© Siemens, 2008 All documents may only be used for rendering services on Siemens Healthcare Products. Any document in electronic form may be printed once. Copy and distribution of electronic docu- ments and hardcopies is prohibited. Offenders will be liable for damages. All other rights are re- served. Print No.: S i e m e n s W a l z SOMATOM Definition Hardcopy Cameras Installation and Startup CT 2008 CT00-000.814.30.02.02 Replaces: CT00-000.814.30.01.02 English Doc. Gen. Date: 05.09 n.a. H WS CS SD CO Camera information with SOMARIS/7 VA20;VA26 © Siemens, 2008 This document is valid for the following CT systems: SOMATOM Definition SOMATOM Definition AS SOMATOM Definition AS+ SOMATOM Definition AS20 07740769 10430603
260

CT00-000.814.30.02.02.pdf

Jan 19, 2016

Download

Documents

w.h.n
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: CT00-000.814.30.02.02.pdf

SiemensWalz

SOMATOM Definition

Hardcopy Cameras

Installation and Startup

© Siemens, 20All documents may only bservices on Siemens Healdocument in electronic foonce. Copy and distributioments and hardcopies is will be liable for damages. served.

CT

08e used for renderingthcare Products. Anyrm may be printedn of electronic docu-prohibited. OffendersAll other rights are re-

2008

Print No.: CT00-000.814.30.02.02Replaces: CT00-000.814.30.01.02

ED

nglishoc. Gen. Date: 05.09

n.a.H WS CS SD CO

Camera information with SOMARIS/7 VA20;VA26

© Siemens, 2008This document is valid for the following CT systems:SOMATOM DefinitionSOMATOM Definition ASSOMATOM Definition AS+SOMATOM Definition AS20

0774076910430603

Page 2: CT00-000.814.30.02.02.pdf

2 Copyright / Version / Disclaimer

1Copyright / Version / Disclaimer

Copyright

“© Siemens, 2008“ refers to the copyright of a Siemens entity such as Siemens Aktienge-sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or Siemens Healthcare Diagnostics Inc. - USA.

Document Version

Siemens reserves the right to change its products and services at any time.

In addition, manuals are subject to change without notice. The hardcopy documents corre-spond to the version at the time of system delivery and/or printout. Versions to hardcopy documentation are not automatically distributed. Please contact your local Siemens office to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer

Siemens provides this documentation “as is“ without the assumption of any liability under any theory of law.

The installation and service of equipment described herein requires superior understand-ing of our equipment and may only be performed by qualified personnel who are specially trained for such installation and/or service.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 2 of 260

Page 3: CT00-000.814.30.02.02.pdf

Table of Contents 3

0Table of Contents

1 _______ General Hardcopy Camera Information ______________________________ 7

HCC Information for SOMATOM Definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Scope of applicability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Image quality and optical density check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18General image quality information (LUT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Density verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Reference films . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Description of Correction LUT Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Additional and unnecessary LUT files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Postscript Hardcopy Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Restrictions and Known Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Postscript Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Look-Up Tables (LUT). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

2 _______ Specific Camera Information - Agfa________________________________ 30

Agfa Drystar AXYS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . . 33Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Agfa Drystar 5503. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 41Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Agfa Drystar 5500. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 49Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Agfa Drystar 5302. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 57Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Agfa Drystar 5300. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 63Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

Agfa Drystar 3000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 70Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 3 of 260

Page 4: CT00-000.814.30.02.02.pdf

4 Table of Contents

Agfa LR5200 / MG3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 76Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Agfa LR5200 / LR DICOM Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 82Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

3 _______ Specific Camera Information - Codonics____________________________ 87

Codonics Horizon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 90Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

4 _______ Specific Camera Information - Fuji_________________________________ 94

Fuji Drypix 7000 / 5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 97Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100

Fuji Drypix 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 105Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109

Fuji Drypix 3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 112Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

Fuji Drypix 2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 119Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123

Fuji FM DPL/FN-PS551 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 128Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132

5 _______ Specific Camera Information - Kodak _____________________________ 134

Kodak DryView 6800. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 4 of 260

Page 5: CT00-000.814.30.02.02.pdf

Table of Contents 5

SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 137Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140

Kodak DryView 5800 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 145Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149

Kodak DryView 8900 (small border) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 154Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156

Kodak DryView 8700-Plus / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 159Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163

Kodak DryView 8700-Dual / PACS Link MIM 9410. . . . . . . . . . . . . . . . . . . . . . . . . . . . 164General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 166Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168

Kodak DryView 8200 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 172Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

Kodak DryView 8100 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 179Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

Kodak DryView 8150 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 187Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

Kodak KELP 2180/PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 194Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197

6 _______ Specific Camera Information - Konica Minolta ______________________ 199

Konica Minolta DryPro 832. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 202

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 5 of 260

Page 6: CT00-000.814.30.02.02.pdf

6 Table of Contents

Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207

Konica Minolta DryPro 793 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 212Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215

Konica Minolta DryPro 771 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 219Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224

Konica DryPro 751/752 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 227Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230

7 _______ Specific Camera Information Sony________________________________ 232

Sony UP-DF 550 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 235Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238

Sony UP-DF 500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 242Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244

8 _______ Specific Printer Information Epson _______________________________ 245

Epson AcuLaser C 4200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245Printer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246Modality Configuration (Postscript Driver) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246Hardcopy Printer Identification (provided by OEM) . . . . . . . . . . . . . . . . . . . . . . . . . 249Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

9 _______ Specific Printer Information Hewlett Packard _______________________ 252

Hewlett Packard Color Laser Jet 4700N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252Printer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Modality Configuration (Postscript Driver) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Hardcopy Printer Identification (provided by OEM) . . . . . . . . . . . . . . . . . . . . . . . . . 257Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258

10 ______ Changes to Previous Version ____________________________________ 260

Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 6 of 260

Page 7: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 7

1- 1General Hardcopy Camera InformationHCC Information for SOMATOM Definition 0

General information 0

Acronyms and abbreviations

Block Diagram SOMATOM Definition

Fig. 1:

System references

• Refer to Siemens Medical Solutions Intranet, UPTIME Services

AE Title Application entity title

BU Business Unit (Modality)

DCS DICOM Conformance Statement

DICOM Digital Imaging and Communication in Medicine

HW/SW Hardware/Software

HCC Hardcopy Camera

LAN Local Area Network

LUT Lookup Table

OEM Original Equipment Manufacturer

SCP Service Class Provider

SCU Service Class User

TCP/IP Transfer Control Protocol/Internet Protocol

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 7 of 260

Page 8: CT00-000.814.30.02.02.pdf

8 General Hardcopy Camera Information

• Product-specific technical documentation

Hardware components

• Ethernet switch ATI-FS708 (part no.: 30 79 332)

• Cable, straight-through, S/UTP RJ45 (5m.; part no.: 30 74 515)

• Cable, straight-through, S/UTP RJ45 (30 m.; part no.: 30 74 523)

or

• Refer to Siemens Medical Solutions Intranet, UPTIME Services

• Product-specific technical documentation

Restrictions and known effects

NOTE Sending film sizes that are not available at the customer sitemay cause problems at the camera site. For this reason, it isrecommended that you delete these film sizes for portraitand landscape formats.

NOTE The Grayscale Standard Display Function (GSDF) is imple-mented for printing and imaging. No Presentation LUTs areavailable. For this reason, the DICOM Presentation LUT is notreleased.

NOTE The use of the presentation LUT has to be disabled by localservice for configuring LUT files.

NOTE The density values for the image quality were measured witha calibrated densitometer X-Rite 301. The result of the mea-surement from other densitometers used may differ from theabsolute density values as described in the tables.

NOTE Print the 17 graystep test image to check the modality andcamera adjustment according to the description. Verify theresults using the reference values for max. density 2.80 listedin the table. This is also part of the acceptance test.

NOTE For additional Postscript Hardcopy Restrictions and KnownEffects, refer to the Chapter on Postscript Hardcopy Informa-tion as well. (Restrictions and known effects / p. 8)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 8 of 260

Page 9: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 9

Errors and warning messages

Only certain errors and warnings have been tested for each hardcopy camera.

These errors and warnings are described in the corresponding camera information.

.

Scope of applicability 0

Modality/system

Hardcopy Cameras

This table includes all hardcopy cameras using DICOM basic print.

Device type Modality (SCU)

Product name SOMATOM Definition

Modality host name/console CT host name

ICS (syngo Acquisition Workplace)

IES (syngo CT Workplace; optional)

IRS (Image Reconstruction System)

Manufacturer Siemens Med CT

SW revision Somaris 7 VA20A (syngo-based VE32A)

Somaris 7 VA26 (syngo-based VE32B)

Conformance statement EDM 10094810-EKL-YK0-01

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 9 of 260

Page 10: CT00-000.814.30.02.02.pdf

10 General Hardcopy Camera Information

Manufac-turer

HCC nameSoftware validity

VA20 VA26 V.... V....

AGFA Drystar AXYS X X

AGFA Drystar 5503 X X

AGFA Drystar 5500 X X

AGFA Drystar 5302 X X

AGFA Drystar 5300 X X

AGFA Drystar 3000 X X

AGFA LR5200 / LR DICOM Controller X X

AGFA LR5200 / MG 3000 X X

CODONICS Horizon X X

FUJI Drypix 7000/5000 X X

FUJI Drypix 4000 X X

FUJI Drypix 3000 X X

FUJI Drypix 2000 X X

FUJI FM DPL / FN-PS 551 X X

KODAK DryView 6800 X X

KODAK DryView 5800 X X

KODAK DryView 8900 (small border) X X

KODAK DryView 8700-Plus /

PACS Link MIM 200X X

KODAK DryView 8700-Dual /

PACS Link 9410X X

KODAK DryView 8200 /

PACS Link MIM 200X

X

KODAK DryView 8150 X X

KODAK DryView 8100 /

PACS Link MIM 200X X

KODAK KELP 2180 /

PACS Link MIM 200X X

KONICA MINOLTA

DryPro 832X X

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 10 of 260

Page 11: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 11

X = hardcopy camera is released

N/A = hardcopy camera is not released

NOTE The use of non-released hardcopy cameras in conjunctionwith BU products is prohibited and does not meet therequirements specified.

Hardcopy Printer

This table includes all hardcopy cameras using Postscript.

For configuration of one of these printers - refer to the specific documents for these printers

X = hardcopy printer is released

N/A = hardcopy printer is not released

NOTE The use of a non-released hardcopy printer in conjunctionwith CT products is prohibited and does not meet the require-ments specified.

KONICA MINOLTA

DryPro 793X X

KONICA MINOLTA

DryPro 771X X

KONICA DryPro 751/752 X X

SONY UP-DF550 X X

SONY UP-DF500 X X

Manufac-turer HCC name

Software validity

VA20 V... V.... V....

EPSON AcuLaser C4200 X X

Hewlett Packard

Color Laserjet 4700NX X

Xerox Phaser 6360 DP n.a. n.a.

Manufac-turer HCC name

Software validity

VA20 VA26 V.... V....

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 11 of 260

Page 12: CT00-000.814.30.02.02.pdf

12 General Hardcopy Camera Information

DICOM associations

A DICOM association is a network connection between 2 DICOM applications. The asso-ciation is established and released by the SCU and may be aborted by the SCU or SCP. The maximum number of associations for each HCC device is defined in the DCS. The maximum number of associations depends on the configuration and may be extended by increasing the number of gateways connected to the printer device.

In a DICOM network environment, there are two ways to connect a modality (SCU) to an HCC device (SCP).

Case 1: The association will always be opened. Assuming that this behavior applies to all modalities (connected to the same HCC device), the number of modalities has to be equal to or less than the maximum number of supported associations.

Case 2: The association will be opened (and closed) on demand. Assuming that this behavior applies to all modalities (connected to the same HCC device), the number of modalities may be greater than the maximum number of supported associations. This num-ber will be shared by the modalities.

The association open-close behavior is unknown to other non-SIEMENS modalities. Under optimum circumstances, the number of modalities is equal to or less than the maximum number of supported associations.

The association is always opened by the SCU (case 1)

If the associations are always opened by the SCU for each configured HCC device, the association is reserved for this SCU only.

If the maximum number of associations = 1, and another modality never releases the asso-ciation, this HCC device is always reserved for this modality.

The association is opened and closed on demand (case 2)

The associations will be opened (and closed) on demand by the SCU for each configured HCC device via the product name.

Special cases (compare also to case 1):

If the number of modalities configured in the HCC device is greater than the maximum num-ber of supported associations, some modalities will not be able to print. The printing func-tion depends on the access time and release mechanism of each connected modality (e.g., the maximum number of supported associations = 1, the number of modalities = 2. During access by the first modality, access by the second modality is denied and connectivity fails, and vice versa).

If other modalities do not release the associations, the shareable number of associations decreases.

Modality 0

Hardware and software prerequisites

All hardware and software prerequisites are included in the <modality> product delivery volume.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 12 of 260

Page 13: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 13

Parameters

For software configuration, HCC data/parameters and possibly additional film matrix prop-erties (syngo: filming properties) are required for software configuration.

Modality data and configuration parameters

HCC Data and Configuration Parameters1

1 The values for these parameters are part of the specific HCC information

Example: Kodak DryView 8300

Properties Example Supplier

Port number 104 Default entry

AE title PR_”Hostname” Default entry

TCP/IP address xxx.x.x.xxx Network administrator

Application Entity

PropertiesDisplay Supplier

AE TitleK8300

Network administrator or HCC service

Port number 5040 HCC Service

HC Device /

General settings Display Supplier

Type DICOM printer Default entry

Class (printer spec file name)

kodak8300 Default entry

DICOM Node dv8300 Default entry

Filming properties Display Supplier

Hold printed Film Jobs 10 Default entry

Min. density 20 Default entry

Pixel Size [1/1000 mm] 85.0 x 85.0 Default entry

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

8" x 10"

2256 x 2676Default entry

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

8" x 10"

2736 x 2136Default entry

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 13 of 260

Page 14: CT00-000.814.30.02.02.pdf

14 General Hardcopy Camera Information

Film matrix properties

The settings for filming properties depend on the class selected (DICOM camera). A pull-down menu is provided for DICOM-class cameras. Select your DICOM camera type. A specific file for this type is loaded and the "filming properties" are shown. Changing the "filming properties" will affect the camera site and may not meet the requirements. Do not make any changes, unless otherwise instructed.

NOTE Filming properties are "pixel size" and "media sheet size";this information is required for image scaling and cannot bechanged.

Software configuration

• The following information is required:

- Host Name

- TCP/IP address

- Application entity title

- Port number of the DICOM remote host (DICOM laser camera)

The AE title, port number and camera type have to be provided by the local camera service.

syngo software configuration and laser camera

1. Select options in the menu line

2. Navigate to local service

The Siemens Service Software Authentication platform is displayed

3. Enter the password and click ”OK”

The Home Menu Platform of the syngo Service Software is displayed

4. Click ”Configuration”

The Local Service Configuration List of system options is displayed

5. Click DICOM ”General”

6. Click ”>”

The local Application Entity Title platform is displayed

Medium type Blue film Default entry

Film destination Processor Default entry

Color appearance Grayscale 8 bit Default entry

Background Black Default entry

Transformation No magnification Default entry

Filming properties Display Supplier

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 14 of 260

Page 15: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 15

7. Provide the Local Camera Service with the information in the field ”Print (SCU) AE Title PR_”Hostname”

8. Click ”Save”; ”>”; ”Finish”

The Configuration Setup platform is displayed

9. Click Local Host ”TCP/IP LAN”

The local Service Configuration TCP/IP LAN platform is displayed

10. Provide the Local Camera Service with the information in the ”IP address” field (scroll down menu for IP address but do not select it)

11. Click ”Finish”

The Configuration Setup platform is displayed

12. Select DICOM Print Devices

The Local Service Hardcopy Device platform is displayed

13. Enter the host name and TCP/IP address of the laser camera (provided by the camera manufacturer)

14. Click ”Test” to test IP communication (ping)

15. Click ”Save”

Host properties successfully saved

16. Click ”OK”

17. Click ”>”

The General Note Properties Dialog platform is displayed

18. Under logical name: Enter the logical name of the laser camera

19. Select the host and choose the previously defined host name in the hardcopy device platform

20. Under application entity properties: Enter the AE title in the edit AE title field (provided by the camera manufacturer)

21. Enter the Port Number (provided by the camera manufacturer)

22. Click ”Add”

23. Click ”Save”

DICOM node properties successfully saved

24. Click ”OK”

25. Select the logical name again and click ”Verification” to test the DICOM communication (mc3echo)

26. Click ">"

The camera-specific parameter platform is displayed.

27. HC device: Enter the camera name

28. General settings: /type: DICOM printer

29. /class: e.g., agfadrystar3000, or refer to specific HCC information

30. Filming properties:

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 15 of 260

Page 16: CT00-000.814.30.02.02.pdf

16 General Hardcopy Camera Information

- Hold printed film job, 10 = default; (max. = 10) (films can be printed again)

- Min. density 20 = default; (can be changed in increments of 1)

- Pixel size (changes not allowed)

- Film Sheet Formats Portrait (do not modify and/or add film sizes, select and delete film sizes that are not available at the customer site)

- Medium type, blue film = default; clear film; paper

- Film destination, processor = default; magazine

- Color appearance, monochrome = default; (select color for printers that support and use color media only, e.g., Drystar2000; Codonics)

- Background, black =default; white

- Transformation no magnification = default;

do not change this value; except for Kodak DV8900; replicate;

(do not change any other values under filming properties)

31. Click "Save"

Caution: Please wait for the message "device properties successfully saved"

32. Click ”OK”

33. Click ”Finish”

The Configuration Setup platform is displayed

34. Click ”>” to proceed and select the next marked DICOM ”HC Overview”

35. Click ”Finish”

The Configuration Setup platform is displayed

36. Click LUT files

The LUT platform is displayed.

NOTE If new LUT files for one or more modalities are available on aremovable medium, e.g., floppy or CD, these files can beselected on the source drive and copied to the dependentLUT directories on the local host. Unnecessary LUT files canbe removed from the LUT directories on the local host.

37. Click ">"

Usage of Presentation LUT platform is displayed

38. Select disable

39. Modifications are necessary for additional image impressions only.

NOTE Only the topic CT image and SC image are relevant. Depend-ing on the camera, specific entries have to be made. If noth-ing else is mentioned, set the CT image and SC image to"CTLUT" and "2.80" for maximum density. The entries for BUimage and SC image have to be identical.

40. Overall maximum density:

- 3.00 = default (depends on the max. possible value of the max. density of all image types)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 16 of 260

Page 17: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 17

41. Interpolation for printing:

- Bilinear mode = default; #Smoothing Type

- Replicate #Smoothing Type

- Cubic 0, 2, 3, 4 = concerns cubic mode only

42. When changes are made, click "Save"

43. Click "Finish"

44. Click "Home"; the system automatically restarts the application

NOTE This may take some time. Confirm warnings with "OK."

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 17 of 260

Page 18: CT00-000.814.30.02.02.pdf

18 General Hardcopy Camera Information

Image quality and optical density check 1.1

General image quality information (LUT) 0

Only standard LUT files are loaded. For importing/removing additional/unnecessary LUT files (see section ”Additional/unnecessary LUT files”) (LINK).

The image data from the modality to the HCC is transformed/corrected by the LUTs, once in the modality and once in the HCC. As as result, the LUT has to be set on both sides:

• Refer to the Siemens Medical Solutions Intranet, UPTIME Services

• Product-Specific Technical Documentation

Correction LUT is selected

Density verification 0

There are two ways to check the LUT via the film density:

The modality is configured to a specific correction LUT and the HCC is configured to linear curves. As an alternative, the modality is configured to a linear LUT and the HCC is config-ured to linear curves.

Modality Image Type Correction LUT Max. density

CT Image CT LUT 2.80

SC Image CT LUT 2.80

HCCRefer to the specific Camera Information "DICOM printer settings".

The HCC Service performs the setting "LINEAR LUT"

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 18 of 260

Page 19: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 19

Density verification for CT

Since these two checks are equivalent; it is sufficient to check the “specified density”.

Checking the “HCC linearity” may be necessary if the desired density verification (1) shows deviations and the HCC setting needs to be changed.

NOTE Even if the HCC service does not require a test image of themodality for the HCC setting, it is easier to check this setting- and possibly further HCC corrections - with the HCC linear-ity verification (2) than with the desired density verification(1). Only the HCC linearity LUT is effective. A graphical dis-play would now show the current HCC behavior only. Thiscould be helpful for HCC Service.

Density values for CT

• Measure the density values in 17 gray levels in the IEC test image (Fig. 2 / p. 19) (for each film size).

W:256; C:0;

Fig. 2: Test image

Check no.

Verification for Modality adjustmentReference

valueCT

ApplicationImage Type Correction

LUTMax.

Density

1Desired density CT Image CT LUT 2.80 Refer to the

table for CTLUT

2HCC linearity CT Image Linear 2.80 Refer to the

table for lin-ear LUT

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 19 of 260

Page 20: CT00-000.814.30.02.02.pdf

20 General Hardcopy Camera Information

The density difference of the BU LUT (Ref. LUT) between steps 1, 2, and 3 has to be a minimum density value of 0.02. Test image 3 (SMPTE image) can be used for visual veri-fication (the 95% gray square has to be visible).

Reference films 0

Reference film

To create a reference film with the desired LUT adjustment, proceed as follows:

1. The layout is configurable under Option/Configuration/Filming layout

2. Use format 3.2; default center/width, for the film sheet

Step

CT LUT Linear LUT

Ref. LUTMaximum Density

Minimum Density Ref. LUT

Maximum density

Minimum density

1 0.20 0.23 0.17 0.20 0.23 0.17

2 0.23 0.26 0.20 0.36 0.40 0.33

3 0.28 0.32 0.24 0.53 0.56 0.49

4 0.35 0.39 0.31 0.69 0.73 0.64

5 0.43 0.48 0.38 0.85 0.90 0.80

6 0.52 0.57 0.47 1.01 1.06 0.96

7 0.63 0.69 0.57 1.18 1.23 1.12

8 0.75 0.81 0.69 1.34 1.40 1.28

9 0.88 0.95 0.82 1.50 1.57 1.44

10 1.04 1.11 0.97 1.66 1.73 1.59

11 1.21 1.28 1.14 1.83 1.90 1.75

12 1.41 1.49 1.33 1.90 2.07 1.91

13 1.64 1.72 1.56 2.15 2.23 2.07

14 1.88 1.97 1.79 2.31 2.40 2.23

15 2.13 2.22 2.04 2.48 2.57 2.38

16 2.44 2.54 2.34 2.64 2.73 2.54

17 2.80 2.90 2.70 2.80 2.90 2.70

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 20 of 260

Page 21: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 21

3. Select the "Viewing" tab, select Patient Browser, navigate to the folder "CT Reference Images" and view the test images:

Use test images (technical images) without text and the three CT Medical Ref images (cerebrum, pancreas, (LWS) lumbar spine) with text and print the film sheet. (the DICOM command to the printer is always "magnification none")

4. Measure IEC test image 1 (Fig. 2 / p. 19) and verify the CTLUT using the table "Density values for modality CT”

Segment layout:

5. Repeat this procedure with the linear LUT adjustment

6. If the reference images are not available, please load the reference images from

H:\SITE_DATA\Reference-images\* in the storage device (import from offline)

Description of Correction LUT Files 0

Description of Correction LUT Files for CT

1 2

3 4

5 6

Segment Image Window Center

1 IEC test image 1 (17 gray levels) 256 0

2 IEC test image 2 (50% gray field) 256 0

3 Test image 3 (SMPTE test pattern) 100 0

4 Pancreas (Abdomen) 520 104

5 Cerebrum (Head) 80 53

6 LWS (Spine) 440 56

Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80

Linear Linear LUT

CT LUT Standard LUT; default for all cameras; adapts Dmin. of 0.2

CT02 vs. CT LUTDark area = less contrast; darker; softer; white area = more con-trast

CT03 vs. CT02Dark area = less contrast; darker; softer; white area = more con-trast

CT04 vs. CT LUTDark area = more contrast; darker; sharper; white area: no changes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 21 of 260

Page 22: CT00-000.814.30.02.02.pdf

22 General Hardcopy Camera Information

Additional and unnecessary LUT files 0

If new, released LUT files are available on a removable medium (e.g., floppy or CD). These files may be selected on the source drive and copied to the dependent LUT directories on the local system. Unnecessary LUT files may be removed from LUT directories on the local system.

