
Fig. 9:
Available film sizes
Format Release status Film Type Medium type
14" x 17"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
14" x 14"Portrait Yes
Scopix LT2B DLA1
Blue-based
Landscape No No No
11" x 14"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
10” x 12”Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
8" x 10"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
Other n.a. N/A n.a.
Released: Yes - Film size is available and released
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 73 of 260
Page 74
74 Specific Camera Information - Agfa
Performance and Connections
Associations
Maximum number of associations supported: 4
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly.
NOTE No message will be displayed in case of manual input of filmsinto developer.
Known effects
NOTE The camera does support format 14”X14” landscape.
SCU / Modality Software Configuration 0
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
Time 228 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title LR5200 Yes
Port number 104 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 74 of 260
Page 75
Specific Camera Information - Agfa 75
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfalr5200 No
DICOM Node LR5200 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 80.0 x 80.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4224 x 4839
14" x 14"
4224 x 3897
11" x 14"
3268 x 3924
8" x 10"
2344 x 2647
In “define new”
Do not modify/add
Select only
Delete onlyFilm Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5140 x 3924
11" x 14"
4224 x 2962
8" x 10"
2948 x 2046
Medium type Blue film Yes
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 75 of 260
Page 76
76 Specific Camera Information - Agfa
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Device type: DICOM print server/printer
Device name: MG 3000
Manufacturer: AGFA
SW Revision: MeGa 3.32,
dcm 32957
Physical Interface: 10 base - TP Ethernet
Network protocol: TCP/IP
DICOM Conformance Statement: Rev.1.4, April 20, 2001
Image memory: 128 Mbyte
Hard disk: 4.1 Gbyte
Device type: DICOM camera
Device name: LR 5200
Manufacturer: AGFA
SW Revision: LTP 01806, 240698
Physical Interface: FOI optic
Network protocol: None
DICOM Conformance Statement: None
Image memory: n.a.
Hard disk: None
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 76 of 260
Page 77
Specific Camera Information - Agfa 77
Printer settings
Error / Warning Messages (examples)
File Name Content Remarks
pms.ini PR_hostname
IDF=
CFG=D:\MEDCOM2.PMS
No .idf file defined
PMS.IDF is the default
PMS.IDF default STANDARD\1,1 Agfa layout ID:
14x17P ID1411
14x17L ID1511
14x14P ID1611
11x14P ID1211
11x14L ID1311
8x10P ID1011
8x10L ID1111
Medcom2.pms 00
pms_virgin_cfg 0
n-event_enabled 0
trim_flag 0
utl_id 1
is used (DCS), no *.idf file defined
AGFA default confirm Level is 0 (warnings)
Disable N-events messages
Disable trim around images
LINEAR LUT
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Camera switched off or camera switched to offline
OFFLINE OFFLINE Camera is switched off-line
Manual input of films in developer
NORMAL NORMAL No message dis-played
Camera is switched on
PRINTER INIT PRINTER INIT Camera is warming up
Supply magazine empty
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 77 of 260
Page 78
78 Specific Camera Information - Agfa
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 331.5 0.45
Y 380.0 383.0 0.79
14" x 17"
Landscape
X 400.0 403.0 0.75
Y 300.0 301.0 0.33
14" x 14"
Portrait
X 330.0 332.0 0.61
Y 300.0 302.0 0.67
14" x 14"
Landscape
X No No No
Y No No No
11" x 14"
Portrait
X 250.0 252.0 0.80
Y 300.0 301.0 0.33
11" x 14"
Landscape
X 330.0 331.5 0.45
Y 220.0 221.5 0.68
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X 170.0 170.5 0.29
Y 200.0 201.5 0.75
8" x 10"
Landscape
X 230.0 232.0 0.87
Y 160.0 160.5 0.31
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 78 of 260
Page 79
Specific Camera Information - Agfa 79
Agfa LR5200 / LR DICOM Controller 2.7
General information 0
Block diagram
Fig. 10:
Available film sizes
Format Release status Film Type Medium type
14" x 17"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
14" x 14"Portrait Yes
Scopix LT2B DLA1
Blue-based
Landscape No No No
11" x 14"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
10” x 12”Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
8" x 10"
Portrait YesScopix LT2B
DLA1Blue-based
Landscape YesScopix LT2B
DLA1Blue-based
Other n.a. N/A n.a.
Released: Yes - Film size is available and released
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 79 of 260
Page 80
80 Specific Camera Information - Agfa
Performance and Connections
Associations
Maximum number of associations supported: 15
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
Known effects
NOTE For page mode magnification, set type "NONE" and activatethe mark.
NOTE The camera does support format 14”X14” landscape.
SCU / Modality Software Configuration 0
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
Time 197 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title LRDICOM Yes
Port number 104 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 80 of 260
Page 81
Specific Camera Information - Agfa 81
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfalr5200 No
DICOM Node LR5200 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 80.0 x 80.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4224 x 4839
14" x 14"
4224 x 3897
11" x 14"
3268 x 3924
8" x 10"
2344 x 2647
In “define new”
Do not modify/add
Select only
Delete onlyFilm Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5140 x 3924
11" x 14"
4224 x 2962
8" x 10"
2948 x 2046
Medium type Blue film Yes
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 81 of 260
Page 82
82 Specific Camera Information - Agfa
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM print server/printer
Device name: Agfa LR DICOM Controller
Manufacturer: AGFA
SW Revision: FLC1.01.14
Physical Interface: Gateway
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. 1.2, July 31, 2003
Image memory: 128 MB RAM
Hard disk: 20 Gbyte
Device type: DICOM camera
Device name: LR 5200
Manufacturer: AGFA
SW Revision: LTP 01806.240698
Physical Interface: FOI optic
Network protocol: None
DICOM Conformance Statement: None
Image memory: n.a.
Hard disk: None
Selected Variable name Setting
n.a. Calling AE title PR_Hostname
n.a. Nickname CT hostname
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 82 of 260
Page 83
Specific Camera Information - Agfa 83
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID Deselected
No Medium type Deselected
n.a. Print even if Film Size ID / Medium Type is not supported
No
No Kernel Site
Magnification type: cubicHighRes
Smooth Factor: - 0.7
Yes Magnification Type: None
Max Custom scale factor: 1.0
No Pixel Size Ratio: 1.0 / 1.0
No Perception LUT Linear
No Kanamori-like value 0
No Custom Value Deselected
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2)
10
No Border density (OD x 100) Black
No Empty image density (OD x 100)
Black
No Minimum density Empty
No Maximum density Site 320
n.a. Header annotation
No Content: Edit custom: none
Courier 12
n.a. Footer annotation
No Content: Edit custom: none
Courier 12
Selected Variable name Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 83 of 260
Page 84
84 Specific Camera Information - Agfa
n.a. Window (%) Empty
n.a. Level (%) Empty
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo Site no.
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN
Yes
n.a. ALLOW EXPLICIT VR LIT-TLE ENDIAN
Yes
n.a. ALLOW EXPLICIT BIG ENDIAN
Yes
n.a. Resolution 315
n.a. Supported abstract syntax Basic Film Box_SOP_Class
Basic gray PM Meta_SOP_Class
Print Queue Management_SOP_INSTANCE
Printer_SOP_Class
Basic Color Image Box_SOP_CLass
Basic Grey Image_Box_SOP_Class
Basic Film Session_SOP_Class
Print Job_SOP_Class
Presentation LUT_SOP_Class
Basic Annotation Bo_ SOP_Class
Verification service class
Print Queue Management_SOP_Class
Selected Variable name Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 84 of 260
Page 85
Specific Camera Information - Agfa 85
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Action that results in camera error mes-
sages
Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP
Camera switched off or
Camera switched to offline
PRINTER OFFLINE
PRINTER OFFLINE
Camera offline
Camera start-up PRINTER INIT PRINTER INIT Camera initializing
Manual input of film in developer
CHECK PRINTER
CHECK PRINTER
Check camera
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 331.5 0.45
Y 370.0 373.0 0.81
14" x 17"
Landscape
X 400.0 403.0 0.75
Y 300.0 301.5 0.50
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 330.0 332.5 0.76
14" x 14"
Landscape
X No No No
Y No No No
11" x 14"
Portrait
X 250.0 252.0 0.80
Y 300.0 301.0 0.33
11" x 14"
Landscape
X 330.0 331.0 0.30
Y 220.0 221.5 0.68
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 85 of 260
Page 86
86 Specific Camera Information - Agfa
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X 170.0 170.0 0.00
Y 200.0 201.5 0.75
8" x 10"
Landscape
X 230.0 231.5 0.65
Y 160.0 160.5 0.31
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 86 of 260
Page 87
Specific Camera Information - Codonics 87
3- 3Specific Camera Information - CodonicsCodonics Horizon 0
General information 0
Block diagram
Fig. 11:
Available film sizes
Format Release status Film Type Medium type
14" x 17"
Portrait Yes DirectVistaGrayscale
Blue-based
Landscape Yes DirectVistaGrayscale
Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait No No No
Landscape No No No
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"
Portrait Yes DirectVistaGrayscale
Blue-based
Landscape Yes DirectVistaGrayscale
Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 87 of 260
Page 88
88 Specific Camera Information - Codonics
Performance and Connections
Associations
Maximum number of associations supported: More than 30
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE Film size 8” x 10” is not released for subdivision 1x1 with lin-ear LUT.
Film size 8” x 10” is not released also for subdivisionssmaller than 6 on 1 (4 on 1; 2 on 1; etc) with CT LUT.
NOTE Print-outs on paper are not released.
Known Effects
n/a
SCU / Modality Software Configuration 0
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
Time 95 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title HORIZON Yes
Port number 104 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 88 of 260
Page 89
Specific Camera Information - Codonics 89
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)codonicshorizon No
DICOM Node Horizon Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.4 x 79.4 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4322 x 5025
11" x 14"
3382 x 4022
8" x 10"
2406 x 2790
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5025 x 4322
11" x 14"
4022 x 3382
8" x 10"
2790 x 2406
Medium type Blue film Yes
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black No
Transformation No magnification No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 89 of 260
Page 90
90 Specific Camera Information - Codonics
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Manufacturer / device type: DICOM Printer (SCP)
Device name: Horizon
Manufacturer: Codonics
SW Revision: V 1.9.0
Physical type: Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: D.C.S., V1.9.0
Image memory: 256 MB RAM, 6 GB-HD
Hard disk: 10 GB
Content:
DV FILM BLUE Gamma:
Contrast:
Dmax:
Dmin:
Film View:
Illumination:
Reflected Light:
LUT:
Polarity:
Background:
Scaling:
Rotate:
Image Warnings:
Captions:
Border Fill:
Crop Anchor:
Scale Factor:
Coverage:
1.0
0
300
0
Front
2000
10
linear
Positive
Black
Bilinear
Auto
On
Off
On
Center
0.0
Standard
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 90 of 260
Page 91
Specific Camera Information - Codonics 91
Error / Warning Messages (examples)
Step in Service Mode (with "<" and "Menu")
select Job Settings
Select:
PR_hostname (AE title of modality)
Name:
Media
Receive Tray:
Priority:
Gamma:
Contrast:
Dmax:
Dmin:
Film View:
Illumination:
Reflected Light:
LUT:
Polarity:
Saturation:
MCM:
Background:
Scaling:
Rotate:
Image Warnings:
Captions:
Border Fill:
Crop Anchor:
Scale Factor:
Coverage:
PR_hostname
DV FILM BLUE 14X17
Default
Default
1.0
0
Default
Default
Front
Default
Default
linear
Positive
Default
Default
Black
Bilinear
Auto
On
Off
On
Center
Default
Standard
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Printer switched to service mode
PRINTER OFFLINE PRINTER OFFLINE Printer offline
Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open
Content:
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 91 of 260
Page 92
92 Specific Camera Information - Codonics
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Send job when no magazine is inserted
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
14”x17” blue film sup-ply empty
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.5 0.15
Y 380.0 380.5 0.13
14" x 17"
Landscape
X 390.0 390.5 0.13
Y 330.0 330.5 0.15
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 92 of 260
Page 93
Specific Camera Information - Codonics 93
8" x 10"
Portrait
X 170.0 170.0 0.00
Y 210.0 210.5 0.24
8" x 10"
Landscape
X 200.0 200.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 93 of 260
Page 94
94 Specific Camera Information - Fuji
4- 4Specific Camera Information - FujiFuji Drypix 7000 / 5000 0
General information 0
Block diagram
Fig. 12:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film Type Medium type
14" x 17"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
14" x 14"Portrait Yes DI-HL Blue-based
Landscape n.a. n.a. n.a.
11" x 14"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
10" x 12"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
8" x 10"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 94 of 260
Page 95
Specific Camera Information - Fuji 95
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Device is only class A, EN55011(Hospital use).
NOTE 14INX14IN landscape not supported by camera.
NOTE Film job that contains more than 10 films will not be printedout completely. (with SW-Revision 13.0-0001 only)
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 66 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY7000 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)fujidrypix7000 No
DICOM Node dry7000 Yes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 95 of 260
Page 96
96 Specific Camera Information - Fuji
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 100.0 x 100.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
3500 x 4170
14" x 14"
3500 x 3410
11" x 14"
2500 x 3495
10” x 12”
2460 x 2916
8" x 10"
1954 x 2410
In “define new”
Do not modify/add
Select only
Delete onlyFilm Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
4220 x 3360
11" x 14"
3600 x 2450
10” x 12”
2972 x 2404
8" x 10"
2466 x 1898
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 96 of 260
Page 97
Specific Camera Information - Fuji 97
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM Printer (SCP)
Device name: Drypix 7000
Manufacturer: FUJI
SW Revision: V 13.0-0001 (for VA20)
V 13.5 (for VA26)
Physical type: Ethernet, 10/100 base T/Tx, RJ45
Network protocol: TCP/IP
DICOM Conformance Statement: September 2003, 1st Edition.
Image memory: 512 MB RAM (standard)
Hard disk: 20 GB
File Name Content Remarks
Protocol: Client: PR_hostname
0107H warning Indicate
0116H warning Indicate
B604H warning Indicate
N-event report RQ Disabled
Use Extention format: 0: No
ID
Use default annotation: 0: No
Change film: 272:0110H[F]
Use System Timeout: 1:Yes
Presentation LUT 0:No
Precode picking up: 0:No
Use Max/Min Density 1:Modality
Logging Client: 0:No
Reprint: 0:No
Magnify:
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 97 of 260
Page 98
98 Specific Camera Information - Fuji
Procedure SSM
Default Smoothing Type SHARP
Adjustment Param. SHARP
Edge Detection 1:on (dimmed)
Detection level 600 (dimmed)
Sharpness 130 (dimmed)
Char. Mode White (dimmed)
Adjustment param. MEDIUM
Edge Detection 1:On (dimmed)
Detection level 600 (dimmed)
Sharpness 80 (dimmed)
Char. Mode 0:White (dimmed)
Adjustment param. SMOOTH
Edge Detection 1:On (dimmed)
Detection level 800 (dimmed)
Sharpness 30 (dimmed)
Char. Mode 0:White (dimmed)
Default Requested Decimate Crop Behavior:
1:Decimate
LUT:
Procedure SAR
Default LUT 1
LUT 1
r table number 17
Max. density 300
Min. density 0
Contrast 100 (dimmed)
Number of tuning points 0
Dmax3.6 0:Standard
Output format:
Border density 300
File Name Content Remarks
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 98 of 260
Page 99
Specific Camera Information - Fuji 99
Error / Warning Messages (examples)
Trim density 0
Trim 0:Off
Number of Copies 1
Margin between images 20
Image layout 0:Concentration
Film size 4:14 IN X 17 IN
Medium Type (base color): 1:Blue
Film orientation 1:Portrait
Mirror 0:Off
Film Destination:
Upper Tray: 1:use
Annotation
Annotation Height: 1:1
Annotation System:System
Annotation 14:14
Upper Left
Upper Center
Upper Right
Lower Left
Lower Center
Lower Right
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Front cover open COVER OPEN COVER OPEN Camera cover, drawer or door open
Camera in utility mode
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Camera in service mode (M-utility mode)
TCP/IP INIT ERROR TCP/IP INIT ERROR Connection to cam-era is not available
File Name Content Remarks
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 99 of 260
Page 100
100 Specific Camera Information - Fuji
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.5 0.15
Y 400.0 399.0 0.25
14" x 17"
Landscape
X 410.0 409.0 0.24
Y 320.0 320.5 0.16
14" x 14"
Portrait
X 330.0 330.5 0.15
Y 330.0 329.0 0.30
14" x 14"
Landscape
X No No No
Y No No No
11" x 14"
Portrait
X 230.0 230.0 0.00
Y 340.0 340.5 0.15
11" x 14"
Landscape
X 350.0 350.0 0.00
Y 230.0 230.0 0.00
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 280.0 280.5 0.18
10” x 12”
Landscape
X 290.0 290.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 190.0 190.0 0.00
Y 230.0 229.5 0.22
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 100 of 260
Page 101
Specific Camera Information - Fuji 101
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 101 of 260
Page 102
102 Specific Camera Information - Fuji
Fuji Drypix 4000 4.1
General information 0
Block diagram
Fig. 13:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film type Medium type
14" x 17"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
14" x 14"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
11" x 14"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
10" x 12"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
8" x 10"Portrait Yes DI-HL Blue-based
Landscape Yes DI-HL Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 102 of 260
Page 103
Specific Camera Information - Fuji 103
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE CISPR11 is Class A only (hospital use).
