Source: CTC/CTM © Siemens AG, 1998 Medical Engineering Group, Computed Tomography CT Basics 1 Siemens Medical Systems Computed Tomography (Basics)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 1
Siemens Medical Systems
Computed Tomography (Basics)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 2
The Discovery of X-Rays...
100 years ago,
Wilhelm Conrad Roentgen,
a German scientist,
discovered x-rays...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 3
Look Inside The Human Body...
« But anatomic structures were superimposed « And soft tissue couldn’t be differentiated
This allowed people for thefirst time to be able to viewthe anatomy of the human
body noninvasively
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 4
In 1972,
two scientists -
Hounsfield and Ambrose-
presented the first
clinical CT image ...
CT Broke the Barrier...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 5
♠ But it was time consuming (10 min. / image)
♠ And the resolution needed to be improved
So we couldsee
tomographic anatomy&
density differences
The Human Body Slice by Slice...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 6
What Does a CT Look Like?
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 7
The First Siemens CT Scanner...
Acquisition time 7 min., image matrix 80x80 pixels,scan field 25 cm , spatial resolution 1,3 mm (4LP/cm)
SIRETOM (in 1974)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 8
The Progress in Image Quality...
SIRETOM (1974)
SOMATOM Plus 4 UFC(1996)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 9
What Does a CT Look Like?
♣ Gantry♣ Table♣ Generator♣ Console♣ Computer
From the outside...
Console
Computer
Table
Gantry
Generator
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 10
What Does a CT Look Like?From the inside...
♣ Tube♣ Detector♣ DAS*
Tube
Detector
DAS
* Data Acquisition System
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 11
How Does CT Work?
Recon & postprocessingRecon & postprocessing
Data acquisitionData acquisition
X-ray generationX-ray generation
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 12
Image Generation - The “Slice”X-rays pass through a collimator therefore onlypenetrating an axial layer of the object, calleda "slice"
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 13
The slice is artificiallydivided into small volumeelements called "voxels"with a square base,inside which theattenuation is measuredas a constant value.And in plane, the pictureelements are called “pixels”
Image Generation - The “Voxel”
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 14
The attenuation of radiation values is measured,encoded and transferred to a computer.
Image Generation - The “Matrix”
35 36 3934 3331 34 33 35 3231 80 85 9078
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 15
The numerical matrix is converted into a black andwhite image in a corresponding gray scale.
CT Image Generation - A/D/A*
35 36 3934 3331 3433 35 3231 80 85 9078
*Analog - Digital - Analog
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 16
Criteria for CT Image Quality
CT Image Quality
Spatial resolution
Contrast Detectability
Artifacts
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 17
Spatial Resolution...The ability to resolve High Contrast Objects,(also called “High Contrast Resolution”)
This is influenced by system geometry,and determines image definition
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 18
Definition
Scan time
AlgorithmOperating mode
Slice thickness
Image display
Definition...
Image definition means the sharpness of anobject relative to surrounding tissue. Itdepends on:
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 19
When Small Contrast Differences are crucial(also called “Low Contrast Resolution”)
This is influenced by image definition & noise
Contrast Detectability...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 20
Noise is superimposed on the image and results ina "grainy" impression, as is the case with poor TVreception.
Noise?
206 mA 60 mA
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 21
Noise...
Noise
mAs
kV
Algorithm
Slice thickness
Patient size
Operating mode
Image display
Image “noise” is determined by the number ofx-ray quanta that reach the detector and thencontribute to the image. It depends on:
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 22
Artifacts...The various structures or patterns that appear in aCT image, but are not found in the original object.
Artifacts
Scan time
Operating modeSlice thickness
They depend on: ...
Patient
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 23
Influences on CT Image Quality?
User
PatientSystem
CT Image Quality
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 24
System Efficiency...
100 mAs 100 mAs
The complete system design determines howefficiently x-rays are finally converted toelectrical signal as the detector output, afterpassing through the patient.
The generator, tube, geometry, filtration,collimation and detector design all play a role.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 25
mAs...
Noise
high
low
highlow
Rule of thumb:The higher the dose,the lower the noise.
Rule of thumb:The higher the dose,the lower the noise.
mAs
noisedose
1
* Noise reduced by factor 1.4while dose is doubled.
Tube current and scan time, determine dose.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 26
Low mAs High mAs
mAs...
Image 1: Low mAs value - high noise level
Image 2: 4 times the mAs value- half the noise level
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 27
In the case of a soft tissue study, it is
most important to keep noise to a minimum
by using higher mAs. The lower the noise
level, the easier it is to recognize structures
with minute differences in density. But for
bone or lung studies, higher mAs is not
necessary.
Recommendations for mAs...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 28
Tube Voltage - kV
The dose level, depends very strongly on thevoltage applied to the tube.
Water
20 cm
kV Relative dose
140 100%
120 58%
80 12%
kV Relative dose
140 100%
120 58%
80 12%
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 29
Tube Voltage - kVThe higher the voltage, the more the radiationspectrum is shifted to a higher energy level,resulting in decreased radiation attenuation. Thisis most noticeable in bone and contrast media.140 kV 80 kV
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 30
Algorithms...
Image definition
Sharp
Smooth
low HighNoise
ULTRA HIGH
HIGH
STANDARD
SOFT
SOFT DETAIL
... provide the recipe for mathematical imagecalculation
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 31
An edge-enhancing (HighRes) algorithm
produces good edge definition, but also a
high noise level, while a smoothing
algorithm produces a low noise level, but
also poorer edge definition. For routine
studies, a standard algorithm is normally
recommended.
Algorithms
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 32
Soft Algorithm HighRes Algorithm
Algorithms
Soft algorithms provide better contrastdetectability with less noise.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 33
Standard Algorithm HighRes Algorithm
Algorithms
HighRes algorithms provide better spatialresolution, but with more noise.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 34
Selecting Slice Thickness
A thick slicemeans:A thick slicemeans:
A thin slicemeans:A thin slicemeans:
Selecting a suitable slice thickness is a balancebetween edge definition and noise because oftheir mutually offsetting effects.
low noise better contrast resolution poorer edge definition partial volume artifacts
high noisepoorer contrast resolutionbetter edge definitionno partial volume artifacts
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 35
Slice Thickness...
Slice thickness
Noise
10 mm
High
Low
1 2 3 5
Slice thickness
Image definition = spatial resolution
High
Low
10 mm1 2 3 5
slice thickness 1/2 noise level 1.4no. of x-ray quanta 50%
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 36
3 mm Slice 10 mm Slice
Slice Thickness...
Thicker slices give less noise & bettercontrast detectability for soft tissue
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 37
5 mm Slice 1 mm Slice
Slice Thickness
Thinner slices give better spatial resolutionfor bony structures.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 38
Patient Size...
Noise
400%
300%
200%
100%
30 34 38 42 46 cmPatient diameter
Rule of thumb:The noise level doubles for every 8 cmincrease in patient diameter.
Rule of thumb:The noise level doubles for every 8 cmincrease in patient diameter.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 39
Patient Size...
45 cm 28.8 cm
An attenuation by a factor of 2 results fromeach 4 cm increase in patient thickness, thusincreasing the pixel noise.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 40
Water
Mamma
Air
Bone
Spleen
Fat
Pancreas
Lung
Kidneys
AdrenalGland
Blood
Heart
Liver
Intestine
Tumor
Bladder
3000
60
40
0
-100
-200
-900
-1000
Image Display - Windowing
This is the so-called CT number in Hounsfield unit (HU)
Rule of thumb:
The CT value of water is 0 andair -1000. The relative values ofthe other tissues are calculatedrelative to that of water
Rule of thumb:
The CT value of water is 0 andair -1000. The relative values ofthe other tissues are calculatedrelative to that of water
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 41
The range of CT density values is defined from -1000 to+3000, but the human eye can distinguish only 30 - 40 grayscales at best.
So, the window settings must be in accordancewith the structures to be visualized
Lung Window Mediastinum Window
Image Display - Windowing
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 42
Window width (W): the density range represented within the gray scale.
Window center (C): the center of the density range.
Image Display - WindowingHounsfieldunit
+3000
-1000
0
Window width W
Window center C
Gray scaledisplay
White
Black
CT Windowing
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 43
Narrow Window Width Broad Window Width
Narrow window width:High-contrast image, butstructures outside that windowrange may be inadequatelyrepresented or overlooked.
Broad window width:Minor density differencesappear homogeneous andmay be masked.
Image Display - Windowing
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 44
Hounsfieldunit
+3000
-1000
0
Window 1
Gray scaledisplay
White
Black
CT WindowingWindow 2
Double Window Technique
This is used for the display of two tissue typesdiffering greatly in their density values, such asthe lung & the mediastinum.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 45
Double Window Technique*
To see bothlungs andmediastinumwithin imagesimutanously
*Double window is not recommended for diagnosis.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 46
Lung Window Mediastinum WindowDouble Window
Image Display - Windowing
Image 2:Only lungvisible
Image 1:Both lung,thorax wall &mediastinumvisible
Image 3:Only thorax wall& mediastinumvisible
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 47
Extended CT ScaleNormally, CT values can be measured from -1024 to +3071,but with SOMATOM Plus 4, this can be extended from -10240to +30710 to visualize metals of high attenuation. Therefore,it is always possible to measure the real CT value no matterwhere and how the window is positioned.
Post operative femoralhead replacement - theCT value is 6000 HU
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 48
Reviewed Magnified
Review vs. Magnification
Magnification:a purely opticalmagnification of imagedata which may result inblurred appearance
Review:A zoom reconstruction fromraw data to enhancesharpness of details
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 49
Image Artifacts - Origins
Image Artifacts
Motion
Technicaldefects
Partial volume Beam hardening
Metal
Operatorerror
...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 50
Image Artifacts - Appearance
Image artifacts
Streak Cupping
Dark bar Ring
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 51
Motion Artifact & Scan TimeMotion artifacts
Severe
Moderate
Short LongScan time (s)
Rule of thumb:The shorter the scan time, theless likely motion artifacts areoccur.
Rule of thumb:The shorter the scan time, theless likely motion artifacts areoccur.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 52
Motion Artifact & Correction
w/o correction w/ correction
Motion artifacts can be compensated for bythe Motion Correction Algorithm (MCA)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 53
Metal Artifact
Metals, such as gold, absorbx-radiation almost completely,thus producing
“radiation shadows”,leading to pronounced streakartifacts over the entirereconstructed image
This can only be avoided via a gantry tilt that excludesthe disturbing metallic objects from the slice plane.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 54
Partial Volume Effect
5 mm
Streak-like artifacts, also called partial volume artifact,occur most frequently in the bony structuresat the base of the skull and the petrous bone region.
That is because thevery dense structures(bones) are onlypartially included in theslice, resulting in highcontrast errors.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 55
5 mm 2 mm
Selecting a thinner slice prevents such artifacts fromoccurring, since high contrast structures are lessfrequently partially included, but this inherentlyincreases the noise level, thus degrading contrastresolution.
Partial Volume Effect
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 56
VAR - Volume Artifact Reduction2 x 2 mm
Combines several thin slices (which reduces the partialvolume artifact) to provide a thicker slice (which reducesthe pixel noise and offers good soft tissue discrimination).
5 mm
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 57
Beam Hardening EffectThe x-ray photons emitted from the x-ray tube donot all have the same energy. As they penetratethe irradiated object, the spectrum is shifted tohigher energies - called “beam hardening”. In theimage, streak artifacts or the so-called “cuppingeffect” can be seen.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 58
Beam Hardening & Correction
w/o correction w/ correction
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 59
Beam Hardening & CorrectionThe “cupping” effect can be compensated forby means of “beam hardening correction”
w/o correction w/ correction
severe cupping homogenuous CT values
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 60
Technical DefectsThe individual detector elements of a detectorsystem may not produce the same signal for thesame irradiation.
When a detector elementoutputs an erroneous signal,ring artifacts appear.
This can be eliminated bycalibration, if not, calltechnical service!
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 61
Adaptive Filter
w/o A.F. w/ A.F.
For a non-circular object, x-ray attenuation isgreater along the long axis than along the shortaxis, therefore directional noise is seen.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 62
Artifacts & CorrectionsThe reasons for artifacts are quite diverse. Whatwe do is to perform corresponding corrections inorder to avoid them. But sometimes artifacts cannot be compensated for completely.
Nothing is perfect...
But we keep on working to reducethem as much as possible!
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 63
Since the influence on HC resolution & LCresolution by changing parameters can becontradictory, it is necessary to differentiatebetween:
For the Clinical Routine...
¨ A Soft tissue study (contrast detectability) ( >90% of routine studies, normally)
¨ A Bone study (spatial resolution) (<10% of routine studies, normally)
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 64
CT is really great!
But there are still some problems...
Clinical Requirements...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 65
Conventional CT...
I.S.D* I.S.D
I.S.D I.S.D I.S.D I.S.D
* Inter Scan Delay
Standard Scan: - Longer cycle time - but instant display
Dynamic Scan: - Fast acquisition - but delayed display
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 66
Moderate inspiration
Misregistration due to different levels of respiration
Problems of Conventional CT...
Deep inspiration
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 67
Breathing Was the Problem...
Sorry, my dear patient, thelesion was missed becauseyou did not keep the sameinspiration level for eachscan...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 68
Best case
Worst case
Problems of Conventional CT...Partial Volume Effect and slice location causemisregistration and/or misdiagnosis
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 69
Location Was the Problem...
Sorry, the lesion wasmissed because its locationdid not fit my sliceposition...
One mouse...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 70
Deep inspiration Moderate inspiration
l Scan influenced by respiration
l Image can’t be reconstructed anywhere as desired
Problems of Conventional CT...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 71
How can we improve this?
l Scan influenced by respiration
l Image can’t be reconstructed anywhere as desired
Make it as fast as possible within onebreathhold
Change "slice" into "volume"
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 72
w/o I.S.D w/ I.S.D
2D Slice 3D Volume
Conventional vs. Spiral CT ...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 73
Continuously rotating tube/detector system
Continuous radiation
Continuous data acquisition
Continuous table feed
Spiral CT = Volume scan
The 4 “C”s?
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 74
“You said that images will contain artifacts if thereis any motion during the scan, but a spiral scan isperformed during constant table movement!”
Yes, therefore we need special techniques forimage reconstruction:
Spiral AlgorithmsWIDE, SLIM & SLIM 2
Spiral CT...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 75
Wide (360°) Algorithm2 x 360° (2 full rotations) data acquisition is usedfor 1 image reconstruction. Both are measured data.
measureddata
table positionslice
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 76
Slim/Slim 2 (180°) Algorithm2 x 180° (2 half rotations) data acquisitions areused for 1 image reconstruction. Measured dataare weighted for Slim, and interpolated for Slim 2.
measureddata
table positionslice
complementarydata
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 77
Different Interpolations?
♣ The in-plane resolution for a homogenousobject in Z-direction is the same for WIDE,SLIM & SLIM 2
♣ SLIM & SLIM 2 give better Z-directionresolution than WIDE when pitch 0.5.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 78
Effective Slice Thickness (FWHM*)
WIDE makes FWHM wider, while SLIM/SLIM 2 keepFWHM narrow.
0 0.5 1 1.5 2 2.5 30
0.5
1
1.5
2
2.5
3
3.5
WIDE
SLIM
Pitch
Slic
e T
hic
knes
s (r
elat
ive)
* Full Width Half Maximum
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 79
Noise AmplitudesCompared with Sequence scans, WIDE gives less noise, andSLIM/SLIM 2 give more. But for all of them, noise amplitudedoes not depend on pitch.
0 0.5 1 1.5 2 2.5 30
0.2
0.4
0.6
0.8
1
1.2
Pitch
No
ise
(rel
ativ
e)
WIDE
SLIM / SLIM 2SEQ.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 80
Noise Amplitudes
WIDE
Image reconstructed with WIDE algorithm shows less noisethan those with SLIM & SLIM 2, but for SLIM & SLIM 2they are about the same.
SLIM 2SLIM
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 81
Slim: use it with HighRes algorithm when onlyspatial resolution is most important (e.g. bone andlung study)
Slim2: use it with algorithms from smooth tostandard when the contrast detectability is mostimportant (e.g. soft tissue study)
Wide: use it to achieve low noise only if z-directionresolution is not important.
Which Algorithm to Choose?
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 82
Spiral ParametersPitch: Table feed per rotation divided by slice thickness
Pitch =Table feed / rotation
Slice thickness
It is physically not possible to scan a volumewith gaps because x-rays always irradiate thewhole volume.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 83
Increment:Determines degree of overlap between successiveimages.
Increment = 5 mmIncrement = 3 mm
*slice = 10 mm
Spiral Parameters
Increment = 10 mm
The smaller the increment, the morethe images are overlapped.
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 84
u Scan a whole volume in one breath hold
u Reduce partial volume effects
u No gaps
u Overlapping images can be reconstructed without additional dose
u High quality data for 3D-rendering
Advantages of Spiral CT...
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 85
Why Perform Spiral CT?u Fast scanning of large anatomical volumesu Gapless data acquisition during one breathholdu Optimum utilization of contrast mediumu Retrospective reconstruction with arbitrary slice increments
Overlapping reconstruction givesbetter z-axis resolution
Spiral CT ScanConventional CT Scan
Source: CTC/CTM© Siemens AG, 1998Medical Engineering Group, Computed TomographyCT Basics 86
u Use Spiral CT for contrast studiesu Use Spiral CT for all regions of the bodyu Use it for pediatrics and trauma patients, who
require quick scanningu Use it for long anatomical ranges
Always use Spiral if you intend toperform 3D postprocessing (e.g. CTA) !
When to Use Spiral CT?