Importing LUT files

1. In Local Service Configuration platform, select LUT files

2. Select the (source) drive and the path where the appropriate LUT files are located

3. Select the destination LUT directory (correction LUT directory) which depends on the image type (CTImage).

4. By pushing the copy button, the selected LUT file will be transferred to CTImage type.

5. Repeat this procedure using the appropriate LUT files for SC Images.

6. Select the destination LUT directory (correction LUT directory) which depends on the image type (SCImage).

7. Pushing the copy button transfers the selected LUT file to SCImage type.

8. Confirm the message "Files successfully copied" by clicking "OK"

Removing LUT files

Unnecessary LUT files can be removed from LUT directories on the local host by selecting the appropriate LUT file and clicking the "Remove" button.

• Click ">"

CT05 vs. CT LUTBrighter; no changes in contrast for large window width; more contrast for head images (small window width)

CT06 vs. CT LUT Brighter than CT5; white area = less contrast

cthead1 LUT for head images, images smoother

cthead2 vs. cthead1 Images smoother

CT015Adapts Dmin of 0.15 (for cameras which are not supported Dmin. command)

CT017Adapts Dmin. of 0.17 (for cameras which are not supported Dmin. command)

CT LUTdryLUT for dry cameras; adapts Dmin. of 0.22 (5% visible in SMPTE test pattern)

CT Lung LUT for lung images

Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 22 of 260

Page 23: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 23

Real Size Scaling

This procedure should be performed after system installation and when the Hardcopy Camera is ready to print. The real size behavior has to be performed with each connected Hardcopy Camera.

Real size scaling verification

1. Load the test image (water image phantom) into the viewer (e.g. 20 cm water phantom, generated with the constancy procedure by CT).

2. Insert Crosshair (Viewing Card: Tools) and copy the image to the film sheet (“copy image to film sheet” button).

3. Select film sheet format 1 X 1.

4. Select "Original Image" in the filming card and print the image.

5. Use a calibrated ruler to measure the crosshairs of the printed image.

The results should correspond to the printed value underneath the scale with a tolerance of ±1% in both directions.

Refer to the specific camera information.

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 23 of 260

Page 24: CT00-000.814.30.02.02.pdf

24 General Hardcopy Camera Information

Postscript Hardcopy Information 1.2

Restrictions and Known Effects 0

Restrictions

NOTE Paper printouts are not allowed for medical diagnosis.

NOTE Only one Adobe Postscript Printer can be installed on a64-bit computer.

More than one Postscript Printer Installation will not work.

NOTE The adobe printer driver with the printer-specific descriptionfile has to be used. The paper printer is not linear in greyscaleimage print-out.

Therefore the respective PSLUT Files has to be used to opti-mize the image quality for the print-out.

The PSLUT files are:

• EPSON_C4200_correction.lut

• HP_LaserJet4700_linear.lut

• Kyocera_FS_C5020N_Cor.lut

• inv_CTLUT.lut

NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g calcium scoring) usethe report of the applications to print out the images of theresults.

Known Effects

n.a.

Postscript Driver 0

General information

The Printer Driver Installation of Somaris software includes the particular printer descrip-tion file. Both have to be performed directly in WINDOWS XP. The generic Adobe Post-Script Printer Driver is already preconfigured and also visible in Somaris as “SIEMENS.PD” printer. Only the postscript description file (ppd - file) for the specific printer has to be installed additionally.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 24 of 260

Page 25: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 25

Missing Installation of Adobe Postscript Driver V5.2 in Windows XP

Installation of Adobe Postscript

Alternatives for special printer setting

Missing Installation of Adobe Postscript Driver V.5.2 in Windows XP

If the “SIEMENS.PD” is not available (e.g. deleted), a new installation of this generic Adobe driver is required

Log into your system with your administrator account.

Navigate to folder: C:\Somaris\Tools\Postscript\XP64.

Double-click the patch procedure:

Install_Printer_Driver.bat

Installation of Adobe Postscript Printer in Windows XP

Log into your system with your administrator account.

Navigate to folder: C:\Somaris\Tools\Postscript.

Double-click the file PostScriptInst.exe

click <browse>

----> "browse for folder " window appears.

Select the appropriate post script printer name

1. HP Color Laser Jet 4700N

select folder PPD\hp and click OK:

A list of available HP - printers appears

Select "hpmcpd25.ppd" and press Go

The window is closed

Restart system

Login as administrator

2. Epson Aculaser C4200

select folder PPD\epson and click OK:

A list of available EPSON printers appears

Select "epalc420.ppd” and press Go

The window is closed

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 25 of 260

Page 26: CT00-000.814.30.02.02.pdf

26 General Hardcopy Camera Information

“Configure the postscript printer driver as described in the part of printer informa-tion specific for the respective PS printer”.

Restart system

Login as administrator

Configuration of Adobe Postscript Printer in Windows XP

Start --> Settings --> Printers and Faxes

click “SIEMENS.PD” with the right mouse button

select properties

change in “General” the name “SIEMENS.PD” to the appropriate printer name (e.g. HP4700N)

click <OK>

select e.g. "HP4700N"

and with right mouse key

select "Properties"

The "HP4700N Properties"- window appears.

Select: "Ports" and click <Add Port>

"Printer Ports" window

Select: "StandardTCP/IP Port" and "New Port"

"Add Standard TCP/IP Printer Port Wizard" window appears,

click <Next>

Enter the IP address of the printer in the field:

Printer Name or IP Address: xxx.xxx.xxx.xxx

Port Name:

Let Windows choose the port name.

<Next>

"Add Standard TCP/IP printer Port Wizard" window

"Standard" and "Generic Network Card" is selected

Click : <Next>

Click : <Finish>

<Close> "Printer port" window

<Close> "HP4700N Properties" window

2. Epson Aculaser C4200

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 26 of 260

Page 27: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 27

Configure in Local Service Configuration

Select "Paper Printer" in Configuration Menu.

Select “Local service” -> configuration -> Paper Printer

Deselect all paper printer media that the printer does not support.

<save>,

click “>”

The “Printer Paper size Mapping” window appears

select the appropriate printer in Configured Printers

a table for mapping of paper sizes appears

Assign the real names of the printer paper sizes to the fixed syngo paper size names.

e.g. Printer Paper Size Mapping

Configured Printer e.g. HP 4700N

Syngo Paper Size Printer Paper Size

Letter Small --> Letter

A4 --> A4

Configure in Local Service Configuration

<save>,

<finish>,

Select "Local service" -> configuration -> PS LUT Files

Go to the PSLUT directory and select the specific Postscript LUT file for this printer (names of PS-LUTs, refer to the description specific to the printer):

Select: "Use LUT for Postscript Printers"

Select: “Use DICOM LUT for Postscript Printers”

Select: "Use floating Zoom factors"

Click : <Save>

Click : <Finish>

Click : <Home>

Restart Somaris

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 27 of 260

Page 28: CT00-000.814.30.02.02.pdf

28 General Hardcopy Camera Information

Look-Up Tables (LUT) 0

CT Reference Film and LUT

To create a reference, film the images of the patient: “Reference Images” have to be used. It contains 3 test images and 3 medical images. Concerning the positions of the images on the paper, the order is from top left to right bottom. Test images cannot contain image text, medical images need the full text.

The measured values are described in the specific Printer Information included in this doc-ument.

Image Attributes for syngo-based modalities

Precondition

Segment Image Window Center

1 IEC test image 1 (17 gray levels) 256 0

2 IEC test image 2 (50% gray field) 256 0

3 Test image 3 (SMPTE test pattern) 100 0

4 Pancreas (Abdomen) 520 104

5 Cerebrum (Head) 80 53

6 LWS (Spine) 440 56

Max. Density 2.8

Max. Density overall 3.0

Interpolation for printing Bilinear

Transformation No magnification

Segment lines no

Page number on print-out Not selected

Layout divisions 2x3

Orientation portrait

Image order Horizontal

Paper size A4

Reference image Not selected

Modality

The modality has to be configured for image types “CT Image” and for “SC Image” to CTLUT.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 28 of 260

Page 29: CT00-000.814.30.02.02.pdf

General Hardcopy Camera Information 29

Postscript Printer LUT

Real size scaling

Load the test image (water image phantom) into the viewer (e.g. 20 cm water phantom, generated with the CT constancy procedure ). Insert crosshair with tool in viewer. Send this image from the viewer to the filming tab card and expose it on paper.

Evaluate the printed paper. Search 2 scale lines of the crosshair with the largest possible distance to each other, separate in the x and y direction and noted as the desired value.

Measure the distance with the ruler and compare the measured value with the described values written in the specific Printer Information included in this document.

Image attributes for syngo-based modalities

Printer

The printer must be switched to black & white.

In "Local service" -> configuration -> PS LUT Files

The respective PS LUT file has to be selected

Postscript Printer PS LUT file name

Epson AcuLaser C4200 EPSON_C4200_correction.lut

Hewlett Packard Color Laserjet 4700 HP_LaserJet4700_linear.lut

Kyocera Ecosys FS-C5020N Kyocera_FS_C5020N_Cor.lut

Xerox Phaser 6350 DP inv_CTLUT.lut

Transformation No magnification

Interpolation for printing Bilinear

Segment lines no

Page number on print-out Not selected

Layout divisions 1x1

Orientation Portrait/landscape

Paper size A4/Letter small

Aspect ratio Original image

Scale factor 100%

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 29 of 260

Page 30: CT00-000.814.30.02.02.pdf

30 Specific Camera Information - Agfa

2- 2Specific Camera Information - AgfaAgfa Drystar AXYS 0

General Information 0

Block Diagram

Fig. 3:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film type Medium type

14" x 17"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

14" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

11" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

10” x 12”Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

8" x 10"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 30 of 260

Page 31: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 31

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE European EMC Classification is Class A.

NOTE Slight inhomogeneity visible in gray field

NOTE Typical slight line structure of the thermo head is visible intechnical image.

NOTE Two edges of the diagnostic area are cut (approx 1 mm) for8x10, 10x12, 11x14,14x17.

NOTE The background command does not work properly. The back-ground is always white.

Known effects

NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.

SCU / modality software configuration 0

Time 74 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title AXYS Yes

Port number 104 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 31 of 260

Page 32: CT00-000.814.30.02.02.pdf

32 Specific Camera Information - Agfa

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfadrystar5503_AX

YSNo

DICOM Node DSAXYS Yes

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 50.0 x 50.0 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

6922 x 8368

14" x 14"

6882 x 6882

11" x 14"

5376 x 6922

10" x 12"

4880 x 5860

8" x 10"

3852 x 4880

In “define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17"

8368 x 6922

14" x 14"

6882 x 6882

11" x 14"

6922 x 5376

10" x 12"

5860 x 4880

8" x 10"

4880 x 3852

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 32 of 260

Page 33: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 33

NOTE Do not modify or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / hardcopy camera configuration (provided by OEM) 0

Identification

Printer Settings

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM printer

Device name: DRYSTAR AXYS

Manufacturer: AGFA

SW Revision: R 1.60C1

Physical interface: Ethernet 10/100 Base-Tx

Network protocol: TCP/IP DICOM

DICOM conformance statement: Document No.: 001150 Revision 1.0

Image memory: 65 GByte

Hard disk: 70 GByte

Selected Variable name Setting

n.a. Nickname CT hostname

n.a. Use this profile only for incoming call-ing AE Title

PR_hostname

n.a. Use this profile only for local SCP server

DEFAULT_SCP

n.a. Host profile type GENERIC-GENERIC (customized)

No Number of Copies 1

No Print priority Low

No Polarity Normal

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 33 of 260

Page 34: CT00-000.814.30.02.02.pdf

34 Specific Camera Information - Agfa

No Film orientation Portrait

No Trim No

No Film Size ID (empty)

No Medium type (empty)

n.a. Print even if Film Size ID / Medium Type is not supported

No

n.a. Print even if Application Type is not supported

No

No Initial memory size allocated per print job (Mbytes):

100

No Kernel Site, Edit Custom:

Interpolation type;

Cubic high res. -smoothing factor -0.7

No Magnification Magn. none

No Decimate/Crop behavior: Disabled

No Pixel Size ratio 1.0 / 1.0

No Perception LUT Linear

n.a. Kanamori-like value 0

n.a. Custom Value empty

No Illumination (cd/m) 2000

No Reflected ambient light (cd/m2) 10

No Border density (OD x 100) Black

No Empty image density (OD x 100) Black

No Minimum density empty

No Maximum density Site 320

n.a. Header annotation Contents: Edit custom:

None

Courier 12

n.a. Footer annotation Contents: Edit custom:

None

Courier 12

n.a. Window (%) Empty

n.a. Level (%) Empty

Selected Variable name Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 34 of 260

Page 35: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 35

Error / warning messages (examples)

n.a. Image orientation r0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo modality site No

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT BIG ENDIAN Selected

n.a. Resolution 508

n.a. Supported SOP classes Presentation LUT SOP class

Print job SOP class

Printer SOP class

Basic annotation box SOP class

Verification service class

Basic film box SOP class

Basic film session SOP class

Basic color PM Meta SOP class

Print Queue Management SOP INSTANCE

Basic Grey Image Box SOP Class

Basic color image box SOP class

Basic gray PM Meta SOP class

Print queue management SOP class

n.a. LR like off

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open

Tray open or front door open

BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 35 of 260

Page 36: CT00-000.814.30.02.02.pdf

36 Specific Camera Information - Agfa

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Send print job to camera if the camera is in service mode

PRINTER OFFLINE PRINTER OFFLINE Camera offline

Sent print job to cam-era with empty input tray

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 390.0 391.0 0.26

14" x 17"

Landscape

X 410.0 411.0 0.24

Y 320.0 320.0 0.00

14" x 14"

Portrait

X 330.0 331.0 0.30

Y 320.0 320.0 0.00

14" x 14"

Landscape

X 330.0 330.0 0.00

Y 330.0 331.0 0.30

11" x 14"

Portrait

X 260.0 261.0 0.38

Y 330.0 330.5 0.15

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 250.0 251.0 0.40

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 271.0 0.37

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 36 of 260

Page 37: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 37

10” x 12”

Landscape

X 280.0 280.5 0.18

Y 220.0 220.0 0.00

8" x 10"

Portrait

X 180.0 180.0 0.00

Y 220.0 220.0 0.00

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 37 of 260

Page 38: CT00-000.814.30.02.02.pdf

38 Specific Camera Information - Agfa

Agfa Drystar 5503 2.1

General information 0

Block diagram

Fig. 4:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film Type Medium type

14" x 17"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

14" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

11" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

10” x 12”Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

8" x 10"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 38 of 260

Page 39: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 39

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

NOTE European EMC Classification is Class A.

NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.

NOTE Typical line structure of the thermo head is visible.

Known effects

NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.

SCU / Modality Software Configuration 0

Time 70 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY5503 Yes

Port number 104 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 39 of 260

Page 40: CT00-000.814.30.02.02.pdf

40 Specific Camera Information - Agfa

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfadrystar5503_AX

YSNo

DICOM Node DRY5503 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 50.0 x 50.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

6922 x 8368

14" x 14"

6882 x 6882

11" x 14"

5376 x 6922

10" x 12"

4880 x 5860

8" x 10"

3852 x 4880

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

8368 x 6922

14" x 14"

6882 x 6882

11" x 14"

6922 x 5376

10" x 12"

5860 x 4880

8" x 10"

4880 x 3852

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 40 of 260

Page 41: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 41

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM printer

Device name: DRYSTAR 5503

Manufacturer: AGFA

SW Revision: R 5.0.0 B2

Physical Interface: Ethernet 10/100 Base-Tx

Network protocol: TCP/IP DICOM

DICOM Conformance Statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5503, SW ver-sion 5.0

Image memory: 2 * 256 Mbyte (RAM memory)

Hard disk: min. 18 Gbyte

Selected Variable name Setting

n.a. Nickname CT hostname

n.a. Use this profile only for incoming call-ing AE Title

PR_hostname

n.a. Use this profile only for local SCP server

DEFAULT_SCP

n.a. Host profile type GENERIC-GENERIC (customized)

No Number of Copies 1

No Print priority Low

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 41 of 260

Page 42: CT00-000.814.30.02.02.pdf

42 Specific Camera Information - Agfa

No Polarity Normal

No Film destination output tray n.a.

n.a. Tray 1 Selected

n.a. Tray 2 Deselected

n.a. Tray 3 Deselected

n.a. Tray 4 Deselected

No Film orientation Portrait

No Trim No

No Film Size ID (empty)

No Medium type “empty”

n.a. Print even if Film Size ID / Medium Type is not supported

No

n.a. Print even if Application Type is not supported

No

No Initially allocated memory size per print job (Mbytes):

100

No Kernel Site, Edit Custom:

Interpolation type;

Cubic high res. -smoothing factor -0.7

No Magnification Magn. none

No Pixel Size ratio 1.0 / 1.0

No Perception LUT Linear

n.a. Kanamori-like value 0

n.a. Custom Value “empty”

No Illumination (cd/m) 2000

No Reflected ambient light (cd/m2) 10

No Border density (OD x 100) Black

No Empty image density (OD x 100) Black

No Minimum density “empty”

No Maximum density Site 320

n.a. Header annotation Contents: Edit custom:

None

Courier 12

Selected Variable name Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 42 of 260

Page 43: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 43

n.a. Footer annotation Contents: Edit custom:

None

Courier 12

n.a. Window (%) “empty”

n.a. Level (%) “empty”

n.a. Image orientation r0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo modality site, No

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT BIG ENDIAN Selected

n.a. Resolution 508

n.a. Supported abstract syntax Presentation LUT SOP class

Print job SOP class

Printer SOP class

Basic annotation box SOP class

Verification service class

Basic film box SOP class

Basic film session SOP class

Basic color PM Meta SOP class

Print Queue Management SOP INSTANCE

Basic Grey Image Box SOP Class

Basic Color Image Box SOP Class

Basic gray PM Meta SOP class

Print queue management SOP class

n.a. Early processing allowed Yes

n.a. LR like off

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 43 of 260

Page 44: CT00-000.814.30.02.02.pdf

44 Specific Camera Information - Agfa

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Open front door COVER OPEN COVER OPEN Camera cover, drawer or door open

Tray open or front door open

BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine

Sent print job to cam-era with empty input tray

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Camera is warming up

PRINTER INIT PRINTER INIT Camera initializing

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 390.0 389.5 0.13

14" x 17"

Landscape

X 410.0 409.5 0.12

Y 330.0 330.0 0.00

14" x 14"

Portrait

X 330.0 330.0 0.00

Y 320.0 319.5 0.16

14" x 14"

Landscape

X 330.0 329.5 0.15

Y 330.0 330.0 0.00

11" x 14"

Portrait

X 250.0 250.0 0.00

Y 320.0 320.0 0.00

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 250.0 250.0 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 44 of 260

Page 45: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 45

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 270.0 0.00

10” x 12”

Landscape

X 280.0 280.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 180.0 180.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 45 of 260

Page 46: CT00-000.814.30.02.02.pdf

46 Specific Camera Information - Agfa

Agfa Drystar 5500 2.2

General Information 0

Block Diagram

Fig. 5:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film type Medium type

14" x 17"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

14" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

11" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

10” x 12”Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

8" x 10"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 46 of 260

Page 47: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 47

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.

NOTE Typical line structure of the thermo head is visible.

Known effects

NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.

SCU / Modality Software Configuration 0

Time 69 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY5500 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 47 of 260

Page 48: CT00-000.814.30.02.02.pdf

48 Specific Camera Information - Agfa

Class

(Printer spec file name)agfadrystar5500 No

DICOM Node DRY5500 Yes

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 50.0 x 50.0 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

6922 x 8368

14" x 14"

6882 x 6882

11" x 14"

5376 x 6922

10" x 12"

4880 x 5860

8" x 10"

3852 x 4880

In “define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17"

8368 x 6922

14" x 14"

6882 x 6882

11" x 14"

6922 x 5376

10" x 12"

5860 x 4880s

8" x 10"

4880 x 3852

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

HC Device /

General settingsDisplay Changes allowed

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 48 of 260

Page 49: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 49

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer Settings

Device type: DICOM printer

Device name: DRYSTAR 5500

Manufacturer: AGFA

SW Revision: V3.0.0 C1

Physical interface: Ethernet (RJ45)

Network protocol: TCP/IP

DICOM conformance statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5500, Document No.: 737

Image memory: 2 * 256 Mbyte

Hard disk: 18 GByte

Selected Variable name Setting

n.a. Nickname CT hostname

n.a. Use this profile only for incoming call-ing AE Title

PR_hostname

n.a. Use this profile only for local SCP server

DEFAULT_SCP

n.a. Host profile type GENERIC-GENERIC (customized)

No Number of Copies 1

No Print priority Low

No Polarity Normal

No Film destination output tray n.a.

n.a. Tray 1 Selected

n.a. Tray 2 Deselected

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 49 of 260

Page 50: CT00-000.814.30.02.02.pdf

50 Specific Camera Information - Agfa

n.a. Tray 3 Deselected

n.a. Tray 4 Deselected

No Film orientation Portrait

No Trim No

No Film Size ID (empty)

No Medium type (empty)

n.a. Print even if Film Size ID / Medium Type is not supported

No

n.a. Print even if Application Type is not supported

No

No Initial memory size allocated per print job (Mbytes):

100

No Kernel Site, Edit Custom:

Interpolation type;

Cubic high res. -smoothing factor -0.7

No Magnification Magn. none

No Pixel Size ratio 1.0 / 1.0

No Perception LUT Linear

n.a. Kanamori-like value 0

n.a. Custom Value “empty”

No Illumination (cd/m) 2000

No Reflected ambient light (cd/m2) 10

No Border density (OD x 100) Black

No Empty image density (OD x 100) Black

No Minimum density “empty”

No Maximum density Site 320

n.a. Header annotation Contents: Edit custom:

None

Courier 12

n.a. Footer annotation Contents: Edit custom:

None

Courier 12

n.a. Window (%) Empty

Selected Variable name Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 50 of 260

Page 51: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 51

n.a. Level (%) Empty

n.a. Image orientation r0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo modality site, No

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT BIG ENDIAN Selected

n.a. Resolution 508

n.a. Supported abstract syntax Presentation LUT SOP class

Print job SOP class

Printer SOP class

Basic annotation box SOP class

Verification service class

Basic film box SOP class

Basic film session SOP class

Basic color PM Meta SOP class

Print Queue Management SOP INSTANCE

Basic Grey Image Box SOP Class

Basic Color Image Box SOP Class

Basic gray PM Meta SOP class

Print queue management SOP class

n.a. Early processing allowed Yes

n.a. LR like off

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 51 of 260

Page 52: CT00-000.814.30.02.02.pdf

52 Specific Camera Information - Agfa

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Sent print job to cam-era with front door open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Tray open or front door open

BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supply magazine

Sent print job to cam-era with empty input tray

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Printer warming up PRINTER INIT PRINTER INIT Camera initializing

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 400.0 399.0 0.25

14" x 17"

Landscape

X 410.0 409.0 0.24

Y 330.0 330.0 0.00

14" x 14"

Portrait

X 330.0 329.5 0.15

Y 330.0 330.0 0.00

14" x 14"

Landscape

X 330.0 330.0 0.00

Y 330.0 329.5 0.15

11" x 14"

Portrait

X 250.0 249.5 0.20

Y 330.0 330.0 0.00

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 250.0 249.5 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 52 of 260

Page 53: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 53

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 269.5 0.19

10” x 12”

Landscape

X 280.0 279.5 0.18

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 180.0 179.5 0.28

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 169.5 0.29

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 53 of 260

Page 54: CT00-000.814.30.02.02.pdf

54 Specific Camera Information - Agfa

Agfa Drystar 5302 2.3

General information 0

Block diagram

Fig. 6:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 11

Format Release status Film type Medium type

14" x 17"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

14" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

11" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

10” x 12”Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

8" x 10"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 54 of 260

Page 55: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 55

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

NOTE If the film size was not included in the configuration of thecamera, the connection to the camera is not available.

Known effects

NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.

SCU / Modality Software Configuration 0

Time 67 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY5302 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec. file name)agfadrystar5302 No

DICOM Node Agfa5302 Yes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 55 of 260

Page 56: CT00-000.814.30.02.02.pdf

56 Specific Camera Information - Agfa

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 79.375 x 79.375 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4358 x 5232

14" x 14"

4358 x 4303

11" x 14"

3348 x 4358

10” x 12”

3070 x 3653

8" x 10"

2376 x 3070

In “Define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5232 x 4358

14" x 14"

4303 x 4358

11" x 14"

4358 x 3348

10” x 12”

3653 x 3070

8" x 10"

3070 x 2376

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 56 of 260

Page 57: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 57

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Device type: DICOM printer

Device name: DRYSTAR 5302

Manufacturer: AGFA

SW Revision: V1.80 c1

Physical Interface: Ethernet (RJ45)

Network protocol: TCP/IP

DICOM Conformance Statement: Revision 1.1, June 2005

Image memory: 512 Mbyte (internal memory)

Hard disk: 35 GB

Selected Variable name Setting

n.a. Nickname CT hostname

n.a. Use this profile only for incoming "Calling AE Title":

PR_hostname

n.a. Use this profile only for local SCP server DEFAULT_SCP

n.a. Host Profile Type GENERIC-GENERIC (custom-ized)

No Number of Copies 1

No Print priority Low

No Polarity Normal

No Film orientation Portrait

No Trim No

No Film Size ID Deselected

No Medium type Deselected

n.a. Print even if Film Size ID / Medium Type is not supported

No

n.a. Print even if Application Type is not sup-ported:

No

No Initial allocated memory size per print job (Mbytes):

100

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 57 of 260

Page 58: CT00-000.814.30.02.02.pdf

58 Specific Camera Information - Agfa

No Kernel SITE (*) Edit Custom:

Interpolation type: Cubic high res. -smoothing factor -0.7

No Magnification Magn: Max. - custom scale fac-tor 1.0

No Pixel Size Ratio: 1.0 / 1.0

No Perception LUT Linear

n.a. Kanamori-like value 0

n.a. Custom Value Deselected

No Illumination (cd/m2) 2000

No Reflected ambient light (cd/m2) 10

No Border density (OD x 100) Black

No Empty image density (OD x 100) Black

No Minimum density Empty

No Maximum density Site 300

n.a. Header annotation Contents: Edit custom:

None

Courier 12

n.a. Footer annotation Contents: Edit custom:

None

Courier 12

n.a. Window (%) Empty

n.a. Level (%) Empty

n.a. Image orientation R0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo modality Site, No.

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN Selected

n.a. ALLOW EXPLICIT VR LITTLE ENDIAN Selected

n.a. ALLOW EXPLICIT BIG ENDIAN Selected

Selected Variable name Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 58 of 260

Page 59: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 59

Error / Warning Messages (examples)

n.a. Resolution 320

n.a. Supported SOP classes: Edit

n.a. Supported abstract syntax Presentation LUT SOP class

Print job SOP class

Printer SOP class

Basic annotation box SOP class

Verification service class

Basic film box SOP class

Basic film session SOP class

Basic color PM Meta SOP class

Print Queue Management SOP INSTANCE

Basic Grey Image Box SOP Class

Basic Color Image Box SOP Class

Basic gray PM Meta SOP class

Print queue management SOP class

n.a. LR-like Off

Action that results in camera error mes-sages

Printer messagesMessage displayed on

SCUDefined DCS (*1)

Sent by SCP (*2)

Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open

Open film supply tray BAD SUPPLY MGZ

BAD SUPPLY MGZ

Problem with supply magazine

Empty input tray SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Sent wrong film size n.a. n.a. Status -> aborted

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 59 of 260

Page 60: CT00-000.814.30.02.02.pdf

60 Specific Camera Information - Agfa

NOTE If film size is not inserted in configuration of camera, connec-tion to camera is not available.

Image Quality and Optical Density 0

No specific parameter available.

Real size scaling

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 390.0 391.0 0.26

14" x 17"

Landscape

X 410.0 411.0 0.24

Y 330.0 330.0 0.00

14" x 14"

Portrait

X 330.0 331.0 0.30

Y 330.0 330.0 0.00

14" x 14"

Landscape

X 330.0 330.0 0.00

Y 330.0 331.0 0.30

11" x 14"

Portrait

X 250.0 250.5 0.20

Y 330.0 330.0 0.00

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 250.0 251.0 0.40

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 271.0 0.37

10” x 12”

Landscape

X 280.0 281.0 0.36

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 170.0 170.5 0.29

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 170.0 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 60 of 260

Page 61: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 61

Agfa Drystar 5300 2.4

General information 0

Block diagram

Fig. 7:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 11

Format Release status Film Type Medium type

14" x 17"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

14" x 14"Portrait n.a. N/A n.a.

Landscape n.a. N/A n.a.

11" x 14"Portrait Yes DT2B Blue-based

Landscape Yes DT2B Blue-based

10” x 12”Portrait n.a. N/A n.a.

Landscape n.a. N/A n.a.

8" x 10"Portrait n.a. N/A n.a.

Landscape n.a. N/A n.a.

Other n.a. N/A n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 61 of 260

Page 62: CT00-000.814.30.02.02.pdf

62 Specific Camera Information - Agfa

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

Known effects

NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.

SCU / Modality Software Configuration 0

Time 67 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY5300 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfadrystar5300 No

DICOM Node DS5300 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 24 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 62 of 260

Page 63: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 63

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Pixel Size [1/1000 mm] 79.375 x 79.375 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4256 x 5174

11" x 14"

3300 x 4256

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5174 x 4256

11" x 14"

4256 x 3300

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM print server/printer

Device name: DRYSTAR 5300

Manufacturer: AGFA

SW Revision: 1.82 C1

Physical Interface: 10/100 Base - T

Network protocol: TCP/IP

DICOM Conformance Statement: Rev.1.2, October 24, 2005

Image memory: 512 Mbyte

Hard disk: None

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 63 of 260

Page 64: CT00-000.814.30.02.02.pdf

64 Specific Camera Information - Agfa

Printer settings

Selected Variable name Setting

n.a. Nickname CT hostname

n.a. Use this profile only for incoming “Calling AE title”

PR_hostname

n.a. Use this profile only for local SCP server

All

n.a. Host Profile Type GENERIC-GENERIC (customized)

No Number of Copies 1

No Print priority Low

No Polarity Normal

No Film orientation Portrait

No Trim No

No Film Size ID Deselected

No Medium type Deselected

n.a. Print even if Film Size ID / Medium Type is not supported

No

No Kernel SITE

Interpolation type: Cubic High Res - Smooth Factor -2.5

No Magnification Magn: Max. - custom scale factor 1.0

No Pixel Size Ratio: 1.0 / 1.0

No Perception LUT Linear

No Kanamori-like value 0

No Custom Value Deselected

No Illumination (cd/m) 2000

No Reflected ambient light (cd/m2) 10

No Border density (OD x 100) Black

No Empty image density (OD x 100) Black

No Minimum density Empty

No Maximum density Site 300

No. Annotation 1 None

No Annotation 2 None

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 64 of 260

Page 65: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 65

n.a. Window (%) Empty

n.a. Level (%) Empty

n.a. Image orientation R0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo Modality site: No.

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT VR LITTLE ENDIAN

Selected

n.a. ALLOW EXPLICIT BIG ENDIAN Selected

n.a. Resolution 320

n.a. Supported SOP Classes Presentation LUT SOP class

Print job SOP class

Printer SOP class

Basic annotation box SOP class

Verification service class

Basic film box SOP class

Basic film session SOP class

Basic Color PM Meta SOP Class

Print Queue Management SOP INSTANCE

Basic Grey Image Box SOP Class

Basic Color Image Box SOP Class

Basic gray PM Meta SOP class

Print queue management SOP class

n.a. Initial allocated memory size per print job (MBytes)

23

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 65 of 260

Page 66: CT00-000.814.30.02.02.pdf

66 Specific Camera Information - Agfa

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Action that results in camera error messages

Printer messages Message dis-played on SCU

(*2)Defined DCS (*1)

Sent by SCP

Open input tray BAD SUPPLY MGZ

WARNING\BAD SUPPLY MGZ

Camera info.: WARNING

Top cover open COVER OPEN WARNING\COVER OPEN Camera info.: WARNING

Empty input tray SUPPLY EMPTY

WARNING\SUPPLY EMPTY

Camera info.: WARNING

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 400.0 401.0 0.25

14" x 17"

Landscape

X 400.0 401.5 0.37

Y 320.0 320.0 0.00

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X 250.0 250.5 0.20

Y 320.0 320.0 0.00

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 250.0 250.5 0.20

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 66 of 260

Page 67: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 67

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 67 of 260

Page 68: CT00-000.814.30.02.02.pdf

68 Specific Camera Information - Agfa

Agfa Drystar 3000 2.5

General information 0

Block diagram

Fig. 8:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 4

Format Release status Film Type Medium type

14" x 17"Portrait Yes DT1B Blue-based

Landscape Yes DT1B Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait No No No

Landscape No No No

10” x 12”Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 68 of 260

Page 69: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 69

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

Known effects

NOTE Changing the film size requires manual modifications at thecamera site.

SCU / Modality Software Configuration 0

Time 160 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY3000 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfadrystar3000 No

DICOM Node DRY3000 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 69 of 260

Page 70: CT00-000.814.30.02.02.pdf

70 Specific Camera Information - Agfa

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Pixel Size [1/1000 mm] 79.4 x 79.4 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4224 x 4839

11" x 14"

3268 x 3924

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5142 x 3924

11" x 14"

4224 x 2962

Medium type Blue film Yes

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM printer

Device name: DRYSTAR 3000

Manufacturer: AGFA

SW Revision: MeGa 3.32, cad 32948, ecu 32950, DI301p95, R1.95, aos 32947, dcm 32957, mps 32941, OEM 49535

Physical Interface: 10 base - TP Ethernet

Network protocol: TCP/IP

DICOM Conformance Statement: Rev.2.3, July 31, 2003

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 70 of 260

Page 71: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 71

Printer settings

Error / Warning Messages (examples)

Image memory: 32 Mbyte

Hard disk: 1.2 Gbyte

File Name Content Remarks

pms.ini PR_hostname

IDF=

CFG=D:\MEDCOM2.PMS

No .idf file defined

PMS.IDF is the default

PMS.IDF default STANDARD\1,1 Agfa layout ID:

14x17P ID1411

14x17L ID1511

14x14P ID1611

11x14P ID1211

11x14L ID1311

8x10P ID1011

8x10L ID1111

Medcom2.pms 00

pms_virgin_cfg 0

n-event_enabled 0

trim_flag 0

utl_id 1

is used (DCS), no *.idf file defined

AGFA confirm Level is 0 (warnings)

Disable N-events messages

Disable trim around images

LINEAR LUT

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Switch camera from online to offline

PRINTER OFFLINE Association rejected

Connection to cam-era is not available

Open cover (lower) SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Top cover open n.a. CHECK PRINTER Check camera

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 71 of 260

Page 72: CT00-000.814.30.02.02.pdf

72 Specific Camera Information - Agfa

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 380.0 377.5 0.66

14" x 17"

Landscape

X 400.0 397.0 0.75

Y 300.0 300.0 0.00

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X No No No

Y No No No

11" x 14"

Landscape

X No No No

Y No No No

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 72 of 260

Page 73: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 73

Agfa LR5200 / MG3000 2.6

General information 0

Block diagram

Fig. 9:

Available film sizes

Format Release status Film Type Medium type

14" x 17"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

14" x 14"Portrait Yes

Scopix LT2B DLA1

Blue-based

Landscape No No No

11" x 14"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

10” x 12”Portrait n.a. N/A n.a.

Landscape n.a. N/A n.a.

8" x 10"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

Other n.a. N/A n.a.

Released: Yes - Film size is available and released

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 73 of 260

Page 74: CT00-000.814.30.02.02.pdf

74 Specific Camera Information - Agfa

Performance and Connections

Associations

Maximum number of associations supported: 4

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly.

NOTE No message will be displayed in case of manual input of filmsinto developer.

Known effects

NOTE The camera does support format 14”X14” landscape.

SCU / Modality Software Configuration 0

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

Time 228 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title LR5200 Yes

Port number 104 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 74 of 260

Page 75: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 75

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfalr5200 No

DICOM Node LR5200 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 80.0 x 80.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4224 x 4839

14" x 14"

4224 x 3897

11" x 14"

3268 x 3924

8" x 10"

2344 x 2647

In “define new”

Do not modify/add

Select only

Delete onlyFilm Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5140 x 3924

11" x 14"

4224 x 2962

8" x 10"

2948 x 2046

Medium type Blue film Yes

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 75 of 260

Page 76: CT00-000.814.30.02.02.pdf

76 Specific Camera Information - Agfa

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Device type: DICOM print server/printer

Device name: MG 3000

Manufacturer: AGFA

SW Revision: MeGa 3.32,

dcm 32957

Physical Interface: 10 base - TP Ethernet

Network protocol: TCP/IP

DICOM Conformance Statement: Rev.1.4, April 20, 2001

Image memory: 128 Mbyte

Hard disk: 4.1 Gbyte

Device type: DICOM camera

Device name: LR 5200

Manufacturer: AGFA

SW Revision: LTP 01806, 240698

Physical Interface: FOI optic

Network protocol: None

DICOM Conformance Statement: None

Image memory: n.a.

Hard disk: None

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 76 of 260

Page 77: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 77

Printer settings

Error / Warning Messages (examples)

File Name Content Remarks

pms.ini PR_hostname

IDF=

CFG=D:\MEDCOM2.PMS

No .idf file defined

PMS.IDF is the default

PMS.IDF default STANDARD\1,1 Agfa layout ID:

14x17P ID1411

14x17L ID1511

14x14P ID1611

11x14P ID1211

11x14L ID1311

8x10P ID1011

8x10L ID1111

Medcom2.pms 00

pms_virgin_cfg 0

n-event_enabled 0

trim_flag 0

utl_id 1

is used (DCS), no *.idf file defined

AGFA default confirm Level is 0 (warnings)

Disable N-events messages

Disable trim around images

LINEAR LUT

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Camera switched off or camera switched to offline

OFFLINE OFFLINE Camera is switched off-line

Manual input of films in developer

NORMAL NORMAL No message dis-played

Camera is switched on

PRINTER INIT PRINTER INIT Camera is warming up

Supply magazine empty

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 77 of 260

Page 78: CT00-000.814.30.02.02.pdf

78 Specific Camera Information - Agfa

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 331.5 0.45

Y 380.0 383.0 0.79

14" x 17"

Landscape

X 400.0 403.0 0.75

Y 300.0 301.0 0.33

14" x 14"

Portrait

X 330.0 332.0 0.61

Y 300.0 302.0 0.67

14" x 14"

Landscape

X No No No

Y No No No

11" x 14"

Portrait

X 250.0 252.0 0.80

Y 300.0 301.0 0.33

11" x 14"

Landscape

X 330.0 331.5 0.45

Y 220.0 221.5 0.68

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X 170.0 170.5 0.29

Y 200.0 201.5 0.75

8" x 10"

Landscape

X 230.0 232.0 0.87

Y 160.0 160.5 0.31

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 78 of 260

Page 79: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 79

Agfa LR5200 / LR DICOM Controller 2.7

General information 0

Block diagram

Fig. 10:

Available film sizes

Format Release status Film Type Medium type

14" x 17"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

14" x 14"Portrait Yes

Scopix LT2B DLA1

Blue-based

Landscape No No No

11" x 14"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

10” x 12”Portrait n.a. N/A n.a.

Landscape n.a. N/A n.a.

8" x 10"

Portrait YesScopix LT2B

DLA1Blue-based

Landscape YesScopix LT2B

DLA1Blue-based

Other n.a. N/A n.a.

Released: Yes - Film size is available and released

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 79 of 260

Page 80: CT00-000.814.30.02.02.pdf

80 Specific Camera Information - Agfa

Performance and Connections

Associations

Maximum number of associations supported: 15

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

Known effects

NOTE For page mode magnification, set type "NONE" and activatethe mark.

NOTE The camera does support format 14”X14” landscape.

SCU / Modality Software Configuration 0

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

Time 197 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title LRDICOM Yes

Port number 104 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 80 of 260

Page 81: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 81

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)agfalr5200 No

DICOM Node LR5200 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 80.0 x 80.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4224 x 4839

14" x 14"

4224 x 3897

11" x 14"

3268 x 3924

8" x 10"

2344 x 2647

In “define new”

Do not modify/add

Select only

Delete onlyFilm Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5140 x 3924

11" x 14"

4224 x 2962

8" x 10"

2948 x 2046

Medium type Blue film Yes

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 81 of 260

Page 82: CT00-000.814.30.02.02.pdf

82 Specific Camera Information - Agfa

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Device type: DICOM print server/printer

Device name: Agfa LR DICOM Controller

Manufacturer: AGFA

SW Revision: FLC1.01.14

Physical Interface: Gateway

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. 1.2, July 31, 2003

Image memory: 128 MB RAM

Hard disk: 20 Gbyte

Device type: DICOM camera

Device name: LR 5200

Manufacturer: AGFA

SW Revision: LTP 01806.240698

Physical Interface: FOI optic

Network protocol: None

DICOM Conformance Statement: None

Image memory: n.a.

Hard disk: None

Selected Variable name Setting

n.a. Calling AE title PR_Hostname

n.a. Nickname CT hostname

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 82 of 260

Page 83: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 83

n.a. Host profile type GENERIC-GENERIC (customized)

No Number of Copies 1

No Print priority Low

No Polarity Normal

No Film orientation Portrait

No Trim No

No Film Size ID Deselected

No Medium type Deselected

n.a. Print even if Film Size ID / Medium Type is not supported

No

No Kernel Site

Magnification type: cubicHighRes

Smooth Factor: - 0.7

Yes Magnification Type: None

Max Custom scale factor: 1.0

No Pixel Size Ratio: 1.0 / 1.0

No Perception LUT Linear

No Kanamori-like value 0

No Custom Value Deselected

No Illumination (cd/m) 2000

No Reflected ambient light (cd/m2)

10

No Border density (OD x 100) Black

No Empty image density (OD x 100)

Black

No Minimum density Empty

No Maximum density Site 320

n.a. Header annotation

No Content: Edit custom: none

Courier 12

n.a. Footer annotation

No Content: Edit custom: none

Courier 12

Selected Variable name Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 83 of 260

Page 84: CT00-000.814.30.02.02.pdf

84 Specific Camera Information - Agfa

n.a. Window (%) Empty

n.a. Level (%) Empty

n.a. Image orientation r0

n.a. Conformance level 1

n.a. N-event enabled Off

n.a. Mammo Site no.

n.a. Association time-out(s) 0

n.a. Image time-out(s) 0

n.a. ALLOW IMPLICIT VR LITTLE ENDIAN

Yes

n.a. ALLOW EXPLICIT VR LIT-TLE ENDIAN

Yes

n.a. ALLOW EXPLICIT BIG ENDIAN

Yes

n.a. Resolution 315

n.a. Supported abstract syntax Basic Film Box_SOP_Class

Basic gray PM Meta_SOP_Class

Print Queue Management_SOP_INSTANCE

Printer_SOP_Class

Basic Color Image Box_SOP_CLass

Basic Grey Image_Box_SOP_Class

Basic Film Session_SOP_Class

Print Job_SOP_Class

Presentation LUT_SOP_Class

Basic Annotation Bo_ SOP_Class

Verification service class

Print Queue Management_SOP_Class

Selected Variable name Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 84 of 260

Page 85: CT00-000.814.30.02.02.pdf

Specific Camera Information - Agfa 85

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Action that results in camera error mes-

sages

Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP

Camera switched off or

Camera switched to offline

PRINTER OFFLINE

PRINTER OFFLINE

Camera offline

Camera start-up PRINTER INIT PRINTER INIT Camera initializing

Manual input of film in developer

CHECK PRINTER

CHECK PRINTER

Check camera

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 331.5 0.45

Y 370.0 373.0 0.81

14" x 17"

Landscape

X 400.0 403.0 0.75

Y 300.0 301.5 0.50

14" x 14"

Portrait

X 330.0 331.0 0.30

Y 330.0 332.5 0.76

14" x 14"

Landscape

X No No No

Y No No No

11" x 14"

Portrait

X 250.0 252.0 0.80

Y 300.0 301.0 0.33

11" x 14"

Landscape

X 330.0 331.0 0.30

Y 220.0 221.5 0.68

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 85 of 260

Page 86: CT00-000.814.30.02.02.pdf

86 Specific Camera Information - Agfa

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X 170.0 170.0 0.00

Y 200.0 201.5 0.75

8" x 10"

Landscape

X 230.0 231.5 0.65

Y 160.0 160.5 0.31

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 86 of 260

Page 87: CT00-000.814.30.02.02.pdf

Specific Camera Information - Codonics 87

3- 3Specific Camera Information - CodonicsCodonics Horizon 0

General information 0

Block diagram

Fig. 11:

Available film sizes

Format Release status Film Type Medium type

14" x 17"

Portrait Yes DirectVistaGrayscale

Blue-based

Landscape Yes DirectVistaGrayscale

Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait No No No

Landscape No No No

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"

Portrait Yes DirectVistaGrayscale

Blue-based

Landscape Yes DirectVistaGrayscale

Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 87 of 260

Page 88: CT00-000.814.30.02.02.pdf

88 Specific Camera Information - Codonics

Performance and Connections

Associations

Maximum number of associations supported: More than 30

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The background command does not work properly. The back-ground is always white.

NOTE Film size 8” x 10” is not released for subdivision 1x1 with lin-ear LUT.

Film size 8” x 10” is not released also for subdivisionssmaller than 6 on 1 (4 on 1; 2 on 1; etc) with CT LUT.

NOTE Print-outs on paper are not released.

Known Effects

n/a

SCU / Modality Software Configuration 0

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

Time 95 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title HORIZON Yes

Port number 104 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 88 of 260

Page 89: CT00-000.814.30.02.02.pdf

Specific Camera Information - Codonics 89

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)codonicshorizon No

DICOM Node Horizon Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 79.4 x 79.4 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4322 x 5025

11" x 14"

3382 x 4022

8" x 10"

2406 x 2790

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5025 x 4322

11" x 14"

4022 x 3382

8" x 10"

2790 x 2406

Medium type Blue film Yes

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black No

Transformation No magnification No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 89 of 260

Page 90: CT00-000.814.30.02.02.pdf

90 Specific Camera Information - Codonics

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Manufacturer / device type: DICOM Printer (SCP)

Device name: Horizon

Manufacturer: Codonics

SW Revision: V 1.9.0

Physical type: Ethernet 10/100 base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: D.C.S., V1.9.0

Image memory: 256 MB RAM, 6 GB-HD

Hard disk: 10 GB

Content:

DV FILM BLUE Gamma:

Contrast:

Dmax:

Dmin:

Film View:

Illumination:

Reflected Light:

LUT:

Polarity:

Background:

Scaling:

Rotate:

Image Warnings:

Captions:

Border Fill:

Crop Anchor:

Scale Factor:

Coverage:

1.0

0

300

0

Front

2000

10

linear

Positive

Black

Bilinear

Auto

On

Off

On

Center

0.0

Standard

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 90 of 260

Page 91: CT00-000.814.30.02.02.pdf

Specific Camera Information - Codonics 91

Error / Warning Messages (examples)

Step in Service Mode (with "<" and "Menu")

select Job Settings

Select:

PR_hostname (AE title of modality)

Name:

Media

Receive Tray:

Priority:

Gamma:

Contrast:

Dmax:

Dmin:

Film View:

Illumination:

Reflected Light:

LUT:

Polarity:

Saturation:

MCM:

Background:

Scaling:

Rotate:

Image Warnings:

Captions:

Border Fill:

Crop Anchor:

Scale Factor:

Coverage:

PR_hostname

DV FILM BLUE 14X17

Default

Default

1.0

0

Default

Default

Front

Default

Default

linear

Positive

Default

Default

Black

Bilinear

Auto

On

Off

On

Center

Default

Standard

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Printer switched to service mode

PRINTER OFFLINE PRINTER OFFLINE Printer offline

Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open

Content:

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 91 of 260

Page 92: CT00-000.814.30.02.02.pdf

92 Specific Camera Information - Codonics

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Send job when no magazine is inserted

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

14”x17” blue film sup-ply empty

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.5 0.15

Y 380.0 380.5 0.13

14" x 17"

Landscape

X 390.0 390.5 0.13

Y 330.0 330.5 0.15

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 92 of 260

Page 93: CT00-000.814.30.02.02.pdf

Specific Camera Information - Codonics 93

8" x 10"

Portrait

X 170.0 170.0 0.00

Y 210.0 210.5 0.24

8" x 10"

Landscape

X 200.0 200.0 0.00

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 93 of 260

Page 94: CT00-000.814.30.02.02.pdf

94 Specific Camera Information - Fuji

4- 4Specific Camera Information - FujiFuji Drypix 7000 / 5000 0

General information 0

Block diagram

Fig. 12:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film Type Medium type

14" x 17"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

14" x 14"Portrait Yes DI-HL Blue-based

Landscape n.a. n.a. n.a.

11" x 14"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

10" x 12"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

8" x 10"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 94 of 260

Page 95: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 95

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Device is only class A, EN55011(Hospital use).