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 80 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY4000 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec. file name)fujidrypix4000 No
DICOM Node dry4000 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 100.0 x 100.0 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 103 of 260
Page 104
104 Specific Camera Information - Fuji
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
3500 x 4170
14" x 14"
3500 x 3410
11" x 14"
2510 x 3500
10" x 12"
2460 x 2916
8" x 10"
1954 x 2410
In “Define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
4220 x 3400
14" x 14"
3500 x 3410
11" x 14"
3600 x 2460
10" x 12"
2972 x 2404
8" x 10"
2466 x 1898
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 104 of 260
Page 105
Specific Camera Information - Fuji 105
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM Printer (SCP)
Device name: Drypix 4000
Manufacturer: FUJI
SW Revision: V 3.0 (for VA20)
V 3.5 (for VA26)
Physical type: Ethernet, (RJ45)
Network protocol: TCP/IP
DICOM Conformance Statement: 1st Edition, October 2004
Image memory: 512 MB RAM
Hard disk: 40 GB (25 GB for image storage)
File Name
Content / Remarks
Fuji printer PR_hostname
Protocol: Attribute List Error (0107H warning) 1:Indicate
Attribute Value Out of Range (0116H warning) 1:Indicate
Image size is larger than image box (B604H warning) 1:Indicate
N-event report RQ 0:Disabled
Use extension format 0:No
Use extension format --> ID (empty)
Use extension format--> Use Default Annotation 0:No
Change film 272:0110H[F]
Use System Timeout 1:Yes
Presentation LUT 0:No
Precede Picking up 1:Yes
Use Max/Min Density 1:Modality
Logging Client 0:No
Reprint 0:No
Change Base Color (Disable:110H[F]) 1:Enable
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 105 of 260
Page 106
106 Specific Camera Information - Fuji
Define Film Size of 11" x 14'' 1: 25.7 x 36.4 cm
Magnify: Procedure 0:A-VR
Default Smoothing Type 4:modality
Adjustment Parameter -> SHARP -> Edge Detection 1:On
Adjustment Parameter -> SHARP -> Detection level 600
Adjustment Parameter -> SHARP -> Sharpness 130
Adjustment Parameter -> SHARP -> Character Color 0:White
Adjustment Parameter -> MEDIUM -> Edge Detection 1:On
Adjustment Parameter -> MEDIUM -> Detection level 600
Adjustment Parameter -> MEDIUM -> Sharpness 80
Adjustment Parameter -> MEDIUM -> Character Color 0:White
Adjustment Parameter -> SMOOTH -> Edge Detection 1:On
Adjustment Parameter -> SMOOTH -> Detection level 800
Adjustment Parameter -> SMOOTH -> Sharpness 30
Adjustment Parameter -> SMOOTH -> Character Color 0:White
Default Requested Decimate / Crop Behavior 1:DECIMATE
Magnification Type 2:modality
Requested Image Size (0=None) 0
Requested Image Size (0=None) ' Default Requested Image Size
1:modality
LUT: Procedure 0:SAR
Default LUT 1
Default LUT -> Configuration Information (Default LUT#)
0:printer
Adjustment Parameter -> LUT1 -> r table number 17
Adjustment Parameter -> LUT1 -> Max. Density 300
Adjustment Parameter -> LUT1 -> Min. Density 20
Adjustment Parameter -> LUT1 -> 100
File Name
Content / Remarks
Fuji printer PR_hostname
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 106 of 260
Page 107
Specific Camera Information - Fuji 107
Adjustment Parameter -> LUT1 -> Number of tuning points
0
Dmax 3.6 0:Standard
Illumination 2000
Illumination -> Default Illumination 1:modality
Reflective Ambient Light 10
Reflective Ambient Light -> Default Ambient Light 1:modality
Output format:
Border density 280
Border density -> Default border density 1:modality
Polarity (Pos./Neg.) 1:normal (Pos.)
Trim density 0
Trim 0:Off
Trim Width 1
Number of Copies 1
Number of Copies -> Default Number of Copies 1:modality
Margin between images 20
Image layout 0:Concentration
Default film size 1:modality
Default Film Size -> Film Size 4: 14 IN X 17 IN
Default Medium Type 0:printer
Default Medium Type -> Medium Type (base color) 1: blue
Film orientation 3:modality
Mirror 0:OFF (Normal)
Processing Type 0:Standard
Print priority 1:HIGH
Resolution 0:Standard
Default film destination 0:printer
Processsor 1: use
File Name
Content / Remarks
Fuji printer PR_hostname
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 107 of 260
Page 108
108 Specific Camera Information - Fuji
Error / Warning Messages (examples)
Bin No 1 0: not use
Bin No 2 0: not use
Bin No 3 0: not used
Bin No 4 0: not used
Annota-tion
Annotation Height 5:modality
Annotation Font System:System
Annotation Size 14:14
Upper Left (empty)
Upper Center (empty)
Upper Right (empty)
Lower Left (empty)
Lower Center (empty)
Lower Right (empty)
Camera warning or error messages
Printer message Message displayed on SCU
Defined DCS (*1) Sent by SCP (*2)
Camera in utility mode
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Camera in service mode (M-utility mode)
TCP/IP INIT ERROR
TCP/IP INIT ERROR
Connection to cam-era is not available
Front door open COVER OPEN COVER OPEN Camera cover, drawer or door open
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
File Name
Content / Remarks
Fuji printer PR_hostname
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 108 of 260
Page 109
Specific Camera Information - Fuji 109
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 332.0 0.61
Y 400.0 402.5 0.63
14" x 17"
Landscape
X 410.0 413.0 0.73
Y 330.0 332.5 0.61
14" x 14"
Portrait
X 330.0 332.5 0.76
Y 330.0 332.5 0.76
14" x 14"
Landscape
X 330.0 332.0 0.61
Y 330.0 332.5 0.76
11" x 14"
Portrait
X 230.0 232.0 0.87
Y 340.0 342.0 0.59
11" x 14"
Landscape
X 350.0 351.5 0.43
Y 230.0 232.0 0.87
10” x 12”
Portrait
X 230.0 232.0 0.87
Y 280.0 281.5 0.54
10” x 12”
Landscape
X 290.0 291.5 0.52
Y 230.0 232.0 0.87
8" x 10"
Portrait
X 190.0 191.0 0.53
Y 230.0 232.0 0.87
8" x 10"
Landscape
X 230.0 232.0 0.87
Y 170.0 171.0 0.59
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 109 of 260
Page 110
110 Specific Camera Information - Fuji
Fuji Drypix 3000 4.2
General information 0
Block diagram
Fig. 14:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 6
Format Release status Film Type Medium type
14" x 17"Portrait Yes DI-AT Blue-based
Landscape Yes DI-AT Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 110 of 260
Page 111
Specific Camera Information - Fuji 111
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The DI-AT film type requires a specific densitometer charac-teristic. Density measurements using the X-Rite 301 densito-meter are only possible if the FUJI/X-Rite colorcompensating filter kit CM500 is installed on the X-Rite 301and calibrated as described in the X-Rite letter.
NOTE Additional annotation and trim have been switched off.
NOTE The background command does not work properly. The back-ground is always white.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 150 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRYPIX3000 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 111 of 260
Page 112
112 Specific Camera Information - Fuji
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Class
(Printer spec file name)fujidrypix3000 No
DICOM Node drypix3000 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 84.7 x 84.7 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4072 x 4891
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
4972 x 3993
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
HC Device /
General settingsDisplay Changes allowed
Device type: DICOM Printer (SCP)
Device name: Drypix 3000
Manufacturer: FUJI
SW Revision: A03 Main Unit / A04 LAN Board
Physical type: 100Base-Tx /10 Base T
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 112 of 260
Page 113
Specific Camera Information - Fuji 113
Printer settings
Network protocol: TCP/IP
DICOM Conformance Statement: 1st Edition March 31, 2001.
Image memory: Approx. 56 MB
Hard disk: None
File Name Content Remarks
None FUJI PRINTER Provided by FUJI
Function Properties
Protocol information Protocol = DICOM
Normal AE title = DRYPIX 3000
Role = SCP
Port no. = 104
Data transfer time-out = 300
Annotation Annotation size = 1:half
Association count 6
Logging DICOM 0 : None
Log Auto Save 0 : No
Default client 0 : No
Printer Model = DRYPIX
Manufacturer = FUJI PHOTO FILM CO, LTD
CT hostname None
Protocol: Attribute list error (0107H warning) = indicate
Attribute value out of range (0116H warning) = indicate
Image size larger than image box (B604H warning) = 1: Indicate
N-Event-Report-RQ = 0 disable
Change film size = 0: 0116H ()W
Use system time-out = 1: Yes
Use extension format Use default annotation = 1: No
Device type: DICOM Printer (SCP)
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 113 of 260
Page 114
114 Specific Camera Information - Fuji
Error / Warning Messages (examples)
Magnify: Procedure = 1:SSM
Default smoothing type = 1:SHARP
LUT: Procedure = 0: SAR
Default LUT = 1
r table no. = 17
Max. density = 300
Min. density = 0
Output format: Border density = 300
Polarity (Pos/Neg) = 1 : normal(Pos.)
Trim density = 0
Trim = 0: Off
Trim with = 1
Number of copies = 1
Margin between images = 20
Image layout = 0: Concentration
Film size = 1: 35x43 (14 in. x17 in.)
Medium type (base color) = 1 blue
Film orientation = 1: PORTRAIT
Mirror = 0: OFF(Normal)
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1)
Sent by SCP
Top cover open n/a UNKNOWN Unspecified problem with camera.
Remove tray n/a UNKNOWN Unspecified problem with camera.
Supply magazine empty n.a. SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
File Name Content Remarks
None FUJI PRINTER Provided by FUJI
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 114 of 260
Page 115
Specific Camera Information - Fuji 115
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 398.0 0.50
14" x 17"
Landscape
X 410.0 408.5 0.37
Y 320.0 320.0 0.00
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 115 of 260
Page 116
116 Specific Camera Information - Fuji
Fuji Drypix 2000 4.3
General Information 0
Block Diagram
Fig. 15:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film type Medium type
14" x 17"Portrait Yes DI-HT Blue-based
Landscape Yes DI-HT Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait Yes DI-HT Blue-based
Landscape Yes DI-HT Blue-based
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait Yes DI-HT Blue-based
Landscape Yes DI-HT Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 116 of 260
Page 117
Specific Camera Information - Fuji 117
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Test of line structure and artifacts not passed. Small horizon-tal and vertical line structure visible in technical images andfor all film sizes.
NOTE Electro Magnetic Compatibility(EN55011, CISPR 11) classifi-cation is not available. An installation is suggested only forIndustrial Areas (Hospital).
Known Effects
NOTE The measured values of the X-Rite 301 Densitometer must betranslated via the Fuji Conversion Table (Operating Manual).
NOTE Sporadic transportation problems with film size 8x10 and11x14. Install magazine at the top side to prevent this prob-lem.
SCU / Modality Software Configuration 0
Time 118 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DP2000 Yes
Port number 104 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 117 of 260
Page 118
118 Specific Camera Information - Fuji
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec. file name)fujidrypix2000 No
DICOM Node DRY2000 Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 84.7 x 84.7 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
4107 x 4858
11" x 14"
2938 x 4094
8" x 10"
2280 x 2812
In “Define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
4932 x 4016
11" x 14"
4188 x 2872
8" x 10"
2876 x 2215
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 118 of 260
Page 119
Specific Camera Information - Fuji 119
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer Settings
Device type: DICOM Printer (SCP)
Device name: Drypix 2000
Manufacturer: FUJI
SW Revision: V2.0
Physical type: Ethernet 10/100 Base T/Tx
Network protocol: TCP/IP
DICOM conformance statement: 1st Edition, June 2006
Image memory: 256 MB RAM
Hard disk: None
Fuji printer Content
Protocol: Attribute List Error (0107H warning) Indicate
N-event report RQ Disabled
Attribute Value Out of Range (0116H warning) Indicate
Image size is larger than image box (B604H warning) Indicate
Change film size 272:0110H[F]
Use System Timeout Yes
Presentation LUT No
Precede picking up No
Use max/min density Modality
Use extension format --> ID SM
Use extension format--> Use Default Annotation No
Change base color (disable:110H[F]) Enable
Define Film Size of 11" x 14'' 25.7 x 36.4 cm
AE Title PR_hostname
Magnify:
Default Smoothing Type sharp
Default requested decimate / crop behavior decimate
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 119 of 260
Page 120
120 Specific Camera Information - Fuji
Magnification Type cubic
Requested image size (0=None) 0
Default Magnification Type Modality
Default Smoothing Type Modality
Default Requested Image Size (0=None) Modality
Default requested decimate / crop behavior Modality
AE Title PR_hostname
Procedure AV-R
SHARP -> Edge Detection On
SHARP -> Detection level 600
SHARP -> Sharpness 130
SHARP -> Character Color White
SMOOTH -> Edge Detection On
SMOOTH -> Detection level 800
SMOOTH -> Sharpness 30
SMOOTH -> Character Color White
MEDIUM -> Edge Detection On
MEDIUM -> Detection level 600
MEDIUM -> Sharpness 80
MEDIUM -> Character Color White
LUT Com-mon:
Default LUT 1
Illumination 2000
Ambient light 10
Default LUT -> Configuration Information (Default LUT#) printer
Default Illumination Modality
Default Ambient Light Modality
AE Title PR_hostname
Procedure SAR
LUT1 -> Max. Density 280
Fuji printer Content
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 120 of 260
Page 121
Specific Camera Information - Fuji 121
LUT1 -> Min. Density 18
AE Title PR_hostname
LUT1 -> r table number 17
LUT1 -> Contrast 100
Number of tuning points 0
P1_Density 120
P2_Density 120
P3_Density 120
P4_Density 120
P5_Density 120
P1_Shift 0
P2_Shift 0
P3_Shift 0
P4_Shift 0
P5_Shift 0
P1_Contrast 100
P2_Contrast 100
P3_Contrast 100
P4_Contrast 100
P5_Contrast 100
Output format:
AE Title PR_hostname
Film Size 14INx17IN
Medium type blue
Film orientation portrait
Border density 280
Polarity (Pos./Neg.) normal(Pos.)
