  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    1/20

    CALIBRATION OF PH METER

    What is Calibration of pH Meter?It is a process of preparing the pH Meter in checking its

    accuracy reading using buffer solutions..What is Benefits?

    Accuracy of pH reading

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    2/20

    Frequency of Calibration?Weekly.How the process works?1.Prepare the following itemsa.) Buffer 4 Solutionb.) Buffer 7 Solutionc.) Paper Toweld.) Alcohole.) 2 pcs. of 40 ml beaker

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    3/20

    2.Clean pH electrode and temperature probe usingalcohol then wipe using tube wiper.3.Rinse the pH electrode and temperature probe using D.I.

    water then blot dry using paper towel.4.Let the pH electrode and temperature probe hang untilset-up of pH meter is done.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    4/20

    5.Press standby.6.Press the set-up twice.7.Press enterto clear all existing buffers.8.Continue press the set-upto display the desired buffergroup. (2 4 7 10 12)..

    9.Press enterto select the group and return to measurescreen..

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    5/20

    10.Continue press the set-upagain to display the degreesCelsius temperature icon.

    11.Press enterto acceptC temperature unit and return to

    the measure screen..12.Continue press the set-upagain to view the two decimalpoint resolution selection.(88.88)

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    6/20

    13.Press enterto accept the resolution and return to themeasure screen.14.Prepare the 2 samples buffer solutions to 40 ml beaker..

    15.Immerse the pH electrode and temperature probe tobuffer 4 solution into the center of the solution so that thebulb of the electrode is completely covered..

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    7/20

    16.Wait until stableappears on the screen then record theBEFORE CALIBRATIONon calibration checklist.17.Press the standardtwice then wait to appear the phrase

    of GOODELECTRODEindicated below of the meter thenrecord the AFTERCALIBRATIONon calibration checklist.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    8/20

    18.Rinse the pH electrode and temperature probe using D.I.water then blot dry using paper towel.

    19.Do the same procedure from number 15-18 to buffer 7.20.Return the 2 buffer solutions and alcohol to its properplace..

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    9/20

    Buffer 4 and Buffer 7 Solution

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    10/20

    pH Meter and Conductivity Meter

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    11/20

    CALIBRATION OF CONDUCTIVITY

    METER

    What is checking of conductivity meter?It is a process of preparing the conductivity meter in

    checking its accuracy reading using conductivity standardsolution.What is benefits?

    Accuracy of conductivity reading..

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    12/20

    Frequency of Calibration?

    Weekly..How the process works?1.Prepare the following itemsa.)Conductivity solutionsb.)40 ml beaker

    c.)Alcohold.)Paper Towel

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    13/20

    2.Press standbyat conductivity meter.3.Clean the conductivity and temperature probe usingalcohol then wipe using paper towel.4.Rinse the conductivity and temperature probe using D.I.

    water then blot dry using paper towel.5.Let the conductivity and temperature probe hang untilset-up of conductivity meter is done.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    14/20

    6.Press set-upagain to view the measuring conductivity in

    S then press enter.7.Press set-upagain to view the nominal constant of 1.000

    then press enter..8.Press set-up again to view the 2 cell constant (series ofdashes_ _ _ _). then press enter.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    15/20

    9.Pres set-upagain to view the temperature coefficient of1.5 then press enter.10.Press set-upagain to view the temperature unit C thenpress enter..

    11.Immerse conductivity probe into a portion of standardsolution.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    16/20

    12.Record the BEFORE CALIBRATION readingcorresponding with the temperature value of 25 degrees.13.Check for the temperature reading if the value matchthe standard solution displayed.(the conductivity andtemperature value can be viewed at conductivity solution

    bottle..14.Press the standard twice to access the standardizationscreen.

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    17/20

    15.Record the AFTER CALIBRATION reading correspondingwith the temperature value into calibration checklist

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    18/20

    Conductivity Solution

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    19/20

    pH Meter and Conductivity Meter

  • 8/11/2019 Calibration of PH Meter and Conductivity Meter

    20/20

    Thank You!!RAFAEL MACARAIG