Top Banner
GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the standard ISO/IEC 17025:2017 ''General Requirements for the Competence of Testing & Calibration Laboratories'' for its facilities at PLOT NO. 126, POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA in the field of CALIBRATION Certificate Number: CC-2205 Issue Date: 17/01/2020 Valid Until: 17/05/2021 This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continued satisfactory compliance to the above standard & the relevant requirements of NABL. (To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org) Signed for and on behalf of NABL N. Venkateswaran Chief Executive Officer
80

CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

Mar 10, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORSGROUP

has been assessed and accredited in accordance with the standard

ISO/IEC 17025:2017''General Requirements for the Competence of Testing &

Calibration Laboratories''for its facilities at

PLOT NO. 126, POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

in the field of

CALIBRATIONCertificate Number: CC-2205

Issue Date: 17/01/2020 Valid Until: 17/05/2021

This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continuedsatisfactory compliance to the above standard & the relevant requirements of NABL.

(To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org)

Signed for and on behalf of NABL

N. VenkateswaranChief Executive Officer

Page 2: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

Permanent Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 1 kHz 100 µA to 200 µA 0.26% to 0.39% Using 6 ½ DigitMultimeter 8846A byDirect Method

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz 10 A to 3200 A 0.4% Using Standard CurrentTransformer with 6 ½Digit Multimeter 8846Aby Direct Method

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

0.2 mA to 100 mA 0.39% to 0.18% Using 6 ½ DigitMultimeter 8846A byDirect Method

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

100 mA to 10 A 0.18% to 0.29% Using 6 ½ DigitMultimeter 8846A byDirect Method

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 1 kHz to 5kHz

3 mA to 3 A 0.16% to 3.54% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 1 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 3: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz 20 A to 1000 A 0.79% Using MPF 5502 A withCurrent Coil by DirectMethod

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

10 A to 20 A 0.17% to 0.36% Using Multi ProductCalibrator 5502A byDirect Method

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

3 A to 10 A 0.16% to 0.17% Using Multi ProductCalibrator 5502A byDirect Method

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1KHz

30 uA to 3 mA 1.44% to 0.16% Using Multi ProductCalibrator 5502A byDirect Method

10 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10 uA to 100 mA 0.36% to 0.064% Using 6 ½ DigitMultimeter 8846A byDirect Method

11 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100 mA to 3 A 0.064% to 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 2 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 4: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

12 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 3 A to 10 A 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

13 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 µA to 100 µA 0.26% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

14 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 100 uA to 3 mA 0.044% to 0.031% Using Multi ProductCalibrator 5502A byDirect Method

15 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 20 A to 1000 A 0.64% Using MPF 5502 AWith Current Coil byDirect Method

16 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 A to 20 A 0.11% to 0.38% Using Multi ProductCalibrator 5502A byDirect Method

17 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 mA to 3 A 0.031% to 0.11% Using Multi ProductCalibrator 5502A byDirect Method

18 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(PhaseError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 3.34min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

19 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(RatioError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 0.047% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 3 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 5: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

20 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC HIGH VOLTAGE 1 kV to 40 kV 3.2% Using HV Divider withKV Meter

21 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 1 kV to 5 kV 3.2% to 3.8% Using Udeyraj makeHV Divider withindicator by DirectMethod

22 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 5 kV to 25 kV 3.8% to 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

23 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 1 kHz 1 mV to 20 mV 4.72% to 0.33% Using 6 ½ DigitMultimeter 8846A byDirect Method

24 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 50 Hz to 1kHz

20 mV to 1000 V 0.54% to 0.10% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 4 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 6: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

25 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

26 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A/5ASecondary from 1% to120% of rated current

2000 A to 7500 A 3.4min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

27 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

28 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

29 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 0.05% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 5 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 7: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

30 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A/5ASecondary from 1% to120% of Rated Current

2000 A to 7500 A 0.085% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

31 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

32 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

33 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 1 kV to 5 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

34 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 5 kV to 25 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 6 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 8: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

35 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 mOhm to 1 Ohm 0.42% to 0.36% Using 6 ½ DigitMultimeter 8846A byV/I Method

36 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 MOhm to 1000MOhm

0.013% to 2.31% Using 6 ½ DigitMultimeter 8846A byDirect Method

37 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 Ohm to 1 MOhm 0.36% to 0.013% Using 6 ½ DigitMultimeter 8846A byDirect Method andVoltage/ CurrentMethod

38 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 1 mV to 100 mV 0.42% to 0.085% Using 6 ½ DigitMultimeter 8846A byDirect Method

39 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 100 mV to 1000 V 0.085% to 0.007% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 7 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 9: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

40 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/110 4.03min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

41 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/63.5 3.16min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

42 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/110 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

43 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/63.5 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

44 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Voltage Ratio 0.8 to 2021 ratio 0.16% Using 6 ½ DigitMultimeter 8846A andFluke 5502A by DirectMethod

45 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Power 50 [email protected] UPF 120/2240 V 0.1A to 20 A

1.2 W to 4800 W 3.47% to 0.23% Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 8 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 10: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

46 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Vertical deflectionHorizontal Deflection 1kHz

5m V to 50 V 2.64% to 0.40% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

47 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz 2 mV to 10 mV 1.76% to 0.41% Using Multi ProductCalibrator 5502A byDirect Method

48 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz to 10kHz

30 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

49 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

10 mV to 300 mV 0.41% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

50 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 mV to 300 V 0.044% to 0.1% Using Multi ProductCalibrator 5502A byDirect Method

51 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 9 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 11: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

52 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @ 1 kHz 10 pF to 1 nF 2.98% to 1.2% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

53 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1 kHz 9 nF to 1 mF 1.26% to 0.80% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

54 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1kHz 1 nF to 9 nF 1.7% to 1.26% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

55 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Power @ 1 V to600 V, 1A to 20 A

1 W to 12 KW 0.37% to 0.25% Using MultiproductCalibrator 5502A withPower Option by DirectMethod.

56 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 mohm to 1 mohm 0.6% to 0.6% Using Shunt By DirectMethod

57 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 Ohm to 100 Ohm 0.25% to 0.072% Using Multi ProductCalibrator 5502A byDirect Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 10 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 12: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

58 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 kOhm to 100MOhm

0.021% to 0.58% Using Multi ProductCalibrator 5502A byDirect Method

59 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 ohm to 100 kOhm 0.072% to 0.021% Using Multi ProductCalibrator 5502A byDirect Method

60 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 1 mV to 100 mV 0.36% to 0.013% Using Multi ProductCalibrator 5502A byDirect Method.

61 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 100 mV to 30 V 0.013% to 0.0064% Using Multi ProductCalibrator 5502A byDirect Method.

62 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 30 V to 1000 V 0.0064% to 0.007% Using Multi ProductCalibrator 5502A byDirect Method.

63 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

High Resistance 100 kohm to 1 Tohm 1.2% to 2.4% Using DecadeResistance Box byDirect Method

64 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Inductance @1 kHz 1 mH to 10 H 1.53% to 2.03% Using DecadeInductance Box byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 11 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 13: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

65 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics(Current) upto 41st

100 mA to 6 A 0.75% Using FlukeMultiProduct Calibrator5502A

66 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics upto41st Voltage

100 mV to 60 V 0.75% Using FlukeMultiproduct Calibrator5502A

67 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Oscilloscope DCFunction

2 mV to 33 V 6.59% to 0.33% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

68 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Power Factor 50 Hz 0.1 to UPF 0.05PF Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

69 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage ratio 0.8 to 400 0.5% to 0.3% Using TransformerRatio Standard byComparison Method

70 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage Ratio 400 to 2021 0.5% to 0.3% Using TransformerRatio Standard byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 12 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 14: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

71 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-)200 °C to 800 °C 0.51°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

72 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)C-Type

0 °C to 2315 °C 1.4°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

73 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)E-Type

(-)270 °C to 1300 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

74 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)J-Type

(-)200 °C to 1200 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 13 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 15: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

75 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)K-Type

(-)250 °C to 1137 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

76 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)L-Type

(-)200 °C to 600 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

77 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)N-Type

(-)270 °C to 1300 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

78 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)R-Type

(-)50 °C to 1767 °C 0.9°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 14 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 16: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

79 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)S-Type

(-)50 °C to 1767 °C 0.8°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

80 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)T-Type

(-)270 °C to 400 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

81 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)U-Type

(-)200 °C to 600 °C 0.5°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

82 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-) 200 °C to 800 °C 0.27°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 15 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 17: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

83 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)EType

(-)200 °C to 1000 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

84 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)JType

(-)210 °C to 1200 °C 0.37°C Using Multi ProductCalibrator 5502A byDirect Method

85 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)KType

(-)200 °C to 1372 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

86 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)NType

(-)200 °C to 1300 °C 0.61°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 16 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 18: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

87 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)RType

0 °C to 1700 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

88 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)SType

0 °C to 1700 °C 0.58°C Using Multi ProductCalibrator 5502A byDirect Method

89 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)TType

(-)200 °C to 400 °C 0.70°C Using Multi ProductCalibrator 5502A byDirect Method

90 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)UType

(-)200 °C to 600 °C 0.65°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 17 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 19: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

91 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)BType

600 °C to 1820 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

92 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)CType

200 °C to 2315 °C 0.95°C Using Multi ProductCalibrator 5502A byDirect Method

93 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 MHz 0.084% to 0.014% Using 6 ½ DigitMultimeter 8846A byDirect Method

94 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 0.1 sec to 1 hr 0.013sec to 2.32sec Using Digital TimeInterval Meter By DirectMethod

95 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 hr to 24 hr 2.32sec to 5.18sec Using Digital TimeInterval Meter Bycomparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 18 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 20: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

96 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Bandwidth 50 MHz to 250 MHz 6.21% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

97 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 10 Hz to 300 kHz 0.086% to 0.0094% Using Multi ProductCalibrator 5502A byDirect Method

98 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 300 kHz to 2 MHz 0.0094% to 0.49% Using Multi ProductCalibrator 5502A byDirect Method

99 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Time Base 2 n Sec to 5 Sec 0.3% to 0.67% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

100 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

10 RPM to 5000 RPM 2.1RPM Using Tachometer.

101 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

5000 RPM to 30000RPM

19RPM Using Tachometer.

102 MECHANICAL-ACOUSTICS

Acoustic Pressure -Sound Level Meter @ 1kHz

114 db 1.1db Using Sound LevelCalibrator.

103 MECHANICAL-ACOUSTICS

Acoustic Pressure -Sound Level Meter @ 1kHz

94 db 0.96db Using Sound LevelCalibrator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 19 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 21: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

104 MECHANICAL-DENSITY ANDVISCOSITY

Hydrometer - DensityHydrometer, TwaddleHydrometer, BaumeHydrometer, SpecificGravity Hydrometer,BrixHydrometer,Lactometer, Alcometer.

0.6 g/ml to 2.000 g/ml 0.00025g/ml Using WeighingBalance with resolution0.01 mg by Hydrostaticweighing method asper NISP SP 250-78Standard.

105 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Bevel ProtractorL.C.: 1' 0 ° to 360 ° 3.1arc min Using Angle GaugeBlock & Surface Plate.

106 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Bore Gauge(Transmission Only)L.C.: 1 µm

Up to 1 mm 3.1µm Using Dial CalibrationTester.

107 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

1000 mm to 2000 mm 22.5µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

108 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

0 to 1000 mm 14µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 20 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 22: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

109 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

0 to 600 mm 10µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

110 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Coating ThicknessGauge(Digimatic/Analog)L.C.:0.1 µm

0 to 6 mm 3.55µm Using Thickness Foils.

111 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Comparator Stand 300 mm to 300 mm 3.7µm Using Gauge Block;Elec. Probe & SurfacePlate.

112 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cube Mould 0 to 300 mm, 0 to 360° 16.8 µm, 3.6 min Using Profile Projector& Digital Verniercaliper, BevelProtractor byComparison Method

113 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical MeasuringPin

0.5 mm to 25 mm 1.3µm Using Gauge Block Set& Electronic Probe.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 21 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 23: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

114 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical SettingMasterDiameterVariation

>100 mm to 160 mm 3.3µm Using Gauge Block Set& Electronic Probe

115 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical SettingMasterDiameterVariation

Up to 100 mm 2.8µm Using Gauge Block Set& Electronic Probe.

116 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Degree Protractor /Combination Set L.C.:1°

0 ° to 360 ° 20arc min Using Slip GaugeBlock, Sine Bar; AngleGauge Block & SurfacePlate.

117 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth Caliper(Vernier/Dial/Digital)L.C.: 10 µm

0 to 300 mm 8µm Using Gauge Block;Long Gauge Block &Surface Plate

118 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth MicrometerL.C. :1 µm

0 to 300 mm 4.0µm Using Gauge BlockSet; Long Gauge Block& Surface Plate.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 22 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 24: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

119 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial CalibrationTesterL.C.: 0.1 µm

0 to 25 mm 1.12µm Using Electronic Probe

120 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ComparatorL.C.:0.5 µm

-50 µm to 50 µm 1.5µm Using Gauge Block Set& Comparator Stand.

121 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ThicknessGaugeL.C.: 1 µm

0 to 10 mm 0.8µm Using Gauge BlockSet.

122 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ThicknessGaugeL.C.: 10 µm

0 to 100 mm 8µm Using Gauge BlockSet.

123 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

ElectronicComparatorL.C.: 0.1µm

0 to 25 mm 0.9µm Using Gauge BlockSet.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 23 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 25: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

124 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Electronic HeightGaugeL.C.: 0.5 µm

0 to 600 mm 8µm Using Gauge Block;Long Gauge & SurfacePlate.

125 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 1 µm

>25 mm to 300 mm 4.0µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

126 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 1 µm

0 to 25 mm 0.8µm Using Mic.Check Set.

127 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

>300 mm to 600 mm 12.9µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

128 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

>600 mm to 1000 mm 14.0µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 24 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 26: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

129 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

0 to 300 mm 6.9µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

130 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Digital)L.C.: 0.1µm

0 to 25 mm 0.4µm Using Gauge BlockSet.

131 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Feeler Gauge Up to 2 mm 1.5µm Using DigimaticMicrometer.

132 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

FlankinessGauge/ElongationGauge/Receiver Gauge

0 to 100 mm 16.8 µm Using Profile Projector& Digital VernierCaliper

133 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 1000 mm 20µm Using Caliper Checker;Gauge Block & SurfacePlate.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 25 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 27: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

134 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 600 mm 11µm Using Caliper Checker;Gauge Block & SurfacePlate.

135 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Caliper TypeL.C. : 1µm

5 mm to 50 mm 3.0µm Using Gauge Block Set& Gauge BlockAccessories.

136 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Caliper TypeL.C. : 10µm

5 mm to 300 mm 5.0µm Using Gauge Block Set& Gauge BlockAccessories.

137 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Stick Micrometer (2points)Basic Travel ofMicrometer (Plain)L.C.:10 µm

50 mm to 63 mm 4.2µm Using Gauge BlockSet; Long Gauge BlockSet & Mic Check Set &Electronic Probe withComparator.

138 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Stick Micrometer (2points)Overall LengthAccuracy withExtension Rod(Stick)L.C.: 10 µm

50 mm to 1050 mm 12.0µm Using Gauge BlockSet; Long Gauge BlockSet & Mic Check Set &Electronic Probe withComparator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 26 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 28: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

139 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Internal CaliperGaugeL.C.: 10 µm

2.5 mm to 250 mm 7.1µm Using Gauge Block Set& Gauge BlockAccessories.

140 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :1 µm

Up to 0.14 mm 1.6µm Using Dial CalibrationTester.

141 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :10 µm

Up to 1 mm 5.9µm Using Dial CalibrationTester.

142 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :2 µm

Up to 2 mm 1.9µm Using Dial CalibrationTester.