Trim Off
Trim Width 1
Trim density 0
Fuji printer Content
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 121 of 260
Page 122
122 Specific Camera Information - Fuji
Number of Copies 1
Print priority high
Default Number of Copies Modality
Default Print Priority Printer
Default Medium Type Printer
Default Film Orientation Modality
Default Film Size Modality
Default Border Density Printer
Default Trim printer
Default Polarity Modality
AE Title PR_hostname
Mirror OFF(Normal)
Margin between images 20
Image layout Concentration
Processing Type 2
Image Crop UPDWN Area (8x10) for CR Image both
Image Crop Side Area (8x10) for CR Image both
Image Crop UPDWN Area (10x14) for CR Image both
Image Crop Side Area (10x14) for CR Image both
Image Crop UPDWN Area (14x17) for CR Image both
Image Crop UPDWN Area (10x12) for CR Image both
Image Crop Side Area (10x12) for CR Image both
Annotation
Upper left empty
Upper center empty
Upper right empty
Lower left empty
Lower center empty
Lower right empty
Fuji printer Content
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 122 of 260
Page 123
Specific Camera Information - Fuji 123
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Camera warning or error messages
Printer message Message displayed on SCU
Defined DCS (*1) Sent by SCP (*2)
Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open
Remove Tray COVER OPEN COVER OPEN Camera cover, drawer or door open
Supply empty SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Film jam FILM JAM FILM JAM Film jam
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 398.0 0.50
14" x 17"
Landscape
X 410.0 407.5 0.61
Y 330.0 330.0 0.00
14" x 14"
Portrait
X n/a n/a n/a
Y n/a n/a n/a
14" x 14"
Landscape
X n/a n/a n/a
Y n/a n/a n/a
11" x 14"
Portrait
X 230.0 229.5 0.22
Y 340.0 340.0 0.00
11" x 14"
Landscape
X 350.0 350.0 0.00
Y 230.0 229.5 0.22
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 123 of 260
Page 124
124 Specific Camera Information - Fuji
10” x 12”
Portrait
X n/a n/a n/a
Y n/a n/a n/a
10” x 12”
Landscape
X n/a n/a n/a
Y n/a n/a n/a
8" x 10"
Portrait
X 190.0 189.5 0.26
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 124 of 260
Page 125
Specific Camera Information - Fuji 125
Fuji FM DPL/FN-PS551 4.4
General information 0
Block diagram
Fig. 16:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 6
Format Release status Film Type Medium type
14" x 17"Portrait Yes DI-AL Blue-based
Landscape Yes DI-AL Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait Yes DI-AL Blue-based
Landscape Yes DI-AL Blue-based
8" x 10"Portrait Yes DI-AL Blue-based
Landscape Yes DI-AL Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 125 of 260
Page 126
126 Specific Camera Information - Fuji
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE 2 cameras (each with a separate channel) can be connectedto PS551. Only one camera connection to FN-PS551 wasused for the qualification.
NOTE The European Standard EN55011 is supported only withClass A (Hospital use).
Known Effects
NOTE In a multi-modality environment one additional hardwarechannel must be installed at FM-DPL and a second printserver PS551 has to be connected for Mammography Modal-ities.
NOTE When the Laser Imager connected to the print server is in theoffline status being disabled to receive image data, FAILUREwill be returned.
SCU / Modality Software Configuration 0
Time 158 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title FMDPL Yes
Port number 104 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 126 of 260
Page 127
Specific Camera Information - Fuji 127
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)fujifmdpl No
DICOM Node FMDPL Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 100 x 100 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
3500 x 4170
11" x 14"
2538 x 3522
8" x 10"
1952 x 2410
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
4240 x 3442
11" x 14"
3600 x 2460
8" x 10"
2464 x 1898
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 127 of 260
Page 128
128 Specific Camera Information - Fuji
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM print server/printer
Device name: FN-PS551
Manufacturer: FUJI Photo Film
SW Revision: A 07
Physical Interface: 100/10 base T Ethernet, UTP/STP
Network protocol: TCP/IP
DICOM Conformance Statement: FN-PS551 May 2001, 4th edition
Image memory: 261.56 MB
Hard disk: 3 Gbyte
Device type: DICOM camera
Device name: FM - DPL
Manufacturer: FUJI Photo Film
SW Revision: W01-02 (Main Unit) Panel: Unit W00-00
Physical Interface: SCSI Interface
Network protocol: None
DICOM Conformance Statement: None
Image memory: 96 MB
Hard disk: None
Film size 14 IN X 17 IN
Image type: Select: Standard scan
Gradation type: Select: 3.0
File Name Content Remarks
None FUJI service utility setting Provided by FUJI
Function Properties
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 128 of 260
Page 129
Specific Camera Information - Fuji 129
Protocols: DICOM Properties Network port no. = 104
Data transfer time-out = 600 sec.
Number of associations = 6
Association establishment restriction by the memory = enable
Print server application entity = one to each printer
Data receive if the printer is offline = enable
Printer FMDPL Properties
Name: AE title of DICOM = FMDPL
Fine-printing = off
Interface: Interface = E-i/f
Pixel clock = 1.0 microsec/pixel
Version = type 1
Film size = 14 x 17 (dimmed)
Film medium = blue (dimmed)
Model: Model = FN-PS551
Manufacturer = FUJI PHOTO FILM CO, LTD
Clients: DICOM: CT hostname Properties
Name: Hostname = CT hostname
IP address = (from system administrator)
File Name Content Remarks
None FUJI service utility setting Provided by FUJI
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 129 of 260
Page 130
130 Specific Camera Information - Fuji
Protocol: Protocol = DICOM
Application Entity Title = PR_hostname
Attribute list error (0107H warning) = indi-cate
Attribute value out of range (0116H warn-ing) = indicate
Image size larger than image box (B604H warning) = indicate
N-event report RQ = disabled
Use presentation LUT = No
Data transfer time-out = enabled
Use extension format = No
Magnify: Procedure = SSM
Default smoothing type = SHARP
LUT: Procedure = SAR
Default LUT = 1 (linear)
Adj. parameter LUT = 1
r table no. = 17
Max. density = 300
Min. density = 0
Number of tuning points = 0
Printer: fixed, FMDPL
File Name Content Remarks
None FUJI service utility setting Provided by FUJI
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 130 of 260
Page 131
Specific Camera Information - Fuji 131
Output format 1: Border density = 300
Polarity = normal(Pos.)
Trim density = 0
Trim = Off
Trim width = 1 pixel
Number of copies = 1
Film size = 14x17
Medium type = blue
Film orientation = portrait
Mirror = OFF(normal)
Overlay threshold density = 120
Output format 2: Margin between images = 20 pixel
Image layout = concentration
Annotation: Print server Upper left = (nothing selected)
Upper center = (nothing selected)
Upper right = (nothing selected)
Upper left = (nothing selected)
Lower center = (nothing selected)
Lower right = (nothing selected)
Logging: Network = (Not selected)
Printer = (Not selected)
DICOM = normal
Output image to file = off
Precedence = Input
Sorting: Sorting conditions = (nothing selected)
2 cameras (each with a separate channel) can be connected to PS551. Only one printer connection is used for the test.
File Name Content Remarks
None FUJI service utility setting Provided by FUJI
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 131 of 260
Page 132
132 Specific Camera Information - Fuji
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Camera in utility mode
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Camera is switched off
n.a. UNKNOWN Unspecified problem with camera
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 340.0 339.0 0.29
Y 400.0 400.5 0.12
14" x 17"
Landscape
X 410.0 410.5 0.12
Y 330.0 329.0 0.30
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X 240.0 240.5 0.21
Y 340.0 338.5 0.44
11" x 14"
Landscape
X 350.0 348.5 0.43
Y 230.0 230.2 0.09
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 132 of 260
Page 133
Specific Camera Information - Fuji 133
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X 190.0 189.0 0.53
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.5 0.22
Y 180.0 179.0 0.56
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 133 of 260
Page 134
134 Specific Camera Information - Kodak
5- 5Specific Camera Information - KodakKodak DryView 6800 0
General Information 0
Block Diagram
Fig. 17:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point-to-point) was used.
Application Status Film type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
11" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
10" x 12"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
8" x 10"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
Others n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 134 of 260
Page 135
Specific Camera Information - Kodak 135
Restrictions
NOTE For restrictions not related to HCC, refer to (Restrictions andknown effects / p. 8).
NOTE For SW version V 1.00 only. (internal built-in MIM Core swv2.1.b12) The LUT file CTLUTdry has to be used to reach therequired image quality resolution in white areas.
NOTE No error message is sent if film size of job to be printed is notinserted in camera.
NOTE Min. Density is always fog on film.
NOTE A Film Jam error will not be sent to modality as specified.
NOTE The serrated border line and some pixels of the diagnosticarea are cut because of rounded corners.
NOTE The standard EN55011 (EMC) is supported with Class A (Hos-pital use) only.
Known effects
n.a.
SCU / modality software configuration 0
Time 65 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DV68001
1. Do not use an AE title such as “NER_...”
Yes
Port number 5040 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 135 of 260
Page 136
136 Specific Camera Information - Kodak
HC Device /
General settings Display Changes allowed
Type DICOM printer No
Class
(Printer spec file name)kodak6800_smallB No
DICOM node DV6800 Yes
Filming properties Display Changes allowed
Hold Printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.154 x 78.154 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
4488 x 5464
14" x 14"
4448 x 4448
11" x 14"
3514 x 4488
10" x 12"
3184 x 3828
08" x 10"
2528 x 3184
In “Define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17”
5464 x 4488
14" x 14"
4488 x 4488
11" x 14"
4488 x 3514
10" x 12"
3828 x 3184
08" x 10"
3184 x 2528
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale, 8-bit No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 136 of 260
Page 137
Specific Camera Information - Kodak 137
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Printer Settings
The following table describes the level (L = level) in the Kodak configuration tool “Kodak Weblink”.
Background Black Yes
Transformation Replicate No
Filming properties Display Changes allowed
Manufacturer / device type: DICOM Printer (SCP) (internal print server)
Device name: DryView 6800
Manufacturer: Kodak
SW Revision: V 1.00 (internal built-in MIM Core sw v2.1.b12);(for VA20)
V 2.00 (internal built-in MIM Core sw v2.1.b14) (for VA26)
Physical interface: Ethernet 10/100/1000 base T/Tx,
Network protocol: TCP/IP
DICOM conformance statement: Rev. 5.0, September 26, 2007
Image memory: 786 MB RAM
Hard disk: 40 GB
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 137 of 260
Page 138
138 Specific Camera Information - Kodak
L1 L2 L3 L4 Sub-Level of L4
Configuration
DICOM SCP
SCP Configuration
Image Optimization
Services Provided to SCU
- Negotiate PLUT / NO
- Negotiate Grayscale Print / YES
- Negotiate Color Print / NO
Image quality
- Tonescaling Method / Perception LUT - NO
- TFT Set Name / wrkstn2a.w87 - Always Apply
- Polarity / NORMAL - NO
- Interpolation / REPLICATE - NO
- Density Min / 0.20 - NO
- Density Max / 2.80 - NO
- Smoothing / 0 - NO
- Perception LUT Contrast / 3 - Always Apply
- Curve Shape Contrast / 0 - NO
- Curve Shape Value / 429 - NO
- Wet Pivot Density / 1.2 - NO
- Illumination / 2000 - NO
- Reflected Ambient Light / 10 - NO
Film and Image Options
Film and Page Options
- Film Size / 14”x17” – NO
- Film Base / BLUE FILM – NO
- Border Density / 3.10 – NO
- Sorter Bin / BIN 1 – NO
- Film Orientation / PORTRAIT – NO
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 138 of 260
Page 139
Specific Camera Information - Kodak 139
- Trim / NO – NO
Image Resizing
- Method / MINIFY-Scale – NO
- Text Box Location / NONE – Always Apply
Other Options
Status Reporting
- Send Warnings / YES - Always Apply
- Send N-Events to SCU / NO - Always Apply
Annotation Elements
- Print Date / NO - Always Apply
- Print Hospital Name / NO - Always Apply
- Print Modality ID / NO - Always Apply
- Print Time / NO - Always Apply
- Print Page Number / NO - Always Apply
- Common Text Override / NO - Always Apply
- Film Session Label / “empty” /NO
Printer and Other Parameters
- Printer Priority / NORMAL - NO
- Body Part / DEFAULT - NO
SCP Config AutoSetup
Modality Identification
- Source ID / 3.2
- Modality ID / xxxxxxx6800xxxxxx
- Modality Type / Other - NO
Modality Defaults
L1 L2 L3 L4 Sub-Level of L4
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 139 of 260
Page 140
140 Specific Camera Information - Kodak
Error and warning messages
Image Quality and Optical Density 0
No specific parameter available.
- Manufacturer / "empty"
- Type / "empty"
- Model / "empty"
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Camera cover, drawer or door open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Unlock magazine PRINTER OFFLINE
PRINTER OFFLINE
Camera offline
Send film to empty 14x17 supply maga-zine
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
Camera info: EMPTY 14X17 BLUE
Camera is warming up
PROC INIT PROC INIT Processor initializing
Camera in "power save" or "sleep mode"
PRINTER OFFLINE
PRINTER OFFLINE
Camera offline
Send job for a film size not inserted in camera
n.a. n.a. “nothing”
Camera is in calibra-tion
PRINTER OFFLINE
PRINTER OFFLINE
Camera offline
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
L1 L2 L3 L4 Sub-Level of L4
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 140 of 260
Page 141
Specific Camera Information - Kodak 141
Real Size Scaling
Film size / Direction Desired (mm) Measured (mm)
Difference obtained from desired value
in %
14" x 17"
Portrait
X 340.0 338.5 0.44
Y 400.0 400.0 0.00
14" x 17"
Landscape
X 400.0 400.2 0.05
Y 340.0 338.5 0.44
14" x 14"
Portrait
X 300.0 298.5 0.50
Y 300.0 300.0 0.00
14" x 14"
Landscape
X 300.0 300.0 0.00
Y 300.0 298.5 0.50
11" x 14"
Portrait
X 260.0 260.5 0.19
Y 340.0 338.5 0.44
11" x 14"
Landscape
X 340.0 338.5 0.44
Y 260.0 260.5 0.19
10" x 12"
Portrait
X 240.0 238.5 0.62
Y 280.0 280.2 0.07
10" x 12"
Landscape
X 280.0 280.2 0.07
Y 240.0 238.5 0.62
8" x 10"
Portrait
X 160.0 160.0 0.00
Y 240.0 238.5 0.62
8" x 10"
Landscape
X 240.0 238.5 0.62
Y 160.0 160.0 0.00
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 141 of 260
Page 142
142 Specific Camera Information - Kodak
Kodak DryView 5800 5.1
General information 0
Block Diagram
Fig. 18:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Application Status Film type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
11" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
10" x 12"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
8" x 10"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
Others n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 142 of 260
Page 143
Specific Camera Information - Kodak 143
Restrictions
NOTE For restrictions not related to HCC, refer to (Restrictions andknown effects / p. 8).
NOTE Min. Density is always fog on film.
NOTE Some dust artifacts (less then five) visible on film.
NOTE The EN55011 standard is class A (Hospital use) only.
NOTE No error message is sent if film size of job to be printed is notinserted in camera.
NOTE If the film supply is empty, no messages are displayed in“Film Task Status”
Known effects
n.a.
SCU / modality software configuration 0
Time 92 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV58001
1. Do not use an AE title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settings Display Changes allowed
Type DICOM printer No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 143 of 260
Page 144
144 Specific Camera Information - Kodak
Class
(Printer spec file name)kodak5800 No
DICOM node DV5800 Yes
Filming properties Display Changes allowed
Hold Printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.154 x 78.154 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
4412 x 5384
14" x 14"
4412 x 4412
11" x 14"
3436 x 4412
10" x 12"
3104 x 3752
8" x 10"
2452 x 3104
In “Define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17”
5384 x 4412
14" x 14"
4412 x 4412
11" x 14"
4412 x 3436
10" x 12"
3752 x 3104
8" x 10"
3104 x 2452
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale, 8-bit No
Background Black Yes
Transformation No magnification No
HC Device /
General settings Display Changes allowed
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 144 of 260
Page 145
Specific Camera Information - Kodak 145
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Printer Settings
The following table describes the Kodak configuration tool “Kodak Weblink”.