NOTE 14INX14IN landscape not supported by camera.

NOTE Film job that contains more than 10 films will not be printedout completely. (with SW-Revision 13.0-0001 only)

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 66 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY7000 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)fujidrypix7000 No

DICOM Node dry7000 Yes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 95 of 260

Page 96: CT00-000.814.30.02.02.pdf

96 Specific Camera Information - Fuji

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 100.0 x 100.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

3500 x 4170

14" x 14"

3500 x 3410

11" x 14"

2500 x 3495

10” x 12”

2460 x 2916

8" x 10"

1954 x 2410

In “define new”

Do not modify/add

Select only

Delete onlyFilm Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

4220 x 3360

11" x 14"

3600 x 2450

10” x 12”

2972 x 2404

8" x 10"

2466 x 1898

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 96 of 260

Page 97: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 97

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Device type: DICOM Printer (SCP)

Device name: Drypix 7000

Manufacturer: FUJI

SW Revision: V 13.0-0001 (for VA20)

V 13.5 (for VA26)

Physical type: Ethernet, 10/100 base T/Tx, RJ45

Network protocol: TCP/IP

DICOM Conformance Statement: September 2003, 1st Edition.

Image memory: 512 MB RAM (standard)

Hard disk: 20 GB

File Name Content Remarks

Protocol: Client: PR_hostname

0107H warning Indicate

0116H warning Indicate

B604H warning Indicate

N-event report RQ Disabled

Use Extention format: 0: No

ID

Use default annotation: 0: No

Change film: 272:0110H[F]

Use System Timeout: 1:Yes

Presentation LUT 0:No

Precode picking up: 0:No

Use Max/Min Density 1:Modality

Logging Client: 0:No

Reprint: 0:No

Magnify:

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 97 of 260

Page 98: CT00-000.814.30.02.02.pdf

98 Specific Camera Information - Fuji

Procedure SSM

Default Smoothing Type SHARP

Adjustment Param. SHARP

Edge Detection 1:on (dimmed)

Detection level 600 (dimmed)

Sharpness 130 (dimmed)

Char. Mode White (dimmed)

Adjustment param. MEDIUM

Edge Detection 1:On (dimmed)

Detection level 600 (dimmed)

Sharpness 80 (dimmed)

Char. Mode 0:White (dimmed)

Adjustment param. SMOOTH

Edge Detection 1:On (dimmed)

Detection level 800 (dimmed)

Sharpness 30 (dimmed)

Char. Mode 0:White (dimmed)

Default Requested Decimate Crop Behavior:

1:Decimate

LUT:

Procedure SAR

Default LUT 1

LUT 1

r table number 17

Max. density 300

Min. density 0

Contrast 100 (dimmed)

Number of tuning points 0

Dmax3.6 0:Standard

Output format:

Border density 300

File Name Content Remarks

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 98 of 260

Page 99: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 99

Error / Warning Messages (examples)

Trim density 0

Trim 0:Off

Number of Copies 1

Margin between images 20

Image layout 0:Concentration

Film size 4:14 IN X 17 IN

Medium Type (base color): 1:Blue

Film orientation 1:Portrait

Mirror 0:Off

Film Destination:

Upper Tray: 1:use

Annotation

Annotation Height: 1:1

Annotation System:System

Annotation 14:14

Upper Left

Upper Center

Upper Right

Lower Left

Lower Center

Lower Right

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Front cover open COVER OPEN COVER OPEN Camera cover, drawer or door open

Camera in utility mode

PRINTER OFFLINE PRINTER OFFLINE Camera offline

Camera in service mode (M-utility mode)

TCP/IP INIT ERROR TCP/IP INIT ERROR Connection to cam-era is not available

File Name Content Remarks

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 99 of 260

Page 100: CT00-000.814.30.02.02.pdf

100 Specific Camera Information - Fuji

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.5 0.15

Y 400.0 399.0 0.25

14" x 17"

Landscape

X 410.0 409.0 0.24

Y 320.0 320.5 0.16

14" x 14"

Portrait

X 330.0 330.5 0.15

Y 330.0 329.0 0.30

14" x 14"

Landscape

X No No No

Y No No No

11" x 14"

Portrait

X 230.0 230.0 0.00

Y 340.0 340.5 0.15

11" x 14"

Landscape

X 350.0 350.0 0.00

Y 230.0 230.0 0.00

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 280.0 280.5 0.18

10” x 12”

Landscape

X 290.0 290.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 190.0 190.0 0.00

Y 230.0 229.5 0.22

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 100 of 260

Page 101: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 101

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 101 of 260

Page 102: CT00-000.814.30.02.02.pdf

102 Specific Camera Information - Fuji

Fuji Drypix 4000 4.1

General information 0

Block diagram

Fig. 13:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film type Medium type

14" x 17"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

14" x 14"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

11" x 14"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

10" x 12"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

8" x 10"Portrait Yes DI-HL Blue-based

Landscape Yes DI-HL Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 102 of 260

Page 103: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 103

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE CISPR11 is Class A only (hospital use).

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 80 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRY4000 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec. file name)fujidrypix4000 No

DICOM Node dry4000 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 100.0 x 100.0 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 103 of 260

Page 104: CT00-000.814.30.02.02.pdf

104 Specific Camera Information - Fuji

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

3500 x 4170

14" x 14"

3500 x 3410

11" x 14"

2510 x 3500

10" x 12"

2460 x 2916

8" x 10"

1954 x 2410

In “Define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

4220 x 3400

14" x 14"

3500 x 3410

11" x 14"

3600 x 2460

10" x 12"

2972 x 2404

8" x 10"

2466 x 1898

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 104 of 260

Page 105: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 105

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Device type: DICOM Printer (SCP)

Device name: Drypix 4000

Manufacturer: FUJI

SW Revision: V 3.0 (for VA20)

V 3.5 (for VA26)

Physical type: Ethernet, (RJ45)

Network protocol: TCP/IP

DICOM Conformance Statement: 1st Edition, October 2004

Image memory: 512 MB RAM

Hard disk: 40 GB (25 GB for image storage)

File Name

Content / Remarks

Fuji printer PR_hostname

Protocol: Attribute List Error (0107H warning) 1:Indicate

Attribute Value Out of Range (0116H warning) 1:Indicate

Image size is larger than image box (B604H warning) 1:Indicate

N-event report RQ 0:Disabled

Use extension format 0:No

Use extension format --> ID (empty)

Use extension format--> Use Default Annotation 0:No

Change film 272:0110H[F]

Use System Timeout 1:Yes

Presentation LUT 0:No

Precede Picking up 1:Yes

Use Max/Min Density 1:Modality

Logging Client 0:No

Reprint 0:No

Change Base Color (Disable:110H[F]) 1:Enable

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 105 of 260

Page 106: CT00-000.814.30.02.02.pdf

106 Specific Camera Information - Fuji

Define Film Size of 11" x 14'' 1: 25.7 x 36.4 cm

Magnify: Procedure 0:A-VR

Default Smoothing Type 4:modality

Adjustment Parameter -> SHARP -> Edge Detection 1:On

Adjustment Parameter -> SHARP -> Detection level 600

Adjustment Parameter -> SHARP -> Sharpness 130

Adjustment Parameter -> SHARP -> Character Color 0:White

Adjustment Parameter -> MEDIUM -> Edge Detection 1:On

Adjustment Parameter -> MEDIUM -> Detection level 600

Adjustment Parameter -> MEDIUM -> Sharpness 80

Adjustment Parameter -> MEDIUM -> Character Color 0:White

Adjustment Parameter -> SMOOTH -> Edge Detection 1:On

Adjustment Parameter -> SMOOTH -> Detection level 800

Adjustment Parameter -> SMOOTH -> Sharpness 30

Adjustment Parameter -> SMOOTH -> Character Color 0:White

Default Requested Decimate / Crop Behavior 1:DECIMATE

Magnification Type 2:modality

Requested Image Size (0=None) 0

Requested Image Size (0=None) ' Default Requested Image Size

1:modality

LUT: Procedure 0:SAR

Default LUT 1

Default LUT -> Configuration Information (Default LUT#)

0:printer

Adjustment Parameter -> LUT1 -> r table number 17

Adjustment Parameter -> LUT1 -> Max. Density 300

Adjustment Parameter -> LUT1 -> Min. Density 20

Adjustment Parameter -> LUT1 -> 100

File Name

Content / Remarks

Fuji printer PR_hostname

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 106 of 260

Page 107: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 107

Adjustment Parameter -> LUT1 -> Number of tuning points

0

Dmax 3.6 0:Standard

Illumination 2000

Illumination -> Default Illumination 1:modality

Reflective Ambient Light 10

Reflective Ambient Light -> Default Ambient Light 1:modality

Output format:

Border density 280

Border density -> Default border density 1:modality

Polarity (Pos./Neg.) 1:normal (Pos.)

Trim density 0

Trim 0:Off

Trim Width 1

Number of Copies 1

Number of Copies -> Default Number of Copies 1:modality

Margin between images 20

Image layout 0:Concentration

Default film size 1:modality

Default Film Size -> Film Size 4: 14 IN X 17 IN

Default Medium Type 0:printer

Default Medium Type -> Medium Type (base color) 1: blue

Film orientation 3:modality

Mirror 0:OFF (Normal)

Processing Type 0:Standard

Print priority 1:HIGH

Resolution 0:Standard

Default film destination 0:printer

Processsor 1: use

File Name

Content / Remarks

Fuji printer PR_hostname

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 107 of 260

Page 108: CT00-000.814.30.02.02.pdf

108 Specific Camera Information - Fuji

Error / Warning Messages (examples)

Bin No 1 0: not use

Bin No 2 0: not use

Bin No 3 0: not used

Bin No 4 0: not used

Annota-tion

Annotation Height 5:modality

Annotation Font System:System

Annotation Size 14:14

Upper Left (empty)

Upper Center (empty)

Upper Right (empty)

Lower Left (empty)

Lower Center (empty)

Lower Right (empty)

Camera warning or error messages

Printer message Message displayed on SCU

Defined DCS (*1) Sent by SCP (*2)

Camera in utility mode

PRINTER OFFLINE PRINTER OFFLINE Camera offline

Camera in service mode (M-utility mode)

TCP/IP INIT ERROR

TCP/IP INIT ERROR

Connection to cam-era is not available

Front door open COVER OPEN COVER OPEN Camera cover, drawer or door open

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

File Name

Content / Remarks

Fuji printer PR_hostname

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 108 of 260

Page 109: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 109

Image Quality and Optical Density 0

No specific parameter available.

Real size scaling

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 332.0 0.61

Y 400.0 402.5 0.63

14" x 17"

Landscape

X 410.0 413.0 0.73

Y 330.0 332.5 0.61

14" x 14"

Portrait

X 330.0 332.5 0.76

Y 330.0 332.5 0.76

14" x 14"

Landscape

X 330.0 332.0 0.61

Y 330.0 332.5 0.76

11" x 14"

Portrait

X 230.0 232.0 0.87

Y 340.0 342.0 0.59

11" x 14"

Landscape

X 350.0 351.5 0.43

Y 230.0 232.0 0.87

10” x 12”

Portrait

X 230.0 232.0 0.87

Y 280.0 281.5 0.54

10” x 12”

Landscape

X 290.0 291.5 0.52

Y 230.0 232.0 0.87

8" x 10"

Portrait

X 190.0 191.0 0.53

Y 230.0 232.0 0.87

8" x 10"

Landscape

X 230.0 232.0 0.87

Y 170.0 171.0 0.59

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 109 of 260

Page 110: CT00-000.814.30.02.02.pdf

110 Specific Camera Information - Fuji

Fuji Drypix 3000 4.2

General information 0

Block diagram

Fig. 14:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 6

Format Release status Film Type Medium type

14" x 17"Portrait Yes DI-AT Blue-based

Landscape Yes DI-AT Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 110 of 260

Page 111: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 111

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The DI-AT film type requires a specific densitometer charac-teristic. Density measurements using the X-Rite 301 densito-meter are only possible if the FUJI/X-Rite colorcompensating filter kit CM500 is installed on the X-Rite 301and calibrated as described in the X-Rite letter.

NOTE Additional annotation and trim have been switched off.

NOTE The background command does not work properly. The back-ground is always white.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 150 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DRYPIX3000 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 111 of 260

Page 112: CT00-000.814.30.02.02.pdf

112 Specific Camera Information - Fuji

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Class

(Printer spec file name)fujidrypix3000 No

DICOM Node drypix3000 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 84.7 x 84.7 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4072 x 4891

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

4972 x 3993

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

HC Device /

General settingsDisplay Changes allowed

Device type: DICOM Printer (SCP)

Device name: Drypix 3000

Manufacturer: FUJI

SW Revision: A03 Main Unit / A04 LAN Board

Physical type: 100Base-Tx /10 Base T

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 112 of 260

Page 113: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 113

Printer settings

Network protocol: TCP/IP

DICOM Conformance Statement: 1st Edition March 31, 2001.

Image memory: Approx. 56 MB

Hard disk: None

File Name Content Remarks

None FUJI PRINTER Provided by FUJI

Function Properties

Protocol information Protocol = DICOM

Normal AE title = DRYPIX 3000

Role = SCP

Port no. = 104

Data transfer time-out = 300

Annotation Annotation size = 1:half

Association count 6

Logging DICOM 0 : None

Log Auto Save 0 : No

Default client 0 : No

Printer Model = DRYPIX

Manufacturer = FUJI PHOTO FILM CO, LTD

CT hostname None

Protocol: Attribute list error (0107H warning) = indicate

Attribute value out of range (0116H warning) = indicate

Image size larger than image box (B604H warning) = 1: Indicate

N-Event-Report-RQ = 0 disable

Change film size = 0: 0116H ()W

Use system time-out = 1: Yes

Use extension format Use default annotation = 1: No

Device type: DICOM Printer (SCP)

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 113 of 260

Page 114: CT00-000.814.30.02.02.pdf

114 Specific Camera Information - Fuji

Error / Warning Messages (examples)

Magnify: Procedure = 1:SSM

Default smoothing type = 1:SHARP

LUT: Procedure = 0: SAR

Default LUT = 1

r table no. = 17

Max. density = 300

Min. density = 0

Output format: Border density = 300

Polarity (Pos/Neg) = 1 : normal(Pos.)

Trim density = 0

Trim = 0: Off

Trim with = 1

Number of copies = 1

Margin between images = 20

Image layout = 0: Concentration

Film size = 1: 35x43 (14 in. x17 in.)

Medium type (base color) = 1 blue

Film orientation = 1: PORTRAIT

Mirror = 0: OFF(Normal)

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1)

Sent by SCP

Top cover open n/a UNKNOWN Unspecified problem with camera.

Remove tray n/a UNKNOWN Unspecified problem with camera.

Supply magazine empty n.a. SUPPLY EMPTY Camera out of film

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

File Name Content Remarks

None FUJI PRINTER Provided by FUJI

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 114 of 260

Page 115: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 115

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 400.0 398.0 0.50

14" x 17"

Landscape

X 410.0 408.5 0.37

Y 320.0 320.0 0.00

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 115 of 260

Page 116: CT00-000.814.30.02.02.pdf

116 Specific Camera Information - Fuji

Fuji Drypix 2000 4.3

General Information 0

Block Diagram

Fig. 15:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 11

Format Release status Film type Medium type

14" x 17"Portrait Yes DI-HT Blue-based

Landscape Yes DI-HT Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait Yes DI-HT Blue-based

Landscape Yes DI-HT Blue-based

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait Yes DI-HT Blue-based

Landscape Yes DI-HT Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 116 of 260

Page 117: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 117

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Test of line structure and artifacts not passed. Small horizon-tal and vertical line structure visible in technical images andfor all film sizes.

NOTE Electro Magnetic Compatibility(EN55011, CISPR 11) classifi-cation is not available. An installation is suggested only forIndustrial Areas (Hospital).

Known Effects

NOTE The measured values of the X-Rite 301 Densitometer must betranslated via the Fuji Conversion Table (Operating Manual).

NOTE Sporadic transportation problems with film size 8x10 and11x14. Install magazine at the top side to prevent this prob-lem.

SCU / Modality Software Configuration 0

Time 118 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DP2000 Yes

Port number 104 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 117 of 260

Page 118: CT00-000.814.30.02.02.pdf

118 Specific Camera Information - Fuji

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec. file name)fujidrypix2000 No

DICOM Node DRY2000 Yes

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 84.7 x 84.7 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

4107 x 4858

11" x 14"

2938 x 4094

8" x 10"

2280 x 2812

In “Define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17"

4932 x 4016

11" x 14"

4188 x 2872

8" x 10"

2876 x 2215

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 118 of 260

Page 119: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 119

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer Settings

Device type: DICOM Printer (SCP)

Device name: Drypix 2000

Manufacturer: FUJI

SW Revision: V2.0

Physical type: Ethernet 10/100 Base T/Tx

Network protocol: TCP/IP

DICOM conformance statement: 1st Edition, June 2006

Image memory: 256 MB RAM

Hard disk: None

Fuji printer Content

Protocol: Attribute List Error (0107H warning) Indicate

N-event report RQ Disabled

Attribute Value Out of Range (0116H warning) Indicate

Image size is larger than image box (B604H warning) Indicate

Change film size 272:0110H[F]

Use System Timeout Yes

Presentation LUT No

Precede picking up No

Use max/min density Modality

Use extension format --> ID SM

Use extension format--> Use Default Annotation No

Change base color (disable:110H[F]) Enable

Define Film Size of 11" x 14'' 25.7 x 36.4 cm

AE Title PR_hostname

Magnify:

Default Smoothing Type sharp

Default requested decimate / crop behavior decimate

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 119 of 260

Page 120: CT00-000.814.30.02.02.pdf

120 Specific Camera Information - Fuji

Magnification Type cubic

Requested image size (0=None) 0

Default Magnification Type Modality

Default Smoothing Type Modality

Default Requested Image Size (0=None) Modality

Default requested decimate / crop behavior Modality

AE Title PR_hostname

Procedure AV-R

SHARP -> Edge Detection On

SHARP -> Detection level 600

SHARP -> Sharpness 130

SHARP -> Character Color White

SMOOTH -> Edge Detection On

SMOOTH -> Detection level 800

SMOOTH -> Sharpness 30

SMOOTH -> Character Color White

MEDIUM -> Edge Detection On

MEDIUM -> Detection level 600

MEDIUM -> Sharpness 80

MEDIUM -> Character Color White

LUT Com-mon:

Default LUT 1

Illumination 2000

Ambient light 10

Default LUT -> Configuration Information (Default LUT#) printer

Default Illumination Modality

Default Ambient Light Modality

AE Title PR_hostname

Procedure SAR

LUT1 -> Max. Density 280

Fuji printer Content

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 120 of 260

Page 121: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 121

LUT1 -> Min. Density 18

AE Title PR_hostname

LUT1 -> r table number 17

LUT1 -> Contrast 100

Number of tuning points 0

P1_Density 120

P2_Density 120

P3_Density 120

P4_Density 120

P5_Density 120

P1_Shift 0

P2_Shift 0

P3_Shift 0

P4_Shift 0

P5_Shift 0

P1_Contrast 100

P2_Contrast 100

P3_Contrast 100

P4_Contrast 100

P5_Contrast 100

Output format:

AE Title PR_hostname

Film Size 14INx17IN

Medium type blue

Film orientation portrait

Border density 280

Polarity (Pos./Neg.) normal(Pos.)

Trim Off

Trim Width 1

Trim density 0

Fuji printer Content

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 121 of 260

Page 122: CT00-000.814.30.02.02.pdf

122 Specific Camera Information - Fuji

Number of Copies 1

Print priority high

Default Number of Copies Modality

Default Print Priority Printer

Default Medium Type Printer

Default Film Orientation Modality

Default Film Size Modality

Default Border Density Printer

Default Trim printer

Default Polarity Modality

AE Title PR_hostname

Mirror OFF(Normal)

Margin between images 20

Image layout Concentration

Processing Type 2

Image Crop UPDWN Area (8x10) for CR Image both

Image Crop Side Area (8x10) for CR Image both

Image Crop UPDWN Area (10x14) for CR Image both

Image Crop Side Area (10x14) for CR Image both

Image Crop UPDWN Area (14x17) for CR Image both

Image Crop UPDWN Area (10x12) for CR Image both

Image Crop Side Area (10x12) for CR Image both

Annotation

Upper left empty

Upper center empty

Upper right empty

Lower left empty

Lower center empty

Lower right empty

Fuji printer Content

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 122 of 260

Page 123: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 123

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available.

Real size scaling

Camera warning or error messages

Printer message Message displayed on SCU

Defined DCS (*1) Sent by SCP (*2)

Top cover open COVER OPEN COVER OPEN Camera cover, drawer or door open

Remove Tray COVER OPEN COVER OPEN Camera cover, drawer or door open

Supply empty SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Film jam FILM JAM FILM JAM Film jam

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 400.0 398.0 0.50

14" x 17"

Landscape

X 410.0 407.5 0.61

Y 330.0 330.0 0.00

14" x 14"

Portrait

X n/a n/a n/a

Y n/a n/a n/a

14" x 14"

Landscape

X n/a n/a n/a

Y n/a n/a n/a

11" x 14"

Portrait

X 230.0 229.5 0.22

Y 340.0 340.0 0.00

11" x 14"

Landscape

X 350.0 350.0 0.00

Y 230.0 229.5 0.22

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 123 of 260

Page 124: CT00-000.814.30.02.02.pdf

124 Specific Camera Information - Fuji

10” x 12”

Portrait

X n/a n/a n/a

Y n/a n/a n/a

10” x 12”

Landscape

X n/a n/a n/a

Y n/a n/a n/a

8" x 10"

Portrait

X 190.0 189.5 0.26

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 124 of 260

Page 125: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 125

Fuji FM DPL/FN-PS551 4.4

General information 0

Block diagram

Fig. 16:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 6

Format Release status Film Type Medium type

14" x 17"Portrait Yes DI-AL Blue-based

Landscape Yes DI-AL Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait Yes DI-AL Blue-based

Landscape Yes DI-AL Blue-based

8" x 10"Portrait Yes DI-AL Blue-based

Landscape Yes DI-AL Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 125 of 260

Page 126: CT00-000.814.30.02.02.pdf

126 Specific Camera Information - Fuji

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE 2 cameras (each with a separate channel) can be connectedto PS551. Only one camera connection to FN-PS551 wasused for the qualification.

NOTE The European Standard EN55011 is supported only withClass A (Hospital use).

Known Effects

NOTE In a multi-modality environment one additional hardwarechannel must be installed at FM-DPL and a second printserver PS551 has to be connected for Mammography Modal-ities.

NOTE When the Laser Imager connected to the print server is in theoffline status being disabled to receive image data, FAILUREwill be returned.

SCU / Modality Software Configuration 0

Time 158 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title FMDPL Yes

Port number 104 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 126 of 260

Page 127: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 127

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

HC Device /

General settingsDisplay Changes allowed

Type DICOMprinter No

Class

(Printer spec file name)fujifmdpl No

DICOM Node FMDPL Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 100 x 100 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

3500 x 4170

11" x 14"

2538 x 3522

8" x 10"

1952 x 2410

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

4240 x 3442

11" x 14"

3600 x 2460

8" x 10"

2464 x 1898

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 127 of 260

Page 128: CT00-000.814.30.02.02.pdf

128 Specific Camera Information - Fuji

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Device type: DICOM print server/printer

Device name: FN-PS551

Manufacturer: FUJI Photo Film

SW Revision: A 07

Physical Interface: 100/10 base T Ethernet, UTP/STP

Network protocol: TCP/IP

DICOM Conformance Statement: FN-PS551 May 2001, 4th edition

Image memory: 261.56 MB

Hard disk: 3 Gbyte

Device type: DICOM camera

Device name: FM - DPL

Manufacturer: FUJI Photo Film

SW Revision: W01-02 (Main Unit) Panel: Unit W00-00

Physical Interface: SCSI Interface

Network protocol: None

DICOM Conformance Statement: None

Image memory: 96 MB

Hard disk: None

Film size 14 IN X 17 IN

Image type: Select: Standard scan

Gradation type: Select: 3.0

File Name Content Remarks

None FUJI service utility setting Provided by FUJI

Function Properties

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 128 of 260

Page 129: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 129

Protocols: DICOM Properties Network port no. = 104

Data transfer time-out = 600 sec.

Number of associations = 6

Association establishment restriction by the memory = enable

Print server application entity = one to each printer

Data receive if the printer is offline = enable

Printer FMDPL Properties

Name: AE title of DICOM = FMDPL

Fine-printing = off

Interface: Interface = E-i/f

Pixel clock = 1.0 microsec/pixel

Version = type 1

Film size = 14 x 17 (dimmed)

Film medium = blue (dimmed)

Model: Model = FN-PS551

Manufacturer = FUJI PHOTO FILM CO, LTD

Clients: DICOM: CT hostname Properties

Name: Hostname = CT hostname

IP address = (from system administrator)

File Name Content Remarks

None FUJI service utility setting Provided by FUJI

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 129 of 260

Page 130: CT00-000.814.30.02.02.pdf

130 Specific Camera Information - Fuji

Protocol: Protocol = DICOM

Application Entity Title = PR_hostname

Attribute list error (0107H warning) = indi-cate

Attribute value out of range (0116H warn-ing) = indicate

Image size larger than image box (B604H warning) = indicate

N-event report RQ = disabled

Use presentation LUT = No

Data transfer time-out = enabled

Use extension format = No

Magnify: Procedure = SSM

Default smoothing type = SHARP

LUT: Procedure = SAR

Default LUT = 1 (linear)

Adj. parameter LUT = 1

r table no. = 17

Max. density = 300

Min. density = 0

Number of tuning points = 0

Printer: fixed, FMDPL

File Name Content Remarks

None FUJI service utility setting Provided by FUJI

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 130 of 260

Page 131: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 131

Output format 1: Border density = 300

Polarity = normal(Pos.)

Trim density = 0

Trim = Off

Trim width = 1 pixel

Number of copies = 1

Film size = 14x17

Medium type = blue

Film orientation = portrait

Mirror = OFF(normal)

Overlay threshold density = 120

Output format 2: Margin between images = 20 pixel

Image layout = concentration

Annotation: Print server Upper left = (nothing selected)

Upper center = (nothing selected)

Upper right = (nothing selected)

Upper left = (nothing selected)

Lower center = (nothing selected)

Lower right = (nothing selected)

Logging: Network = (Not selected)

Printer = (Not selected)

DICOM = normal

Output image to file = off

Precedence = Input

Sorting: Sorting conditions = (nothing selected)

2 cameras (each with a separate channel) can be connected to PS551. Only one printer connection is used for the test.

File Name Content Remarks

None FUJI service utility setting Provided by FUJI

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 131 of 260

Page 132: CT00-000.814.30.02.02.pdf

132 Specific Camera Information - Fuji

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Camera in utility mode

PRINTER OFFLINE PRINTER OFFLINE Camera offline

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Camera is switched off

n.a. UNKNOWN Unspecified problem with camera

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 340.0 339.0 0.29

Y 400.0 400.5 0.12

14" x 17"

Landscape

X 410.0 410.5 0.12

Y 330.0 329.0 0.30

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X 240.0 240.5 0.21

Y 340.0 338.5 0.44

11" x 14"

Landscape

X 350.0 348.5 0.43

Y 230.0 230.2 0.09

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 132 of 260

Page 133: CT00-000.814.30.02.02.pdf

Specific Camera Information - Fuji 133

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X 190.0 189.0 0.53

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 230.5 0.22

Y 180.0 179.0 0.56

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 133 of 260

Page 134: CT00-000.814.30.02.02.pdf

134 Specific Camera Information - Kodak

5- 5Specific Camera Information - KodakKodak DryView 6800 0

General Information 0

Block Diagram

Fig. 17:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Performance

Use medical image 4 of the "Reference Images" patient, maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point-to-point) was used.

Application Status Film type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

11" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

10" x 12"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

8" x 10"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

Others n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 134 of 260

Page 135: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 135

Restrictions

NOTE For restrictions not related to HCC, refer to (Restrictions andknown effects / p. 8).

NOTE For SW version V 1.00 only. (internal built-in MIM Core swv2.1.b12) The LUT file CTLUTdry has to be used to reach therequired image quality resolution in white areas.

NOTE No error message is sent if film size of job to be printed is notinserted in camera.

NOTE Min. Density is always fog on film.

NOTE A Film Jam error will not be sent to modality as specified.

NOTE The serrated border line and some pixels of the diagnosticarea are cut because of rounded corners.

NOTE The standard EN55011 (EMC) is supported with Class A (Hos-pital use) only.

Known effects

n.a.

SCU / modality software configuration 0

Time 65 sec

Application Entity

PropertiesDisplay Changes allowed

AE Title DV68001

1. Do not use an AE title such as “NER_...”

Yes

Port number 5040 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 135 of 260

Page 136: CT00-000.814.30.02.02.pdf

136 Specific Camera Information - Kodak

HC Device /

General settings Display Changes allowed

Type DICOM printer No

Class

(Printer spec file name)kodak6800_smallB No

DICOM node DV6800 Yes

Filming properties Display Changes allowed

Hold Printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.154 x 78.154 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

4488 x 5464

14" x 14"

4448 x 4448

11" x 14"

3514 x 4488

10" x 12"

3184 x 3828

08" x 10"

2528 x 3184

In “Define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17”

5464 x 4488

14" x 14"

4488 x 4488

11" x 14"

4488 x 3514

10" x 12"

3828 x 3184

08" x 10"

3184 x 2528

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale, 8-bit No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 136 of 260

Page 137: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 137

NOTE Do not modify or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

SCP / hardcopy camera configuration (provided by OEM) 0

Identification

Printer Settings

The following table describes the level (L = level) in the Kodak configuration tool “Kodak Weblink”.

Background Black Yes

Transformation Replicate No

Filming properties Display Changes allowed

Manufacturer / device type: DICOM Printer (SCP) (internal print server)

Device name: DryView 6800

Manufacturer: Kodak

SW Revision: V 1.00 (internal built-in MIM Core sw v2.1.b12);(for VA20)

V 2.00 (internal built-in MIM Core sw v2.1.b14) (for VA26)

Physical interface: Ethernet 10/100/1000 base T/Tx,

Network protocol: TCP/IP

DICOM conformance statement: Rev. 5.0, September 26, 2007

Image memory: 786 MB RAM

Hard disk: 40 GB

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 137 of 260

Page 138: CT00-000.814.30.02.02.pdf

138 Specific Camera Information - Kodak

L1 L2 L3 L4 Sub-Level of L4

Configuration

DICOM SCP

SCP Configuration

Image Optimization

Services Provided to SCU

- Negotiate PLUT / NO

- Negotiate Grayscale Print / YES

- Negotiate Color Print / NO

Image quality

- Tonescaling Method / Perception LUT - NO

- TFT Set Name / wrkstn2a.w87 - Always Apply

- Polarity / NORMAL - NO

- Interpolation / REPLICATE - NO

- Density Min / 0.20 - NO

- Density Max / 2.80 - NO

- Smoothing / 0 - NO

- Perception LUT Contrast / 3 - Always Apply

- Curve Shape Contrast / 0 - NO

- Curve Shape Value / 429 - NO

- Wet Pivot Density / 1.2 - NO

- Illumination / 2000 - NO

- Reflected Ambient Light / 10 - NO

Film and Image Options

Film and Page Options

- Film Size / 14”x17” – NO

- Film Base / BLUE FILM – NO

- Border Density / 3.10 – NO

- Sorter Bin / BIN 1 – NO

- Film Orientation / PORTRAIT – NO

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 138 of 260

Page 139: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 139

- Trim / NO – NO

Image Resizing

- Method / MINIFY-Scale – NO

- Text Box Location / NONE – Always Apply

Other Options

Status Reporting

- Send Warnings / YES - Always Apply

- Send N-Events to SCU / NO - Always Apply

Annotation Elements

- Print Date / NO - Always Apply

- Print Hospital Name / NO - Always Apply

- Print Modality ID / NO - Always Apply

- Print Time / NO - Always Apply

- Print Page Number / NO - Always Apply

- Common Text Override / NO - Always Apply

- Film Session Label / “empty” /NO

Printer and Other Parameters

- Printer Priority / NORMAL - NO

- Body Part / DEFAULT - NO

SCP Config AutoSetup

Modality Identification

- Source ID / 3.2

- Modality ID / xxxxxxx6800xxxxxx

- Modality Type / Other - NO

Modality Defaults

L1 L2 L3 L4 Sub-Level of L4

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 139 of 260

Page 140: CT00-000.814.30.02.02.pdf

140 Specific Camera Information - Kodak

Error and warning messages

Image Quality and Optical Density 0

No specific parameter available.

- Manufacturer / "empty"

- Type / "empty"

- Model / "empty"

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Camera cover, drawer or door open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Unlock magazine PRINTER OFFLINE

PRINTER OFFLINE

Camera offline

Send film to empty 14x17 supply maga-zine

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

Camera info: EMPTY 14X17 BLUE

Camera is warming up

PROC INIT PROC INIT Processor initializing

Camera in "power save" or "sleep mode"

PRINTER OFFLINE

PRINTER OFFLINE

Camera offline

Send job for a film size not inserted in camera

n.a. n.a. “nothing”

Camera is in calibra-tion

PRINTER OFFLINE

PRINTER OFFLINE

Camera offline

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

L1 L2 L3 L4 Sub-Level of L4

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 140 of 260

Page 141: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 141

Real Size Scaling

Film size / Direction Desired (mm) Measured (mm)

Difference obtained from desired value

in %

14" x 17"

Portrait

X 340.0 338.5 0.44

Y 400.0 400.0 0.00

14" x 17"

Landscape

X 400.0 400.2 0.05

Y 340.0 338.5 0.44

14" x 14"

Portrait

X 300.0 298.5 0.50

Y 300.0 300.0 0.00

14" x 14"

Landscape

X 300.0 300.0 0.00

Y 300.0 298.5 0.50

11" x 14"

Portrait

X 260.0 260.5 0.19

Y 340.0 338.5 0.44

11" x 14"

Landscape

X 340.0 338.5 0.44

Y 260.0 260.5 0.19

10" x 12"

Portrait

X 240.0 238.5 0.62

Y 280.0 280.2 0.07

10" x 12"

Landscape

X 280.0 280.2 0.07

Y 240.0 238.5 0.62

8" x 10"

Portrait

X 160.0 160.0 0.00

Y 240.0 238.5 0.62

8" x 10"

Landscape

X 240.0 238.5 0.62

Y 160.0 160.0 0.00

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 141 of 260

Page 142: CT00-000.814.30.02.02.pdf

142 Specific Camera Information - Kodak

Kodak DryView 5800 5.1

General information 0

Block Diagram

Fig. 18:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Performance

Use medical image 4 of the "Reference Images" patient, maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Application Status Film type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

11" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

10" x 12"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

8" x 10"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

Others n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 142 of 260

Page 143: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 143

Restrictions

NOTE For restrictions not related to HCC, refer to (Restrictions andknown effects / p. 8).

NOTE Min. Density is always fog on film.

NOTE Some dust artifacts (less then five) visible on film.

NOTE The EN55011 standard is class A (Hospital use) only.

NOTE No error message is sent if film size of job to be printed is notinserted in camera.

NOTE If the film supply is empty, no messages are displayed in“Film Task Status”

Known effects

n.a.

SCU / modality software configuration 0

Time 92 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV58001

1. Do not use an AE title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settings Display Changes allowed

Type DICOM printer No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 143 of 260

Page 144: CT00-000.814.30.02.02.pdf

144 Specific Camera Information - Kodak

Class

(Printer spec file name)kodak5800 No

DICOM node DV5800 Yes

Filming properties Display Changes allowed

Hold Printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.154 x 78.154 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

4412 x 5384

14" x 14"

4412 x 4412

11" x 14"

3436 x 4412

10" x 12"

3104 x 3752

8" x 10"

2452 x 3104

In “Define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17”

5384 x 4412

14" x 14"

4412 x 4412

11" x 14"

4412 x 3436

10" x 12"

3752 x 3104

8" x 10"

3104 x 2452

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale, 8-bit No

Background Black Yes

Transformation No magnification No

HC Device /

General settings Display Changes allowed

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 144 of 260

Page 145: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 145

NOTE Do not modify or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

SCP / hardcopy camera configuration (provided by OEM) 0

Identification

Printer Settings

The following table describes the Kodak configuration tool “Kodak Weblink”.

Manufacturer / device type: DICOM Printer (SCP) (internal print server)

Device name: DryView 5800

Manufacturer: Carestream (Kodak)

SW Revision: V1.01.b01

(internal built-in MIM Core SW V3.0.b11)

Physical interface: Ethernet 10/100 base T/Tx, RJ45 connector

Network protocol: TCP/IP

DICOM conformance statement: Rev. 2.0, December 12, 2007

Image memory: 512 MB RAM

Hard disk: 80 GB

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 145 of 260

Page 146: CT00-000.814.30.02.02.pdf

146 Specific Camera Information - Kodak

Name (on Configura-tion)

Setting

System, Network

Host Name DV5800DRE

IPAddr accordingly

Subnet Mask accordingly

Default Gateway accordingly

Connection Name LAN

DICOM SCP, Communications

Dicom Source Communications

Port 5040

Association Limit 12

DICOM SCP, Destination

Identification AETiltle=DV5800

Media:

Media Type avail./PageExtent(x,y)/ImageExtent(x,y)/ImageOrigin(x,y)

8x10, blue&clear selected 2590x3264, 2452x3107, 74, 69

10x12, blue&clear selected 3264x3890, 3107x3752, 69, 64

11x14, blue&clear selected 3575x4576, 3437x4412, 74, 69

14x14, blue&clear selected 4576x4550, 4412x4412, 69, 64

14x17, blue&clear selected 4576x5525, 4412x5387, 69, 64

Page Image Options:

Min Vert Sep=2, Min Horiz Sep=2

Max Band Page=10, Max Img Band=10

Pixel Pitch=12.795, Border DMax=3.10

Image Dmin=0.00, Image DMax=3.10

Lines Image Annot=1, Lines Page Annot=2

Image Annot Height=52, Page Annot Height=52

Max Magnif.Factor=20, Max Minif.Factor=0.1

Trim Width=2, Accept Annotations=Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 146 of 260

Page 147: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 147

Accept Color Image=Yes, Accept BW Image=Yes

Mix Aspect Ratio Page=Y, Mix Aspect Ratio Row=Y

Mix Image Size Std Fmt=Y, Mix Image Size Row=Y

Mix Req.Size Page=Y, Minify Image To Fit=Y

Scale Image To Fit=Y, Crop Image To Fit=Y

Req.Size OK=Y, Configuration Notes=empty

DICOM SCP, select specific modality, Modify:

SCP Configuration, Image Optimization

Services Provided to SCU

Negotiate Presentation LUT

No

Negotiate Color Print No

Image quality

Smoothing Type 0 - Always Apply

Interpolation NONE - No

TFT SeT Wrkstn2a.W87 -Always Apply

Perception LUT Con-trast

1 - Always Apply

Image Polarity Normal - No

Image DMin 0.00 - No

Image DMax 3.10 - No

Illumination 2000 - No

Reflected Ambient Light 10 - No

SCP Configuration, Film and Image Options

Film and Page Options

Film size 14IN x 17IN - No

Film Base Blue Film - No

Film orientation Portrait - No)

Name (on Configura-tion)

Setting

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 147 of 260

Page 148: CT00-000.814.30.02.02.pdf

148 Specific Camera Information - Kodak

Border density 3.10 - No

Trim No - No

Image Resizing

Method Minify/Scale - No

SCP Configuration, Other Options

Status Reporting

Report Warnings YES - Always Apply

Send N-Events to SCU No Always Apply

Film Session Elements

Capture Date No

Capture Time No

Hosp.Name No

Study Page Number No

Modality ID No

Film Session Label Override

No

Printer and Other Parameters

Printer Priority Normal - No

SCP Configuration, Modality Preferences

Modality Identification

Modality ID PR_ Hostname

Modality Type Other - No

Modality Defaults

Manufacturer “empty”

Type “empty”

Model “empty”

Name (on Configura-tion)

Setting

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 148 of 260

Page 149: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 149

Error and warning messages

Image Quality and Optical Density 0

No specific parameter available.

Real Size Scaling

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Camera is warming up PROC INIT PROC INIT Processor initializing

Film 14x17 supply empty

EMPTY 14x17 BLUE

NORMAL Camera is ready

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Send job for a film size not inserted in camera

n/a n/a Camera is ready

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Difference obtained from desired value

in %

14" x 17"

Portrait

X 320.0 320.0 0.00

Y 410.0 410.0 0.00

14" x 17"

Landscape

X 410.0 410.0 0.00

Y 330.0 329.5 0.15

14" x 14"

Portrait

X 330.0 330.0 0.00

Y 330.0 330.5 0.15

14" x 14"

Landscape

X 330.0 330.5 0.15

Y 330.0 329.5 0.15

11" x 14"

Portrait

X 260.0 260.0 0.00

Y 330.0 330.0 0.00

11" x 14"

Landscape

X 320.0 320.0 0.00

Y 260.0 260.5 0.19

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 149 of 260

Page 150: CT00-000.814.30.02.02.pdf

150 Specific Camera Information - Kodak

10" x 12"

Portrait

X 230.0 230.0 0.00

Y 280.0 280.5 0.18

10" x 12"

Landscape

X 280.0 280.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 180.0 180.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 230.0 229.5 0.22

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Difference obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 150 of 260

Page 151: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 151

Kodak DryView 8900 (small border) 5.2

General information 0

Block diagram

Fig. 19:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Format Release status Film Type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

11" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

10" x 12"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

8" x 10"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

Other N/A N/A N/A

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 151 of 260

Page 152: CT00-000.814.30.02.02.pdf

152 Specific Camera Information - Kodak

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.

NOTE The round edge of the film sheet cuts the diagnostic area.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 71 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV89001

1. Please do not use an AE Title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)kodak8900_smallB No

DICOM Node DV8900 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 152 of 260

Page 153: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 153

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.154 x 78.154 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4488 x 5464

14" x 14"

4488 x 4488

11" x 14"

3514 x 4488

10" x 12"

3184 x 3828

8" x 10"

2528 x 3184

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5464 x 4488

14" x 14"

4488 x 4488

11" x 14"

4488 x 3514

10" x 12"

3828 x 3184

8" x 10"

3184 x 2528

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation Replicate No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 153 of 260

Page 154: CT00-000.814.30.02.02.pdf

154 Specific Camera Information - Kodak

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

Manufacturer / device type: DICOM Print Server (SCP)

Device name: DryView 8900

Manufacturer: Kodak

SW Revision: V 3.17, MIM 5.6.b14 with WIN 2000

Physical type: Ethernet 10/100 /1000 base T/Tx, RJ45

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. C, January 28, 2004.

Image memory: 786 MB RAM

Hard disk: 40 GB

Content

8900 setup: (Provided by Kodak)

under Configuration, System, Network:Host Name=DV8900, IPAddr=accordinglySubnet Mask=accordingly, Gateway Addr=accordinglyunder Configuration, DICOM SCP, Communications:Port=5040, Association Limit=12, Source Name=DICOMunder Configuration, Destination:General Remarks: Log. Name=8900Media types: Media=35x43cm, 9100x11037, 10781, anyMedia types: Media=35x35 cm, 9100x9087, 8831, AnyMedia types: Media=11"x14", 7137x9100, 8844, AnyMedia types: Media=10"x12", 6490x7767, 7511, AnyMedia types: Media=8"x10", 5167x6490, 6234, AnyFormatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=OnFormatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=OnFormatting: DICOM Annotation Box=On, Req. Image Size=OnFormatting: Req. Size Scale=On, Req. Size Mix Page=OnFormatting: Req. Size Mix Row=On, Trim=On,Formatting: True Landscape=On, Rotation=OnFormatting: Minify/Crop=On, Tone Scaling on Image Basis=OnFormatting: Max Img Band=20, Max Band Page=20, Max Img Cols=0Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4Formatting: Min Vert Sep=4, Trim Width Enabled=4Formatting: Trim Width Disabled=0, Maximum DMAX=3.30

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 154 of 260

Page 155: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 155

Error / Warning Messages (examples)

Content

8900 setup: (Provided by Kodak)

under Configuration, DICOM SCP, SCP Services,Negotiate Presentation LUT=NO,TFT/ULUT= Wrkstn2a.w87 (Always=Y)Contrast= 3 (Always=Y),Smoothing type=0 (Always=Y),Polarity= Normal, (Always=N)Image Dmin=0.00 (Always=N),Image Dmax=3.00 (Always=N)NeventReports=No (Always=Y), Warnings=YesPrint Priority=Normal (Always=N),Border=3.99 (Always=N)2mmBorder=yesFilm Clip Border = NoFilm Base=Blue Film (Always=N),Film Size=14INx17IN (Always=N)Sorter Bin=Bin1 (Always=N),Include print date in annotation=NoInclude print time in annotation=NoInclude page number in annotation=NoInclude hospital name in annotation=NoInclude modality ID in annotation=NoModality ID=PR_ xxx.xxx.xxxAnnotation Override=No,Density Patch = NoImage Resizing=Minify/Scale (Always=N),Text Box Location=None, Source ID=DI016

Action that results in camera error messages

Printer messagesMessages displayed

on SCU (*2)Defined DCS (*1) Sent by SCP

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Camera is warming up

PROC INIT PROC INIT Processor initializing

Supply magazine empty

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

Camera info: EMPTY 14X17 BLUE

Magazine with damaged chip inserted

BAD SUPPLY MGZ BAD SUPPLY MGZ Problem with supply magazine

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 155 of 260

Page 156: CT00-000.814.30.02.02.pdf

156 Specific Camera Information - Kodak

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Action that results in camera error messages

Printer messages Messages displayed on SCU (*2)Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 340.0 340.0 0.00

Y 410.0 411.0 0.24

14" x 17"

Landscape

X 400.0 401.0 0.25

Y 330.0 330.0 0.00

14" x 14"

Portrait

X 340.0 339.5 0.15

Y 330.0 330.0 0.00

14" x 14"

Landscape

X 340.0 340.0 0.00

Y 330.0 330.0 0.00

11" x 14"

Portrait

X 270.0 270.0 0.00

Y 330.0 330.0 0.00

11" x 14"

Landscape

X 340.0 340.0 0.00

Y 250.0 250.0 0.00

10” x 12”

Portrait

X 240.0 240.0 0.00

Y 280.0 280.0 0.00

10” x 12”

Landscape

X 290.0 290.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Portrait

X 190.0 190.0 0.00

Y 230.0 230.0 0.00

8" x 10"

Landscape

X 240.0 240.0 0.00

Y 180.0 180.0 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 156 of 260

Page 157: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 157

Kodak DryView 8700-Plus / PACS Link MIM 200 5.3

General information 0

Block diagram

Fig. 20:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 13

Format Release status Film Type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 157 of 260

Page 158: CT00-000.814.30.02.02.pdf

158 Specific Camera Information - Kodak

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Camera does not support minimum density command.

NOTE The background command does not work properly. Cameradoesn't support background white.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 120 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV87001

1. Please do not use an AE title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)imation8700 No

DICOM Node DV8700 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 158 of 260

Page 159: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 159

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.0 x 78.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4096 x 5220

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5220 x 4096

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Device type: DICOM Print Server (SCP)

Device name: PACS Link MIM 200

Manufacturer: Kodak

SW Revision: V 6.1.1

Physical Interface: Ethernet (RJ45)

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. J., June 20, 2005

Image memory: 256 MB

Hard disk: 4 GB

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 159 of 260

Page 160: CT00-000.814.30.02.02.pdf

160 Specific Camera Information - Kodak

Printer settings

Device type: Printer

Device name: DryView 8700

Manufacturer: Kodak

SW Revision: 1.34

Physical Interface: Fiber optic

Network protocol: None

DICOM Conformance Statement: None

Image memory: 32 MB

Hard disk: n.a.

Content: Configuration

MIM 200 setup

Under Configure, Network:

Node Name=MIM, IPAddr=accordingly

Subnet Mask=accordingly, Gateway Addr=accordingly

under Configure, Input, Network, Add:

Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

under Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 160 of 260

Page 161: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 161

General: Log.Name=8700General: Database Name=8700, Model=8700, Memory=32Media info.:Media=35x43cm, 4096x5223, 5133, blueOutput parameters.:Cmd=SuperSet, Baud=19200, Data=8,Stop=1, Parity=NoneOutput parameters.:Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=OnFormat types: Custom=On, Slide=On, Row Symmetric=OnFormat types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1M952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7

Under Configure, SCP Services, select the specific modality, modify:

Content: Configuration

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 161 of 260

Page 162: CT00-000.814.30.02.02.pdf

162 Specific Camera Information - Kodak

Error / Warning Messages (examples)

Image resizingResizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT= Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=YInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic Color=N

8700 Plus setup

under COMM Menu: 969 default

under IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14

under EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Time-out=60

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Magazine empty SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Open Top Cover COVER OPEN COVER OPEN Camera cover, drawer or door open

Camera switched off PRINTER DOWN PRINTER DOWN Camera down

No supply magazine inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

Camera is warming up

PRINTER BUSY PRINTER BUSY Camera initializing

Film jam in camera FILM JAM FILM JAM Film jam

Content: Configuration

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 162 of 260

Page 163: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 163

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 310.0 310.5 0.16

Y 390.0 390.0 0.00

14" x 17"

Landscape

X 390.0 390.5 0.13

Y 310.0 310.5 0.16

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 163 of 260

Page 164: CT00-000.814.30.02.02.pdf

164 Specific Camera Information - Kodak

Kodak DryView 8700-Dual / PACS Link MIM 9410 5.4

General information 0

Block diagram

Fig. 21:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 10

Format Release status Film Type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 164 of 260

Page 165: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 165

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Camera does not support minimum density command.

NOTE The background command does not work properly. Cameradoesn't support background white.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 128 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title 9410 Yes

Port number 1024 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)imation8700 No

DICOM Node DV8700 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 165 of 260

Page 166: CT00-000.814.30.02.02.pdf

166 Specific Camera Information - Kodak

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Pixel Size [1/1000 mm] 78.0 x 78.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4096 x 5220

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5220 x 4096

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Manufacturer / device type: DICOM Print Server (SCP) Printer

Device name: PACS Link 9410 DryView 8700Dual

Manufacturer: Kodak Kodak

SW Revision: V 1.6 1.34

Physical type: 100/10 base T Ethernet, UTP/STP Fiber Optic

Network protocol: TCP/IP n.a.

DICOM Conformance State-ment:

Version 2.0, June 5, 2000 n.a.

Image memory: 128 MB 32 MB

Hard disk: 4.1 GB n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 166 of 260

Page 167: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 167

Printer settings

Error / Warning Messages (examples)

Content: Configuration

MIM 9410 setup

under Device Setup, LOCAL PRINTERS::

Printer=8700, Model=8700 Dual, Output=Fiber,

Baud=19200, Parity=None, Data=8, Stop=1, Command Set=Superset

Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz

Film Mismatch=Print Always, Interpolation Type=831

Interp Num=2, Max.Density=300, Contrast=6, Border=0

TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87

under Device Setup, SCP PRINTERS, Current Providers:

Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 10

under Device Setup, SCP PRINTERS, Registered Users:

SCU AETitle= PR_cthostname, IPAddr=accordingly

SCP=8700, Copies=1, Contrast Table=1, TFT/ULUT= Wrkstn2a.w87

Max.Density= 280, Border Density=0, Mag.Type=8, Smoothing=0

Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low

Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No

Trim=No, Attribute Warning=Yes

Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus setup

under COMM Menu: 969 default

under IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14

under EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Time-out=60

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Camera switched off PRINTER DOWN PRINTER DOWN Camera down

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

No supply magazine inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 167 of 260

Page 168: CT00-000.814.30.02.02.pdf

168 Specific Camera Information - Kodak

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Camera is in calibra-tion

CALIBRATING ERR

CALIBRATING ERR

Problem when cali-brating Film quality may not be optimal

Camera is warming up

PRINTER INIT PRINTER INIT Camera initializing

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 300.0 300.5 0.17

Y 400.0 400.2 0.05

14" x 17"

Landscape

X 400.0 400.2 0.05

Y 300.0 300.5 0.17

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 168 of 260

Page 169: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 169

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 169 of 260

Page 170: CT00-000.814.30.02.02.pdf

170 Specific Camera Information - Kodak

Kodak DryView 8200 / PACS Link MIM 200 5.5

General information 0

Block diagram

Fig. 22:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Format Release status Film Type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

11" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 170 of 260

Page 171: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 171

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The minimum density command is not supported.

NOTE The background command does not work properly. The bor-der is always black.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 228 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV82001

1. Please do not use an AE title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class (printer spec file name)

kodak8200 No

DICOM Node Kodak8200 Yes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 171 of 260

Page 172: CT00-000.814.30.02.02.pdf

172 Specific Camera Information - Kodak

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.0 x 78.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4360 x 5220

14" x 14"

4360 x 4244

11" x 14"

3272 x 4360

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5220 x 4360

14" x 14"

4244 x 4360

11" x 14"

4360 x 3272

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Device type: DICOM Print Server (SCP)

Device name: PACS Link MIM 200

Manufacturer: Kodak

SW Revision: V 6.1.1

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 172 of 260

Page 173: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 173

Printer settings

Physical Interface: Ethernet 10/100 base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: DCM, Rev. J, June 20, 2005

Image memory: 256 MB RAM

Hard disk: 4 GB

Device type: Printer

Device name: DryView 8200

Manufacturer: Kodak

SW Revision: 3.5

Physical Interface: 100/10 base T Ethernet

Network protocol: TCP/IP

DICOM Conformance Statement: n.a.

Image memory: 50 MB

Hard disk: n.a.

Content: Configuration

MIM 200 setup

Under Configure, Network:

Node Name=MIM, IPAddr=accordingly

Subnet Mask=accordingly, Gateway Addr=accordingly

under Configure, Input, Network, Add:

Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

under Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:

Device type: DICOM Print Server (SCP)

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 173 of 260

Page 174: CT00-000.814.30.02.02.pdf

174 Specific Camera Information - Kodak

General Remarks:Log.Name=8200, IPAddr=100.100.100.40General Remarks:Database Name=8200, Model=8200, Memory 32Media info.:Media=35x43cm, 4361x5223, 5133, Any (Blue, Clear)Media=35x35cm, 4361x4247, 4157, any (blue, clear)Media=11inx14in, 3272x4361, 4271, any (blue, clear)Supported item:Collation=On, Text=On, Tone Scaling=On,Supported item:Remote Image Deletion=On, Rotation=OnInterpolation:No Magnification=On, Bilin=On, Repl=On,Interpolation:CubicSpline=On, Cubic Convolution 2=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1SCU:Density Space=On, Variable Border=On, PLUT Service=OnSCU:Curve Shape Tone Scaling=On, Perception LUT Tone Scaling=OnM952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7

Content: Configuration

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 174 of 260

Page 175: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 175

Error / Warning Messages (examples)

Under Configure, SCP Services, select the specific modality, modify:

Image Resizing, Resizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT=Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=NInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic Color=N

8200 setup under the CONFIGURATION menu:

On IMAGE QUALITY-> TFT FILES menu: TFT=Ver693c0.w87

under the LOCAL PANEL Menu: Contrast=6, Density=3.00

under the DICOM Menu:

SCU config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0

DICOM config.: AE=ANY_ANY, Port=1025

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Supply magazine not inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

Less than 5 films in supply magazine

SUPPLY LOW SUPPLY LOW Film supply low

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Camera is warming up

PRINTER INIT

PROC INIT

PRINTER INIT

PROC INIT

Camera initializing

Processor initializing

Camera switched off PRINTER DOWN PRINTER DOWN Camera down

Content: Configuration

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 175 of 260

Page 176: CT00-000.814.30.02.02.pdf

176 Specific Camera Information - Kodak

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 400.0 398.5 0.38

14" x 17"

Landscape

X 390.0 389.5 0.13

Y 330.0 330.0 0.00

14" x 14"

Portrait

X 330.0 330.0 0.00

Y 320.0 319.5 0.16

14" x 14"

Landscape

X 320.0 319.5 0.16

Y 330.0 330.0 0.00

11" x 14"

Portrait

X 250.0 249.5 0.20

Y 330.0 330.5 0.15

11" x 14"

Landscape

X 330.0 330.0 0.00

Y 240.0 240.0 0.00

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 176 of 260

Page 177: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 177

Kodak DryView 8100 / PACS Link MIM 200 5.6

General information 0

Block diagram

Fig. 23:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Format Release status Film Type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 177 of 260

Page 178: CT00-000.814.30.02.02.pdf

178 Specific Camera Information - Kodak

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The minimum density command is not supported.

NOTE The background command does not work properly. The Bor-der Density is always black.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 234 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV81001

1. Please do not use an AE title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)kodak8100 No

DICOM Node Kodak8100 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 178 of 260

Page 179: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 179

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.0 x 78.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4096 x 5216

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5216 x 4096

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Device type: DICOM Print Server (SCP)

Device name: PACS Link MIM 200

Manufacturer: Kodak

SW Revision: V 6.1.1

Physical Interface: Ethernet 10/100base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. J, June 20, 2005

Image memory: 256 MB RAM

Hard disk: 4 GB

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 179 of 260

Page 180: CT00-000.814.30.02.02.pdf

180 Specific Camera Information - Kodak

Printer settings

Device type: Printer

Device name: DryView 8100

Manufacturer: Kodak

SW Revision: 2.7

Physical Interface: n.a.

Network protocol: n.a.

DICOM Conformance Statement: n.a.

Image memory: 32 MB

Hard disk: None

Content: Configuration

MIM 200 setup

Under Configure, Network:

Node Name=MIM, IPAddr=accordingly

Subnet Mask=accordingly, Gateway Addr=accordingly

under Configure, Input, Network, Add:

Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

under Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 180 of 260

Page 181: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 181

General Remarks:Log.Name=8100,General Remarks:Database Name=8100, Model=8100, Memory 32Media info.:Media=35x43cm, 4096x5223, 5133, blueOutput parameters.:Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=EvenOutput parameters.:Pixel=12, Header/Line, Transfer=1000 KHz, Acq=60Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation:NoMag=On, Bilin=On, Repl=On, CubicSpline=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=0Formatting:Pixel Pitch=12.8205Image processing:Gray scale=On, Minimize=On, Max Image Dmax=3.1M952 usage:Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7

Under Configure, SCP Services, select the specific modality, modify:

Content: Configuration

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 181 of 260

Page 182: CT00-000.814.30.02.02.pdf

182 Specific Camera Information - Kodak

Error / Warning Messages (examples)

Image resizingResizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=PL (PerceptionLUT), Always Apply=YCurve Shape/TFT, TFT/ULUT=Wrkstn2a.w87, Always Apply=YContrast (dry)=01, Always Apply=YInterp. /Smoothing, Smoothing Type=0, Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=3.00, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffBasic color=N

8100 setup under the CONFIGURATION menu:

On SIB board (i.e. external switch below control plugs):

SW1=Down

Under the COMMUNICATIONS Menu:

Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,... P6=0

On IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off

On IMAGE QUALITY-> TFT FILES menu: TFT=Ver693c0.w87

under the LOCAL PANEL Menu: Contrast=6, Density=3.00

Under the SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361

Under the DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Top cover or door open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Supply magazine empty

SUPPLY EMPTY SUPPLY EMPTY Camera out of film

Content: Configuration

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 182 of 260

Page 183: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 183

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Supply magazine not inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 310.0 310.0 0.00

Y 400.0 401.5 0.37

14" x 17"

Landscape

X 390.0 391.5 0.38

Y 310.0 310.5 0.16

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 183 of 260

Page 184: CT00-000.814.30.02.02.pdf

184 Specific Camera Information - Kodak

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 184 of 260

Page 185: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 185

Kodak DryView 8150 5.7

General information 0

Block diagram

Fig. 24:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 12

Format Release status Film type Medium type

14" x 17"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

14" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

11" x 14"Portrait Yes DVB Blue-based

Landscape Yes DVB Blue-based

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 185 of 260

Page 186: CT00-000.814.30.02.02.pdf

186 Specific Camera Information - Kodak

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The European Standard EN55011 is supported only withClass A (hospital use).

NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 146 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DV81501

1. Please do not use an AE Title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settings Display Changes allowed

Type DICOM Printer No

Class

(Printer spec file name)kodak8150 No

DICOM node DV8150 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 186 of 260

Page 187: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 187

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.12 x 78.12 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4438x5390

14" x 14"

4438x4414

11" x 14"

3440x4438

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5390x4438

14" x 14"

4414x4438

11" x 14"

4438x3440

Medium type Blue film Yes

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

Manufacturer / device type: DICOM Printer (SCP)

Device name: DryView 8150

Manufacturer: Kodak

SW Revision: V1.1.b04 / MIM_V5.7.b12

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 187 of 260

Page 188: CT00-000.814.30.02.02.pdf

188 Specific Camera Information - Kodak

Printer settings

Physical type: Ethernet 10/100 base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. 1.0, July 27, 2004

Image memory: 786 MB RAM

Hard disk: 40 GB

DICOM SCP / Communications

Port number 5040

Association Limit 12

Source Name DICOM

DICOM SCP / Cropping

No settings available

DICOM SCP / SCP Services / Image Optimization

Negotiate Presentation LUT No

Negotiate Color Print No

TFT Set Name LINODS.w87

Always Apply TFT Set Yes

Contrast 4

Always Apply Contrast Yes

Smoothing Type 0

Always Apply Smoothing Type No

Image DMin 0.00

Always Apply Image DMin No

Image DMax 3.00

Always Apply Image DMax No

Image Polarity Normal

Always Apply Image Polarity No

Manufacturer / device type: DICOM Printer (SCP)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 188 of 260

Page 189: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 189

DICOM SCP / SCP Services / Other Options

Send N-Event Reports No

Always Apply Send N-Event Reports Yes

Send Warnings Yes

Print priority Normal

Always Apply Print Priority No

Border Type 3.99

Always Apply Border Type No

2mm Border No

Film Clip Border No

Film Base Blue Film

Always Apply Film Base No

Film Size 14 IN X 17 IN

Always Apply Film Size No

Include Print date in Annotation No

Include Print time in Annotation No

Include Page Number in Annotation No

Include Hospital Name in Annotation No

Include modality ID in Annotation No

Modality ID PR_hostname

Annotation Override No

Density Patch No

Image Resizing Minify/Scale

Always Apply Image Resizing No

Text Box Location None

Source ID DI000

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 189 of 260

Page 190: CT00-000.814.30.02.02.pdf

190 Specific Camera Information - Kodak

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Supply magazine empty

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

Camera info:

EMPTY 14X17 BLUE

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Magazine with dam-aged chip inserted

BAD SUPPLY MGZ BAD SUPPLY MGZ Problem with supply magazine

Camera in calibration PRINTER DOWN PRINTER DOWN Camera offline

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 328.5 0.45

Y 410.0 409.5 0.12

14" x 17"

Landscape

X 410.0 410.0 0.00

Y 330.0 328.5 0.45

14" x 14"

Portrait

X 330.0 327.5 0.76

Y 330.0 330.0 0.00

14" x 14"

Landscape

X 330.0 330.0 0.00

Y 330.0 327.5 0.76

11" x 14"

Portrait

X 250.0 250.0 0.00

Y 330.0 327.5 0.76

11" x 14"

Landscape

X 330.0 327.5 0.76

Y 250.0 250.0 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 190 of 260

Page 191: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 191

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 191 of 260

Page 192: CT00-000.814.30.02.02.pdf

192 Specific Camera Information - Kodak

Kodak KELP 2180/PACS Link MIM 200 5.8

General information 0

Block diagram

Fig. 25:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 11

Format Release status Film Type Medium type

14" x 17"Portrait Yes EHN-10 Blue-based

Landscape Yes EHN-10 Blue-based

14" x 14"Portrait Yes DHG-10 Blue-based

Landscape Yes DHG-10 Blue-based

11" x 14"Portrait Yes DHG-10 Blue-based

Landscape Yes DHG-10 Blue-based

10" x 12"Portrait N/A N/A N/A

Landscape N/A N/A N/A

8" x 10"Portrait Yes DHG-10 Blue-based

Landscape Yes DHG-10 Blue-based

Other N/A N/A N/A

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 192 of 260

Page 193: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 193

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 115 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title K21801

1. Please do not use an AE title such as “NER_...”

Yes

Port number 5040 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class (printer spec file name)

kodak2180 No

DICOM Node Kelp 2180 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 79.0 x 79.0 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 193 of 260

Page 194: CT00-000.814.30.02.02.pdf

194 Specific Camera Information - Kodak

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4096 x 5156

14" x 14"

4096 x 4132

11" x 14"

3236 x 4132

8" x 10"

2284 x 2836

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5156 x 4096

14" x 14"

4132 x 4096

11" x 14"

4132 x 3236

8" x 10"

2836 x 2286

Medium type Blue film Yes

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM Print Server (SCP)

Device name: PACS Link MIM 200

Manufacturer: Kodak

SW Revision: V 6.1.1

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 194 of 260

Page 195: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 195

Printer settings

Physical Interface: Ethernet 10/100 base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: Rev. J, June, 2005

Image memory: 256 MB RAM

Hard disk: 4 GB

Manufacturer / device type: Printer

Device name: KELP 2180

Manufacturer: Kodak

SW Revision: V 2.17H

Physical type: None

Network protocol: None

Manufacturer´s specifications: None

Image memory: 24 MB

Hard disk: None

Content: Configuration

MIM 200 setup

Under Configure, Network:

Computer Name=e.g., Kodak2180, IPAddr=accordingly

Subnet Mask=accordingly, Gateway Addr=accordingly

under Configure, Input, Network, Add:

Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Under Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:

Device type: DICOM Print Server (SCP)

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 195 of 260

Page 196: CT00-000.814.30.02.02.pdf

196 Specific Camera Information - Kodak

General Remarks:Log.Name=KELP2180, Database Name=2180General Remarks:Model=KELP2180, Memory 24, Node ID=6 or accordinglyMedia info.:Media=35x43cm, 4136x5160, 5030, (blue)Media=35x35cm, 4136x4136, 4006, (blue)Media=11"x14", 3240x4136, 4006, (Blue)Media=8"x10", 2290x2840, 2710, (Blue)Supported item:Collation=On, Text=On, Tone Scaling=On, Rotation=OnInterpolation:No Mag=On, Bil=On, Repl=On, Cubic=On, Cubic 2=OnFormat types:Custom=On, Slide=On, Row Symmetric=OnFormat types:1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=OnFormatting:Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=OnFormatting:DICOM Annotation Box=On, Req.Size Scale=OnFormatting:Req.Size Mix in Row=On, Trim=OnFormatting:Aspect Ratio Mix in Row=On, Image Size Mix in Row=OnFormatting:Req.Image Size=On, Req.Size Mix on Page=OnFormatting:True Landscape=OnFormatting:Max Img per Band=20, Max Band per Page=20, Max Img Col=0Formatting:Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2Formatting:Trim Width Enabled=2, Trim Width Disabled=2Formatting:Pixel Pitch=12.6582Image processing:Gray Scale=On, Minimize=On, Max Image Dmax=3.2M952 usage:Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7

Under Configure, SCP Services, select the specific modality, modify:

Content: Configuration

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 196 of 260

Page 197: CT00-000.814.30.02.02.pdf

Specific Camera Information - Kodak 197

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Image Resizing, Resizing <M/S>=Minify/Scale, Always Apply=NRelocate Text Box, Location = noneN-EVENT Reporting, Reporting=N, Always Apply=YWarning Status=YPLUT=NTone Scaling=CS (Curve Shape), Always Apply=YCurve Shape/TFT, Curve Shape=0.0, Always Apply=YContrast (wet)=+0, Always Apply=NInterpolation=No Mag., Always Apply=NImage Dmin=0.00, Always Apply=N,Image Dmax=2.80, Always Apply=NPolarity=Normal, Always Apply=NBorder=3.99, Always Apply=NSize=14"x17", Always Apply=NBase=Blue Film, Always Apply=NPriority=Normal, Always Apply=NPage Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hospital Name=Off, Print Date=Off, Print Time=Off, Page Number=OffFilm Output = 9(no. of sorter), Always Apply=NBasic color = N

2180 setup:

Optical Interface Board will automatically configure to 2180, no setup needed

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

The switch “Access magazines” is acti-vated

PRINTER OFFLINE PRINTER OFFLINE Camera offline

Send film size which is not inserted in camera

CHECK PRINTER CHECK PRINTER Check camera

Printing is stopped PRINTER OFFLINE PRINTER OFFLINE Camera offline

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Content: Configuration

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 197 of 260

Page 198: CT00-000.814.30.02.02.pdf

198 Specific Camera Information - Kodak

Real Size Scaling

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 310.0 310.0 0.00

Y 390.0 389.5 0.13

14" x 17"

Landscape

X 390.0 389.5 0.13

Y 320.0 320.0 0.00

14" x 14"

Portrait

X 310.0 310.0 0.00

Y 320.0 319.5 0.16

14" x 14"

Landscape

X 310.0 309.5 0.16

Y 320.0 320.0 0.00

11" x 14"

Portrait

X 240.0 240.0 0.00

Y 320.0 319.5 0.16

11" x 14"

Landscape

X 310.0 309.5 0.16

Y 240.0 240.0 0.00

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X 170.0 170.0 0.00

Y 210.0 209.5 0.24

8" x 10"

Landscape

X 210.0 209.5 0.24

Y 170.0 1700 0.00

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 198 of 260

Page 199: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 199

6- 6Specific Camera Information - Konica MinoltaKonica Minolta DryPro 832 0

General Information 0

Block Diagram

Fig. 26:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 9

Format Release status Film type Medium type

14" x 17"Portrait Yes DRYPRO SD-Q Blue-based

Landscape Yes DRYPRO SD-Q Blue-based

14" x 14"Portrait Yes DRYPRO SD-Q Blue-based

Landscape Yes DRYPRO SD-Q Blue-based

11" x 14"Portrait Yes DRYPRO SD-Q Blue-based

Landscape Yes DRYPRO SD-Q Blue-based

10" x 12"Portrait Yes DRYPRO SD-Q Blue-based

Landscape Yes DRYPRO SD-Q Blue-based

8" x 10"Portrait Yes DRYPRO SD-Q Blue-based

Landscape Yes DRYPRO SD-Q Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 199 of 260

Page 200: CT00-000.814.30.02.02.pdf

200 Specific Camera Information - Konica Minolta

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Magnification "NONE" not supported therefore KonicaServer needs a fixed setting:

Magnification "CUBIC" with smoothing type 7.

NOTE If “Camera cover open”, no message is displayed. If the cam-era is ready, jobs will be queued but not printed.

NOTE If a film is not inserted, but configured in camera server, thequeue is stopped with a camera software problem. This jobmust be deleted from the camera queue.

Known effects

NOTE The DICOM Print Server PRINTLINK 5 is not 510k registered.

NOTE Two film sizes can be installed at Drypro 832. To install otherfilm sizes the Konica-Minolta Service Technician is needed.

