Top Banner

of 30

Biopotentials.ppt

Jan 18, 2016

Download

Documents

Fidaa Jaafrah
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • Chapter 74: Biopotentials and Electrophysiology MeasurementTeemu [email protected]

    butler.cc.tut.fi/~malmivuo/bem/bembook/

  • Agenda1st halfIntroduction to biopotentialsMeasurement methodsTraditional: ECG, EEG, EMG, EOGNovell: VCG2nd halfMeasurement considerationsElectronicsElectrodesPracticesQ&A

  • What are biopotentialsBiopotential: An electric potential that is measured between points in living cells, tissues, and organisms, and which accompanies all biochemical processes. Also describes the transfer of information between and within cellsThis book focuses strictly on the measurement of potentials

  • Mechanism behind biopotentials 1/2Concentration of potassium (K+) ions is 30-50 times higher inside as compared to outsideSodium ion (Na+) concentration is 10 times higher outside the membrane than insideIn resting state the member is permeable only for potassium ions

    Potassium flows outwards leaving an equal number of negative ions insideElectrostatic attraction pulls potassium and chloride ions close to the membraneElectric field directed inward formsElectrostatic force vs. diffusional forceNernst equation:

    Goldman-Hodgkin-Katz equation:

  • Mechanism behind biopotentials 2/2When membrane stimulation exceeds a threshold level of about 20 mV, so called action potential occurs:

    Sodium and potassium ionic permeabilities of the membrane changeSodium ion permeability increases very rapidly at first, allowing sodium ions to flow from outside to inside, making the inside more positiveThe more slowly increasing potassium ion permeability allows potassium ions to flow from inside to outside, thus returning membrane potential to its resting valueWhile at rest, the Na-K pump restores the ion concentrations to their original valuesThe number of ions flowing through an open channel >106/secBody is an inhomogeneous volume conductor and these ion fluxes create measurable potentials on body surface

  • Electrocardiography (ECG)Measures galvanically the electric activity of the heartWell known and traditional, first measurements byAugustus Waller using capillary electrometer (year 1887)Very widely used method in clinical environmentVery high diagnostic value

  • ECG basicsAmplitude:1-5 mVBandwidth:0.05-100 Hz

    Largest measurement error sources:Motion artifacts50/60 Hz powerline interference

    Typical applications:Diagnosis of ischemiaArrhythmiaConduction defects

  • 12-Lead ECG measurementMost widely used ECG measurement setup in clinical environmentSignal is measured non-invasively with 9 electrodesLots of measurement data and international reference databasesWell-known measurement and diagnosis practicesThis particular method was adopted due to historical reasons, now it is already rather obsolete

    Einthoven leads: I, II & IIIGoldberger augmented leads: VR, VL & VF Precordial leads: V1-V6

  • Why is 12-lead system obsolete?Over 90% of the hearts electric activity can be explained with a dipole source model

    Only 3 orthogonal components need to be measured, which makes 9 of the leads redundantThe remaining percentage, i.e. nondipolar components, may have some clinical value

    This makes 8 truly independent and 4 redundant leads12-lead system does, to some extend, enhance pattern recognition and gives the clinician a few more projections to choose from

    but.If there was no legacy problem with current systems, 12-lead system wouldve been discarded ages ago

  • Electroencephalography (EEG) Measures the brains electric activity from the scalpMeasured signal results from the activity of billions of neurons

    Amplitude:0.001-0.01 mVBandwidth:0.5-40 Hz

    Errors:Thermal RF noise50/60 Hz power linesBlink artifacts and similar

    Typical applications:Sleep studiesSeizure detectionCortical mapping

  • EEG measurement setup10-20 Lead system is most widely clinically acceptedCertain physiological featuresare used as reference pointsAllow localization of diagnostic features in the vicinity of the electrodeOften a readily available wire or rubber mesh is usedBrain research utilizes even 256 or 512 channel EEG hats

  • Electromyography (EMG)Measures the electric activity of active muscle fibersElectrodes are always connected very close to the muscle group being measuredRectified and integrated EMG signal gives rough indication of the muscle activityNeedle electrodes can be used to measure individual muscle fibers

    Amplitude:1-10 mVBandwidth:20-2000 Hz

    Main sources of errors are 50/60 Hz and RF interference

    Applications: muscle function, neuromuscular disease, prosthesis

  • Electrooculography (EOG)Electric potentials are created as a result of the movement of the eyeballsPotential varies in proportion to the amplitude of the movementIn many ways a challenging measurement with some clinical value

    Amplitude: 0.01-0.1 mVBandwidth:DC-10 Hz

    Primary sources of error include skin potential and motion

    Applications: eye position, sleep state, vestibulo-ocular reflex

  • Vectorcardiogram (VCG or EVCG)Instead of displaying the scalar amplitude (ECG curve) the electric activation front is measured and displayed as a vector (dipole model, remember?)

    It has amplitude and directionDiagnosis is based on the curve that the point of this vector draws in 2 or 3 dimensionsThe information content of the VCG signal is roughly the same as 12-lead ECG system. The advantage comes from the way how this information is displayedA normal, scalar ECG curve can be formed from this vectro representation, although (for practical reasons) transformation can be quite complicated Plenty of different types of VCG systems are in use

    No legacy problem as such

  • Short break,Kahvia ja pullaa!

  • The biopotential amplifierSmall amplitudes, low frequencies, environmental and biological sources of interference etc.Essential requirements for measurement equipment:High amplificationHigh differential gain, low common mode gain high CMRRHigh input impedanceLow NoiseStability against temperature and voltage fluctuationsElectrical safety, isolation and defibrillation protection

  • The Instrumentation AmplifierPotentially combines the best features desirable for biopotential measurementsHigh differential gain, low common mode gain, high CMRR, high input resistanceA key design component to almost all biopotential measurements!Simple and cheap, although high-quality OpAmps with high CMRR should be used

  • Application-specific requirementsECG amplifierLower corner frequency 0.05 Hz, upper 100HzSafety and protection: leakage current below safety standard limit of 10 uAElectrical isolation from the power line and the earth groundProtection against high defibrillation voltagesEEG amplifierGain must deal with microvolt or lower levels of signalsComponents must have low thermal and electronic noise @ the front endOtherwise similar to ECGEMG amplifierSlightly enhanced amplifier BW sufficesPost-processing circuits are almost always needed (e.g. rectifier + integrator)EOG amplifierHigh gain with very good low frequency (or even DC) responseDC-drifting electrodes should be selected with great careOften active DC or drift cancellation or correction circuit may be necessary

  • Electrical Interference ReductionPower line interference (50 or 60 Hz) is always around usConnects capacitively and causes common mode interferenceThe common mode interference would be completely rejected by the instrumentation amplifier if the matching would be idealOften a clever driven right leg circuit is used to further enhance CMRR

    Average of the VCM is inverted and driven back to the body via reference electrode

  • FilteringFiltering should be included in the front end of the InstrAmpTransmitters, motors etc. cause also RF interference

    Small inductorsor ferrite beadsin the lead wiresblock HF frequencyEM interferenceRF filtering withsmall capacitors

  • 50 or 60 Hz notch filterSometimes it may be desirable to remove the power line interferenceOverlaps with the measurement bandwidth

    May distort the measurement result and have an affect on the diagnosis!Option often available with EEG & EOG measuring instruments

    Twin Tnotch filter

  • Artifact reductionElectrode-skin interface is a major source of artifactChanges in the junction potential causes slow changes in the baselineMovement artifacts cause more sudden changes and artifactsDrifting in the baseline can be detected by discharging the high-pass capacitor in the amplifier to restore the baseline

  • Electrical isolationElectrical isolation limits the possibility of passage of any leakage current from the instrument in use to the patientSuch passage would be harmful if not fatal!

    TransformerTransformers are inherently high frequency AC devicesModulation and demodulation needed

    Optical isolationOptical signal is modulated in proportion to the electric signal and transmitted to the detectorTypically pulse code modulated to circumvent the inherent nonlinearity of the LED-phototransistor combination

  • Defibrillation ProtectionMeasuring instruments can encounter very high voltagesE.g. 15005000V shocks from defibrillatorFront-end must be designed to withstand these high voltages

  • Electrodes BasicsHigh-quality biopotential measurements requireGood amplifier designUse of good electrodes and their proper placement on the patientGood laboratory and clinical practicesElectrodes should be chosen according to the applicationBasic electrode structure includes:The body and casingElectrode made of high-conductivity materialWire connectorCavity or similar for electrolytic gelAdhesive rimThe complexity of electrode design often neglected

  • Electrodes - BasicsSkin preparation by abrasion or cleansingPlacement close to the source being measuredPlacement above bony structures where there is less muscle massDistinguishing features of different electrodes:How secure? The structure and the use of strong but less irritant adhesivesHow conductive? Use of noble metals vs. cheaper materialsHow prone to artifact? Use of low-junction-potential materials such as Ag-AgClIf electrolytic gel is used, how is it applied? High conductivity gels can help reduce the junction potentials and resistance but tend to be more allergenic or irritating

    Baseline drift due to thechanges in junctionpotential or motion artifactsChoice of electrodes Muscle signalinterference Placement Electromagneticinterference Shielding

  • Ag-AgCl, Silver-Silver Chloride ElectrodesThe most commonly used electrode typeSilver is interfaced with its salt silver-chlorideChoice of materials helps to reduce junction potentialsJunction potentials are the result of the dissimilar electrolytic interfacesElectrolytic gel enhances conductivity and also reduces junction potentialsTypically based on sodium or potassium chloride, concentration in the order of 0.1 M weak enough to not irritate the skinThe gel is typically soaked into a foam pad or applied directly in a pocket produced by electrode housingRelatively low-cost and general purpose electrodeParticularly suited for ambulatory or long term use

  • Gold ElectrodesVery high conductivity suitable for low-noise meas.Inertness suitable for reusable electrodesBody forms cavity which is filled with electrolytic gelCompared to Ag-AgCL: greater expense, higherjunction potentials and motion artifactsOften used in EEG, sometimes in EMG

    Conductive polymer electrodesMade out of material that is simultaneously conductive and adhesivePolymer is made conductive by adding monovalent metallic ionsAluminum foil allows contact to external instrumentationNo need for gel or other adhesive substanceHigh resistivity makes unsuitable for low-noise meas.Not as good connection as with traditional electrodes

  • Metal or carbon electrodesOther metals are seldom used as high-quality noblemetal electrodes or low-cost carbon or polymericelectrodes are so readily availableHistorical value. Bulky and awkward to useCarbon electrodes have high resistivity and are noisier but they are also flexibleand reusableApplications in electrical stimulation and impedance plethysmography

    Needle electrodesObviously invasive electrodesUsed when measurements have to be taken from the organ itselfSmall signals such as motor unit potentials can be measuredNeedle is often a steel wire with hooked tip

  • Thats it,Now for Q&ASQUID = Superconducting Quantum Interference Device