Manufacturer / device type: DICOM Printer (SCP) (internal print server)
Device name: DryView 5800
Manufacturer: Carestream (Kodak)
SW Revision: V1.01.b01
(internal built-in MIM Core SW V3.0.b11)
Physical interface: Ethernet 10/100 base T/Tx, RJ45 connector
Network protocol: TCP/IP
DICOM conformance statement: Rev. 2.0, December 12, 2007
Image memory: 512 MB RAM
Hard disk: 80 GB
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 145 of 260
Page 146
146 Specific Camera Information - Kodak
Name (on Configura-tion)
Setting
System, Network
Host Name DV5800DRE
IPAddr accordingly
Subnet Mask accordingly
Default Gateway accordingly
Connection Name LAN
DICOM SCP, Communications
Dicom Source Communications
Port 5040
Association Limit 12
DICOM SCP, Destination
Identification AETiltle=DV5800
Media:
Media Type avail./PageExtent(x,y)/ImageExtent(x,y)/ImageOrigin(x,y)
8x10, blue&clear selected 2590x3264, 2452x3107, 74, 69
10x12, blue&clear selected 3264x3890, 3107x3752, 69, 64
11x14, blue&clear selected 3575x4576, 3437x4412, 74, 69
14x14, blue&clear selected 4576x4550, 4412x4412, 69, 64
14x17, blue&clear selected 4576x5525, 4412x5387, 69, 64
Page Image Options:
Min Vert Sep=2, Min Horiz Sep=2
Max Band Page=10, Max Img Band=10
Pixel Pitch=12.795, Border DMax=3.10
Image Dmin=0.00, Image DMax=3.10
Lines Image Annot=1, Lines Page Annot=2
Image Annot Height=52, Page Annot Height=52
Max Magnif.Factor=20, Max Minif.Factor=0.1
Trim Width=2, Accept Annotations=Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 146 of 260
Page 147
Specific Camera Information - Kodak 147
Accept Color Image=Yes, Accept BW Image=Yes
Mix Aspect Ratio Page=Y, Mix Aspect Ratio Row=Y
Mix Image Size Std Fmt=Y, Mix Image Size Row=Y
Mix Req.Size Page=Y, Minify Image To Fit=Y
Scale Image To Fit=Y, Crop Image To Fit=Y
Req.Size OK=Y, Configuration Notes=empty
DICOM SCP, select specific modality, Modify:
SCP Configuration, Image Optimization
Services Provided to SCU
Negotiate Presentation LUT
No
Negotiate Color Print No
Image quality
Smoothing Type 0 - Always Apply
Interpolation NONE - No
TFT SeT Wrkstn2a.W87 -Always Apply
Perception LUT Con-trast
1 - Always Apply
Image Polarity Normal - No
Image DMin 0.00 - No
Image DMax 3.10 - No
Illumination 2000 - No
Reflected Ambient Light 10 - No
SCP Configuration, Film and Image Options
Film and Page Options
Film size 14IN x 17IN - No
Film Base Blue Film - No
Film orientation Portrait - No)
Name (on Configura-tion)
Setting
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 147 of 260
Page 148
148 Specific Camera Information - Kodak
Border density 3.10 - No
Trim No - No
Image Resizing
Method Minify/Scale - No
SCP Configuration, Other Options
Status Reporting
Report Warnings YES - Always Apply
Send N-Events to SCU No Always Apply
Film Session Elements
Capture Date No
Capture Time No
Hosp.Name No
Study Page Number No
Modality ID No
Film Session Label Override
No
Printer and Other Parameters
Printer Priority Normal - No
SCP Configuration, Modality Preferences
Modality Identification
Modality ID PR_ Hostname
Modality Type Other - No
Modality Defaults
Manufacturer “empty”
Type “empty”
Model “empty”
Name (on Configura-tion)
Setting
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 148 of 260
Page 149
Specific Camera Information - Kodak 149
Error and warning messages
Image Quality and Optical Density 0
No specific parameter available.
Real Size Scaling
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Camera is warming up PROC INIT PROC INIT Processor initializing
Film 14x17 supply empty
EMPTY 14x17 BLUE
NORMAL Camera is ready
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Send job for a film size not inserted in camera
n/a n/a Camera is ready
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Difference obtained from desired value
in %
14" x 17"
Portrait
X 320.0 320.0 0.00
Y 410.0 410.0 0.00
14" x 17"
Landscape
X 410.0 410.0 0.00
Y 330.0 329.5 0.15
14" x 14"
Portrait
X 330.0 330.0 0.00
Y 330.0 330.5 0.15
14" x 14"
Landscape
X 330.0 330.5 0.15
Y 330.0 329.5 0.15
11" x 14"
Portrait
X 260.0 260.0 0.00
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 320.0 320.0 0.00
Y 260.0 260.5 0.19
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 149 of 260
Page 150
150 Specific Camera Information - Kodak
10" x 12"
Portrait
X 230.0 230.0 0.00
Y 280.0 280.5 0.18
10" x 12"
Landscape
X 280.0 280.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 229.5 0.22
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Difference obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 150 of 260
Page 151
Specific Camera Information - Kodak 151
Kodak DryView 8900 (small border) 5.2
General information 0
Block diagram
Fig. 19:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Format Release status Film Type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
11" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
10" x 12"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
8" x 10"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
Other N/A N/A N/A
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 151 of 260
Page 152
152 Specific Camera Information - Kodak
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.
NOTE The round edge of the film sheet cuts the diagnostic area.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 71 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV89001
1. Please do not use an AE Title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)kodak8900_smallB No
DICOM Node DV8900 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 152 of 260
Page 153
Specific Camera Information - Kodak 153
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.154 x 78.154 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4488 x 5464
14" x 14"
4488 x 4488
11" x 14"
3514 x 4488
10" x 12"
3184 x 3828
8" x 10"
2528 x 3184
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5464 x 4488
14" x 14"
4488 x 4488
11" x 14"
4488 x 3514
10" x 12"
3828 x 3184
8" x 10"
3184 x 2528
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation Replicate No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 153 of 260
Page 154
154 Specific Camera Information - Kodak
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Manufacturer / device type: DICOM Print Server (SCP)
Device name: DryView 8900
Manufacturer: Kodak
SW Revision: V 3.17, MIM 5.6.b14 with WIN 2000
Physical type: Ethernet 10/100 /1000 base T/Tx, RJ45
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. C, January 28, 2004.
Image memory: 786 MB RAM
Hard disk: 40 GB
Content
8900 setup: (Provided by Kodak)
under Configuration, System, Network:Host Name=DV8900, IPAddr=accordinglySubnet Mask=accordingly, Gateway Addr=accordinglyunder Configuration, DICOM SCP, Communications:Port=5040, Association Limit=12, Source Name=DICOMunder Configuration, Destination:General Remarks: Log. Name=8900Media types: Media=35x43cm, 9100x11037, 10781, anyMedia types: Media=35x35 cm, 9100x9087, 8831, AnyMedia types: Media=11"x14", 7137x9100, 8844, AnyMedia types: Media=10"x12", 6490x7767, 7511, AnyMedia types: Media=8"x10", 5167x6490, 6234, AnyFormatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=OnFormatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=OnFormatting: DICOM Annotation Box=On, Req. Image Size=OnFormatting: Req. Size Scale=On, Req. Size Mix Page=OnFormatting: Req. Size Mix Row=On, Trim=On,Formatting: True Landscape=On, Rotation=OnFormatting: Minify/Crop=On, Tone Scaling on Image Basis=OnFormatting: Max Img Band=20, Max Band Page=20, Max Img Cols=0Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4Formatting: Min Vert Sep=4, Trim Width Enabled=4Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 154 of 260
Page 155
Specific Camera Information - Kodak 155
Error / Warning Messages (examples)
Content
8900 setup: (Provided by Kodak)
under Configuration, DICOM SCP, SCP Services,Negotiate Presentation LUT=NO,TFT/ULUT= Wrkstn2a.w87 (Always=Y)Contrast= 3 (Always=Y),Smoothing type=0 (Always=Y),Polarity= Normal, (Always=N)Image Dmin=0.00 (Always=N),Image Dmax=3.00 (Always=N)NeventReports=No (Always=Y), Warnings=YesPrint Priority=Normal (Always=N),Border=3.99 (Always=N)2mmBorder=yesFilm Clip Border = NoFilm Base=Blue Film (Always=N),Film Size=14INx17IN (Always=N)Sorter Bin=Bin1 (Always=N),Include print date in annotation=NoInclude print time in annotation=NoInclude page number in annotation=NoInclude hospital name in annotation=NoInclude modality ID in annotation=NoModality ID=PR_ xxx.xxx.xxxAnnotation Override=No,Density Patch = NoImage Resizing=Minify/Scale (Always=N),Text Box Location=None, Source ID=DI016
Action that results in camera error messages
Printer messagesMessages displayed
on SCU (*2)Defined DCS (*1) Sent by SCP
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Camera is warming up
PROC INIT PROC INIT Processor initializing
Supply magazine empty
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
Camera info: EMPTY 14X17 BLUE
Magazine with damaged chip inserted
BAD SUPPLY MGZ BAD SUPPLY MGZ Problem with supply magazine
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 155 of 260
Page 156
156 Specific Camera Information - Kodak
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Action that results in camera error messages
Printer messages Messages displayed on SCU (*2)Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 340.0 340.0 0.00
Y 410.0 411.0 0.24
14" x 17"
Landscape
X 400.0 401.0 0.25
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 340.0 339.5 0.15
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 340.0 340.0 0.00
Y 330.0 330.0 0.00
11" x 14"
Portrait
X 270.0 270.0 0.00
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 340.0 340.0 0.00
Y 250.0 250.0 0.00
10” x 12”
Portrait
X 240.0 240.0 0.00
Y 280.0 280.0 0.00
10” x 12”
Landscape
X 290.0 290.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 190.0 190.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 240.0 240.0 0.00
Y 180.0 180.0 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 156 of 260
Page 157
Specific Camera Information - Kodak 157
Kodak DryView 8700-Plus / PACS Link MIM 200 5.3
General information 0
Block diagram
Fig. 20:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 13
Format Release status Film Type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 157 of 260
Page 158
158 Specific Camera Information - Kodak
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Camera does not support minimum density command.
NOTE The background command does not work properly. Cameradoesn't support background white.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 120 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV87001
1. Please do not use an AE title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)imation8700 No
DICOM Node DV8700 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 158 of 260
Page 159
Specific Camera Information - Kodak 159
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.0 x 78.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4096 x 5220
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5220 x 4096
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Device type: DICOM Print Server (SCP)
Device name: PACS Link MIM 200
Manufacturer: Kodak
SW Revision: V 6.1.1
Physical Interface: Ethernet (RJ45)
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. J., June 20, 2005
Image memory: 256 MB
Hard disk: 4 GB
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 159 of 260
Page 160
160 Specific Camera Information - Kodak
Printer settings
Device type: Printer
Device name: DryView 8700
Manufacturer: Kodak
SW Revision: 1.34
Physical Interface: Fiber optic
Network protocol: None
DICOM Conformance Statement: None
Image memory: 32 MB
Hard disk: n.a.
Content: Configuration
MIM 200 setup
Under Configure, Network:
Node Name=MIM, IPAddr=accordingly
Subnet Mask=accordingly, Gateway Addr=accordingly
under Configure, Input, Network, Add:
Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
under Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 160 of 260
Page 161
Specific Camera Information - Kodak 161
General: Log.Name=8700General: Database Name=8700, Model=8700, Memory=32Media info.:Media=35x43cm, 4096x5223, 5133, blueOutput parameters.:Cmd=SuperSet, Baud=19200, Data=8,Stop=1, Parity=NoneOutput parameters.:Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=OnFormat types: Custom=On, Slide=On, Row Symmetric=OnFormat types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1M952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
Under Configure, SCP Services, select the specific modality, modify:
Content: Configuration
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 161 of 260
Page 162
162 Specific Camera Information - Kodak
Error / Warning Messages (examples)
Image resizingResizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT= Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=YInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic Color=N
8700 Plus setup
under COMM Menu: 969 default
under IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
under EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Time-out=60
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Magazine empty SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Open Top Cover COVER OPEN COVER OPEN Camera cover, drawer or door open
Camera switched off PRINTER DOWN PRINTER DOWN Camera down
No supply magazine inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
Camera is warming up
PRINTER BUSY PRINTER BUSY Camera initializing
Film jam in camera FILM JAM FILM JAM Film jam
Content: Configuration
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 162 of 260
Page 163
Specific Camera Information - Kodak 163
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 310.0 310.5 0.16
Y 390.0 390.0 0.00
14" x 17"
Landscape
X 390.0 390.5 0.13
Y 310.0 310.5 0.16
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 163 of 260
Page 164
164 Specific Camera Information - Kodak
Kodak DryView 8700-Dual / PACS Link MIM 9410 5.4
General information 0
Block diagram
Fig. 21:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film Type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 164 of 260
Page 165
Specific Camera Information - Kodak 165
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Camera does not support minimum density command.
NOTE The background command does not work properly. Cameradoesn't support background white.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 128 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title 9410 Yes
Port number 1024 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)imation8700 No
DICOM Node DV8700 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 165 of 260
Page 166
166 Specific Camera Information - Kodak
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Pixel Size [1/1000 mm] 78.0 x 78.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4096 x 5220
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5220 x 4096
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Manufacturer / device type: DICOM Print Server (SCP) Printer
Device name: PACS Link 9410 DryView 8700Dual
Manufacturer: Kodak Kodak
SW Revision: V 1.6 1.34
Physical type: 100/10 base T Ethernet, UTP/STP Fiber Optic
Network protocol: TCP/IP n.a.
DICOM Conformance State-ment:
Version 2.0, June 5, 2000 n.a.
Image memory: 128 MB 32 MB
Hard disk: 4.1 GB n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 166 of 260
Page 167
Specific Camera Information - Kodak 167
Printer settings
Error / Warning Messages (examples)
Content: Configuration
MIM 9410 setup
under Device Setup, LOCAL PRINTERS::
Printer=8700, Model=8700 Dual, Output=Fiber,
Baud=19200, Parity=None, Data=8, Stop=1, Command Set=Superset
Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
Film Mismatch=Print Always, Interpolation Type=831
Interp Num=2, Max.Density=300, Contrast=6, Border=0
TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
under Device Setup, SCP PRINTERS, Current Providers:
Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 10
under Device Setup, SCP PRINTERS, Registered Users:
SCU AETitle= PR_cthostname, IPAddr=accordingly
SCP=8700, Copies=1, Contrast Table=1, TFT/ULUT= Wrkstn2a.w87
Max.Density= 280, Border Density=0, Mag.Type=8, Smoothing=0
Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
Trim=No, Attribute Warning=Yes
Contrast Test Mode=No, Contrast Test Mode Density=0
8700 Plus setup
under COMM Menu: 969 default
under IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
under EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Time-out=60
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Camera switched off PRINTER DOWN PRINTER DOWN Camera down
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
No supply magazine inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 167 of 260
Page 168
168 Specific Camera Information - Kodak
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera is in calibra-tion
CALIBRATING ERR
CALIBRATING ERR
Problem when cali-brating Film quality may not be optimal
Camera is warming up
PRINTER INIT PRINTER INIT Camera initializing
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 300.0 300.5 0.17
Y 400.0 400.2 0.05
14" x 17"
Landscape
X 400.0 400.2 0.05
Y 300.0 300.5 0.17
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 168 of 260
Page 169
Specific Camera Information - Kodak 169
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 169 of 260
Page 170
170 Specific Camera Information - Kodak
Kodak DryView 8200 / PACS Link MIM 200 5.5
General information 0
Block diagram
Fig. 22:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Format Release status Film Type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
11" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 170 of 260
Page 171
Specific Camera Information - Kodak 171
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The minimum density command is not supported.
