04.12.2014 1 PD Dr. Zöllner I Folie 155 I 9/9/2014 Blood Flow Measurement Techniques ! not only blood pressure but blood flow important information ! describes the supply of blood in organs and tissue ! abnormal changes in the blood flow or blood velocity gives rise to malformatuon of vessels ! blood flow Q: volume of blood per time [ml/min] ! perfusion: blood supply in capillary bed [ml/min/100g] PD Dr. Zöllner I Folie 156 I 9/9/2014 Typical values for blood flow [cm/s]: Aorta 100 – 250 Abdominal 100 Carotis commuis 80 – 150 Vena Cava 5 – 40 Blood Flow Values Body
13

Blood Flow Measurement Techniques Typical values for blood flow ...

Feb 11, 2017

Download

Documents

dokiet
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

1

PD Dr. Zöllner I Folie 155 I 9/9/2014

Blood Flow Measurement Techniques

!  not only blood pressure but blood flow important information

!  describes the supply of blood in organs and tissue

!  abnormal changes in the blood flow or blood velocity gives rise to malformatuon of vessels

!  blood flow Q: volume of blood per time [ml/min]

!  perfusion: blood supply in capillary bed [ml/min/100g]

PD Dr. Zöllner I Folie 156 I 9/9/2014

Typical values for blood flow [cm/s]:

Aorta 100 – 250 Abdominal 100 Carotis commuis 80 – 150 Vena Cava 5 – 40

Blood Flow Values Body

Page 2: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

2

PD Dr. Zöllner I Folie 157 I 9/9/2014

Blood Flow and Pressure Waves Abdominal Vasculature

Nic

ols

& O

‘Rou

rke,

McD

onal

d‘s

Blo

od F

low

in A

rterie

s

PD Dr. Zöllner I Folie 158 I 9/9/2014

Electromagnetic flow meter

!  based on Faraday‘s law of induction

!  blood is a conductor of electricity

!  can be ‚aligned‘ by external magnet field

Nicols & O‘Rourke, McDonald‘s Blood Flow in Arteries

Page 3: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

3

PD Dr. Zöllner I Folie 159 I 9/9/2014

Blood velocity measurement using US

PD Dr. Zöllner I Folie 160 I 9/9/2014

Typical Values

speed of sound:

velocity of blood:

transmitted US-frequency:

Doppler-frequency

" Audio Range

Page 4: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

4

PD Dr. Zöllner I Folie 161 I 9/9/2014

Continuous Wave (CW) - US

base wave

echo

product

Doppler

sender

mixer

bandpass display

sender

receiver

flow signal

fS

fE

IM

fS-fE

PD Dr. Zöllner I Folie 162 I 9/9/2014

Pulse Wave (PW) - US

Page 5: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

5

PD Dr. Zöllner I Folie 163 I 9/9/2014

US Doppler Probes - Invasive

Nicols & O‘Rourke, McDonald‘s Blood Flow in Arteries

PD Dr. Zöllner I Folie 164 I 9/9/2014

MR Blood Flow Measurements (1)

Lotz et al. , RadioGraphics 2002

#  spins moving along a magnetic field gradient acquire a difference in the phase (right)

#  amount of phase difference proportional to the velocity of the moving spins

Page 6: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

6

PD Dr. Zöllner I Folie 165 I 9/9/2014

MR Blood Flow Measurements (2)

Lotz et al. , RadioGraphics 2002

#  flow encoding by means of difference in the bipolar gradients

#  acquisition of two images #  phase difference by substraction

PD Dr. Zöllner I Folie 166 I 9/9/2014

MR Blood Flow Measurements (3)

Phase difference expressed in formulars:

γ gyromagnetic ratio Δm difference of the first moment of the

gradient time curve ν velocity

Page 7: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

7

PD Dr. Zöllner I Folie 167 I 9/9/2014

MR Blood Flow Measurements (4)

#  ΔM relates to the velocity encoding (VENC) #  gives sensitivity of measurable velocities #  relates to the gradient strength

PD Dr. Zöllner I Folie 168 I 9/9/2014

Diagram of MR PC sequence

Page 8: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

8

PD Dr. Zöllner I Folie 169 I 9/9/2014

PC-MR Images

Flow encoding needs an operator defined value:

VENC [cm/sec]

-VENC < v < VENC

#  receive 2 images

# magnitude image with velocity

#  phase image with velocity and direction

PD Dr. Zöllner I Folie 170 I 9/9/2014

#  flow encoding is directional

#  depends on the gradient axis applied

#  therefore: 3D flow measurements possible in each voxel

3D PC MRI

Page 9: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

9

PD Dr. Zöllner I Folie 171 I 9/9/2014

Courtesey of J Henning, Freiburg

Examples

PD Dr. Zöllner I Folie 173 I 9/9/2014

MIP Projection – Renal Artery

Page 10: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

10

PD Dr. Zöllner I Folie 174 I 9/9/2014

RAS, Secondary Hypertension, etc.

Parameters to detect status: !  Flow reduction !  Flow velocity ! Kidney size ! Overall and split renal

function

!  RAS leading cause for secondary hypertension " CKD

!  Reduced flow triggers auto-regulation of systemic circulation

!  Loss of renal parenchyma in the stenosed kidney

!  Microangiopathy in the contralateral kidney

Allow for stenosis grading and better selection of patients for PTA

PD Dr. Zöllner I Folie 175 I 9/9/2014

Example - Acquisition Protocol

ECG gated PC cine MR sequence with:

!  TE=4ms, TR=37ms, FA=30°

! VENC=100 cm/s

!  ca. 20 images / cardiac cycle

! Resolution 0.5 x0.5 mm

! Slice thickness 6mm

1.5T Siemens Symphony Vision

Page 11: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

11

PD Dr. Zöllner I Folie 176 I 9/9/2014

Example – Slice Positioning

Schönberg et al. , Radiology 1997

a.  Abdominal aorta

b.  Images acquired from MR angiography

c.  Image slice through the vessel

d.  Imaging plane for PC-MRI (perpendicular to c)

PD Dr. Zöllner I Folie 177 I 9/9/2014

Clinical Interpretation of Flow Profile

Schönberg et al. , Radiology 1997

#  4 section within RR interval

#  Systolic peak (I)

#  Incision (II) followed by midsystolic peak (III)

#  Diastolic phase (IV)

Page 12: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

12

PD Dr. Zöllner I Folie 178 I 9/9/2014

Flow Profile in Stenoic Patient

Schönberg et al. , Radiology 1997

#  First systolic peak and incision disappear with increasing degree of stenosis

PD Dr. Zöllner I Folie 179 I 9/9/2014

Flow Measurement in Stenosis Model

Scheuer et al. Z Med Phys 2010

measurement setup stenosis model at different flow rates

Page 13: Blood Flow Measurement Techniques Typical values for blood flow ...

04.12.2014

13

PD Dr. Zöllner I Folie 180 I 9/9/2014

Flow Measurement in Stenosis Model

Scheuer et al. Z Med Phys 2010

flow profiles from 3D PC-MRI at different locations along the stenosis model

PD Dr. Zöllner I Folie 181 I 9/9/2014

Summary

!  Basic laws to describe pressure and flow: Bernoulli & Poiseuille

!  Blood flow in normal regimes laminar flow

#  Reynolds number

#  Parabolic shape of flow in tubes

!  Blood pressure measurement techniques

#  Korotkof sounds

#  Cuff & stetoscope

!  Blood flow measurement techniques

#  EM flow meters

#  Doppler US

#  PC-MRI