11/26/2004, Coll. WG Meeting E. B. Holzer 1 BLM Data from the Collimator tests in SPS and TT40 B. Dehning, F. Ferioli, E.B. Holzer , L. Jensen, L. Ponce and the Collimator Team 26.11.2004 Collimation WG Meeting, CERN
17

BLM Data from the Collimator tests in SPS and TT40

Feb 10, 2016

Download

Documents

Lajos

BLM Data from the Collimator tests in SPS and TT40. B. Dehning, F. Ferioli, E.B. Holzer , L. Jensen, L. Ponce and the Collimator Team 26.11.2004 Collimation WG Meeting, CERN. SPS BA5. 8 Ionization Chambers. B. B. D. D. Left out Right in. C. A. Left in / Right in Smaller gap. - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 1

BLM Data from the Collimator tests in SPS and TT40

B. Dehning, F. Ferioli, E.B. Holzer, L. Jensen, L. Ponce and the Collimator Team

26.11.2004Collimation WG Meeting, CERN

Page 2: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 2

SPS BA5

Page 3: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 3

8 Ionization Chambers

Page 4: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 4

Page 5: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 5

Page 6: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 6

A

C

DB

A

B

C

D

Left in / Right out

Left out Right in

Left in /

Right in

Smaller gapLeft in Right in

Page 7: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 7A

C

DB

AC

DB

Left out / Right in

Left in / Right out

Left in Right in

Left in

Right in

Smaller gap

Page 8: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 8

Page 9: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 9t [s]

Activation?

Factor 5000 reduction in signal after 4 seconds

Page 10: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 10

Activation measurements in the SPS• Lit: G. Ferioli, R.L. Keizer, Analysys of the induced radioactivity in the SPS

extraction channels during 1994, SL/Note 95-06 (BT)• Ionisation chamber based measurements during the fast extraction (SPS

LSS6) and after to determine the remaining activation • Location of monitors: near kicker magnets and septum's• Ratio between fluence of losses and activation 2000

Page 11: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 11

Decay constants• Two decay constants

– t1: 1.75 s– t2: 278 s

Page 12: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 12

TT40

Page 13: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 13

• BLM1 – BLM4: ionization chambers• BLM5: SEM (no calibration)

BLM1BLM2

BLM3

BLM4

BLM5

Page 14: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 14

Number of electron pairs produced (35 eV) per litre by 4X72 bunches, at floor level. Scale in cm. From M. Magistris.

Measured Simulated

BLM1 3E13 ~ same

BLM2 4E13 ~ same

BLM3 8E13 (scaled) ~ same

BLM4 2E14 (scaled) ~ OK

Signal for 4 batches:

Page 15: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 15

1 - 4 Batches on both Collimator Jaws and the TED

0

500

1000

1500

2000

2500

5.00E+12 1.00E+13 1.50E+13 2.00E+13 2.50E+13 3.00E+13 3.50E+13

BLM1 C-CBLM2 C-CBLM3 C-CBLM5 C-CBLM1 graphiteBLM2 graphiteBLM3 graphiteBLM5 graphite

# protons

BLM

sig

nal

0

500

1000

1500

2000

2500

5.00E+12 1.00E+13 1.50E+13 2.00E+13 2.50E+13 3.00E+13 3.50E+13

BLM1 TED

BLM2 TED

BLM3 TED

BLM4 TED

BLM5 TED

# protons

BLM

sig

nalTED

Collimator

Page 16: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 16

0

500

1000

1500

2000

2500

0 2 4 6

BLM1 C-CBLM2 C-CBLM3 C-CBLM5 C-CBLM1 graphiteBLM2 graphiteBLM3 graphiteBLM5 graphite

0.00E+00

1.00E-11

2.00E-11

3.00E-11

4.00E-11

5.00E-11

6.00E-11

7.00E-11

8.00E-11

0 1 2 3 4 5 6

BLM1 C-CBLM2 C-CBLM3 C-CBLM5 C-CBLM1 graphiteBLM2 graphiteBLM3 graphiteBLM5 graphite

Scanning the Impact Position on the Jaws

Distance from surface [mm]

BLM

sig

nal /

# p

roto

ns

BLM

sig

nal

Distance from surface [mm]

Page 17: BLM Data from the Collimator tests in SPS and TT40

11/26/2004, Coll. WG Meeting E. B. Holzer 17

Summary• No activation observed (Signal decreases by at

least a factor 5000 when collimator moves out).• “Tails” in distributions are from the beam.• BLM signal is linear with proton intensity.• Left-right asymmetry of the shower depends on

the collimator gap size and gap position.• Slight top-bottom asymmetry?• BLM signal depends on the impact position on

the jaw.• Compares ~ OK with simulations (TT40).• Further analysis …