NOTE The background command does not work properly. The bor-der is always black.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 228 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV82001
1. Please do not use an AE title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class (printer spec file name)
kodak8200 No
DICOM Node Kodak8200 Yes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 171 of 260
Page 172
172 Specific Camera Information - Kodak
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.0 x 78.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4360 x 5220
14" x 14"
4360 x 4244
11" x 14"
3272 x 4360
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5220 x 4360
14" x 14"
4244 x 4360
11" x 14"
4360 x 3272
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Device type: DICOM Print Server (SCP)
Device name: PACS Link MIM 200
Manufacturer: Kodak
SW Revision: V 6.1.1
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 172 of 260
Page 173
Specific Camera Information - Kodak 173
Printer settings
Physical Interface: Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: DCM, Rev. J, June 20, 2005
Image memory: 256 MB RAM
Hard disk: 4 GB
Device type: Printer
Device name: DryView 8200
Manufacturer: Kodak
SW Revision: 3.5
Physical Interface: 100/10 base T Ethernet
Network protocol: TCP/IP
DICOM Conformance Statement: n.a.
Image memory: 50 MB
Hard disk: n.a.
Content: Configuration
MIM 200 setup
Under Configure, Network:
Node Name=MIM, IPAddr=accordingly
Subnet Mask=accordingly, Gateway Addr=accordingly
under Configure, Input, Network, Add:
Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
under Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
Device type: DICOM Print Server (SCP)
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 173 of 260
Page 174
174 Specific Camera Information - Kodak
General Remarks:Log.Name=8200, IPAddr=100.100.100.40General Remarks:Database Name=8200, Model=8200, Memory 32Media info.:Media=35x43cm, 4361x5223, 5133, Any (Blue, Clear)Media=35x35cm, 4361x4247, 4157, any (blue, clear)Media=11inx14in, 3272x4361, 4271, any (blue, clear)Supported item:Collation=On, Text=On, Tone Scaling=On,Supported item:Remote Image Deletion=On, Rotation=OnInterpolation:No Magnification=On, Bilin=On, Repl=On,Interpolation:CubicSpline=On, Cubic Convolution 2=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1SCU:Density Space=On, Variable Border=On, PLUT Service=OnSCU:Curve Shape Tone Scaling=On, Perception LUT Tone Scaling=OnM952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
Content: Configuration
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 174 of 260
Page 175
Specific Camera Information - Kodak 175
Error / Warning Messages (examples)
Under Configure, SCP Services, select the specific modality, modify:
Image Resizing, Resizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT=Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=NInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic Color=N
8200 setup under the CONFIGURATION menu:
On IMAGE QUALITY-> TFT FILES menu: TFT=Ver693c0.w87
under the LOCAL PANEL Menu: Contrast=6, Density=3.00
under the DICOM Menu:
SCU config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
DICOM config.: AE=ANY_ANY, Port=1025
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Supply magazine not inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
Less than 5 films in supply magazine
SUPPLY LOW SUPPLY LOW Film supply low
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Camera is warming up
PRINTER INIT
PROC INIT
PRINTER INIT
PROC INIT
Camera initializing
Processor initializing
Camera switched off PRINTER DOWN PRINTER DOWN Camera down
Content: Configuration
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 175 of 260
Page 176
176 Specific Camera Information - Kodak
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 398.5 0.38
14" x 17"
Landscape
X 390.0 389.5 0.13
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 330.0 0.00
Y 320.0 319.5 0.16
14" x 14"
Landscape
X 320.0 319.5 0.16
Y 330.0 330.0 0.00
11" x 14"
Portrait
X 250.0 249.5 0.20
Y 330.0 330.5 0.15
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 240.0 240.0 0.00
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 176 of 260
Page 177
Specific Camera Information - Kodak 177
Kodak DryView 8100 / PACS Link MIM 200 5.6
General information 0
Block diagram
Fig. 23:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Format Release status Film Type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 177 of 260
Page 178
178 Specific Camera Information - Kodak
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The minimum density command is not supported.
NOTE The background command does not work properly. The Bor-der Density is always black.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 234 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV81001
1. Please do not use an AE title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)kodak8100 No
DICOM Node Kodak8100 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 178 of 260
Page 179
Specific Camera Information - Kodak 179
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.0 x 78.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4096 x 5216
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5216 x 4096
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Device type: DICOM Print Server (SCP)
Device name: PACS Link MIM 200
Manufacturer: Kodak
SW Revision: V 6.1.1
Physical Interface: Ethernet 10/100base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. J, June 20, 2005
Image memory: 256 MB RAM
Hard disk: 4 GB
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 179 of 260
Page 180
180 Specific Camera Information - Kodak
Printer settings
Device type: Printer
Device name: DryView 8100
Manufacturer: Kodak
SW Revision: 2.7
Physical Interface: n.a.
Network protocol: n.a.
DICOM Conformance Statement: n.a.
Image memory: 32 MB
Hard disk: None
Content: Configuration
MIM 200 setup
Under Configure, Network:
Node Name=MIM, IPAddr=accordingly
Subnet Mask=accordingly, Gateway Addr=accordingly
under Configure, Input, Network, Add:
Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
under Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 180 of 260
Page 181
Specific Camera Information - Kodak 181
General Remarks:Log.Name=8100,General Remarks:Database Name=8100, Model=8100, Memory 32Media info.:Media=35x43cm, 4096x5223, 5133, blueOutput parameters.:Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=EvenOutput parameters.:Pixel=12, Header/Line, Transfer=1000 KHz, Acq=60Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation:NoMag=On, Bilin=On, Repl=On, CubicSpline=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1M952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
Under Configure, SCP Services, select the specific modality, modify:
Content: Configuration
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 181 of 260
Page 182
182 Specific Camera Information - Kodak
Error / Warning Messages (examples)
Image resizingResizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT=Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=YInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic color=N
8100 setup under the CONFIGURATION menu:
On SIB board (i.e. external switch below control plugs):
SW1=Down
Under the COMMUNICATIONS Menu:
Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,... P6=0
On IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
On IMAGE QUALITY-> TFT FILES menu: TFT=Ver693c0.w87
under the LOCAL PANEL Menu: Contrast=6, Density=3.00
Under the SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
Under the DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Top cover or door open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Supply magazine empty
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Content: Configuration
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 182 of 260
Page 183
Specific Camera Information - Kodak 183
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Supply magazine not inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 310.0 310.0 0.00
Y 400.0 401.5 0.37
14" x 17"
Landscape
X 390.0 391.5 0.38
Y 310.0 310.5 0.16
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 183 of 260
Page 184
184 Specific Camera Information - Kodak
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 184 of 260
Page 185
Specific Camera Information - Kodak 185
Kodak DryView 8150 5.7
General information 0
Block diagram
Fig. 24:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 12
Format Release status Film type Medium type
14" x 17"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
14" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
11" x 14"Portrait Yes DVB Blue-based
Landscape Yes DVB Blue-based
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 185 of 260
Page 186
186 Specific Camera Information - Kodak
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The European Standard EN55011 is supported only withClass A (hospital use).
NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 146 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DV81501
1. Please do not use an AE Title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settings Display Changes allowed
Type DICOM Printer No
Class
(Printer spec file name)kodak8150 No
DICOM node DV8150 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 186 of 260
Page 187
Specific Camera Information - Kodak 187
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.12 x 78.12 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4438x5390
14" x 14"
4438x4414
11" x 14"
3440x4438
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5390x4438
14" x 14"
4414x4438
11" x 14"
4438x3440
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
Manufacturer / device type: DICOM Printer (SCP)
Device name: DryView 8150
Manufacturer: Kodak
SW Revision: V1.1.b04 / MIM_V5.7.b12
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 187 of 260
Page 188
188 Specific Camera Information - Kodak
Printer settings
Physical type: Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. 1.0, July 27, 2004
Image memory: 786 MB RAM
Hard disk: 40 GB
DICOM SCP / Communications
Port number 5040
Association Limit 12
Source Name DICOM
DICOM SCP / Cropping
No settings available
DICOM SCP / SCP Services / Image Optimization
Negotiate Presentation LUT No
Negotiate Color Print No
TFT Set Name LINODS.w87
Always Apply TFT Set Yes
Contrast 4
Always Apply Contrast Yes
Smoothing Type 0
Always Apply Smoothing Type No
Image DMin 0.00
Always Apply Image DMin No
Image DMax 3.00
Always Apply Image DMax No
Image Polarity Normal
Always Apply Image Polarity No
Manufacturer / device type: DICOM Printer (SCP)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 188 of 260
Page 189
Specific Camera Information - Kodak 189
DICOM SCP / SCP Services / Other Options
Send N-Event Reports No
Always Apply Send N-Event Reports Yes
Send Warnings Yes
Print priority Normal
Always Apply Print Priority No
Border Type 3.99
Always Apply Border Type No
2mm Border No
Film Clip Border No
Film Base Blue Film
Always Apply Film Base No
Film Size 14 IN X 17 IN
Always Apply Film Size No
Include Print date in Annotation No
Include Print time in Annotation No
Include Page Number in Annotation No
Include Hospital Name in Annotation No
Include modality ID in Annotation No
Modality ID PR_hostname
Annotation Override No
Density Patch No
Image Resizing Minify/Scale
Always Apply Image Resizing No
Text Box Location None
Source ID DI000
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 189 of 260
Page 190
190 Specific Camera Information - Kodak
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Supply magazine empty
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
Camera info:
EMPTY 14X17 BLUE
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Magazine with dam-aged chip inserted
BAD SUPPLY MGZ BAD SUPPLY MGZ Problem with supply magazine
Camera in calibration PRINTER DOWN PRINTER DOWN Camera offline
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 328.5 0.45
Y 410.0 409.5 0.12
14" x 17"
Landscape
X 410.0 410.0 0.00
Y 330.0 328.5 0.45
14" x 14"
Portrait
X 330.0 327.5 0.76
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 327.5 0.76
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 330.0 327.5 0.76
11" x 14"
Landscape
X 330.0 327.5 0.76
Y 250.0 250.0 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 190 of 260
Page 191
Specific Camera Information - Kodak 191
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 191 of 260
Page 192
192 Specific Camera Information - Kodak
Kodak KELP 2180/PACS Link MIM 200 5.8
General information 0
Block diagram
Fig. 25:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film Type Medium type
14" x 17"Portrait Yes EHN-10 Blue-based
Landscape Yes EHN-10 Blue-based
14" x 14"Portrait Yes DHG-10 Blue-based
Landscape Yes DHG-10 Blue-based
11" x 14"Portrait Yes DHG-10 Blue-based
Landscape Yes DHG-10 Blue-based
10" x 12"Portrait N/A N/A N/A
Landscape N/A N/A N/A
8" x 10"Portrait Yes DHG-10 Blue-based
Landscape Yes DHG-10 Blue-based
Other N/A N/A N/A
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 192 of 260
Page 193
Specific Camera Information - Kodak 193
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 115 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title K21801
1. Please do not use an AE title such as “NER_...”
Yes
Port number 5040 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class (printer spec file name)
kodak2180 No
DICOM Node Kelp 2180 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.0 x 79.0 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 193 of 260
Page 194
194 Specific Camera Information - Kodak
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4096 x 5156
14" x 14"
4096 x 4132
11" x 14"
3236 x 4132
8" x 10"
2284 x 2836
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5156 x 4096
14" x 14"
4132 x 4096
11" x 14"
4132 x 3236
8" x 10"
2836 x 2286
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM Print Server (SCP)
Device name: PACS Link MIM 200
Manufacturer: Kodak
SW Revision: V 6.1.1
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 194 of 260
Page 195
Specific Camera Information - Kodak 195
Printer settings
Physical Interface: Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: Rev. J, June, 2005
Image memory: 256 MB RAM
Hard disk: 4 GB
Manufacturer / device type: Printer
Device name: KELP 2180
Manufacturer: Kodak
SW Revision: V 2.17H
Physical type: None
Network protocol: None
Manufacturer´s specifications: None
Image memory: 24 MB
Hard disk: None
Content: Configuration
MIM 200 setup
Under Configure, Network:
Computer Name=e.g., Kodak2180, IPAddr=accordingly
Subnet Mask=accordingly, Gateway Addr=accordingly
under Configure, Input, Network, Add:
Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
Under Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
Device type: DICOM Print Server (SCP)
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 195 of 260
Page 196
196 Specific Camera Information - Kodak
General Remarks:Log.Name=KELP2180, Database Name=2180General Remarks:Model=KELP2180, Memory 24, Node ID=6 or accordinglyMedia info.:Media=35x43cm, 4136x5160, 5030, (blue)Media=35x35cm, 4136x4136, 4006, (blue)Media=11"x14", 3240x4136, 4006, (Blue)Media=8"x10", 2290x2840, 2710, (Blue)Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation:No Mag=On, Bil=On, Repl=On, Cubic=On, Cubic 2=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=2Formatting:Pixel Pitch=12.6582Image processing:Gray Scale=On, Minimize=On, Max Image Dmax=3.2M952 usage:Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
Under Configure, SCP Services, select the specific modality, modify:
Content: Configuration
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 196 of 260
Page 197
Specific Camera Information - Kodak 197
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Image Resizing, Resizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=CS (Curve Shape), Always Apply=YCurve Shape/TFT, Curve Shape=0.0, Always Apply=YContrast (wet)=+0, Always Apply=NInterpolation=No Mag., Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=2.80, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffFilm Output = 9(no. of sorter), Always Apply=NBasic color = N
2180 setup:
Optical Interface Board will automatically configure to 2180, no setup needed
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
The switch “Access magazines” is acti-vated
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Send film size which is not inserted in camera
CHECK PRINTER CHECK PRINTER Check camera
Printing is stopped PRINTER OFFLINE PRINTER OFFLINE Camera offline
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Content: Configuration
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 197 of 260
Page 198
198 Specific Camera Information - Kodak
Real Size Scaling
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 310.0 310.0 0.00
Y 390.0 389.5 0.13
14" x 17"
Landscape
X 390.0 389.5 0.13
Y 320.0 320.0 0.00
14" x 14"
Portrait
X 310.0 310.0 0.00
Y 320.0 319.5 0.16
14" x 14"
Landscape
X 310.0 309.5 0.16
Y 320.0 320.0 0.00
11" x 14"
Portrait
X 240.0 240.0 0.00
Y 320.0 319.5 0.16
11" x 14"
Landscape
X 310.0 309.5 0.16
Y 240.0 240.0 0.00
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X 170.0 170.0 0.00
Y 210.0 209.5 0.24
8" x 10"
Landscape
X 210.0 209.5 0.24
Y 170.0 1700 0.00
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 198 of 260
Page 199
Specific Camera Information - Konica Minolta 199
6- 6Specific Camera Information - Konica MinoltaKonica Minolta DryPro 832 0
General Information 0
Block Diagram
Fig. 26:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 9
Format Release status Film type Medium type
14" x 17"Portrait Yes DRYPRO SD-Q Blue-based
Landscape Yes DRYPRO SD-Q Blue-based
14" x 14"Portrait Yes DRYPRO SD-Q Blue-based
Landscape Yes DRYPRO SD-Q Blue-based
11" x 14"Portrait Yes DRYPRO SD-Q Blue-based
Landscape Yes DRYPRO SD-Q Blue-based
10" x 12"Portrait Yes DRYPRO SD-Q Blue-based
Landscape Yes DRYPRO SD-Q Blue-based
8" x 10"Portrait Yes DRYPRO SD-Q Blue-based
Landscape Yes DRYPRO SD-Q Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 199 of 260
Page 200
200 Specific Camera Information - Konica Minolta
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Magnification "NONE" not supported therefore KonicaServer needs a fixed setting:
Magnification "CUBIC" with smoothing type 7.
