Top Banner
BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON Sat 31 st Aug 2013 Session 2 / Talk 1 10:34 – 10:55 ABSTRACT In this study we assessed a Works-in-Progress sequence developed by Siemens for evaluation of the peripheral arterial tree without contrast agent - Quiescent Interval Single Shot MR Angiography (QISS) sequence. Previously we have evaluated other non-contrast enhanced angiographic techniques (Native Space and Native TrueFisp) which worked well in the majority of regions but still presented some limitations. It was hoped that the new technique using the inflow of non-saturated blood spins & acquiring single-shot 2D TrueFISP readout images might improve detection and accuracy of lesions, particularly in the abdominal and pelvic region. A total of 50 volunteers were recruited between March 1 st 2011 and December 12 th 2011. All studies were acquired on a 1.5T Magnetom Avanto scanner (Siemens, Erlangen, Germany). The first two participants were healthy volunteers, followed by 48 clinical patients. The images were independently reviewed by 3 Vascular Radiologists blinded to patient details. Diseased arterial segments were assessed as normal, stenosed or occluded. Stenosis severity was assessed as mild (< 50% diameter loss), moderate (50-70% diameter loss) and severe (> 70% diameter loss). Stenosis and occlusion length were also assessed. These findings were then compared to the “gold standard” CE-MRA to allow assessment of accuracy of the QISS sequence in assessing arterial disease severity and extent. The QISS sequence produced images that corresponded very well with the contrast enhanced images of the same anatomical regions. Stenoses and occlusions were accurately assessed in the majority of cases. We present our findings and show some examples of the strengths and weaknesses of this Works-in-Progress technique.
30

BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Feb 24, 2016

Download

Documents

ling

Sat 31 st Aug 2013 Session 2 / Talk 1 10:34 – 10:55. BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON. ABSTRACT - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

BROOKLYN 3MRI USER GROUP

Anna Marie LYNDON

Sat 31st Aug 2013Session 2 / Talk 1

10:34 – 10:55

ABSTRACTIn this study we assessed a Works-in-Progress sequence developed by Siemens for evaluation of the peripheral arterial tree without contrast agent - Quiescent Interval Single Shot MR Angiography (QISS) sequence.Previously we have evaluated other non-contrast enhanced angiographic techniques (Native Space and Native TrueFisp) which worked well in the majority of regions but still presented some limitations. It was hoped that the new technique using the inflow of non-saturated blood spins & acquiring single-shot 2D TrueFISP readout images might improve detection and accuracy of lesions, particularly in the abdominal and pelvic region.A total of 50 volunteers were recruited between March 1st 2011 and December 12th 2011. All studies were acquired on a 1.5T Magnetom Avanto scanner (Siemens, Erlangen, Germany). The first two participants were healthy volunteers, followed by 48 clinical patients.The images were independently reviewed by 3 Vascular Radiologists blinded to patient details. Diseased arterial segments were assessed as normal, stenosed or occluded. Stenosis severity was assessed as mild (< 50% diameter loss), moderate (50-70% diameter loss) and severe (> 70% diameter loss). Stenosis and occlusion length were also assessed.These findings were then compared to the “gold standard” CE-MRA to allow assessment of accuracy of the QISS sequence in assessing arterial disease severity and extent.The QISS sequence produced images that corresponded very well with the contrast enhanced images of the same anatomical regions. Stenoses and occlusions were accurately assessed in the majority of cases.We present our findings and show some examples of the strengths and weaknesses of this Works-in-Progress technique.

Page 2: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Evaluation of QISS

Non-CE MRA technique

Anna-Maria Lydon, Andrew Holden Centre for Advanced MRI

University of Auckland, New Zealand

Page 3: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Background Other non CE-MRA techniques QISS – how it works Recruitment Challenges & limitations Results Case examples Conclusion

SIEMENS WIP sequence - QISS

Page 4: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Conventional CE-MRA sequences

Compares favourably with invasive catheter angiography

Still requires contrast media

Page 5: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

SIEMENS non-CE MRA sequences

Native SPACEA technique that relies on the inherent difference in signal between fast flowing blood in systole and the slower flowing blood in the diastolic phase of the cardiac cycle

It is a high spatial resolution 3D TSE sequence with contrast optimized variable flip-angle trains and inherent flow sensitivity

Page 6: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Cine Scout Mean curve analysis

Page 7: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Femoro-popliteal Station

CE-MRA

NATIVE SPACE

DSA

Page 8: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Tibial Station

CE-MRA

NATIVE SPACE

Page 9: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Aorto-iliac Station All 4 cases assessed as poor quality Poor vessel signal intensity obtained from

the aorta and iliac vessels but femoral vessels well seen (S:N issues?)

CE-MRA

NATIVESPACE

Page 10: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

SIEMENS WIP sequence - QISS Quiescent-Interval Single Shot Magnetic

Resonance Angiography (QISS) Rapid, sequential two-dimensional (2D) steady-state free

precession acquisition acquired using ECG-gating Acquires one slice per heartbeat

Figure 1. Pulse sequence diagram of the QISS sequence (Edelman et al, MRM 2009).

