Page 1: Biospect 21 Manual_eng
Page 2: Biospect 21 Manual_eng

Research Scheme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Save to disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Reprinter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Comparative analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Entropy analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44NLS Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Contents:

2

Patient's card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Determination of resonance frequency of system modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Test (research) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Administrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Program “Multiscreening Ultra”. . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Page 3: Biospect 21 Manual_eng

test of nutritional supplements.

4

Program "Multiscreening Ultra"Program "Multiscreening Ultra" is designed for the computer nonlinear analysis and virtual information

Page 4: Biospect 21 Manual_eng

Program Multiscreening Ultra

When the program is activated, the Main Menuappears as a group of buttons.

When you press the "Reception of patients" button, you can start working with the program.

When you press the "Settings" button,the program settings appear. There

on or off, change the font size andcorrect the processor deviation if required.How to make the correction of the processordeviation: move the triangle mark above thescale and correct the evaluation on the point located on the image of an organ in case if the point's evaluation deviates because of

the specific characteristics of the processor compared to the processor accepted as a standard one.

The “Select doctor ” button allows you to select an operator, who will be performing the analysis, from the list of operators.

5

or Deutsch), as well as switch the sound you can select the language (Russian, English

Page 5: Biospect 21 Manual_eng

The "Exit" button enables to exit the program.

Here you can: By pressing the “ Color images” button, make all pictures in colour, or by pressing this button repeatedly, make them black and white. The “ Color in card-index” button enables to colour the names of the organs with different functional conditions in the list of the completed tests:

Green - no expressed changes in the functional condition.Red - insignificant changes in the functional condition.Brown - with significant functional and organic changes.

6

The drobdown list "Signal generation speed" enables to change generation speed of researches.

But it should be taken into account that increase of generation speed decreases reliability of reserches.

Page 6: Biospect 21 Manual_eng

The “ Administrator” button enables the owner of the device to go tothe administrator mode that gives him/her an administrative right to:- set (change) an access password in the administrator mode with the“ Change password” button;- enable or disable other users to view the patients' card index with the "Close card-index" button;- turn on or off the mode of keeping the records of payments made by the patient for the performed tests with the "Record Fees" button;

When you press the"Delete card file" buttonin the right column, the card of ahighlighted patient is deleted,and with the button“ Delete research”located above the list of tests,the individual test is deleted.

Administrator

7

Page 7: Biospect 21 Manual_eng

Administrator

By pressing the “Doctor selection” button, you will get the list of operators - specialists who examine patients with this device.When you select the name of a specialist, the card index of his/her patients appears.When you press the "New" button in this catalog,you can expand the list of operators -users of this device; and in the same way, you can delete a user from the list with the "Delete" button,and at the same time delete the card index of his/her patients. With the "Patient exchange" button you can move the patient's card from one specialist to another.

The “ Delete by date” buttonallows to delete the information onthe tests of all patients carried outuntil the certain date. Set the datein the " Delete by date" screen; all testscarried out until this date to be deleted.

.

8

Page 8: Biospect 21 Manual_eng

Administrator

The button “ Report” allows to make a statisticalanalysis on the tests time and the quantityof patients as well as on the number of testscarried out by every researcher and by the specialistsoverall, and also to make a financial analysisof the performance of the researchers group.

9

The line “ Search” allows to find a card file on the first letter of the patient's last name.

Page 9: Biospect 21 Manual_eng

Before the test, the patient is seated in front of the device, the headphones with the magnetic

inductors have to be placed according to the marking on them: R on the right temporal point,

10

L on the left.

The program "Multiscreening Ultra" has a general sheme for forms.

There are reference information in the top line.

The left side of the line shows the name of the formThe right side of the line is the button "Exit".

Control buttons are made in the form of icons, a description of their functions appears below the icon when you aim

at it the mouse cursor.

Page 10: Biospect 21 Manual_eng

After the button "Research" is clicked, "The patient's card" appears, where theinformation about a patient is stored,including: the last name, first name,middle name, age, sex, address, and the phone number if available.

information about the performed testof a selected patient.

you have to fill in according to the requirements: last name, first name, middle name, age, address, phone; and select the sex from the drop-down list.