143 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Limit Gauge/CDGauge/PCD Gauge

0 to 200 mm 16.8 µm Using Profile Projector,LMM & 2Height Gaugeby Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 27 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 29: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

144 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Scale 0 to 2000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit

145 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Tape 0 to 100000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit.

146 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Tape (PiTape)

0 to 15000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit.

147 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Micrometer Setting Rod 300 mm to 1000 mm 10.0µm Using Gauge BlockSet; Long Gauge BlockSet & ElectronicComparator withProbe.

148 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Micrometer Setting Rod Up to 300 mm 4.5µm Using Gauge BlockSet; Long Gauge BlockSet & ElectronicComparator withProbe.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 28 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 30: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

149 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Parallel Thread PlugGauge

2 mm to 100 mm 4.1µm Using FCDM,Cylindrical SettingMaster & ThreadMeasuring Wires.

150 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Pistol CaliperL.C.: 50µm

Up to 100 mm 26.1µm Using Gauge BlockSet.

151 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Plug Gauge 1 mm mm to 100 mm 2.0µm Using Gauge Block Set& Electronic Probe.

152 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Plug Gauge 100 mm to 160 mm 5.0µm Using Gauge Block Set& Electronic Probe.

153 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Ring Gauge 3 mm to 100 mm 2.0µm Using ULM & MasterRing.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 29 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 31: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

154 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plunger DialGaugeL.C. : 1 µm

0 mm to 25 mm 1.7µm Using Dial CalibrationTester & Gauge BlockSet.

155 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plunger DialGaugeL.C. : 1 µm

0 mm to 50 mm 2.2µm Using Dial CalibrationTester & Gauge BlockSet.

156 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Radius Gauge/RadiusTemplete

Up to 40 mm 15.2µm Using Profile Projector.

157 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Screw / Thread PitchGauge - Angle

55° & 60° 9.5Min of Arc Using Profile Projector

158 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Screw / Thread PitchGauge - Pitch

Up to 7 mm 12µm Using Profile Projector

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 30 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 32: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

159 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Sine Bar Up to 300 mm 9arc sec Using Gauge Block;Angle Gauge Block;Elec. Probe & SurfacePlate.

160 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Snap Gauge >100 mm to 200 mm 6.0µm Using Gauge BlockSet.

161 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Snap Gauge 1 mm to 100 mm 3.0µm Using Gauge Block Set

162 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Spirit / PrecisionLevelL.C.: 20 µm/m

Up to 300 mm 14.4µm/m Using Electronic Level.

163 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Surface Plate 1000 mm X 1000 mm 1.2 xv(L+W)/125 L&Win mm

Using Electronic Level,L.C.1µm/m.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 31 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 33: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

164 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Taper Scale 0.1 mm to 16 mm 13.6µm Using Profile Projector.

165 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Taper Thread PlugGauge

2 mm to 100 mm 4.1µm Using FCDM,Cylindrical SettingMaster & ThreadMeasuring Wires.

166 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Templates/Plain WorkPiece/Inspection JigandFixture/Moulds/LeverArm/ MasterConnecting Rod

0 to 600 mm, 0 to 360° 16.8 µm, 3.6 min Using Gauge Block set,2D Height Gauge,Profile Projector, LLM,Electronic Comparator,Digital Vernier Caliper,Digital Micrometer,Bevel Protractor, AngleGauge Set, Sine Bar,Gauge Block set andElectronic Comparatorby Comparison Method

167 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Test Sieves Up to 125 mm 13µm Using Profile Projector& Digital VernierCaliper

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 32 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 34: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

168 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thickness Foils 0.01 mm to 12 mm 1.12µm Using Electronic Probe.

169 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thread MeasuringWires

0.17 mm to 6.35 mm 1.3µm Using Electronic Probe.

170 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thread Ring Gauge 3 mm to 100 mm 2.0µm Using ULM & MasterRing.

171 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Ultrasonic ThicknessGauge.L.C.: 0.01 mm

0 to 200 mm 72µm Using Steel MasterRod.

172 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Vickers/Knoop/Rockwell Diamond ConeIndenter

0 to 100 mm, 0 to 360° 16.8 µm, 3.6 min Using Profile Projector

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 33 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 35: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

173 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Weld Fillet Gauge Up to 100 mm 16.8 µm Using Profile Projectorby Comparison method

174 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Weld/Hi-Lo Gauge,Bridge Cam Gauge

0 to 60 mm, 0 to 60° 16.8 µm, 3.6 min Using Profile Projector,Gauge Block Set, Tapeand Scale calibrator byComparison method

175 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge / HeightSetting Master

>100 mm to 160 mm 5.0µm Using Gauge Block Set& Electronic Probe.

176 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge / HeightSetting Master

0 mm to 100 mm 2.0µm Using Gauge Block Set& Electronic Probe.

177 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge/PaddleGauge

0 to 200 mm 16.8 µm Using Profile ProfileProjector, Gauge BlockSet, & Electronic Probe

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 34 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 36: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

178 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Wire Gauge 0.025 mm to 8 mm 12µm Using Profile Projector.

179 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Electronic LevelL.C.: 1µm/m

Up to 150 mm 7.2µm/m Using Electronic Level.