Page 11: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

QISS – CAMRI Experience 50 volunteers studied from March 1st to

December 12th 2011

Of these - 3 normal volunteers- 47 clinical patients

1.5T Magnetom Avanto scanner (Siemens, Erlangen, Germany) using a dedicated peripheral vascular coil and body matrix coils as required

Image quality assessed and graded from non-diagnostic to excellent

Page 12: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

QISS – CAMRI Experience

Page 13: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

3 volunteers 2 Female – all stations

1 Male – foot and ankle

47 clinical patients 21 Female26 Male

Stations – Tibial (Std) - 16 Hi-res trifurcation - 6

Fem-pop - 17Aorto-iliac - 14

Heart rate - Range 50-120bpm

QISS – Patient distribution

Page 14: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Claudication – 31 Ulcers – 3 Other - ?PAD (specific level)

- pulses (weak/absent)- 1 x Type B aortic dissection- 2 x aneurysms (AAA and iliac)- 1 x toe numbness- Previous grafts/ PTA- 1 x amputee with weak pulses

remaining limb- 5 x TKJR, 1 x THJR

QISS – Clinical Indications

Page 15: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Venous Contamination - Our first clinical patient had the QISS sequence acquired in the abdominal (aorto-iliac) region post contrast.

QISS – Limitations & challenges

QISS #03 data set MIP

Page 16: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

In-Plane Signal Loss - be seen when the vessel orientation runs in-plane with the slice orientation. CE-MRA MIP QISS MIP QISS angled slab

showing signal loss 

QISS – Limitations & challenges

Page 17: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Abdominal (aortio-iliac) region – - The initial healthy volunteers struggled with the breath-holds- By adapting the abdominal (aorto-iliac) regions to a single concatenation with 2-3 averages this sequence could be acquired with free breathing

QISS – Limitations & challenges

MIP QISS data set Ce-MRA MIP

Page 18: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Patient movement

MIP’s often showed small steps between the stations. This was due to slight patient movement between slabs

QISS – Limitations & challenges

Page 19: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

QISS MIP Ce-MRA MIP

Metal Artifact – 5 of the clinical volunteers had a total knee joint replacements and 1 had a total hip joint replacement.

QISS – Limitations & challenges

Page 20: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Fast AF - QISS MIP

Arrhythmia & poor ECG - 6 patients presented with arrhythmia, eg, atrial fibrillation and bigeminy. In addition, 2 patients had poor ECG traces1x patient had tachycardia (HR 115bpm)

QISS – Limitations & challenges

Fast AF - CE-MRA MIP

Slow AF - QISS MIP

Slow AF CE-MRA MIP

Page 21: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Images independently reviewed by 3 vascular radiologists Imaged are segments graded for image quality:

Grade 1: non-diagnostic Grade 2: poor quality Grade 3: diagnostic Grade 4: excellent quality

Stenoses were colour coded as according to assessed severity

QISS – Image Assessment

Page 22: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Grade 1: non-diagnostic Grade 2: poor quality Grade 3: diagnostic Grade 4: excellent quality

QISS – Image Quality

Grade NumberN=14

1 2 2 2 3 4 4 6

Grade NumberN=17

1 1 2 3 3 6 4 7

Grade NumberN=16

1 0 2 2 3 4 4 10

Aorto-iliac Femoro-popliteal Tibial

Page 23: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

QISS – Stenosis Severity

Correlation with CE-MRA Poor Poor Excellent Excellent

Image quality Number Non-diagnostic 3 Poor 7 Diagnostic 14 Excellent 23

QISS sequences compared to “gold standard” CE-MRA by 1 reviewer

QISS images of excellent and diagnostic quality compared well with CE-MRA, independent of site

Page 24: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Case #38 Aorto-iliac station (stations 6-8)73yo MaleClaudication both calvesFast AF – HR 115bpm QISS MIP CE-MRA

QISS – Case examples

Page 25: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

QISS – Case examples

QISS MIP CE-MRA MIP

Case #09 Femoro-popliteal stationsMale? SFA occlusionHR 62bpm

Page 26: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

56yo MaleSevere left leg claudicationAF 90 -120 bpmHi-res trifurcation tibial station

QISS – Case examplesCase #46

QISS MIP CE-MRA

Page 27: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Case #47Male volunteerResearcher High-res foot & ankle station

QISS – Case examples

Page 28: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

In our experience the QISS sequence has been a robust and relatively easy sequence to run We have found it quick and easy to use Potentially is of great use in cases where patients are unable to have Gadolinium contrast agent. However – there are pitfalls to be aware of.

Conclusion

Page 29: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

References Bi, X & Glielmi, C (2010), ‘Non contrast-enhanced,

Quiescent Interval Single Shot (QISS) MR Angiography of the Peripheral Arteries’, Siemens Applications Guide (Works-in-Progress # 592).

Eldelman RR, Sheehan JJ, Dunkle E, Schindler N, Carr J, Koktzoglou I (2010), ‘Quiescent-Interval Single-Shot Unenhanced Magnetic Resonance Angiography of Peripheral Vascular Disease: Technical Considerations and Clinical Feasibility’, Magnetic Resonance in Medicine, vol. 63, pp. 951-8.

Page 30: BROOKLYN 3 MRI USER GROUP Anna Marie LYNDON

Acknowledgements Dr Andrew Holden, Dr Brett Cowan,

Kate Handley, Hilary McIntyre, Rachel Heron and all the team at CAMRI

Dr Peter Schmitt, Dr Andreas Greiser & the CV development team at Siemens, Erlangen

All our patients who volunteered