Top Banner

of 21

Chapter 0010

May 29, 2018

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/9/2019 Chapter 0010

    1/21

    Measurement Errors

    The measurement of any quantity plays very important role not only in science but in

    all branches of engineering, medicine and in almost all the human day to day activities.

    The technology of measurement is the base of advancement of science. The role of science

    and engineering is to discover the new phenomena, new relationships, the laws of nature

    and to apply these discoveries to human as well as other scientific needs. The science and

    engineering is also responsible for the design of new equipments. The operation, controland the maintenance of such equipments and the processes is also one of the important

    functions of the science and engineering branches. All these activities are based on the

    proper measurement and recording of physical, chemical, mechanical, optical and many

    other types of parameters.

    The measurement means, to monitor a process or a operation and using an instrument,

    express the parameter, quantity or a variable in terms of meaningful numbers. Such a

    measurement gives in depth knowledge of the process and the parameter and helps in

    further modifications, if required. Thus the measurement provides us with a means of

    expressing a natural phenomena or the various processes, in quantitative terms. The

    feedback information is possible with the help of measurement techniques, which helps in

    achieving goals and objectives of various engineering processes and systems.

    The measurement of a given parameter or quantity is the act or result of a quantitative

    comparison between a predefined standard and an unknown quantity to be measured.

    For the result to be meaningful, there are two basic requirements :-

    1. The comparison standard is accurately defined and commonly accepted, and

    2. The procedure and the instrument used for obtaining the comparison must be

    provable.

    The major problem with any measuring instrument is the error. Hence, it is necessary

    to select the appropriate measuring instrument and measurement procedure which

    minimises the error. The measuring instrument should not affect the quantity to be

    measured.

  • 8/9/2019 Chapter 0010

    2/21

    This is the actual value of the unknown resistance.

    At - A 5.462 -5.333ii i ) % error = At m X 100 5.462 x 100

    = 2.36 %

    iv ) The relative accuracy,

    % A (1 -Ierrorl) x lO O = (1 - 0.0236) x lO O

    97.63 %

    1 . W ha t is mea surem ent? W hat ar e the two basic requirements o f an y measurement?

    2. List the advantages of an elec tronic measurement.

    3. De fine and cxplain tilc t erm 'Calibration '.

    - t . How the performance characteristics o f a n i nstrumcnt ar e classified?

    5. Definc and e xplain the following static characteristics of an instrulllcnt :

    i) Accuracy ii) Prec ision iii) Static error iv) Resolution

    v) Sen sitiVity v/) Th reshold vii) Z er o drift viii ) Rcp rodu cibility

    ix) Lillearit y a nd x) Stability

    6. Exp lain how the accuracy can be spec ified for a n instm ment.

    7. Distinguish clea rl y b etwe en aCCliracy and pr ec ision.

    S. State an d explain the cha rac teristics o f precision.

    9. Exp laill t i le terms relative erro r a n d re lative perc entage error.

    10 . What is scale s pan of an instrument?

    11 . Define a dynamic response o f an instrument.

    12 . D efin t' the following terms,i) Sp ee d o f r es po ns e ii) Lag ii i) F id eli ty ivY Dyna mic error .

    13. Define and e xplain the types of errors possible in an instrument.

    14 . Define limiting errors. Derive the e xpression for relative limiting error.

    1 5 . A mov ing coil vo ltm ete r h as a u nifo rm s ca le with 100 divisions, the full scale reading is 200 V

    an d 1/10 of sca le d ivisio n ca n be estim ated w ith a fa ir deg ree of certainity . Determine the

    reso lution of the in strument i n volt . [Ans. : 0.2 V]

    16. A digital vol tmeter has a read out range fr om 0 -9 999 counts. Determine the resolution of the

    instrum ent in vo lt w hen the ful l scale reading is 9.999 \ I. [Ans. : 1 mV]

    17. A t ru e value o f volta ge acro ss res ister is 50 V. T he i nstru men t r eads 49 V. Calculate

    i) ab solute error ii) percentage error iii) percentage aCCliracy

    [Ans. : 1 V, 2%, 98%]

  • 8/9/2019 Chapter 0010

    3/21

    Voltmeters and Multimeters

    The measurement of a given quantity is the result of comparison between the quantity

    to be measured and a definite standard. The instruments which are used for such

    measurements are called measuring instruments. The three basic quantities in the electricnlmC,lsurement are current, voltage and power. The measurement of these quantities is

    important as it is used for obtaining measurement of some other quantity or used to test

    the performance of some electronic circuits or components etc.

    The necessary requirements for any measuring instruments are:

    1) With the introduction of the instrument in the circuit, the circuit conditions

    should not be altered. Thus the quantity to be measured should not get affected

    due to the instrument used.

    2) The power consumed by the instruments for their operation should be as small as

    possible.

    The instrument which measures the current flowing in the circuit is called ammeter

    while the instrument which measures the voltage across any two points of a circuit is

    c,ll1ed voltmeter. But there is no fundamental diff erence in the operating principle of

    analog \'oltmeter and ammeter. The action of almost all the analog ammeters and

    \oltmeters depends on the deflecting torque produced by an electric current. In ammeters

    such c 1 torque is proportional to the current to be measured. In voltmeters this torque is

    decided by a current which is proportional to the voltage to be measured. Thus all the

    analog ammeters and voltmeters are basically current measuring devices.

    A basic d.c. meter uses a motoring principle for its operation. It stntes that any current

    carrying coil placed in a magnetic field experiences a force, which is proportional to the

    magnitude of current passing through the coil. This movement of coil is called D'Arsonval

    movement and basic meter is called D'Arsonval galvanometer. Adding various other

    elements to the basic meter, various practical instruments can be obtained. These

    instruments are classified as,

  • 8/9/2019 Chapter 0010

    4/21

    a) Using shunt resistance, d.c. current can be measured. The instrument is d .c .

    microammeter, milliammeter or ammeter.

    b) Using series resistance called multiplier, d.c. voltage can be measured. The

    instrument is d.c. millivoltmeter, voltmeter or kilovoltmeter.

    c) Using a battery and resistive network, resistance can be measured. The instrument

    is ohmmeter.

    a) Using a rectifier, a.c. voltages can be measured, at power and audio frequencies.

    The instrument is a.c. voltmeter.

    b) Using a thermocouple type meter radio frequency (RF) voltage or current can be

    measured.

    c) Using a thermistor in a resistive bridge network, expanded scale for power line

    voltage can be obtained.

    The basic d.c. voltmeter is nothing but a

    permanent magnet moving coil (PMMC)

    0' Arsonval galvanometer. The resistance is

    required to be connected in series with the

    basic meter to use it as a voltmeter. This

    series resistance is called a multiplier. The

    main function of the multiplier is to limit the

    current through the basic meter so that the

    meter current does not exceed the full scale

    deflection value. The voltmeter measures the

    voltage across the two points of a circuit or a

    voltage across a circuit component. The basic d.c. voltmeter is shown in the Fig. 2.1.

    The voltmeter must be connected across the two points or a component, to measure

    the potential difference, with the proper polarity.

    The multiplier resistance can be calculated as :

    Rm

    Basicmeter

    Let Rm internal resistance of coil i.e. meter

    R s = series multiplier resistance

    1m full scale deflection current

    V = full range voltage to be measured

    From Fig. 2.1, V = I In (R il l + R s)

  • 8/9/2019 Chapter 0010

    5/21

    ~v

    L_'_-IIll_-

    R

    _

    Ill

    _

    The multiplying factor for multiplier is the ratio of full range voltage to be I11casurC'd

    and the drop across the basic cleter.