180 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >25 mm to 50 mm 0.10µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

181 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >50 mm to 75 mm 0.13µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

182 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >75 mm to 100 mm 0.15µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

183 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide 0.5 mm to 25 mm 0.09µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 35 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 37: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

184 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >25 mm to 50 mm 0.15µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

185 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >50 mm to 75 mm 0.18µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

186 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >75 mm to 100 mm 0.22µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

187 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel 0.5 mm to 25 mm 0.12µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

188 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Length MeasuringMachine.L.C.:0.0001mm

0 to 100 mm 0.66µm Using Slip Gauge SetGrade 'k' byComparison Method.

189 MECHANICAL-DUROMETER

Rubber HardnessTester

0 to 100 Shore A 0.5Shore A Rubber hardnessCalibration (MicrometerHead with Fixture) ByDepth IndentationMethod as per ISO18898

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 36 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 38: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

190 MECHANICAL-DUROMETER

Rubber HardnessTester

0 to 100 Shore D 0.5Shore D Rubber hardnessCalibration (MicrometerHead with Fixture) ByDepth IndentationMethod as per ISO18898

191 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 1000 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

192 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 250 bar 0.05bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

193 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 700 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

194 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 37 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 39: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

195 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices.

0 mbar to -200 mbar 0.25mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

196 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices,PressureTransmitter/Switches.

0 bar to -0.95 bar 1.3mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

197 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressureGauge, PressureIndicating Devices &PressureTransmitter/Switches

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

198 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressurePressure IndicatingDevice & PressureTransmitter/Switches

0 bar to 2 bar 1mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 38 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 40: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

199 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial/Digital LowPressure IndicatingDevices & PressureTransmitter.

0 mbar to 200 mbar 0.25mbar Using DigitalManometer withVacuum / PneumaticHand Pressure Pumpby Comparison Methodas per DKD- R - 6 -1

200 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Pressure IndicatingDevices(MaghnelicGauges, Low PressureGauges, Calibrators)

0 Pa to 500 Pa 0.35 Pa Using Digital PressureCalibrator byComparison MethodDKD- R - 6 -1

201 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

0.1 Nm to 10 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

202 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

10 Nm to 200 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

203 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

200 Nm to 1000 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 39 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 41: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

204 MECHANICAL-VOLUME

Glass Pipettes &Burette

>10 ml to 100 ml 5µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

205 MECHANICAL-VOLUME

Glass Pipettes &Burette

0.1 ml to 10 ml 2.4µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

206 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>10 ml to 100 ml 19.5µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

207 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>100 ml to 500 ml 21.3µl Using Weighingbalance of d:1 mg &Distilled water byGravimetric methodbased on ISO 4787

208 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>2 L to 5 L 316.4µl Using Weighingbalance of d:10 mg &Distilled water byGravimetric methodbased on ISO 4787

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 40 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 42: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

209 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>500 ml to 2 L 19.5µl to 249ul Using Weighingbalance of d:10 mg &Distilled water byGravimetric methodbased on ISO 4787

210 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

0.1 ml to 10 ml 2.9µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

211 MECHANICAL-VOLUME

Micro pipettes (PistonOperated-Apparatus)

1 ml to 5 ml 4µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

212 MECHANICAL-VOLUME

Micro pipettes(PistonOperated-Apparatus)

100 µl to 1000 µl 0.35µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

213 MECHANICAL-VOLUME

Micro pipettes(PistonOperated-Apparatus)

10 µl to 100 µl 0.30µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 41 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 43: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

214 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 g)

0 kg to 5 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

215 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 mg)

0 g to 100 g 0.03mg Using Weights ofaccuracy class E2 asper OIML R-76

216 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 g)

0 kg to 30 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

217 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 mg)

0 g to 200 g 0.10mg Using Weights ofaccuracy class E2 asper OIML R-76

218 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =1 mg)

0 kg to 1 kg 1.0mg Using Weights ofaccuracy class E2 asper OIML R-76

219 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

2 kg 8.6mg Using E2 class weightsand balance of d: 0.01g as per OIML R-111.Based on ABBAMethod

220 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

5 kg 9mg Using E2 class weightsand balance of d: 0.01g as per OIML R-111.Based on ABBAMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 42 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 44: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

221 MECHANICAL-WEIGHTS

Weights (ConventionalMass)M1 Class &Coarser

10 kg 80mg Using E2 & F1 classweights and balance ofd: 0.1 g as per OIML R-111. Based on ABBAMethod

222 MECHANICAL-WEIGHTS

Weights (ConventionalMass)M1 Class &Coarser

20 kg 100mg Using E2 & F1 classweights and balance ofd: 0.1 g as per OIML R-111. Based on ABBAMethod

223 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

224 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 kg 1.2mg Using E2 class weightsand balance of d: 0.001g as per OIML R-111.Based on ABBAMethod

225 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

10 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 43 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 45: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

226 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

10 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

227 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

100 g 0.08mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

228 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

100 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

229 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

2 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 44 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 46: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

230 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

2 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

231 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

20 g 0.05mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

232 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

20 mg 0.01mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

233 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

200 g 0.15mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 45 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 47: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

234 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

5 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

235 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

5 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

236 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

50 g 0.05mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

237 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

50 mg 0.01mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

238 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

500 g 1.0mg Using E2 class weightsand balance of d: 0.001g as per OIML R-111.Based on ABBAMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 46 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 48: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

239 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

500 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

240 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

200 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

241 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mgas per OIML R-111.Based on ABBAMethod.

242 THERMAL-MISCELLANEOUS

Temperature Gauge 30°C to 250°C0.30°C Using PRT with

6.5DMM orMUltifunction Calibratorand Liquid bath bycomparison method

243 THERMAL-TEMPERATURE

Glass Thermometers -80 °C to 250 °C 0.24°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath by ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 47 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 49: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

244 THERMAL-TEMPERATURE

Infrared/Non-contactType Thermometer

>300 °C to 650 °C 6.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorAnd Black body sourceby Comparison Method

245 THERMAL-TEMPERATURE

Infrared/Non-contactType Thermometer

50 °C to 300 °C 5.2°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorAnd Black body sourceby Comparison Method

246 THERMAL-TEMPERATURE

Temperature Gauge 0 °C to 600 °C 0.24°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath by ComparisonMethod

247 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>1000 °C to 1200 °C 2.44°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

248 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>400 °C to 650 °C 0.16°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 48 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 50: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

249 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

250 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>250 °C to 400 °C 0.21°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

251 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>50 °C to 250 °C 0.19°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

252 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

-80 °C to 50 °C 0.22°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 49 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 51: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

253 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>1000 °C to 1200 °C 2.43°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths byComparison Method

254 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>250 °C to 400 °C 0.21°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

255 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>400 °C to 650 °C 0.21°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

256 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>50 °C to 250 °C 0.19°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 50 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 52: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

257 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths byComparison Method

258 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

-80 °C to 50 °C 0.22°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 51 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 53: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

Site Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 1 kHz 100 µA to 200 µA 0.26% to 0.39% Using 6 ½ DigitMultimeter 8846A byDirect Method

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz 10 A to 3200 A 0.4% Using Standard CurrentTransformer with 6 ½Digit Multimeter 8846Aby Direct Method

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

0.2 mA to 100 mA 0.39% to 0.18% Using 6 ½ DigitMultimeter 8846A byDirect Method

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

100 mA to 10 A 0.18% to 0.29% Using 6 ½ DigitMultimeter 8846A byDirect Method

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 1 kHz to 5kHz

3 mA to 3 A 0.16% to 3.54% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 52 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 54: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz 20 A to 1000 A 0.79% Using MPF 5502 A withCurrent Coil by DirectMethod

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

10 A to 20 A 0.17% to 0.36% Using Multi ProductCalibrator 5502A byDirect Method

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

3 A to 10 A 0.16% to 0.17% Using Multi ProductCalibrator 5502A byDirect Method

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1KHz

30 uA to 3 mA 1.44% to 0.16% Using Multi ProductCalibrator 5502A byDirect Method

10 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10 uA to 100 mA 0.36% to 0.064% Using 6 ½ DigitMultimeter 8846A byDirect Method

11 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100 mA to 3 A 0.064% to 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 53 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 55: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

12 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 3 A to 10 A 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

13 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 µA to 100 µA 0.26% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

14 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 100 uA to 3 mA 0.044% to 0.031% Using Multi ProductCalibrator 5502A byDirect Method

15 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 20 A to 1000 A 0.64% Using MPF 5502 AWith Current Coil byDirect Method

16 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 A to 20 A 0.11% to 0.38% Using Multi ProductCalibrator 5502A byDirect Method

17 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 mA to 3 A 0.031% to 0.11% Using Multi ProductCalibrator 5502A byDirect Method

18 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(PhaseError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 3.34min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

19 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(RatioError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 0.047% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 54 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 56: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

20 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC HIGH VOLTAGE 1 kV to 40 kV 3.2% Using HV Divider withKV Meter

21 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 1 kV to 5 kV 3.2% to 3.8% Using Udeyraj makeHV Divider withindicator by DirectMethod

22 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 5 kV to 100 kV 3.8% to 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

23 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 1 kHz 1 mV to 20 mV 4.72% to 0.33% Using 6 ½ DigitMultimeter 8846A byDirect Method

24 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 50 Hz to 1kHz

20 mV to 1000 V 0.54% to 0.10% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 55 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 57: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

25 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

26 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A/5ASecondary from 1% to120% of rated current

2000 A to 7500 A 3.4min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

27 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

28 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

29 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 0.05% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 56 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 58: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

30 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A/5ASecondary from 1% to120% of Rated Current

2000 A to 7500 A 0.085% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

31 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

32 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

33 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 1 kV to 5 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

34 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 5 kV to 40 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 57 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 59: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

35 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 mOhm to 1 Ohm 0.42% to 0.36% Using 6 ½ DigitMultimeter 8846A byV/I Method

36 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 MOhm to 1000MOhm

0.013% to 2.31% Using 6 ½ DigitMultimeter 8846A byDirect Method

37 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 Ohm to 1 MOhm 0.36% to 0.013% Using 6 ½ DigitMultimeter 8846A byDirect Method andVoltage/ CurrentMethod

38 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 1 mV to 100 mV 0.42% to 0.085% Using 6 ½ DigitMultimeter 8846A byDirect Method

39 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 100 mV to 1000 V 0.085% to 0.007% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 58 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 60: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

40 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/110 4.03min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

41 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/63.5 3.16min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

42 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/110 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

43 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/63.5 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

44 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Voltage Ratio 0.8 to 2021 ratio 0.16% Using 6 ½ DigitMultimeter 8846A andFluke 5502A by DirectMethod

45 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Power 50 [email protected] UPF 120/2240 V 0.1A to 20 A

1.2 W to 4800 W 3.47% to 0.23% Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 59 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 61: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

46 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Vertical deflectionHorizontal Deflection 1kHz

5m V to 50 V 2.64% to 0.40% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

47 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz 2 mV to 10 mV 1.76% to 0.41% Using Multi ProductCalibrator 5502A byDirect Method

48 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz to 10kHz

30 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

49 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

10 mV to 300 mV 0.41% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

50 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 mV to 300 V 0.044% to 0.1% Using Multi ProductCalibrator 5502A byDirect Method

51 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 60 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 62: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

52 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @ 1 kHz 10 pF to 1 nF 2.98% to 1.2% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

53 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1 kHz 9 nF to 1 mF 1.26% to 0.80% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

54 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1kHz 1 nF to 9 nF 1.7% to 1.26% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

55 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Power @ 1 V to600 V, 1A to 20 A

1 W to 12 KW 0.37% to 0.25% Using MultiproductCalibrator 5502A withPower Option by DirectMethod.

56 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 mohm to 1 mohm 0.6% to 0.6% Using Shunt By DirectMethod

57 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 Ohm to 100 Ohm 0.25% to 0.072% Using Multi ProductCalibrator 5502A byDirect Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 61 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 63: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

58 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 kOhm to 100MOhm

0.021% to 0.58% Using Multi ProductCalibrator 5502A byDirect Method

59 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 ohm to 100 kOhm 0.072% to 0.021% Using Multi ProductCalibrator 5502A byDirect Method

60 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 1 mV to 100 mV 0.36% to 0.013% Using Multi ProductCalibrator 5502A byDirect Method.

61 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 100 mV to 30 V 0.013% to 0.0064% Using Multi ProductCalibrator 5502A byDirect Method.

62 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 30 V to 1000 V 0.0064% to 0.007% Using Multi ProductCalibrator 5502A byDirect Method.

63 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

High Resistance 100 kohm to 1 Tohm 1.2% to 2.4% Using DecadeResistance Box byDirect Method

64 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Inductance @1 kHz 1 mH to 10 H 1.53% to 2.03% Using DecadeInductance Box byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 62 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 64: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

65 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics(Current) upto 41st

100 mA to 6 A 0.75% Using FlukeMultiProduct Calibrator5502A

66 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics upto41st Voltage

100 mV to 60 V 0.75% Using FlukeMultiproduct Calibrator5502A

67 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Oscilloscope DCFunction

2 mV to 33 V 6.59% to 0.33% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

68 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Power Factor 50 Hz 0.1 to UPF 0.05PF Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

69 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage ratio 0.8 to 400 0.5% to 0.3% Using TransformerRatio Standard byComparison Method

70 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage Ratio 400 to 2021 0.5% to 0.3% Using TransformerRatio Standard byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 63 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 65: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

71 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-)200 °C to 800 °C 0.51°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

72 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)C-Type

0 °C to 2315 °C 1.4°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

73 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)E-Type

(-)270 °C to 1300 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

74 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)J-Type

(-)200 °C to 1200 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 64 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 66: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

75 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)K-Type

(-)250 °C to 1137 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

76 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)L-Type

(-)200 °C to 600 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

77 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)N-Type

(-)270 °C to 1300 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

78 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)R-Type

(-)50 °C to 1767 °C 0.9°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 65 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 67: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

79 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)S-Type

(-)50 °C to 1767 °C 0.8°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

80 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)T-Type

(-)270 °C to 400 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

81 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)U-Type

(-)200 °C to 600 °C 0.5°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

82 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-) 200 °C to 800 °C 0.27°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 66 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 68: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

83 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)EType

(-)200 °C to 1000 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

84 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)JType

(-)210 °C to 1200 °C 0.37°C Using Multi ProductCalibrator 5502A byDirect Method

85 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)KType

(-)200 °C to 1372 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

86 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)NType

(-)200 °C to 1300 °C 0.61°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 67 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 69: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

87 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)RType

0 °C to 1700 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

88 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)SType

0 °C to 1700 °C 0.58°C Using Multi ProductCalibrator 5502A byDirect Method

89 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)TType

(-)200 °C to 400 °C 0.70°C Using Multi ProductCalibrator 5502A byDirect Method

90 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)UType

(-)200 °C to 600 °C 0.65°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 68 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 70: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

91 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)BType

600 °C to 1820 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

92 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)CType

200 °C to 2315 °C 0.95°C Using Multi ProductCalibrator 5502A byDirect Method

93 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 MHz 0.084% to 0.014% Using 6 ½ DigitMultimeter 8846A byDirect Method

94 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 0.1 sec to 1 hr 0.013sec to 2.32sec Using Digital TimeInterval Meter By DirectMethod

95 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 hr to 24 hr 2.32sec to 5.18sec Using Digital TimeInterval Meter Bycomparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 69 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 71: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

96 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Bandwidth 50 MHz to 250 MHz 6.21% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

97 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 10 Hz to 300 kHz 0.086% to 0.0094% Using Multi ProductCalibrator 5502A byDirect Method

98 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 300 kHz to 2 MHz 0.0094% to 0.49% Using Multi ProductCalibrator 5502A byDirect Method

99 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Time Base 2 n Sec to 5 Sec 0.3% to 0.67% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

100 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

10 RPM to 5000 RPM 2.1RPM Using Tachometer.

101 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

5000 RPM to 30000RPM

19RPM Using Tachometer.

