Characterization And Testing Of

2m X 2mGlass Resistive Plate Chamber

Mohammed Salim M

Aligarh Muslim University

Page 2: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Topics

Introduction Principles of Operation Gas Mixture Fabrication Testing Trigger Results

Page 3: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Introduction

INO (India based Neutrino Experiment) is India’s biggest experiment to study the properties of weakly interacting, electrically neutral fundamental particle, neutrinos

Underground experiment will study the properties of ‘Atmospheric Neutrinos’.

Resistive Plate Chambers (RPC) are the active detectors used in the proposed INOICAL (Iron Calorimeter)

Page 4: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Introduction (contd..)

Page 5: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Introduction (contd..)

Two 3mm thick float glassesSeparated by 2mm spacers

2mm thick spacer

Glass plates

Graphite coating on the outer surfaces of glasses

Pickup strips

Page 6: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Various layers of RPC

Page 7: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Principle of operation

3

Two modes of operation

o Avalanche mode

o Streamer mode

l

A

A

lRC

s 2)10 x (8.85 x 4 x )10 x ( -1210

Page 8: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Gas mixture

The gas mixture fixes the working mode of RPC

Three gases are used in avalanche mode

R134A( as Freon) 95.4% Isobutane 4.3% SF6 0.3%

Page 9: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Avalanche and Streamer pulses Taken from CRO

Page 10: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Fabrication

1. Glass cutting and cleaning2. Conductive coating3. Gas leak test4. Connecting High Voltage Cables5. Placing pick up panels6. Connecting Pre Amplifier boards

Page 11: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Surface resistance measurement

1 div~33cm

Uniform surface resistance of 1 MΩ is expected.

Page 12: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

VI Characteristics

Page 13: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Trigger

• Scintillator paddles are used for triggering the RPC pulse

Page 14: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Block diagram of lab set up

Page 15: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Results of 2m X 2m RPC

Efficiency Noise rate at various voltages

Fixed operating voltage at 9.6 kV

Page 16: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Effect of SF6

Page 17: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

TDC and QDC plots

TDC plot of the main strip (strip no.32) ADC plot of the main strip

Page 18: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Calibration of the MFC

The gas is fluxed into the tube and then water is allowed to flow through the tube. The gas apply the pressure on the water and water bubble flows through the tube and we can determine the rate of flow by measuring the change in volume in some fixed time with the help of stop watch.

Water flow

Glass tube Gas flow

Scale

Page 19: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Calibration plots of MFC

Page 20: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Calibration plots of MFC

Page 21: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.

Snap shots from our work shop

Page 22: Characterization And Testing Of 2m X 2m Glass Resistive Plate Chamber Mohammed Salim M Aligarh Muslim University.