Calibration of high-frequency wattmeters used for standby power testing to IEC 62301 August 2011 I. Budovsky and D. Georgakopoulos National Measurement Institute, Australia Low Frequency Electrical Standards
Mar 30, 2015
Calibration of high-frequency wattmeters used for standby power testing to IEC
62301
August 2011I. Budovsky and D. Georgakopoulos
National Measurement Institute, Australia Low Frequency Electrical Standards
Overview
• IEC 62301 definition of standby power• Motivation for this work• Calibration waveforms• System description• Traceability• Experimental results• Summary
Standby power definition according to IEC 62301
“the lowest power consumption mode which cannot be switched off (influenced) by the user and that may persist for an indefinite time when an appliance is connected to the main electricity supply and used in accordance with the manufacturer’s instructions.”
Motivation• Electrical appliances operating in standby mode
consume small amount of electrical power• The current in standby mode can take the form of
pulses or spikes• Required measurement uncertainty by IEC 62301
< 2%, measured power > 0.5 W< 0.01 W, measured power < 0.5 W
• The IEC standard specifies current waveforms with CF up to 10
RMSI
ICF max CF=1.41 for a sinewave
CF=1.00 for a square wave
… Motivation
• Because of the high CF specified by the IEC standard, power meters for standby power cannot be calibrated with sinusoidal waveforms
• Hence there is a need to develop techniques to traceably verify power meters for the specific conditions described in IEC 62301
Calibration waveforms• Voltage
– sinewave with maximum distortion of 2% (up to and including the 13th harmonic) –crest factor from 1.34 to 1.49– use of a sinewave (THD <0.02%) for calibration is good enough
• Current– specification only for CF (up to 10)– no distortion specified– recommends the instrument must measure up to at least the 50th harmonic
Current waveforms
-1.5
-1
-0.5
0
0.5
1
1.5
0 0.2 0.4 0.6 0.8 1 1.2
Normalised time (s)
Am
plit
ude
I sinc
I pulse
Same CF and fundamental magnitudeBut different • harmonic content• peak values and• rms values when the harmonics are included
00.020.040.060.08
0.10.120.140.160.18
0 20 40 60 80 100
Harmonic number
Har
mon
ic m
agni
tude Pulse
Sinc
Selecting current waveform
We are looking for a current waveform to test the current channel of a power meter which:
• satisfies the IEC requirements or recommendations– CF requirement– contain pulses or spikes– high bandwidth
• is physically realisable
In our calibration system we use a pulsed current
Advantages of using pulsed current• A pulse is a wideband signal and is a more
challenging test for the measurement system
I-V converter Amplifier LPF S/H ADC
• Sampling frequency > 2 maximum signal frequency• The channel frequency response known for the signal frequencies
• A crest factor of 10 is more easily achieved with a pulsed current waveform while maximizing the measured power
• Satisfies the 50th harmonic recommendation of the IEC standard
• Simple relation of the CF with the power -20
0
20
40
60
80
100
120
1 10 100 1000 10000
Harmonic number
Mag
nit
ud
e (%
of
fun
dam
enta
l)
Fs -fx
LPF, Fs>2×fmax
fx Fs ff
H(j)
fmax
-1.5
-1
-0.5
0
0.5
1
1.5
0 1 2 3 4 5 6 7
U (t)
I
Measured power as a function of CF
t1 t2 t3 t4 T t
i(t)
Im
-Im
3412
43
21
243
243
2424
ttttt
tTttTt
tTttTt
p
pp
pp
t1 t2 t3 t4 T t
i(t)
Im
-Im
3412
43
21
243
243
2424
ttttt
tTttTt
tTttTt
p
pp
pp
prms
m
t
T
I
ICF
2
2)(2sin
2
CFIVP mm
Zero power factor
tp: pulse widthT: pulse period
22 2sinc
2sin
2
CF
nnI
CFI mn
Harmonics
CF T/tp
5 50
10 200
20 800
Calibration waveforms (… continued)V
I 0°
cos11IVP • The power for the selected waveforms is concentrated only on the fundamental
• The calibration system must be characterized at power frequencies
1 5 9 13 17 21 25 29 33 37 41 45 49 53 57 61 65 69 73 77
Harmonic voltage
Mag
nit
ud
e o
f vo
ltag
e an
d c
urr
ent
har
mo
nic
s
IV
0˚ 90˚45˚
Calibration system
Power Generation Power Measurement
System descriptionHigh frequency Thermal Power Comparator (TPC)Measures the difference between the unknown ac power (ACV and ACI) and the known dc (DS1 and DS2)
TPC: multi-junction thermal converters, no aliasing
UUT
Rs
VD
TPC
DCS1
DCS2
ACV
ACI
DCV
DCI
Dual Channel VoltageSource
VoltageAmplifier
TCA
10 MHz
i(t)
t
TCA minimum Distortion, highbandwidth
Thermal Power Comparator
Controlled Switches Amplifiers
U~
I~
Uu
Ui
Temperature Compensation
VD
R
TC1
TC2
OutputAmplifier
S1
S2
S3
S4
S5
S6
S7
Uu1
Uu2
Ui1
Ui2
I
II
III
IV
Uc1
Uc 2o
o
S
Differential
Sum-and-Difference
Y
Traceability
• Thermal Power Comparator• Current shunts• Voltage divider • Voltage measurement error (dc only)• Frequency (negligible)
Each of these components has been characterized and is traceable to the NMIA standards of voltage, current, resistance, frequency and electrical power
Magnitude andPhase errors
Power measurement
Uncertainty budgetComponent Distribution
Ui
(W/VA) ki ci
ui
(W/VA) ciui
(W/VA) i
TPC AC-DC difference Normal 67 2 1 33.6 33.6 30
Shunt Normal 10 2 1 5.0 5.0 30
Resistive voltage divider Normal 20 2 1 10.0 10.0 30
DVM1 Normal 8 1 1 8.0 8.0 30
DVM2 Normal 8 1 1 8.0 8.0 31
Type A Normal 50 2 1 25.0 25.0 19
Source adjustment Normal 10 2 1 5.0 5.0 30
Combined standard uncertainty, uc (W/VA) 45.1
Effective degrees of freedom, ueff 65 Coverage factor, k 2.00
Expanded uncertainty U = kuc (W/VA) 90.1
Typical uncertainties of commercial wattmeters compared to IEC 62301
requirementsCurrent Range Current Applied Current Phase Angle Nominal Power Uncertainty
Uncertainty requiredby IEC62301
(Apk) (Apk) Crest Factor (degrees) (W) (W) (W)
3 0 112.564 ± 0.06 ± 2.3 3 + 60 56.282 ± 0.03 ± 1.1 3 - 60 56.282 ± 0.04 ± 1.1 5 0 40.703 ± 0.03 ± 0.81 10 0 10.182 ± 0.06 ± 0.20 3 0 22.513 ± 0.012 ± 0.45 3 + 60 11.256 ± 0.006 ± 0.23 3 - 60 11.256 ± 0.006 ± 0.23 5 0 8.141 ± 0.005 ± 0.16 10 0 2.036 ± 0.012 ± 0.04 20 0 0.509 ± 0.006 ± 0.01
5 3
1 0.6
Summary• A traceable measurement system for calibration of
wattmeters used in standby electrical power testing has been developed
• The measurement system is based on a thermal power comparator, precision current shunts, inductive and resistive voltage dividers and precision amplifiers
• Each of these components has been evaluated in a traceable way
• The expanded uncertainty of the system is better than 100 W/VA for a crest factor of 3
• The system has been tested for current waveforms with crest factors up to 20
National Measurement InstituteBradfield RoadWest Lindfield NSW 2070Australia
Phone: + 61 2 8467 3600
Email: [email protected]