102 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

1000 mm to 2000 mm 22.5µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 70 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 72: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

103 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial CalibrationTesterL.C.: 0.1 µm

0 to 25 mm 1.12µm Using Electronic Probe

104 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Electronic HeightGaugeL.C.: 0.5 µm

0 to 600 mm 8µm Using Gauge Block;Long Gauge & SurfacePlate.

105 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 1000 mm 20µm Using Caliper Checker;Gauge Block & SurfacePlate.

106 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 600 mm 11µm Using Caliper Checker;Gauge Block & SurfacePlate.

107 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

MetallurgicalMicroscope

100 x to 1000 x 0.3% Using Micro Scale ofL.C. 10 µm.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 71 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 73: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

108 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Straight Edge Upto 2000 mm 1.8 x v(L/125) µm L inmm

Using Electronic LevelL.C.1µm/m & GaugeBlock.

109 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Surface Plate 4000 mm X3000 mm 1.2 xv(L+W)/125 L&Win mm

Using Electronic Level,L.C.1 µm/m

110 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Length MeasuringMachine.L.C.:0.0001mm

0 to 100 mm 0.66µm Using Slip Gauge SetGrade 'k' byComparison Method.

111 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineAngleL.C.:1sec

0 ° to 360 ° 6.0arc sec. Using AngularGraticule / AngleGauge Set.

112 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineLinear : L.C.:0.1 µm

0 to 200 mm 4.8µm Using Glass Scale./Gauge Block.

113 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineMagnification

Up to 100 X 1.2% Using Glass Scale &Digimatic Caliper.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 72 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 74: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

114 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 1000 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

115 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 250 bar 0.05bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

116 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 700 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

117 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

118 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices.

0 mbar to -200 mbar 0.25mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 73 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 75: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

119 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices,PressureTransmitter/Switches.

0 bar to -0.95 bar 1.3mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

120 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressureGauge, PressureIndicating Devices &PressureTransmitter/Switches

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

121 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressurePressure IndicatingDevice & PressureTransmitter/Switches

0 bar to 2 bar 1mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

122 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial/Digital LowPressure IndicatingDevices & PressureTransmitter.

0 mbar to 200 mbar 0.25mbar Using DigitalManometer withVacuum / PneumaticHand Pressure Pumpby Comparison Methodas per DKD- R - 6 -1

123 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Pressure IndicatingDevices(MaghnelicGauges, Low PressureGauges, Calibrators)

0 Pa to 500 Pa 0.35 Pa Using Digital PressureCalibrator byComparison MethodDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 74 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 76: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

124 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 g)

0 kg to 5 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

125 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 mg)

0 g to 100 g 0.03mg Using Weights ofaccuracy class E2 asper OIML R-76

126 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 g)

0 kg to 30 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

127 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 mg)

0 g to 200 g 0.10mg Using Weights ofaccuracy class E2 asper OIML R-76

128 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =1 mg)

0 kg to 1 kg 1.0mg Using Weights ofaccuracy class E2 asper OIML R-76

129 THERMAL-MISCELLANEOUS

Temperature Gauge 30°C to 250°C0.30°C Using PRT with

6.5DMM orMUltifunction Calibratorand Liquid bath bycomparison method

130 THERMAL- SPECIFICHEAT & HUMIDITY

Humidity chamber/climatic chamber/Environmentalchamber

10 °C to 60 °C 0.31°C Using Temperature &humidity Data loggerswireless by Multi PointCalibration

131 THERMAL- SPECIFICHEAT & HUMIDITY

Humidity chamber/climatic chamber/Environmentalchamber @25°C

15% RH to 95% RH 2.44%RHRH Using Temperature &humidity Data loggerswireless by Multi PointCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 75 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 77: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

132 THERMAL-TEMPERATURE

Deep Freezer,Refrigerators,Incubator, Oven,Temperature Bath,Chamber, Furnace

-80 °C to 300 °C 2.37°C UsingRTD/Thermocouple(K/J/N/ other type)sensor with MultiChannel Data Loggermulti Point Calibration

133 THERMAL-TEMPERATURE

Glass Thermometers 30 °C to 250 °C 0.25°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath byComparison Method

134 THERMAL-TEMPERATURE

Infrared /Non-contactType Thermometer

>300 °C to 650 °C 6.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorand Black body sourceBy Comparison Method

135 THERMAL-TEMPERATURE

Infrared /Non-contactType Thermometer

50 °C to 300 °C 5.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorand Black body sourceBy Comparison Method

136 THERMAL-TEMPERATURE

Temperature Bath,Chamber & Furnace.

300 °C to 1200 °C 2.95°C UsingRTD/Thermocouple(K/J/N/ other type)sensor with MultiChannel Data Loggermulti Point Calibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 76 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 78: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

137 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>400 °C to 650 °C 1.70°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

138 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

139 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Furnace.

>1000 °C to 1200 °C 2.44°C°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

140 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>50 °C to 250 °C 0.30°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 77 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 79: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

141 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath /Incubators.

-80 °C to 50 °C 0.25°C°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

142 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Oven /Furnace

>250 °C to 400 °C 0.65°C°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

143 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>1000 °C to 1200 °C 2.44°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

144 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>250 °C to 400 °C 0.65°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 78 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 80: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

145 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>400 °C to 650 °C 1.70°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

146 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>50 °C to 250 °C 0.30°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

147 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

148 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

-35 °C to 50 °C 0.25°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 79 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -