Calibration Report Job number: IND201100148 Report number: IND200900105/4 Date: 06/07/2011 Pg 1 of 2 Customer Mr. Concrete Kodigamam road, Puloly South Point Pedro. Unit under Test Digital Caliper Make Serial no. /10 no. Range Division Precise Not Available / MT031 0-150 mm 0.01 mm Method of test Test was carried out refered to the international standard ISO 6906 : 1984. Uncertainty calculations were done Refered to EA-4/02, NABL 141 Standard Laboratory working standard Grade 1 gauge blocks set no 090013 traceable to NIST via 821/276375 - 08, PTB Via 4937 PTB 06 Temperature and Humidity meter: SN:1 0745936 Place of test performed : SGS Lanka Calibration Lab Results of Calibration Temperature during test Note: The calliper reading was checked at three different positions of the measuring jaw,viz. tip, mid portion, and the base. Gauge setting Caliper reading, mm mm Tip Mid portion Base Zero setting 0.00 20.00 20.00 20.00 20.01 40.00 40.00 40.00 40.00 50.00 50.00 50.01 50.00 70.00 69.99 69.99 70.00 100.00 99.99 99.99 100.00 120.00 119.99 120.00 120.00 150.00 150.00 150.00 150.00 (Company No. PV 19226) SGS Lanka 1M Ltd Industrial Services, AEC Building, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e [email protected]www.sgs.com Head Office, 3< Floor, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e [email protected]www.sgs.com Member of the SGS Group
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Customer Mr. ConcreteKodigamam road, Puloly SouthPoint Pedro.
Unit under Test Digital CaliperMakeSerial no. /10 no.RangeDivision
PreciseNot Available / MT0310-150 mm0.01 mm
Method of testTest was carried out refered to the international standard ISO 6906 : 1984. Uncertainty calculations were doneRefered to EA-4/02, NABL 141
StandardLaboratory working standard Grade 1 gauge blocks set no 090013 traceable to NIST via 821/276375 - 08,PTB Via 4937 PTB 06
Temperature and Humidity meter: SN:1 0745936
Place of test performed : SGS Lanka Calibration Lab
Results of Calibration
Temperature during test
Note: The calliper reading was checked at three different positions of the measuring jaw,viz. tip, mid portion,
and the base.
Gauge setting Caliper reading, mmmm Tip Mid portion Base
SGS Lanka 1M Ltd Industrial Services, AEC Building, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e [email protected] www.sgs.com
Head Office, 3< Floor, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e [email protected] www.sgs.comMember of the SGS Group
Method of test IThe accuracy of this irhlstrument was checked in accordance with Linear measurements were madeby Linear Scale. Weight is applide only to remove wrinkles
StandardThis certificate provides traceability of measurement to national standards through NIST Japan andIndustrial Technology Institute Sri Lanka
Place of test performed:
Results of Calibration
Condition room at the SGS Lanka Pvt. Ltd.
Temperature during test
Zero Error:
Range 1 Value 1(mm) (mm)0-10 1000-20 2000-30 3000-40 4000-50 4990-60 5990-70 6990-80 7990-90 8990-100 998
C I b 2 S . L nka t + 94 (0) 11 5395118 I + 94 (0) 114717878 e [email protected] www.sgs.comI d t' IS' es AEC Building 140 Vauxhall Street, oom 0 - , n a ,
SGS Lanka 1M) Ltd HnudsrOlaffiCee~~cFI~or 140 vau~all s;reet, Colombo _2. Sri Lanka, t + 94 (0) 115376280 f + 94 (0) 115376281 e [email protected].