NOTE If “Camera cover open”, no message is displayed. If the cam-era is ready, jobs will be queued but not printed.
NOTE If a film is not inserted, but configured in camera server, thequeue is stopped with a camera software problem. This jobmust be deleted from the camera queue.
Known effects
NOTE The DICOM Print Server PRINTLINK 5 is not 510k registered.
NOTE Two film sizes can be installed at Drypro 832. To install otherfilm sizes the Konica-Minolta Service Technician is needed.
SCU / modality software configuration 0
Time 62 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DPRO832 Yes
Port number 104 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 200 of 260
Page 201
Specific Camera Information - Konica Minolta 201
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec. file name)konicaDryPro832_largeB No
DICOM Node Konica832 Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.6 x 78.6 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
4344 x 5196
14” x 14”
4344 x 4230
11" x 14"
3390 x 4230
10” x 12”
3046 x 3580
8" x 10"
2398 x 2932
In “Define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
5260 x 4268
14” x 14”
4294 x 4268
11" x 14"
4306 x 3327
10” x 12”
3644 x 2970
8" x 10"
2996 x 2322
Medium type Blue film No
Film destination Processor Yes
Color appearance Gray scale 8 bit No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 201 of 260
Page 202
202 Specific Camera Information - Konica Minolta
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Printer Settings
Web-Servicetool for Printlink 5
Configuration Parameter:
select input channel in "Input Parameter Setup Screen" window:
Background Black No
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM Printer (SCP)
Device name: DRYPRO 832
Manufacturer: Konica Minolta
SW Revision: V1.02R00 (for Printer & DICOM Server)
Physical type: Ethernet Card 100 / 1000
Network protocol: TCP/IP
DICOM conformance statement: Konica-Minolta, Printlink5-IN, Ver. 1.0 2006.10
Image memory: Printer: 64 MB; DS: 1 GB-RAM
Hard disk: Printer: None; DS: 80 GB
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 202 of 260
Page 203
Specific Camera Information - Konica Minolta 203
Configuration LUT:
select input channel in "LUT Setup Screen" window:
Content Printer Setting Value
Input Channel CT_hostname
Number of Copies 1
Print priority LOW
Film destination BIN_1
Film orientation PORTRAIT
Magnification Type CUBIC (checkbox selected)
Smooth Type 7 (checkbox selected)
Border density BLACK
Min. Dens. 0.20
Max. Dens.: 3.00
Trim: NO
Config. Information KC_LUT=1 (checkbox selected)
Density 0
Contrast 0
Flip NORMAL
Polarity NORMAL
Medium type BLUE FILM
Film Size ID 14" X 17"
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 203 of 260
Page 204
204 Specific Camera Information - Konica Minolta
Content Printer Setting Value
Input Channel CT_hostname
LUT No 1
Display Mode 16Step - Density
Drag Mode One
Illumination 2000
Refl. Ambient 10
Basic Pattern (linear LUT):
Previous Setting 1=20 / 2=39 / 3=57 / 4=76 / 5=95 / 6=113 / 7=132 / 8=151 / 9=169 / 10=188 / 11=207 / 12=225 / 13=244 / 14=263 / 15=281 / 16=300)
Test Image Convert P-Value Convert
Polarity NORMAL
Configuration Maintenance
Host name LENOVO-0B9651EF
IP address xxx.xxx.xxx.xx
AE Title SCP PLNK5
AE Title SCU PLNK5
Port No. 104, 6000, 5200
select input channel in "Printlink 5 Maintenance" window:
Input Channel 4
Input Name CT_hostname
SCU IP-Adr. xxx.xxx.xxx.xx
SCU AET title PR_hostname
Protocol Name DICOM PRINT IN
Maker KonicaMinolta (1.03)
Model RS
"double-click" Ch4, SOM64B
Channel Name CT_hostname
Default Channel (checkbox deselected)
Protocol Name DICOM PRINT IN
Default Format 1x1
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 204 of 260
Page 205
Specific Camera Information - Konica Minolta 205
IP Adr./H. Name xxx.xxx.xxx.xx
AE Title SCU PR_hostname
Common Input ... ("not relevant for printing")
DRYPRO 832 Output Setup:
Luminance Conversion ON
Border Size CT/MRI
Format Type EVEN
Pat.Name Stamp Type OFF
DICOM PRINT IN
N-Event Port No. 0
Empty Notice LV Standard (checkbox selected)
Error Level M/M Abort
Max PDU 64
Upper Stamp Use OFF
Printlink 5 Maintenance IMAGER:
"double click" Imager Ch1 DPRO832
Imager Output channel setup screen
Channel Name DPRO832
Protocol Name DICOM PRINT OUT
IP-Adr./Hostname: Localhost
DRYPRO832 IP-Adr. xxx.xxx.xxx.xx
AET-SCP KC_DRO4_P001
Port No 63101
Common Output
Connect Imager DRYPRO832
Output Pixels 12bit
Convert LUT ON
Stretch Base No Stretch
Stretch Mode BiLinear
Content Printer Setting Value
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 205 of 260
Page 206
206 Specific Camera Information - Konica Minolta
Error / warning messages (examples)
Warning Out ON
Layout DcmPrtLayOut832.ini
Output Name ASCII + KANJI + KANA
Output Image Color Grayscale
Study Output Order Normal
Annotation Box Not send
Calculate Sampling Pitch (checkbox deselected)
Imager Default Setting:
Default Film Size 1 14x17
Def. Media Type 1 BLUE FILM
Def. Film Type 2 11x14
Def. Media Type 2 BLUE FILM
Action that results in cam-era error mes-
sages
Printer messagesMessage displayed
on SCUDefined DCS (*1) Sent by SCP (*2)
Film Jam FILM JAM FILM JAM Film jam
Open cover and no film is printing
COVER OPEN Normal * "nothing is displayed, no film is printed, all films are stored in the camera. After closing the cover all stored films were printed out.
10"x12" film Sup-ply empty
EMPTY 10x12 BLUE EMPTY 10x12 BLUE 10x12 blue film supply empty
Switch off cam-era but Konica DICOM Server is running
PRINTER DOWN PRINTER DOWN Camera down
Content Printer Setting Value
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 206 of 260
Page 207
Specific Camera Information - Konica Minolta 207
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Send a film with a size which is not inserted in camera (14"x17", 8"x10" inserted, 10"x12" is sent)
not defined ELEC SWERROR ** Camera software problem - Queue stopped.
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
* In case of "Camera cover open", no message is displayed. If the camera is ready, jobs are queued only and not printed.
** If a film is not inserted and configured in the camera server, the queue is stopped with a camera software problem. This job must be deleted in the camera queue.
Action that results in cam-era error mes-
sages
Printer messagesMessage displayed
on SCUDefined DCS (*1) Sent by SCP (*2)
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 391.5 0.38
14" x 17"
Landscape
X 410.0 412.0 0.49
Y 320.0 320.5 0.16
14" x 14"
Portrait
X 300.0 301.0 0.33
Y 300.0 301.0 0.33
14" x 14"
Landscape
X 300.0 301.0 0.33
Y 300.0 301.0 0.33
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 320.0 321.0 0.31
11" x 14"
Landscape
X 330.0 331.0 0.30
Y 250.0 250.5 0.20
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 207 of 260
Page 208
208 Specific Camera Information - Konica Minolta
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 271.0 0.37
10” x 12”
Landscape
X 270.0 271.0 0.37
Y 220.0 220.5 0.23
8" x 10"
Portrait
X 170.0 170.0 0.00
Y 220.0 221.0 0.45
8" x 10"
Landscape
X 230.0 231.0 0.43
Y 170.0 170.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 208 of 260
Page 209
Specific Camera Information - Konica Minolta 209
Konica Minolta DryPro 793 6.1
General information 0
Block diagram
Fig. 27:
Available film sizes
Format Release status Film type Medium type
14" x 17"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
14" x 14"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
11" x 14"
Portrait YesDRYPRO SD-P
NewBlue-based
Landscape YesDRYPRO SD-P
NewBlue-based
10" x 12"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
8" x 10"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 209 of 260
Page 210
210 Specific Camera Information - Konica Minolta
Performance and Connections
Associations
Maximum number of associations supported: 32
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 105 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DPRO793 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec. file name)konicaDryPro793_largeB No
DICOM Node Konica793 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 210 of 260
Page 211
Specific Camera Information - Konica Minolta 211
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 87.5 x 87.5 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
3902 x 4668
14" x 14"
3902 x 3799
11" x 14"
3044 x 3799
10” x 12”
2745 x 3219
8" x 10"
2160 x 2643
In “Define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
4725 x 3834
14" x 14"
3856 x 3834
11" x 14"
3868 x 2988
10” x 12”
3275 x 2677
8" x 10"
2699 x 2092
Medium type Blue film No
Film destination Processor Yes
Color appearance Gray scale 8 bit No
Background Black No
Transformation No magnification No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 211 of 260
Page 212
212 Specific Camera Information - Konica Minolta
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
The configuration has to be performed via the printer keypad.
Konica Service has to install the modality name so that the printer settings can be assigned to this name.
Switch to Service mode
Select "Print" Mode
Device type: DICOM Printer (SCP)
Device name: DRYPRO 793
Manufacturer: Konica Minolta Medical & Graphics, Inc.
SW Revision: V1.00R05
Physical type: Ethernet 10 base-T / 100 base-TX / 1000 base-T
Network protocol: TCP/IP
DICOM Conformance Statement: Ver.1.02, 2005.02
Image memory: 512 MB
Hard disk: 80 GB
Setting: Print Setting for CT modality Value
LUT NO LUT1: LINEAR
DENSITY 0
CONTRAST 0
SMOOTHING 7
MAX DENSITY 300
MIN DENSITY 20
LUMINOUS INTENSITY 2000
AMBIENT LIGHT 10
ROTATION OFF
SORTER 1
ORIENTATION PORTRAIT
IMAGE POLARITY POSITIVE
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 212 of 260
Page 213
Specific Camera Information - Konica Minolta 213
BORDER BLACK
TRIM ON/OFF OFFS
TAM
P
TIME/DATE (Not selected)
COPY NO. (Not selected)
SCU ID/PAGE NO. (Not selected)
STAMP MESSAGE (Not selected)
PATIENT ID / NAME (Not selected)
PATIENT DATA ASCII (Not selected)
STAMP CONTENT (EMPTY)
Setting: Print Setting for CT modality Value
STAMP POSITION
EDGE ENHANCE OFF
BORDER SIZE CT/MRI
EXPANSION CHARACTER OFF
MANAGE INFO. PRINT ON
REQ.IMAGESIZE EFFECTIVE
REQ:BEHAVIOR DECIMATE
INTEGRAL MULTIPLE RATIO 2
P VALUE RECEPTION OFF
FORMAT EVEN
STANDARD DMIN/DMAX DICOM
Setting: Print Setting for CT modality Value
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 213 of 260
Page 214
214 Specific Camera Information - Konica Minolta
Select “Service 1” Mode
DIC
OM
Prio
rity
DEFAULT LUT (Not selected)
SMOOTH (Not selected)
DENSITY MAX (Not selected)
DENSITY MIN (Not selected)
POLARITY (Not selected)
ORIENTATION (Not selected)
TRIM (Not selected)
BORDER (Not selected)
SORTER (Not selected)
Setting: SCU SETUP for CT modality Value
CHANNEL USE ON
SCU NAME CT-hostname
IP ADDRESS xxx.xxx.xxx.xxx
HOST NAME CT-hostname
PORT NUMBER 104
AE TITLE PR_hostname
ERROR LEVEL STANDARD
SIZE TYPE CHECK ON
ENFORCE SPECIFIC FILM OFF
STANDARD MAGAZINE 14x17B
EMPTY NOTICE
SIZE TYPE 14x17B ON
SIZE TYPE 8x10B ON
SIZE TYPE 10x12B ON
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 214 of 260
Page 215
Specific Camera Information - Konica Minolta 215
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Action that results in cam-era error mes-
sages
Printer messagesMessage displayed
on SCUDefined DCS (*1) Sent by SCP (*2)
Switch to service mode
CHECK PRINTER CHECK PRINTER Check camera
14x17 Magazine empty
EMPTY 14x17 BLUE EMPTY 14x17 BLUE 14x17 blue film supply empty
Front door open CHECK PRINTER CHECK PRINTER Check camera
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.5 0.15
Y 390.0 390.0 0.00
14" x 17"
Landscape
X 400.0 400.0 0.00
Y 320.0 321.0 0.31
14" x 14"
Portrait
X 330.0 330.5 0.15
Y 320.0 320.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 320.0 320.5 0.16
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 320.0 320.5 0.16
11" x 14"
Landscape
X 330.0 330.5 0.15
Y 250.0 250.0 0.00
10” x 12”
Portrait
X 230.0 230.0 0.00
Y 270.0 270.5 0.19
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 215 of 260
Page 216
216 Specific Camera Information - Konica Minolta
10” x 12”
Landscape
X 270.0 270.5 0.19
Y 220.0 220.0 0.00
8" x 10"
Portrait
X 170.0 170.0 0.00
Y 220.0 220.5 0.23
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 180.0 180.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 216 of 260
Page 217
Specific Camera Information - Konica Minolta 217
Konica Minolta DryPro 771 6.2
General information 0
Block diagram
Fig. 28:
Available film sizes
Format Release status Film Type Medium type
14" x 17"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
14" x 14"Portrait Yes DRYPRO SD-P Blue-based
Landscape No No No
11" x 14"
Portrait YesDRYPRO SD-P
NewBlue-based
Landscape YesDRYPRO SD-P
NewBlue-based
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 217 of 260
Page 218
218 Specific Camera Information - Konica Minolta
Performance and Connections
Associations
Maximum number of associations supported: 16
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE 14INX14IN landscape not supported by camera.
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 112 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DP771 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)konicaDyPro771_largeB No
DICOM Node KonicaDP771 Yes
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 218 of 260
Page 219
Specific Camera Information - Konica Minolta 219
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 78.6 x 78.6 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4350 x 5202
14" x 14"
4350 x 4236
11" x 14"
3384 x 4236
In “define new”
Do not modify/add
Select only
Delete onlyFilm Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5318 x 4236
11" x 14"
4310 x 3282
Medium type Blue film No
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black No
Transformation No magnification No
Device type: DICOM Printer (SCP)
Device name: DRYPRO MODEL 771
Manufacturer: KonicaMinolta
SW Revision: V1.00R03
Physical type: RJ45
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 219 of 260
Page 220
220 Specific Camera Information - Konica Minolta
Printer settings
The configuration has to be performed via the printer keypad.
Konica service has to install the modality name and assign it to a channel number. For this camera validation, the modality name "specific Modality name" is installed and assigned to channel: "CH3". Use Maintenance Mode to view/modify Print Server Configuration.
Select U30 Print Cond.
Select CH3 "specific modality name"
Enter
Network protocol: DICOM, TCP/IP
DICOM Conformance Statement: Vers. 1.00, Sep.01. 2003
Image memory: 128 MB / Standard
Hard disk: 80 GB / Standard
Device type: DICOM Printer (SCP)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 220 of 260
Page 221
Specific Camera Information - Konica Minolta 221
Exit
Select U90 - Password
Enter
INPUT PASSWORD
Enter
Select U80 Service Mode
S00 DICOM SCU displayed
Select CH3 "specific modality name"
Enter
Variable name Content
LUT NO 1 LINEAR
DENSITY 0
CONTRAST 0
SMOOTHING 2
MAX DENSITY 300
MIN DENSITY 20
ORIENTATION PORTRAIT
IMAGE POLARITY POSI
TRIM ON/OFF Off
BORDER Black
ILLUMINATION 2000
AMBIENT LIGHT 10
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 221 of 260
Page 222
222 Specific Camera Information - Konica Minolta
Exit
Select S30 DICOM PRIORITY
Select CH3 "specific modality name"
Enter
Exit
Select S20 Print Cond.