Patient's card

11

In the main table, there is the complete

When you press the button "New card", there is a form that

By pressing the button "ok" we go to the form "catalogue".

Page 11: Biospect 21 Manual_eng

appears, where the information aboutall persons tested by this specialist is stored, including the last name, first name, middle name, age, address, sex,and the phone number if available.Note, that the administrator can closethis form if necessary. The quick search of the requiredpatient's card, can be made in this form, if patient's last name is entered in

first several letters from the lastname are enough to enter).The similar search method can be usedin the form of "Patient's card".

Patient's card

12

When you press the button “Select card" the form "Patients' card-index"

the“ Search” (sometimes only

The button "Research" allows to start a new test, to continue further test during the day or start control test.

a few patients randomly selected from the catalogs at once.

This button opens a window in which the physician may write an opinion on the health status of the patient.

This option may be necessary for the better understanding the specifics of lesions in specific diseases

Finally, you can print it by clicking "Print".

The button "Compare patients" allows to compare the state of a certain tested system (by pairwise comparison)

Page 12: Biospect 21 Manual_eng

By pressing the button “Printresearch” we go to a form, which

enables us to select the print parameters.The program allows to print out up to4 images of size 100X100 mm on one A4 sheet.There is also a possibility to lookthrough the earlier "Epicrisis"by bringing the standard processeswith big spectral charts taken from the patient's test (if the spectraldifference coefficient (D) is less than 0.425), up for the test and for printing (the button "Recomendations" and "Print").

The program allows to print outthe list of the most effective nutritionalsupplements, selected by a specialist according to the results of thecurrently performed test or to make

corrections of the earlier selected preparations by exiting the pint parameters mode to the test mode (the button "Analysis").

Patient's card

13

Page 13: Biospect 21 Manual_eng

The main elements to manage the "Patient Card" form:

"Current Analysis" - the current analysis of the results, with the possibilityto make the post check of a patient with the device or without it(then the button "View Results").

"Comparative Analysis" - the comparative analysis of the results of several tests following the dynamical observation of a patient.

“ Research”- when you press this button, the form “Scheme ofinvestigation” appears.

Patient's card

14

The “ Print comments" button allows to print out additional recommendations and notes for the patientmade by the doctor.The “ Make pause between pages" button allows to pause the printing in order to print out the resultson the both sides of a sheet.

The “ Color ” button allows to make images in color, and when it is not activated, in black and white.

The “ Print ”

button - starts printing.

"Save to disk":allows to record the information on the test of a selected patient on the disk,

with the opportunity to the further copy it to the program.

Page 14: Biospect 21 Manual_eng

is designed to preparethe object for the test.

15

Catalogue

The form "Catalogue"

The button "Options"allows to change some

parameters of research.

The button "Text mode"allows to display all

systems in text.

The button “ Automatic/Manual choice" allows to make an automatic/manual selection for more detailed

separate elements.test of anatomical, histological and cytological structures depending on the given changes in its

Page 15: Biospect 21 Manual_eng

Research Types:Express - makes tests on the full topographical images without detail. Normal - performs a study of several biological structuresif only there is a pathological change. Detailed - evaluates the structure of all biological formations that may be necessary for the high-quality scientific research.

Scheme investigation

the left of the name of the image with the help of the left mouse button.The button "Uncheck" removes the selection of all images.The button “ Restore” restores the selection of all images.The button “ Alphabet order ” sorts all images in the alphabetical order.

The button "Preparation"- the making of preparation is performed automatically for the images, where the primary siteswere already evaluated. After pressing of this button you have to make appropriate settings in the "Reprinter" form,and press the button "Start preparation".

16

The test scheme of images in the list “Catalogue” is determined by setting and removing flags on

The button “ Analysis” gives the current analysis of test results.

The button “Ectomies” allows to eliminate an attempt to research ectomies' organs.

Page 16: Biospect 21 Manual_eng

Scheme of investigation

17

By pressing the button "Select a system" on the screen displays 9 buttons with the image of organs specific to

the relevant anatomical structures . When choosing a system to research there are available only organs of the choosen system,

on the top line appears its name. To return to the base catalogue need to re-press the button corresponding to the selected system .

"Acupunctura" - this button allows

to enter a research of biological

active points. On the screen appears images

for evaluation of acupuncture points.

"UltraStructure" allows to use

UltraStructure's mode. In this mode on

the organ's image appears the icons, which

allow to enter the test of another object. This

object is anatomically or functionally related

To step back press the button to the basic image.

Page 17: Biospect 21 Manual_eng

Scheme of investigation

18

The "Ultrastructure" mode allows to enablehistological, cytological and molecular analysis.Moving the cursor on the projection of the organ,selects the structure image, whose name and

image appears on the right.To start the research of choosen structure at a givenpoint press the left mouse button.

expanded by clicking on it.It can be carried out a several ultrastructure's researcheswith different localization (in one organ), graduallymoving from histological sections to cytological preperations and molecular structures.

After the end of the research icon appears on organ image with the number of research, which can be

Page 18: Biospect 21 Manual_eng

19

When you press the button "Savethe form appears in the to disk",

card index. It enables you to copythe test data of specific patients either on a disk or from the disk to the appropriate directory of patients.

At first, the copy mode either "From program" or To program" to be set.

"From program " - the test data of specific patients s copied on the disk.

Save to disk

This list may consist of a few patients.

When the list will be ready, press the button "Copy to disk" and indicate the path where you want to have a copy.

(in the bottom part of the form displays the list of tests f a given patient). The highlighted patient's name will be moved

to the right partof the form.

The list of copied patients is created by highlighting patient's name and pressing the button "Copy"

Page 19: Biospect 21 Manual_eng

Save to disk

"To program" - copies test data from the disk to a directory of patients available in the computer. When you go to this mode, it is necessary to specify the source of data (diskette or disk), after that in the right part of the form,the list of persons and their tests available for copying to the computer will appear.Then after pressing the button "Copy", the contents of the disk will be copied to the program.

The button "Clear", clears the complete list of persons and teststo copy (without deleting them from the disk).

“ Exit ” - exits from the copy mode.

20

Page 20: Biospect 21 Manual_eng

Determination of resonance frequency of system modulation

There is a Fleindler's logarithmicpolychrome scale in the lower partof the screen.

Pulsing image of an organ.

Chart of an amplitude distributionof measured signal in the standardfrequency 1.8 - 8.2 Hz.

The button “ Pause/Resume” - stops the test . When you press the button, it turns to the "Resume" buttonpress this button to continue the automatic test. "Exit" - enables to go back to the "Research Scheme".

Relative scale of a noise level in the system in decibels.

Conventional scale in points of the valid signal/noise ratio.

Scale of a relative pulse duration of the measured signal in percent.

In the mode “Determination of resonance frequency of system modulation ” , there is an opportunity to evaluatethe spectral characteristics of the nidus; all main buttons are similar to the form "Determination of resonance frequency of system modulation".

21

Page 21: Biospect 21 Manual_eng

In the mode “ Research” , the measurement of the representation of a destructive process in the control points, located onthe image in the standard way, ismade. The results are evaluated on thesix-point Fleindler's polychrome scale,its buttons are located in the lower right

part of the screen.

The button “ Pause/Resume”enables to pause the test.

The button “ Exit” - exit to the mode“ Scheme of investigation ” .

Evaluation in points of the NLS Analysis results.

1. The level of a latent functional activity.2. The level of balanced regulation.3. The shift of the characteristics to the higher level, stressed condition of regulatory systems.4. Asthenia of regulatory mechanisms.5. Compensated failures of adaptation mechanisms.6. Decompensation of adaptation mechanisms, represented destruction processes.

