Performance Validation for Explosive Trace Detection Forensics@NIST 2012, Gaithersburg, Maryland, November 28-30, 2012 Certain commercial equipment, instruments, or materials may be identified in this presentation in order to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by the National Institute of Standards and Technology, nor is it intended to imply that the materials or equipment identified are necessarily the best available for the purpose. Mike Verkouteren, Research Chemist, Materials Measurement Division, NIST-DOC, Gaithersburg, MD USA “This material is based upon work supported by the Science and Technology Directorate of the U.S. Department of Homeland Security under Award Number HSHQDC-10-X-00552.”
20

Certain commercial equipment, instruments, or materials ...

Jun 01, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Certain commercial equipment, instruments, or materials ...

Performance Validation for Explosive Trace Detection

Forensics@NIST 2012, Gaithersburg, Maryland, November 28-30, 2012

Certain commercial equipment, instruments, or materials may be identified in this presentation in order to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by the National Institute of Standards and Technology,

nor is it intended to imply that the materials or equipment identified are necessarily the best available for the purpose.

Mike Verkouteren, Research Chemist, Materials Measurement Division,

NIST-DOC, Gaithersburg, MD USA

“This material is based upon work supported by the Science and Technology Directorate of the U.S. Department of Homeland

Security under Award Number HSHQDC-10-X-00552.”

Page 2: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Outline • Trace detection and challenges • Role of metrology and standards • Inkjet printing, ASTM limit of detection, cloud computing, and standard dirt

Page 3: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Trace Detection

Trace detection involves quantities of substance invisible to the unaided eye… typically less than a microgram.

Most commercial explosive trace detectors (ETDs) can detect in the low nanogram

range (a single crystal much smaller than the

width of a human hair!)

Page 4: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Trace Detection Challenges: Sampling

Trace chemical targets are vanishingly small

RDX crystals in C4

http://www.itstactical.com/centcom/news/youll-never-look-at-hay-bales-the-same/

http://showmehowto.net/wp-content/uploads/2012/07/788suitcase.jpg

Page 5: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Trace Detection Challenges: Specificity/Sensitivity

Contamination is ubiquitous, compositionally variable, and can interfere with detection

Page 6: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

0

100

200

300

400

500

600

700

800

900

1000

0 20 40 60 80 100 120

ETD Response Curves M

axim

um a

mpl

itude

(a.u

.)

Analyte mass (ng)

Vendor-supplied test materials

Page 7: Certain commercial equipment, instruments, or materials ...

Ink Jet Printing

Page 8: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Inkjet Metrology and Application to Trace Explosive Validation

Anal Chem 2009, 81, 8577

Langmuir 2011, 27, 9644

Anal Chem 2010, 82, 8519

Page 9: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

MicroFab Gravimetric Printer (JL4-XLB)

Production of Test Materials Using Ink Jet Printing

Accuracy 5 % Repeatability 0.5 % Deposition Range (<1 min)

1 pg to 1 μg 70 pL to 70 μL

Spatial Resolution

100 μm

Page 10: Certain commercial equipment, instruments, or materials ...

ASTM Limit of Detection &

Cloud Computing

http://www.v3.co.uk/IMG/473/149473/cloud-computing-security-services-provider-370x229.jpg

Page 11: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

LD: a fundamental yet misunderstood analytical metric We define the LD-90 as the lowest mass of a particular compound – introduced to the sampling inlet of a well-functioning contraband detection system – for which 90 % of independent measurements result in true detection, while the true non-detection probability is at least 90 % when measuring independent process blank samples. LD-90 influenced by combination of detector sensitivity, response repeatability, and specificity for the analyte against background interferences.

ASTM E54.01 Proposed Standard on

Limit of Detection for ETDs

http://www.chemometry.com/Research/LOD.html

Page 12: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

LOD Method for ETDs

Limit of detection determination for censored samples. Rukhin, A.L.; Samarov, D.V. (2011) 105, 188-194.

NIST method designed specifically for trace

explosive detectors as implemented by ASTM E54.01 subtask group*

* S. Leigh, A. Rukhin, J. Yen, J. Staymates, M. Verkouteren

Page 13: Certain commercial equipment, instruments, or materials ...

(1)

(2)

(3)

(4)

(5)

(6)

(7)

(8)

(9)

(10)

(11)

(12)

(13)

(14)

(15)

(16)

(17)

Page 14: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Web-based LOD Calculator • http://pubapps.nist.gov:8444/loda

– Input measurement data

– Data quality check

– LD-90 estimate & supporting information

ASTM Subtask Group: A. Heckert, K. Kwiatek, M. Verkouteren

Page 15: Certain commercial equipment, instruments, or materials ...

Standard Dirt

Page 16: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Standard “Dirts” Natural Matrix NIST SRMs

• Soils, sediments, dusts, air particulates, sludges, leaves…

• Immediately and internationally available

• Represent reasonable contamination sources

• Highly characterized

• Certified

• Homogeneous

• Stabilized

Page 17: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Standard Interferent Material (SIMdirt-1)

SRM 2704 (Buffalo River Sediment) 266.5 mg SRM 2585 (Household Dust) 83.7 mg SRM 2709a (San Joaquin Soil) 87.5 mg SRM 1650 (Diesel Particulate Matter) 5.1 mg 2-propanol (HPLC grade) 100 mL

One drop from squeeze dropper deposits 100 μg of SIMdirt-1 (U = 4%)

Page 18: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Limit of Detection Comparison An atypical surprise with SIMdirt-1

0

500

1000

1500

2000

2500

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5

Dose (ng)

Det

ecto

r res

pons

e (a

.u.)

0

500

1000

1500

2000

2500

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5

Dose (ng)

Det

ecto

r res

pons

e (a

.u.)

Analyses repeated, with 100 μg SIMdirt-1 added to each substrate

Detector responses from analyte on clean substrates

Page 19: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

SRM Preparers and Analysts M.M. Schantz

D.L. Poster S.J. Christopher

J.R. Sieber D.G. Friend

S.S. Vander Pol R.M. Lindstrom

R.O. Spatz S.E. Long

R.S. Popelka-Filcoff E.A. Mackey

B.E. Tomlin A.F. Marlow

L.J. Wood S.A. Wise

R.L. Watters J.R. Kucklick

S.D. Leigh

M.P. Cronise C.N. Fales P. Schubert J.M. Keller B.J. Porter H.M. Stapleton K.E. Murphy L.L. Yu R.L. Paul R. Zeisler S.A. Rabb A.P. Lindstrom G.C. Turk L.J. Wood L.L. Yu R.R. Greenberg E.A. Mackey B.S. McDonald

Page 20: Certain commercial equipment, instruments, or materials ...

Forensics@NIST, November 28-30, 2012

Summary • Several tools have been developed at NIST suitable

for performance testing and validation of trace detectors • Inkjet technology for accurate and precise printing

of pg-to-μg of compounds for reliable production of test materials and reference substrates

• Proposed ASTM method for determining LD-90, aided with a web-based calculator

• Standard interferent material (SIMdirt-1) for ETD interference response testing

[email protected]