CP series INCLINED LIQUID COLUMN MANOMETERS The CP 25 inclined liquid column manometer, developed and manufactured by KIMO measures slight variations in pressure, depression or differential pressure of air or gas. It is specially designed for checking clogging of filters in painting booths. ● Graduation scale via coloured zones for fast and easy reading ● Safety reservoir enabling momentaneous overshooting of the scale ● Supplied with two 487 connectors and a bottle of AWS 10 liquid KEY POINTS 0 – 25 mm CE 5 mm MEASURING RANGE SNESITIVITY SCALE for 1 mm CE DIMENSIONS Pressure / Depression 76 mm 152 mm 85 mm 15 mm Ø 5 mm Cylindrical 1/8 gas connector Ø 6.2 mm ribbing [email protected]www.digiparts.ch DigiParts GmbH
24
Embed
CP series INCLINED LIQUID COLUMN MANOMETERS · For a pressure measurement ... HP series INCLINED LIQUID COLUMN MANOMETERS The HP range of inclined liquid column manometers, developed
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
CP seriesINCLINED LIQUID
COLUMN MANOMETERS
The CP 25 inclined liquid column manometer, developed and manufactured by KIMO measures slight variations in pressure, depression or differential pressure of air or gas.It is specially designed for checking clogging of filters in painting booths.
● Graduation scale via coloured zones for fast and easy reading● Safety reservoir enabling momentaneous overshooting of the scale● Supplied with two 487 connectors and a bottle of AWS 10 liquid
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduation
Positioning
Manometric liquidReservoir capacityConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 15 mm thick Altuglas.Entirely bored in the solid block, Ø 4 mm.Directly silk-screened onto the rear face with different colour zones :
from 0 to 5 mm CE : whitefrom 5 to 10 mm CE : greenfrom 10 to 15 mm CE : yellowfrom 15 to 25 mm CE : red (Variations possible on request).
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mmAWS 10 red oil - density 0.87 at 15°C.20 mlØ 5x8 mm semi-rigid crystal tube, on Ø 6.2 mm ribbed, nickel-plated brass connectors, 1/8 gas thread.With or without white PVC support
MOUNTING
1. Mount on a wall or a vertical partition wall with two maximum Ø 5 x 25 mm screws (supplied).2. Set horizontality using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. Connect the manometer with the Ø 5 X 8 mm crystal tube to the pressure or depression source to be checked.
NOTES : For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)
MAINTENANCE :
The CP 25 manometer requires no special maintenance other than simply changing the reading liquid once a year.
The GF range of vertical liquid column manometers, developed and manufactured by KIMO, are mainly for checking pressures in gas networks within measurement ranges which vary according to the type of manometric liquid used : VF 1 or MERCURY.(See table below)
● ''U''-shaped column for measuring consecutively positive and negative pressures● Measurement by addition of values read on each column● For fixed and portable use● Zero adjustment by moving the slide strip● Possibility of resisting static pressures over 10 bars● Altuglas column sunk into the solid block● Comes with bottle of liquid, 2 screws and rawplugs
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid column
Graduated slide stripZero adjustment
Manometric liquidConnection
Wall mounting
From +5 to +30°CFrom -30 to +60°C14 bars20 mm thick PVCTube Ø 4 X 10 mm in altuglas for mercuryTube Ø 6 X 10 mm for VF1Transparent altiglas. Cross-section 54 X 3 mmBy moving the graduated slide strip, travel 20 mm.Fixed via milled, nickel-plated brass screwVF1 liquid, density 13,545 or mercury, density 1Ø 5x8 mm semi-rigid crystal tube on Ø 6.2 Delrin ribbed connectors with M 7x100 thread.
2 screws Ø 5 X 25 mm
Ref.
abc
Distance between tubes
Weight
1. Mount the manometer on a wall or partition wall with two maximum Ø 5 x 25 mm screws.2. Unscrew the left-hand connector and slowly pour the manometric liquid to zero point on the graduation3. Remount the connector without overtightening.4. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
Note :
For a pressure measurement.............................. Connect the crystal tube to one of the two connectorsFor a depression measurement.......................... Connect the crystal tube to one of the two connectors.
The HP range of inclined liquid column manometers, developed and manufactured by KIMO, measure slight variations in pressure, depression or differential pressure of air or gas.They are particularly recommended for measuring overpressures in clean rooms, operating blocks, laboratories, aseptic rooms, laminar flows....
● High measurement sensitivity● Very low pressure ranges● Zero adjustment by moving the slide strip.● Integrated spirit level for adjusting horizontality.● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectorsand a bottle of AWS 10 liquid.
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduated slide strip
Positioning
Manometric liquidReservoir capacityConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 20 mm thick Altuglas.Entirely bored in the solid block, Ø 4 mm.Transparent Altuglas. Cross-section 20 x 2 mm.
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.AWS 10 red oil, density 0.87 at 15°C.20 mlØ 5x8 mm semi-rigid crystal tube, on Ø 6.2 mm ribbed, nickel-plated brass connectors, 1/8 gas thread.With or without white PVC support.
MOUNTING
1. Mount on a wall or a vertical partition wall with two maximum Ø 5 x 25 mm screws (supplied).2. Set horizontality using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. onnect the manometer with the Ø 5 X 8 mm crystal tube to the pressure or depression source to be checked.
NOTES : For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)
MAINTENANCE :
HP manometers require no special maintenance other than simply changing the reading liquid once a year.
Ref. F
Tang
– HP
– 01
/99 A
- RC
S 24
(24)
Pér
igueu
x 349
282 0
95 N
on-co
ntrac
tual d
ocum
ent –
We r
eser
ve th
e righ
t to m
odify
the c
hara
cteris
tics o
f our
pro
ducts
with
out p
rior n
otice
.
Zero adjustment By moving the graduated slide strip, travel 12 mmFixed in place via milled, nickel-plated brass screw.
The KM vertical liquid column portable manometer, developed and manufactured by KIMO, measures low pressures in gas networks.
● Easy to carry.● "U"- shaped column for pressure and depression measurement.● Direct read-off by moving the gratuated slide strip.● Safety valves actuation for momentaneous overshooting of the scale.● Fitted with valve connectors and mounting hook.● Supplied with connection sleeves, a bottle of VOLT 1S liquid and carrying case.
Possible range of usePression statique maximumManometer body
Graduated slide stripZero adjustment
Manometric liquidConnection
From +5 to +30°CFrom -30 to +60°C8 bars15 mm thick transparent Altuglas.
By moving the gratuated slide stripFixed in place via milled, nickel-plated brass screw. VOLT 1S, density 1.86 at 20°C.On Ø 6.2 nickel-plated brass valve connectors, with 1m long neoprene tube fitted with dedicated end-pieces for gas equipment.
1. Dismount one of the 2 connectors using a no.12 spanner and slacken the milled head of the other connector by one turn.2. Check beforehand that the slide strip is at its lowest level.3. Pour the liquid in the column using the spout.4. Do not overfill. Never go beyond the NL line at the middle of the slide strip.5. Remount the connector and screw the milled head of the other connector back down.
Altuglas transparent
TECHNICAL FEATURES
MOUNTING
OPERATION
1. Hang up the manometer vertically by the mounting hook or hold manually.2. Open to the air by slackening the milled heads of the 2 valves (one turn is sufficient).3. Push one of the ends of the connecting tube firmly onto the right-hand valve. Push the other end of the tube onto the pressure point of the pipeline or the instrument which has to be checked.4. The liquid, under gas pressure, decreases in the right column and rises in the left one. If the gas flow occurs too hard and plays the safety valve, repeat the operation by pinching the connecting tube more or less strongly to admit the gas more slowly (if the safety valve is working again, it is because the pressure control exceeds the measuring range of the manomters)5. When the liquid has settled, slide the graduated strip so as to bring the zero mark opposite the right-hand tube's liquid level (lowest level).6. The graduation corresponding to the height of the liquid in the left-hand tube indicates the exact gas pressure.7. Close off the 2 valves securely after operation
Ref. F
Tang
– KM
–20/
10/11
- RC
S 24
(24)
Pér
igueu
x 349
282 0
95 N
on-co
ntrac
tual d
ocum
ent –
We r
eser
ve th
e righ
t to m
odify
the c
hara
cteris
tics o
f our
pro
ducts
with
out p
rior n
otice
.
IMPORTANT :• Only VOLT1S liquid will ensure precise measurement (slide scale graduation corresponding to the density of this liquid). • Maximum static pressure : 8 bars
The KX range of inclined liquid column manometers, developed and manufactured by KIMO, measure slight variations in pressure, depression or differential pressure of air or gas.They are for use in the following applications : treatment of air, ventilation, air conditioning, heating, dust elimination...
● Scale with centered zero.● Possibility for measuring positive and negative pressures.● Zero adjusting by moving the slide strip.● Integrated spirit level for adjusting horizontality● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectors and a bottle of AWS 10 liquid.
