nice NeÖtech Medical Systems P
nice NeÖtech Medical Systems Pvt. Ltd
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice BUBBLE CPAP
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology , we always strive
for growth and delighted customers
Timeline-nice Neotech
2005•First New Factory Unit 1 Opened
•Increase in production Line
2001•Registered as Pvt ltd
•Production started for different models
2013•Second New Factory Unit 2 Opened
1999• Neotech Logo
Designed
• Designed first incubator & open
care prototype
2003• Export to
Vietnam
2009-2011
• CE certification for products
2015• 125+skilled
employee strong firm
• Multiple range of NICU products
1997•Started Neotech Medical Systems with 2 employees
•First Infant Radiant Warmer
What is a Bubble CPAP?
The Bubble CPAP System helps in breathing of infants requiring respiratory support. It is incorporated with unique feature of humidification technology and Air/Oxygen Blender.
CPAP works by maintaining positive pressure in the airway during spontaneous breathing, thereby increasing functional residual capacity and improving oxygenation in infants with RDS. CPAP does this by stabilizing airspaces that have a tendency to collapse during expiration due to surfactant deficiency.
Why Bubble CPAP is Used?
Non invasive ventilation
Delivering continuous positive airway pressure
Deliver blended and humidified Oxygen
Bubble CPAP
Introducing nice Bubble CPAP
Designed to perfection
Different Models of 5060 series Bubble CPAP
Features
Consistent and accurate delivery of CPAPServo controlled Humidifier attachedEase of useOptimum HumidificationBubble generator
Operation
Gas Supply Flow rate. Setting the FIO2 Percentage. Setting the Humidifier. Setting the Positive End Expiratory Pressure (PEEP) at the PAP Valve. Fixing the Patient interface. Check during operation
Gas Supply Flow rate
Set the Flow rate at 5 LPM or required in the Flow Meter.
Setting the FIO2 Percentage
Set the required FIO2 percentage on the dial knob (21% to 100%).
Warning: The oxygen Concentration must be monitored with a calibrated oxygen measuring unit.
Setting the Humidifier.
Select the desired mode by using mode key.
Select heater wire ON/OFF by using Heater wire Key.
Setting the Positive End Expiratory Pressure (PEEP) at the PAP Valve.
Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level in the PAP Valve. To fill the water reservoir Remove the lid from the jar. Fill the jar to the mean fill line with sterile water.
Note: To assure consistency, water level must be between minimum and maximum lines without the drop tube beneath the surface of the water.
Note: The PAP Valve reference marks indicate PAP level at the mean water line at 6 LPM. Replace the lid on to the jar
To attach the PAP Valve to a mounting clamp
Slide the equipment mount into the corresponding equipment mount holder. Use of this type of mounting system will assure that the device remains in the upright (vertical) position.
Exhalation Port ConnectionSet the device to the minimum position by rotating the PAP Adjustment lever to the full anticlockwise position Adjust the gas flow to the desired level and connect the expiratory end of the breathing circuit to the gas inlet connector
PAP Setting
Using clockwise rotation, adjust the device to the desired back pressure [Positive Airway Pressure (PAP)]
Monitoring with Manometer
During use, continuously monitor airway pressure with an accurate pressure monitoring device and adjust the device accordingly
Water Level Adjustment Port
A water level adjustment port is provided to adjust water level without interruption of therapy. To use, remove the luer cap and connect a syringe to the port, adding or removing water as required
Fixing the Patient Interface
Check during Operation
Regularly observe that the water is feeding into the humidification chamber.Should the water level exceed the maximum level marked on the humidification chamber, the chamber should be replaced.Check all connections are tight before use and after adjustment.Ensure air flow is presence at all times. If air flow is interrupted, turn off the humidifier.Regularly observe the circuit for condensate, drain as requiredRegularly observed the PAP Valve for bubbling. If bubbling is not observed check for a minimize air leaks in the system and at the patient end. If air leaks have been minimized, air flow may be increased to achieve continues bubbling.Regularly observe the water level in the PAP valve refill the PAP valve if the water level drops below minimum water level line.Monitor the Patient oxygen level.Always use pressure monitoring to verify that the patient is receiving the prescribed CPAP level.
TROUBLESHOOTING
AND MAINTENANCE
Cleaning & Maintenance
General Clean external surfaces of the Bubble CPAP and Gas Supply line using a damp cloth and mild soapywater or Isopropyl Alcohol. Dry all surfaces after cleaning with a clean soft cloth or paper towel. The Bubble CPAP should require minimal servicing or maintenance when used under normal conditions. At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity. If any part of the equipment is damaged or fails to operate correctly, take it out of service immediately and contact a qualified Service Engineer to ensure operational safety. A Service Manual enabling qualified technical personnel to carry out routine service schedules, fault finding and repairs is available from nice Neotech. Always carry out a functional test after cleaning or maintenance to ensure safety and operationalintegrity
Cleaning
The reusable humidifier chamber can be autoclavable.
All other components may be disinfected using an approved chemical disinfectant solution.
When all components are completely dry, assemble the system again.
How nice Bubble CPAP fare over competitors?
nice Bubble CPAP
Low cost
Inbuilt Manometer
Reusable gas pressure regulator
PEEP set manually by rotating knob
Trouble Shooting
Certifications and Approvals
Electrical & Product Safety : IEC 60601-1
EMC Safety : IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85, Krishna Industrial Estate
Mettukuppam, Vannagaram,
Chennai – 600095, Tamil Nadu,
INDIA
Tel: +91-44-24762594,24764608.
Fax: +91-44-24766920
Web: www.niceneotech.com.
E-mail: [email protected] , [email protected]
Overseas Enquiry: [email protected]
Toll Free : +91-1800-425-2594.