    1 I" Vmu tlP VIng J,Ktor = -" - ~

    IIll(RIll+R,)

    '111 Rill

    Thus to increase thE r

  • 8/9/2019 Chapter 0010

    6/21

    (20 mA) Rm

    10 n

    20x 10 -' x 10200-20x 10 -'

    Thi: is the required shunt.

    ii) For llsing it as a v oltmeter,

    500 V

    ~-R1

    111

    m

    500 -1020x ]0--'

    The range of the basic d.c. voltmeter can be extended by using number of multipliers

    clnd a selector switch. Such a meter is called multirange voltmeter and is shown in

    the Fig. 2.2.

    Fig. 2.2 Multirange voltmeter

    The R:, R2, R) and R~ are the four series multipliers. When connected in series with

    the meter, they can give four diff erent voltage ranges as V], V2,V-, and V~. The "elector

  • 8/9/2019 Chapter 0010

    7/21

  • 8/9/2019 Chapter 0010

    8/21

    __ R_ :! R_T_-_(R_,,_n_+_R_3_+_R_~)__ JIn position V1, the multiplier is R1 + R:! + R3 + R~

    VIRT =

    Jm

    R1 + R2 + R \ + R~ = F'I - Rill

    RT - ( Rill + R2 + R3 + R~) JUsing the equations (1), (2), (3) and (4) multipliers can he dcsigncci. Tlw ,ld \ ,mtage of

    this t

  • 8/9/2019 Chapter 0010

    9/21

    v ,.-:- - Rm.till

    _5_0 _ 50 - 49502x1O 1

    _1_00 _ 50 - 4950 - 200002xlO 1

    For position Vj = 500 V, multiplier is ( R1 + R2 + R, + R~ ).

    VI(R1 + R2 + R, + R~) 1:- Rm

    500 _ 50 - 4950 - 25000 - 200002xl0 3

    R1 = 200 kQ

    Thus Rj, R2, R, and R~ forms a series string of multipliers.

    Tn a multirange voltmeter, the ratio of the total resistance R r to the vollag(' r,1ng:'

    remains same.((rhis ratio is nothing but the reciprocal of the full scale deflection current ,)f

    the meter)i.e. 1/101. This value is called sensitivity of the voltmeter.

    Thus the sensitivity of the voltmeter is defined as,

    S = I

    Full scale deflection current

    1S = r:D./V or kO .!V

    Key Point: The sensiti()it y range is specified 0/1 the nleter dial and it ii7dicntes the res iston ce

    of the 11Ieterfor n one volt range.

    The internal resistance of the voltmeter is not the same in each of its ranges. The

    higher is the range of the voltmeter, greater is its internal resistance. Internal resistance of

    a voltmeter can be obtained from its sensitivity as,

    Internal resistance of voltmeter = Maximum voltage (range) x Sensitivity in D/V

  • 8/9/2019 Chapter 0010

    10/21

    The ~~'nsitivity i~ useful in calculating the resistance of a multiplier in d.c voltmeter.

    Con-.;ider the prClcticcll multirange voltmeter circuit shown in the Fig. 2.5.

    R1 R2 R3 R4

    V2+

    V1 t RmV4S BasIc

    meter

    Fig. 2.5

    sensitivity rating in 0.jV

    where VI, V2, V ., cllld V-I (lre the required voltage ranges.

    Key Point: This lIlethod is called the sensitivity method of cnlClllating thl' lIlultiplier

    /"l'sistallcl's.

    1

    2x10-3

    = 500 OjV

    VI 500 V, V2 = 1 0 0 V, VI = 50 V, V-I=10 V

    R-l S V-l - Rm = 500 x 1 0 - 5 0

    4.951

  • 8/9/2019 Chapter 0010

    11/21

    111* Example 2.5: Calcula te the va lue of the mul tiplier re sistanc e 011 the 5UU V ra nge of 1 1 d.c.

    utiltmeta, that us es 5U IlA meta movement with an in ternal resistance of 200 D.

    s = r : , -50X~0 6 =20000 D/V

    11I* Example 2.6 : The meter A has. a ral lgc of 0 - 100 V a nd m u ltip lie r r es istance of 25 ill.

    TllC met('r B h a s a I"IIn ge 0 - 1000 V a nd a m u lt ipli('r re sistance of 150 kD.. Both metas /1I11'C

    [Ia sic m ete r r es ista nc e o f 1 l\fl. Whi ch meter is 111 0resensitive?

    Solution : ~or meter A, Rs = 25 ill, Rm= 1 ill, V = 100 V

    Now R, SV- Rm

    25 x 1() 3 S x 100 - lxlO 3

    S2600.jV

    For meter B, R, 150 ill, Rm= 1 ko., V = 1000 V

    R, SV- Rm

    150 x 103 S x 1000 - 1 x 103

    S 151 D/V

  • 8/9/2019 Chapter 0010

    12/21

    While selecting a m eter for a particular measurement, the sensitivity rating IS very

    important. A low sensitive meter may give the accurate reading in low resistance circuit

    but will produce totally inaccurate reading in high resistance circuit.

    The voltmeter is always connected across the two points between which the potential

    difference is to be measured. If it is connected across a low resistance then as voltmeter

    resistance is high, most of the current will pass through a low resistance and will produce

    the voltage drop which will be nothing but the true reading. But if the voltmeter is

    connected tcr s arf used all the 150 V

    + 1

    250 V

    j

    2 50 2 "-(20 + 25) x .J

  • 8/9/2019 Chapter 0010

    13/21

    Req R2 I I Rv

    25x 75

    (25 + 75)

    Hence the voltage across Rt.q is,

    V = Req x 250(Req + R1)

    Thus first voltmeter will read 120.96 V.

    Case ii) S = 10,000 {1IV

    The voltmeter resistance will be

    Rv S V

    1000,0 x 150

    1.5 1 \.1 0

    Req R2 I I Rv

    25x1.5xl06 xl03

    (25xl03 + 1.5xl06

    )

    Hence the voltage across Req is,

    Req 250 24.59 ~50V = (Req + R1) x = (24.59 + 20) xL

    Thus the second voltmeter reads more accurately.Key Point: Thus the high sensitivit y voltm eter gives more accurate read ing, though the

    voltage range for both the meters is same.

  • 8/9/2019 Chapter 0010

    14/21

    4.167 -4.132

    4.167 x 100

    0.84%

    e) The percentage accuracy can be obtained as :

    99.16%

    Thus voltmeter 2 is 99.16% accurate while voltmeter 1 is 85.7% accurate.

    2.5.2 Precautions to be taken while using a Voltmeter

    The following precautions must be taken while using a voltmeter:1) The voltmeter resistance is very high and it should always be connected across the

    circuit or component whose voltage is to be measured.

    2) The polarities must be observed correctly. The wrong polarities deflect the pointer

    in the opposite direction against the mechanical stop and this may damage the

    pointer.

    3) While using the multirange voltmeter, first use the highest range and then decrease

    the voltage range until the sufficient deflection is obtained.

    4) Take care of the loading effect. The effect can be minimised by using high

    sensitivity voltmeters.

    2.5.3 Requirements of a Multiplier

    1) Their resistance should not change with time.

    2) The change in their resistance with temperature should be small.

    3) They should be non-inductively wound f or a.c. meters.

    Commonly used resistive materials for construction of multiplier are manganin and

    constantan.

    The PMMC movement used in d.c. voltmeters can be effectively used in a.c.

    voltmeters. The rectifier is used to convert a.c. voltage to be measured, to d.c. This d.c., if

    required is amplified and then given to the PMMC movement. The PMMC movement

    gives the deflection proportional to the quantity to be measured.

    lt is important to study some basic definitions related to the a.c. quantities, before

    studying the operation of the a.c. voltmeters. The a.c. meters are usually calibrated to read

    Lm.S. value of an alternating quantity to be measured.

  • 8/9/2019 Chapter 0010

    15/21

    The r.m.s. value of an alternating quantity is given by that steady current (d.c.) which

    when flowing through a given circuit for a given time produces the same amount of heat

    as produced by the alternating current which when flowing through the same circuit for

    the same time. The r.m.s value is calculated by measuring the quantity at equal intervals

    for one complete cycle. Then squaring each quantity, the average of squared v,llues is

    obtained. The square root of this average value is the r.m.s. value. The r.m.s means

    root-mean square i.e. squaring, finding the mean i.e. average and finally root.

    If the waveform is continuous then instead of squaring and calculating mC,lll, the

    integratioll is used. Mathematically the r.m.s. value of the continuous a.c. voltage having

    time period T is given by,

    1TT f V}, dt

    o

    The + term indicates the mean value or average value.

    Most of the a.c. vo1tmeters are r.m.s. responding or average responding type, with

    scale calibrated interms of the r.m.s. value of a sine wave.

    The average value of an a.c. quantity is another important parameter. The average

    value is defined as that value which is obtained by averaging all the instantaneous values

    over a period of a half cycle. For the symmetrical a.c. quantity, the average value over a

    complete cycle is zero as both positive and negative half cycles are exactly identical. Hence

    average value is calculated over a half cycle. If the a.c. quantity is continuous then average

    value can be expressed mathematically using an integration as,

    T/2

    Vav = ~ fVin dt

    o

    The interval T/2 indicates the average over half a cycle.

    For purely sinusoidal quantity,

    2= - Vm = 0.636

    IT

    As mentioned earlier, the average responding meter scale is also calibrated in terms of

    r.m.s. values. To achieve such calibration, a pure sine wave ,,-ith ":'.m.s. value of 1 V is

  • 8/9/2019 Chapter 0010

    16/21

    applied. Then the deflection of meter is adjusted to IV reading. For this, a particular factor

    is required to be considered. This factor is called Form Factor.

    The form factor is the ratio of r.m.s. value to the average value of an alternating

    quantity.

    r. m. s. value f f - - - - - - = orm actoraverage value

    Key Point: For purely sinusoidal wave form theform factor is 1.11.

    Thus while calibrating average responding meter interms of r.m.s. values the markings

    are actually corrected by a factor of 1.11.

    r. m. s. value

    Kf

    Some meter scales are calibrated in terms of peak values of the input. In such cases

    another factor relating peak value and the r.m.s. value becomes important. This factor isGllled Peak Factor or Crest Factor.

    The peak factor or crest factor is the ratio of peak (maximum) value to the r.m.s. value

    of an alternating quantity.

    K _ maximum valuep - r. m. s. value

    Key Point: For purely sinusoidal a.c. quant ity the crest factor is 1.414.

    The question is, why to use these factors to correct the readings by measuring averageand peak values, when the true r.m.s. voltmeter can give direct r.m.s reading. The reason

    behind this is that the average and peak responding meters are less in cost and very

    simple in construction as compared to true r.m.s. voltmeters.

    A.C. voltmeters can be designed in two ways:

    i) First rectifying the a.c. signal and then amplifying.

    ii) First amplifying the a.c. signal and then rectifying.

    2.6.1 First Rectifying and then Amplifying A.C. Signal

    In this arrangement simple diode

    rectifier circuit precedes the amplifier and

    the meter. This is shown in the Fig. 2.8.

    The a.c. input voltage if first rectified

    using the diode D. This rectified signal is

    then applied to the amplifier of gain A.

    The amplified signal is then given to the

    basic PMMC meter to obtain the

    deflection.

  • 8/9/2019 Chapter 0010

    17/21

    This

  • 8/9/2019 Chapter 0010

    18/21

    diode 01 , the meter is shunted with a resistance Rsh' This ensures high current through

    aiode and its linear behaviour.

    When the a.c. input is applied, for the positive half cycle, the diode 01 conducts and

    causes the meter deflection proportional to the average value of that half cycle.In the negative cycle, the diode O2 conducts and 01 is reverse biased. The current

    through the meter is in opposite direction and hence meter movement is bypassed.

    Thus due to diodes, the rectifying action produces pulsating d.c. and lile meter

    indicates the average value of the input.

    R Basicm meter

    The a.c. voltmeter using half

    wave rectifier is achieved byintroducing a diode in a basic d.c.

    voltmeter. This is shown in the

    Fig. 2.12.

    The diode 0 conducts only

    during positive half cycle. Let us

    compare the sensitivities of d.c.

    and a.c. voltmeters.

    The sensitivity of d.c. voltmeter is,

    I 5". = II Ifsd

    Let I fsd be 1 mA, hence the sensitivity becomes 1 kO/volt. The series resistance Rs is

    10 kO hence the 10 V d.c. input would cause exactly the full scale deflection, when

    connected with proper polarity.

    Let purely sinusoidal input of 1 0 V r.m.s. is applied.

    Erms 10 V

    Ep peak value = J 2 . Erms

    Now the rectified d.c. is pulsating d.c. hence meter will deflect proportional to the

    average value.

    But the diode conducts only for half cycle and meter movement is bypassed for

    another cycle. Hence it responds to half the average value of the a.c. input.

  • 8/9/2019 Chapter 0010

    19/21

    = 8.99", 4 5 V2 .

    Thus pointer will deflect for full scale if 10 V d.c. is applied and 4.5 V when 10 V

    r.m.s. sinusoidal input is applied.

    I Ede = 0.45 Erms I

    Thus the vi:l1ue of series multiplier can be obtained for a.c. voltmeters as,

    Rs

    = E ] de :... Rm

    de

    OASE rms----- RmIde

    The a.c. voltmeter using full wave rectifier is achieved by using bridge rectifier

    consisting of four diodes, as shown in the Fig. 2.13.

    Fig. 2.13 A.C. voltmeter using full wave rectifier

    Let 10 V r.m.s. purely sinu?oidal input be applied.

    0.636 Ep = 8.99 V

    "" 9 V

    Now this meter uses full wave rectifier and hence the average value of output over a

    cycle is same as average of the input over a cycle i.e. 9 V.

    Thus, the 10 V r.m.s. voltage is equal to 9 V d . c. for full scale def lection Thus the

    jPointer will deflect to 90% of full scale.

  • 8/9/2019 Chapter 0010

    20/21

    R, =Elk -

    RrnIde

    )1. Example 2.9 : All a.c. 'uoltmeter IIses half wave rectifier alld the basic I/ /e ter with filII

    scale d{:flection current of 1 mA and tire meter resistallce of 200 D. Calclliate the 1IlIIItiplier

    resistance for a 10 V 1'.111.5.rallge on the voltmeter.

    Solution: The meter uses half wave rectifier and input is 10 V r.m.s.

    Eav ~ (Eav over a cycle of input)

    0.6 Ep = 8.99 '" 9 V

    1"2 x 9 = 4.5 V

    Eo e R-J-- III

    ~

    0.45x JO _ 200

    Ix 10'

    )1. Example 2.10: All a.c. voltmeter IIses 11filII wave bridge rectifier and tire basic I//eter

    ll'itlr filII scale deflection current of 2 I1IA I 1l1a tire meter resistance of 500 D. Calclliate the

    1IlIIItiplier resistl1nce for 11 10 V 1'./11.5. rallge all the voltmeter.

    Solution : The meter uses f ull wave rectifier,

    0.9x 10 _ 500 = 4000 D2x 10 3

  • 8/9/2019 Chapter 0010

    21/21

    As the name indicates, this type of meter responds to the peak value of the