SCU / modality software configuration 0

Time 62 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DPRO832 Yes

Port number 104 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 200 of 260

Page 201: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 201

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec. file name)konicaDryPro832_largeB No

DICOM Node Konica832 Yes

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.6 x 78.6 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

4344 x 5196

14” x 14”

4344 x 4230

11" x 14"

3390 x 4230

10” x 12”

3046 x 3580

8" x 10"

2398 x 2932

In “Define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17"

5260 x 4268

14” x 14”

4294 x 4268

11" x 14"

4306 x 3327

10” x 12”

3644 x 2970

8" x 10"

2996 x 2322

Medium type Blue film No

Film destination Processor Yes

Color appearance Gray scale 8 bit No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 201 of 260

Page 202: CT00-000.814.30.02.02.pdf

202 Specific Camera Information - Konica Minolta

NOTE Do not modify or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

SCP / hardcopy camera configuration (provided by OEM) 0

Identification

Printer Settings

Web-Servicetool for Printlink 5

Configuration Parameter:

select input channel in "Input Parameter Setup Screen" window:

Background Black No

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM Printer (SCP)

Device name: DRYPRO 832

Manufacturer: Konica Minolta

SW Revision: V1.02R00 (for Printer & DICOM Server)

Physical type: Ethernet Card 100 / 1000

Network protocol: TCP/IP

DICOM conformance statement: Konica-Minolta, Printlink5-IN, Ver. 1.0 2006.10

Image memory: Printer: 64 MB; DS: 1 GB-RAM

Hard disk: Printer: None; DS: 80 GB

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 202 of 260

Page 203: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 203

Configuration LUT:

select input channel in "LUT Setup Screen" window:

Content Printer Setting Value

Input Channel CT_hostname

Number of Copies 1

Print priority LOW

Film destination BIN_1

Film orientation PORTRAIT

Magnification Type CUBIC (checkbox selected)

Smooth Type 7 (checkbox selected)

Border density BLACK

Min. Dens. 0.20

Max. Dens.: 3.00

Trim: NO

Config. Information KC_LUT=1 (checkbox selected)

Density 0

Contrast 0

Flip NORMAL

Polarity NORMAL

Medium type BLUE FILM

Film Size ID 14" X 17"

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 203 of 260

Page 204: CT00-000.814.30.02.02.pdf

204 Specific Camera Information - Konica Minolta

Content Printer Setting Value

Input Channel CT_hostname

LUT No 1

Display Mode 16Step - Density

Drag Mode One

Illumination 2000

Refl. Ambient 10

Basic Pattern (linear LUT):

Previous Setting 1=20 / 2=39 / 3=57 / 4=76 / 5=95 / 6=113 / 7=132 / 8=151 / 9=169 / 10=188 / 11=207 / 12=225 / 13=244 / 14=263 / 15=281 / 16=300)

Test Image Convert P-Value Convert

Polarity NORMAL

Configuration Maintenance

Host name LENOVO-0B9651EF

IP address xxx.xxx.xxx.xx

AE Title SCP PLNK5

AE Title SCU PLNK5

Port No. 104, 6000, 5200

select input channel in "Printlink 5 Maintenance" window:

Input Channel 4

Input Name CT_hostname

SCU IP-Adr. xxx.xxx.xxx.xx

SCU AET title PR_hostname

Protocol Name DICOM PRINT IN

Maker KonicaMinolta (1.03)

Model RS

"double-click" Ch4, SOM64B

Channel Name CT_hostname

Default Channel (checkbox deselected)

Protocol Name DICOM PRINT IN

Default Format 1x1

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 204 of 260

Page 205: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 205

IP Adr./H. Name xxx.xxx.xxx.xx

AE Title SCU PR_hostname

Common Input ... ("not relevant for printing")

DRYPRO 832 Output Setup:

Luminance Conversion ON

Border Size CT/MRI

Format Type EVEN

Pat.Name Stamp Type OFF

DICOM PRINT IN

N-Event Port No. 0

Empty Notice LV Standard (checkbox selected)

Error Level M/M Abort

Max PDU 64

Upper Stamp Use OFF

Printlink 5 Maintenance IMAGER:

"double click" Imager Ch1 DPRO832

Imager Output channel setup screen

Channel Name DPRO832

Protocol Name DICOM PRINT OUT

IP-Adr./Hostname: Localhost

DRYPRO832 IP-Adr. xxx.xxx.xxx.xx

AET-SCP KC_DRO4_P001

Port No 63101

Common Output

Connect Imager DRYPRO832

Output Pixels 12bit

Convert LUT ON

Stretch Base No Stretch

Stretch Mode BiLinear

Content Printer Setting Value

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 205 of 260

Page 206: CT00-000.814.30.02.02.pdf

206 Specific Camera Information - Konica Minolta

Error / warning messages (examples)

Warning Out ON

Layout DcmPrtLayOut832.ini

Output Name ASCII + KANJI + KANA

Output Image Color Grayscale

Study Output Order Normal

Annotation Box Not send

Calculate Sampling Pitch (checkbox deselected)

Imager Default Setting:

Default Film Size 1 14x17

Def. Media Type 1 BLUE FILM

Def. Film Type 2 11x14

Def. Media Type 2 BLUE FILM

Action that results in cam-era error mes-

sages

Printer messagesMessage displayed

on SCUDefined DCS (*1) Sent by SCP (*2)

Film Jam FILM JAM FILM JAM Film jam

Open cover and no film is printing

COVER OPEN Normal * "nothing is displayed, no film is printed, all films are stored in the camera. After closing the cover all stored films were printed out.

10"x12" film Sup-ply empty

EMPTY 10x12 BLUE EMPTY 10x12 BLUE 10x12 blue film supply empty

Switch off cam-era but Konica DICOM Server is running

PRINTER DOWN PRINTER DOWN Camera down

Content Printer Setting Value

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 206 of 260

Page 207: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 207

Image Quality and Optical Density 0

No specific parameter available.

Real size scaling

Send a film with a size which is not inserted in camera (14"x17", 8"x10" inserted, 10"x12" is sent)

not defined ELEC SWERROR ** Camera software problem - Queue stopped.

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

* In case of "Camera cover open", no message is displayed. If the camera is ready, jobs are queued only and not printed.

** If a film is not inserted and configured in the camera server, the queue is stopped with a camera software problem. This job must be deleted in the camera queue.

Action that results in cam-era error mes-

sages

Printer messagesMessage displayed

on SCUDefined DCS (*1) Sent by SCP (*2)

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 390.0 391.5 0.38

14" x 17"

Landscape

X 410.0 412.0 0.49

Y 320.0 320.5 0.16

14" x 14"

Portrait

X 300.0 301.0 0.33

Y 300.0 301.0 0.33

14" x 14"

Landscape

X 300.0 301.0 0.33

Y 300.0 301.0 0.33

11" x 14"

Portrait

X 250.0 250.0 0.00

Y 320.0 321.0 0.31

11" x 14"

Landscape

X 330.0 331.0 0.30

Y 250.0 250.5 0.20

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 207 of 260

Page 208: CT00-000.814.30.02.02.pdf

208 Specific Camera Information - Konica Minolta

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 271.0 0.37

10” x 12”

Landscape

X 270.0 271.0 0.37

Y 220.0 220.5 0.23

8" x 10"

Portrait

X 170.0 170.0 0.00

Y 220.0 221.0 0.45

8" x 10"

Landscape

X 230.0 231.0 0.43

Y 170.0 170.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 208 of 260

Page 209: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 209

Konica Minolta DryPro 793 6.1

General information 0

Block diagram

Fig. 27:

Available film sizes

Format Release status Film type Medium type

14" x 17"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

14" x 14"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

11" x 14"

Portrait YesDRYPRO SD-P

NewBlue-based

Landscape YesDRYPRO SD-P

NewBlue-based

10" x 12"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

8" x 10"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 209 of 260

Page 210: CT00-000.814.30.02.02.pdf

210 Specific Camera Information - Konica Minolta

Performance and Connections

Associations

Maximum number of associations supported: 32

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE The minimum density command is not supported. The mini-mum density value is the value of the fog on the film.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 105 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DPRO793 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec. file name)konicaDryPro793_largeB No

DICOM Node Konica793 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 210 of 260

Page 211: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 211

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes that are not available at the customer site.

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 87.5 x 87.5 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

3902 x 4668

14" x 14"

3902 x 3799

11" x 14"

3044 x 3799

10” x 12”

2745 x 3219

8" x 10"

2160 x 2643

In “Define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

4725 x 3834

14" x 14"

3856 x 3834

11" x 14"

3868 x 2988

10” x 12”

3275 x 2677

8" x 10"

2699 x 2092

Medium type Blue film No

Film destination Processor Yes

Color appearance Gray scale 8 bit No

Background Black No

Transformation No magnification No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 211 of 260

Page 212: CT00-000.814.30.02.02.pdf

212 Specific Camera Information - Konica Minolta

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Printer settings

The configuration has to be performed via the printer keypad.

Konica Service has to install the modality name so that the printer settings can be assigned to this name.

Switch to Service mode

Select "Print" Mode

Device type: DICOM Printer (SCP)

Device name: DRYPRO 793

Manufacturer: Konica Minolta Medical & Graphics, Inc.

SW Revision: V1.00R05

Physical type: Ethernet 10 base-T / 100 base-TX / 1000 base-T

Network protocol: TCP/IP

DICOM Conformance Statement: Ver.1.02, 2005.02

Image memory: 512 MB

Hard disk: 80 GB

Setting: Print Setting for CT modality Value

LUT NO LUT1: LINEAR

DENSITY 0

CONTRAST 0

SMOOTHING 7

MAX DENSITY 300

MIN DENSITY 20

LUMINOUS INTENSITY 2000

AMBIENT LIGHT 10

ROTATION OFF

SORTER 1

ORIENTATION PORTRAIT

IMAGE POLARITY POSITIVE

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 212 of 260

Page 213: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 213

BORDER BLACK

TRIM ON/OFF OFFS

TAM

P

TIME/DATE (Not selected)

COPY NO. (Not selected)

SCU ID/PAGE NO. (Not selected)

STAMP MESSAGE (Not selected)

PATIENT ID / NAME (Not selected)

PATIENT DATA ASCII (Not selected)

STAMP CONTENT (EMPTY)

Setting: Print Setting for CT modality Value

STAMP POSITION

EDGE ENHANCE OFF

BORDER SIZE CT/MRI

EXPANSION CHARACTER OFF

MANAGE INFO. PRINT ON

REQ.IMAGESIZE EFFECTIVE

REQ:BEHAVIOR DECIMATE

INTEGRAL MULTIPLE RATIO 2

P VALUE RECEPTION OFF

FORMAT EVEN

STANDARD DMIN/DMAX DICOM

Setting: Print Setting for CT modality Value

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 213 of 260

Page 214: CT00-000.814.30.02.02.pdf

214 Specific Camera Information - Konica Minolta

Select “Service 1” Mode

DIC

OM

Prio

rity

DEFAULT LUT (Not selected)

SMOOTH (Not selected)

DENSITY MAX (Not selected)

DENSITY MIN (Not selected)

POLARITY (Not selected)

ORIENTATION (Not selected)

TRIM (Not selected)

BORDER (Not selected)

SORTER (Not selected)

Setting: SCU SETUP for CT modality Value

CHANNEL USE ON

SCU NAME CT-hostname

IP ADDRESS xxx.xxx.xxx.xxx

HOST NAME CT-hostname

PORT NUMBER 104

AE TITLE PR_hostname

ERROR LEVEL STANDARD

SIZE TYPE CHECK ON

ENFORCE SPECIFIC FILM OFF

STANDARD MAGAZINE 14x17B

EMPTY NOTICE

SIZE TYPE 14x17B ON

SIZE TYPE 8x10B ON

SIZE TYPE 10x12B ON

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 214 of 260

Page 215: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 215

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available.

Real size scaling

Action that results in cam-era error mes-

sages

Printer messagesMessage displayed

on SCUDefined DCS (*1) Sent by SCP (*2)

Switch to service mode

CHECK PRINTER CHECK PRINTER Check camera

14x17 Magazine empty

EMPTY 14x17 BLUE EMPTY 14x17 BLUE 14x17 blue film supply empty

Front door open CHECK PRINTER CHECK PRINTER Check camera

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.5 0.15

Y 390.0 390.0 0.00

14" x 17"

Landscape

X 400.0 400.0 0.00

Y 320.0 321.0 0.31

14" x 14"

Portrait

X 330.0 330.5 0.15

Y 320.0 320.0 0.00

14" x 14"

Landscape

X 330.0 330.0 0.00

Y 320.0 320.5 0.16

11" x 14"

Portrait

X 250.0 250.0 0.00

Y 320.0 320.5 0.16

11" x 14"

Landscape

X 330.0 330.5 0.15

Y 250.0 250.0 0.00

10” x 12”

Portrait

X 230.0 230.0 0.00

Y 270.0 270.5 0.19

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 215 of 260

Page 216: CT00-000.814.30.02.02.pdf

216 Specific Camera Information - Konica Minolta

10” x 12”

Landscape

X 270.0 270.5 0.19

Y 220.0 220.0 0.00

8" x 10"

Portrait

X 170.0 170.0 0.00

Y 220.0 220.5 0.23

8" x 10"

Landscape

X 230.0 230.0 0.00

Y 180.0 180.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 216 of 260

Page 217: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 217

Konica Minolta DryPro 771 6.2

General information 0

Block diagram

Fig. 28:

Available film sizes

Format Release status Film Type Medium type

14" x 17"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

14" x 14"Portrait Yes DRYPRO SD-P Blue-based

Landscape No No No

11" x 14"

Portrait YesDRYPRO SD-P

NewBlue-based

Landscape YesDRYPRO SD-P

NewBlue-based

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 217 of 260

Page 218: CT00-000.814.30.02.02.pdf

218 Specific Camera Information - Konica Minolta

Performance and Connections

Associations

Maximum number of associations supported: 16

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE 14INX14IN landscape not supported by camera.

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 112 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DP771 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)konicaDyPro771_largeB No

DICOM Node KonicaDP771 Yes

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 218 of 260

Page 219: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 219

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 78.6 x 78.6 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4350 x 5202

14" x 14"

4350 x 4236

11" x 14"

3384 x 4236

In “define new”

Do not modify/add

Select only

Delete onlyFilm Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5318 x 4236

11" x 14"

4310 x 3282

Medium type Blue film No

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black No

Transformation No magnification No

Device type: DICOM Printer (SCP)

Device name: DRYPRO MODEL 771

Manufacturer: KonicaMinolta

SW Revision: V1.00R03

Physical type: RJ45

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 219 of 260

Page 220: CT00-000.814.30.02.02.pdf

220 Specific Camera Information - Konica Minolta

Printer settings

The configuration has to be performed via the printer keypad.

Konica service has to install the modality name and assign it to a channel number. For this camera validation, the modality name "specific Modality name" is installed and assigned to channel: "CH3". Use Maintenance Mode to view/modify Print Server Configuration.

Select U30 Print Cond.

Select CH3 "specific modality name"

Enter

Network protocol: DICOM, TCP/IP

DICOM Conformance Statement: Vers. 1.00, Sep.01. 2003

Image memory: 128 MB / Standard

Hard disk: 80 GB / Standard

Device type: DICOM Printer (SCP)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 220 of 260

Page 221: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 221

Exit

Select U90 - Password

Enter

INPUT PASSWORD

Enter

Select U80 Service Mode

S00 DICOM SCU displayed

Select CH3 "specific modality name"

Enter

Variable name Content

LUT NO 1 LINEAR

DENSITY 0

CONTRAST 0

SMOOTHING 2

MAX DENSITY 300

MIN DENSITY 20

ORIENTATION PORTRAIT

IMAGE POLARITY POSI

TRIM ON/OFF Off

BORDER Black

ILLUMINATION 2000

AMBIENT LIGHT 10

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 221 of 260

Page 222: CT00-000.814.30.02.02.pdf

222 Specific Camera Information - Konica Minolta

Exit

Select S30 DICOM PRIORITY

Select CH3 "specific modality name"

Enter

Exit

Select S20 Print Cond.

Select CH3 "specific modality name"

Enter

S21 STAMP SELECT

Variable name Content

CHANNEL USE ON

SCU NAME CT-hostname

N-EVENT PORT NO BORD=0

IP ADDRESS xxx.xxx.xxx.xxx

AE TITLE PR_hostname

HOST NAME CT-hostname

BORDER SIZE REGIUS

FILM SIZE TYPE NO CHECK

EMPTY NOTICE ON

DICOM ERROR LEVEL STANDARD

FRAME LAYOUT EVEN

Dmax/Dmin DICOM

P-VALUE RCV OFF

Variable name Content

DEFAULT LUT DICOM

SMOOTH DICOM

DMAX DICOM

DMIN DICOM

POLARITY DICOM

ORIENTATION DICOM

TRIM DICOM

BORDER DICOM

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 222 of 260

Page 223: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 223

Exit

Data was changed

Save the data? Yes

Enter

Exit

Error / Warning Messages (examples)

Variable name Content

TIME/DATE OFF

Copy no. OFF

Patient ID OFF

SCU ID / Page OFF

STAMP MESSAGE OFF

STAMP MESSAGE

STAMP DIRECTION LOWER

YEAR/FORMAT YY

MONTH/FORMAT TEXT

DATE/FORMAT TIME+DATE

REQ.IMAGESIZE EFFECTIVE

EXPANSION CHAR OFF

EDGE ENHANCE OFF

REQ:BEHAVIOR DECIMATE

FLIP ON/OFF OFF

Action that results in camera error mes-

sages

Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP

14x17 Magazine empty EMPTY 14x17 BLUE

EMPTY 14x17 BLUE

14x17 blue film supply empty

Front door open COVER OPEN CHECK PRINTER

Check camera

Switch to service mode n.a. CHECK PRINTER

Check camera

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 223 of 260

Page 224: CT00-000.814.30.02.02.pdf

224 Specific Camera Information - Konica Minolta

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Action that results in camera error mes-

sages

Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.5 0.15

Y 400.0 401.0 0.25

14" x 17"

Landscape

X 410.0 411.0 0.24

Y 320.0 320.5 0.16

14" x 14"

Portrait

X 330.0 330.0 0.00

Y 320.0 321.0 0.31

14" x 14"

Landscape

X No No No

Y No No No

11" x 14"

Portrait

X 250.0 250.0 0.00

Y 320.0 320.5 0.16

11" x 14"

Landscape

X 330.0 330.5 0.15

Y 250.0 250.5 0.20

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 224 of 260

Page 225: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 225

Konica DryPro 751/752 6.3

General information 0

Block diagram

Fig. 29:

Available film sizes

Format Release status Film Type Medium type

14" x 17"

Portrait YesDRYPRO SD-P

NewBlue-based

Landscape YesDRYPRO SD-P

NewBlue-based

14" x 14"Portrait Yes DRYPRO SD-P Blue-based

Landscape Yes DRYPRO SD-P Blue-based

11" x 14"

Portrait YesDRYPRO SD-P

NewBlue-based

Landscape YesDRYPRO SD-P

NewBlue-based

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 225 of 260

Page 226: CT00-000.814.30.02.02.pdf

226 Specific Camera Information - Konica Minolta

Performance and Connections

Associations

Maximum number of associations supported: 16

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

Known Effects

n.a.

SCU / Modality Software Configuration 0

Time 98 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title DPRO752 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)konicaDryPro752 No

DICOM Node DPRO752 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 226 of 260

Page 227: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 227

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Pixel Size [1/1000 mm] 80.0 x 80.0 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4272 x 5112

14" x 14"

4272 x 4162

11" x 14"

3336 x 4162

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5224 x 4162

14" x 14"

4272 x 4162

11" x 14"

4274 x 3224

Medium type Blue film No

Film destination Processor Yes

Color appearance Grayscale 8 bit No

Background Black No

Transformation No magnification No

Filming properties Display Changes allowed

Device type: DICOM Printer (SCP)

Device name: DRYPRO 752

Manufacturer: Konica

SW Revision: V1.00R7

Physical type: Ethernet RJ45 10/100 Base TX

Network protocol: TCP/IP

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 227 of 260

Page 228: CT00-000.814.30.02.02.pdf

228 Specific Camera Information - Konica Minolta

Printer settings

DICOM Conformance Statement: V 1.03, Jun.18.2003

Image memory: 256 MB RAM

Hard disk: 20 GB

File Name Parameters Content

Maintenance 1/3

Print Condition CT host name

LUT LINEAR

DENSITY 0

CONTRAST 0

SMOOTH TYPE 7

MAX DENSITY 300

MIN DENSITY 020

L0 0

La 0

SORTER BIN 01

ORIENTATION Portrait

POLARITY Posi

TRIM Off

BORDER Black

FLIP Off

STAMP SELECT DATE TIME No

Copy no. No

Patient ID No

SCU ID/page no. No

Device type: DICOM Printer (SCP)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 228 of 260

Page 229: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 229

STAMP MESSAGE No

STAMP DIRECTION Stamp position in graphic: top

Maintenance 2/3

SERVICE n.a.

DICOM SCU CT hostname

SCU SETUP n.a.

SCU CH USE ON

SCU NAME CT hostname

N-EVENT PORT NO 0

IP ADDRESS xxx.xxx.xxx.xxx

AE TITLE PR_hostname

HOST NAME CT hostname

SIZE TYPE CHECK ON

BORDER SIZE REGIUS

EMPTY NOTICE n.a.

14x17 B ON

14x14 B ON

EDGE ENHANCE ON

FRAME LAYOUT CENTER

DICOM ERROR LEVEL STANDARD

DEFAULT Dmax/Dmin DICOM

Magnification 02

File Name Parameters Content

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 229 of 260

Page 230: CT00-000.814.30.02.02.pdf

230 Specific Camera Information - Konica Minolta

Error / Warning Messages (examples)

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Action that results in camera error mes-

sages

Printer messages Message displayed on SCU (*2)Defined DCS (*1) Sent by SCP

Camera switched off n/a TCP init error, tim-eout

Connection to camera is not available

14x17 Magazine empty EMPTY 14x17 BLUE

EMPTY 14x17 BLUE

14x17 blue film supply empty

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 331.0 0.30

Y 380.0 381.0 0.26

14" x 17"

Landscape

X 400.0 401.0 0.25

Y 320.0 321.0 0.31

14" x 14"

Portrait

X 330.0 331.0 0.30

Y 320.0 321.0 0.31

14" x 14"

Landscape

X 330.0 331.0 0.30

Y 320.0 321.0 0.31

11" x 14"

Portrait

X 250.0 251.0 0.40

Y 320.0 321.0 0.31

11" x 14"

Landscape

X 330.0 331.0 0.30

Y 240.0 241.0 0.42

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 230 of 260

Page 231: CT00-000.814.30.02.02.pdf

Specific Camera Information - Konica Minolta 231

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 231 of 260

Page 232: CT00-000.814.30.02.02.pdf

232 Specific Camera Information Sony

7- 7Specific Camera Information SonySony UP-DF 550 0

General information 0

Block diagram

Fig. 30:

Available film sizes

Performance and Connections

Associations

Maximum number of associations supported: 8

Format Release status Film Type Medium type

14" x 17"Portrait Yes UPT-517BL Blue-based

Landscape Yes UPT-517BL Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait Yes UPT-514BL Blue-based

Landscape Yes UPT-514BL Blue-based

10" x 12"Portrait Yes UPT-512BL Blue-based

Landscape Yes UPT-512BL Blue-based

8" x 10"Portrait Yes UPT-510BL Blue-based

Landscape Yes UPT-510BL Blue-based

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 232 of 260

Page 233: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 233

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, format 4x5, Copies=1, default center/width, magnification type none.

The optimal connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Horizontal line structure visible with all film sizes.

NOTE Background of film is always white.

Known Effects

NOTE The UPT-517BL film type requires a specific densitometercharacteristic. Density measurements using the X-Rite 301densitometer are only possible if special optics 301-34SO areinstalled on the X-Rite 301 and calibrated as described in theX-Rite letter.

NOTE If a film size is sent to the camera and the required film sizeis not inserted in the camera, then nothing is displayed, nei-ther in Somaris nor on the camera display. Jobs stay in thequeue until the required magazine is inserted. If an attempt ismade to switch off the camera before all jobs are printed, theunprinted jobs are displayed - and the camera is not switchedoff.

SCU / Modality Software Configuration 0

Time 112 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title UPDF550 Yes

Port number 104 No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 233 of 260

Page 234: CT00-000.814.30.02.02.pdf

234 Specific Camera Information Sony

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)SonyUPDF550 No

DICOM Node Sony550 Yes

Filming properties Display Changes allowed

Hold printed Film Jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 79.4 x 79.4 No

Film Sheet Formats,

Portrait

Number of Pixels

[columns x rows]

14" x 17"

4360 x 5232

11” x 14”

3426 x 4286

10” x 12”

3100 x 3630

8” x 10”

2452 x 3004

In “define new”

Do not modify/add

Select only

Delete only

Film Sheet Formats,

Landscape

Number of Pixels

[columns x rows]

14" x 17"

5232 x 4360

11” x 14”

4286 x 3426

10” x 12”

3630 x 3100

8” x 10”

3004 x 2452

Medium type Blue film No

Film destination Processor No

Color appearance Grayscale 8 bit No

Background Black Yes

Transformation No magnification No

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 234 of 260

Page 235: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 235

NOTE Do not modify and/or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / Hardcopy Camera Configuration (provided by OEM) 0

Identification

Manufacturer / device type: DICOM Printer (SCP)

Device name: Film Station UP-DF 550

Manufacturer: SONY

SW Revision: (for VA20) (for VA26)

MA-FW: 1.05 TST7;

DICOM: V3.21;

DR-FW: V1.01;

CT-FW: V1.00,

ICTAC: V1.00;

IDMAC: V25.0

MA-FW: 1.07 TST1;

DICOM: V3.21;

DR-FW: V1.01;

CT-FW: V1.00,

ICTAC: V1.00;

IDMAC: V25.0

Physical Interface: Ethernet 10/100 base T/Tx

Network protocol: TCP/IP

DICOM Conformance Statement: Sony UP-DF550, Rev. 1.1, Feb.19, 2007

Image memory: 3.5 GB

Hard disk: No

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 235 of 260

Page 236: CT00-000.814.30.02.02.pdf

236 Specific Camera Information Sony

Printer settings

File Name Content Remarks

No file;

Keypad settings, press "Menu”

Information>

IP ADDRESS xxx.xxx.xxx.xxx

TOTAL PRINTS

Network address

No file;

Keypad settings, press "Menu"

PRINT Quality>

SELECT GAMMA

/SL32_14C/ [EXEC]

SHARPNESS OFF

PATTERN PRINT

1: 14x17 [EXEC]

SELECT POINT

95%.....[25%]

SELECT POINT

5%.....[75%]

SAVE GAMMA [EXEC]

Gamma table 8 (Linear LUT)

No file;

Keypad settings, press "Menu"

CALIBRATION>

FULL RANGE CAL

1: 14x17 [EXEC]

AUTO CALIBRATION

[On/Off]

CAL RESET

14x17 [EXEC]

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 236 of 260

Page 237: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 237

Error / Warning Messages (examples)

No file;

Keypad settings, press "Menu"

GAMMA SEL>

DEFAULT14x17

/SL32_14C/

DEFAULT11x14

/SL32_14C/´

DEFAULT10x12

/SL32_14C/

DEFAULT 8x10

/SL32_14C/

Linear LUT

Linear LUT

Linear LUT

Linear LUT

No file;

Keypad settings, press "Menu"

SYSTEM SETTINGS>

NETWORK SETTINGS>

IP ADDRESS

xxx.xxx.xxx.xxx

SUBNETMASK

xxx.xxx.xxx.xxx

DEFAULT Gateway

xxx.xxx.xxx.xxx

STRICT SIZE .. ON

SAVE SETTINGS >EXEC

Network address

Subnet mask

Default Gateway

ON = job stored if film size is not available

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Supply magazine not inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

Supply magazine empty

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

14x17 blue film sup-ply empty

Film jam FILM JAM FILM JAM Film jam

File Name Content Remarks

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 237 of 260

Page 238: CT00-000.814.30.02.02.pdf

238 Specific Camera Information Sony

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

Magazines with 14x17 and 8x10 films inserted, a job with film size 11x14 is sent

Nothing NORMAL Nothing is displayed, neither in Somaris nor on the camera display. Jobs stay in queue until the 11x14 magazine is inserted. If an attempt is made to switch off the cam-era before all jobs are printed, the unprinted jobs are displayed - and the camera is not switched off.

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 320.0 320.0 0.00

Y 400.0 399.5 0.12

14" x 17"

Landscape

X 400.0 399.5 0.12

Y 320.0 320.0 0.00

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X 260.0 260.0 0.00

Y 320.0 319.2 0.25

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 238 of 260

Page 239: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 239

11" x 14"

Landscape

X 320 319.0 0.31

Y 260.0 260.0 0.00

10” x 12”

Portrait

X 240.0 239.8 0.08

Y 260.0 259.5 0.19

10” x 12”

Landscape

X 280.0 279.5 0.18

Y 220.0 220.0 0.00

8" x 10"

Portrait

X 180.0 180.0 0.00

Y 220.0 218.8 0.55

8" x 10"

Landscape

X 220.0 218.8 0.55

Y 180.0 180.0 0.00

Film size / Direction Desired (mm) Measured (mm)

Deviation obtained from desired value

in %

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 239 of 260

Page 240: CT00-000.814.30.02.02.pdf

240 Specific Camera Information Sony

Sony UP-DF 500 7.1

General Information 0

Block Diagram

Fig. 31:

Available Film Sizes

Performance and Connections

Associations

Maximum number of associations supported: 8

Format Release status Film type Medium type

14" x 17"Portrait Yes UPT-517BL Blue-based

Landscape Yes UPT-517BL Blue-based

14" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

11" x 14"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

10" x 12"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

8" x 10"Portrait n.a. n.a. n.a.

Landscape n.a. n.a. n.a.

Other n.a. n.a. n.a.

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 240 of 260

Page 241: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 241

Performance

Use medical image 4 of patient "Reference Images", maximum film size, positive image, 4x5 format, Copies=1, default center/width, magnification type none.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Line structure is visible. 3% modulation is visible with a lenseonly.

NOTE The UPT-517BL film type requires a specific densitometercharacteristic. Density measurements using the X-Rite 301densitometer are only possible if special optics 301-34SO areinstalled on the X-Rite 301 and calibrated as described in theX-Rite letter.

Known effects

n.a.

SCU / modality software configuration 0

Time 133 sec.

Application Entity

PropertiesDisplay Changes allowed

AE Title UPDF500 Yes

Port number 104 No

HC Device /

General settingsDisplay Changes allowed

Type DICOM printer No

Class

(Printer spec file name)SonyUPDF500 No

DICOM Node Sony UP_DF500 Yes

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 241 of 260

Page 242: CT00-000.814.30.02.02.pdf

242 Specific Camera Information Sony

NOTE Do not modify or add film sizes.

Do not modify pixel sizes.

Delete film sizes not available at the customer site.

SCP / hardcopy camera configuration (provided by OEM) 0

Identification

Filming properties Display Changes allowed

Hold printed film jobs 10 Yes

Min. density 20 Yes

Pixel Size [1/1000 mm] 79.4 x 79.4 No

Film sheet formats,

Portrait

Number of pixels

[columns x rows]

14" x 17"

4224 x 5060

In “define new”

Do not modify/add

Select only

Delete only

Film sheet formats,

Landscape

Number of pixels

[columns x rows]

14" x 17"

5060 x 4224

Medium type Blue film No

Film destination Processor No

Color appearance Gray scale 8 bit No

Background Black Yes

Transformation No magnification No

Manufacturer / device type:

DICOM Printer (SCP)

Device name: Film Station UP-DF 500

Manufacturer: SONY

SW Revision: (for VA20) (for VA26)

MA: 3.04

DR: 3.03

CT: 1.03

IDMAC: 25.0

DICOM: 01.67

MA: 4.00

DR: 4.02

CT: 2.00

IDMAC: 25.0

ICTAG: 1.00

DICOM: 3.21

Physical interface: Ethernet (RJ45)

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 242 of 260

Page 243: CT00-000.814.30.02.02.pdf

Specific Camera Information Sony 243

Printer Setting

Error / warning messages (examples)

Network protocol: TCP/IP

DICOM conformance state-ment:

Rev. 2.0, May 21, 2003.

Image memory: 12 GB

Hard disk: 20 GB

File Name Content Remarks

No file;

Keypad settings, press "Menu"

Information>

IP ADDRESS xxx.xxx.xxx.xxx Network address

No file;

Keypad settings, press "Menu"

PRINT Quality>

SHARPNESS OFF

GAMMA SEL

Default GAMMA

SL32_14C

Interpolation

Linear LUT

No file;

Keypad settings, press "Menu"

Density ADJ>

SELECT POINT

95% ...[25]

95%...[9]

5 % ...[75]

(for MA3.04)

(for MA4.00)

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Cover or door of camera open

COVER OPEN COVER OPEN Camera cover, drawer or door open

Supply magazine not inserted

NO SUPPLY MGZ NO SUPPLY MGZ Film supply not avail-able

Supply magazine empty

EMPTY 14X17 BLUE

EMPTY 14X17 BLUE

14x17 blue film sup-ply empty

Film jam FILM JAM FILM JAM Film jam

Manufacturer / device type:

DICOM Printer (SCP)

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 243 of 260

Page 244: CT00-000.814.30.02.02.pdf

244 Specific Camera Information Sony

Image Quality and Optical Density 0

No specific parameter available

Real Size Scaling

(*1) refer to all messages in the DCS of the hardcopy camera

(*2) refer to all messages in the DCS of the modality

Camera warning or error messages

Printer message Message displayed on SCU (*2)

Defined DCS (*1) Sent by SCP

Film size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

14" x 17"

Portrait

X 330.0 330.0 0.00

Y 380.0 378.5 0.39

14" x 17"

Landscape

X 390.0 388.5 0.38

Y 320.0 320.0 0.00

14" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

14" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

11" x 14"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

10” x 12”

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Portrait

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

8" x 10"

Landscape

X n.a. n.a. n.a.

Y n.a. n.a. n.a.

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 244 of 260

Page 245: CT00-000.814.30.02.02.pdf

Specific Printer Information Epson 245

8- 8Specific Printer Information EpsonEpson AcuLaser C 4200 0

General information 0

Block diagram

Fig. 32:

Available film sizes

Performance

Use medical color image A4, positive image, format 3x4, copies=1, default center/width. The modality has to be configured to CTLUT and the respective PS LUT file for this printer.

The optimum connection (point to point) was used.

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

Format Release status Paper Type Medium type

A4 YesBusiness Laser Copier Paper

80g/m2 Paper

Letter YesBusiness Laser Copier Paper

80g/m2 Paper

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

Time 58 sec

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 245 of 260

Page 246: CT00-000.814.30.02.02.pdf

246 Specific Printer Information Epson

NOTE Paper print-outs are not allowed for medical diagnoses.

NOTE 510K premarket notification/approval not available.

NOTE The Adobe printer driver with the printer-specific descriptionfile has to be used.

The paper printer is not linear in the grayscale imageprint-out. As a result, the special LUT"EPSON_C4200_correction.lut" in the configuration of thePSLUT files has to be used to optimize the image quality ofthe print-out.

NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g. calcium scoring), usethe report of the applications to print out the result images.

Known Effects

NOTE Line structure visible on grey field.

NOTE Saturated colors do not have any contrast (approx 3 steps ofa 17-step test image). Smooth transitions from black to a sat-urated color are not continuous. Effective resolution on theXerox Business Paper is only approx. 200 DPI. Displacementof the primary colors visible (light vertical shift in portrait ori-entation)

Printer Configuration 0

Only the configuration has been performed with the generic “winsteng.exe” and the post-script description file “epalc420ppd” version.

NOTE Version of Adobe Driver: V 5.2

Modality Configuration (Postscript Driver) 0

The default setting of the postscript driver has to be optimized at the modality site.

These settings are:

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 246 of 260

Page 247: CT00-000.814.30.02.02.pdf

Specific Printer Information Epson 247

Printer --> Properties --> Device Settings:

AdobePS EPSON AL-C4200 PS3 Device Settings

--> Form To Tray Assignment.

--> MP Tray: Letter

--> Cassette 1: A4

--> Cassette 2: Not Available

--> Cassette 3: Not Available

--> Cassette 4: Not Available

--> Manual Feed 1st Page: Letter

--> Manual Feed Each Page: Letter

--> Font Substitution Table

--> Arial: Helvetica

--> Courier New: Courier

--> Symbol: Symbol

--> Times New Roman: Times

--> xxxx: <Don’t Substitute>

--> xxxx: For all other fonts of the substitu-tion table

--> Available PostScript Memory: 249 KB

--> Output Protocol: ASCII

--> Send CTRL-D Before Each Job: No

--> Send CTRL-D After Each Job: Yes

--> Convert Gray Text to PostScript Gray: No

--> Convert Gray Graphics to PostScript Gray: No

--> Add Euro Currency Symbol To PostScript Fonts: No

--> Job Timeout: 30 seconds

--> Wait Timeout: 40 seconds

--> Minimum Font Size to Downloads as Outline: 100 pixel(s)

--> Maximum Font Size to Download as Bitmap: 600 pixel(s)

--> Installable Options

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 247 of 260

Page 248: CT00-000.814.30.02.02.pdf

248 Specific Printer Information Epson

--> Installed Memory: 128 MB

--> Lower Cassette Unit: None

Printer --> Printing Preferences --> Advanced: (*1) and

Printer --> Properties --> Advanced --> Printing Defaults --> Advanced:(*2)

AdobePS EPSON AL-C4200 PS3 Advanced Document Settings

--> Paper/Output --> Paper Size: A4

--> Copy Count: 1 Copy

--> Graphic

--> Print Quality: Quality

--> Image Color Management

--> --> ICM Method: ICM Disabled

--> --> ICM Intent: Pictures

--> Scaling: 100%

--> True Type Font: Substitute with Device Font

--> Document Options

--> Advanced Printing Features: Enabled

--> Postscript Options

--> --> PostScript Output Option: Optimize for Speed

--> --> True Type Font Download Option: Automatic

--> --> PostScript Language Level: 3

--> --> Send PostScript Error Handler: Yes

--> --> Mirrored Output: No

--> Printer Features

Color Model RGB

Color Mode Color

RGB Color Mode Vivid

Color Screen Increase Gradation

Pure Black Mode Text

Auto Mono Mode Off

RITech On

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 248 of 260

Page 249: CT00-000.814.30.02.02.pdf

Specific Printer Information Epson 249

Hardcopy Printer Identification (provided by OEM) 0

Identification

Printer Keypad Settings

The default setting of the printer should be used.

Error / Warning Messages (examples)

Toner Save Mode Off

Image Protect Off

Start Page Front

Rotate by 180° Off

(*1/2) Both settings must be identical

Device type: Color Laser Printer

Device name: AcuLaser C4200

Manufacturer: Epson

SW Revision or Firmware Version: OS: 22710 22618 Fonts: 00005208 LAN: 02.20

Number of Input Interface: 3 (Parallel, USB and Network)

Physical Interface: Network: 10/100 Base Tx

USB: Hi-Speed USB 2.0

Parallel: IEEE 1284

Network protocol: TCP/IP 10/100

Image memory: 128 MB

Hard disk: None

Connector for Network: RJ45

Documentation: - User Guide (SAP 14403347 ASD 01S 00)- Admin-istrator Guide (SAP 14403347 ASD 02S 00)

Action to provide error messages of printer Message displayed on Modality

Printer switched off Offline

No paper inserted in magazine,

some print jobs sent

Out of Paper

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 249 of 260

Page 250: CT00-000.814.30.02.02.pdf

250 Specific Printer Information Epson

Image Quality and Optical Density 0

Optical Density

NOTE The PSLUT file has to be configured:

“ EPSON_C4200_correction”

At this time, no reference values have been defined for paper printers.

Densitometer model/type X-Rite 528 was used.

For this measurement result, the Xerox Buisness Laser Copier Inkjet Paper was used.

It is also recommended to add some additional white paper below the print-out to improve results during the measurement.

Before testing the image quality, calibrate the densitometer. Afterwards check the white field of the densitometer in the "paper" mode (dot area). Note the values for V.

Top cover of printer is open,some print jobs sent Door open

Tray for paper not inserted Out of Paper

Action to provide error messages of printer Message displayed on Modality

Values of "Paper" & "Sample" mode V

Offset of the white field in the "Paper" mode (absolute value) 0.06

Value of the white field for density in the "Solid" mode ( relative value)

0.00

StepMeasured Rela-

tive Value

1 0.01

2 0.08

3 0.14

4 0.18

5 0.22

6 0.27

7 0.31

8 0.38

9 0.45

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 250 of 260

Page 251: CT00-000.814.30.02.02.pdf

Specific Printer Information Epson 251

NOTE The PS Printer has to be set to Black&White for this measure-ment

Real Size Scaling

10 0.51

11 0.60

12 0.70

13 0.87

14 1.07

15 1.28

16 1.44

17 1.57

Paper size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

A4

Portrait

X 160.0 159.5 0.31

Y 280.0 280.0 0.00

A4

Landscape

X 280.0 280.5 0.18

Y 160.0 160.0 0.00

Letter

Portrait

X 200.0 199.5 0.25

Y 240.0 239.5 0.21

Letter

Landscape

X 240.0 239.5 0.21

Y 160.0 159.5 0.31

Step Measured Rela-tive Value

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 251 of 260

Page 252: CT00-000.814.30.02.02.pdf

252 Specific Printer Information Hewlett Packard

9- 9Specific Printer Information Hewlett PackardHewlett Packard Color Laser Jet 4700N 0

General information 0

Block diagram

Fig. 33:

Available film sizes

Performance and Connections

Performance

Use medical color image A4, positive image, format 3x4, copies=1, default center/width. The modality has to be configured to CTLUT and the respective PS LUT file for this printer.

The optimal connection (point to point) was used.

Format Release status Paper Type Medium type

A4 YesBusiness Laser Copier Paper

80g/m2 Paper

Letter YesBusiness Laser Copier Paper

80g/m2 Paper

Released: Yes - Film size is available and released

Released: No - Film size is available but not released

Released: n.a. - Film size is not available

Time 23 sec

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 252 of 260

Page 253: CT00-000.814.30.02.02.pdf

Specific Printer Information Hewlett Packard 253

Restrictions

NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).

NOTE Paper print-outs are not allowed for medical diagnoses.

NOTE 510K premarket notification/approval not available.

NOTE For pure black&white images the configuration of the driverfor the printer should be set to black&white to optimize theimage quality on paper.

NOTE No resolution (contrast) in the white area between step 1 and2 of the grayscale image

NOTE Color graphic lines and markers (arrows) are difficult to iden-tify on paper. For CT applications (e.g. calcium scoring), usethe report of the applications to print out the result images.

Known Effects

NOTE Resolution of 600DPI not visually reached on the printout.1-Pixel stripes are smoothed.

NOTE Color displacement of the primaries visible on the printout.

NOTE Line structure visible on flat fields.

NOTE Color gradient from white to a light color is not continuous.

NOTE No contrast at saturated colors.

NOTE The toner can be smeared off the paper, this lightens theimage and affects image quality.

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 253 of 260

Page 254: CT00-000.814.30.02.02.pdf

254 Specific Printer Information Hewlett Packard

Printer Configuration 0

Only the configuration has been performed with the generic “winsteng.exe” and the “hpmcpd25.ppd” postscript description file version.

NOTE Version of Adobe Driver: V 5.2

Modality Configuration (Postscript Driver) 0

The default setting of the postscript driver has to be optimized at the modality site.

These settings are:

Printer --> Properties --> Device Settings:

AdobePS HP Universal PrintingPS Device Settings

--> Form To Tray Assignment.

--> Printer Auto Select: A4

--> Manual Feed in Tray1: Letter

--> Tray 1: Letter

--> Tray 2: A4

--> Tray 3: Letter

--> Tray 4: Letter

--> Tray 5: Letter

--> Tray 6: Letter

--> Tray 7: Letter

--> Tray 8: Letter

--> Tray 9: Letter

--> Ex Tray (MP5): Letter

--> Ex Tray (MP6): Letter

--> Ex Tray (MP7): Letter

--> Ex Tray (MP8): Letter

--> Ex Tray (MP9): Letter

--> Ex Tray (MP10): Letter

--> Envelope Feeder: Letter

--> Font Substitution Table

--> Courier New: Courier

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 254 of 260

Page 255: CT00-000.814.30.02.02.pdf

Specific Printer Information Hewlett Packard 255

--> xxxx: <Don’t Substitute>

--> xxxx: For all other fonts of the substitu-tion table

--> Available PostScript Memory: 6144KB

--> Output Protocol: ASCII

--> Send CTRL-D Before Each Job: No

--> Send CTRL-D After Each Job: Yes

--> Convert Gray Text to PostScript Gray: No

--> Convert Gray Graphics to PostScript Gray: No

--> Add Euro Currency Symbol To PostScript Fonts: Yes

--> Job Timeout: 0 seconds

--> Wait Timeout: 300 seconds

--> Minimum Font Size to Downloads as Outline: 100 pixel(s)

--> Maximum Font Size to Download as Bitmap: 600 pixel(s)

--> Installable Options

--> Accessory Output Bin: Not Installed

--> Printer Memory: 384-494 MB

--> Printer Hard Disk: Installed

--> Job Storage: Enabled

--> Postscript Passthrough: Enabled

--> Copier Mode: Disabled

--> Automatic Configuration: Off

--> HPOutputBinPMLRange: 4-11

--> HPPrnPropResourceData: hpzhl3xr.cab

--> HPInstallableHCO: 3000Stacker-C8084

--> HPMOutputBinHCOMap: Auto

--> HPMOutputBinHCOPMLMap: 0-0

Printer --> Printing Preferences --> Advanced(*1): and

Printer --> Properties --> Advanced --> Printing Defaults --> Advanced(*2):

AdobePS HP Universal PrintingPS Advanced Document Settings

--> Paper/Output --> Paper Size: A4

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 255 of 260

Page 256: CT00-000.814.30.02.02.pdf

256 Specific Printer Information Hewlett Packard

--> Copy Count: 1 Copy

--> Graphic

--> Image Color Management

--> --> ICM Method: ICM Disabled

--> --> ICM Intent: Pictures

--> Scaling: 100%

--> True Type Font: Substitute with Device Font

--> Document Options

--> Advanced Printing Features: Enabled

--> Postscript Options

--> --> PostScript Output Option: Optimize for Speed

--> --> True Type Font Download Option: Automatic

--> --> PostScript Language Level: 2

--> --> Send PostScript Error Handler: Yes

--> --> Mirrored Output: No

--> Printer Features

--> --> Orientation Rotated 180 Degrees: Off

--> --> PostScriptCustomPageSize: Off

--> --> Output Bin: Automatically Select

--> --> Stapling Options: None

--> --> HPCovers: Other Pages

--> --> Print All Text as Black: Disabled

--> --> PrintQualityGroup: Best

--> --> JRHDInstalled: JRHDOff

--> --> HPDocPropResourceData: hpzhl3xr.cab

--> --> JRConstraints: JRCHDFull

--> --> HPSmartHub: _

--> --> Print Color as Gray: Off

(*1/2) Both settings must be identical

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 256 of 260

Page 257: CT00-000.814.30.02.02.pdf

Specific Printer Information Hewlett Packard 257

Hardcopy Printer Identification (provided by OEM) 0

Identification

Printer Keypad Settings

The default setting of the printer should be used.

Error / Warning Messages (examples)

Device type: Color Laser Printer

Device name: Color Laser Jet 4700N

Manufacturer: Hewlett-Packard

SW Revision or Firmware Version: 20051204 46.029.3

Number of Input Interface: 2 (Network and USB 2.0))

Physical Interface: Ethernet and USB-B and IEEE 1284-C-compatible bidirectional parallel interface

Network protocol: TCP/IP and Windows USB 2.0

Image memory: 384 MB

Hard disk: Yes

Connector for Network: RJ 45 / USB-B / Parallel port

Documentation: - “Technical Data” EMEA 08/05 5983-2739DEE

SAP: 14403335 GDB 01S 00

- "HP Color LaserJet 4700 Operator Manual“

SAP: 14403335 ASD 01S 00

- “Configuration page” (can be printed in the printer`s LCD menu)

SAP: 14403335 GDB 02S 00

Action to provide error messages of printer Message displayed on Modality

Printer switched off Offline

No paper inserted in magazine, some print jobs sent Out of paper

Top cover of printer is open, some print jobs sent Door Open

Tray for paper not inserted correctly Offline

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 257 of 260

Page 258: CT00-000.814.30.02.02.pdf

258 Specific Printer Information Hewlett Packard

Image Quality and Optical Density 0

Optical Density

NOTE The PSLUT file has to be configured:

“ HP_LaserJet4700_linear”

At this time, no reference values have been defined for paper printers.

Densitometer model/type X-Rite 528 was used.

For this measurement result, the Xerox Buisness Laser Copier Inkjet Paper was used.

It is also recommended to add some additional white paper below the print-out to improve the results during the measurement.

Before testing the image quality, calibrate the densitometer. Afterwards check the white field of the densitometer in the "paper" mode (dot area). Note the values for V.

Values of "Paper" & "Sample" mode V

Offset of the white field in the "Paper" mode (absolute value) 0.06

Value of the white field for density in the "Solid" mode ( relative value)

0.00

Step Measured Rela-tive Value

1 0.01

2 0.02

3 0.05

4 0.10

5 0.16

6 0.22

7 0.31

8 0.41

9 0.50

10 0.63

11 0.74

12 0.87

13 1.02

14 1.19

15 1.31

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 258 of 260

Page 259: CT00-000.814.30.02.02.pdf

Specific Printer Information Hewlett Packard 259

NOTE The PS Printer has to be set to Black&White for this measure-ment

Real Size Scaling

16 1.54

17 1.92

Paper size / Direction Desired (mm)Measured

(mm)

Deviation obtained from desired value

in %

A4

Portrait

X 160.0 160.0 0.00

Y 280.0 280.0 0.00

A4

Landscape

X 280.0 279.5 0.18

Y 160.0 159.5 0.31

Letter

Portrait

X 160.0 160.5 0.31

Y 240.0 240.5 0.21

Letter

Landscape

X 240.0 240.0 0.00

Y 160.0 160.0 0.00

Step Measured Rela-tive Value

© Siemens, 2008 CT00-000.814.30.02.02 SOMATOM Definition05.09 H WS CS SD CO

Page 259 of 260

Page 260: CT00-000.814.30.02.02.pdf

260 Changes to Previous Version

10- 10Changes to Previous VersionChanges to Previous Version 0

Chapter Section Action

4 Fuji Drypix 4000 Identification updated

4 Fuji Drypix 7000/5000 Identification updated

5 Kodak DryView 6800 Identification updated

7 Sony UP-DF 550 Identification updated

Printer setting updated

7 Sony UP-DF 500 Restrictions updated

Identification updated

Printer setting updated

SOMATOM Definition CT00-000.814.30.02.02 © Siemens, 200805.09 H WS CS SD CO

Page 260 of 260