Biotin Interference in the Siemens Centaur TnI Assays Effective June 1 st , 2016 Providence Health & Services Oregon Laboratories will be including the following comment in the Siemens Centaur TnI assays: “Excessive Biotin may affect TnI results. Serial monitoring is not affected. Contact Client Services at 503-215-6660, if there are any questions.” In a recent communication, Siemens Healthcare Laboratory Diagnostics informed us that Biotin causes an increase in the Centaur TnI assay as shown in the table below. Increase in TnI-Ultra Result per Biotin Concentration (ng/mL) Biotin Concentration (ng/mL) TnI-Ultra Sample Result (ng/mL) 1 2.5 5 10 20 0.014 0.000 0.006 0.003 0.006 0.006 0.030 0.008 0.008 0.007 0.007 0.006 0.049 0.000 0.002 0.008 0.008 0.010 0.066 0.003 0.005 0.009 0.013 0.012 0.095 0.005 0.009 0.013 0.016 0.014 0.158 0.004 0.011 0.018 0.032 0.028 0.292 0.007 0.023 0.040 0.052 0.044 0.416 0.020 0.034 0.048 0.064 0.061 0.587 0.006 0.040 0.064 0.097 0.082 0.704 0.017 0.042 0.083 0.105 0.095 1.453 0.065 0.096 0.184 0.257 0.221 The ability of the TnI assay to detect serial increases is not affected by the biotin interference since the bias is proportionate across TnI concentrations. Siemens has declared that they are actively investigating the root cause of this issue and looking for solutions on a new formulation of reagent. We interpret this interference as having low clinical impact; however, we are required to share this information to increase awareness. For more information, please contact Drs. Veronica Luzzi, June Olson or Alec Saitman at Providence Lab Services at (503) 215-6660.
1

Biotin Interference in the Siemens Centaur TnI Assays/media/Files/Providence OR...Biotin Interference in the Siemens Centaur TnI Assays Effective June 1st, 2016 Providence Health &

Jul 05, 2018

Download

Documents

phungtuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Biotin Interference in the Siemens Centaur TnI Assays/media/Files/Providence OR...Biotin Interference in the Siemens Centaur TnI Assays Effective June 1st, 2016 Providence Health &

Biotin Interference in the Siemens Centaur TnI Assays

Effective June 1st, 2016 Providence Health & Services Oregon Laboratories will be including the following comment in the Siemens Centaur TnI assays:

“Excessive Biotin may affect TnI results. Serial monitoring is not affected. Contact Client Services at 503-215-6660, if there are any questions.”

In a recent communication, Siemens Healthcare Laboratory Diagnostics informed us that Biotin causes an increase in the Centaur TnI assay as shown in the table below.

Increase in TnI-Ultra Result per Biotin Concentration (ng/mL) Biotin Concentration (ng/mL) TnI-Ultra Sample Result (ng/mL)

1 2.5 5 10 20

0.014 0.000 0.006 0.003 0.006 0.006 0.030 0.008 0.008 0.007 0.007 0.006 0.049 0.000 0.002 0.008 0.008 0.010 0.066 0.003 0.005 0.009 0.013 0.012 0.095 0.005 0.009 0.013 0.016 0.014 0.158 0.004 0.011 0.018 0.032 0.028 0.292 0.007 0.023 0.040 0.052 0.044 0.416 0.020 0.034 0.048 0.064 0.061 0.587 0.006 0.040 0.064 0.097 0.082 0.704 0.017 0.042 0.083 0.105 0.095 1.453 0.065 0.096 0.184 0.257 0.221

The ability of the TnI assay to detect serial increases is not affected by the biotin interference since the bias is proportionate across TnI concentrations. Siemens has declared that they are actively investigating the root cause of this issue and looking for solutions on a new formulation of reagent.

We interpret this interference as having low clinical impact; however, we are required to share this information to increase awareness. For more information, please contact Drs. Veronica Luzzi, June Olson or Alec Saitman at Providence Lab Services at (503) 215-6660.