Select CH3 "specific modality name"
Enter
S21 STAMP SELECT
Variable name Content
CHANNEL USE ON
SCU NAME CT-hostname
N-EVENT PORT NO BORD=0
IP ADDRESS xxx.xxx.xxx.xxx
AE TITLE PR_hostname
HOST NAME CT-hostname
BORDER SIZE REGIUS
FILM SIZE TYPE NO CHECK
EMPTY NOTICE ON
DICOM ERROR LEVEL STANDARD
FRAME LAYOUT EVEN
Dmax/Dmin DICOM
P-VALUE RCV OFF
Variable name Content
DEFAULT LUT DICOM
SMOOTH DICOM
DMAX DICOM
DMIN DICOM
POLARITY DICOM
ORIENTATION DICOM
TRIM DICOM
BORDER DICOM
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 222 of 260
Page 223
Specific Camera Information - Konica Minolta 223
Exit
Data was changed
Save the data? Yes
Enter
Exit
Error / Warning Messages (examples)
Variable name Content
TIME/DATE OFF
Copy no. OFF
Patient ID OFF
SCU ID / Page OFF
STAMP MESSAGE OFF
STAMP MESSAGE
STAMP DIRECTION LOWER
YEAR/FORMAT YY
MONTH/FORMAT TEXT
DATE/FORMAT TIME+DATE
REQ.IMAGESIZE EFFECTIVE
EXPANSION CHAR OFF
EDGE ENHANCE OFF
REQ:BEHAVIOR DECIMATE
FLIP ON/OFF OFF
Action that results in camera error mes-
sages
Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP
14x17 Magazine empty EMPTY 14x17 BLUE
EMPTY 14x17 BLUE
14x17 blue film supply empty
Front door open COVER OPEN CHECK PRINTER
Check camera
Switch to service mode n.a. CHECK PRINTER
Check camera
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 223 of 260
Page 224
224 Specific Camera Information - Konica Minolta
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Action that results in camera error mes-
sages
Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.5 0.15
Y 400.0 401.0 0.25
14" x 17"
Landscape
X 410.0 411.0 0.24
Y 320.0 320.5 0.16
14" x 14"
Portrait
X 330.0 330.0 0.00
Y 320.0 321.0 0.31
14" x 14"
Landscape
X No No No
Y No No No
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 320.0 320.5 0.16
11" x 14"
Landscape
X 330.0 330.5 0.15
Y 250.0 250.5 0.20
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 224 of 260
Page 225
Specific Camera Information - Konica Minolta 225
Konica DryPro 751/752 6.3
General information 0
Block diagram
Fig. 29:
Available film sizes
Format Release status Film Type Medium type
14" x 17"
Portrait YesDRYPRO SD-P
NewBlue-based
Landscape YesDRYPRO SD-P
NewBlue-based
14" x 14"Portrait Yes DRYPRO SD-P Blue-based
Landscape Yes DRYPRO SD-P Blue-based
11" x 14"
Portrait YesDRYPRO SD-P
NewBlue-based
Landscape YesDRYPRO SD-P
NewBlue-based
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 225 of 260
Page 226
226 Specific Camera Information - Konica Minolta
Performance and Connections
Associations
Maximum number of associations supported: 16
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
Known Effects
n.a.
SCU / Modality Software Configuration 0
Time 98 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title DPRO752 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)konicaDryPro752 No
DICOM Node DPRO752 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 226 of 260
Page 227
Specific Camera Information - Konica Minolta 227
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Pixel Size [1/1000 mm] 80.0 x 80.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4272 x 5112
14" x 14"
4272 x 4162
11" x 14"
3336 x 4162
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5224 x 4162
14" x 14"
4272 x 4162
11" x 14"
4274 x 3224
Medium type Blue film No
Film destination Processor Yes
Color appearance Grayscale 8 bit No
Background Black No
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM Printer (SCP)
Device name: DRYPRO 752
Manufacturer: Konica
SW Revision: V1.00R7
Physical type: Ethernet RJ45 10/100 Base TX
Network protocol: TCP/IP
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 227 of 260
Page 228
228 Specific Camera Information - Konica Minolta
Printer settings
DICOM Conformance Statement: V 1.03, Jun.18.2003
Image memory: 256 MB RAM
Hard disk: 20 GB
File Name Parameters Content
Maintenance 1/3
Print Condition CT host name
LUT LINEAR
DENSITY 0
CONTRAST 0
SMOOTH TYPE 7
MAX DENSITY 300
MIN DENSITY 020
L0 0
La 0
SORTER BIN 01
ORIENTATION Portrait
POLARITY Posi
TRIM Off
BORDER Black
FLIP Off
STAMP SELECT DATE TIME No
Copy no. No
Patient ID No
SCU ID/page no. No
Device type: DICOM Printer (SCP)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 228 of 260
Page 229
Specific Camera Information - Konica Minolta 229
STAMP MESSAGE No
STAMP DIRECTION Stamp position in graphic: top
Maintenance 2/3
SERVICE n.a.
DICOM SCU CT hostname
SCU SETUP n.a.
SCU CH USE ON
SCU NAME CT hostname
N-EVENT PORT NO 0
IP ADDRESS xxx.xxx.xxx.xxx
AE TITLE PR_hostname
HOST NAME CT hostname
SIZE TYPE CHECK ON
BORDER SIZE REGIUS
EMPTY NOTICE n.a.
14x17 B ON
14x14 B ON
EDGE ENHANCE ON
FRAME LAYOUT CENTER
DICOM ERROR LEVEL STANDARD
DEFAULT Dmax/Dmin DICOM
Magnification 02
File Name Parameters Content
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 229 of 260
Page 230
230 Specific Camera Information - Konica Minolta
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Action that results in camera error mes-
sages
Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP
Camera switched off n/a TCP init error, tim-eout
Connection to camera is not available
14x17 Magazine empty EMPTY 14x17 BLUE
EMPTY 14x17 BLUE
14x17 blue film supply empty
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 331.0 0.30
Y 380.0 381.0 0.26
14" x 17"
Landscape
X 400.0 401.0 0.25
Y 320.0 321.0 0.31
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 320.0 321.0 0.31
14" x 14"
Landscape
X 330.0 331.0 0.30
Y 320.0 321.0 0.31
11" x 14"
Portrait
X 250.0 251.0 0.40
Y 320.0 321.0 0.31
11" x 14"
Landscape
X 330.0 331.0 0.30
Y 240.0 241.0 0.42
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 230 of 260
Page 231
Specific Camera Information - Konica Minolta 231
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 231 of 260
Page 232
232 Specific Camera Information Sony
7- 7Specific Camera Information SonySony UP-DF 550 0
General information 0
Block diagram
Fig. 30:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 8
Format Release status Film Type Medium type
14" x 17"Portrait Yes UPT-517BL Blue-based
Landscape Yes UPT-517BL Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait Yes UPT-514BL Blue-based
Landscape Yes UPT-514BL Blue-based
10" x 12"Portrait Yes UPT-512BL Blue-based
Landscape Yes UPT-512BL Blue-based
8" x 10"Portrait Yes UPT-510BL Blue-based
Landscape Yes UPT-510BL Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 232 of 260
Page 233
Specific Camera Information Sony 233
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Horizontal line structure visible with all film sizes.
NOTE Background of film is always white.
Known Effects
NOTE The UPT-517BL film type requires a specific densitometercharacteristic. Density measurements using the X-Rite 301densitometer are only possible if special optics 301-34SO areinstalled on the X-Rite 301 and calibrated as described in theX-Rite letter.
NOTE If a film size is sent to the camera and the required film sizeis not inserted in the camera, then nothing is displayed, nei-ther in Somaris nor on the camera display. Jobs stay in thequeue until the required magazine is inserted. If an attempt ismade to switch off the camera before all jobs are printed, theunprinted jobs are displayed - and the camera is not switchedoff.
SCU / Modality Software Configuration 0
Time 112 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title UPDF550 Yes
Port number 104 No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 233 of 260
Page 234
234 Specific Camera Information Sony
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)SonyUPDF550 No
DICOM Node Sony550 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.4 x 79.4 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4360 x 5232
11” x 14”
3426 x 4286
10” x 12”
3100 x 3630
8” x 10”
2452 x 3004
In “define new”
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5232 x 4360
11” x 14”
4286 x 3426
10” x 12”
3630 x 3100
8” x 10”
3004 x 2452
Medium type Blue film No
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 234 of 260
Page 235
Specific Camera Information Sony 235
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Manufacturer / device type: DICOM Printer (SCP)
Device name: Film Station UP-DF 550
Manufacturer: SONY
SW Revision: (for VA20) (for VA26)
MA-FW: 1.05 TST7;
DICOM: V3.21;
DR-FW: V1.01;
CT-FW: V1.00,
ICTAC: V1.00;
IDMAC: V25.0
MA-FW: 1.07 TST1;
DICOM: V3.21;
DR-FW: V1.01;
CT-FW: V1.00,
ICTAC: V1.00;
IDMAC: V25.0
Physical Interface: Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
DICOM Conformance Statement: Sony UP-DF550, Rev. 1.1, Feb.19, 2007
Image memory: 3.5 GB
Hard disk: No
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 235 of 260
Page 236
236 Specific Camera Information Sony
Printer settings
File Name Content Remarks
No file;
Keypad settings, press "Menu”
Information>
IP ADDRESS xxx.xxx.xxx.xxx
TOTAL PRINTS
Network address
No file;
Keypad settings, press "Menu"
PRINT Quality>
SELECT GAMMA
/SL32_14C/ [EXEC]
SHARPNESS OFF
PATTERN PRINT
1: 14x17 [EXEC]
SELECT POINT
95%.....[25%]
SELECT POINT
5%.....[75%]
SAVE GAMMA [EXEC]
Gamma table 8 (Linear LUT)
No file;
Keypad settings, press "Menu"
CALIBRATION>
FULL RANGE CAL
1: 14x17 [EXEC]
AUTO CALIBRATION
[On/Off]
CAL RESET
14x17 [EXEC]
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 236 of 260
Page 237
Specific Camera Information Sony 237
Error / Warning Messages (examples)
No file;
Keypad settings, press "Menu"
GAMMA SEL>
DEFAULT14x17
/SL32_14C/
DEFAULT11x14
/SL32_14C/´
DEFAULT10x12
/SL32_14C/
DEFAULT 8x10
/SL32_14C/
Linear LUT
Linear LUT
Linear LUT
Linear LUT
No file;
Keypad settings, press "Menu"
SYSTEM SETTINGS>
NETWORK SETTINGS>
IP ADDRESS
xxx.xxx.xxx.xxx
SUBNETMASK
xxx.xxx.xxx.xxx
DEFAULT Gateway
xxx.xxx.xxx.xxx
STRICT SIZE .. ON
SAVE SETTINGS >EXEC
Network address
Subnet mask
Default Gateway
ON = job stored if film size is not available
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Supply magazine not inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
Supply magazine empty
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
14x17 blue film sup-ply empty
Film jam FILM JAM FILM JAM Film jam
File Name Content Remarks
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 237 of 260
Page 238
238 Specific Camera Information Sony
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Magazines with 14x17 and 8x10 films inserted, a job with film size 11x14 is sent
Nothing NORMAL Nothing is displayed, neither in Somaris nor on the camera display. Jobs stay in queue until the 11x14 magazine is inserted. If an attempt is made to switch off the cam-era before all jobs are printed, the unprinted jobs are displayed - and the camera is not switched off.
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 320.0 320.0 0.00
Y 400.0 399.5 0.12
14" x 17"
Landscape
X 400.0 399.5 0.12
Y 320.0 320.0 0.00
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X 260.0 260.0 0.00
Y 320.0 319.2 0.25
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 238 of 260
Page 239
Specific Camera Information Sony 239
11" x 14"
Landscape
X 320 319.0 0.31
Y 260.0 260.0 0.00
10” x 12”
Portrait
X 240.0 239.8 0.08
Y 260.0 259.5 0.19
10” x 12”
Landscape
X 280.0 279.5 0.18
Y 220.0 220.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 220.0 218.8 0.55
8" x 10"
Landscape
X 220.0 218.8 0.55
Y 180.0 180.0 0.00
Film size / Direction Desired (mm) Measured (mm)
Deviation obtained from desired value
in %
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 239 of 260
Page 240
240 Specific Camera Information Sony
Sony UP-DF 500 7.1
General Information 0
Block Diagram
Fig. 31:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 8
Format Release status Film type Medium type
14" x 17"Portrait Yes UPT-517BL Blue-based
Landscape Yes UPT-517BL Blue-based
14" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
11" x 14"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
10" x 12"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
8" x 10"Portrait n.a. n.a. n.a.
Landscape n.a. n.a. n.a.
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 240 of 260
Page 241
Specific Camera Information Sony 241
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Line structure is visible. 3% modulation is visible with a lenseonly.
NOTE The UPT-517BL film type requires a specific densitometercharacteristic. Density measurements using the X-Rite 301densitometer are only possible if special optics 301-34SO areinstalled on the X-Rite 301 and calibrated as described in theX-Rite letter.
Known effects
n.a.
SCU / modality software configuration 0
Time 133 sec.
Application Entity
PropertiesDisplay Changes allowed
AE Title UPDF500 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
Class
(Printer spec file name)SonyUPDF500 No
DICOM Node Sony UP_DF500 Yes
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 241 of 260
Page 242
242 Specific Camera Information Sony
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.4 x 79.4 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
4224 x 5060
In “define new”
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
5060 x 4224
Medium type Blue film No
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
Manufacturer / device type:
DICOM Printer (SCP)
Device name: Film Station UP-DF 500
Manufacturer: SONY
SW Revision: (for VA20) (for VA26)
MA: 3.04
DR: 3.03
CT: 1.03
IDMAC: 25.0
DICOM: 01.67
MA: 4.00
DR: 4.02
CT: 2.00
IDMAC: 25.0
ICTAG: 1.00
DICOM: 3.21
Physical interface: Ethernet (RJ45)
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 242 of 260
Page 243
Specific Camera Information Sony 243
Printer Setting
Error / warning messages (examples)
Network protocol: TCP/IP
DICOM conformance state-ment:
Rev. 2.0, May 21, 2003.
Image memory: 12 GB
Hard disk: 20 GB
File Name Content Remarks
No file;
Keypad settings, press "Menu"
Information>
IP ADDRESS xxx.xxx.xxx.xxx Network address
No file;
Keypad settings, press "Menu"
PRINT Quality>
SHARPNESS OFF
GAMMA SEL
Default GAMMA
SL32_14C
Interpolation
Linear LUT
No file;
Keypad settings, press "Menu"
Density ADJ>
SELECT POINT
95% ...[25]
95%...[9]
5 % ...[75]
(for MA3.04)
(for MA4.00)
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Cover or door of camera open
COVER OPEN COVER OPEN Camera cover, drawer or door open
Supply magazine not inserted
NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able
Supply magazine empty
EMPTY 14X17 BLUE
EMPTY 14X17 BLUE
14x17 blue film sup-ply empty
Film jam FILM JAM FILM JAM Film jam
Manufacturer / device type:
DICOM Printer (SCP)
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 243 of 260
Page 244
244 Specific Camera Information Sony
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning or error messages
Printer message Message displayed on SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 380.0 378.5 0.39
14" x 17"
Landscape
X 390.0 388.5 0.38
Y 320.0 320.0 0.00
14" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
14" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
11" x 14"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
10” x 12”
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Portrait
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
8" x 10"
Landscape
X n.a. n.a. n.a.
Y n.a. n.a. n.a.
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 244 of 260
Page 245
Specific Printer Information Epson 245
8- 8Specific Printer Information EpsonEpson AcuLaser C 4200 0
General information 0
Block diagram
Fig. 32:
Available film sizes
Performance
Use medical color image A4, positive image, format 3x4, copies=1, default center/width. The modality has to be configured to CTLUT and the respective PS LUT file for this printer.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
Format Release status Paper Type Medium type
A4 YesBusiness Laser Copier Paper
80g/m2 Paper
Letter YesBusiness Laser Copier Paper
80g/m2 Paper
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
Time 58 sec
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 245 of 260
Page 246
246 Specific Printer Information Epson
NOTE Paper print-outs are not allowed for medical diagnoses.
NOTE 510K premarket notification/approval not available.
NOTE The Adobe printer driver with the printer-specific descriptionfile has to be used.
The paper printer is not linear in the grayscale imageprint-out. As a result, the special LUT"EPSON_C4200_correction.lut" in the configuration of thePSLUT files has to be used to optimize the image quality ofthe print-out.
NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g. calcium scoring), usethe report of the applications to print out the result images.
Known Effects
NOTE Line structure visible on grey field.
NOTE Saturated colors do not have any contrast (approx 3 steps ofa 17-step test image). Smooth transitions from black to a sat-urated color are not continuous. Effective resolution on theXerox Business Paper is only approx. 200 DPI. Displacementof the primary colors visible (light vertical shift in portrait ori-entation)
Printer Configuration 0
Only the configuration has been performed with the generic “winsteng.exe” and the post-script description file “epalc420ppd” version.
NOTE Version of Adobe Driver: V 5.2
Modality Configuration (Postscript Driver) 0
The default setting of the postscript driver has to be optimized at the modality site.
These settings are:
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 246 of 260
Page 247
Specific Printer Information Epson 247
Printer --> Properties --> Device Settings:
AdobePS EPSON AL-C4200 PS3 Device Settings
--> Form To Tray Assignment.
--> MP Tray: Letter
--> Cassette 1: A4
--> Cassette 2: Not Available
--> Cassette 3: Not Available
--> Cassette 4: Not Available
--> Manual Feed 1st Page: Letter
--> Manual Feed Each Page: Letter
--> Font Substitution Table
--> Arial: Helvetica
--> Courier New: Courier
--> Symbol: Symbol
--> Times New Roman: Times
--> xxxx: <Don’t Substitute>
--> xxxx: For all other fonts of the substitu-tion table
--> Available PostScript Memory: 249 KB
--> Output Protocol: ASCII
--> Send CTRL-D Before Each Job: No
--> Send CTRL-D After Each Job: Yes
--> Convert Gray Text to PostScript Gray: No
--> Convert Gray Graphics to PostScript Gray: No
--> Add Euro Currency Symbol To PostScript Fonts: No
--> Job Timeout: 30 seconds
--> Wait Timeout: 40 seconds
--> Minimum Font Size to Downloads as Outline: 100 pixel(s)
--> Maximum Font Size to Download as Bitmap: 600 pixel(s)
--> Installable Options
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 247 of 260
Page 248
248 Specific Printer Information Epson
--> Installed Memory: 128 MB
--> Lower Cassette Unit: None
Printer --> Printing Preferences --> Advanced: (*1) and
Printer --> Properties --> Advanced --> Printing Defaults --> Advanced:(*2)
AdobePS EPSON AL-C4200 PS3 Advanced Document Settings
--> Paper/Output --> Paper Size: A4
--> Copy Count: 1 Copy
--> Graphic
--> Print Quality: Quality
--> Image Color Management
--> --> ICM Method: ICM Disabled
--> --> ICM Intent: Pictures
--> Scaling: 100%
--> True Type Font: Substitute with Device Font
--> Document Options
--> Advanced Printing Features: Enabled
--> Postscript Options
--> --> PostScript Output Option: Optimize for Speed
--> --> True Type Font Download Option: Automatic
--> --> PostScript Language Level: 3
--> --> Send PostScript Error Handler: Yes
--> --> Mirrored Output: No
--> Printer Features
Color Model RGB
Color Mode Color
RGB Color Mode Vivid
Color Screen Increase Gradation
Pure Black Mode Text
Auto Mono Mode Off
RITech On
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 248 of 260
Page 249
Specific Printer Information Epson 249
Hardcopy Printer Identification (provided by OEM) 0
Identification
Printer Keypad Settings
The default setting of the printer should be used.
Error / Warning Messages (examples)
Toner Save Mode Off
Image Protect Off
Start Page Front
Rotate by 180° Off
(*1/2) Both settings must be identical
Device type: Color Laser Printer
Device name: AcuLaser C4200
Manufacturer: Epson
SW Revision or Firmware Version: OS: 22710 22618 Fonts: 00005208 LAN: 02.20
Number of Input Interface: 3 (Parallel, USB and Network)
Physical Interface: Network: 10/100 Base Tx
USB: Hi-Speed USB 2.0
Parallel: IEEE 1284
Network protocol: TCP/IP 10/100
Image memory: 128 MB
Hard disk: None
Connector for Network: RJ45
Documentation: - User Guide (SAP 14403347 ASD 01S 00)- Admin-istrator Guide (SAP 14403347 ASD 02S 00)
Action to provide error messages of printer Message displayed on Modality
Printer switched off Offline
No paper inserted in magazine,
some print jobs sent
Out of Paper
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 249 of 260
Page 250
250 Specific Printer Information Epson
Image Quality and Optical Density 0
Optical Density
NOTE The PSLUT file has to be configured:
“ EPSON_C4200_correction”
At this time, no reference values have been defined for paper printers.
Densitometer model/type X-Rite 528 was used.
For this measurement result, the Xerox Buisness Laser Copier Inkjet Paper was used.
It is also recommended to add some additional white paper below the print-out to improve results during the measurement.
Before testing the image quality, calibrate the densitometer. Afterwards check the white field of the densitometer in the "paper" mode (dot area). Note the values for V.
Top cover of printer is open,some print jobs sent Door open
Tray for paper not inserted Out of Paper
Action to provide error messages of printer Message displayed on Modality
Values of "Paper" & "Sample" mode V
Offset of the white field in the "Paper" mode (absolute value) 0.06
Value of the white field for density in the "Solid" mode ( relative value)
0.00
StepMeasured Rela-
tive Value
1 0.01
2 0.08
3 0.14
4 0.18
5 0.22
6 0.27
7 0.31
8 0.38
9 0.45
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 250 of 260
Page 251
Specific Printer Information Epson 251
NOTE The PS Printer has to be set to Black&White for this measure-ment
Real Size Scaling
10 0.51
11 0.60
12 0.70
13 0.87
14 1.07
15 1.28
16 1.44
17 1.57
Paper size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
A4
Portrait
X 160.0 159.5 0.31
Y 280.0 280.0 0.00
A4
Landscape
X 280.0 280.5 0.18
Y 160.0 160.0 0.00
Letter
Portrait
X 200.0 199.5 0.25
Y 240.0 239.5 0.21
Letter
Landscape
X 240.0 239.5 0.21
Y 160.0 159.5 0.31
Step Measured Rela-tive Value
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 251 of 260
Page 252
252 Specific Printer Information Hewlett Packard
9- 9Specific Printer Information Hewlett PackardHewlett Packard Color Laser Jet 4700N 0
General information 0
Block diagram
Fig. 33:
Available film sizes
Performance and Connections
Performance
Use medical color image A4, positive image, format 3x4, copies=1, default center/width. The modality has to be configured to CTLUT and the respective PS LUT file for this printer.
The optimal connection (point to point) was used.
Format Release status Paper Type Medium type
A4 YesBusiness Laser Copier Paper
80g/m2 Paper
Letter YesBusiness Laser Copier Paper
80g/m2 Paper
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
Time 23 sec
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 252 of 260
Page 253
Specific Printer Information Hewlett Packard 253
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE Paper print-outs are not allowed for medical diagnoses.
NOTE 510K premarket notification/approval not available.
NOTE For pure black&white images the configuration of the driverfor the printer should be set to black&white to optimize theimage quality on paper.
NOTE No resolution (contrast) in the white area between step 1 and2 of the grayscale image
NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g. calcium scoring), usethe report of the applications to print out the result images.
Known Effects
NOTE Resolution of 600DPI not visually reached on the printout.1-Pixel stripes are smoothed.
NOTE Color displacement of the primaries visible on the printout.
NOTE Line structure visible on flat fields.
NOTE Color gradient from white to a light color is not continuous.
NOTE No contrast at saturated colors.
NOTE The toner can be smeared off the paper, this lightens theimage and affects image quality.
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 253 of 260
Page 254
254 Specific Printer Information Hewlett Packard
Printer Configuration 0
Only the configuration has been performed with the generic “winsteng.exe” and the “hpmcpd25.ppd” postscript description file version.
NOTE Version of Adobe Driver: V 5.2
Modality Configuration (Postscript Driver) 0
The default setting of the postscript driver has to be optimized at the modality site.
These settings are:
Printer --> Properties --> Device Settings:
AdobePS HP Universal PrintingPS Device Settings
--> Form To Tray Assignment.
--> Printer Auto Select: A4
--> Manual Feed in Tray1: Letter
--> Tray 1: Letter
--> Tray 2: A4
--> Tray 3: Letter
--> Tray 4: Letter
--> Tray 5: Letter
--> Tray 6: Letter
--> Tray 7: Letter
--> Tray 8: Letter
--> Tray 9: Letter
--> Ex Tray (MP5): Letter
--> Ex Tray (MP6): Letter
--> Ex Tray (MP7): Letter
--> Ex Tray (MP8): Letter
--> Ex Tray (MP9): Letter
--> Ex Tray (MP10): Letter
--> Envelope Feeder: Letter
--> Font Substitution Table
--> Courier New: Courier
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 254 of 260
Page 255
Specific Printer Information Hewlett Packard 255
--> xxxx: <Don’t Substitute>
--> xxxx: For all other fonts of the substitu-tion table
--> Available PostScript Memory: 6144KB
--> Output Protocol: ASCII
--> Send CTRL-D Before Each Job: No
--> Send CTRL-D After Each Job: Yes
--> Convert Gray Text to PostScript Gray: No
--> Convert Gray Graphics to PostScript Gray: No
--> Add Euro Currency Symbol To PostScript Fonts: Yes
--> Job Timeout: 0 seconds
--> Wait Timeout: 300 seconds
--> Minimum Font Size to Downloads as Outline: 100 pixel(s)
--> Maximum Font Size to Download as Bitmap: 600 pixel(s)
--> Installable Options
--> Accessory Output Bin: Not Installed
--> Printer Memory: 384-494 MB
--> Printer Hard Disk: Installed
--> Job Storage: Enabled
--> Postscript Passthrough: Enabled
--> Copier Mode: Disabled
--> Automatic Configuration: Off
--> HPOutputBinPMLRange: 4-11
--> HPPrnPropResourceData: hpzhl3xr.cab
--> HPInstallableHCO: 3000Stacker-C8084
--> HPMOutputBinHCOMap: Auto
--> HPMOutputBinHCOPMLMap: 0-0
Printer --> Printing Preferences --> Advanced(*1): and
Printer --> Properties --> Advanced --> Printing Defaults --> Advanced(*2):
AdobePS HP Universal PrintingPS Advanced Document Settings
--> Paper/Output --> Paper Size: A4
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 255 of 260
Page 256
256 Specific Printer Information Hewlett Packard
--> Copy Count: 1 Copy
--> Graphic
--> Image Color Management
--> --> ICM Method: ICM Disabled
--> --> ICM Intent: Pictures
--> Scaling: 100%
--> True Type Font: Substitute with Device Font
--> Document Options
--> Advanced Printing Features: Enabled
--> Postscript Options
--> --> PostScript Output Option: Optimize for Speed
--> --> True Type Font Download Option: Automatic
--> --> PostScript Language Level: 2
--> --> Send PostScript Error Handler: Yes
--> --> Mirrored Output: No
--> Printer Features
--> --> Orientation Rotated 180 Degrees: Off
--> --> PostScriptCustomPageSize: Off
--> --> Output Bin: Automatically Select
--> --> Stapling Options: None
--> --> HPCovers: Other Pages
--> --> Print All Text as Black: Disabled
--> --> PrintQualityGroup: Best
--> --> JRHDInstalled: JRHDOff
--> --> HPDocPropResourceData: hpzhl3xr.cab
--> --> JRConstraints: JRCHDFull
--> --> HPSmartHub: _
--> --> Print Color as Gray: Off
(*1/2) Both settings must be identical
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 256 of 260
Page 257
Specific Printer Information Hewlett Packard 257
Hardcopy Printer Identification (provided by OEM) 0
Identification
Printer Keypad Settings
The default setting of the printer should be used.
Error / Warning Messages (examples)
Device type: Color Laser Printer
Device name: Color Laser Jet 4700N
Manufacturer: Hewlett-Packard
SW Revision or Firmware Version: 20051204 46.029.3
Number of Input Interface: 2 (Network and USB 2.0))
Physical Interface: Ethernet and USB-B and IEEE 1284-C-compatible bidirectional parallel interface
Network protocol: TCP/IP and Windows USB 2.0
Image memory: 384 MB
Hard disk: Yes
Connector for Network: RJ 45 / USB-B / Parallel port
Documentation: - “Technical Data” EMEA 08/05 5983-2739DEE
SAP: 14403335 GDB 01S 00
- "HP Color LaserJet 4700 Operator Manual“
SAP: 14403335 ASD 01S 00
- “Configuration page” (can be printed in the printer`s LCD menu)
SAP: 14403335 GDB 02S 00
Action to provide error messages of printer Message displayed on Modality
Printer switched off Offline
No paper inserted in magazine, some print jobs sent Out of paper
Top cover of printer is open, some print jobs sent Door Open
Tray for paper not inserted correctly Offline
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 257 of 260
Page 258
258 Specific Printer Information Hewlett Packard
Image Quality and Optical Density 0
Optical Density
NOTE The PSLUT file has to be configured:
“ HP_LaserJet4700_linear”
At this time, no reference values have been defined for paper printers.
Densitometer model/type X-Rite 528 was used.
For this measurement result, the Xerox Buisness Laser Copier Inkjet Paper was used.
It is also recommended to add some additional white paper below the print-out to improve the results during the measurement.
Before testing the image quality, calibrate the densitometer. Afterwards check the white field of the densitometer in the "paper" mode (dot area). Note the values for V.
Values of "Paper" & "Sample" mode V
Offset of the white field in the "Paper" mode (absolute value) 0.06
Value of the white field for density in the "Solid" mode ( relative value)
0.00
Step Measured Rela-tive Value
1 0.01
2 0.02
3 0.05
4 0.10
5 0.16
6 0.22
7 0.31
8 0.41
9 0.50
10 0.63
11 0.74
12 0.87
13 1.02
14 1.19
15 1.31
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO
Page 258 of 260
Page 259
Specific Printer Information Hewlett Packard 259
NOTE The PS Printer has to be set to Black&White for this measure-ment
Real Size Scaling
16 1.54
17 1.92
Paper size / Direction Desired (mm)Measured
(mm)
Deviation obtained from desired value
in %
A4
Portrait
X 160.0 160.0 0.00
Y 280.0 280.0 0.00
A4
Landscape
X 280.0 279.5 0.18
Y 160.0 159.5 0.31
Letter
Portrait
X 160.0 160.5 0.31
Y 240.0 240.5 0.21
Letter
Landscape
X 240.0 240.0 0.00
Y 160.0 160.0 0.00
Step Measured Rela-tive Value
© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO
Page 259 of 260
Page 260
260 Changes to Previous Version
10- 10Changes to Previous VersionChanges to Previous Version 0
Chapter Section Action
4 Fuji Drypix 4000 Identification updated
4 Fuji Drypix 7000/5000 Identification updated
5 Kodak DryView 6800 Identification updated
7 Sony UP-DF 550 Identification updated
Printer setting updated
7 Sony UP-DF 500 Restrictions updated
Identification updated
Printer setting updated
SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO