chromatography

Page 2: chromatography

Chromatographic Process

Chromatographic separations are based on transporting the liquid carrying the analyte mixture (mobile ph. ) through the porous media (stationary ph.) and the differences in the interactions of analytes with the surface of this porous media resulting in different migration times (retention times) for a mixture components.

Page 3: chromatography
Page 4: chromatography

Provides the analyte transport.

Immobile phase.

Mixture of components dispersed

in the mobile phase.

Mobile phase

Analyte

Stationary phase

Page 5: chromatography

Classification

The nature of the stationary and the mobile phases, with the mode of the transport through the column, is the basis for the classification of chromatographic methods.

Page 6: chromatography

HPLC system

Page 7: chromatography
Page 8: chromatography

Solvent Reservoirs: Storage of HPLC solvents (mobile phase) and could be equipped with an online degassing system and special filters.

pumpProvides the constant and continuous flow of the mobile phase through the system. Modern pump allow mixing of different solvents from different reservoirs.Injector: Allows the introduction (injection) of the analytes mixture into the stream of the mobile phase before it enters the column. Modern injectors are autosamplers.

Page 9: chromatography

Column:It is a device that hold a stationary phase in place, allowing the mobile phase to carry an injected sample through and allowing analytes to interact with available surface.Guard column:A short column introduced before the analytical column to increase its life by removing particulate matter and contaminants.

Detectors: A device for continuous registration of specific physical properties of the column effluent

Page 10: chromatography

Stationary phase

It is the column packing material that are the media producing the separation.

The properties of this media are important for successful separations.

Page 11: chromatography
Page 12: chromatography

chromatogram

Use HPLC to separate the compounds physically

Qualitatively by?

Quantitivly by?

Page 13: chromatography

HPLC parameters

Dead volume. Retention volume (VR). Retention time (tR). Void time (t₀). Capacity factor (K). Selectivity (ᾳ) Efficiency. Resolution (R). Peak symmetry factor (s).

Page 14: chromatography

Dead volume: Volume of mobile phase inside the column .

V₀= V column – V particle + V pores.

=0.65 (µD²L/4) where dead volume equal 65% of empty column volume

Retention volume (VR).Total volume of eluent (in ml) require to elute certain substance.

VR= tR * F

Page 15: chromatography

Retention time (tR).Time from injection point to maximum detector response for corresponding compound.

Void time (t₀).Time require for non retaine substance to pass through the column.

Capacity factor (K)Use to describe the migration rate of solutes on columns.K=(tR -t₀ )/t₀ Recommended to be 2-10 Selectivity (ᾳ)Use to describe the separation of band centreᾳ =K2/K1 should be > 1.2

Page 16: chromatography

Efficiency.It is measured in number of theoretical plates. N= 16 ( tR/w)². H=L/N*H plate height.*column length.

Resolution (R).

R = 2(t R2 –t R1) / (w2 +w1)Should be 1.5

Peak symmetry factor (s)

Page 17: chromatography
Page 18: chromatography

problems

1-On given 200 cm column the sample required 4.80 min. to emerge, an air bubble required 0.5 min. ,and the time required for the sample pass the detector was 0.40min. Calculate N and platelet height?2- the adjusted retention distance for a lipid sample on HPLC chromatogram with a 1.5m column was 59.6 cm. the recorder was calibrated at 3.60cm/min. the width of the peak at the base was 12.0 cm calculate N?

Page 19: chromatography

The following data apply to a column for a lipid chromatography:

Length of packing 24.7 cm Flow rate 0.313ml/min. Vm 1.37ml Vs 0.164 ml

A chromatogram of a mixture of species A,B,C,D provided the following data;

Width of peak base

( w()min)

Retention time (min)

- 3.1 Non retained0.41 5.4 A1.07 13.3 B1.16 14.1 C1.72 21.6 D

Page 20: chromatography

Calculate 1- the number of plates from each peak?2- the plate height for the column?

From the previous data calculate for A,B,C, and D:- the capacity factor?

Calculate for B and C:-the resolution?-the selectivity factor?

Calculate for C and D:- The resolution?

Page 21: chromatography

Determination of paracetamol concentration in pediatric

drops

Quantitative method. Using external standard method (more basic and

common) . AUC is directly relashion to the concentration. It calculated by calibration curve or by matching

point.

Page 22: chromatography

Procedure:Stock standard :-weight 0.1 g of pure paracetamol powder and put it in 100 ml volumetric flask and complete it with deionized water.

-From stock take 1 ml and put it into 100 ml volumetric flask and complete it with deionized water and called it solution A.

-From solution A prepare serial dilution 2,4,6,8,10 ml in 25 ml volumetric flask and complete it with mobile phase. Mobile phase:25% acetonitrate +75% acetic acid 1%

Page 23: chromatography

Sample preparation:

-From bottle take 1 ml ( contain 100mg) and put it in 100 ml volumetric flask with deionized water and called it B.

-from B take 1ml and put it in 100ml volumetric flask and complete it with deionized water and called it C.

- From C take 5 ml put it in 25 volumetric flask and complete it by mobile phase.

Calculation: