COMPUTED TOMOGRAPHY INSTRUMENTATION AND

OPERATION

Nchanji NKEH [email protected]/[email protected]

671459765/662695118

B.TECH/HPD, MDIRT

CHM Virtued Academy India (Ongoing)

Leader

Radiology Department

Level 300 MDI

2016/2017 Academic Year

ST. Louis UNIHEBS mile 3 Nkwen ,Bamenda- Cameroon

1

Page 2: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMPUTED TOMOGRAPHY INSTRUMENTATION AND OPERATION

2

Page 3: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

3

Page 4: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

4

Page 5: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

5

Page 6: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

6

Page 7: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

7

Page 8: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

8

Page 9: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

9

Page 10: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

10

Page 11: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

11

Page 12: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

12

Page 13: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

13

Page 14: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

14

Page 15: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

15

Page 16: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

OUTLINE CT SYSTEM COMPONENTS – DEFINITION OF

A SCANNER SCANNER COORDINATE SYSTEM – XYZ,

ISOCENTER IMAGING SYSTEM COMPUTER SYSTEM DISPLAY, RECORDING, AND STORAGE

SYSTEMS

16

Page 17: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT MAIN SYSTEMS

IMAGING SYSTEM COMPUTER SYSTEM DISPLAY, RECORDING, STORAGE SYSTEM DATA ACQUISITION SYSTEM

17

Page 18: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT SYSTEMGANTRY

DETECTORS

SAMPLE HOLD UNIT

ADC

ARRAY PROCESSOR

HOST COMPUTER

STORAGECONSOLE

SCAN CONTROLLER

DAC

GANTRY CONTROL

HIGH VOLTAGE GENERATOR

X-RAY TUBE18

Page 19: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SCANNER 19

Page 20: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

Some CT hardware 20

Page 21: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SCANNER

GANTRY PATIENT COUCH

21

Page 22: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

22

Page 23: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

23

Page 24: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

24

Page 25: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

GANTRY HOUSES:

X-RAY TUBE GENERATOR (LOW VOLTAGE DESIGN) COLLIMATORS DETECTORS

25

Page 26: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

GANTRY CHARACTERISTICS

APERTURE TILTING RANGE

26

Page 27: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

MOST OF THE SCANNERS HAVE 70CM APERTURE

27

Page 28: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

70 CM

28

Page 29: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COORDINATE SYSTEM

X

29

Page 30: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COORDINATE SYSTEM

Y

30

Page 31: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COORDINATE SYSTEM

Z

31

Page 32: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

ISOCENTER 32

Page 33: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

TILTING RANGE OF MOST SCANNERS- +30 TO -30

DEGREES

33

Page 34: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

PATIENT COUCH :

450 LBS (204 KG) DISTRIBUTED WEIGHT LIMIT

34

Page 35: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SCANNABLE RANGE:

COVERAGE FROM HEAD TO THIGH (162CM)

35

Page 36: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

MAX. SCANNABLE RANGE 36

Page 37: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGING SYSTEM

PRODUCTION OF X-RAYS SHAPING OF X-RAY BEAM ENERGY FILTERING X-RAY BEAM

37

Page 38: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGING SYSTEM COMPONENTS

X-RAY TUBE GENERATOR –HIGH VOLTAGEHIGH VOLTAGE COLLIMATORS FILTER DETECTORS DETECTOR ELECTRONICS

38

Page 39: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

X-RAY TUBE AND X-RAY PRODUCTION

39

Page 40: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

40

Page 41: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CATHODE --------MADE OF TUNGSTEN

IN CT – STILL SMALL AND LARGE

41

Page 42: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

THERMIONIC EMISSION

CATHODE HEATED UP TO AT LEAST 2,200 DEG. CELSIUS TO LIBERATE ELECTRONS FOR TRANSIT TO ANODE

42

Page 43: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

FOCAL SPOT- CT UTILIZES DIFFERENT FOCAL SPOTS

THE FILAMENT SIZE – LENGTH – FOCAL SPOTFOCAL SPOT

SMALLER FOCAL SPOT - Low mA

SMALLER FOCAL SPOT – sharper image

43

Page 44: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

ANODE +++++ MADE OF TUNGSTEN AND MOLYBDENUM

TUNGSTENTARGET

TARGET MADE OF TUNGSTEN AND

RHENIUM

44

Page 45: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

mA – tube current

The number of electrons flowing from cathode to anode

45

Page 46: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

kVp

Potential difference between cathode and anode (Volts) kilo means 1,000 x.

46

Page 47: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

S –time of exposure

mAs tube current for certain length of time

47

Page 48: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

X-RAY PRODUCTION RESULTS IN A LOT OF HEAT AND VERY LITTLE X-RAYS BEING GENERATEDHEAT UNITS CALCULATION

HU= kVp X mA x time

MOST CT TUBES HEAT CAPACITY 3-5 MILLION HU

48

Page 49: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

REDUCTION OF HEAT UNITS – TECHNIQUE COMPENSATION kVp

mA

Time

INCREASED NOISE

49

Page 50: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

TOO LOW OF kVp:

NOISE !!!!NOISE !!!!

50

Page 51: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

X-RAY EMISSION 51

Page 52: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

TUBE CURRENT CHANGE

CURRENTINTENSITY

ENERGY – NO CHANGE

2 * mA = 2 * number of photons4 * mA = 4 * number of photons

52

Page 53: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

Why changing mA or time Avoiding motion – mA time Pediatric technique modification Reducing noise - mAs

NOISE MOTION

53

Page 54: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

Tube voltage (kVp) CHANGE

kVp

INTENSITY -

ENERGY –

15% INCREASE OF KVP = 2 * mAs

54

Page 55: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

kVp IN CT 80-140 TOO LOW – NOISE (NOT ENOUGH PENETRATION OF THE PATIENT ) PHOTON STARVATION - NOISE!!!!!PHOTON STARVATION - NOISE!!!!!

55

Page 56: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

HIGH VOLTAGE GENERATOR –(HVG)

GENERATES HIGH VOLTAGE POTENTIAL BETWEEN CATHODE AND ANODE OF AN X-RAY TUBE

56

Page 57: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT GENERATOR

5-50 kHz 30-60 kW

KVP SELECTION:

80, 100, 120, 130,140

mA selection:

30, 50, 65, 100, 125, 150, 175, 200, 400

57

Page 58: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COLLIMATION IN CT

ADC

PRE-PATIENT COLLIMATION

POST-PATIENT COLLIMATION

58

Page 59: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

BASIC DATA AQUSITION SCHEME IN CT

ADC

FILTRATION

59

Page 60: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

FILTRATION CHANGE

FILTRATION

INTENSITY

ENERGY –

60

Page 61: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

FILTRATION MATERIAL ALUMINIUM ( SPECIAL FILTER IN CT)

BOWTIE

TO MAKE THE BEAM HARDER AND MORE MONOENERGETIC

61

Page 62: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

Filter

Patient

DEFINES SLICE THICKNESS

REDUCES SCATTER RECHING THE PATIENT

Detector

62

Page 63: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT DETECTORS 63

Page 64: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CONCEPT OF MDCT 64

Page 65: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SDTCT AND MDCT 65

Page 66: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

DETECTOR TYPES: SCINTILLATION

PM TUBE

S. CRYSTALS. CRYSTAL

PHOTODIODE

66

Page 67: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SCINTILLATION CRYSTALS USED WITH PM TUBES: SODIUM IODIDE –AFTERGLOW + LOW DYNAMIC AFTERGLOW + LOW DYNAMIC

RANGE RANGE ( USED IN THE PAST)( USED IN THE PAST)

CALCIUM FLUORIDE

BISMUTH GERMANATE

67

Page 68: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

S. CRYSTAL USED WITH PHOTODIODE

CALCIUM TUNGSTATE RARE EARTH OXIDES - CERAMIC

68

Page 69: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

DETECTOR TYPE: GAS IONIZATION

XENON GAS

30 ATM

69

Page 70: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

EFFICIENCY OF DETECTORS- QDE

SCINTILLATION – 95% - 100%- COMMONLY COMMONLY USED IN III & IV GENERATION SCANNERSUSED IN III & IV GENERATION SCANNERS

GAS – 50% - 60%

70

Page 71: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMPUTER SYSTEM

RECONSTRUCTION AND POSTPROCESSING CONTROL OF ALL SCANNER COMPONENTS CONTROL OF DATA ACQUSITION, PROCESSING,

DISPLAY. DATA FLOW DIRECTION

71

Page 72: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMPUTER SYSTEM IN CT

MINICOMPUTERS

72

Page 73: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMPUTER SYSTEM COMPOSED OF:

HARDWARE SOFTWARE

73

Page 74: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMPUTER PROCESSING IN CT

SEQUENTIAL PROCESSING MULTITASKING MULTIPROCESSING

74

Page 75: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SOFTWARE –PROGRAM (S) HELPING CT USER TO COMMUNICATE WITH THE CT SYSTEM

75

Page 76: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT OPERATING SYSTEM-PROGRAMS THAT CONTROL THE HARDWARE COMPONENTS AND THE OVERALL OPERATION OF THE CT COMPUTER

76

Page 77: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT OPERATING SYSTEM

UNIX WINDOWS

77

Page 78: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

HOST COMPUTER

CONTROL OF ALL COMPONENTS CONTROL OF DATA ACQUSITION, PROCESSING,

DISPLAY. DATA FLOW DIRECTION

78

Page 79: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

ARRAY PROCESSOR

TAKES DETECTOR MEASUREMENTS FROM HUNDREDS OF

PROJECTIONS. RESPONSIBLE FOR RETROSPECTIVE

RECONSTRUCTION AND POSTPROCESSING OF DATA.

THE MORE PROCESSORS IN THE COMPUTER THE SHORTER THE RECONSTRUCTION TIME

79

Page 80: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

DATA ACQUISITION SYSTEM (DAS)

SET OF ELECTRONICS BETWEEN DETECTORS AND HOST COMPUTER.

IT CONTAINS: AMPLIFIER, ADC, DAC, GENERATOR, S/H.

80

Page 81: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

AMPLIFIER

SIGNAL FROM DETECTORS GOES TO AMPLIFIERS FOR SIGNAL MAGNIFICATION AND THEN IS SENT TO SAMPLE/HOLD UNIT

81

Page 82: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

ADC

CONVERTS ANALOG SIGNAL OUTPUT FROM THE SCANNING EQUIPMENT TO A DIGITAL SIGNAL SO IT CAN BE PROCESSED BY A COMPUTER.

82

Page 83: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

SAMPLE/HOLD UNIT (S/H)

LOCATED BETWEEN AMPLIFIERAMPLIFIER AND ADC ADC PERFORMS SAMPLING AND ASSIGNS SHADES OF GRAY TO THE PIXELS IN THE DIGITAL MATRIX CORRESPONDING TO THE STRUCTURES

83

Page 84: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

DASGANTRY

DETECTORS

S/H

ADC

ARRAY PROCESSOR

HOST COMPUTER

STORAGECONSOLE

SCAN CONTROLLER

DAC

GANTRY CONTROL

HIGH VOLTAGE GENERATOR

X-RAY TUBE84

Page 85: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGE DISPLAY, RECORDING, STORAGE

DISPLAYS IMAGE ( OUTPUT FROM COMPUTER) PROVIDES HARD COPY OF THE IMAGE FACILITATES THE STORAGE AND RETRIEVAL OF

DIGITAL DATA COMMUNICATES IMAGES IN THE NETWORK

85

Page 86: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGE DISPLAY

86

Page 87: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGE RECORDING SYSTEMS(LASER PRINTERS) SOLID STATE

LASER PRINTERS GAS LASER

PRINTERS

87

Page 88: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

HARD COPY 88

Page 89: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

IMAGE STORAGE MEDIA

MAGNETIC TAPES MAGNETO-OPTICAL DISK (MOD) CD

89

Page 90: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

COMMUNICATION

PACS

90

Page 91: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

OVERREAD NETWORK

WHILE MOST TELERADIOLOGY SYSTEMS PURCHASED OVER THE LAST DECADE WERE INTENDED FOR ON-CALL PURPOSES, THE PAST TWO YEARS HAVE SEEN A RAPID INCREASE IN THE USE OF TELERADIOLOGY TO LINK HOSPITALS AND AFFILIATED SATELLITE FACILITIES, OTHER PRIMARY HOSPITALS, AND IMAGING CENTERS. A NUMBER OF THE ENABLING TECHNOLOGIES NEEDED FOR EFFECTIVE OVERREAD NETWORKS, SUCH AS MORE AFFORDABLE HIGH-SPEED TELECOMMUNICATIONS NETWORKS AND IMPROVED DATA COMPRESSION TECHNIQUES, HAVE MATURED IN RECENT YEARS.

91

Page 92: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

NightHawk Radiology Services has developed an innovative approach to the delivery of radiology services by operating centralized, state-of-the-art reading centers in Sydney, Australia and Zurich, Switzerland. Staffing U.S.-trained, board-certified radiologists specializing in emergency radiology, these locations are ideally situated for U.S. care because when it’s the middle of the night in Boston, it’s daytime “Down Under.” When it’s early morning in Los Angeles, it’s daytime in the Alps. From the centralized reading centers, NightHawk radiologists interpret exams and report the results to attending physicians in real-time, usually less than 20 minutes.

92

Page 93: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

CT ROOM LAYOUT 93

Page 94: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

How much does a CT scanner cost?

94

Page 95: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

How much ??4-slice scanner: $85,000 - $115,000

16-slice scanner: $145,000 - $225,000

64-slice scanner: $250,000 - $450,000

Service contracts:$100,000 - $135,000

95

Page 96: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

96

Page 97: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

97

Page 98: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

Lecture delivered with use of resources from ICRP, IAEA,

AAPM and GE

98

Page 99: Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOUIS UNIVERSITY, MILE 3 NKWEN- BAMENDA TOWN, CAMEROON

NEXT LECTURE:

1. ULTRASONOGRAPHIC INSTRUMENTATION

2. MAMMORAPHIC INSTRUMENTATION

3. FLUOROSCOPIC INSTRUMENTATION