Cryocooler comprehensive catalog 172E2018041000TLC Head Office and Factory 1222-1 yabata,Chigasaki,Kanagawa,253-0085,Japan TEL:+81-467-85-9366 FAX:+81-467-85-9356 Kyoto Cryogenic Technology R&D Center/Kyoto Factory 140-1,Megawa,Makishima-cho,Uji,kyoto,611-0041,Japan TEL:+81-774-28-5595 FAX:+81-774-20-2201 Osaka Branch Office Uemura Nissay Bldg.3-3-31 Miyahara,Yodogawa-Ku,Osaka,532-0003,Japan TEL:+81-(0)6-6397-0112 FAX:+81-(0)6-6397-0126 www.ulvac-cryo.com Contact Us ■ The content of this catalog is subject to change without notice for the purpose of performance improvement etc.
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Cryocoolercomprehensive catalog
172E2018041000TLC
Head Office and Factory1222-1 yabata,Chigasaki,Kanagawa,253-0085,Japan
Millimeter wave measurement optical measurement NMR (nuclear magnetic resonance)
ESR / ERP (electron paramagnetic resonance) X ray diffraction Hall effect
SQUID (superconducting quantum interference device) Mössbauer superconducting material
J - J device magnetic cooling surface analysis radio telescope materials research
Compact refrigerator for satellites
The study
Applicat ion of cryogenic technology
1 2
CRYOCOOLER CRYOCOOLER
3 4
φ40
8-M4(φ56) 8-M4(φ33)
(244)
(126)
1stStage 2ndStage
5
147.5
161.5
16(133)
(442)
8
φ64
φ109
(128)
(309)
Flange2-φ6.5(φ96)
22.5゚
(282)
φ52
618
φ71
495
φ109
346
(146)
165
Flange
1stStage8-M4 (φ65)
2ndStage8-M4 (φ42)
2-φ6.5 (φ96)
22.5゚
84
22.5゚
30゚
8-φ5.2(φ170)Flange
1stStage 2ndStage8-M5(φ110) 6-M5(φ54)
(392.5)
φ64
6
φ125φ90
64189.5
φ180φ88.5
(580)
236.5
156
5(187)
SUPPLY
RETURN
(146)
98180
RETURN
SUPPLY
(304)
6.5
6
SUPPLY
RETURN
(146)
98 RETURN
SUPPLY
(304)
(187)
5156
236.5
(580)
φ180
6
189.5
64
φ90φ125
φ68
(392.5)
6.5
22.5゚
30゚
8-φ5.2(φ170)Flange
2ndStage1stStage8-M5(φ110) 6-M5(φ58)
6
180
φ88.5
4K GM refrigerator HE01
4K GM refrigerator HE05
4K GM refrigerator UHE 10
4K GM refrigerator UHE 15
1 stage temperature(K)
2 stage temperature (K)
1 stage temperature(K)
2 stage temperature (K)
1 stage temperature(K)
2 stage temperature (K)
Examples of Performance curve (coldhead pointing-down)
60Hz50Hz
1 stage temperature(K)
2 stage temperature (K)
60Hz50Hz
1 stage temperature(K)
2 stage temperature (K)
1 stage temperature(K)
2 stage temperature (K)
60Hz50Hz
1 stage temperature(K)
2 stage temperature (K)
1 stage temperature(K)
2 stage temperature (K)
60Hz50Hz
02468101214161820
35 45 55 650W 3W 4W
1.5W2.0W
2.5W
5W
1.0W0.3W0.1W
1W
0.2W0.5W
0W 2W02468101214161820
35 45 55 65 7070
0.5W1.0W
0.3W0.2W0.1W0W 1W 2W 4W
1.5W
2.0W
3W0W
2.0
2.5
3.0
3.5
4.0
4.5
5.0
10 20 30 40 50 60 70
0.3W
10W
0.5W
0.1W0W0W 5W
0.7W
15W
2.0
2.5
3.0
3.5
4.0
4.5
5.0
10 20 30 40 50 60 70
0.3W
0.5W
0.1W0W
0.7W
10W5W 15W20W
20 30 40 50 60 70
56789101112
80W60W40W20W0W
8W
23456 4W
1W0W
1 stage load
2 stage load
20 30 40 50 60 70 80 90
56789101112
80W60W40W20W0W
23456 4W
1W
0W
8W
1 stage load
2 stage load
1 stage load
2 stage load
1 stage load
2 stage load
1 stage load
2 stage load
1 stage load
2 stage load
20 40 60 80 100 120
5W
1.5W
0W
80W60W40W
20W0W
10W
1112
23456789101112
23
20 40 60 80 1000W
5W
1.5W
80W60W40W20W0W
10W
1112
234567891011
2
1 stage load 1 stage load
2 stage load
2 stage load
50Hz 60HzPower frequency
126(W) x 244(D) x 442(H)mm Dimensions
10,000 hoursMaintenance interval * 2
7.5kgWeight
Ultimate temperature
Cooldown time (4.2 K) * 1
First stage
Second stage
3W/60K
0.1W/4.2K
Second stage
Refrigerationcapacity※1
<2.7K
5W/60K
140min 120min
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.
Specification
50Hz 60HzPower frequency
146(W) x 282(D) x 495(H)mm
10,000 hoursMaintenance interval * 2
11kg
10W/45K
0.5W/4.2KRefrigerationcapacity※1
<2.7K
10W/40K
75min 60min
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.
Examples of Performance curve (coldhead pointing-down)
Dimensions
Weight
Cooldown time (4.2 K) * 1
First stage
Second stage
Second stage
Specification
50Hz 60HzPower frequency
180(W) x 304(D) x 578(H)mm
10,000 hoursMaintenance interval * 2
17kg
40W/43K
1.0W/4.2K
<2.7K
50W/43K
60min
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.※ Flexible hose standard length 20 A × 10 m
Examples of Performance curve (coldhead pointing-down)
Dimensions
Weight
Cooldown time (4.2 K) * 1
First stage
Second stage
Second stage
Refrigerationcapacity※1
Specification
50Hz 60HzPower frequency
180(W) x 304(D) x 580(H)mm
10,000 hoursMaintenance interval * 2
18kg
35W/50K
1.5W/4.2KRefrigerationcapacity※1
<2.7K
45W/50K
60min
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.※ Flexible hose standard length 20 A × 10 m
Examples of Performance curve (coldhead pointing-down)
Dimensions
Weight
Cooldown time (4.2 K) * 1
First stage
Second stage
Second stageUltimate temperature
Specification
Ultimate temperature
Ultimate temperature
4K GM CRYOCOOLERS 4K GM CRYOCOOLERS
5 6
φ39.7
300.0
185.9
157.2
282.5
468.4
223.8199.9147.7
67.5
φ76.2
8-φ5.5φ141.3
4-M4φ33.3
φ152.4
6-M4φ66.8
347.5
185.9
271.3247.4195.2
67.5
182.0
333.0
518.9
φ57
φ89
φ152.4
8-φ5.5φ141.3
6-M4φ44.5
8-M4φ79
185.9
247.4195.2281.5
182.0
φ57
71.5
φ152.4
φ108.8
357.7
8-φ5.5φ141.3
518.9
333.0
6-M4φ44.5
12-M4φ100.6
347.5
185.9
271.3247.4195.2
67.5
182.0
313.0
499.0
φ57
φ152.4
φ89
8-φ5.5φ141.3
6-M4φ44.5
8-M4φ79
347.5
185.9
67.5
182.0
271.3247.4195.2
367.9
φ152.4
φ89
8-φ5.5φ141.3
8-M4φ79
185.9
247.4195.2281.5
186.3372.2
71.5
φ152.4
357.7
φ108.8
8-φ5.5φ141.3
12-M4φ100.6 φ108.8
φ152.4
264.7281.5316.9
403.1
224.3
12.7(178.8)
SUPPLY
70.3
30゚
30゚
30゚
30゚
8-φ5.5φ141.3
393.1
12-M4φ100.6
300.0
185.9
157.2343.1
223.8199.9147.7
67.5
φ152.4
φ76.2
8-φ5.5φ141.3
φ66.86-M4
φ152.4
φ106φ68
50
50.5
φ52
6
5
173.6
φ180
347.5
200
147.5
6.5
(324)
(189)
(537)
(148)
(183)
Flange8-φ5(φ170)
2ndStage6-M5(φ44)
1stStage8-M5(φ98)
6.1
22.5゚
22.5゚
30゚
4K GM refrigerator UR4K0420T GM refrigerator RM series
GM Refrigerator RMS Series
1 stage temperature(K)
2 stage temperature (K)
1 stage temperature(K)
2 stage temperature (K)
60Hz50Hz
2.0
2.5
3.0
3.5
4.0
4.5
5.0
10 20 30 40 50 60 70 80 90 100 110
0.3W
35W
15W
0.45W
0.15W
0W
0.6W
25W0W
2 stage load
1 stage load
2.0
2.5
3.0
3.5
4.0
4.5
5.0
10 20 30 40 50 60 70 80 90
0.3W
10W
0.45W
0.15W
0W0W
0.6W
40W20W 30W
2 stage load
1 stage load
50Hz 60Hz
183(W) x 324(D) x 537(H)mm
10,000 hours
13kg
15W/40K
0.45W/4.2K
<3.5K
20W/40K
70min 60min
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.※ Flexible hose standard length 20 A × 10 m
* Refrigerating capacity, reachable temperature, cooling down time are the combination when the refrigerator and compressor are 1: 1 / ambient temperature: value at 24 ° C / no load. * When the power frequency is 60 Hz.
RM series (2 stage type) RM10 RM20 RM50 RM80TRefrigerator system
* Refrigerating capacity, reachable temperature, cooling down time are the combination when the refrigerator and compressor are 1: 1 / ambient temperature: value at 24 ° C / no load. * When the power frequency is 60 Hz.
First stageUltimate temperatureCooldown timeWeight (refrigerator)
RM series (1stage type)
Examples of Performance curve (coldhead pointing-down)
Power frequency
Dimensions
Maintenance interval * 2
Weight
Ultimate temperature
Cooldown time (4.2 K) * 1
First stage
Second stage
Second stage
Refrigerationcapacity※1
4K GM CRYOCOOLERS GM CRYOCOOLERS RM/RMS SERIES
11 12
Solvay refrigerator S 050 Compressor unit
Dimensions
Maintenance interval * 2
Weight
Cooldown time (77K) * 1
* 1 Value at no load. The performance depends on the installation environment of the refrigerator and the cooling object.* 2 Maintenance interval is recommended, not guaranteed time.
50Hz 8 W @ 77 K 12 W @ 77 K 4.0 W @ 77 K 9 W @ 77 K 10 W @ 77 K
60Hz 8 W @ 77 K 4.5 W @ 77 K
1 W @ 20 K
1 W @ 20 K
5 W @ 20 K
5 W @ 20 K
9 W @ 77 K 10 W @ 77 K
50Hz -
-
-
-
-
12 W @ 77 K
-
10 K
60Hz
50Hz 40 K 40 K
40 K
10 K
10 K
10 K
30 K
30 K
60Hz 40 K
50Hz <30 min (20K) <45 min (20K)
60Hz <30 min (20K) <50 min (20K)
<20 min (77K) <30 min (77K)
<20 min (77K) <30 min (77K)
<30 min (77K)
<30 min (77K)
1.5 kg 4.5 kg 2.5 kg 6.3 kg 3 kg
○
○
○
○
○
○
○
○
○
○
SA112
SW112
SA115
SW115
UW404
Type expression
C30HVRT
C30HMVRT-380
C30VR
C30PVRT
C10
First stagerefrigerationcapacity
Second stagerefrigerationcapacity
Cooldown time
Reachingtemperature
Weight
GM (RM Series)
RM20 RM50 RM80T RMS10 RMS50 RMS80T RMS150T
50Hz 150 W @ 77 K
60Hz 165 W @ 77 K
50Hz -
-
40 K
40 K
60Hz
50Hz
60Hz
50Hz
60Hz
<15 min (77K)
<13 min (77K)
20 kg
30 W @ 77 K 55 W @ 77 K 80 W @ 77 K 29 W @ 77 K 67 W @ 77 K 120 W @ 77 K
35 W @ 77 K 65 W @ 77 K 100 W @ 77 K 35 W @ 77 K 85 W @ 77 K 130 W @ 77 K
10 W @ 20 K 5 W @ 20 K 8 W @ 20 K ---
---
35 K30 K30 K
35 K30 K30 K
12 W @ 20 K 7 W @ 20 K 8 W @ 20 K
12 K 12 K 12 K
12 K 12 K 12 K
<35 min (20K) <45 min (20K) <40 min (20K) <35 min (77K) <25 min (77K)
<30 min (20K) <40 min (20K) <35 min (20K) <30 min (77K) <20 min (77K)
<20 min (77K)
<18 min (77K)
14 kg 14 kg 16 kg 10 kg 13 kg 15 kg
○ ○
○
○
○
○ ○
SA112
SW112
SA115
SW115
UW404
Type expression
C30HVRT
C30HMVRT-380
C30VR
C30PVRT
C10
First stagerefrigerationcapacity
Second stagerefrigerationcapacity
Cooldown time
Reachingtemperature
Weight
4K GM 20K GM GM (RM Series)
HE01 HE05 UHE10 UHE15 UR4K0420T RSC50T RM10
50Hz 3 W @ 60 K 10 W @ 45 K 40 W @ 43 K 35 W @ 50 K 15 W @ 40 K 40 W @ 20 K 10 W @ 77 K
60Hz 5 W @ 60 K 10 W @ 40 K 50 W @ 43 K 45 W @ 50 K 20 W @ 40 K 45 W @ 20 K 13 W @ 77 K
3 W @ 20 K
3.5 W @ 20 K
50Hz 0.1 W @ 4.2 K 0.5 W @ 4.2 K 1.0 W @ 4.2 K 1.5 W @ 4.2 K 0.45 W @ 4.2 K -
60Hz 0.1 W @ 4.2 K 0.5 W @ 4.2 K 1.0 W @ 4.2 K 1.5 W @ 4.2 K 0.45 W @ 4.2 K -
50Hz 2.7 K 2.7 K 2.7 K 2.7 K 3.5 K 13 K 12 K
60Hz 2.7 K 2.7 K 2.7 K 2.7 K 3.5 K 13 K 12 K
50Hz <140 min (4.2K) <75 min (4.2K) <60 min (4.2K) <60 min (4.2K) <70 min (4.2K)
60Hz <120 min (4.2K) <60 min (4.2K) <60 min (4.2K) <60 min (4.2K) <60 min (4.2K)
<35 min (20K)
<40 min (20K)
<50 min (20K)
<40 min (20K)
7.5 kg 11 kg 17 kg 18 kg 13 kg 23 kg 11 kg
○
○
○ ○
○ ○ ○
○
○
First stagerefrigerationcapacity
Second stagerefrigerationcapacity
Cooldown time
SA112
SW112
SA115
SW115
UW404
Type expression
C30HVRT
C30HMVRT-380
C30VR
C30PVRT
C10
Reachingtemperature
Weight
■When applying our refrigerator to a cryocooler for optical sensors, the cryocooler falls under row 6.A.2.d.2 of the control list established by The Wassenaar Arrangement, which is equal to row 10(2) of appended table 1 of Japan’s Export Trade Control Order. Customers must follow all related rules and regulations such as Foreign Exchange and Foreign Trade Act and take appropriate procedures when exporting or re-exporting our refrigerators.
S030 S050
30 W @ 77 K 60 W @ 77 K
33 W @ 77 K 62 W @ 77 K
-
-
-
-
22 K
22 K
25 K
25 K
<25 min (77K)
<25 min (77K)
<20 min (77K)
<20 min (77K)
5.9 kg 6.5 kg
○
○
○
○
○
※1
※1
※2
※2
* 1 Using M501 motor unit * 2 Using M502 motor unit