BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 2
Dec 19, 2015
BME 560Medical Imaging: X-ray, CT, and
Nuclear Methods
X-ray Instrumentation Part 2
Today
• Anti-scatter devices
• X-ray screen-film systems
• Other methods of X-ray detection
X-ray System
Source
Restrictor (Collimator)
FilterSubject
Anti-scatter Detector
Produces X-rays from electrical energy
Determines size and shape of beam
Tailors X-ray spectrum
Selectively removes scattered photons
Converts X-rays to light and records
Total linear attenuation coefficient
Nx N0e x
Rayleigh photoelectric compton pair productin
Soft Tissue
At X-ray energies, most photons that interact in the patient are Compton-scattered.
X-ray Scatter
Object
Grid
Detector
Tube Scattered radiation comes into detector from all directions.
Result is a relatively uniform background “fog” that reduces dynamic range of the detector available to image true signal.
Would like some way to reduce scattered radiation without blocking much direct radiation.
Anti-scatter Strategies
• Collimation of the beam at the front end
• Air gaps
• Grids
• Scanning Slits
Air Gap
Moving the patient away from the detector reduces the scatter reaching the detector.
Square-law
Solid angle
What price do we pay for this?
Anti-scatter Grids
Construct a device to collimate photons after they leave the patient.
Thin lead strips must be precisely aligned.
Performance depends on the grid ratio
What is the price paid for a high grid ratio?
Typical grid ratios: 5:1 to 16:1 (lower for mammography)
Anti-scatter Grids
• A stationary grid will leave line artifacts in the image.
• A Potter-Bucky diaphragm is a movable grid that basically blurs the grid lines during exposure.
• The grid also blocks some primary radiation in the system.
Anti-scatter Grids
Tradeoff between scatter penetrating the grid and primary radiation detected
Anti-scatter Grids
• Thickness of strips determines the likelihood of penetration.– Less scatter penetration =
less primary radiation
• Low-angle scatter may still get through.
• Multiply-scattered photons may get through.
Anti-scatter Grids
• Additional exposure is needed to maintain same detector exposure level when using grid.
• Grid Conversion Factor
• Typically 3 < GCF < 8
• May avoid grid for small body parts or low energies.
mAs with grid for exposure E
mAs without grid for exposure EGCF
Scanning SlitsMoving source and collimator
Moving slit
Stationary detector
Move source, collimator, and slit together.
Only takes one part of image at a time.
Very high scatter reduction
Slow
X-ray Detectors
• Film-Screen
• Image Intensifiers
• Panel Detectors
Film-Screen Detectors
• Roentgen’s first X-rays exposed a photographic plate directly.– But photographic film has very low stopping
power (a couple of percent).– To expose the film to its full dynamic range
(contrast) would require high dose and most would be wasted.
• Augment this with an intensifying screen that converts X-ray photons to visible light.
Screen-film System
• Double emulsion film sandwiched between pair of intensifying screens
• Phosphor particles (high Z) covert X-ray into light photons
• Screen enhances contrast but lowers resolution
• Engineering tradeoff: Phosphor thickness
Film
CoatingPhosphor
Reflective Layer
Base
Screen-film System
• Reflective layer reflects light back into the film
• Base for mechanical support• Phosphor layer material choice:
– More fluorescent than phosphorescent
– High linear attenuation coefficient = stopping power
• Conversion efficiency: total light energy per unit incident X-ray energy (usually 5 – 20%)– Energy dependent
Film
CoatingPhosphor
Reflective Layer
Base
Film
• Very similar to photographic film; must be developed to fix the image
• Two components:– Base: Plastic sheet, dimensionally stable (size and
shape do not change under environmental and processing conditions)
– Emulsion: Crystals of silver bromide suspended in gelatin substance; on one side (single-emulsion), or both sides (double-emulsion) of base.
• Image is formed in the silver bromide crystals.
Screen-film System
• The screen-film combination usually has a speed quoted– More sensitive (= fewer X-
ray photons to result in a given image density) pairs have higher speed
– At a particular energy!
SpeedSensitivity
(mR)
1200 0.1
800 0.16
400 0.32
200 0.64
100 1.28
50 2.56
25 5.0
12 10.0
Exposure to get a standard level of film density
From Sprawls
Radiographic Cassette
• Ensures firm and uniform contact between intensifying screens and film sandwiched in between
• Optical mirrors located outside screens to direct light towards film, maximize light conversion efficiency
• Contains ID card and loaded only one way into X-ray machine
Image source: The Essential Physics of Medical Imaging
Film Density
• Density describes the overall blackness of the radiograph
Image source: http://www.nurseslearning.com/courses/fice/fde0030/Imaging_terms.htm
Screen-film System
Thicker screens result in higher sensitivity but increase image blur
Other Detection Schemes
• Detection is a result of radiation interaction with matter. Radiation interaction results in emission of by products, e.g. electrons, electromagnetic radiation, that can be sensed by instrumentation and recorded by data acquisition systems– Gas-Filled Detectors
– Scintillation Detectors
– Flat-panel detectors
– PSP plates
– Solid State Detectors
Gas Filled Detectors
• Radiation ionizes the gas. Charges freed by ionization produce a current.
Positive Power Supply
Radiation Ammeter Negative Resistance R Gas Chamber
Gas Filled Detectors• Radiation interacts with
gas and ionizes its atoms• Freed electrons interact
with gas and ionize more atoms - amplification
– Ionization chamber: No amplification
– Proportional counter: Amplification up to 106
times– Geiger-Muller counter:
Very strong avalanche
Ionization Proportional Geiger Mueller Chambers Counters Counters
Collected Charge
Voltage
Spatial sensitivity is lacking – Not used for imaging
Scintillation Detectors
• Interaction of X-rays with some materials (CsI, cadmium zinc telluride - CZT) produces ‘scintillation’ or “flash of light”.
scintillator
photomulitplierX-ray
Visible light Electrical pulse
The pulse can tell you about the energy of the incident photon.
Not capable of handling high photon flux.
Photomultiplier Tube
• Photomultiplier tube (PMT) converts light into electric current by photoelectric effect
Dynodes Photocathode grid Anode photons
Flat-panel Detectors
Varian Medical Systems
Scintillator
Light coupling
Light-sensitive digital detector (CCD array)
Photostimulable Phosphor Plates
• PSP plates
• X-rays excite electrons which are trapped in the material lattice.
• Later, the plate is scanned by a laser in a “plate reader” which frees the electrons locally and digitizes the image.
• The plate can be reused.
• Plugs in to the film-screen cassette slot.
Solid State Detectors
• They are compact semiconductors. Electrical conductivity of semiconductor is sensitive to impurities. The depletion layer is sensitive to radiation and electric current flow through, thus the measured current is a measure of radiation.
+ _ n-type depletion layer p-type - Radiation or incident particles
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
+ + + + + + + + + + + + + + + + + + + + + + + + + + + + + +
X-ray Image Detection
• Screen-film: Still in use• PSP Plates: Displacing screen-film in many
applications• Flat-panel: Increasing use but expensive• Solid state: Still in development for X-ray• Scintillation detectors: Not fast enough for X-ray
imaging, but still important research tools.– SPECT imaging
• Gas counter: Not useful for imaging but used for active measurement of patient exposure.