22

Page 22: Biospect 21 Manual_eng

Please note that the suggested evaluation in points gives to the greater extent the dynamics of increasesor decreases of adaptation reserves; any stable destructive conditions without expressed dynamical changesare poorly registered on this scale.

Reprinter

Natural frequencies taken from the nidus are converted into theopposite in polarity and identical inform frequencies, and then they arerecorded on a matrix, such as water, ethyl alcohol or sugar, as converted and amplified information. Water is recommended to be usedonly to cure any acute processes.The duration of the efficiency ofthe preparations recorded on wateris about 2-3 weeks, then it is lost.Alcohol is used to treat subacuteand chronic processes.The alcohol solutions of phytoprepara-tions are recommended for use forelimination of chronic processesThis helps to prolong the validitytime of preparations.

23

Page 23: Biospect 21 Manual_eng

The preparations recorded on alcohol are valid up to 2-3 months, and on alcoholic solutions- 4-6 months. It is possible to treat chronic disorders using preparations recorded on lactose, in this case the preparation is validfor up to 6-9 months, but note that lactose used is supposed to be moistened with alcohol or water. For treatment of acute processes, take 4 to 8 drops or "peas" at a time, three times a day. For children under 14 years, the dose should be 2-3 drops less.For treatment of subacute or chronic disorders, take 2-4 drops (peas) once or twice a day. For children under 14 years,the dose should be 1-2 drops less.Paraffin is used for treatment of pathological processes on skin and peripheral nervous system disturbances(radiculitis, neuralgia, lumbalgia) by means of application.For treatment of acute processes, it is possible to record from 2 to 4 preparations on one matrix at a time.For treatment of subacute chronic processes, the number of preparations can be increased to 6-8. Please note that the more is the voltage of recorded preparations the fewer preparations can be recorded on one matrix.

The button “ Invert” enables to invert the output signal into the one of opposite polarity.In this mode, there is the opportunity to succeeding record of preparations from several pathology nidi existing in one organ (part of the body). To switch from one nidus to another, point the arrow to the appropriate nidusand press the left mouse button.

Reprinter

24

The button "Start preparation " starts making a preparation, and

The button "Stop preparation" interrupts the process of making a preparation.

Page 24: Biospect 21 Manual_eng

Etalon

Every cell have its own specific oscillations that are stored in the computermemory. They can be displayed onthe screen as a specific chart, whichexpresses the conditions of informationalexchange between the tissue and theenvironment. Any pathological process hasits own, peculiar only to it, specific chart.A great deal of pathological processwith regard to the extent of their representation, as well as to the age,sex and other variations, is recorded inthe computer memory.By reading the frequency characteristics

of the tissue, the device can compare them by the spectral similarity tothe standard processes (healthy tissue, tissue with pathological changes,infection agents) and detect the closest pathological process or tendency forits origination. By means of analysis, you can follow the input signal, displayed in red (S),and output signal, displayed in blue (N) on the chart.

25

Page 25: Biospect 21 Manual_eng

Etalon

By viewing the charts, you can determine which standard process is the closest, as well as to determine the degreeof spectral similarity between the standard process and the chart, read from a patient.

"The list of Groups"- contents thelist of groups of standards.

ORGANOPREPARATIONS contains the standards of healthytissues' spectral characteristics.

In the "Etalon list" below, the etalonsare arranged in the descendingorder according to the spectralsimilarity to the object of the research.Healthy tissues essentially have nodissociation with the incoming andoutgoing signals (correspondingred and blue graphs).The less is the difference betweenan organopreparation and an object,the greater is the intact of the tissuein the body under the test, the greater - the greater is the damage of the tissue.

26

Page 26: Biospect 21 Manual_eng

Etalon

Enables to makes qualitative assessment of the main biochemicalindictors by wave functions of tissues.The assessment of biochemicalhomeostasis is made in the NLS-analysismode.Note, that the lowest value of ferments (hormones) concentration wit-hin the standard limits on the chartcorresponds to 2, the upper valuecorresponds to 6. Values 3, 4, 5correspond to the value mode,and boundary values such as 1 and 7

show that biochemical indexes are outside the standard physiological limits,lower and higher respectively. Results of the computer biochemical analysis will be true only if the standard

regulations for conducting of a biochemical analysis in a hospital usingconventional methods are observed.

BIOCHEMICAL HOMEOSTASIS

27

Page 27: Biospect 21 Manual_eng

Etalon

PATHOMORPHOLOGY

Contains the list of standards of pathological processes.This catalog contains most ofpathomorphological conditions featured by certain body tissues.Every pathological processhas a unique chart.

28

Page 28: Biospect 21 Manual_eng

Etalon

This catalog contains all the main features of the infection agents, including bacteria, viruses mycoplasma ricketsias,fungi and helminthes. As a rule, they have high peaks of dissociation within the frequency range where the natural tissue frequencies are represented, and toward which they have a tropism.

MICROORGANISMS AND HELMINTHES

29

Page 29: Biospect 21 Manual_eng

Etalon

frequency of a urinogenital system.

ALLOPATHY group contains wave characteristics of main chemical (synthetic) medical drugs used in theconventional medical practice.

HOMEOPATHYThis group contains wave characteristics of homeopathic preparations.

30

E.g., opisthorhiasis has a dissociation on the 4.9 Hz

Page 30: Biospect 21 Manual_eng

PHYTOTHERAPYThis group contains wave characteristics of medical plants that grow in the Midwest of Russia.

NUTRICEUTICALSThis group contains wave characteristics of nutritional supplements (BAD) produced by main Russian and foreign producers of nutritional supplements.

ALLERGENSThis group contains wave characteristics of food, domestic, vegetable and industrial allergens.

Etalon

Natural frequencies of tissues lie within the following standard frequency range.1.8 Osseous system.2.6 Rough connective tissue, joints, and heart valves.2.6-3.4 Loose connective tissue, striated musculature, cardiac muscle.3.4 Unstrained musculature.4.2 Tessellated epithelium of the digestive tract.4.9 Stratifies squamous and columnar epithelium. Parenchymatous liver tissue and tissue of thebiliary tract.4.9-5.8 Kidney tissue epithelium and reproductive organs.5.8 Lymphoid ring of the pharynx, upper section of the respiratory tract, lymphatic system, spleen,ovaries, prostate.6.6 Peripheral nervous system, bronchus epithelium, adrenals, thyroid. 7.4 Central sections of sensory analyzers except the optic one, subcortical structures of the brain,pons; cerebellum, limbic system and lungs parenchyma.8.2 Retina, optic nerve, cerebral cortex.

31

Page 31: Biospect 21 Manual_eng

Etalon

Next to the chart there are the following control buttons: “ Object ” - displays a chart of the examined biological object, organ or tissue plotted in the courseof the test.“ Etalon”- represents a chart of the selected standard process.“ Spectrum”- by pressing this button the chartis enlarged.“ Optimum”- a yellow chart that shows standard (Gaussian) distribution of the signal in thestandard frequency range.“ Model ” - displays a chart of a virtual model.“ Invert ” - enables to invert the chart into the opposite polarity.

The button “ Clear”, above the list of standards shuts off the opportunity of dispersion analysis simultaneously for all groups of standards.

32

Page 32: Biospect 21 Manual_eng

This program gives a unique opportunity to make a comparison of allpreparations by the extent of their spectralsimilarity to a pathology nidus. The marked arearepresents the values of spectral differences (D)between the standards and the object.If the value is the less than 0,425, it means that the spectral similarity to the examined objectis over 95%, and this standard is marked red.It is clinically significant, i.e. there is a statisticallyproven expression of processes, if this value is not greater than 0,750(the similarity to the examined object is not less then 65%).

Etalon

A math addition of spectral characteristics of information preparations gives an opportunity to obtain the bestcombinations of remedies, the closest to the spectral characteristics of pathological process, and to choose the mosteffective remedy. The possibility to combine frequencies of pathological agents gives an opportunity to experimentallymake virtual models of various pathological processes. It is essential to make virtual models in selecting groupsof remedies that give the best results when combined.

33

Page 33: Biospect 21 Manual_eng

Etalon

The combination of standards with the purposeto make a virtual model is done by pressingthe left mouse button in the first columnof the list of standards against the selected etalon.The elimination of a standard from a virtualmodel is done by means ofthe right mouse button.Turning off the virtual modeling is done by placing the mark [X] in the first column of the listof standards against the virtual model. If the spectral differences value (D) of any addedpreparations and the object taking separately ismuch higher than that of the virtual model of theircomposition there is the synergism of preparations.If the value is lower there is antagonism in their interaction.

Control buttons of the "Etalon" form

The "Adjustment" button - virtual activization of the pathological process.

The process may be in the remission stage but we are interested in the active process.When this button is pressed, the selected reference process is adjusted to thespectral characteristics of the examined object.

"Adjust all"- all reference processes stored in the standards directory are adjusted.

34

Page 34: Biospect 21 Manual_eng

Etalon

Sign for other standards of the model indicates that the share of the control standard in themodel structure is minimized.

"Reprinter"- enables to record any preparation from the list of standards on a matrix(water, alcohol, sugar, paraffin).

“ View report”- displays the list of standards selected for printing on the screen.

35

"Vegeto test"- the use of the mode of internal vegetative resonance test where the preparation is selected from the list of standards.

“ Entropy analysis”- carries out a two-fold factorial entropy analysis of a pathological process.

“ NLS-analysis”- carries out a multifactorial entropy analysis of a pathological process(biochemical homeostasis of the system).

“ Etalon-Object”- carries out a dispersion analysis with reference to the randomly selected standart withthe object functions assigned to it.

“ To print”- indicates the standards with a high spectral similarity to the examined object in the printout.

“ Imperative model ”- this function is similar to the "Absolute model", except that one of the standards(control standard) is selected randomly and it is included into the calculation of the virtual modelto evaluate its spectrum in the construction of the model.

“ Absolute model ”- displays an extremely optimized virtual model on the basis of the development of anintricate program algorithm that enables to handle all kinds of standart combinations.

"Bacterial analysis" - (available only if the spectral similarity index of the standard of "microorganisms" is less

than 0,750, which might suggest that the presence of inactivity of the microorganism in biological tissues) start

up the mode of Bacterial analysis.

Page 35: Biospect 21 Manual_eng

The buttons above the list of standards:

“ Fix”- allows fixing of three lines in the upper part of the list:Name of the biological object.Optimum distribution.Virtual model.

“ Sort”- allows to arrange the standards in the alphabetical order.“ Select”- allows to select a group of standards with the similar last word in their names.The search box in the lower left corner, under the list of standards enables to find the required standard very quickly by typing a few first letters of its name.

Etalon

36

Elimination - allows for multifactorial eliminate analysis due to exclusion from the spectrum of the object

the frequency of individual agents (usually bacterial or viral nature), which allows to detect masked by these agents, the basic process.

“Etalon description”- displays a detailed description of the selected standard.

“Gropups”- allows to determine the number of groups of standards in the same system, differing in their latest titles.In this mode you can select multiple groups of standards (right-click) which will be displayed after clicking the button Filter.

Page 36: Biospect 21 Manual_eng

Analysis

Press the button

mouse button is pressed on the cross, the message will be printed.Tick the boxes on the left to the textof the message in the displayed form to print.

This mode gives an opportunity

related to this test with the helpof icons.

Press the button "Icons" in order to do this. The icons will appear on the image. To be able to seewhich organ is represented by the icon, place the mouse arrow on the icon, in the box beside, the name of the image will appear; and if the mouse button is pressed on the icon, the corresponding organ will be displayed for examination.

37

Buttons to the right of the organ'simage

“Describe Structure” - enables to receive information about

any fragments of the image."Describe

Structure" then green crosses signs appear on he image. Place the mouse arrow on the cross to read the text in

to move to the examination of any organs

the square beside it. If the left

The button “ Describe functions”- enables to read the description of the biological object shown on the image.

Page 37: Biospect 21 Manual_eng

It is also possible to display (hide or switch to) the nidus by pressing the left mouse button with the arrow pointing to the nidus.

Then it is necessary to click on the image and by moving the mouse to extend the frame, and click the mouse againto highlight the enlarged fragment. To restore the full image from the fragment, press this mouse button again.

38

The button “ Nidus ”- enable to display or hide the pre-selected nidi on the image of an organ.

Press the button in the upper left corner of the image to enlarge a fragment of the image.

"Ultrastructura" button shows the histological, molecular and bacterial analysis. You can start up ultrastructural

analysis by clicking on the appropriate icon.

The program allows to accurately determine the limits of the nidus, placed by any number of points on the projection

of the organ which will be further evaluated by pressing the button "More precise nidus".

Page 38: Biospect 21 Manual_eng

"Comperative analysis" - a comparative analysis of examination results under dynamic observation of the patient.

"Meta-therapy" - by pressing this button you will enter the mode, which enables to produce a correcting effect by using the infrared scannerTo do this place the influence center on the image and press the button "Start".

Analysis

39

Page 39: Biospect 21 Manual_eng

"VegetoTest" button.

"VegetoTest"

VegetoTest using the preparation that is not found in the list of standards of nutritional supplement's allergens, we should go

40

provided there is information influence of the preparation under test.

The use of the vegetative resonance test mode gives wide opportunities and allowsintroducing into the measurements of the patient's natural electromagnetic oscillations,electromagnetic oscillations of the standards from the test-set, the internal VegetoTest,

tolerance of a nutritional supplement as well as an allergic sensitivity of the patient's body. To make the or taken by means of the resonance chamber - external VegetoTest , and determine the efficiency and

to the "Analysis" mode, put the preparation under examination in the resonance chamber, press the button"VegetoTest" , and enter the name of the tested preparation, then press the button "Ok". The dynamicsof the vegetative reaction of the body to this preparation is best evaluated in the "Comparative analysis" mode, which determines he intensification or relaxation of the body's compensatory reactions

The internal VegetoTest is carried out in the "Etalon" mode, the preparation is selected from the list of standards where it has lready been recorded as a digital model, by means of the left mouse button/

The mode is started by the

Page 40: Biospect 21 Manual_eng

ReprinterReprinter enables to record any nutritional supplement on the matrix. In order to do this, select from the mode "Etalon"a preparation proven effective for this patient (the value "D" is not greater than 0.425) and press the button "Reprinter".Note, that all medical preparations have been recorded in an inverted form so before making a preparation re-inverting is supposed to be done by pressing the button "Invert" in the chart, then a matrix should be selected just like it is

done in making nozodes, after that the button "Start preparation" should be pressed.

Analysis

41

Page 41: Biospect 21 Manual_eng

by spectrum frequencies.

The button “ Evaluate region ” -allows making themore thorough evaluation of themarked area on the enlargedimage of the organ.

“ Localize” - defines the boundariesof the nidus with the roughest changes

in the organ's structure.To define the boundaries, click the leftmouse button on the organ's imageto single out the area that is of interest,clicking the left mouse buttonagain closes the frame.

“ Evalute region” - the nidi with mostpronounced pathological changes

in the organ selected by thedoctor are evaluated. The evalutionis automatic, unmarked organs andorgans free of nidi being skipped.

“ Delete selected ” - allows to deletethe selected nidi with the right mouse button.

42

Press the "Test" button, and display the form "Etalon".

The button "Spectrum" displays an enlarged chart with the values of signal amplitudes distributed

Page 42: Biospect 21 Manual_eng

Comperative analysis

Comparative analysis allows to assign the values dynamically to the results of the conducted adjustement, andalso to evaluate the condition of the compensatory reactions of the body following the vegetative testing of the preparations. The left image features the initial condition; the right one gives a dynamic representation or characterizes the condition after a period of time. The button "Auto Search" allows to automatically retrieve the organs to be compared in the card index. The buttons "Improvement" and "Aggravation" allow to focus attention on the parts of the tissue structure where some changes have taken place.

43

Page 43: Biospect 21 Manual_eng

Entropy analysis is accessible from the mode "Etalon"

Entropy (two-factor) analysis allows tobuild a mathematical model of a pathological process by taking thehealthy tissue (organopreparation) asthe initial (zero) phase, and the clinicallyexpressed from of a pathological processas the final phase while making amathematical calculation for the numberof charts of some intermediate condi-tions. In the course of the analysis,the highest spectral similarity to anyintermediate condition or extremeconditions is determined, and thus the maturity of this process is determined.

Note that the values 1, 2 of the entropy factor indicate that there are no tendencies in thedevelopment of the tested process; the values 3, 4 indicate that there are preclinical phases in the process development and the values 5,6 indicate that the process is mature.

The low spectral similarity to the standard (correlation over 1), in case that the entropy factor is as high as 5, 6, indicates a remission state of the pathological process against low adaptive reactions of the tissue.

44

Page 44: Biospect 21 Manual_eng

Entropy analysis

A small difference in the spectral similarity (dispersion) over the whole range of entropy factors from 1 to 6 indicatesan acute process.

The most similar chart in terms of the spectrum is marked by a thick line at the chart, and the value is marked red, in this case it is 0.760.

To view the charts separately or together, use two buttons "Hide N graph" and "Hide S graph" that hide the corresponding chart.

To view the chart in the mode "Surfaces" , press the button "Surfaces"; to get backto the mode "Lines", release this button.

45

values for this process.The button "Object" displays the optimum

“Vera-Test” represents the chart with the closest spectral similarity.

Page 45: Biospect 21 Manual_eng

Entropy analysis

46

It is possible to trace the dynamics of the process by pressing the button"Dynamic" above the chart.

The button "Spectrum" displays an enlarged chart.

"Exit" allows to exit from the entropy analysis.

"Intendance - Screening" - allows to determine the extent of destructionof different tissues by a developing pathological process. It is possible to trace the representation of pathological changes either in all groups of tissue structures at a time (button " A "), or in the mode of separate morphological groups.

"A" (red colour) represents the arterial system;The button “V” (blue colour) the venous system;The button

The button “N” (yellow colour) the nervous system;The button “L” (green colour) the lymphatic system;The button “M” (orange colour) the locomotor apparatus:The button “O” (lilac colour) all other groups of tissues that are not included in the above mentioned list.

Page 46: Biospect 21 Manual_eng

47

NLS Analysis

The NLS Analysis - is accessible from "Etalon".

The NLS-analysis (multidimensional) isthe mode

that in the NLS-analysis all intermediateidentical to the entropy analysis except

stages are recorded as standardsof the biological object in the differentstages of the developing process.This allows building the more accuratedevelopment model of the pathology.This is a very complicated and laborioustype of the process record so the NLS-analysis is used solely to evaluate bioche-mical factors.The increase of the amplitu-de of the output signal (the blue line in the chart) indicates the intensification ofthe compensatory mechanisms. Whenthe maximum value of the input signal inthe chart is reached, it may dropabruptly with high amplitude values of the inputsignal (red line), which indicates a failurein the adaptations mechanisms and thedevelopment of irreversible conditions.

Page 47: Biospect 21 Manual_eng

NLS Analysis

The NLS Analysis - is accessible from the mode "Etalon".The NLS-analysis (multidimensional) isidentical to the entropy analysis exceptthat in the NLS-analysis all intermediatestages are recorded as standardsof the biological object in the differentstages of the developing process.This allows building the more accuratedevelopment model of the pathology.This is a very complicated and laborioustype of the process record so the NLS-analysis is used solely to evaluate bioche-mical factors.The increase of the amplitu-de of the output signal (the blue line in the chart) indicates the intensification ofthe compensatory mechanisms. Whenthe maximum value of the input signal inthe chart is reached, it may dropabruptly with high amplitude values of the inputsignal (red line), which indicates a failurein the adaptations mechanisms and thedevelopment of irreversible conditions.

48