Top Banner
Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy of meningioma using integrated stereotactic cone-beam-CT and adaptive re-planning (a-gkFSRT)” Strahlentherapie und Onkologie, July, pp. 1-5. 76 year old female, refused surgery Benign meningioma; Large, WHO Grade 1, left petrous bone Adaptive fractionated treatment with mask, chosen due to larger tumor volume, location and proximity to critical structures [Document reference number: 4513 371 1500]
5

Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

Apr 12, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

Case Report:

ADAPTIVE RADIOSURGERY

Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy of meningioma using integrated

stereotactic cone-beam-CT and adaptive re-planning (a-gkFSRT)”

Strahlentherapie und Onkologie, July, pp. 1-5.

• 76 year old female, refused surgery

• Benign meningioma; Large, WHO Grade 1, left petrous bone

• Adaptive fractionated treatment with mask, chosen due to larger tumor

volume, location and proximity to critical structures

[Document reference number: 4513 371 1500]

Page 2: Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

2 | Focus where it matters

WORKFLOW

MRI-based

pre-plan

7 days prior to treatment

Repeated at fraction 2-5

Treatment: 25Gy in 5 fractions over 5 days

CT: verify tumor position.

Co-registration MRI / CT.

Pre-plan adaptation.

CBCT to verify actual skull position.

Automatic co-registration to determine daily shift in

translation and rotation.

Automatic adaption by TPS to daily position;

recalculation of dose distribution.

Mask-based

immobilization

Page 3: Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

3 | Focus where it matters

TREATMENT

Outcome

• Prescribed to 46% isodose line for improved

coverage and optimal treatment time

• Total treatment time for each session around 20 mins

- Patient positioning 0.8mins

- CBCT plus acquisition 1.65mins

- CT data processing and adaptive planning 2.66mins

- Treatment 15.6mins

• Differences for the five daily CBCTs compared to

reference:

- For rotation:-0.59±0.49º/0.18±0.20º/0.05±0.36º

- For translation:0.94±0.52mm/-0.08±0.08mm/-1.13±0.89mm

- The TPS automatically adapted the shot positions to the daily

position and recalculated the dose distribution (online adaptive

planning)

• Over all fractions an intrafractional movement of 0.13

± 0.04mm noted

› ‘Adaptive daily re-planning was accurate

and yielded quality measures, e. g.,

coverage, selectivity, and gradient for the

delivered dose identical to the initial

values’

› Treatment well tolerated (slight headache

after 1st treatment resolved with

dexamethasone)

› No neurological symptoms at

6 weeks and 4 month follow-up

Details

Page 4: Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

4 | Focus where it matters

CONE BEAM CT IMAGES AND INTRAFRACTIONAL MOVEMENT

Intra-fractional movements

Shown are the measurements (in mm) of the HDMM

(high-definition motion management) over time.

On days 2 and 3, no intrafractional movement was

detected.

Dose distribution based on Cone Beam CT

The isodose lines represent the following doses (outer

to inner lines): 5 Gy,12 Gy,25 Gy, and 50 Gy.

Page 5: Case Report: ADAPTIVE RADIOSURGERY517a7b90-aaab-4aee-8a19...Case Report: ADAPTIVE RADIOSURGERY Stieler et al. (2016) “Adaptive fractionated stereotactic Gamma Knife radiotherapy

Thank you