Recommended range of useTempérature d'utilisation possibleMaximum static pressureManometer bodyLiquid columnGraduated slide stripZero adjustment
Positioning
Manometric liquidReservoir capacityConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 15 mm thick Altuglas.entirely bored in the solid block, Ø 4mm.Transparent Altuglas. Cross-section 20 x 2 mm.By moving the graduated slide strip, travel 12 mm.Fixed in place via milled, nickel-plated brass screw.Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.AWS 10 red oil, density 0.87 at 15°C20 mlØ 5x8 mm semi-rigit crystal tube, on Ø 6.2 mm ribbed, nickel-plated brass connectors, 1/8 gas thread.With or without white PVC support.
DIMENSIONS
Reference KX 205 KX404
abcdef
weight
178 mm110 mm117 mm25 mm
174 mm109 mm
420 g
219 mm
25 mm
151 mm154 mm
215 mm129 mm
660 g
MOUNTING
1. Mount the manometer on a wall or a vertical partition wall with two maximum Ø 5 x 25 mm screws.2. Set horizontality by using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
NOTE :
For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)
MAINTENANCE :KX manometers require no special maintenance other than simply changing the reading liquid once a year.
The LU range of vertical liquid column manometers, developed and manufactured by KIMO, are mainly for measuring variations of pressure, depression or differential pressure of air or gas within measurement ranges which vary according to the type of manometric liquid used : AWS 10, VOLT 1S or MERCURY.(See table below)
● ''U''-shaped column for measuring consecutively positive and negative pressures● Measurement by addition of values read on each column● Zero adjustment by moving the slide strip● For fixed use regardless of the manometric liquid used● For portable use with the LU model using VOLT 1S liquid● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectors and bottle of liquid
MEASURING RANGE
KEY POINTS
Pressure / Depression
Liquide AWS 10
Liquide VOLT 1S
Mercure
Measuring range
Measuring range
Measuring range
mm CE
mm CE
mm HG
mbar
mbar
mbar
LU 100 LU 20050 – 0 – 50 100 – 0 – 100 200 – 0 – 200
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduated slide stripZero adjustment
Manometric liquidConnection
Wall mounting
From +5 to +30°CFrom -30 to +60°C6 barsTransparent Altuglas of 15 mm thickØ 4 mm tube in extruded Altuglas.Altuglas transparent. Section 40 X 2 mmBy moving the graduated slide strip, travel 10 mm.Fixed via milled, nickel-plated brass screwAWS 10 liquid, density 0.86, VOLT 1S liquid, density 1.86 or mercury, density 13.545Ø 5x8 mm semi-rigid crystal tube on Ø 6.2 nickel-plated brass ribbed connectors (for manometers using AWS 10 or VOLT 1S liquid) or in Delrin (for mercury manometer), 1/8 gas threadWith or without white PVC support.
REF. LU 100
abcde
1. Mount the manometer on a wall or partition wall with two maximum Ø 5 screws.2. Unscrew the left-hand connector and slowly pour the manometric liquid to zero point on the graduation3. Remount the connector without overtightening.4. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
REMARQUES :For a pressure measurement..............................Connect the crystal tube to one of the two connectorsFor a depression measurement..........................Connect the crystal tube to one of the two connectors.For a differential pressure measurement...........Connect both tubes to both connectors
The MG range of inclined liquid column manometers, developed and manufactured by KIMO, measure slight variations in pressure, depression or differential pressure of air or gas.They are for use in the following applications : treatment of air, ventilation, air conditioning, heating, dust elimination...
● Zero adjustement by moving the slide strip● Safety reservoir enabling momentaneous overshooting of the scale● Integrated spirit level for adjusting horizontality● Supplied with a white PVC support, two screws and two rawplugs, two 487 connectors and a bottle of AWS 10 liquid
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduated slide strip
Positionning
Manometric liquidReservoir capacityConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 15 mm thick Altuglas Entirely bored in the solid block, Ø 4 mmTransparent Altuglas. Cross-section 20 X 2 mm
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.AWS 10 red oil, density 0.87 at 15°C20 mlØ 5 X 8 mm semi-rigid crystal tube, on Ø 6,2 mm ribbed, nickel-plated brass connectors, 1/8 gas threadWith or without white PVC support
MOUNTING
1. Mount on a wall or a vertical partition wall with two maximum Ø 5 x 25 mm screws (supplied).2. Set horizontality using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. Connect the manometer with the Ø 5 X 8 mm crystal tube to the pressure or depression source to be checked.
NOTE :For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)MAINTENANCE :MG manometers require no special maintenance other than simply changing the reading liquid once a year.
Ref. F
Tang
– M
G – 1
7/08/1
1 - R
CS 2
4 (24
) Pér
igueu
x 349
282 0
95 N
on-co
ntrac
tual d
ocum
ent –
We r
eser
ve th
e righ
t to m
odify
the c
hara
cteris
tics o
f our
pro
ducts
with
out p
rior n
otice
.
Zero adjustment By moving the graduated slide strip, travel 20 mm.Fixed via milled, nickel-plated brass screw
The MT range of inclined liquid column portable manometers, developed and manufactured by KIMO, are particulary recommended for those in the heating trade for checking pressures in chimney flues, combustion chambers, filters...
● Easy to carry● Safety reservoir enabling momentaneous overshooting of the scale● Zero adjustment by moving the slide strip● Integrated spirit level for adjusting horizontality.● Equipped with valve connectors, magnetic fixations, support with base plate● Can be used for measuring speeds with Pitot tube.● Supplied with connecting sleeves, bottle of VOLT 1 S liquid and travelling case.
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduated slide strip
Positionning
Manometric liquidReservoir capacityConnection
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 15 mm thick Altuglas Entirely bored in the solid block, Ø 4 mmTransparent Altuglas. Cross-section 20 X 2 mm
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.VOLT 1S oil - density 1.86 at 20°C.20 mlon valve connectors in nickel-plated brass Ø 6.2 mm neoprene tubes with connecting sleeves.
MOUNTING
1. Place the manometer on a horizontal surface or a vertical partition wall by using the magnetic fixations.2. Set horizontality by using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slacken the milled wheel on the other connecter by one turn.4. Slowly pour the manometric liquid to zero point on the graduation.5. Remount the connector without overtightening.6.Connect the manometer with the tube provided to the pressure or depression source to be checked.
NOTE :For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)MAINTENANCE :MT40 manometer requires no special maintenance other than simply changing the reading liquid once a year.
Ref. F
Tang
– M
T – 1
7/08/1
1 - R
CS 2
4 (24
) Pér
igueu
x 349
282 0
95 N
on-co
ntrac
tual d
ocum
ent –
We r
eser
ve th
e righ
t to m
odify
the c
hara
cteris
tics o
f our
pro
ducts
with
out p
rior n
otice
.
Zero adjustment By moving the graduated slide strip, travel 20 mm.Fixed via milled, nickel-plated brass screw
DIMENSIONS
a
d
e
Cylindrical 1/8 gas connectorØ 6.2 mm ribbing
c
b
ReferenceMT 40
a b c d e Weight193 mm 208 mm 93 mm 37 mm 80 mm 800 g
The TJ range of vertical liquid column manometers, developed and manufactured by KIMO, are mainly for measuring variations of pressure, depression or differential pressure of air or gas within measurement ranges which vary according to the type of manometric liquid used : AWS 10, VOLT 1S or MERCURY.(See table below)
● "J"- shaped column for direct measuring.● Zero adjustment by moving the slide strip.● For fixed use regardless of the manometric liquid used.● For portable use with the TJ model using VOLT 1S liquid.● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectors and a bottle of liquid.
Recommended range of usePossible range of useMaximum static pressureMAnometer boyLiquid columnGraduated slide stripZero adjustment
Manometric liquidConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 15 mm thick Altuglas Ø 4mm tube in extruded Altuglas.Transparent Altuglas. Cross-section 23 x 3 mm.moving the gratuated slide strip, travel 15 mmFixed in place via milled, nickel-plated brass screw.
AWS 10 liquid, density 0.86 ; VOLT 1S liquid, density 1.86 or mercury, density 13.545Ø 5x8 mm semi-rigid crystal tube on Ø 6.2 nickel-plated brass ribbed connectors (for manometers using AWS 10 or VOLT 1S liquid) or in Delrin (for mercury manometer), 1/8 gas thread.With or without white PVC support.
REF. TJ 100
abcde
1. Mount the manometer on a wall or partition wall with two maximum Ø 5 screws.2. Unscrew one of the two connectors and slowly pour the manometric liquid to zero point on the graduation.3. Remount the connector without overtightening4. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
NOTE :
For a pressure measurement..................Connect the crystal tube to the right-hand connector (+)For a depression measurement..............Connect the crystal tube to the left-hand connector (-)For a differential pressure......................Connect the highest pressure to the right-hand connector(+)
and the lowest pressure to the left hand connector (-)
The TX range of inclined liquid column manometers, developed and manufactured by KIMO, measure slight variations in pressure, depression or differential pressure.They are particulary recommended for measuring overpressures and depressions in clean rooms, operating blocks, laboratories, aseptic rooms, laminar flows...
● Scale with centered zero● Possibility for measuring positive and negative pressures● High measurement sensitivity● Very low pressure ranges● Zero adjusting by moving the slide strip● Integrated spirit level for adjusting horizontality● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectors and a bottle of AWS 10 liquid.
KEY POINTS
MEASURING RANGE
Resolution
0.1 mm CE or 1 Pa0.1 mm CE or 1 Pa
Sensitivity scaleFor 1 mm CE or 10 PA
20 mm15 mm50 – 0 – 50
25 – 0 – 25Pascal
2.5 – 0 – 2,55 – 0 – 5
mm CEMeasuring range
Reference
TX 25TX 50
0.1 mm CE or 1 Pa10 mm75 – 0 – 757.5 – 0 – 7.5TX 75
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduated slide strip
Positioning
Manometric liquidReservoir capacityConnection
Wall-mounted
From +5 to +30°CFrom -30 to +60°C6 barsTransparent 20 mm thick AltuglasEntirely bored in the solid block, Ø 4mm.Transparent Altuglas. Cross-section 20 x 2 mm
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.AWS 10 red oil, density 0.87 at 15°C.20 mlConnection : Ø 5x8 mm semi-rigid crystal tube, on Ø 6.2 mm ribbed, nickel-plated brass connectors, 1/8 gas threadWith or without white PVC support.
MOUNTING
1. Mount the manometer on a wall or a vertical partition wall with two maximum Ø 5 x 25 mm screws.2. Set horizontality by using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
NOTE : For a pressure measurement..............................Connect the crystal tube to the right-hand connector (+)For a depression measurement..........................Connect the crystal tube to the left-hand connector (-)For a differential pressure..................................Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)
MAINTENANCE :
TX manometers require no special maintenance other than simply changing the reading liquid once a year.
Ref. F
Tang
– TX
– 02
/99 A
- RC
S 24
(24)
Pér
igueu
x 349
282 0
95 N
on-co
ntrac
tual d
ocum
ent –
We r
eser
ve th
e righ
t to m
odify
the c
hara
cteris
tics o
f our
pro
ducts
with
out p
rior n
otice
.
Zero adjustment By moving the graduated slide strip, travel 12 mm.Fixed in place via milled, nickel-plated brass screw.
The VH range of inclined liquid column manometers, developed and manufactured by KIMO, measure slight variations in pressure, depression or differential pressure of air or gas.They are particulary recommended for checking clogging of filters in the ventilation and dust elimination industry.
● Horizontale V-shaped liquid column.● Dual measuring range● Different sensitivity on the two measuring ranges● Zero adjustment via float● Compact dimensions● Integrated spirit level for adjusting horizontality● Supplied with a white PVC support, 2 screws and 2 rawlplugs, two 487 connectors and a bottle of AWS 10 liquid.
KEY POINTS
MEASURING RANGE
Resolution
1 mm CE or 10 Pa
Sensitivity scaleFor 1 mm CE or 10 PA
3.5 mm
Pascalmm CEMeasuring range
Reference
VH 50
Pressure / Depression
Total 1st column 2nd column Total 1st column0 – 50 0 – 16 19 – 50 0 – 500 0 – 160 190 – 500 7 mm
Recommended range of usePossible range of useMaximum static pressureManometer bodyLiquid columnGraduation
Positioning
Manometric liquidReservoir capacityConnection
Wall-mount
From +5 to +30°CFrom -30 to +60°C1 barTransparent 15 mm thick Altuglas.Entirely bored in the solid block, Ø 4 mm.Directly silk-screened onto the rear face.
Horizontal positioning via integrated spirit level and milled, nickel-plated brass adjusting screw, vertical travel 12 mm.AWS 10 red oil, density 0.87 at 15°C.20 mlØ 5x8 mm semi-rigid crystal tube, on Ø 6.2 mm ribbed, nickel-plated brass connectors,1/8 gas thread.With or without white PVC support.
MOUNTING
1. Mount the manometer on a wall or a vertical partition wall with two maximum Ø 5 screws.2. Set horizontality by using the integrated level and the milled adjusting screw.3. Unscrew the connector on the reservoir and slowly pour the manometric liquid to zero point on the graduation.4. Remount the connector without overtightening.5. Connect the manometer with the Ø 5x8 mm crystal tube to the pressure or depression source to be checked.
NOTE :For a pressure measurement.............................. Connect the crystal tube to the right-hand connector (+)For a depression measurement.......................... Connect the crystal tube to the left-hand connector (-)For a differential pressure.................................. Connect the highest pressure to the right-hand connector (+) and the lowest pressure to the left
hand connector (-)
MAINTENANCE :
VH manometers require no special maintenance other than simply changing the reading liquid once a year.
Zero adjustment By moving the Altuglas float and the milled, nickel-plated brass screw, travel 10 mm.